Hospital: dedicated modules To secure power distribution and monitoring solution for operating theatres
IECOur solution complies with international standard IEC 60364-7-710 and national standards and regulations
Operating rooms require extremely high availability and quality of electric power to offer patients maximum safety. For this reason, the standards lay down very strict rules to ensure the continuity of service of electrical installations.
What do the standards say?> In group 2 rooms for medical use, the
medical IT system should be used for the circuits powering medical electrical equipment and systems for survival and surgical applications, and the other equipment located in the environment of the patient.
> An audible and visual alarm must be provided in the room in question to alert medical personnel.
> Operating activities must have continuity of electric power supply.
> For the satisfactory operation of medical equipment, prevention of electromagnetic disturbances may be necessary.
> Monitoring of overload and high temperature for the medical IT transformer is required.
> An alarm should take place if the earth connection or the connection to the system is disconnected.
IM10-H and HRP (Hospital Remote Panel) for the "Classic" solution
> Graphic display
> Bargraph
> Smart HMI
> 8 languages
> Earth and injection connection monitoring
IM20-H as a brick of the advanced solution
> Modbus Communication
> Datalog with timestamping of all events
> Transformer management: - secondary load current display - alarm on threshold (in % of nominal current) - alarm on temperature through the sensor
(bimetal)
Simple and efficent> Audible and visual alarm for an insulation or
electrical fault (transformer overload or circuit breaker tripping)
> Testing of the insulation monitoring system
> Audible alarm stoppage
> 24 V DC power supply
> Antibacterial
> Tested with Anios products (disinfection products)
> Complies with IEC 60601-1 (medical equipment)
Vigilohm_Catalogue_Part_2.indd version: 1.0 17
Vigilohm IM10-H and IM20-HInsulation monitoring devices for hospitals
PB
1063
75_5
5
Vigilohm IM10-H and IM20-HType of installation to be monitored
LV AC / DC IT systems Phase-to-phase voltage y 230 V AC +15 %Frequency 50/60 Hz
Electrical characteristicsRange for insulation resistance readings 1 kΩ to 10 MΩFault signalling Number of thresholds 1 (password protected)
Thresholds 50 kΩ to 500 kΩResponse time y 1 sDevice operating test YesFailsafe feature (1) As standard Internal impedance At 50 Hz 110 kΩAccuracy 5 %Output contact Number 1 Standard or failsafe
Type ChangeoverBreaking AC 250 V 6 ACapacity DC 12 to 24 V 6 A
Input contact Voltage supplied 24 V
UseThese modules are dedicated to hospital IT networks.
OperationInjection of a low-frequency AC voltage between the
network and the earth.
MeasurementInsulation measurement through the earth leakage
current in the IMD.
IndicationSatisfactory insulation resistance (green light).Drop in insulation resistance below the fault
threshold (orange light).Earth or injection connection lost.
Display (8 languages (3))Insulation resistance.Thresholds. Alarms with dedicated pop-up windows.
(3) French, English, Spanish, Italian, Portuguese, German, Russian, Chinese.
Additional functions with IM20-HModbus communication.Datalog with timestamping of all events.Transformer management:secondary load current display (in %)alarm on a threshold (in % of nominal current)alarm on temperature through the sensor (bimetal).
AccessoriesHospital Remote Panel HRP (ref 50168) see pages 36 and 37.Or as part of the solution for operating theatres.
b
b
bb
b
bbb
bbbvvv
Transformer bimetal Minimum load 5 mAAuxiliary supply voltage 50/60 Hz 110/230 V AC ±15 %
DC 125/250 V DC ±15 %Cable size 0.2 to 2.5 mm2
Maximum device consumption 12 VAMeasurement voltage 25 V maxMeasurement current 0.9 mADielectric strength 4000 V AC / 5500 V DCMechanical characteristics
Weight 0.25 kgThermoplastic case Mounting Panel mount or DIN
railDegree of protection Front IP52Installation Cat. III, pollution 2, moulded case,
disconnectable assembly, symetrical or embedded
Other characteristicsTemperature range For operation -25 ºC to +55 ºC
For storage -40 ºC to +70 ºCClimatic conditions (2)
Standards Product IEC 61557-8Safety IEC 61010-1Installation IEC 60364-7-710
(1) Failsafe: the relay is deactivated on occurrence of a fault or if the auxiliary supply voltage accidentally fails.(2) Suitable for use in all climates:b damp heat, equipment not operating (IEC 60068-2-30)b damp heat, equipment operating (IEC 60068-2-56)b salt mist (IEC 60068-2-52).
Commercial references:IMD-IM10-H IMD-IM20-H
bb
Vigilohm_Catalogue_Part_5.inddversion: �.036
PB
�059
43_5
8
PresentationThis accessory provides the interface for an electrical installation monitoring system in hospitals.Installed in the operating theatre, the Vigilohm HRP informs hospital personnel, in real time, of correct operation or the presence of a fault:
faulty insulation of operating theatre equipmentelectrical fault following a circuit breaker trip or transformer overload.
UseInsulation fault signal light (orange).Electrical fault signal light (red).Correct operation signal light (green).Push button to test installation insulation.Push button to stop alarm buzzer.
bb
bbbbb
Vigilohm HRPMechanical characteristics
Weight 0.5 kgCase Plastic Vertical mountingDegree of protection IP54
IK08Dimensions Height �70 mm
Width �70 mmDepth 20 mm
Buzzer Factory setting 80 db Electrical characteristicsAuxiliary supply voltage 24 V DC 65 mA EnvironmentOperating temperature 0 °C to 40 °CStorage temperature -25 °C to +70 °CMaximum relative humidity 90 %Altitude 2000 m StandardsComplies with IEC 60364-7-7�0 Medical locations
IEC 6�557-8 Electrical safetyIEC 6060�-� Medical electrical
equipment
Tested with Anios products (disinfection, antiseptis and detergency products).
Commercial reference: 50168.
PB
�059
45 Functions identificationVisual alarm on insulation fault
Right operation lamp
Visual alarm on electrical fault (overload or overheating transformer or circuit-breaker fault tripping)
�Test button of the insulation monitoring system (daily test in compliance with IEC 60364-7-7�0 standard)
Audible alarm stoppage, in case of insulation fault or electrical fault. The alarm volume setting is made at the mounting phase (available in the rear)
Vigilohm HRPHospital Remote Panel
Vigilohm_Catalogue_Part_5.indd version: �.0 37
HRP and IM10-H or IM20-HD
B�2
7897
1 wire electrical fault(CB fault tripping or transformer overload)
2 wires (NC Contact)
2 wires (+24 V 0 V)
2 wires (N, Earth)
CPICabinet 1
IM10-H
ABL8MEM24003
HRP
VigilohmHRP
VigilohmHRP
ElectricalfaultInsulationfault
ElectricalfaultInsulationfault
CPICabinet 2
Example architecture 1 using IM10-H and HRPThis monitoring system makes it possible, with the Vigilohm HRP in the operating room, to gather and display information concerning the alarms generated by the electrical installation.
Existing or specific power supply Ex: ABL8MEM24003
DB
�278
98
2 wires(NC contact)
PLC4 wires(Modbus)
ModbusTCP-IP
IM20-HMagelis HRP
MagelisHRP
MagelisHRP
ElectricalfaultInsulationfault Sensors Sensors
ElectricalfaultInsulationfault
Example architecture 2 using IM20-H and SupervisionThe Modbus communication provides the operating room, and/or the maintenance personnel:
insulation valuetransformer load levelalarmstimestamping of events.
This enables event traceability.
bbbb
Note: It is also possible to have a variant of the architecture 1, with an IM20-H linked through Modbus to a supervisor outside the operating room.
Vigilohm_Catalogue_Part_6.inddversion: �.038
Vigilohm Installation and connectionProduct dimensions
IM9, IM9-OLMounting on symmetrical DIN rail Cutout
DB
�279
05
=
=
= =
2 Ø3.85
71
85
67.544
9045
72.5±0.4
61±0.3
66±0.3
46.5±0.3
IM10, IM20, IM10-H, IM20-HMounting on symmetrical DIN rail Cutout
DB
�279
06
59
46.3
93.7110
96
9645 92.4±0.4
92.4±0.4
XM200,XM300C, XL308, XL316Flush mounting Cutout
DB
�278
99
153
129
9.5192
12
DB
�279
38
XD301, XD312Flush mounting Cutout Surface mounting
DB
�279
36
73.5
45
65
49.537.5
46
57573
937110
6.5
2 Ø4.2
6646
10
2 Ø4.2
66
XD312XD301
XD301 XD312
DB
�279
37
2 Ø4.2
66
7.5
10693
6.57257
4.5
8166
1
81
XD3122 Ø4.2
XD301
Vigilohm_Catalogue_Part_6.indd version: �.0 43
IM10-H, IM20-H
PB
�063
75_3
2
DB
�279
57
1
6
2 3 4 5
7 8 9 10 111213141516
D0 = A’ / Rx-, A / Tx-D1 = B’ / Rx+, B / Tx+
0 V
6 A 6 A
Failsafe wiring(recommended)
Standard wiring
6 7 8
u 12 Vc 100 mAy 250 Va/ 24 Vc6 A
u 12 Vc 100 mAy 250 Va 24 Vc6 A
6 A
6 A
6
IM20-HIM20-H
7 8
U 110....230 Va
L
N
ModbusRS 485
Shortingblock
IT transformer bimetalContact min. y 24 V c 5 mA
XM200
044�
47_S
E_3
2
DB
�279
23
1098765
21
prevention(falsafefeature)
fault
XM200
aa
43
r
11
DB
�279
24
+
4
11
N123MT/BT
cardew C XM200
4
11
123MT/BT
cardew C XM200
4
11XM200
-
Tunnel terminals for 1.5 mm² wiring
Legendaa: AC auxiliary supply a 15 %, +10 %, 50-60 Hzut: applicationr: installationterminal 13: earth connection by round lug (4 mm diameter) to be crimped.
Vigilohm_Catalogue_Part_6.inddversion: �.048
XLI300, XTU300
DB
�279
49
RS485
18
17
16
15 S
V+
B
V-
2
15
1 6
9
1 : GND4 : B'(RD-)5 : B(TD-)8 : A'(RD+)9 : A(TD+)
RS482
5
1 6
9
13
aa
notused
Never connect more than one interface to the Vigilohm System bus.To connect devices to the communication bus (shielded, double twisted pair) it is
recommended to interconnect their respective terminals S, B, V+ and V-.
bb
XLI300 Sub D9RS485� signal ground4 received data -5 transmitted data -8 received data +9 transmitted data +
Example of line connection
DB
�279
35
XLI300 XTU300 XAS
XML308 XML316
XD308cXL308 XL316XM300c
V- B V+ S
V-B V+ S
V- B V+ S
V- B V+ S
V- B V+ S
HRP
PB
�059
43_4
2
DB
�279
09
6
2.5 mm² max.
Electrical fault1
2
3
4
5
6
7
Common
Insulation fault
Auxiliary power supply
N
34 V c 14.5 mA
34 V c 14.5 mA
+24 V c 65 mA
0V c
Installation and connectionProduct connection
Vigilohm
Vigilohm_Catalogue_Part_7.indd version: �.0 49
Vigilohm Vigilohm insulation monitoringCommercial references
Vigilohm commercial referencesProduct Voltage range Reference
PB
�063
70_�
5 IM9 Insulation monitoring deviceIM9 ��0/4�5 V AC 50/60 Hz IMD-IM9
IM9-OL Off-line insulation monitoring deviceIM9-OL ��0/4�5 V AC 50/60 Hz IMD-IM9-OL
PB
�063
74_�
5 IM10 & IM20 Insulation monitoring devicesIM10 ��0/4�5 V AC 50/60 Hz IMD-IM10IM20 ��0/4�5 V AC 50/60 Hz IMD-IM20
IM10-H & IM20-H Insulation monitoring devices for hospitalsIM10-H ��0/240 V AC 50/60 Hz IMD-IM10-HIM20-H ��0/240 V AC 50/60 Hz IMD-IM20-H
044�
47_S
E_2
0 XM200 Insulation monitoring deviceXM200 ��5/�27 V AC 50/60 Hz 50727XM200 220/240 V AC 50/60 Hz 50728XM200 380/4�5 V AC 50/60 Hz 50729
044�
48_S
E_2
0 XM300C Insulation monitoring deviceXM300C ��5/�27 V AC 50/60 Hz 50540XM300C 220/240 V AC 50/60 Hz 50541XM300C 380/4�5 V AC 50/60 Hz 50542
044�
49_S
E_3
0 XML308 & XML316 Monitoring-locating devicesXML308 ��5/�27 V AC 50/60 Hz 50490XML308 220/240 V AC 50/60 Hz 50491XML308 380/4�5 V AC 50/60 Hz 50492XML316 ��5/�27 V AC 50/60 Hz 50322XML316 220/240 V AC 50/60 Hz 50323XML316 380/4�5 V AC 50/60 Hz 50324
044�
44_S
E_2
0 XD301 & XD312 Automatic insulation fault detectorsXD301 ��5/�27 V AC 50/60 Hz 50506XD301 220/240 V AC 50/60 Hz 50507XD301 380/4�5 V AC 50/60 Hz 50508XD312 ��5/�27 V AC 50/60 Hz 50535XD312 220/240 V AC 50/60 Hz 50536XD312 380/4�5 V AC 50/60 Hz 50537
044�
43_S
E_�
2 XD308C Communicating automatic insulation fault detectorXD308C ��5/�27 V AC 50/60 Hz 50723XD308C 220/240 V AC 50/60 Hz 50724XD308C 380/4�5 V AC 50/60 Hz 50725
044�
45_S
E_2
2 XL308 & XL316 Insulation fault locatorsXL308 ��5/�27 V AC 50/60 Hz 50606XL308 220/240 V AC 50/60 Hz 50607XL308 380/4�5 V AC 50/60 Hz 50608XL316 ��5/�27 V AC 50/60 Hz 50615XL316 220/240 V AC 50/60 Hz 50616XL316 380/4�5 V AC 50/60 Hz 50617
Vigilohm_Catalogue_Part_7.inddversion: �.050
Vigilohm commercial referencesProduct Voltage range Reference
05�3
50_S
E_�
0 XGR & XRM & probes Mobile fault locatingKit case: XGR (50282) + XRM + XP15 + XP50 + XP100
50310
XRM 50278XGR ��5/�27 V AC 50/60 Hz 50281
PB
�004
7�_S
E_2
5 XGR 220/240 V AC 50/60 Hz 50282XGR 380/4�5 V AC 50/60 Hz 50283XP15 probe 50494XP50 probe 50498XP100 probe 50499Empty case 50285
044�
4�_S
E_�
5 XLI300 & XTU300 Communication interfacesXLI300 ��5/�27 V AC 50/60 Hz 50515XLI300 220/240 V AC 50/60 Hz 50516XLI300 380/4�5 V AC 50/60 Hz 50517XTU300 ��5/�27 V AC 50/60 Hz 50545 (1)
XTU300 220/240 V AC 50/60 Hz 50546 (1)
XTU300 380/4�5 V AC 50/60 Hz 50547 (1)
05�3
52_S
E_2
5 Toroids Earth leakage current detectorsTA30 50437PA50 50438IA80 50439
0425
89_S
E_2
0 MA120 50440SA200 50441GA300 50442Open POA 50845Open GOA 50486100 m cable (shielded) 50136
04�8
57_S
E_�
0 Auxiliaries(2) Accessories for various installations250 V Cardew C 50170440 V Cardew C 50171660 V Cardew C 501721000 V Cardew C 50183
Cardew C base 50169
044�
39_S
E_�
8 HV-IM20-1.7 subassembly IMD-HV-IM20-1.7ZX limiting impedance 50159PHT1000 additional subassembly 50248HRP 50168
PB
�059
43_2
2
(1) Please consult Schneider Electric for configuration. (2) See choice of Cardew on page 33 and auxiliary selection table on page 35.