Upload
sandrajpatino
View
223
Download
0
Embed Size (px)
Citation preview
8/10/2019 Jo Had Erson Pshielding petet Shielding
http://slidepdf.com/reader/full/jo-had-erson-pshielding-petet-shielding 1/8
AAPM Summer School, 2007 [email protected] 1
PET Shielding andPET Shielding andRadiationRadiation
ProtectionProtection
Jon A. Anderson, PhDJon A. Anderson, PhDDepartment of RadiologyDepartment of Radiology
The University of Texas Southwestern Medical Center at DallasThe University of Texas Southwestern Medical Center at Dallas
July 2007July 2007American Association ofAmerican Association ofPhysicists in MedicinePhysicists in Medicine
Summer SchoolSummer School
AAPM Summer School, 2007 [email protected] 2
What Is Different When You HavePET/CT in Your Facility?
1) 511 keV energy-increases exposure rate from doses,patients-greatly increases thickness of requiredshielding
2)Requirements for patient handlingduring injection and uptake phase
3) Combined modality scanners(PET/CT) require consideration of bothgamma-ray and x-ray hazard
AAPM Summer School, 2007 [email protected] 3
The 18F-Injected Patient as a Source(average of different investigators, 2003 )
0.103 ( Sv/hr)/MBq0.383 (mrem/hr)/mCi
0.104 ( Sv/hr)/MBq0.383 (mrem/hr)/mCi
0.075 ( Sv/hr)/MBq0.279 (mrem/hr)/mCi
0.018 ( Sv/hr)/MBq0.065 (mrem/hr)/mCi
all at 1 m from surface ofbody, average value fromseveral investigators
Anterior
Inferior
Lateral
Superior
compare this to0.014 ( Sv/hr)/MBq or0.05 (mrem/hr)/mCi
for99m
Tc:18
F valuesfactor of 8 larger!
not asanisotropic as itmight seem
AAPM Summer School, 2007 [email protected] 4
A Revealing Comparison of LeadRequirements: X-Ray vs PET
Lead Thickness Requiredmm (in, to next 1/16)
#HVL's
46.0 (1 13 /16)1.366 (< 1/16)1032.5 (1 5/16)0.718 (< 1/16)8
19.0 (3/4)0.278 (< 1/16)49.9 (7/16)0.103 (< 1/16)25.3 (1/4)0.044 (< 1/16)1
PET 2X-ray 1
(average primary forrad room)
1. NCRP 147: Structural Shielding Design for Medical X-Ray Imaging Facilities2. Simpkin, 2004, developed for AAPM Task Group on PET Facility Shielding
Even asingle half-value layerfor PET isanexpensiveproposition!
AAPM Summer School, 2007 [email protected] 5
A Revealing Comparison of LeadRequirements: X-Ray vs PET
Lead Thickness Requiredmm (in, to next 1/16)
#HVL's
46.0 (1 13/16)1.366 (< 1/16)1032.5 (1 5/16)0.718 (< 1/16)8
19.0 (3/4)0.278 (< 1/16)49.9 (7/16)0.103 (< 1/16)25.3 (1/4)0.044 (< 1/16)1
PET 2X-ray 1
(average primary for
rad room)
1. NCRP 147: Structural Shielding Design for Medical X-Ray Imaging Facilities2. Simpkin, 2004, developed for AAPM Task Group on PET Facility Shielding
Even asingle half-value layerfor PET isanexpensiveproposition!
Remember this morning's bottom line: for DX,we can apply 3 models from NCRP 147 and findthat 1/16" is (usually) the answer, with some tospare. We usually calculate the "closest" point.
Not true in PET. As we will see, it is true that
normally we need 1-3 HVL's of shielding. Wetend to put just what we need, due to $$$.
Implication: At every protection point, we needto include all sources that can be contributing tothe dose at that point (i.e. multiple injectionrooms, scan rooms, etc.). Corresponds toadjusting protection limit downward.
AAPM Summer School, 2007 [email protected] 6
Workflow at the PET Center(FDG Whole Body Scans)
Receive dosesArrival of patient
Injection of Pt
Uptake of pharmaceutical
Assay of dose
Transport Pt to scanner
Position Pt
Scan
QA Check of Scan
Pt instruction and prep
Release Pt
Read study
Distribute to PACS or Media
Have Pt empty bladder
30-60 min
10-30 min
5-10 min
**
*
* steps withhighesttechnologistexposure
*
*
8-20 mCi
8/10/2019 Jo Had Erson Pshielding petet Shielding
http://slidepdf.com/reader/full/jo-had-erson-pshielding-petet-shielding 2/8
AAPM Summer School, 2007 [email protected] 7
Technologist Doses
0.000
0.050
0.100
0.150
Reference
D o s e
/ A c
t i v i t y H a n
d l e d
S v
/ M B q
( m r e m
/ m C i )
SI Units 0.018 0.012 0.023 0.019 0.019 0.011 0.009 0.029 0.021 0.018
Conventional Units 0.067 0.044 0.085 0.069 0.069 0.041 0.032 0.107 0.077 0.066
B en et ar Ch is ea Ch is ea M cEl ro y UT SW Ro be rt s Gu il le t B ir an Ye st er Ave ra ge
Magnitude of TechnologistExposure Consistent with conventional nuclear
medicine practice, most of technologistdose comes from positioning, transport,and injection.
average dose/procedure< 10 Sv (1 mrem)
AAPM Summer School, 2007 [email protected] 8
More on Technologist Exposure
1) Technologist dose will probably drop as experienceincreases
Over a two year period with the same technologists, we saw a40% decrease in radiation dose per unit activity handled.
2) For 0.018 S v/(MBq injected)370 MBq (10 mCi) injected/pt
10 pt/day
Yearly: 16.6 mSv (1660 mrem) (<5 rem, >ALARA trigger)9 Months: 12.5 mSv (1250 mrem) (> declared pregnancy limit)
AAPM Summer School, 2007 [email protected] 9
Operating Suggestions to MinimizeTechnologist DoseMinimize handling time. Use unit doses.
Use tungsten syringe shields, employ syringe carriers,transport carts, etc. to minimize handling exposure.
Instruct patient before injection. Minimize contactafterward.
Establish IV access with butterfly infusion set.
Use other personnel for hot patient transport.AAPM Summer School, 2007 [email protected] 10
ScanRoom 1
West Corridor
Office 3
Future CameraRoom
Office 2
Office 1
Injection#2
BreakRoom
Injection#1
Storage
HotToilet
Hot Lab
Dock LaundryScanUtility
ScanRoom 2
SouthCo rridor
Reception &Waiting
BusinessOffice
Control
East Corridor
ReadingRoom
PET Facility Tour: University of TexasSouthwestern Medical Center at Dallas(2001-2007)Radioactive
Materials
Patients
AAPM Summer School, 2007 [email protected] 11
ScanRoom 1
West Corridor
Office 3
Future CameraRoom
Office 2
Office 1
Injection#2
BreakRoom
Injection#1
Storage
HotToilet
Hot Lab
Dock Laundry
ScanUtility
ScanRoom 2
SouthCo rridor
Reception &Waiting
BusinessOffice
Control
East Corridor
ReadingRoom
PET Facility Tour: University of TexasSouthwestern Medical Center at Dallas(2001-2007)Radioactive
Materials
Patients
If you can get involved in thearchitectural planning, you can savesome money, by using distance insteadof lead.
AAPM Summer School, 2007 [email protected] 12
Hot Lab Details: Dose StorageArea
Notes:1) Floorprotection(containersweigh > 66 lbs)
2) Space neededdepends on howoften deliveriesare made; mayhave >100 mCihere at a time,even for onescanner3) Extrashielding maybe required
alternative Wtransport carrier
8/10/2019 Jo Had Erson Pshielding petet Shielding
http://slidepdf.com/reader/full/jo-had-erson-pshielding-petet-shielding 3/8
8/10/2019 Jo Had Erson Pshielding petet Shielding
http://slidepdf.com/reader/full/jo-had-erson-pshielding-petet-shielding 4/8
AAPM Summer School, 2007 [email protected] 19
Site Evaluation for PET ShieldingUses of adjacent spaces (including above and below) andoccupancy factors for them
# patients/weekisotopes to be used, activity/pttypes of PET studies to be performed (brains, WB, cardiac)uptake time and scan time for this equipment/study/center
dose delivery schedule (once a day?, multiples?); maximumactivity on hand
CT technique factors (kVp, mAs/scan [depends of # beds])# scans per patient (additional diagnostic scans?)amount of "non-PET" CT workload expected
AAPM Summer School, 2007 [email protected] 20
Radiation Sources to Include in theShielding PlanDoses (pre-injection) in Hot Lab
Calibration sources for scanner
Patient (post injection, in uptake rm)
Patient (in scanner, hot toilet)
TX Sources in scanner (PET)
CT x-ray source (for PET/CT)
require isotopicworkloadparameters:pts/wk, mCi/ptuptake,scan timesisotope type anddelivery scheds
require CT x-rayworkload factors,techniques,include non-PETCT work
AAPM Summer School, 2007 [email protected] 21
P: Radiation Protection TargetsLimitations ALARA
per 10CFR20 Action LimitRadiation workers 50 mSv/yr 5 mSv/yr
(5000 mrem/yr) (500 mrem/yr)Pregnant worker's fetus 5 mSv/9 mo
(500 mrem/9 mo)
Members of public 1 mSv/yr(from each licensed operation) (100 mrem/yr)in any hour, not to exceed .02 mSv
(2 mrem)
GUIDANCE: ALARA -- As Low As Reasonably Achievable
Targets
controlled areas:100 Sv/wk or10 mrem/wk
uncontrolled areas:20 Sv/wk or2 mrem/wk
AAPM Summer School, 2007 [email protected] 22
PET Facility Throughput Example:90 Minute Uptake, 30 Minute Scan
7:00 8:00 9:00 10:00 11:00 12:00 13:0014:0015:00 16:00 17:00
1
3
5
7
9
11
13
15
P a
t i e n
t N u m
b e r
Time of Day
Phase
Uptake
Scanning
N: Maximum Workload Estimation
#pts/day = (T work - Tuptake )/T scan_rm # uptake areas = T uptake /T scan_rm
This facility needsthree uptake rooms
AAPM Summer School, 2007 [email protected] 23
18F: A Plethora of Dose RateConstants for Point Sources (TG-108)
(mrem/hr) m 2/mCi0.6956( Sv/hr) m 2/MBq0.188Maximum Dose (ANS-1977)
(mrem/hr) m 2/mCi0.6771( Sv/hr) m 2/MBq0.183Deep Dose Equivalent(ANS-1977)
(mrem/hr) m 2/mCi0.5476( Sv/hr) m 2/MBq0.148Tissue Dose Constant
(mrem/hr) m 2/mCi0.5291( Sv/hr) m 2/MBq0.143Effective DoseEquivalent (ANS-1991)
(mrem/hr) m 2/mCi0.4958( Sv/hr) m 2/MBq0.134 Air Kerma RateConstant
(mR/hr) m 2/mCi0.5735( R/hr) m 2/MBq15.5Exposure RateConstant
Conventional UnitsSI Units18 F Rate Constants
TG-108 recommends0.143 ( Sv/hr)/MBq0.53 (mrem/hr)/mCifor F-18 bare source
AAPM Summer School, 2007 [email protected] 24
Dose Rate from 18F Injected Patient at 1 m
0.000
0.100
0.200
0.300
0.400
0.500
0.600
Source
D o s e
R a
t e
[ ( S v
/ h r ) / M B q
]
[ ( m r e m
/ h r ) / m
C i ]
S IUn its 0.075 0.055 0.100 0.150 0.137 0.08866 0.097 0.100
Conventional Units 0.279 0.203 0.370 0.553 0.508 0.328 0.359 0.372
K e ar f o tt 19 92 C h i se a 1 99 7 C r o ni n 1 9 99 B e n et a r 2 0 00 W h it e 2 00 0Yester * 2005(attenuation)
Massoth 2003(unpubl.)
Average
: The 18F-Injected Patient asa Source (retained activity)
sources of variation:delay time to measurement,micturationstatus, exposure-to-dose conversion, etc.
about 20% of dose will be in bladder after 1-2 hours;TG108 uses 15%
TG-108 recommends(0.092 Sv/hr)/MBq(0.34 mrem/hr)/mCi
8/10/2019 Jo Had Erson Pshielding petet Shielding
http://slidepdf.com/reader/full/jo-had-erson-pshielding-petet-shielding 5/8
AAPM Summer School, 2007 [email protected] 25
Correction for Decay During DoseIntegration Period: Reduction Factor
Correction for Decay of F-18
0
0.2
0.4
0.6
0.8
1
1.2
0 20 40 60 80 100 120 140 160 180 200
Time [minutes]
R e
d u c
t i o n
F a c
t o r
R t
R 0
t
TA0
e
T
T half
ln 2( )
−⌠
⌡d
0
t
TA0
⌠
⌡d
= 1.443 * (T 1/2/t) * (1 - exp(-0.693 * t/T 1/2)
AAPM Summer School, 2007 [email protected] 26
Effect of CorrectionsCumulative Dose 1 m from Patient, 370 MBq F-18
0
20
40
60
80
100
120
140
160
180
0 20 40 60 80 100 120 140 160 180 200
Time [minutes]
D o s e
[ m i c r o
S v
]
Uncorrected
Corrected for Physical Decay
Corrected for Physical Decay,Self-ShieldingCorrected for Physical Decay,Self-Shielding, Voiding
Effects of Adding Corrections toDose Calculation
Inject @ 5 minutesVoid @ 65 minutes
A factor of 2= 1/4" of lead
AAPM Summer School, 2007 [email protected] 27
Simplified Task Group Formalism:Uptake Room
B, the required barrier transmission factor, will be calculated as
B =P * d 2
Γ * T * N w * (A0 * F tot * t * R t)
B =10.9 * P[ μSv] * d[m] 2
T * N w * (A0[MBq] * tU[hr] * R tU)
10.9 is 1/ in (hr/ Sv)(MBq/m 2);F tot = 1 (no physical decay prior to injection, no
elimination)R tU = reduction factor for uptake time t U
UptakeRoom:
AAPM Summer School, 2007 [email protected] 28
Simplified Task Group Formalism:Scan Room
B, the required barrier transmission factor, will be calculated as
ScanBay:
12.8 includes the value of and the effects of voiding 15% of theinjected activity before imaging;
FU=exp(-0.693t U /T 1/2), the physical decay of the isotope before thepatient enters scan bay
B =P * d 2
Γ * T * N w * (A0 * F tot * t * R t)
B =12.8 * P[ μSv] * d[m] 2
T * N w * (A0[MBq] * F U * t I[hr] * R tI)
AAPM Summer School, 2007 [email protected] 29
Lead
0.0001
0.0010
0.0100
0.1000
1.0000
0 5 10 15 20 25 30 35 40 45 50
Thickness(mm)
T r a n s m
i s s
i o n
Going from Barrier Transmissionto Shield Thickness
Monte Carlocalculations byDouglasSimpkin (2004)Monte Carlo Simulation
(Broad Parallel Beam)
Constant TVL 16.6 mm
x B( ) 1α γ ln
B γ− β
α
+
1
βα
+:=
Curves and fitting parameters for ironand concrete arealso found in thereport
AAPM Summer School, 2007 [email protected] 30
Fitting Parameters for DifferentMaterials per TG108
γ
[cm -1] [cm -1]Lead 1.543 -0.4408 2.136
Concrete 0.1539 -0.1161 2.0752Iron 0.5704 -0.3063 0.6326
Arch er Parameters
B x( ) 1β
α+
eα γ x β
α−
1−γ
x B( )1
α γln
B γ− β
α+
1β
α+
8/10/2019 Jo Had Erson Pshielding petet Shielding
http://slidepdf.com/reader/full/jo-had-erson-pshielding-petet-shielding 6/8
8/10/2019 Jo Had Erson Pshielding petet Shielding
http://slidepdf.com/reader/full/jo-had-erson-pshielding-petet-shielding 7/8
AAPM Summer School, 2007 [email protected] 37
Grid Calculation: Shielded
Y image
DoseToTargetRatio
No shielding in walls in excess of 5/16" Pb; did require ceiling, floorshielding. Was not necessary to run "box" to ceiling.
AAPM Summer School, 2007 [email protected] 38
Grid Calculation: No Shielding
Sources: Injection room, HL, HWC, Scanner, CT, Cal Source4 pts/day, 1 hr in uptake, 2 hrs in scan room
Scan InjCHT
DoseToTargetRatio
Ratio of calculated dose to t arget dose,adjusted for occupancy
Office/Lab
Reading
Corridor C or r i d or
R e a d i n
g
Of f i c e / L a b
E l e v a t or
E l e v a t or
L o b b y
OOPS's happen with complicatedschemes: Both floor and ceiling
needed lead, installed as lead sheetbonded to plywood panels and heldin brackets fastened to structuralweb, but different thickness aboveand below.
The contractor switched them inspite of drawings and labeled leadpallets.
AAPM Summer School, 2007 [email protected] 39
Example: University of TexasSouthwestern Medical Center 2007
STORAGE2.490
INJECTION42.480F
5 4 "
5 4 "
MANAGER2.591H
CORRIDOR2.481
CORRIDOR2.5
READING22.591D
READING12.591E
SCANNERBAY32.591C
SCANNERBAY22.591B
PHYSICIAN22.440
CORRIDOR2.482
CLINICALSTORES2.480HINJECTION3
2.480EINJECTION2
2.480D'HOT'TOILET
2.480A
CONTROL/OBSERVATION2.591
NURSE2.481B
FILEROOM
2.591G
PHYSICIAN12.450
LAB2.591F
SUB-WAITING2.481A
FUTURESCANNERBAY1
2.591A
SCANUTILITY
2.592
'HOT' TOILET2.480G
INJECTION12.480B
'HOT'LAB2.480C
Overalldesign: Nolead in excessof 3/8". Northwall ofinjectionrooms, hot labshielded with16" of dry-laid, full-densityconcrete block
AAPM Summer School, 2007 [email protected] 40
Look Up, Down, and Sideways
Floor Plan Relative Dose Map on Floor Above
Mechanical Space
Exterior
ElectricalElectrical Elevator,Lobby
Corridor Corridor
M e c h a n i c a l S p a c e
Glass Wash
Duct penetrations in ceilingrequired separate shielding.
AAPM Summer School, 2007 [email protected] 41
Dry-Laid Concrete Block as aFlexible Alternative
Use full density concreteblocks (not standarditems!).
Lay to offset seams.
Provide cosmetic stud-wall w gypsum board toprevent tampering ifarea is not controlled.
AAPM Summer School, 2007 [email protected] 42
An Alternative: Shadow Shielding
from JA Anderson, RJ Massoth, and LL Windedahl, 2003 AAPM
8/10/2019 Jo Had Erson Pshielding petet Shielding
http://slidepdf.com/reader/full/jo-had-erson-pshielding-petet-shielding 8/8
AAPM Summer School, 2007 [email protected] 43
Acknowledgements
• Michael Viguet, CNMT• Dana Mathews, MD• Thomas Lane, PhD• Richard Massoth, PhD• Larry Windedahl• Doug Simpkin, PhD• Mark Madsen, PhD
AAPM Summer School, 2007 [email protected] 44
TheEnd