    Directions for Use

    Document No. 031972

    Revision 01

    October 2003

    LC/MS NitrogenGenerator

    MSQ10LA

  • 7/28/2019 LC MS Nitrogen Generator

    2/15

    MSQ10LA Nitrogen Generator

    Instructions for Use

    MSQ10LA issue 1 1

    2003 by Dionex CorporationAll rights reserved worldwide.

    Printed in the United States of America.

    This publication is protected by federal copyright law. No part of this publication may be

    copied or distributed, transmitted, transcribed, stored in a retrieval system, or transmitted into

    any human or computer language, in any form or by any means, electronic, mechanical,

    magnetic, manual, or otherwise, or disclosed to third parties without the express written

    permission of Dionex Corporation, 1228 Titan Way, Sunnyvale, California 94088-3603

    U.S.A.

    The MSQ10LA Nitrogen Generator is manufactured for Dionex by:

    Peak Scientific Instruments

    Fountain Crescent

    Inchinnan Business Park

    Inchinnan, Renfrewshire

    PA4 9RE

    Scotland, UK

    DISCLAIMER OF WARRANTY AND LIMITED WARRANTY

    THIS PUBLICATION IS PROVIDED AS IS WITHOUT WARRANTY OF ANY KIND.

    DIONEX CORPORATION DOES NOT WARRANT, GUARANTEE, OR MAKE ANY

    EXPRESS OR IMPLIED REPRESENTATIONS REGARDING THE USE, OR THE RESULTS

    OF THE USE, OF THIS PUBLICATION IN TERMS OF CORRECTNESS, ACCURACY,

    RELIABILITY, CURRENTNESS, OR OTHERWISE. FURTHER, DIONEX CORPORATION

    RESERVES THE RIGHT TO REVISE THIS PUBLICATION AND TO MAKE CHANGES

    FROM TIME TO TIME IN THE CONTENT HEREINOF WITHOUT OBLIGATION OF

    DIONEX CORPORATION TO NOTIFY ANY PERSON OR ORGANIZATION OF SUCH

    REVISION OR CHANGES.

    TRADEMARKS

    Molex is a registered trademark of Molex, Inc.

    Swagelok is a registered trademark of Swagelok Co.

    PRINTING HISTORY

    Revision 01, October 2003

    The products of Dionex Corporation are produced under ISO 9001 accredited quality management systems.

    Published by Technical Publications, Dionex Corporation, Sunnyvale, CA 94086.

  • 7/28/2019 LC MS Nitrogen Generator

    3/15

    MSQ10LA Nitrogen Generator

    Instructions for Use

    MSQ10LA issue 1 2

    CONTENTS

    1 INTRODUCTION........................................................... ........................................................... ............... 3

    2 UNPACKING AND INSTALLATION............................. ................................................................ ...... 3

    3 ELECTRICAL CONNECTION.................................. ................................................................ ............ 3

    4 AIR CONNECTION............................... ........................................................... ....................................... 4

    5 PRINCIPLE OF OPERATION................................................................ ............................................... 4

    6 COMMISSIONING...................................................... ........................................................... ................. 5

    7 ROUTINE INSPECTION .............................................................. .......................................................... 6

    8 TROUBLESHOOTING .......................................................... ............................................................ ..... 8

    9 TECHNICAL SPECIFICATIONS ............................................................ ........................................... 10

    10 PARTS LIST ...................................................... ............................................................... ...................... 11

    MAINTENANCE RECORD LOG .......................................................... ....................................................... 14

    IMPORTANT NOTICE TO USERS

    These instructions must be read thoroughly and understood before installation andoperation of your LC/MS Nitrogen Generator. Please follow the appropriate safetystandards for handling gases as determined by the laws and regulations of yourcountry.

    Please save the product packaging for storage or future shipment of the generator.

  • 7/28/2019 LC MS Nitrogen Generator

    4/15

    MSQ10LA Nitrogen Generator

    Instructions for Use

    MSQ10LA issue 1 3

    1 Introduction

    The MSQ10LA is a self-contained nitrogen generator designed specifically for use with Liquid

    Chromatography/Mass Spectrometry (LC/MS) analytical instruments as a source of carrier gas. The generator

    has been designed to operate totally independent of an external air supply and only requires an electrical

    supply to deliver high volume, high pressure, clean, dry, pure nitrogen.

    2 Unpacking and Installation

    Although every precaution has been taken for safe transit and packaging, it is advisable to fully inspect the

    unit for any sign of transit damage.

    ANY DAMAGE SHOULD BE REPORTED IMMEDIATELY TO THE CARRIER AND DIONEX.

    After unpacking and a visual inspection, the unit should be placed in a ventilated area away from direct

    sunlight. Care should be taken not to obstruct the ventilation holes on the sides of the unit or the fan outlet

    at the rear.

    The generator should be placed on a steady and level base. It has been designed to fit under most

    workbenches. Alternatively, the unit may be placed on the workbench adjacent to the machine or in any

    location convenient to the user. A guide for installation is included at the end of the document.

    Performance of the generator (as with any sophisticated equipment) is affected by ambient temperature.Continuous operation in ambient temperatures exceeding 25C will lead to a reduction in capacity and

    prolonged operation in temperatures exceeding 35C will shorten the life of the unit. Note should also be

    taken of the proximity of air conditioning outlets. These can sometimes give rise to pockets of air with

    high relative humidity. Operation of the generator within such a pocket could adversely affect its

    performance.

    3 Electrical Connection

    Important Electrical Notice

    This unit is classified asSAFETY CLASS 1

    equipment.THIS UNIT MUST BE EARTHED

    . Before

    connecting the unit to the mains supply, please check the information on the serial plate. The mains

    supply must be of the stated AC voltage and frequency.

    EARTH/GROUND (E): Green & YellowLIVE (L): BrownNeutral (N): Blue

    FuseThe generator protection fuse is in the pull-out drawer of the mains inlet IEC Euroconnector located on the

    bottom right-hand side of the cabinet, adjacent to the off/on switch. The fuse is rated at 230V 10%

    (50/60Hz ac). The drawer also holds a spare.

    Connect the generator to a single-phase supply using the power cord provided, or the power cord

    supplied with the equipment from Dionex for the country of use.

  • 7/28/2019 LC MS Nitrogen Generator

    5/15

    MSQ10LA Nitrogen Generator

    Instructions for Use

    MSQ10LA issue 1 4

    4 Air Connection

    The MSQ10LA generator has its own built-in air compressor and therefore does not require an external

    compressed air source.

    The generator has one bulkhead connection to the rear of the unit. This is supplied with a 6mm push-in

    fitting. This is the nitrogen outlet fitting and should be coupled to the application. There is no drain on this

    machine. Any moisture liberated by the filter-separator is discharged through the high capacity exhaust

    system, where the sudden reduction in pressure causes instant evaporation. The water vapor is safely

    removed from the unit by the ventilation system.

    5 Principle of OperationThe MSQ10LA Nitrogen Generator utilizes a pressure swing adsorption (PSA) method to extract pure

    nitrogen from air. This is where unwanted gases can be selectively adsorbed from compressed air into a

    porous carbon molecular sieve material (CMS). Dionex instruments utilize a unique single-column system

    where the column is alternately pressurized and vented under a finely tuned timing cycle. The rates of

    pressurization and venting are accurately set, which guarantees higher purity than can be achieved with a

    similarly sized traditional 2-column system.

    10

    14

    Nitrogen Generator

    Standard System

    Parts List

    1. Compressor

    14. Bulkhead Connection

    15. Internal Gauge Rail

    9. Nitrogen Storage Tank

    10. Pressure Regulator

    11. Internal Pressure Gauge

    13. Front Panel Gauge

    8.

    7.

    15.

    12

    116.

    9.

    13

    2. 3.

    7. CMS Column

    8. Non-return Valve

    2. Cooling Coil

    4. Filter / Separator

    5. Exhaust Valve

    6. Exhaust Silencer

    3. Cooling Fan (if fitted)

    16. 16.

    16. Compressor Cooling Fan (if fitted)

    1.

    5. 4.

    12. Flow Regulator

    Air is drawn into the system by the compressor (1) and passed via the cooling coil (3) andthe filter/separator (4) into the CMS column (7). Oxygen molecules in the air are trapped bythe sieve. However, the molecules of nitrogen pass straight through and are collected in the

    nitrogen storage tank (9). After a time interval, the compressor is stopped and the exhaustvalve (5) opens, allowing the sieve column to vent to atmosphere. The trapped oxygen isliberated and escapes to atmosphere via the exhaust valve (5) and the silencer (6). Thegenerated nitrogen in the storage tank is then regulated to the correct pressure and flow

    rate. After another time interval, the exhaust valve shuts and the compressor starts. Thiscycle runs continuously.

  • 7/28/2019 LC MS Nitrogen Generator

    6/15

    MSQ10LA Nitrogen Generator

    Instructions for Use

    MSQ10LA issue 1 5

    6 Commissioning

    With the generator installed as described earlier, connect power to the unit and turn it on. Disconnect the

    nitrogen outlet connection to allow the generator to vent to atmosphere until the unit is stabilized. At

    Switch On, the exhaust valve will open and the generator will commence its Venting Cycle. This is to allow

    venting of any residual pressure in order that the compressor does not start against pressure. The Vent

    Cycle may last up to 90 seconds. At the end of the vent cycle, the compressor should be heard to run

    and the normal operating cycle will begin. Pressure should begin to build, and the gauge on the front

    panel should indicate 75 psig after approximately 10 minutes.

    The generator has been pre-set at the factory to give 10 liters ATP per minute at 85 psig. Once the

    pressure in the nitrogen receiver exceeds that setting, the generator will stabilize and produce purenitrogen. Maximum purity will be achieved after around 45 minutes. At this time, the generator can be re-

    connected to the application.

    The generator is designed to deliver up to 10 liters of nitrogen at 85 psig with an oxygen content of 2% or

    less. Should the demand for nitrogen be less than 10 liters, or should the demand stop, the generator will

    continue to operate without any problems. The generator is protected from over-pressure and its normal

    operating cycle ensures frequent venting.

    Timer Setting

    The cycle timer has been set at the factory and should not require further adjustment. Adjusting the timer

    will affect the volume and purity of the delivered nitrogen; therefore, the timer should NEVER be adjusted

    without reference to the factory. The normal settings for this generator are 175 seconds ON, 85 seconds

    OFF. These settings may vary slightly between machines.

  • 7/28/2019 LC MS Nitrogen Generator

    7/15

    MSQ10LA Nitrogen Generator

    Instructions for Use

    MSQ10LA issue 1 6

    7 RoutineInspection

    Due to the simplicity of the design and the small number of moving parts, the MSQ10LA Nitrogen

    Generator will have a long and trouble-free life. However, as with all scientific and technical equipment,

    the generator should be regularly inspected by a competent person and the following points noted.

    Every 12 Months

    Note: All routine maintenance filter elements are contained in Dionex P/N 061492.

    Compressor Inlet Filter Element

    1. Switch the generatorOFFand disconnect it from the electrical supply. Remove the fuse fromthe Euro-socket to prevent the generator being inadvertently switched on by another person.

    2. Remove the 5-off X-head screws securing the lower front panel. Remove the panel to expose

    the compressor.

    3. The filter is located on top of the compressor and faces toward the rear, as shown below.

    The cover is removed by twisting it one-

    quarter turn and pulling clear. The elementsimply lifts out.

    Part Number 02-4187

    Reassembly is the reverse procedure.

    Inlet Filter

  • 7/28/2019 LC MS Nitrogen Generator

    8/15

    MSQ10LA Nitrogen Generator

    Instructions for Use

    MSQ10LA issue 1 7

    Coalescing Filter and Exhaust Silencer

    1. Switch the generatorOFFand disconnect it from the electrical supply. Remove the fuse from

    the Euro-socket to prevent the generator being inadvertently switched on by another person.

    2. Pull the generator clear of the bench (if located under) to a place where access is possible to

    all four sides.

    3. Remove the 8-off X-Head screws securing the lid. Remove the lid.

    4. Remove the coalescing filter bowl shown below by holding down the black slide and twisting

    the bowl one-quarter turn. The element simply unscrews.

    1. Fit new element and clean, and re-fit bowl.

    2. Unscrew and remove the green plastic cover of the exhaust silencer. Remove the 10mm nut

    and prize off the element.

    1. Fit new element and re-fit plastic cover.

    2. Replace all covers and switch on. Allow 4 hours to recover purity before reconnecting toapplication.

  • 7/28/2019 LC MS Nitrogen Generator

    9/15

    MSQ10LA Nitrogen Generator

    Instructions for Use

    MSQ10LA issue 1 8

    8 Troubleshooting

    The PSA method for nitrogen production is the most reliable of the popular methods. Many thousands of

    PSA systems are currently operating worldwide and have given many years of trouble-free operation.

    Provided the CMS material is protected from oils and oxygen enrichment, the purity of the nitrogen

    produced will remain consistent over the lifetime of the machine.

    Problems associated with the generator will be confined to the compressed air or control systems and will

    probably show up through loss of capacity. Reference should be made to the following Fault Finding

    Chart.

    Insuficient

    N2 Flow

    Compressor

    Cycling On/Off

    Sieve Vents on

    Exhuast Cycle

    Check Timer

    Operating

    Check Timer

    OperationFit New Timer

    Yes

    No

    Ok

    Yes

    No

    Ok

    Not

    Running

    Check PowerRepair Wiring

    Fault

    Ok

    No Power

    Check Power out

    to Solenoid ValveNo

    Yes

    Check Exhaust

    Valve

    Replace Faulty

    Valve

    Check

    Compressor Inlet

    Filter

    clean

    Dirty

    Fit New Filter

    Check Pressure

    @ Filter Bowl

    Check System

    for Leaks Repair Leaks

    Check Gauge

    Rail PressureAdjust Pressure

    < 100 psi

    >110 psiNo

    Yes

    Service

    Compressor

    >100 psi

    < 100 psi

    Check Timer

    Settings

  • 7/28/2019 LC MS Nitrogen Generator

    10/15

    MSQ10LA Nitrogen Generator

    Instructions for Use

    MSQ10LA issue 1 9

    Compressor Requirements

    The intake filter requires periodic replacement. Initial inspection should be after 500 hours operation.

    Most compressor-related problems can be prevented by keeping the filter clean. A dirty inlet filter will

    decrease compressor performance and may shorten its life.

    The compressor head surfaces and motor casing become very HOT during operation. Do not

    touch these parts until the unit has been switched off and allowed to cool.

    Switch the generator OFFand disconnect it from the electrical supply. Remove the fuse from the

    Euro-socket to prevent the generator being inadvertently switched on by another person.

    This is a non-lubricated compressor and should never be oiled. Oiling this compressor will cause

    damage.

    Compressor Replacement

    Remove the generator to a clean dry place with sufficient room to work.

    1. Remove the lower front panel.

    2. Disconnect the braided hose from cone fitting where it is secured to the right-hand side panel.Take care not to disturb the fit between the cone fitting and the bracket assembly.

    3. Disconnect the Molexconnector at the front of the compressor.

    4. Tip the generator onto its back and remove the 4-off nuts securing the compressor to thebase. Support the weight of the compressor with one hand while removing the last nut.

    5. Lift the compressor clear of the generator.

    6. Remove and clean all the fittings from the old compressor and reassemble onto the newmachine using PTEF tape and sealing compound. (Swagelok Teflon Free Pt No MS-TFS-

    50) is recommended.)

    Reassembly is the reverse procedure.

  • 7/28/2019 LC MS Nitrogen Generator

    11/15

    MSQ10LA Nitrogen Generator

    Instructions for Use

    MSQ10LA issue 1 10

    9 Technical Specifications

    General Details

    Minimum Operating Ambient Temperature 5oC (41

    oF)

    Maximum Operating Ambient Temperature 30oC (86

    oF)

    Inlet Conditions(Free of oil and bulk moisture)

    Minimum Air Inlet Pressure N/A

    Maximum Air Inlet Pressure N/A

    Minimum Air Inlet Flow Rate N/A

    Outlet Gas

    Maximum Pressure Drop (Outlet-Inlet) P N/A

    Maximum Gas Outlet Pressure 85 psig

    Maximum Gas Outlet Flow for Specified Purity 10 Liters/min (ATP)

    Pressure Dew point -75oC (-103

    oF) (1.4ppm @ ATP)

    Start up time for Purity 45 minutesParticles 0.01um

    O2 level (maximum) 2.0%

    Electrical Requirements

    @230V ac 10% (50/60Hz ac) 4.2 Amps

    Fuse (220V 50/60Hz ac) 10 Amps

    Electrical Connection IEC Euroconnector

    Plug Type NEMA Code N6/15

    General

    Dimensions H x W x D cm

    (inches)

    63 x 43 x 41

    (24.6 x 17 x 16)

    Weight Kg

    (lbs)

    47.5

    (104.5)

    Shipping Weight Kg

    (lbs)

    69

    (152)

  • 7/28/2019 LC MS Nitrogen Generator

    12/15

    MSQ10LA Nitrogen Generator

    Instructions for Use

    MSQ10LA issue 1 11

    10 Parts List

    Recommended Spares

    MSQ10LA Service Spares

    The following parts are in Dionex P/N 061492

    Item Description Interval Part No. No. Off

    1 Compressor Inlet Filter Element 12 Months 02-4187 1

    2 Coalescing Filter Element 12 Months 02-4509 1

    3 Exhaust Element 12 Months 02-4336 1

    MSQ10LA Maintenance Spares

    Item Description Part No. No. Off

    1 Braided Hose 02-1100 1

    2 Internal Gauge 02-1021 1

    3 Exhaust Valve 02-4289 1

    4 Non-Return Valve 02-1135 1

    5 Timer 04-1019 1

    6 Contactor 04-4439 1

    7 Cooling / Ventilation Fan 04-1021 2

    8 Flow Controller 02-4209 1

    9 Pressure Regulator 02-1110 1

    10 Panel Gauge 02-1020 2

    11 Safety Valve 02-4544 112 Compressor Anti-vibe Mounts 06-1020 4

    13 Compressor 06-5513 1

    14 Back Pressure Relief Valve 02-1213 1

    Additional parts can be ordered from:

    Peak Scientific InstrumentsFountain CrescentInchinnan Business Park

    Inchinnan, RenfrewshirePA4 9REScotland, UK

    Contact information:

    +44-141-812-8100+44-141-812-8200 fax

    [email protected]

  • 7/28/2019 LC MS Nitrogen Generator

    13/15

    MSQ10LA Nitrogen Generator

    Instructions for Use

    MSQ10LA issue 1 12

    14

    15.

    10 11

    12

    16.

    1.

    9.

    6. 5.

    13

    7.

    8.

    11. Internal Pressure Gauge

    Nitrogen GeneratorStandard System

    14. Bulkhead Connection

    15. Internal Gauge Rail

    13. Front Panel Gauge

    16. Compressor Cooling Fan (if fitted)

    12. Flow Regulator

    16.

    4.

    Parts List

    1. Compressor

    9. Nitrogen Storage Tank

    10. Pressure Regulator

    7. CMS Column

    8. Non-return Valve

    2. Cooling Coil

    4. Filter / Separator

    5. Exhaust Valve

    6. Exhaust Silencer

    3. Cooling Fan (if fitted)

    2. 3.

    NL ESwitch

    Fuse

    Fan Motor s

    MSQ10LA Electrical Dia ram

    Terminal Rail

    Com ressor

    Timer

    Valve

    Exhaust

  • 7/28/2019 LC MS Nitrogen Generator

    14/15

    MSQ10LA Nitrogen Generator

    Instructions for Use

    MSQ10LA issue 1 13

    80.00 mm

    800 mm

    PowerOutletPosition

    (MaximumD

    istanceLHS)

    PowerOutletPosition

    (MaximumD

    istanceRHS)

    RearVentilationGap

    Clearan

    ceforMaintenance

    875 mm

    900 mm

    Notes

    1.Accessisrequiredthrough

    FrontPanels&Lidfor

    maintenance.

    2.PlugPowerCordintoacon

    venientoutletwithsufficient

    slacktoallow

    theunittobe

    pulledforwarforaccess.

    3.Allow

    sufficientslackinthe

    GasLinetoallow

    the

    generatortobepulledforwardformaintenance

    MSQ10LA

    InstallationNotes

    NitrogenPressure

    110 mm

    625 mm

    IO

    410 mm

    GaprequiredforPowerCord

  • 7/28/2019 LC MS Nitrogen Generator

    15/15

    MSQ10LA Nitrogen Generator

    Instructions for Use

    11 Maintenance Record Log

    Model- MSQ10LA Serial Number:

    WorkDone

    Remarks Date Name