Upload
others
View
0
Download
0
Embed Size (px)
Citation preview
Steve Bouffard, PhD Agilent Technologies
Product Specialist
Molecular Spectroscopy
Innovations in FTIR and UV Technology for the Rapid Analysis of Liquid Samples
The Agilent Cary 630 FTIR • An entry level FTIR that has changed the way analysts approach
routine infrared measurements
• The World’s smallest FTIR!
• Designed for a wide range of markets, in particular, chemicals QA/QC, pharma QA/QC and teaching
The Agilent Cary 630 FTIR sampling flexibility Solving problems through innovation…
630 Engine DialPath TumblIR
Diamond ATR Diffuse Reflectance Transmission
The traditional approach to liquids analysis- the demountable cell
• Great skill and care is needed to assemble and fill the cell without leakage.
• Cleaning cell in-between each analysis is tedious and laborious.
• 4 samples/hour if you’re lucky
Analagous to ATR sampling with enhanced sensitivity due to increased pathlength
1. Place Sample on
Lower Window
3. Scan the sample
2. Rotate into place
4. Cleaning is easy!
The Agilent Cary 630 approach to liquids analysis- the DialPath and TumblIR sampling modules
The Agilent Cary 630 approach to liquids analysis The TumblIR and the DialPath sampling modules – single pathlength liquid transmission accessory.
A REVOLUTION IN LIQUIDS ANALYSIS
100 μm fixed pathlength
3 pathlengths provide flexibility (30 to 250 μm)
Liquid transmission sampling geometry 3 µ L Fill Volume
250 µ L Fill Volume
µ L Fill Volume
250 µ L Fill Volume
The combination of class-leading performance and rapid scanning capability means that most liquid samples can be measured in < 30 secs. Volatile liquids require > 120 secs. to show any appreciable change in concentration, once placed in the DialPath cell that is in the closed position.
The Agilent Cary 630 approach to liquids analysis DialPath – 3 pathlength liquid transmission cell
• 30, 50, 100 μm fixed pathlengths
• 50, 100, 200 μm fixed pathlengths
• Custom pathlengths available
Pathlength (μm) Typical Conc. Range
30 Neat - 0.1%
50 50% - 500 ppm
100 20% - 100 ppm
200 10 % - 50 ppm
Cary 630 approach to liquids analysis
The DialPath sampling module- 3 choices
ID and Quant of Liquid Raw Materials (FDA)
The challenges for this particular application were:-
1. To confirm the identities of various incoming viscous liquid raw materials (Qualitative)
2. To determine the levels of impurities in these raw materials (across a large concentration range) according to FDA guidelines (Quantitative)
3. To keep the analysis as simple as possible for non-skilled operators, using the minimum number of accessories & measurements for the results
Method
Cary 630 FTIR
Accessory – The DialPath
Collection Conditions: – Resolution: 4 cm-1 – # scans: 64 (~30 sec) – DialPath accessory pathlength:
• 30 um for qualitative • 75 um for quantitative
Results – raw material identification (qualitative) Spectra collected with 30um DialPath setting
glycerol
propylene glycol
dipropylene glyco
triacetin
Spectra are searched against reference database
Results – determination of impurities (quantitative)
Spectra collected with 75um DialPath setting
ethylene glycol in glycerol diethylene glycol in glycerol
R2=0.9895 R2=0.9745
Limit of detection – 0.04% Limit of detection – 0.02%
This fits well with FDA limits of <0.1% impurities in glycerol
Results – determination of impurities (quantitative) The Agilent MicroLab software displays analysis results for the level of ethylene glycol impurity in glycerol. Red color shows that the level of impurity is outside of specification.
Monitoring gasoline in diesel fuel
Figure 1. The IR spectral overlay of gasoline (Blue) and diesel fuel (Red) using the Agilent 5500t FTIR spectrometer, 100 μm pathlength
The chemical differences between gasoline and diesel fuel can be seen in their infrared spectra. Below is a comparison between the infrared spectra of gasoline and diesel fuel highlighting the ethanol and aromatics in gasoline. Neither ethanol nor light aromatics are observed in the diesel fuel spectrum.
Figure 2. The IR absorbance vs. concentration plot for gasoline in diesel fuel, 100 um pathlength
Gasoline absorbance versus concentration plot using gasoline-spiked diesel fuel
UV/Vis Sampling Innovations 1. Fiber optic probe analyses 2. Ultra-microvolume cuvettes Increase productivity… Lower cost of analysis
The Cary 60 UV/Vis system is optimized by design for fiber optic probe analyses • High energy Xenon flash lamp coupled with low-noise
electronics provides uncompromised data quality
• Room light immunity minimizes stray light effects
• Small focused beam provides superior coupling to fiber
What are Fiber Optics?
• A fiber optic probe is essentially a light guide that takes the UV-Visible light from the instrument into a fiber – to the sample – and then back to the instrument.
• Performance depends on: • Efficiency of coupling the light from the instrument to the fiber • Ability to shield the fiber tip from the stray light effects for ambient (room)
light.
• Most commercial spectrophotometers have poor coupling (due to large beam size) and are not room-light immune, thereby, making the use of fibers impractical.
Why use Fiber Optics? No more need for expensive cuvettes! Improve workflow and increase productivity by taking the instrument to the
sample Measure cold, hot, toxic/radioactive, odd-shaped samples Measure samples in situ
Measure cold samples directly from the
refrigerator!
Improve workflow and minimize
sample preparation
Fiber optic couplers for Cary 60 UV/Vis system
The fiber optic dip probe is a trouble-free alternative to a conventional sipper.
The fiber optic coupler allows you to connect a fiber optic probe to suit your application.
Measuring the nitrate content in water
Challenges: Solution: Benefits:
American Public Health Association (APHA) Nitrate method Rapid measurement – high sample throughput Simple measurement and auto report generation
Cary 60 with fiber optic dip probe
Room light immunity enabling simple sampling with fiber optics WinUV software for method creation Control instrument from sampling accessory
Concentration and Fiber Optics
Using fiber optic dip probe reduced
time of analysis by 50% Standards curve shows excellent
linearity for accurate measurements of unknowns.
Excellent repeatability on Raw Abs
data demonstrates the performance of the F.O. probe.
Measuring nitrate levels in waste water
Application Note Publication Number 5990-7932EN
Sample Conc mg/L
Mean Abs
SD %RSD Raw Abs
A 0.145 0.0510 0.0009 1.78 0.0520 0.0504 0.0506
B 0.709 0.1825 0.0025 1.36 0.1797 0.1838 0.1841
Measuring the nitrate content in water
Investigating the photochemical properties of Methylene Blue
Challenges: Solution: Benefits:
To analyze the changes, in situ, of Methylene Blue (a photosenstive compound used in cosmetics and sunscreen products) upon its exposure to UV radiation. UV/Vis systems with continuous sources degrade the sample
Cary 60 with fiber optics probe
Unique Cary 60 flashlamp does not cause photodegradation Ability to measure the sample out of the instrument Fastest scan rate to obtain multiple spectra over a short time period
Investigating the photochemical properties of methylene blue
Scanning kinetics
Measuring photo degradation of methylene blue exposed to high intensity UV light from external
source
Application Note Publication Number 5990-7864EN
Challenges: Solution: Benefits:
To measure biological sample directly from storage (eg. fridge) – without compromise to data quality - saving time and increasing sample throughput.
Cary 60 with fiber optics microprobe
Save time and money – increase no. of samples measured and use significantly less sample volume (< 40 µL) Improve workflow by minimizing sample handling and transfer to cuvettes No compromise in accuracy or reproducibility of data due to unique optical design
Measuring low volumes of DNA at 4°C
Scans of 150 µl samples of DNA at 4 ºC at 2 concentrations showing the characteristic absorbance peak at 260 nm. Note peak absorbance of 1.0 absorbance units for 50 µg/ml DNA versus peak absorbance of 0.5 absorbance units for 25 µg/ml DNA demonstrating linear adherence to the Beer Lambert Law (A= εcl)
Application Note Publication Number 5990-7863EN
Measuring low volumes of DNA at 4°C
The ultra-microvolume cuvette Analogous to ATR use in the FTIR Fiber optic cables in a cuvette Microliters or less of sample needed
Factor 2 mm Cap (ng/µL)
1 mm Cap (ng/µL)
0.2 mm Cap (ng/µL)
0.1 mm Cap (ng/µL)
Total Detection Range (ng/µL)
Factor 5 10 50 100
dsDNA 50 6-425 *2-850 65-4250 125-8500 6-8500
ssDNA 37 5-315 10-630 50-3145 95-6290 5-6290
RNA 40 5-340 10-680 50-3400 100-6800 5-6800
Oligo 33 4-280 8-560 45-2800 85-5610 4-5610
Sample Volume
6 - 10 µL 3 - 5 µL
0.7 - 4 µL
0.5 - 3 µL
Cary 60 with Ultra Microvolume Cuvette
*tested on a Cary 60 Manufacturer recommendations
Challenges: Solution: Benefits:
Measure 4 µL of sample accurately and reproducibly
Cary 60 with ultra-microvolume cuvette Direct measurement at 280nm Plot Abs at 280 vs. concentration using standards Accurate and reproducible results Minimize dilutions and reduce sample preparation errors Preservation of precious/expensive samples
Protein/Nucleic Acid Conc. – Low Sample Volume
Protein Concentration – Low Sample Volume
Multiple wavelength scans of DNA demonstrate the superior reproducibility of the Cary 60
using only 4 µL of DNA sample!
Summary • The Agilent Cary 630 FTIR, in addition to being the World’s
smallest FTIR has unique liquid transmission sampling modules- the DialPath and TumblIR. Transmission sampling provides the ease of ATR with increased sensitivity
• The Agilent Cary 60 UV/Vis System is optimized for fiber optic analyses and for ultra low volume liquids analyses. Its small beam geometry is the key to providing the best sensitivity for these techniques