D. Ugolini, LSC March 2006 1 LIGO-G060061-00-Z Developing a Kelvin Probe for Measuring Surface Charge on LIGO Optics Dennis Ugolini, Robert McKinney Trinity University March LSC Meeting March 22, 2006

LIGO-G060061-00-Z D. Ugolini, LSC March 2006 1 Developing a Kelvin Probe for Measuring Surface Charge on LIGO Optics Dennis Ugolini, Robert McKinney Trinity

Embed Size (px)

Citation preview

Page 1: LIGO-G060061-00-Z D. Ugolini, LSC March 2006 1 Developing a Kelvin Probe for Measuring Surface Charge on LIGO Optics Dennis Ugolini, Robert McKinney Trinity

D. Ugolini, LSC March 2006 1LIGO-G060061-00-Z

Developing a Kelvin Probe for MeasuringSurface Charge on LIGO Optics

Dennis Ugolini, Robert McKinneyTrinity University

March LSC MeetingMarch 22, 2006

Page 2: LIGO-G060061-00-Z D. Ugolini, LSC March 2006 1 Developing a Kelvin Probe for Measuring Surface Charge on LIGO Optics Dennis Ugolini, Robert McKinney Trinity

D. Ugolini, LSC March 2006 2LIGO-G060061-00-Z

The Surface Charge Problem

Surface charge may build up on test masses

Sources of charge buildup» Friction with dust molecules during pumpdown

» Mechanical contact with conductors (earthquake stops)

» Cosmic rays (Moscow State group)

Potential concerns» Fluctuating electric fields can interfere with positioning control

» Source of low-frequency suspension noise

» Dust may be attracted/held to optic surface, increasing absorption

Page 3: LIGO-G060061-00-Z D. Ugolini, LSC March 2006 1 Developing a Kelvin Probe for Measuring Surface Charge on LIGO Optics Dennis Ugolini, Robert McKinney Trinity

D. Ugolini, LSC March 2006 3LIGO-G060061-00-Z

The correlation time for charge mobility, which affects force as:

The effectiveness of charge reduction techniques» Slightly conducting ionic coating» Shield test mass with conductors to terminate electric fields» UV light (next talk)

What We Know/Don’t Know

Optics experience drifts of ~105 e-/cm2/month» Mitrofanov et al., Phys. Lett. A300, 370 (2002).

Negligible effects on mechanical Q» Mortonson et al., Rev. Sci. Inst. 74, 4840 (2003).

What we know:

What we don’t know:

(R. Weiss, LIGO-T960137-00-E)

Page 4: LIGO-G060061-00-Z D. Ugolini, LSC March 2006 1 Developing a Kelvin Probe for Measuring Surface Charge on LIGO Optics Dennis Ugolini, Robert McKinney Trinity

D. Ugolini, LSC March 2006 4LIGO-G060061-00-Z

The Kelvin Probe

The Kelvin probe measures the contact potential difference between the probe and sample

Commercial probes modulate the difference by vibrating the probe tip by PZT or voice coil -- expensive

Instead modulate difference with optical “tuning-fork” chopper

Page 5: LIGO-G060061-00-Z D. Ugolini, LSC March 2006 1 Developing a Kelvin Probe for Measuring Surface Charge on LIGO Optics Dennis Ugolini, Robert McKinney Trinity

D. Ugolini, LSC March 2006 5LIGO-G060061-00-Z

Experimental Setup

samplechopper

probe

adjustable power supply

signal (to amp/filter or spectrum analyzer)

Calibrate with conducting sample Charge insulator, measure change

over time Repeat calibration in vacuum Use XY stage to scan sample,

measure charge vs. position

Probe system is then available as testbed for optical material, charge reduction techniques

Page 6: LIGO-G060061-00-Z D. Ugolini, LSC March 2006 1 Developing a Kelvin Probe for Measuring Surface Charge on LIGO Optics Dennis Ugolini, Robert McKinney Trinity

D. Ugolini, LSC March 2006 6LIGO-G060061-00-Z

Kelvin Probe Version 1

Plexiglass spacer

Grounded ring

Plexiglass backing

Copper tip4 mm

2.5 cm

Grounding ring too large, coil on chopper prevented the probe from getting closer than 1 cm.

Page 7: LIGO-G060061-00-Z D. Ugolini, LSC March 2006 1 Developing a Kelvin Probe for Measuring Surface Charge on LIGO Optics Dennis Ugolini, Robert McKinney Trinity

D. Ugolini, LSC March 2006 7LIGO-G060061-00-Z

Kelvin Probe Version 2

“Tongue” shape allows probe tip to get closer.

Coil on chopper emits large 500 Hz signal, picked up by probe without grounding ring.

Page 8: LIGO-G060061-00-Z D. Ugolini, LSC March 2006 1 Developing a Kelvin Probe for Measuring Surface Charge on LIGO Optics Dennis Ugolini, Robert McKinney Trinity

D. Ugolini, LSC March 2006 8LIGO-G060061-00-Z

Kelvin Probe Version 3

Grounding ring

Bare wire tipChopper coil noise reduced by:

Adding grounding ring

Isolating chopper power supply

Wrapping grounded aluminum foil around probe leads

Encasing chopper in grounded aluminum box

Noise peak reduced to 0.1 mV

Page 9: LIGO-G060061-00-Z D. Ugolini, LSC March 2006 1 Developing a Kelvin Probe for Measuring Surface Charge on LIGO Optics Dennis Ugolini, Robert McKinney Trinity

D. Ugolini, LSC March 2006 9LIGO-G060061-00-Z

Observations

Cannot see signal from conducting sample, with voltage varied from 0 – 15V.

Plexiglass sample rubbed with felt/wig/Allen wrench gives 1-3 mV signal (10-30 times larger than noise).

Signal shows exponential decay, rate varies greatly from day to day.

Page 10: LIGO-G060061-00-Z D. Ugolini, LSC March 2006 1 Developing a Kelvin Probe for Measuring Surface Charge on LIGO Optics Dennis Ugolini, Robert McKinney Trinity

D. Ugolini, LSC March 2006 10LIGO-G060061-00-Z

Charge Decay Over Time

Charge decay over time

0

0.2

0.4

0.6

0.8

1

1.2

1.4

0 5 10 15 20 25 30

Time (minutes)

Pro

be s

ign

al (m

V)

3/9/06

3/17/06

Page 11: LIGO-G060061-00-Z D. Ugolini, LSC March 2006 1 Developing a Kelvin Probe for Measuring Surface Charge on LIGO Optics Dennis Ugolini, Robert McKinney Trinity

D. Ugolini, LSC March 2006 11LIGO-G060061-00-Z

Further Improvements

Use electrostatic voltmeter to check insulating sample potential, calibrate

Improved shielding, grounding to reduce noise peak

Better chopper (drive freq. different than signal freq. – rotary?)

Place in vacuum