Upload
jonas-hill
View
216
Download
0
Embed Size (px)
Citation preview
Lab 3 Static Calibration of Electronic Pressure
Transmitters using Manometers
February 1, 2013
Group 0
Miles Greiner
Lab Instructors:
Michael Goodrick and Alyssa Hawthorne
Abstract• The purpose of this lab is to
– Calibrate an electronic pressure transmitter using an inclined-well monometer, and
– Interpret the manometer manufacturer’s accuracy specification. • The best estimate for the true pressure head in terms of the
transmitter reading is: – h = 1.015hT + 0.007 inch H2O, with a confidence interval of ST,M
= ± 0.010 inch-H2O (68%)– The maximum error in the reading is 0.062 inch H2O, which is
roughly equal to the manufacturer’s specification.• The manufacturer stated manometer uncertainty is 6 times
larger than the value of SM,T, corresponding to a nearly 100% confidence level.
• The manufacturer’s specified uncertainty is in the same order of magnitude of the error measures determined in this experiment.
Table 1 Equipment Specifications and Calibration
• The manometer is used to calibrate the transmitter because, after two years, the manometer’s uncertainty is smaller.
Number I3
Range 0-4 inch H20
Relative Accuracy ±1% FS
Absolute Accuracy ±0.04 inch H2O
Number T6
Range 0-3 inch H2O
Relative Accuracy ±0.25% FSStability 1% FS/year
Absolute Accuracy (after 2 years)
0.06 inch H2O
Tranfer function hT = (3 inch H2O)(IT-4mA)/(16mA)
Transmitter
Manometer
Table 2 Calibration Data
• This table shows one cycle of increasing and decrease pressure calibration data.
• The transmitter pressure head was determined from the measured current using the manufacture specified equation in Table 1
• The transmitter pressure head reading did not return to zero at the end of the descending cycle.
Manometer
Reading, hM
Transmitter
Current, IT
Transmitter
Reading hT
[inch-H2O] [mA] [in H20]
0 4 0.0000.4 6.03 0.3811 9.12 0.960
1.6 12.33 1.5622.4 16.6 2.3633 19.67 2.938
2.6 17.68 2.5651.8 13.47 1.7761.2 10.29 1.1790.6 7.09 0.5790 4.01 0.002
Fig. 1 Calibration Data and Linear Fit
• The transmitter reading is consistently lower than the manometer-indicated pressure.
• Standard errors of the estimates for the transmitter and monometer are both ST,M = 0.010 inch-H2O = SM,T
• The manufacturer stated accuracy (0.06 inch-H2O) for the transmitter is 6 times larger than SM,T, corresponding to a nearly 100% confidence level
Calibration CorrelationhT,F = 0.985hM - 0.0069
0
0.5
1
1.5
2
2.5
3
0 0.5 1 1.5 2 2.5 3
Tran
smitt
er R
eadi
ng, h
T[in
H20
]
Manometer Pressure Head, hM [in H2O]
hT = hMPerfect Insturment
Response
Fig. 2 Error
• Error in manometer reading increases with pressure• Maximum error magnitude (0.062 inch H2O) is roughly equal to
the manufacturer specified accuracy (0.06 inch H2O)
-0.07
-0.06
-0.05
-0.04
-0.03
-0.02
-0.01
0
0.01
0 0.5 1 1.5 2 2.5 3
E =
h T-h
M, [
inch
H20
]
Monometer Pressure Head, hM [inch H2O]
Fig. 3 Calibration Deviation
• ST,M characterizes the deviations over the full range of hM
• The ascending deviations are generally positive while the descending ones are negative, which may be caused by hysteresis.
• There are no systematic deviations form the fit correlation, indicating the instrument response is linear.
-0.020
-0.015
-0.010
-0.005
0.000
0.005
0.010
0.015
0.020
0 0.5 1 1.5 2 2.5 3
e =
h TF
-hT
[inch
H2O
]
hM [inch H2O]
Ascending DataDecending Data
e = -ST,M
e = ST,M
Interpretation of Measurement Question
Transmitter Reading hT,R = 1.9 inch H2O
Expression for Best Estimate h = 1.015hT,R + 0.007 inch H2O
Standard Error of the Estimate for the Measured Pressure
0.010 inch H2O
68% Confidence Interval 1.94 ± 0.01 inch H20
Error in the Reading (if the transmitter had not been
calibratedhT,R - h = -0.04 inch H2O