Upload
syed-muhammed-alee-kazmi
View
227
Download
0
Embed Size (px)
Citation preview
8/11/2019 Lab Report 1 Updated
1/28
SAGE OF OSCILLOSCOPE
LOCATION = RESEARCH 1 EE LABS
2ND SEMESTER
DATE: 21ST FEBR ARY 2014
GRO P 5
CREATOR OF THIS LAB REPORT: ALEE KAZMI
GRO P MEMBERS: DORIN G CLIS , CLEMENCE JOMIMASSA
NSINDO
MAILBO N MBER !2
8/11/2019 Lab Report 1 Updated
2/28
SAGE OF OSCILLOSCOPE
INTRODUCTION TO THE EXPERIMENT
In these experiments we will try to identify and see the basic principles of the oscilloscope. These
experiments involve the charging and discharging of capacitors.
OSCILLOSCOPE
It is a device which displays signals in the form of voltage and time graphs which are then analysed
from the screen of the oscilloscope. They are calibrated and then adjusted so that voltage and the
time can be read directly from a screen connected in it. The experiments show us the methods of
analysing AC and DC signals on an oscilloscope. The simple oscilloscope can be divided into various
sections: The ertical control! The hori"ontal control! Trigger Controls and the display.
Vertical Control
Contains a sensitivity #nob$ AC%DC%&round selector and the ' vertical input plug.
Trigger Control
Contains the start of the x(axis sweep. This also contains the auto set button which automatically
resets the wave to its original state if its display settings have been altered too much.
Horizontal Input
This includes the timebase settings and the hori"ontal sensitivity switch. Another input for another
signal is also available for dual signals.
)scilloscope can be used for multiple tas#s such as *elative measurements$ +hase difference$
,re-uency measurement$ +ulse rise or fall measurements$ Time and voltage measurements.
astly the use of probes with an oscilloscope is very useful. It contains resistance of its own which
can be used to attenuate the signals since they act as a voltage divider. The attenuation can go up
to '/x. At the same time$ this helps to isolate the capacitive load presented by the probe0s cable.
The probe itself has a button on it to adjust the attenuation.
CAPACITOR
A capacitor is a device which stores electrical charge between two conducting plates in an
insulator. Capacitance or rather the capability of a capacitor to store charge is directly proportional
to the surface areas of the plates$ and is inversely proportional to the separation between the
plates. Capacitance also depends on the dielectric constant of the substance separating the plates.
8/11/2019 Lab Report 1 Updated
3/28
SAGE OF OSCILLOSCOPE
SINAL PROPERTIES
This picture very
accurately sums up
the components of an
AC signal. The average
value of this signal.
The amplitude is the
distance from the
max to the / point.
+eriod is the time
ta#en for the wave to
complete one cycle.
The fre-uency is the
number of cycles a
wave completes in
once second
8/11/2019 Lab Report 1 Updated
4/28
SAGE OF OSCILLOSCOPE
EXPERIMENT SETUP AND O!SERVATIONS
APPRATUS USED IN ALL THE EXPERIMENTS'1 Te#tronix )scilloscope
21 3labo
41 '/5ohm resistor
61 +ower supply
71 8ignal &enerator device connected in wor#bench
91 Capacitor '/n,
1 ;readboard%Connecting wires as s#eleton
PART " #$ SAGE OF THE "ERTICAL CONTROL
MAIN O!%ECTIVE
This is a basic experiment in which the use of the vertical control which include the position #nob
for Channel ' and Channel 2$ their menu$ sensitivity #nob
8/11/2019 Lab Report 1 Updated
5/28
SAGE OF OSCILLOSCOPE
e-ual to the one on the probe else the wave does not display correctly. astly the auto set button
is pressed which resets the display of the wave to its original state. 8ince the wave was
propagating$ the *un%8top button was pressed to halt it.
At this moment$ a hard copy of the screen was ta#en. It should be noted that the sensitivity wasautomatically set at './ per graticule division.
The position #nob in the hori"ontal section
8/11/2019 Lab Report 1 Updated
6/28
SAGE OF OSCILLOSCOPE
The value of the voltage can be calculated using the following formula(:Counte) Di*i+ion+ in c, - Scale .actor in V/)i*i+ion 0 Value in *oltage
&'&'& MEASURE VOLTAE USIN THE CURSORS
PROCEDURE CARRIED OUT
The auto set button was pushed to reset the wave to its original setting. @ow the cursor button is
pressed which opens a menu. The type of cursor is selected to voltage in order to measure the
voltage. Although it is unnecessary since we are wor#ing with only one channel source but
nevertheless the source was selected to Channel '. A hard copy was ta#en at this instance.
The values of the upper$ lower cursors$ delta and various minor settings are calculated in the
following table.
&'&'1 MEASURE VOLTAE USIN THE MEASURE .UNCTION
PROCEDURE CARRIED OUTImmediately before the previous experiment$ the auto set button was pressed and then the
measure button. =in$ =axB and +#(+#B mode were selected. A hard copy was ta#en at this
Range
V V V
1.00V 2.2 2.72 0.52 0.1V 00.5V 2.18 2.74 0.5 0.1V !"
Vmin Vmax Vpp Estimated Resolution V/[div] Ve#ti$al Divsions
V/[div]
8/11/2019 Lab Report 1 Updated
7/28
SAGE OF OSCILLOSCOPE
instant with the mode selected. ,or 7//m range(:
,or ' range(:
The pea# to pea# voltage is 9//m for both ranges.
8/11/2019 Lab Report 1 Updated
8/28
SAGE OF OSCILLOSCOPE
&'&'2 S3ITCHIN THE INPUT COUPLIN
PROCEDURE CARRIED OUT
There is a slight modification in this experiment as we are filtering out the DC part of the signal.
This is done by selecting the AC coupling in C?' menu. It should be noted that in case of small
fre-uency it affects out measurements.. The auto set button is not pressed and a sine centred
curve is seen. 8ince the signal was not stable$ the trigger was adjusted with the 33 #nob in the
trigger section.
Range/,V Voltage Pea4 to Pea4/V
'// /.62
&'&'5 Veri67 t8e ,ea+ure,ent+ 9it8 a ,ulti,eter
PROCEDURE CARRIED OUT
The oscilloscope probe is removed from the signal generator and the 3labo multimeter is set to DC
and connected to the signal generator. The ground pin of the 3labo is then connected to pin ' andthe
8/11/2019 Lab Report 1 Updated
9/28
SAGE OF OSCILLOSCOPE
PART : #$ SAGE OF THE HORIZONTAL CONTROL
MAIN O!%ECTIVE
This is a basic experiment in which the use of the hori"ontal control which include the hori"ontal
position #nob for Channel ' and Channel 2$ their menu and the sensitivity #nob
8/11/2019 Lab Report 1 Updated
10/28
SAGE OF OSCILLOSCOPE
A table will be compile at the end. The time between the rising edges of two consecutive
rectangular components was measured using cursors. Delta time is distance between the first and
second cursors to "ero so in our case$ the wanted value
To measure the fre-uency of the rectangular component$ the signal is adjusted to cover the entire
portion of the screen. The scale factor of the time base is also recorded.
The amplitude of the rectangular part is also obtained from here since the wave is big and accurate
enough to be counted and then multiplied by the volts per division.
The fre-uency of the sine component is measured by "ooming enough to scale one sine wave. A
cross point is then used between the signal and the grid as reference points for the cursors. The
pea# to pea# value is measured using hori"ontal cursors. astly$ the falling edge of the rectangular
component is used to measure the time distance of the sine from the beginning of the falling edge
of rectangular component.
,or the rectangular component(:
Cursor'
8/11/2019 Lab Report 1 Updated
11/28
SAGE OF OSCILLOSCOPE
,or the AC component(:
&'1'( Dela7e) an) e;pan)e) *ie9 o6 +ignal+
PROCEDURE CARRIED OUT
The auto set button is pressed and the wave position adjusted. After that$ the hori"ontal menu
button is pressed and the =ain is selected. ,rom the side screen menu$ 0window "one is selected.
Two additional vertical bro#en lines are visible on the screen. 0>indow0 is selected from the side
screen menu and the selected region "ooms in. This is done because the wave is -uite small to be
measured accurately. Fsing the position #nob from the hori"ontal section$ the position of the wave
is adjusted till a little more then one period of the wave covers the graph. A hard copy is ta#en and
properties of the signals recorded.
PART ( #$TRIGGER SECTION
INTRODUCTION TO THE !ASICS
This is a basic experiment in which the use of the trigger which include$ the control set for theamplitude level$ trigger menu$ set level to 7/G$ force trigger$ trigger view and input ext trigger$ is
emphasi"ed and practised on. >e will perform a series of experiments focusing on specific groups
)#e*uen$+ o& t'e sine signal %u#so#1 %u#so#2 Delta,%u#so#- &&set om t'e g#ound ime distan$e o& sine"4.4 504mV 2.2V 2.7"V 0.7"V 2.4V 10.5ms 47.0
Vpp value o& sine $omponent 42mV p3!p3
8/11/2019 Lab Report 1 Updated
12/28
SAGE OF OSCILLOSCOPE
of the hori"ontal section.
The trigger menu contains a type$ source$ slope$ mode and coupling mode.
&'2'( U+ing t8e
8/11/2019 Lab Report 1 Updated
13/28
SAGE OF OSCILLOSCOPE
Trigger Po+ition/,+ Trigger Le*el/V Slope
='==,+ "'>2V Ri+ing
@ow the trigger is shifted slowly from (' to 9. At about /.6(6.$ the signal behaves strangely
and starts moving bac# and forth. At about 6.$ the signal moves correctly again. The vertical
resolution is changed to 7//m%Div and the trigger varied slowly between 2 and 4. At about 2.7
$ the graph moves bac# and forth. After this at about 2.$ it moves to the left no change can be
further detected in it. A hard copy is ta#en. The only difference to the previous image is obviously
that the slope is falling now and the graph shifter to the left.
8/11/2019 Lab Report 1 Updated
14/28
SAGE OF OSCILLOSCOPE
&'2'& Ot8er i,portant trigger control+
PROCEDURE CARRIED OUT
The probe is connection to pin 6 of the signal generator. The coupling is set to AC$DC and ground
coupling which show the AC$ DC and ground part of the wave respectively.At AC coupling$ the
wave is centred. )n DC$ it displays a triangular wave form and on ground there is / signal$ off
course.
The coupling is set to DC and probe connected to pin 9 of the signal generator. The auto set button
is pressed and a stationary signal is seen. After setting the trigger to (4$ the signal starts to move.
The probe tip is disconnected for pin 9 and the mode set to normal. The probe tip is again
connected bac# to pin 9 of the generator. There is no signal now. The trigger is adjusted and the
same signal returns.
PART & #$ SING THE PROBE
I
NTROD CTION TO THE BASICS
The use of probes and their implementation is discussed in this section. >e shall be using passive
probes with ':' and '/:' attenuation. The other #inds such as active probes and differential
probes are not used. The following is the schematic of a probe(:
The probe acts li#e a low pass filter.
8/11/2019 Lab Report 1 Updated
15/28
SAGE OF OSCILLOSCOPE
PROCEDURE CARRIED OUT
The probe is connected to pin 9 on the signal generator and the attenuation set to 'x. The probe
factor on the oscilloscope is also set to 'x. The coupling is set to DC and the auto set button
pressed. The time base is altered from the hori"ontal menu so that only one period is visible on thescreen. The vertical resolution is set to the highest possible range.
&'5': Pro
8/11/2019 Lab Report 1 Updated
16/28
SAGE OF OSCILLOSCOPE
The variable capacitor is turned a further / degrees in cloc#wise direction. ?ori"ontal and vertical
scale is adjusted again and a hard copy ta#en(:
.re?uenc7/MHz Vpp/V
'2./2 .'2
8/11/2019 Lab Report 1 Updated
17/28
SAGE OF OSCILLOSCOPE
&'5': Pro
8/11/2019 Lab Report 1 Updated
18/28
SAGE OF OSCILLOSCOPE
Pea4 to Pea4/V .re?uenc7/MHz
7.// '2./7
PART & #$ DISCHARGING THE CAPACITOR
PROCED RE CARRIED O T
The following circuit is build up from a breadboard.
The attenuation is set to '/x on the probe and the vertical menu. Trigger is set to normal mode$
falling edge. The supply is turned on and the capacitor charges in a few seconds. As soon as the
oscilloscope showed '/$ the supply cable was pulled out of the power source and resultantly the
capacitor started to discharge. The process is repeated a couple of time in order to adjust the time
base. The trigger mode is not set to 0single0 and the curve is adjusted so that it spreads over the
entire screen. A hard copy is ta#en(:
8/11/2019 Lab Report 1 Updated
19/28
SAGE OF OSCILLOSCOPE
Trigger
Po+ition/,+
Re+olution
/V
Mo)e Coupling Horizont
al Scale@+.actor/u
+
Trigger Le*el/V Slope
2./E/ 2.// @ormal 8ingle 7// ./6 ,alling
EVALUATION
"B
IHm posting the table from the first experiment again.
oltage oltage per divisionsJ@umber of divisions
,or './/%div sensitivity(:
min ' J 2.2 2.2
max 'J2. 2.
Range V delta Estimated ResolutionV
1.00V 2.2 2.7 0.5 0.1V
0.5V 1.4 2.4 1 0.1V
Vmin VmaxV/[div] Div Div V/[div]
8/11/2019 Lab Report 1 Updated
20/28
SAGE OF OSCILLOSCOPE
,or /.7%div sensitivity(:
min /.7 J '.6 /.
max /.7J 2.6 '.2
dc
8/11/2019 Lab Report 1 Updated
21/28
SAGE OF OSCILLOSCOPE
Accurac7 o6 o+cillo+cope 6or *ertical po+ition at non#zero#$
,or *ange of 2m%div to 2//m%div(:
e could ta#e the wrong reference point while ta#ing some readings or wrongly measure while
ta#ing the readings. =oreover$ "ero error in oscilloscope can also be an issue. astly$ there is a
probability of damaged probes.
Met8o) : Cur+or+B#$
The "ero error and damaged probes still apply. >e can ta#e wrongs reference points by not
properly
8/11/2019 Lab Report 1 Updated
22/28
SAGE OF OSCILLOSCOPE
Met8o) &U+ing ,ulti,eterB
Internal resistance of multimeter or "ero error in multimeter can account for error.
Fsing a higher resolution would mean mathematically a lower resolution number. This would in
turn decrease the amount of relative error after the formula has been applied. ?ence$ it is better
to use higher resolutions.
AC coupling affects the display of small fre-uencies and may filter small AC fre-uencies out too.
As observed from the calculations above$ the multimeter should be preferred over then
oscilloscope due to its much lower relative error of /.''G as compared to 4.EG.
:B
Error +ource+ 98ile ,ea+uring 6re?uenc7
ets consider the case of cursors. There could be a device error$ an error while aligning the cursors
or maybe a probe error. There is also an error as fre-uency is measured with a delay and an
expanded view. There might also be a device error$ error in the resolution$ effect of the probe or
the "ero error in the oscilloscope.
Accurac7 o6 t8e o+cillo+cope in *ariou+ ,o)e+#$
In single shot and sample mode:
K(
8/11/2019 Lab Report 1 Updated
23/28
SAGE OF OSCILLOSCOPE
able to thus put the cursors very accurately at appropriate areas. This further reduces the
possibility to use a wrong "ero value.
91 >e should use the oscilloscope to measure time and fre-uency because not only is the reading
more user friendly to understand but the accuracy can be increased by following specific steps
such as using the window mode of oscilloscope and increase the resolution and fit the entire
image on the screen as much as possible.
(B
Trigger Po+ition 0 It is the position on the hori"ontal scale of the waveform where trigger occurs.
Trigger Le*el 0 It is theinstantaneous level that a trigger source signal must reach before a sweep
is initiated by the trigger circuit.
Slope 0 =ore commonly #nown as the gradient. It is used to give the nature of a wave.
@ow the trigger is shifted slowly from (' to 9. At about /.6(6.$ the signal behaves strangely
and starts moving bac# and forth. At about 6.$ the signal moves correctly again. The only reason
for this can be either because it changed between the triggered and non(triggered option ormaybe the varying amplitude in the signal.
At the following$ we can see the arrow is at the rising edge.
8/11/2019 Lab Report 1 Updated
24/28
SAGE OF OSCILLOSCOPE
At the following the arrow is pointing to a falling slope.
The differences between auto and normal mode(:
,irst of all$ the trigger level to stabili"e the signal is different in both of the modes. In the normal
mode$ the signal stops to be displayed if the signal drops to a low level whereas in auto mode$ it is
still present. @ormal mode is very easy to handle since the signal gets triggered as the trigger
enters a certain range but there is no signal if it is not triggered. Concluding$ in normal mode one
has to vary the trigger himself at each reading but in auto mode$ the oscilloscope does it for us.
&B
Pro
8/11/2019 Lab Report 1 Updated
25/28
SAGE OF OSCILLOSCOPE
At 'x(:
N
8/11/2019 Lab Report 1 Updated
26/28
SAGE OF OSCILLOSCOPE
41 FF
8/11/2019 Lab Report 1 Updated
27/28
SAGE OF OSCILLOSCOPE
('///('%Constant
Time Constant '%'///
Time Constant /.//'
71 In my opinion$ its best to get the time constant from the graph since the errors in the curve
average out when we ta#en the gradient. If we ta#e it directly from the oscilloscope it presents us
with an unreliable value since ta#ing the exact gradient of a curve at a certain point is difficult and
very error prone without the e-uation of the curve. >e could also of course directly calculate the
time constant as done in the part ' of this series0 evaluation but the tolerance of the capacitor and
the resistor affects the value.
CONCLUSION
The C*) is a device which can display a waveform and give new
perspective to it thereby increasing the scope of the
interpretation of a wave. It can be seen that using the cursors and
the measure function are much more accurate then using the
method of graticule divisions. 8econdly$ we should ta#e maximum
advantage of the units%div switch or sensitivity switch since it
allows a larger wave form which further allows more accurate
cursor placement on curves and more accurate readings.
Triggers are very useful for measuring the -uality of a signal orrather specific parts of the signal. ?owever they need to be
adjusted for different waves and modes. Care should be ta#en as
to use an appropriate form of mode.
+robes allow for greater accuracy since they compensate the
capacitance in the wires. A 'x probe is best for lower fre-uencies
and '/x for higher fre-uencies. +robes must be calibrated before
an experiment else they might produce errors and theattenuation be according to the magnitude of the fre-uency else
8/11/2019 Lab Report 1 Updated
28/28
SAGE OF OSCILLOSCOPE
there might be a significant error.
The last experiment involved calculating the time constant of a
capacitor most accurately by 4 ways but the logarithmic scale oneproved to be the most accurate since it cancels out the systematic
error. ,urthermore$ the curve on the oscilloscope is blurred and
not that accurate to draw a tangent on. The tolerance of the
resistor and capacitor add error to the direct formula method.
Therefore I concluded on this base that the logarithmic scale
method is the best.
RE.ERENCES
'1 Capacitor +icture
http:%%www.tvrepair#its.com%xcart%images%+%aluminumOelectrolyticOcapacitor.jpg
21 >ave Components &raph
http:%%www.electronics(tutorials.ws%accircuits%AC(waveform.html
41 ab =anual
61 )scilloscope