Literature Review of Monitoring Methods for Formaldehyde

  • Upload
    ypzo

  • View
    220

  • Download
    0

Embed Size (px)

Citation preview

  • 7/26/2019 Literature Review of Monitoring Methods for Formaldehyde

    1/26

    NATIONAL UNIVERSITY OF SINGAPORE

    SH5110 Chemical Hazard Evaluation

    Literature Review ofMonitoring Methods for

    Formaldehyde

  • 7/26/2019 Literature Review of Monitoring Methods for Formaldehyde

    2/26

    2 of 26

    Executive Summary

    Formaldehyde is a known human carcinogen, and yet there is the potential for this chemical

    to be present at the workplace, retail spaces and even residential spaces due to its applications

    in many building materials. The aim of this literature review is to improve our understandingof the hazards of formaldehyde, how exposure to formaldehyde is being monitored and

    regulated at the workplace and also to review the application of new technologies in

    measuring airborne formaldehyde concentration.

    The TWA and STEL for formaldehyde is at 0.75 ppm (8 hours) and 2 ppm (15 mins)

    respectively. NEA recommends a maximum indoor air concentration of 0.1 ppm. Current

    standardized methods for compliance measures airborne formaldehyde concentration and

    neglects particulate-bound formaldehyde, which is a common occurrence in wood and textile

    industries. The applicability and limitations of conventional laboratory methods and direct

    reading instruments are discussed. Research into novel methods, which have severaladvantages over the traditional methods for monitoring formaldehyde concentration in the

    field, has also been reviewed.

  • 7/26/2019 Literature Review of Monitoring Methods for Formaldehyde

    3/26

    3 of 26

    Contents

    Executive Summary ................................................................................................................................. 2

    1.0 Introduction ................................................................................................................................ 4

    1.1 Industrial applications ............................................................................................................. 4

    1.2 Synthesis of Formaldehyde ..................................................................................................... 4

    1.3 Safety ...................................................................................................................................... 5

    2.0 Standardized Methods ................................................................................................................ 7

    2.1 NIOSH 3500 ............................................................................................................................. 8

    2.1.1 NIOSH Method 3500 for Formaldehyde Measurement ................................................. 8

    2.1.2 Method Evaluation of NIOSH Method 3500 ................................................................... 9

    2.2 OSHA 1007 ............................................................................................................................ 15

    2.3 NIOSH 5700 ........................................................................................................................... 17

    2.4 Other Methods ...................................................................................................................... 18

    2.4.1 Direct Reading Instruments .......................................................................................... 18

    2.4.2 Emission testing standards ........................................................................................... 18

    2.5 Novel Methods ...................................................................................................................... 21

    2.5.1 Biosensors ..................................................................................................................... 21

    2.5.2 DNAzymes ..................................................................................................................... 22

    2.5.3 Formaldehyde dehydrogenase ..................................................................................... 23

    3.0 Conclusion ................................................................................................................................. 25

    4.0 References ................................................................................................................................ 26

    Abbreviations

    CARBCalifornia Air Resources Board

    DMCDynamic Micro Chamber

    FIDFlame Ionization Detector

    FLECField and Laboratory Emission Cell

    GCGas Chromatography

    HPLCHigh Performance Liquid Chromatography

    MSMass Spectrometry

    NPDNitrogen-specific detector

    UVUltraviolet Detector

    VASVisible Absorption Spectrophotometry

  • 7/26/2019 Literature Review of Monitoring Methods for Formaldehyde

    4/26

    4 of 26

    1.0 Introduction

    Formaldehyde1 is an important precursor to many other materials and chemical compounds.

    In 1996, the installed capacity for the production of formaldehyde was estimated to be 8.7

    million tons per year. It is mainly used in the production of industrial resins, e.g., for particleboard and coatings. It is also used in pressed-wood products, such as particleboard, plywood,

    and fibreboard; glues and adhesives; permanent-press fabrics; paper product coatings; and

    certain insulation materials. In addition, formaldehyde is commonly used as an

    industrialfungicide,germicide, anddisinfectant, and as a preservative in mortuaries and

    medical laboratories. Formaldehyde also occurs naturally in the environment. It is produced

    in small amounts by most living organisms as part of normalmetabolicprocesses.

    In view of its widespread use, toxicity, and volatility, formaldehyde is a significant

    consideration for human health.In 2011, the US National Toxicology Programdescribed

    formaldehyde as "known to be a human carcinogen".

    1.1

    Industrial applications

    Formaldehyde is a common precursor to more complex compounds and materials. In

    approximate order of decreasing consumption, products generated from formaldehyde

    include urea formaldehyde resin, melamine resin, phenol formaldehyde resin,

    polyoxmethylene plastics, 1,4-butanediol and methylene diphenyl diisocyanate.

    The textile industry uses formaldehyde-based resins as finishers to make fabrics crease-

    resistant. Formaldehyde-based materials are key to the manufacture of automobiles, and usedto make components for the transmission, electrical system, engine block, door panels, axles

    and brake shoes. The value of sales of formaldehyde and derivative products was over $145

    billion in 2003.

    1.2

    Synthesis of Formaldehyde

    Two steps in formation of urea-formaldehyde resin, which is widely used in the production of

    particle board.

    1Formaldehyde, World health Organization, 2002

    http://www.cancer.gov/Common/PopUps/popDefinition.aspx?id=CDR0000642499&version=Patient&language=Englishhttp://www.cancer.gov/Common/PopUps/popDefinition.aspx?id=CDR0000642503&version=Patient&language=Englishhttp://www.cancer.gov/Common/PopUps/popDefinition.aspx?id=CDR0000642495&version=Patient&language=Englishhttp://www.cancer.gov/Common/PopUps/popDefinition.aspx?id=CDR0000044056&version=Patient&language=Englishhttps://en.wikipedia.org/wiki/File:UFresinSyn.svghttp://www.cancer.gov/Common/PopUps/popDefinition.aspx?id=CDR0000044056&version=Patient&language=Englishhttp://www.cancer.gov/Common/PopUps/popDefinition.aspx?id=CDR0000642495&version=Patient&language=Englishhttp://www.cancer.gov/Common/PopUps/popDefinition.aspx?id=CDR0000642503&version=Patient&language=Englishhttp://www.cancer.gov/Common/PopUps/popDefinition.aspx?id=CDR0000642499&version=Patient&language=English
  • 7/26/2019 Literature Review of Monitoring Methods for Formaldehyde

    5/26

    5 of 26

    When treated with phenol, urea, or melamine, formaldehyde produces, respectively, hard

    thermoset phenol formaldehyde resin, urea formaldehyde resin, and melamine resin. These

    polymers are common permanent adhesives used in plywood and carpeting. It is used as

    the wet-strength resin added to sanitary paper products such as (listed in increasing

    concentrations injected into the paper machine headstock chest) facial tissue, table napkins,and roll towels. They are also foamed to make insulation, or cast into moulded products.

    Production of formaldehyde resins accounts for more than half of formaldehyde consumption.

    Formaldehyde is also a precursor to polyfunctional alcohols such as pentaerythritol, which is

    used to make paints and explosives. Other formaldehyde derivatives include methylene

    diphenyl diisocyanate, an important component in polyurethane paints and foams,

    and hexamine, which is used in phenol-formaldehyde resins as well as the explosive RDX.

    Formaldehyde has been found as a contaminant in several bath products, at levels from 54

    610 ppm: it is thought to arise from the breakdown of preservatives in the products, most

    frequently diazolidinyl urea. Since 2006, formaldehyde (methylene glycol) is also used inhair smoothing treatments in order to straighten wavy/curly hair and make hair less prone to

    frizz under high humid weather. OSHA Oregon has reported these treatments as unsafe for

    human health.

    1.3 Safety

    Formaldehyde is highly toxic to all animals, regardless of method of intake. Ingestion of 30

    mL of a solution containing 37% formaldehyde has been reported to cause deathin an adult

    human.Water solution of formaldehyde is very corrosive and its ingestion can cause severe

    injury to theupper gastrointestinal tract.

    Occupational exposure to formaldehyde by inhalation is mainly from three types of

    sources:thermalorchemical decompositionof formaldehyde-based resins, formaldehyde

    emission fromaqueoussolutions (for example, embalming fluids), and the production of

    formaldehyde resulting from thecombustionof a variety of organic compounds (for example,

    exhaust gases). Formaldehyde can be toxic, allergenic, and carcinogenic.Because

    formaldehyde resins are used in many construction materials it is one of the more common

    indoor air pollutants.At concentrations above 0.1 ppm in air formaldehyde can irritate the

    eyes andmucous membranes, resulting in watery eyes.Formaldehyde inhaled at this

    concentration may cause headaches, a burning sensation in the throat, and difficulty breathing,

    and can trigger or aggravate asthma symptoms.

    A 1988 Canadian study of houses withurea-formaldehydefoam insulation found that

    formaldehyde levels as low as 0.046 ppm was positively correlated with eye and nasal

    irritation.A recent review of studies has shown a strong association between exposure to

    formaldehyde and the development of childhood asthma.The primary exposure concern is

    for the workers in the industries producing or using formaldehyde.

    The formaldehyde theory of carcinogenesis was proposed in 1978.In 1987 the U.S. EPA

    classified it as aprobable human carcinogen, and after more studies theWHO International

  • 7/26/2019 Literature Review of Monitoring Methods for Formaldehyde

    6/26

    6 of 26

    Agency for Research on Cancer (IARC) in 1995 also classified it as aprobable human

    carcinogen. Further information and evaluation of all known data led the IARC to reclassify

    formaldehyde as aknown human carcinogenassociated with nasal sinus cancer

    andnasopharyngeal cancer.Recent studies have also shown a positive correlation between

    exposure to formaldehyde and the development ofleukaemia, particularlymyeloidleukaemia.Nasopharyngeal and sinonasal cancers are relatively rare, with a combined annual

    incidence in the United States of < 4,000 cases.About 25,000 cases of myeloid leukaemia

    occur in the United States each year.Workplace exposure to inhaled chemicals is among the

    most important risk factors for sinonasal cancers.Professionals exposed to formaldehyde in

    their occupation, such as funeral industry workers andembalmers, showed an increased risk

    of leukaemia and brain cancer compared with the general population. Other factors are

    important in determining individual risk for the development of leukaemia or nasopharyngeal

    cancer.

    In the residential environment, formaldehyde exposure comes from a number of differentroutes; formaldehyde can off-gas from wood products, such as plywood or particle board, but

    it is produced by paints, varnishes, floor finishes, andcigarettesmoking as well.

    Singapore National Environment Agency recommends that the maximum threshold level for

    formaldehyde should not exceed 0.1ppm, based on Guidelines for Good Indoor Air Quality

    in Office Premises.2

    The purpose of this case study is to review the different methods to determine the

    concentration of formaldehyde.

    2IAQ, Indoor Air quality website url: http://www.iaqsg.com/chemical-

    parameters/formaldehyde/ , accessed on 04/04/2016

  • 7/26/2019 Literature Review of Monitoring Methods for Formaldehyde

    7/26

    7 of 26

    2.0 Standardized Methods

    Several standards have been developed to provide guidelines for the sampling and analytical

    methods for determining the concentration of formaldehyde in air. The majority of the

    methods use a sorbent tube to collect formaldehyde in an active sampling train. Somemethods consider the use of a diffusive sampler to collect formaldehyde in the air passively;

    one other method makes use of an impinger to collect formaldehyde in a liquid solution3.

    High performance liquid chromatography (HPLC) using an ultraviolet detector is the

    analytical method of choice for most of the standards. Other standards recommend gas

    chromatography (GC) and visible absorption spectrophotometry (VAS) for the analysis of the

    collected samples.

    In addition to sampling formaldehyde direct from air, NIOSH has documented a method for

    sampling formaldehyde-containing particulates. These particulates are a common occurrencein the wood and textile industries. Legislations has focused on the gas-phase concentration of

    formaldehyde based on established interpretation of results from epidemiological studies.

    The concentration of formaldehyde-containing particulates at the workplace is not regulated

    by either OSHA or the Singapore authorities4.

    The list of approved methods for compliance sampling is presented in Table 1. The NIOSH

    3500, OSHA 1007 and NIOSH 5700 methods are selected for further discussion below.

    Table 1: Approved methods for compliance sampling of formaldehyde in air

    Agency Reference Sampler AnalyticalMethod

    Applications

    NIOSH 2016 Sorbent Tube HPLC-UV

    2539 Sorbent Tube GC-FIDGC-MS

    For screening only

    2541 Sorbent Tube GC-FID

    3500 Glass Midget Impinger VAS

    5700 IOM ParticulateSampler

    HPLC-UV Textile dust or wood dust

    OSHA 52 Sorbent Tube GC-NPD Use of formalin solutions

    1007 Diffusive Sampler HPLC-UVID-205 Diffusive Sampler VAS Not suitable for STELsampling

    EPA IP-6A Sorbent Tube HPLC-UV Environmental sampling(indoor air)

    IP-6C Diffusive Sampler HPLC-UV Environmental sampling(indoor air)

    IP-11A Sorbent Tube HPLC-UV Environmental sampling(ambient / outdoor air)

    ASTM D 5197 Sorbent Tube HPLC-UV

    3NIOSH, NOISH Manual of Analytical Methods, Fourth Ed., 19944NEA, National Environment Agency website, url:www.nea.gov.sg,date retrieved: 29th March 2016

    http://www.nea.gov.sg/http://www.nea.gov.sg/http://www.nea.gov.sg/http://www.nea.gov.sg/
  • 7/26/2019 Literature Review of Monitoring Methods for Formaldehyde

    8/26

    8 of 26

    2.1

    NIOSH 3500

    2.1.1 NIOSH Method 3500 for Formaldehyde Measurement5

    The NIOSH method 3500 is an active sampling approach which uses a sampler attached to a

    pump. After collection of the sample, it is analysed through a visible absorption spectrometry

    of wavelength 580nm.

    Sampler:

    The sampler consist of a filter with a PTFE membrane filter (1-3m pore size) supported by

    an O-ring followed by 2 midget impingers (Fig 1). During sampling, the two impingers are

    filled with 20ml, 1% sodium bisulfite solution. A cassette is attached to the impinger and

    impinger to the sampling pump via a flexible inert tubing. The PTFE membrane is necessary

    if the sampling is used in a dusty environment. The use of dual impingers in series is

    recommended to ensure efficient collection of formaldehyde. After sampling, the contents of

    the impingers will be transferred to a polyethylene bottle for shipping.

    Figure 1: Midget Impinger Figure 2: Midget Impinger with Sampling Flow Pump

    Sample Preparation

    When the impingers solution are brought back, the volume of the solution from the front and

    back up impinger, Vf and Vb are recorded. 4mL of pipetted sampling solutions is transferred

    to a 25mL glass stoppered flask. A 0.1mL 1% chromotropic acid can react with 40

    micrograms of formaldehyde. Addition of 6mL of concentrated sulphuric acid is performed

    slowly and a gentle swirl to mix. A colour develops as illustrated (Fig 3). The sample is then

    5Kennedy, E. R. (1994). FORMALDEHYDE: METHOD 3500, Issue 2. NIOSH Manual of Analytical Methods

    (NMAM), Fourth Edition, 8/15/94. Retrieved April 12, 2016, fromhttp://www.cdc.gov/niosh/docs/2003-154/pdfs/3500.pdf

    http://www.cdc.gov/niosh/docs/2003-154/pdfs/3500.pdfhttp://www.cdc.gov/niosh/docs/2003-154/pdfs/3500.pdfhttp://www.cdc.gov/niosh/docs/2003-154/pdfs/3500.pdfhttp://www.cdc.gov/niosh/docs/2003-154/pdfs/3500.pdfhttp://www.cdc.gov/niosh/docs/2003-154/pdfs/3500.pdfhttp://www.cdc.gov/niosh/docs/2003-154/pdfs/3500.pdf
  • 7/26/2019 Literature Review of Monitoring Methods for Formaldehyde

    9/26

    9 of 26

    placed in a 1cm-cuvette into the spectrophotometer. The reading is recorded at visible

    absorption spectrometry at 580nm. Effect of other aldehydes is minimal on this method.

    Figure 3: Addition of Chromotropic Acid into Sample solution in Glass Stoppered Flask

    Applicability: The working range is 0.02 to 4ppm for an 80L air sample. This is the most

    sensitive method and is able to measure ceiling levels as low as 0.1ppm.

    Inteferences: Oxidisable organic materials may give a positive interference.

    Accuracy, bias and precision of Results

    Method Bias should be less than 10% as required by the NIOSH Guidelines. For the NIOSH

    Method 3500, there are none identified. The precision expected is 0.09

    The accuracy which is determined by the intersections of the bias and precision estimates on

    the parabolic grid (nomogram) is 18%

    2.1.2

    Method Evaluation of NIOSH Method 3500

    NIOSH Guidelines for Air Sampling and Analytical Method Development and

    Evaluation6

    The NIOSH Guide suggest guidelines for the development and evaluation of sampling and

    analytical methods for industrial hygiene monitoring For each method under consideration,

    the objective of this protocol was to decide if the method would provide results, on the

    average, over a concentration range of 0.1 to 2 times the exposure limit, to be within 25% of

    the true concentration with a probability of 0.95 for an individual observation.

    6Kennedy, E. R., Fischbach, T. J., Song, R., Eller, P. M., & Shulman, S. A. (1996). Summary of the NIOSH

    Guidelines for Air Sampling and Analytical Method Development and Evaluation. Analyst, 121. RetrievedApril 4, 2016, fromhttp://www.ncbi.nlm.nih.gov/pubmed/8831274

    http://www.ncbi.nlm.nih.gov/pubmed/8831274http://www.ncbi.nlm.nih.gov/pubmed/8831274http://www.ncbi.nlm.nih.gov/pubmed/8831274http://www.ncbi.nlm.nih.gov/pubmed/8831274
  • 7/26/2019 Literature Review of Monitoring Methods for Formaldehyde

    10/26

    10 of 26

    The sampling of a generated atmosphere is needed to more adequately assess the

    performance of a method. The concentration range of the analyte should be at least 0.1 to 2

    times of the exposure limit. If it is toxic or suspected to be carcinogenic, there could be a

    concentration lower than that calculated which needs to be considered.

    NIOSH Method 3500 was checked for reproducibility by having 3 different analysts in 2

    different laboratories analysed standard samples containing between 1-20 micrograms of

    formaldehyde. The results are approximately the same of 5%.

    Evaluation from the sampling of a generated atmosphere determines the following:

    1) Capacity of the sampler

    2) Efficiency of analyte collection

    3) Repeatability

    4) Bias

    5)

    Interferences in the collection by the sampler

    Generation of the Analyte

    As part of the evaluation method, the collection of samples is needed from an environment

    that is as close to the actual sampling conditions as possible. And to fulfil this, the impact of

    environmental conditions such as temperature, pressure, humidity and interferences needs to

    be in consideration.

    Table 2: Effects of surrounding conditions on the analyte

    S/N Changed conditions Effect

    1 Increased temperature on the

    collection medium

    Decreased capacity of sampler or decompose

    the analyste

    2 Reduced pressure Reduced capacity of a sampler

    3 High relative humidity Reduced sampler capacity

    After the generation of the analyte, its concentration will be verified by a

    gravimetric/volumetric means. Ideally, this independent method of verification should not be

    biased and should provide an accurate estimated of the concentration. Precision and bias is

    also homogenous over the concentration ranges.

    Capacity of the Sampler and Sampling Rate

    To determine the applicability, the capacity of the sampler should be determined by the

    function of flowrate and sampling time.

  • 7/26/2019 Literature Review of Monitoring Methods for Formaldehyde

    11/26

    11 of 26

    Different flow rates typical of the media will be used. Sampling should be performed at 3

    different flow rates. The amount of analyte collected at the lowest flow rate and shortest

    sampling time should be greater than the limit of quantitation of the method. The generated

    analyte should be at least 2 times the highest published exposure limit will be used to

    determine the sample capacity to be used.

    Sampling should be conducted at various temperatures and humidity to check on its effect of

    these parameters on the capacity. 3 replicates at different flow rates should be used to verify

    the capacities at each of the different humidity and temperature

    Table 3: Temperature and humidity conditions

    Parameters High Low

    Temperature Ambient >35 C

  • 7/26/2019 Literature Review of Monitoring Methods for Formaldehyde

    12/26

    12 of 26

    Bias is assumed to be homogenous across the evaluated concentration ranges. This

    assumption should be evaluated to see if there is homogeneity. Bias is estimated at each

    concentration. To fulfil the criterion, method bias should be 10%

    Accuracy is the intersections of these points on the parabolic grid in the graph. If the both the

    upper confidence limits on the accuracy are less than 25%, it fulfils the accuracy criterion.

    Field Evaluation

    Field Evaluation is not required in the NIOSH Guidelines for Air Sampling and Analytical

    Method Development and Evaluation, as conditions which exist in the field are difficult to

    reproduce in the laboratory.

    The field evaluation is recommended to further study the performance of the method. Both

    the collection of area samples and personal samples should be included in the field evaluation

    of the method. Area samples can be used to see if there is field precision and bias. Personalsamples can be used to assess the utility of the method.

    In the following is a field precision study extracted from the article, Field Precision of

    Formaldehyde Sampling and Analysis using NIOSH Method 3500, the American Industrial

    Hygiene Association Journal 58:9,657-6607 The study was designed to examine the field

    precision by collecting and analysing a series of replicate samples over an extended period.

    The field area is the gross anatomy laboratory during their normal activities. There is also

    presence of interferences in the existing work environment (oxidisable organics) where

    embalming solutions were used.

    Method:

    2-4 replicates of airborne samples are obtained in the middle of a gross anatomy lab on 29

    days. The inlets of the flow pumps (Fig 5) were positioned upward all at the same level about

    120cm from the floor which emulate the breathing zone of the workers during dissection.

    Distance between each replicate is about 10cm, as shown below

    Figure 4: Set up of the Experiment

    7Khanzadeh, F. A., & Park, C. K. (1997). Field Precision of Formaldehyde Sampling and Analysis Using

    NIOSH Method 3500.American Industrial Hygiene Association Journal, 58(9), 657-660.doi:10.1080/15428119791012450

    120cm

    3750cm

    750cm350cm

    10 cm

  • 7/26/2019 Literature Review of Monitoring Methods for Formaldehyde

    13/26

    13 of 26

    The room temperature and pressure were measured during every collection of the sample.

    The sample air flow rate was also determined before and after each sampling by a flow meter,

    and the target sampling volume was set to be 0.2-1 L/min. Air volumes are adjusted to 25 deg

    Celsius and 760mmHg accordingly to 1-100L

    Figure 5: SKC PCXR Constant Flow Pump used to collect replicates

    It was decided that the Polytetrafluoroethylene filter would not be used as it is a not a dusty

    environment. The assumption is derived due to the whole dissecting operations being

    performed on moist tissues without any mechanical tools, there is no evident source of

    particulate generation. Even if there is, their effects on the precision will be negligible.

    Results & Discussion

    A total of 98 air samples were collected. 13 samples did not fulfil the target sampling and

    working ranges, 4 of them exceeding sample volume range limit of 100L and 9 samples show

    flow rate fluctuations of 10%. Lastly, 7 samples out of the 85 remaining were single and notreplicates. Hence, the remaining samples left for data analysis is 78.

    The data were entered into statistical analysis, using Statistical Package for Social Sciences.

    The descriptive statistics were used to determine means, standard deviations (SD), and the

    ranges of sampling parameters, concentrations and coefficients of variation (CV, precision)

    of each set of replicates. A t-test was applied to see the difference between the means of 2

    independent groups.

    CV = Standard deviation/Mean. A pooled CV is to demonstrate the overall precision of each

    sampling and analytical method.

  • 7/26/2019 Literature Review of Monitoring Methods for Formaldehyde

    14/26

    14 of 26

    Table I: Sampling Parameters and Formaldehyde concentrations by Number of Sample

    Replicates

    Mean

    (SD)

    Range

    Number of Replicates in Each Set2 3 4 Overall

    Number of

    Samples

    16 18 44 78

    Sampling

    Time (min)

    186(18)

    160-215

    147(26)

    120-187

    156(28)

    115-220

    160(29)

    115-220

    Sampling

    Flow (L/min)

    0.4(0.1)

    0.2-0.5

    Sampling

    Volume (L)

    66(2)

    39-89

    56 (21)

    26-99

    60(19)

    27-100

    60(19)

    26-100

    Concentration

    (ppm)

    1.01(0.48)

    0.05-1.71

    0.84(0.17)

    0.62-1.25

    0.97(0.28)

    0.59-1.72

    0.95(0.31)

    0.05-1.72

    From Table II, the precision of replicate samples ranged from 0.03 to 0.24 with an overall

    precision of 0.09. This is equal to the precision of NIOSH Method 3500.

    The results of the precision also improve as the number of samples in each case increasedfrom 2-4 replicates. Nevertheless, within the preset target sampling and working ranges, the

    results did not demonstrate any significant relationship between the parameters of CV,

    formaldehyde concentrations, flow rate, sampling time or sampling volume.

    Table II: Coefficient of Variation (CV) by number of Sample Replicates Collected in a Gross

    Anatomy Laboratory

    Number of Replicates in Each Set

    2 3 4 Overall

    Number of sets 8 6 11 25

    Range of CV 0.035-0.241 0.026-0.108 0.026-0.201 0.026-0.241

    Pooled CV 0.116 0.079 0.092 0.093

    Conclusion:

    The overall precision of the field sampling and analysis was 0.09 which is equal to the

    precision of Method 3500 as determined in the laboratory.

  • 7/26/2019 Literature Review of Monitoring Methods for Formaldehyde

    15/26

    15 of 26

    2.2

    OSHA 1007

    The OSHA Method 1007 describes the validation of diffusive samplers that are meant to

    provide an alternative to an existing method, ID-205. ID-205, which validates a diffusive

    monitor from 3M (model 3721), is limited by the minimum sampling time of 4 hours, making

    it unsuitable to measure STEL level which requires a sampling time of 15 minutes.

    The OSHA 1007 method has evaluated and fully-validated three brands of diffusive samplers.

    These samplers use 2,4-dinitrophenyl hydrazine (DNPH) to react with formaldehyde in the

    presence of an acid to form a unique derivative. Supelcos sampler is found by the authors of

    this report to have been replaced by a high efficiency (HE) model which is not suitable for

    STEL measurement, and hence not suitable for comparison with the other two brands. Refer

    to the Table 4 for other useful information about the samplers.

    The laboratory analysis of field samples requires a liquid chromatograph equipped with a UV

    detector (set at 365 nm wavelength). Refer to the OSHA document for detailed guidelines for

    the analytical procedure8.

    Table 4: Diffusive samplers validated by OSHA Method 1007

    Specifications

    Assay Technology

    ChemDisk 571

    Aldehyde Monitor

    SKC UMEx 100

    Passive Sampler

    Supelco DSD-DNPH

    Diffusive Sampling

    Device

    High Efficiency (HE)

    model

    Reagent 2,4-dinitrophenyl hydrazine (DNPH)

    Collection medium FiberglassTape (material

    unknown)

    Spherical silica gel

    (105210 m)Capacity 100 ppm-hrs 29 g 150 g for HE model

    Lower

    Detection

    Limits

    15 min 330 ppb 200 ppb (0.24 mg/m3)Unknown; HE model

    not recommended for

    STEL

    8 hr 10 ppb 5 ppb (6 g/m3)

    24 hr N.A. 2 ppb (2 g/m3)

    7 days N.A. 0.2 ppb (0.2 g/m3)

    8OSHA, Method 1007: Formaldehyde (Diffusive Samplers), T-1007-FV-01-0505-M, May 2005

  • 7/26/2019 Literature Review of Monitoring Methods for Formaldehyde

    16/26

    16 of 26

    Limitations

    This method is not suitable for sampling formaldehyde exposure at workplaces where

    formalin solutions, formaldehyde/water solution stabilized with methyl alcohol, is used;

    OSHA recommends the active sampling method (OSHA Method 52) to be used when

    monitoring exposures resulting from the use of formalin solutions.

    This method is not suitable for atmospheres with more than 0.5 ppm ozone concentration due

    to interference of DNPH reaction with formaldehyde. DNPH was found to be resistant to

    interference from other aldehydes such as, acetaldehyde, butyraldehyde, benzaldehyde and

    glutaraldehyde. The samplers also requires at least 10% humidity for best performance and

    should be stored at 4 C before and after sampling.

  • 7/26/2019 Literature Review of Monitoring Methods for Formaldehyde

    17/26

    17 of 26

    2.3

    NIOSH 5700

    In the wood and textile industries, there is potential for airborne formaldehyde to be bound to

    inhalable dust. NIOSH Method 5700 is an active sampling method using a sampler

    developed by the Institute of Occupational Medicine (IOM). The sample and filter is

    designed to collect inhalable dust and comes with a PVC filter with pore size of 5 m.

    An example of the sampler is shown in Figure 6. The sampler is fully evaluated for

    concentration range of between 0.007 and 0.16 mg/m3.

    The recommended analytical method is HPLC using UV detector at 365 nm wavelength.

    However, the VAS analytical method used in NIOSH Method 3500 is applicable provided

    that there are no substances in the sample which can interfere with the chromotropic acid

    analysis. These substances include phenol, oxidizable organic material, other aldehydes and

    alcohols.

    As the results collected by this method are not required for compliance, the NIOSH manual

    cautioned against combining the results from this method together with results collected by

    methods which measures the vapour-phase concentration.

    Figure 6. SKC IOM Sampler

  • 7/26/2019 Literature Review of Monitoring Methods for Formaldehyde

    18/26

    18 of 26

    2.4

    Other Methods

    2.4.1 Direct Reading Instruments

    Direct reading instruments provide fast access to results at the field and is more useful than

    the standardized methods in cases where qualitative measurements are desired in real-time to

    pinpoint hot spots. Current direct reading instruments for airborne formaldehyde makes use

    of electrochemical sensing, photoelectric photometry, colorimetric or photo-ionization

    technologies to detect the target molecules.

    These equipment are primarily used as a screening tool to establish the orders of magnitude

    of the airborne concentration levels of formaldehyde, where higher than desired levels will

    warrant the more costly and time-consuming standardized laboratory methods to be carried

    out. The instruments have many limitations affecting the accuracies of the readings. These

    limitations include: sensitivity to temperature, pressure and humidity; non-selective toformaldehyde; and prone to contamination and adsorption effects. A recent research

    compares two direct reading instruments to NIOSH Method 2016 and found that one of the

    instrument significantly underestimated formaldehyde concentration as compare to the

    laboratory method whereas the other instrument produced results that are not statistically

    significantly different9.

    2.4.2

    Emission testing standards

    Apart from indoor and ambient air concentrations, formaldehyde emissions from

    manufactured products are also regulated in countries in North America, Europe and Asia.Different standards are employed in various countries. In the US, the California Air

    Resources Board (CARB) approved the use of ASTM E 1333 and ASTM D 6007 as the

    standards for emission testing in large and small chambers respectively. A smaller chamber,

    known as the Dynamic Micro Chamber (DMC), although not included in the standards, has

    shown good comparability with large chambers for measuring formaldehyde emissions. In

    Europe, EN 717-1 (large chamber method) and EN 717-2 (gas analysis method) are two

    common standards used in emission testing 10 . Other methods includes the EN 120

    (perforator method), EN 717-3 (flask method) and JIS A 1460 (desiccator method).

    A highly portable and widely tested device, called the field and laboratory emission cell

    (FLEC), has been in use since 1991 to carry out emission testing and quality assurance on

    location. The major advantages are its portability and that the test procedure is nondestructive

    9Hirst, D.V.L; Gressel, M.G.; Flanders, W.D., Short-Term Monitoring of Formaldehyde: Comparison of Two

    Direct-Reading Instruments to a Laboratory-Based Method, Journal of Occupational and Environmental

    Hygiene, Vol. 8, pg 357-36310Bhm, M.; Salem, M.Z.M; Srba, J., Formaldehyde emission monitoring from a variety of solid wood,

    plywood, blockboard and flooring products manufactured for building and furnishing materials, Journal of

    Hazardous Materials, 2012, pg 68-79

  • 7/26/2019 Literature Review of Monitoring Methods for Formaldehyde

    19/26

    19 of 26

    in nature11. The design and properties of this type of emission cell is covered by the ISO

    16000-10 standard.

    Figure 7, 8 and 9 illustrates the various test methods

    Figure 7: European chamber method EN 717-1

    Figure 8: Other emission testing methods

    11

    Salthammer, T.; Mentese, S.; Marutzky, R., Formaldehyde in the Indoor Environment, Chemical Reviews,2010, Vol. 110, pg 2536-2572

  • 7/26/2019 Literature Review of Monitoring Methods for Formaldehyde

    20/26

    20 of 26

    Figure 9: Field and laboratory emission cell (FLEC)

  • 7/26/2019 Literature Review of Monitoring Methods for Formaldehyde

    21/26

    21 of 26

    2.5

    Novel Methods

    Since 2012, the WHO International Agency for Research on Cancer (IARC) has expanded its

    classification of formaldehyde as a known carcinogen (Group 1), with positive links to

    nasopharyngeal cancer, leukaemia, as well as sinonasal cancer12. As such, with the hazards

    of formaldehyde attracting increasing attention, the need for a reliable yet cost effective

    method for measuring accurately formaldehyde levels in the environment is apparent.

    As discussed in the previous chapters, while NIOSH approved methods are the gold standard

    for sampling and analyzing formaldehyde concentration in ambient air for compliance

    monitoring, these are primarily chromatography based methods which are ex-situ, requiring

    analysis in a lab following collection in the field. These methods are hence not suited for

    rapid, real time monitoring of formaldehyde concentration in the environment.

    As such, some research have explored novel methods of on-site quantification of

    formaldehyde in air, which does not use the traditional methods of liquid or gas

    chromatography, but instead uses other methods which fulfil the requirements of being both

    simple and portable, as well as being sufficiently specific and sensitive enough for large scale

    deployment for environment monitoring in-situ.

    While these requirements are partly fulfilled by some currently available direct measurement

    methods which uses electrochemical gas sensors and colorimetric detection tabs 13, these

    direct measurements methods suffer from several limitations, which affects the accuracies of

    the reading. Hence although portable, and able to give fast access to results on the field in

    real time, the equipment are not cost-effective, and in some cases of insufficient sensitivityand selectivity to provide reliable measurements.

    Thus, one of the new novel method which stands out is biosensors, specifically an enzyme

    based detection method, which can give provide readings in-situ, with greater selectivity and

    reliability than current commercially available direct measurement methods.

    2.5.1

    Biosensors

    Biosensor is a general term that describe devices which can be used to detect a specific

    substances, using a combination of biological components and physiochemical detector. Thebiological component of a biosensor, such as enzymes or cell receptors, are usually highly

    specific, reacting or binding only with the specific desired substance. The reaction between

    the biological component and the target substance then generate specific physiochemical

    signals, such as color or florescence, which can then be detect and quantified by the

    physiochemical detector.

    12International Agency for Research on Cancer (IARC), Formaldehyde, IARC Monographs on the Evaluation

    of Carcinogenic Risks to Humans, 2012, vol. 100F13Hirst, D. V., Gressel, M. G., & Flanders, W. D. (2011). Short-Term Monitoring of Formaldehyde:

    Comparison of Two Direct-Reading Instruments to a Laboratory-Based Method.Journal of occupational and

    environmental hygiene, 8(6), 357-363.

  • 7/26/2019 Literature Review of Monitoring Methods for Formaldehyde

    22/26

    22 of 26

    The novel methods of formaldehyde measurement in the environment using biosensors hence

    could potentially provides several advantages over both the approved NIOSH methods and

    commercially available direct measurement methods, These advantages are:

    On-situ measurements and reading

    High specificity of detection

    Cost effective

    Two of such novel formaldehyde detection methods using biosensors will be further

    discussed below. These two methods make use of DNAzymes and formaldehyde

    dehydrogenase respectively, with both having their own which uses and advantages.

    2.5.2

    DNAzymes

    DNAzymes, or catalytic DNA are a kind of artificial enzymes which are capable of specific

    catalytic activities. They are widely used for a variety of biochemical reaction, such as DNAglycosidic bond cleavage and DNA self-modification. They have also been used as the

    biological component of biosensors, being used to detect nucleic acid, proteins and metal ions.

    As such, research has been made to apply the use of DNAzymes into the detection of

    formaldehyde, to develop a low cost, sensitive and selective biosensor.

    In a research paper by Yang and et al14, they discussed the procedure used for optimizing the

    colorimetric property of 2,2-azino-bis(3-ethylben-zothiazoline-6-sulfonic acid)(ABTS) for

    the detection of formaldehyde.

    ABTS+ is a blue-green-colored free-radical cation, and the reaction of ABTS with H2O2 toform ABTS+ is catalyzed by the hermin-G-quadruples complex (DNAzymes). Formaldehyde

    participates in a competing reduction-oxidation reaction with H2O2, which thus prevents the

    formation of the ABTS+. As such, the concentration of formaldehyde present can be deduced,

    through the indirect measurement of ABTS+ using a colorimetric assay.

    One possible advantage of this method of formaldehyde analysis is the selectivity of the

    process compared to available direct reading instruments, as various other possible

    contaminants tested, such as various alcohols (known interferent for the htV formaldehyde

    meter) do not have any significant effect on the measured results.

    Through the use of a smartphone application and a portable set up consisting of a dark room

    with a LED light, the colorimetric assay can be performed in the field, for rapid assessment of

    the results:

    14Yang, X., Wang, Y., Liu, W., Zhang, Y., Zheng, F., Wang, S., & Wang, J. (2016). A portable system for on-

    site quantification of formaldehyde in air based on G-quadruplex halves coupled with A smartphone reader.Biosensors and Bioelectronics, 75, 48-54.

  • 7/26/2019 Literature Review of Monitoring Methods for Formaldehyde

    23/26

    23 of 26

    Figure 9: Schematic of the portable smartphone-based optical reader

    This detector setup was found to be able to detect formaldehyde in a linear range of 0.02 to

    14 ppm, and is thus suitable for practical usage, based on the maximum threshold level of 0.1

    ppm recommended by NEA.

    2.5.3

    Formaldehyde dehydrogenase

    Similarly, in another research paper by Kudo and et al15, the use of another novel biosensor

    for formaldehyde monitoring was demonstrated. Their technique is also enzyme-based,

    similar to the previous one discussed, but for this the fluorescence property of nicotinamide

    adenine dinucleotide (NADH) was used as an indirect measurement of formaldehyde

    concentration.

    Formaldehyde dehydrogenase (FALDH), an enzyme which catalyzes the chemical reaction

    between formaldehyde and NAD+ (oxidized form) to formate and NADH (reduced form),

    was prepared and immobilized on a membrane, which act as the biological component of the

    biosensor. Hence, the amount of NADH measured on the membrane after exposure can be

    correlated with the amount of formaldehyde originally present.

    The membrane was mounted in a flow cell set up, together with a custom fiber-optic NADH

    measurement system which can measure the fluorescent intensity of the membrane, to createa formaldehyde-sensitive optode.

    By connecting the optode to a pump system for circulation of a phosphate buffer solution

    containing NAD+, the membrane can be rinsed, enabling continuous monitoring of

    formaldehyde levels, as compared to the previous biosensor setup:

    15Kudo, H., Suzuki, Y., Gessei, T., Takahashi, D., Arakawa, T., & Mitsubayashi, K. (2010). Biochemical gas

    sensor (bio-sniffer) for ultrahigh-sensitive gaseous formaldehyde monitoring.Biosensors and Bioelectronics,26(2), 854-858.

  • 7/26/2019 Literature Review of Monitoring Methods for Formaldehyde

    24/26

    24 of 26

    Figure 10: Structure of flow-cell and formaldehyde-sensitive optode

    The setup was found to be highly sensitive and selective in comparison to conventional

    formaldehyde gas sensor. The selectivity was to be expected, due to the specificity of the

    FALDH enzyme reaction, and the optode was able to monitor formaldehyde levels

    continuously at very low concentrations from 2.5 ppb to 10 ppm.

  • 7/26/2019 Literature Review of Monitoring Methods for Formaldehyde

    25/26

    25 of 26

    3.0 Conclusion

    Formaldehyde can be an invisible killer, if not for regulations put in place to limit its

    concentration at the workplace. Standardized methods provide a way for enforcing the

    regulation at the workplace where formaldehyde is used. However, it can still creep into ourlives unsuspectingly through the use of shoddy building material emitting high concentration

    of the chemical. Direct reading instruments utilizing the novel technologies provide a quick

    and reliable measurement of airborne formaldehyde concentration. They should be widely

    used to provide assurances for indoor air quality in high risk locations such as newly

    constructed homes, newly opened underground retail spaces, etc.

    The usage of either the direct-reading instruments or the standardized laboratory methods

    depend on the objective of the experiment, the condition of the sample, the equipment

    available and the requirements of the regulations. All the limitations and the advantages of

    the method must be considered before choosing it.

    At the cutting edge, there are several novel methods of formaldehyde detection which are

    being researched and tested in the lab. These methods use differing principles and concepts,

    which have several advantages over the convention and traditional methods, particularly in

    for on-site monitoring in the field, and could be of great utility and benefit if realized in the

    future.

    This literature review exercise has given the group members a more thorough understanding

    about the hazards of formaldehyde and the methods of monitoring exposure to this chemical.

  • 7/26/2019 Literature Review of Monitoring Methods for Formaldehyde

    26/26

    4.0 References

    [1] Formaldehyde, World health Organization, 2002

    [2] IAQ, Indoor Air quality website url: http://www.iaqsg.com/chemical-

    parameters/formaldehyde , accessed on 04/04/2016[3]

    NIOSH, NOISH Manual of Analytical Methods, Fourth Ed., 1994

    [4] NEA, National Environment Agency website, url:www.nea.gov.sg,date retrieved:

    29th March 2016

    [5]

    Kennedy, E. R. (1994). FORMALDEHYDE: METHOD 3500, Issue 2. NIOSH

    Manual of Analytical Methods (NMAM), Fourth Edition, 8/15/94. Retrieved April 12,

    2016, fromhttp://www.cdc.gov/niosh/docs/2003-154/pdfs/3500.pdf

    [6] Kennedy, E. R., Fischbach, T. J., Song, R., Eller, P. M., & Shulman, S. A. (1996).

    Summary of the NIOSH Guidelines for Air Sampling and Analytical Method

    Development and Evaluation. Analyst, 121. Retrieved April 4, 2016, from

    http://www.ncbi.nlm.nih.gov/pubmed/8831274[7] Khanzadeh, F. A., & Park, C. K. (1997). Field Precision of Formaldehyde Sampling

    and Analysis Using NIOSH Method 3500.American Industrial Hygiene Association

    Journal, 58(9), 657-660. doi:10.1080/15428119791012450

    [8] OSHA, Method 1007: Formaldehyde (Diffusive Samplers), T-1007-FV-01-0505-M,

    May 2005

    [9]

    Hirst, D.V.L; Gressel, M.G.; Flanders, W.D., Short-Term Monitoring of

    Formaldehyde: Comparison of Two Direct-Reading Instruments to a Laboratory-

    Based Method, Journal of Occupational and Environmental Hygiene, Vol. 8, pg 357-

    363

    [10]

    Bhm, M.; Salem, M.Z.M; Srba, J., Formaldehyde emission monitoring from avariety of solid wood, plywood, blockboard and flooring products manufactured for

    building and furnishing materials, Journal of Hazardous Materials, 2012, pg 68-79

    [11] Salthammer, T.; Mentese, S.; Marutzky, R., Formaldehyde in the Indoor

    Environment, Chemical Reviews, 2010, Vol. 110, pg 2536-2572

    [12] International Agency for Research on Cancer (IARC), Formaldehyde, IARC

    Monographs on the Evaluation of Carcinogenic Risks to Humans, 2012, vol. 100F

    [13]

    Hirst, D. V., Gressel, M. G., & Flanders, W. D. (2011). Short-Term Monitoring of

    Formaldehyde: Comparison of Two Direct-Reading Instruments to a Laboratory-

    Based Method.Journal of occupational and environmental hygiene, 8(6), 357-363.

    [14]

    Yang, X., Wang, Y., Liu, W., Zhang, Y., Zheng, F., Wang, S., & Wang, J. (2016). Aportable system for on-site quantification of formaldehyde in air based on G-

    quadruplex halves coupled with A smartphone reader.Biosensors and Bioelectronics,

    75, 48-54.

    [15] Kudo, H., Suzuki, Y., Gessei, T., Takahashi, D., Arakawa, T., & Mitsubayashi, K.

    (2010). Biochemical gas sensor (bio-sniffer) for ultrahigh-sensitive gaseous

    formaldehyde monitoring.Biosensors and Bioelectronics, 26(2), 854-858.

    http://www.nea.gov.sg/http://www.nea.gov.sg/http://www.nea.gov.sg/http://www.cdc.gov/niosh/docs/2003-154/pdfs/3500.pdfhttp://www.cdc.gov/niosh/docs/2003-154/pdfs/3500.pdfhttp://www.cdc.gov/niosh/docs/2003-154/pdfs/3500.pdfhttp://www.ncbi.nlm.nih.gov/pubmed/8831274http://www.ncbi.nlm.nih.gov/pubmed/8831274http://www.ncbi.nlm.nih.gov/pubmed/8831274http://www.cdc.gov/niosh/docs/2003-154/pdfs/3500.pdfhttp://www.nea.gov.sg/