Jae Lim Chung MD, MBA Jae Lim Chung MD, MBA 1,3 1,3 ; Jin Pyo Hong MD ; Jin Pyo Hong MD 2,3 2,3 ; Kyoung Yul Seo MD, ; Kyoung Yul Seo MD, PhD PhD 3 ; Eung Kweon Kim MD, PhD ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD ; Tae-im Kim MD, PhD 3 1 Noo Ne Eye Hospital, Seoul, Korea 2 Sae Yan Eye Clinic, Seoul, Korea 3 Department of Ophthalmology, Yonsei University College of Medicine, Seoul, Korea. None of the authors have any commercial or financial interests

Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD 3 1 1 Noo Ne Eye Hospital, Seoul,

Embed Size (px)

DESCRIPTION

 Shin-Nippon Co, Ltd, Japan  Measuring capacity Corneal Radius of Curvature ○ 6.50 ~ 9.40 mm Corneal Keratometric Power ○ 36 ~ 52 D Axis Corneal Astigmatism ○ 0 ~ 180° 3

Citation preview

Page 1: Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD 3 1 1 Noo Ne Eye Hospital, Seoul,

Jae Lim Chung MD, MBAJae Lim Chung MD, MBA1,31,3; Jin Pyo Hong MD; Jin Pyo Hong MD2,32,3; Kyoung Yul Seo MD, ; Kyoung Yul Seo MD, PhDPhD33; Eung Kweon Kim MD, PhD; Eung Kweon Kim MD, PhD33; Tae-im Kim MD, PhD; Tae-im Kim MD, PhD33

11Noo Ne Eye Hospital, Seoul, Korea22Sae Yan Eye Clinic, Seoul, Korea

33Department of Ophthalmology, Yonsei University College of Medicine, Seoul, Korea.

None of the authors have any commercial or financial interests

Page 2: Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD 3 1 1 Noo Ne Eye Hospital, Seoul,

For Successful Outcome Of Toric IOL Precise Measurement Of Preoperative

Corneal Astigmatism Crucial Point !

To Determine Accuracy & Correlation of 6 Different Keratometers & Topographers

2

Page 3: Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD 3 1 1 Noo Ne Eye Hospital, Seoul,

Shin-Nippon Co, Ltd, Japan

Measuring capacityCorneal Radius of

Curvature ○ 6.50 ~ 9.40 mm

Corneal Keratometric Power ○ 36 ~ 52 D

Axis Corneal Astigmatism ○ 0 ~ 180°

3

Page 4: Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD 3 1 1 Noo Ne Eye Hospital, Seoul,

Canon, Japan Measuring capacity

Corneal Radius of Curvature ○ 5.5 ~ 10 mm

Corneal Refractive Power ○ 33.75 ~ 61.25 D

Axis Corneal Astigmatism ○ 0 ~ 180°

4

Page 5: Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD 3 1 1 Noo Ne Eye Hospital, Seoul,

• Carl Zeiss Meditec, Germany

• 2.5 mm measurement diameter• Cf. Manual keratometers ~ 3 mm

• 30 data points per measurement• 6 spots x 5 samples

5 images automatically captured in 0.5 sec

Displayed result is average of 5 captured images

5

Page 6: Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD 3 1 1 Noo Ne Eye Hospital, Seoul,

6

Bausch & Lomb, USA Topography-based Simulated K-readings

(=Sim K’s) c Axis 3 & 5 mm zone

statisticsAverage & SD of Mean &

Astigmatic Power Steep & Flat Axis

Page 7: Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD 3 1 1 Noo Ne Eye Hospital, Seoul,

Oculus, Germany Scheimpflug image-

based Topographic analysis of Corneal Astigmatism

7

Page 8: Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD 3 1 1 Noo Ne Eye Hospital, Seoul,

Tracey Technologies, USA Dual Acquisition Unit

EysSys Corneal Topographer○ Placido ring-based

Wavefront analyzer

8

Page 9: Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD 3 1 1 Noo Ne Eye Hospital, Seoul,

Study PopulationPreoperative visit for Cataract Surgery257 eyes of 141 patients○ Age : 54.5 ± 20.5 yrs○ Sex : Male - 54, Female - 87

Manual Keratometric MeasurementDuplicated examination by 2 skillful

examinersConsidered as Standard

Compared with 5 other Keratometers9

Page 10: Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD 3 1 1 Noo Ne Eye Hospital, Seoul,

Pearson Correlation analysis Between Manual Keratometery Vs Other Methods

10

Vs. Manual Keratometer

Pearson Correlation Coefficient

iTrace 0.947

ORBscan 0.921AutoKeratometer 0.919IOL master 0.857

Pentacam 0.847

Page 11: Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD 3 1 1 Noo Ne Eye Hospital, Seoul,

Pearson Correlation analysis Between Manual Keratometery Vs Other Methods

11

Vs. Manual Keratometer

Pearson Correlation Coefficient

iTrace 0.856

AutoKeratometer 0.852IOL master 0.751ORBscan 0.646

Pentacam 0.633

Page 12: Jae Lim Chung MD, MBA 1,3 ; Jin Pyo Hong MD 2,3 ; Kyoung Yul Seo MD, PhD 3 ; Eung Kweon Kim MD, PhD 3 ; Tae-im Kim MD, PhD 3 1 1 Noo Ne Eye Hospital, Seoul,

Duplicated Manual Keratometric Measurements By 2 Experienced Operators Taken As Standard

iTrace Approximated This Standard The Most

Auto Keratometer, ORB & IOL Master Showed To Be Reasonably Accurate.

Trend For ORB Fail To Read Accurately In Steep Axis

12