    HRRS-10kV SERIES

    High Resistance, 10kV Decade Substituter

    User and Service Manual

    Copyright 2011 IET Labs, Inc.Visit www.ietlabs.com for manual revision updates

    HRRS-10kV im/March 2011

  • TEL: (516) 334-5959 (800) 899-8438 FAX: (516) 334-5988534 Main Street, Westbury, NY 11590

    IET LABS, INC.www.ietlabs.comStandards Decades Strobes Sound Level Meters Bridges

  • TEL: (516) 334-5959 (800) 899-8438 FAX: (516) 334-5988534 Main Street, Westbury, NY 11590

    IET LABS, INC.www.ietlabs.comStandards Decades Strobes Sound Level Meters Bridges

    i

    HRRS-10kV Series

    WARRANTY

    We warrant that this product is free from defects in material and workmanship and, when properly used, will perform in accordance with applicable IET specifi cations. If within one year after original shipment, it is found not to meet this standard, it will be repaired or, at the option of IET, replaced at no charge when returned to IET. Changes in this product not approved by IET or application of voltages or currents greater than those allowed by the specifi cations shall void this warranty. IET shall not be liable for any indirect, special, or consequential damages, even if notice has been given to the possibility of such damages.

    THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR ANY PARTICULAR PURPOSE.

  • ii

    HRRS-10kV Series

    WARNING

    OBSERVE ALL SAFETY RULESWHEN WORKING WITH HIGH VOLTAGES OR LINE VOLTAGES.

    Dangerous voltages may be present inside this instrument. Do not open the caseRefer servicing to qualifi ed personnel

    HIGH VOLTAGES MAY BE PRESENT AT THE TERMINALS OF THIS INSTRUMENT

    WHENEVER HAZARDOUS VOLTAGES (> 45 V) ARE USED, TAKE ALL MEASURES TOAVOID ACCIDENTAL CONTACT WITH ANY LIVE COMPONENTS.

    USE MAXIMUM INSULATION AND MINIMIZE THE USE OF BARECONDUCTORS WHEN USING THIS INSTRUMENT.

    Use extreme caution when working with bare conductors or bus bars.

    WHEN WORKING WITH HIGH VOLTAGES, POST WARNING SIGNS AND KEEP UNREQUIRED PERSONNEL SAFELY AWAY.

    CAUTION

    DO NOT APPLY ANY VOLTAGES OR CURRENTS TO THE TERMINALS OF THISINSTRUMENT IN EXCESS OF THE MAXIMUM LIMITS INDICATED ON

    THE FRONT PANEL OR THE OPERATING GUIDE LABEL.

  • iii

    HRRS-10kV Series

    Table of Contents

    Table of ContentsChapter 1: Introduction

    1.1 Overview ............................................................................................................... 1

    Chapter 2: SpecificationsSpecifications ................................................................................................................. 2

    Ordering Information ..................................................................................................... 3

    Chapter 3: Operation3.1 Initial Inspection and Setup .................................................................................. 4

    3.2 Dial Setting ........................................................................................................... 4

    3.3 Environmental Conditions .................................................................................... 4

    3.4 Storage .................................................................................................................. 4

    Chapter 4: Maintenance4.1 Maintainability and Reliability ............................................................................. 5

    4.2 Preventive Maintenance ........................................................................................ 5

    4.3 Verification of Performance .................................................................................. 5

    4.3.1 Calibration Interval ..................................................................................... 5

    4.3.2 General Considerations ............................................................................... 5

    4.3.3 Calibration Procedure ................................................................................. 6

    4.4 Schematic .............................................................................................................. 6

    4.5 Replaceable Parts .................................................................................................. 7

    Figures and TablesFigure 1-1: HRRS-10kV Series High-Resistance, Decade Substituter ...................1

    Figure 2-1: HRRS-10kV sample label ......................................................................3

    Figure 4-1: HRRS-10kV Series Schematic Diagram ..............................................6

    Table 4-1: Replaceable Parts ..................................................................................7

    Figure 4-2: Replaceable Parts ................................................................................7

  • iv

    HRRS-10kV Series

    Table of Contents

    This page is intentionally left blank

  • 1

    HRRS-10kV Series

    1Introduction

    Chapter 1

    INTRODUCTION

    1.1 Overview

    The High Resistance, 10 kV Decade Substituter (HRRS-10kV Series, Figure 1.1) is a family of instruments that answers the need of the calibration and test community for decades and working stan-dards that go up to 10 T, offering a broad choice of high-range, excellent performance resistance sources. High values of resistance are available for use at high voltages, without sacrificing accuracy, stability, and temperature and power coefficients of resistance.

    The HRRS-10kV Series employs state-of-the-art precision resistors of various types. It is designed specifically for operation at high voltage, with very low leakage to minimize degradation of accuracy.

    The standard models offer a choice of one to nine decades. The panels are clearly labeled showing the step size and maximum allowable applied input for each decade. Custom requirements outside of the standard models can be satisfied.

    The binding posts are insulated with Kel-F high-resistance, non-moisture-absorbing material to ensure low leakage.

    With a resolution as low as 10 and a maximum available resistance of 10 T, the HRRS-10kV Series may be used for exacting precision high-value resis-tance applications that require voltages of up to 10 kV.

    Applications include calibration of meters and meg-ohmmeters, and checking of electrochemical and bio-medical sensors and instruments. These instruments are useful wherever high resistances are required, such as in testing high-impedance amplifiers and the insulation of low-power circuits.

    Figure 1-1: HRRS-10kV Series High-Resistance, High-Voltage Decade Substituter

  • 2

    HRRS-10kV Series

    2 Specifications

    For convenience to the user, the pertinent specifications are given in an OPERATING GUIDE affixed to the case of the instrument. Figure 2.1 shows a typical example.

    SPECIFICATIONS

    Zero Resistance

  • 3

    HRRS-10kV Series

    3Specifications

    ORDERING INFORMATION

    Model Total ResistanceNumber of Decades Resolution

    HRRS-*-3-100k-10kV 111 M 3 100 kHRRS-*-3-1M-10kV 1.11 G 3 1 MHRRS-*-3-10M-10kV 11.1 G 3 10 MHRRS-*-3-100M-10kV 111 G 3 100 MHRRS-*-3-1G-10kV 1.11 T 3 1 GHRRS-*-3-10G-10kV 11.1 T 3 10 GHRRS-*-4-10k-10kV 111.1 M 4 10 kHRRS-*-4-100k-10kV 1.111 G 4 100 kHRRS-*-4-1M-10kV 11.11 G 4 1 MHRRS-*-4-10M-10kV 111.1 G 4 10 MHRRS-*-4-100M-10kV 1.111 T 4 100 MHRRS-*-4-1G-10kV 11.11 T 4 1 GHRRS-*-5-1k-10kV 111.11 M 5 1 kHRRS-*-5-10k-10kV 1.111 1 G 5 10 kHRRS-*-5-100k-10kV 11.111 G 5 100 kHRRS-*-5-1M-10kV 111.11 G 5 1 MHRRS-*-5-10M-10kV 1.111 1 T 5 10 MHRRS-*-5-100M-10kV 11.111 T 5 100 MHRRS-*-6-10-10kV 11.111 1 M 6 10 HRRS-*-6-100-10kV 111.111 M 6 100 HRRS-*-6-1k-10kV 1.111 11 G 6 1 kHRRS-*-6-10k-10kV 11.111 1 G 6 10 kHRRS-*-6-100k-10kV 111.111 G 6 100 kHRRS-*-6-1M-10kV 1.111 11 T 6 1 MHRRS-*-6-10M-10kV 11.111 1 T 6 10 M

    Model Total ResistanceNumber of Decades Resolution

    HRRS-*-7-10-10kV 111.111 1 M 7 10

    HRRS-*-7-100-10kV 1.111 111 G 7 100

    HRRS-*-7-1k-10kV 11.111 11 G 7 1 k

    HRRS-*-7-10k-10kV 111.111 1 G 7 10 k

    HRRS-*-7-100k-10kV 1.111 111 T 7 100 k

    HRRS-*-7-1M-10kV 11.111 11 T 7 1 M

    HRRS-*-8-10-10kV 1,111.111 1 M 8 10

    HRRS-*-8-100-10kV 11.111 111 G 8 100

    HRRS-*-8-1k-10kV 111.111 11 G 8 1 k

    HRRS-*-8-10k-10kV 1,111.111 1 G 8 10 k

    HRRS-*-8-100k-10kV 11.111 111 T 8 100 k

    HRRS-*-9-10-10kV 11,111.111 1 M 9 10

    HRRS-*-9-100-10kV 111.111 111 G 9 100

    HRRS-*-9-1k-10kV 1,111.111 11 G 9 1 k

    HRRS-*-9-10k-10kV 11,111.111 1 G 9 10 k*To specify accuracy, replace * with Q, B, or F as required.

    Options-RM: Rack mountable case for standard 19 rack-K: Kelvin-type 4-terminal posts-RO: Rear output

    Figure 2-1: HRRS-10kV sample label

    IET LABS, INC. 534 Main Street, Westbury, NY 11590 (516) 334-5959 (800) 899-8438 Fax (516) 334-5988HRRS-10kV lbl/03-17-11

    Observe all safety rules when working with high voltages or line voltages. Connect the (G) terminal to earth ground in order to maintain the case at a safe voltage. Whenever hazardous voltages (> 45 V) are used, take all measures to avoid accidental contact with any live components: a) Use maximum insulation and minimize the use of bare con-

    ductors. b) Remove power when adjusting switches. c) Post warning signs and keep personnel safely away.

    HRRS-10kV SERIES HIGH-RESISTANCE, 10kV DECADE SUBSTITUTER OPERATING GUIDE