 Instruction Sheet IC : 15030338 For Analogue Panel Meters type EQ, PQ, BIQ, & BIEQ, PL,VL,VQ,ZQ 1.0 General: This meter complies with International standard IEC 1010-1 Amendment 2 for safety and is conforming to routine tests as per IEC 51. The test equipment used is traceable to National/ International standards. This product must be used after mounting i n panel with hardwa re supplied, with wires permanently connected to terminals as specified. * Due consideration should be given to space behind the meter to allow for bends in connecting  cables. * Ensure that wires do not remain under tension. * The products do not have internal protection against overload & short circuit. Hence external  safety protection should be provided. It is recommended to install protect ive device near  equipment. * PQ meter millivolt ranges :- Please use lead wires having total resistance of 35 milliohms. * For IP-65 :-Meter Should be installed with IP 65 Gasket behind the bezel on the case. * Indoor use * Altitude up to 2000 m o * Te mp.-10 to +55 C o * Maximum relative humidity 80% for temperature up to 31 C decreasing linearly to 50% relative o o  humidity at 40 C & furthe r linear dec rease to 2 5% relati ve humidit y at 55 C * Basic Insulation * Pollution degree2 * Installation Categori es (over voltage category): For EQ, PQ, Size 72, 96 & 144 - 600V CAT III  Size 48 - 300V CAT III  For BIQ, BIEQ, VL, VQ, ZQ PL, - 300V CAT III 2.0 Meaning of Symbols : 3.0 Safety specifications (IEC1010-1 A2) 4.0 Equipment Installation:  As per information on Sticker & Dial of the instrument. * Equipment must be installed & maintained by suitably qualified person. * If equipment is used in a manner not specified by the manufacturer , the protection provided by  the equipment may be impaired. * CTs must not be opened when on load. * Instruments with damaged bezels or window glasses must be disconnected from mains. * Back cover must be snapped into place for protection against accidental contact. * Disconnect supply before replacing scales, bezels or window glasses. * There are no user serviceable parts in this product. Please contact at the address given, for  clarifications if any. 5.0 Operation & Maintenance : 6.0 Equipment Rating : o >10 0 C  Caution hot surface EU conformity mark Connections  AC/DC Vo ltage & ZQ (Frequ ency meter) (directly connected) 72, 96, 144 Size  AC/DC Curren t (directly connected ) 72, 96, 144 Size 17 18 L1/+ N(L2)/- 17 18 L1/+ N(L2)/-  AC current (for u se on current t ransformer) 72, 96, 144 Size 17 18 L1 N(L2) k K L Dimensions For EQ Voltmeters & Ammeters upto 30A/For PQ Voltmeter & Ammeters <6A ZQ, BiQ, BiEQ, PL, VL, VQ For EQ Ammeters >30A/For PQ Ammeters 6- 100 A d b c f e a d b c f e a Type Dimensions (in mm) Bezel Case Depth 48X48* 48 43.5 53 44.5 5.5 +0.6 45 64 C = 62mm, for I >30A & up to 60A /*C = 68mm, for I= 6 to 60 A C = 67mm, for I >60A /*C = 78mm, for I> 60A f = 69.5mm, for I >60A f = 75 mm, for EQ 48,I > 30 A/75 mm /** f = 75mm, for PQ 48,1 >A. 72X72 72 66 53 67.5 5.5 +0.7 68 96X96 144X144 96 90 53 91.5 5.5 +0.8 92 144 136 *only EQ. PQ & VQ. 53 137.5 5.5 +1 138 a b c* d e f** * * ** ** Cutout size Depth with back cover 64 64 64 Weight approx 0.16kg 0.1 kg. 0.2kg. 0.4kg. I For 48 size, terminal numbering with be 11 & 12 instead of 17 & 18 repectively. *

Instruction Manual Analogue Panel Meters Type VQ

Embed Size (px)

DESCRIPTION

fdf

Citation preview

7/18/2019 Instruction Manual Analogue Panel Meters Type VQ

http://slidepdf.com/reader/full/instruction-manual-analogue-panel-meters-type-vq 1/2

Instruction Sheet IC : 15030338

For Analogue Panel Meters type EQ, PQ, BIQ, & BIEQ, PL,VL,VQ,ZQ

1.0 General:

This meter complies with International standard IEC 1010-1 Amendment 2 for safety and isconforming to routine tests as per IEC 51. The test equipment used is traceable to National/International standards.

This product must be used after mounting i n panel with hardware supplied, with wirespermanently connected to terminals as specified.

* Due consideration should be given to space behind the meter to allow for bends in connecting  cables.

* Ensure that wires do not remain under tension.

* The products do not have internal protection against overload & short circuit. Hence external  safety protection should be provided. It is recommended to install protective device near   equipment.

* PQ meter millivolt ranges :- Please use lead wires having total resistance of 35 milliohms.

* For IP-65 :-Meter Should be installed with IP 65 Gasket behind the bezel on the case.

* Indoor use* Altitude up to 2000 m

o* Temp.-10 to+55 C

o* Maximum relative humidity 80% for temperature up to 31 C decreasing linearly to 50% relative

o o  humidity at 40 C & further linear decrease to 25% relative humidity at 55 C* Basic Insulation* Pollution degree2* Installation Categories (over voltage category): For EQ, PQ, Size 72, 96 & 144 - 600V CAT III  Size 48 - 300V CAT III  For BIQ, BIEQ, VL, VQ, ZQ PL, - 300V CAT III

2.0 Meaning of Symbols :

3.0 Safety specifications (IEC1010-1 A2)

4.0 Equipment Installation:

 As per information on Sticker & Dial of the instrument.

* Equipment must be installed & maintained by suitably qualified person.

* If equipment is used in a manner not specified by the manufacturer, the protection provided by  the equipment may be impaired.

* CTs must not be opened when on load.

* Instruments with damaged bezels or window glasses must be disconnected from mains.

* Back cover must be snapped into place for protection against accidental contact.

* Disconnect supply before replacing scales, bezels or window glasses.

* There are no user serviceable parts in this product. Please contact at the address given, for   clarifications if any.

5.0 Operation & Maintenance :

6.0 Equipment Rating :

o>100 C

 

Caution hot surface

EU conformity mark

Connections

 AC/DC Voltage & ZQ (Frequency meter)(directly connected)72, 96, 144 Size

 AC/DC Current (directly connected)72, 96, 144 Size

1718

L1/+

N(L2)/-

1718

L1/+

N(L2)/-

 AC current (for use on current transformer)72, 96, 144 Size

1718

L1

N(L2)

k

K L

Dimensions

For EQ Voltmeters & Ammeters upto30A/For PQ Voltmeter & Ammeters <6AZQ, BiQ, BiEQ, PL, VL, VQ

For EQ Ammeters >30A/ForPQ Ammeters 6- 100 A

d b

cf  e

a d b

cf  e

a

TypeDimensions(in mm)BezelCaseDepth

48X48*

4843.55344.55.5

+0.645

64C = 62mm, for I >30A & up to 60A /*C = 68mm, for I= 6 to 60 AC = 67mm, for I >60A /*C = 78mm, for I> 60Af = 69.5mm, for I >60Af = 75 mm, for EQ 48,I > 30 A/75 mm /** f = 75mm, for PQ 48,1 >A.

72X72

72665367.55.5

+0.768

96X96 144X144

96905391.55.5

+0.892

144136 *only EQ. PQ & VQ.53137.55.5

+1138

abc*de

f********

Cutout sizeDepth withback cover  64 64 64

Weight approx 0.16kg0.1 kg. 0.2kg. 0.4kg.

I

For 48 size, terminal numbering with be 11 & 12 instead of 17 & 18 repectively.*

7/18/2019 Instruction Manual Analogue Panel Meters Type VQ

http://slidepdf.com/reader/full/instruction-manual-analogue-panel-meters-type-vq 2/2

Instruction Sheet IC : 30338

For Analogue Panel Meters type EQ, PQ, BIQ, & BIEQ, PL,VL,VQ,ZQ

1.0 General:

This meter complies with International standard IEC 1010-1 Amendment 2 for safety and isconforming to routine tests as per IEC 51. The test equipment used is traceable to National/International standards.

This product must be used after mounting in panel with hardware supplied, with wirespermanently connected to terminals as specified.

* Due consideration should be given to space behind the meter to allow for bends in connecting  cables.

* Ensure that wires do not remain under tension.

* The products do not ha ve internal protection against overload & short circuit. Hence external  safety protection should be provided. It is recommended to install protective device near   equipment.

* PQ meter millivolt ranges :- Please use lead wires having total resistance of 35 milliohms.

 As per i nforma tion on Sticker & Dial of the ins trumen t.

* Equipment must be installed & maintained by suitably qualified person.

* If equipment is used in a manner not specified by the manufacturer, the protection provided by  the equipment may be impaired.

* CTs must not be opened when on load.

* Instruments with damaged bezels or window glasses must be disconnected from mains.

* Back cover must be snapped into place for protection against accidental contact.

* Disconnect supply before replacing scales, bezels or window glasses.

* There are no user serviceable parts in this product. Please contact at the address given, for   clarifications if any.

* Indoor use* Altitude up to 2000 mo

* Temp.-10 to+55 Co

* Maximum relative humidity 80% for temperature up to 31 C decreasing linearly to 50% relativeo o

  humidity at 40 C & further linear decrease to 25% relative humidity at 55 C* Basic Insulation* Pollution degree2* Installation Categories (over voltage category): For EQ, PQ, Size 72, 96 & 144 - 600V CAT III  Size 48 - 300V CAT III  For BIQ, BIEQ, VL, VQ, ZQ PL, - 300V CAT III

2.0 Meaning of Symbols :

3.0 Safety specifications (IEC1010-1 A2)

4.0 Equipment Installation:

5.0 Operation & Maintenance :

6.0 Equipment Rating :

o>100 C

 

Caution hot surface

EU conformity mark

Instruction Sheet IC : 30338

For Analogue Panel Meters type EQ, PQ, BIQ, & BIEQ, PL,VL,VQ,ZQ

1.0 General:

This meter complies with International standard IEC 1010-1 Amendment 2 for safety and isconforming to routine tests as per IEC 51. The test equipment used is traceable to National/International standards.

This product must be used after mounting in panel with hardware supplied, with wirespermanently connected to terminals as specified.

* Due consideration should be given to space behind the meter to allow for bends in connecting  cables.

* Ensure that wires do not remain under tension.

* The products do not have internal protection against overload & short circuit. Hence external  safety protection should be provided. It is recommended to install protective device near   equipment.

* PQ meter millivolt ranges :- Please use lead wires having total resistance o f 35 milliohms.

 As per information o n Sticke r & Dial of the i nstrument.

* Equipment must be installed & maintained by suitably qualified person.

* If equipment is used in a manner not specified by the manufacturer, the protection provided by  the equipment may be impaired.

* CTs must not be opened when on load.

* Instruments with damaged bezels or window glasses must be disconnected from mains.

* Back cover must be snapped into place for protection against accidental contact.

* Disconnect supply before replacing scales, bezels or window glasses.

* There are no user serviceable parts in this product. Please contact at the address given, for   clarifications if any.

* Indoor use* Altitude up to 2000 mo

* Temp.-10 to+55 Co

* Maximum relative humidity 80% for temperature up to 31 C decreasing linearly to 50% relativeo o

  humidity at 40 C & further linear decrease to 25% relative humidity at 55 C* Basic Insulation* Pollution degree2* Installation Categories (over voltage category): For EQ, PQ, Size 72, 96 & 144 - 600V CAT III  Size 48 - 300V CAT III  For BIQ, BIEQ, VL, VQ, ZQ PL, - 300V CAT III

2.0 Meaning of Symbols :

3.0 Safety specifications (IEC1010-1 A2)

4.0 Equipment Installation:

5.0 Operation & Maintenance :

6.0 Equipment Rating :

o>100 C

 

Caution hot surface

EU conformity mark

Instruction SheetIC : 30338

For Analogue Panel Meters type EQ, PQ, BIQ, & BIEQ, PL,VL,VQ,ZQ

1.0 General:

This meter complies with International standard IEC 1010-1 Amendment 2 for safety and isconforming to routine tests as per IEC 51. The test equipment used is traceable to National/International standards.

This product must be used after mounting in panel with hardware supplied, with wirespermanently connected to terminals as specified.

* Due consideration should be given to space behind the meter to allow for bends in connecting  cables.

* Ensure that wires do not remain under tension.

* The products do not ha ve internal protection against overload & short circuit. Hence external  safety protection should be provided. It is recommended to install protective device near   equipment.

* PQ meter millivolt ranges :- Please use lead wires having total resistance of 35 milliohms.

 As per i nforma tion on Sticker & Dial of the ins trumen t.

* Equipment must be installed & maintained by suitably qualified person.

* If equipment is used in a manner not specified by the manufacturer, the protection provided by  the equipment may be impaired.

* CTs must not be opened when on load.

* Instruments with damaged bezels or window glasses must be disconnected from mains.

* Back cover must be snapped into place for protection against accidental contact.

* Disconnect supply before replacing scales, bezels or window glasses.

* There are no user serviceable parts in this product. Please contact at the address given, for   clarifications if any.

* Indoor use* Altitude up to 2000 m

o* Temp.-10 to+55 C

o* Maximum relative humidity 80% for temperature up to 31 C decreasing linearly to 50% relative

o o  humidity at 40 C & further linear decrease to 25% relative humidity at 55 C

* Basic Insulation* Pollution degree2* Installation Categories (over voltage category): For EQ, PQ, Size 72, 96 & 144 - 600V CAT III  Size 48 - 300V CAT III  For BIQ, BIEQ, VL, VQ, ZQ PL, - 300V CAT III

2.0 Meaning of Symbols :

3.0 Safety specifications (IEC1010-1 A2)

4.0 Equipment Installation:

5.0 Operation & Maintenance :

6.0 Equipment Rating :

o>100 C

 

Caution hot surface

EU conformity mark

Instruction SheetIC : 30338

For Analogue Panel Meters type EQ, PQ, BIQ, & BIEQ, PL,VL,VQ,ZQ

1.0 General:

This meter complies with International standard IEC 1010-1 Amendment 2 for safety and isconforming to routine tests as per IEC 51. The test equipment used is traceable to National/International standards.

This product must be used after mounting in panel with hardware supplied, with wirespermanently connected to terminals as specified.

* Due consideration should be given to space behind the meter to allow for bends in connecting  cables.

* Ensure that wires do not remain under tension.

* The products do not have internal protection against overload & short circuit. Hence external  safety protection should be provided. It is recommended to install protective device near   equipment.

* PQ meter millivolt ranges :- Please use lead wires having total resistance o f 35 milliohms.

 As per information o n Sticke r & Dial of the i nstrument.

* Equipment must be installed & maintained by suitably qualified person.

* If equipment is used in a manner not specified by the manufacturer, the protection provided by  the equipment may be impaired.

* CTs must not be opened when on load.

* Instruments with damaged bezels or window glasses must be disconnected from mains.

* Back cover must be snapped into place for protection against accidental contact.

* Disconnect supply before replacing scales, bezels or window glasses.

* There are no user serviceable parts in this product. Please contact at the address given, for   clarifications if any.

* Indoor use* Altitude up to 2000 m

o* Temp.-10 to+55 C

o* Maximum relative humidity 80% for temperature up to 31 C decreasing linearly to 50% relative

o o  humidity at 40 C & further linear decrease to 25% relative humidity at 55 C

* Basic Insulation* Pollution degree2* Installation Categories (over voltage category): For EQ, PQ, Size 72, 96 & 144 - 600V CAT III  Size 48 - 300V CAT III  For BIQ, BIEQ, VL, VQ, ZQ PL, - 300V CAT III

2.0 Meaning of Symbols :

3.0 Safety specifications (IEC1010-1 A2)

4.0 Equipment Installation:

5.0 Operation & Maintenance :

6.0 Equipment Rating :

o>100 C

 

Caution hot surface

EU conformity mark