Leica SP8 DIC Imaging

Kim Peifley

02/10/17

Page 2: Leica SP8 DIC Imaging - National Cancer Institute · Under Transmitted select DIC. 3. Select TL-Shutter. Gray circle indicates off, yellow circle indicates on. 2 1 3 3. 4. If you

With focus and fluorophore settings already established:

On the microscope touch screen:

1. Select Color Contrast window.

2. Under Transmitted select DIC.

3. Select TL-Shutter. Gray circle indicates off, yellow circle indicates on.

2

1

3 3

Page 3: Leica SP8 DIC Imaging - National Cancer Institute · Under Transmitted select DIC. 3. Select TL-Shutter. Gray circle indicates off, yellow circle indicates on. 2 1 3 3. 4. If you

4. If you need to adjust the intensity of the light – select Microscope

menu on the touch screen and then adjust Intensity.

Page 4: Leica SP8 DIC Imaging - National Cancer Institute · Under Transmitted select DIC. 3. Select TL-Shutter. Gray circle indicates off, yellow circle indicates on. 2 1 3 3. 4. If you

6

6

5. Adjust the focus.

6. Adjust the condenser.

Page 5: Leica SP8 DIC Imaging - National Cancer Institute · Under Transmitted select DIC. 3. Select TL-Shutter. Gray circle indicates off, yellow circle indicates on. 2 1 3 3. 4. If you

7. Start closing the field diaphragm.

8. Adjust the condenser as you do this.

9. Repeat until you can start to see the octagon shape and it

is in focus.

7

8

Page 6: Leica SP8 DIC Imaging - National Cancer Institute · Under Transmitted select DIC. 3. Select TL-Shutter. Gray circle indicates off, yellow circle indicates on. 2 1 3 3. 4. If you

11

10. Removed the pins from their holder on the back of the condenser near the

microscope arm.

11. Put in place and center the field diaphragm.

12. You may need to adjust the condenser [6] and close the field diaphragm [7]

more as you move field diaphragm to center.

10

Page 7: Leica SP8 DIC Imaging - National Cancer Institute · Under Transmitted select DIC. 3. Select TL-Shutter. Gray circle indicates off, yellow circle indicates on. 2 1 3 3. 4. If you

13. Make sure Brightfield/TLD lens is in the “out” position.

Not This.Keep cover on.

This

Page 8: Leica SP8 DIC Imaging - National Cancer Institute · Under Transmitted select DIC. 3. Select TL-Shutter. Gray circle indicates off, yellow circle indicates on. 2 1 3 3. 4. If you

14. Select the sequence you wish to attach DIC.

15. Click the arrow next to TLD to open the

transmitted window.

14

15

Page 9: Leica SP8 DIC Imaging - National Cancer Institute · Under Transmitted select DIC. 3. Select TL-Shutter. Gray circle indicates off, yellow circle indicates on. 2 1 3 3. 4. If you

16

16

16. Click on the OFF button to

turn the transmitted light ON.

Page 10: Leica SP8 DIC Imaging - National Cancer Institute · Under Transmitted select DIC. 3. Select TL-Shutter. Gray circle indicates off, yellow circle indicates on. 2 1 3 3. 4. If you

17. In the Fluo Turret menu select Scan-DIC.

18. Click Live to start scanning.

18

17

17

Page 11: Leica SP8 DIC Imaging - National Cancer Institute · Under Transmitted select DIC. 3. Select TL-Shutter. Gray circle indicates off, yellow circle indicates on. 2 1 3 3. 4. If you

19. Click on the DIC window to select it.

20. Adjust the gain via the control panel or software.

20

1919

20

Page 12: Leica SP8 DIC Imaging - National Cancer Institute · Under Transmitted select DIC. 3. Select TL-Shutter. Gray circle indicates off, yellow circle indicates on. 2 1 3 3. 4. If you

21. Adjust the sheer either from the software in the Microscope DIC

window [A] or from the dial underneath the objective turret [B].

A

B

Page 13: Leica SP8 DIC Imaging - National Cancer Institute · Under Transmitted select DIC. 3. Select TL-Shutter. Gray circle indicates off, yellow circle indicates on. 2 1 3 3. 4. If you

This is an image after sheer adjustment.

Page 14: Leica SP8 DIC Imaging - National Cancer Institute · Under Transmitted select DIC. 3. Select TL-Shutter. Gray circle indicates off, yellow circle indicates on. 2 1 3 3. 4. If you

22. Adjust DIC contrast.