Hand instrument sterilization

Page 2: Infection rbna ystr !

Effective and efficient infection control in the dental office is essential for the safety of patients and to ensure that productivity does not suffer. Infection control programs all include the cleaning and sterilization of reusable dental instruments and devices. Care must be taken by the dental healthcare professional to ensure that all instruments are cleaned prior to sterilization, and that this is carried out in a safe manner to avoid injury

Page 3: Infection rbna ystr !

Instrument Processing : Step by step!

1 - Clean the dental instruments manually or through mechanical means like a thermal washer disinfector or an ultrasonic bath. Infection Control Services says that dental instruments cannot be sterilized properly unless they are thoroughly cleaned beforehand.

Cleaning can be accomplished by two automated approved methods:

Page 4: Infection rbna ystr !

Instrument washers- (which resemble a dishwasher but are made to clean instruments; regular dishwashers are not appropriate nor are they recommended)

Page 5: Infection rbna ystr !

Ultrasonic cleaners

Always use a recommended ultrasonic solution. Also, establish a schedule for testing the effectiveness of the ultrasonic. To do this, simply dip several inches of a sheet of aluminum foil in the ultrasonic and turn on for 20 seconds. If the unit is functioning properly, the foil will be full of indentations and holes. If not, a repair is needed.

Page 6: Infection rbna ystr !

They may also be cleaned manually, but this presents more risk of an occupational exposure.

If you must clean instruments manually,

wear all PPE, use a scrub brush with a long handle,

and scrub just a few at a time while holding

deep in a sink. Rinse and inspect each instrument

before packaging. They should not have any

debris remaining.

Page 7: Infection rbna ystr !
Page 8: Infection rbna ystr !

2 - Afterward cleaning through either

manual or mechanical means, the

dental instruments should be dried

with a disposable cloth.

Page 9: Infection rbna ystr !

3 - Package the dried items into medical grade sterilization wraps or sterilization cassettes to prepare them for sterilization. They must be properly packaged before they are loaded into the heat sterilization unit.

Page 10: Infection rbna ystr !
Page 11: Infection rbna ystr !

4 - Sterilize the instruments using a heat, steam, or chemical process. Heat is the usual method used in a dental setting. Use sterilization equipment that is approved by the

Food and Drug Administration for

dental sterilization work and operate it

according to the manufacturer's

specific instructions.

Page 12: Infection rbna ystr !

5 - Storage the sterilized instruments in their intact packaging. According to the Centers for Disease Control, the packaging must remain sealed and undamaged in order to retain sterile nature of the instrument. If the packaging gets wet, tears or is damaged in some way, the instrument should be sterilized again through the same process.

Page 13: Infection rbna ystr !

Hand piece asepsis.

sterilization between patients with acceptable methods that ensure internal as well as external sterility is recommended.

For proper sterilization of hand piece, the manufacturer’s instructions must be followed.

First, the hand piece should be flushed with water by running it for 20 to 30 seconds, discharging the water into a container.

Page 14: Infection rbna ystr !

Hand piece asepsis.

If recommended by the manufacturer, use ultrasonic cleaner to remove any adherent material, otherwise, it should be scrubbed thoroughly with a detergent and hot water.

Package for sterilization in steam or unsaturated chemical vapour following the manufacturer’s directions. If disinfecting a hand piece that cannot be heat sterilized, spray or saturate with disinfectant recommended by the manufacturer.

Page 15: Infection rbna ystr !

Method of sterilization

 Steam pressure sterilization (autoclave).

Chemical vapor pressure sterilization (Chemiclave).

Dry heat sterilization (Dryclave).

Page 16: Infection rbna ystr !

Steam pressure sterilization (autoclave)

 Time required 15 min time at 2500F (1210C) and 15 lbs of pressure (light load of instruments) Wrapped instruments – 7 min, 2730F (1340C) at 30 pounds of pressure. Preformed in a steam auto clave.

Page 17: Infection rbna ystr !

Steam pressure sterilization (autoclave)

Advantages - Most rapid and effective method for sterilizing cloth surgical packs and towel packs

Disadvantages – Items sensitive to elevated temperature cannot be autoclaved. - Tends to rust carbon steel instruments and burs. - Burs can be protected by submerging in anticorrosive agent- 2% sodium nitrite

Page 18: Infection rbna ystr !

Chemical vapor sterilization (chemiclave)

Performed in a chemiclave - Operate at 2700F (1310C) at 20 lbs for half an hour.

Advantages: - Carbon steel

and burs are said to be sterilized

without rust.

Page 19: Infection rbna ystr !

Chemical vapor sterilization (chemiclave)

Disadvantages:

Items sensitive toelevated temperature

will be damaged.

Towels and heavy cloth wrapping

may not be penetrated to provide

adequate sterilization.

Only dry instruments should be

loaded

Page 20: Infection rbna ystr !

Dry heat sterilization ( Conventional dry heat oven )

Heated at 320oF (1600C) for 30 min. Instruments should be packaged in foil wrap or nylon bags.

Wrapped instruments : 3350- 3450 F for 60-90 mins.

Short cycle high temperature dry heat

Oven :- Sterilization time reduced :

6 min for unwrapped.

12 min for wrapped.

Temperature – 370 – 3750F