M1031C_Eng_24.12.02

Embed Size (px)

Citation preview

  • 7/30/2019 M1031C_Eng_24.12.02

    1/25

    MEASUREMENT EQUIPMENT

    FREE CHLORINE

    RESIDUAL ANALYSER

    M 1031 C

    CONTROLMATIK ABW d.o.o.Fajfarjeva 15, 1230 Domale, Slovenija, EvropaTel.: 01 721 35 52, Fax.: 01 721 93 60http://www.cm-abw.si, Email: [email protected]

  • 7/30/2019 M1031C_Eng_24.12.02

    2/25

    MEASUREMENT EQUIPMENT

    1.GENERAL............................................................................................................................................................4

    2.BASIC PRINCIPLES OF THE AMPERE-METRIC CHLORINE MEASUREMENT...............................5

    2.1 DIAGRAMSHOWINGLEVELOF HOCL ACIDDEPENDINGON PH ANDTEMPERATURE......................................................5

    2.2 DIAGRAMSHOWINGLEVELOF HOCL ACIDDEPENDINGFROMPH ANDTEMPERATURE...................................................5

    3.MOUNTING.........................................................................................................................................................6

    3.1 DETERMININGTHEOPTIMALPLACEOFINSTALLATION................................................................................................6

    3.2 DRILLINGHOLESINTHEWALL .............................................................................................................................6

    3.3 PLACING...........................................................................................................................................................6

    4.THE LEVEL OF FREE-RUNNING SAMPLE WATER................................................................................6

    4.1 POOLS..............................................................................................................................................................6

    4.2 WATERSUPPLIES................................................................................................................................................6

    5.TAKING SAMPLE WATER............................................................................................................................ ..7

    5.1 POOLS..............................................................................................................................................................7

    5.2 TAKINGSAMPLEWATERFROMTHEPIPE ................................................................................................................7

    5.3 IMPROPERTAKINGOFSAMPLEWATER....................................................................................................................75.4 TAKINGSAMPLEWATER......................................................................................................................................8

    6.CONNECTING THE INLET AND OUTLET PIPES OF SAMPLE WATER.............................................8

    7.CONNECTING TO POWER SUPPLY............................................................................................................ .8

    8.CURRENT LOOP................................................................................................................................................9

    9.CURRENT LOOP TABLE.................................................................................................................................9

    10.CAN NETWORK............................................................................................................................................ .10

    10.1SELECAN......................................................................................................................................................10

    10.2SEGMENTALCIRCUIT (RECOMMENDED!)....................................................................................................1010.3TERMINATORS .................................................................................................................................................11

    11.A DESCRIPTION OF KEYS AND LIGHT WARNINGS...........................................................................13

    11.1WHENTHE EDIT LIGHTIS OFF.......................................................................................................................14

    11.2WHENTHE EDIT LIGHTIS ON.........................................................................................................................14

    12.A DESCRIPTION OF THE DISPLAY AND ITS FUNCTIONS................................................................15

    13.BEFORE STARTING......................................................................................................................................15

    14.ENTERING THE PASSWORD.....................................................................................................................15

    15.SETTING THE FREE CHLORINE LEVEL................................................................................................16

    15.1WHATYOUNEEDTOKNOW:..............................................................................................................................16

    15.2HOWTODOIT:................................................................................................................................................17

    15.3SETTINGTHEVALUE:........................................................................................................................................17

    16.SETTING FREE CHLORINE .......................................................................................................................17

    17.SETTING THE REGULATOR ................................................................................................................... ..17

    18.SETTING THE TEMPERATURE.................................................................................................................18

    19.SETTING THE CURRENT LOOP................................................................................................................19

    20.RESTART.........................................................................................................................................................20

    CONTROLMATIK ABW d.o.o.Fajfarjeva 15, 1230 Domale, Slovenija, EvropaTel.: 01 721 35 52, Fax.: 01 721 93 60http://www.cm-abw.si, Email: [email protected]

    2

  • 7/30/2019 M1031C_Eng_24.12.02

    3/25

    MEASUREMENT EQUIPMENT

    21.CLEANING......................................................................................................................................................20

    21.1CLEANINGOFTHEMEASUREMENTCELL...............................................................................................................21

    21.2CLEANINGTHEINLETFILTER..............................................................................................................................21

    22.THE MENU STRUKTURE............................................................................................................................21

    23.DIAGNOSTICS................................................................................................................................................22

    24.DIMENSIONS..................................................................................................................................................22

    25.TECHNICAL DATA.......................................................................................................................................23

    26.M1031C ANALYSER CODES ......................................................................................................................24

    CONTROLMATIK ABW d.o.o.Fajfarjeva 15, 1230 Domale, Slovenija, EvropaTel.: 01 721 35 52, Fax.: 01 721 93 60http://www.cm-abw.si, Email: [email protected]

    3

  • 7/30/2019 M1031C_Eng_24.12.02

    4/25

    MEASUREMENT EQUIPMENT

    1. General

    Before mounting the M 1031 C analyser you must read the instructions. Anauthorized person must carry out mounting of the analyser.Type M 1031 C analyser is a measuring instrument operating on the basis of the

    amperometric method. It is used for measuring free chlorine concentration in water invarious dosing fields such as:

    Drinking water

    Pool water

    Treatment of industrial and sewage water

    The analyser is always purchased together with all accompanying elements that arenecessary for quick and functional installation of the measuring system (armature ofthe sink, installation inserted pieces, screws, ball tap, drilling templates)

    The analyser has been designed for continuous operation.

    Connector

    Outlet pipesInlet filter

    Inlet ofsample water

    Outlet connection

    CONTROLMATIK ABW d.o.o.Fajfarjeva 15, 1230 Domale, Slovenija, EvropaTel.: 01 721 35 52, Fax.: 01 721 93 60http://www.cm-abw.si, Email: [email protected]

    4

  • 7/30/2019 M1031C_Eng_24.12.02

    5/25

    MEASUREMENT EQUIPMENT

    2. Basic principles of the ampere-metric chlorine measurement

    2.1Diagram showing level of HOCL acid depending on PH and temperature

    General:

    Currently there are two methods for determining/measuring chlorine in use:Chlorometric determiningElectronic measuring (amperometrics)

    Chlorometric determining is mainly used for checking and calibration.DPD method is still used in public swimming pools(reagent: N,n-dietil-p-fenildiamin).Chlorometric determining is not appropriate for continuous checking and automaticchlorination. Measuring is in such cases executed via electronic measuring, with theamperometric method, with the measuring cell for chlorine and the appropriatemeasuring and regulating instrument. We need to differentiate between free andcombined chlorine.The term total chlorine is usually related to the chlorine in water, consisting of the

    elemental chlorine Cl2, hypochloric acid (HOCl), hypochloric ion (ClO), and of acombination of the above. With short reaction periods these forms of chlorine arecapable of destroying bacteria, viruses and oxidising organic substances.The term combined chlorine is usually related to the chlorine in water, which ischained to chlorine compounds with ammoniac (NH3) or ammoniac ion (NH4+). Theterm total chlorine means free and combined chlorine together. Free chlorine ismeasured via amperometric measuring. The result of hypochloric acid (HOCl) is thedisinfection of water. This acid is split into hypochloric ion (ClO) and hydrogen ion(H+), depending on the PH value of the water.The diagram below clearly shows that the share of hypochloric acid decreases when

    the PH value increases. E.g., at PH value 9 the share of hypochloric acid drops toapprox. 3 %, as compared to 97 % at PH 6.

    2.2 Diagram showing level of HOCL acid depending from pH andtemperature

    0

    10

    20

    30

    40

    50

    60

    70

    80

    90

    100

    6,00

    6,20

    6,40

    6,60

    6,80

    7,00

    7,20

    7,40

    7,60

    7,80

    8,00

    8,20

    8,40

    8,60

    8,80

    9,00

    9,20

    9,40

    9,60

    9,80

    10,00

    pH

    HOCl(%

    )10C

    20C

    25C

    CONTROLMATIK ABW d.o.o.Fajfarjeva 15, 1230 Domale, Slovenija, EvropaTel.: 01 721 35 52, Fax.: 01 721 93 60http://www.cm-abw.si, Email: [email protected]

    5

  • 7/30/2019 M1031C_Eng_24.12.02

    6/25

    MEASUREMENT EQUIPMENT

    3. Mounting

    3.1Determining the optimal place of installationMount the analyser as close as possible to the place of taking water sample and in

    an easily accessible place in order to make the mounting undisturbed.

    The outlet connection of the armature of the analyser sink must always be higherthan the main drainage outlet (the outlet tap should never rise above the height of thearmature of the analyser sink).

    NELA0006

    3.2 Drilling holes in the wall

    Place the enclosed template (code SABXXXXX) for drilling holes against the wall(pay attention to the first angle). With help of the template drill 3 holes with adiameter d8 and depth 60 mm into the wall. Insert the enclosed PVC wall plugs d8into the holes.Insert three enclosed screws with the length of 70 mm into the wall plugs and leaveapprox. 12 mm of the screw unfastened (unturned).

    3.3 Placing

    Hang the analyser on the three screws and check the stability of the installation.

    4. The level of free-running sample water

    4.1 Pools

    The minimal level of free-running sample water is: 1 m of the water column above thelevel of the analyser.

    When the minimal level of free-running sample water is not possible, take the samplewith the help of a pump.

    4.2 Water supplies

    When you are taking the water sample, of which the pressure exceeds 6 bars, inserta reducing valve and set it to 1 bar.

    CONTROLMATIK ABW d.o.o.Fajfarjeva 15, 1230 Domale, Slovenija, EvropaTel.: 01 721 35 52, Fax.: 01 721 93 60http://www.cm-abw.si, Email: [email protected]

    6

  • 7/30/2019 M1031C_Eng_24.12.02

    7/25

    MEASUREMENT EQUIPMENT

    5. Taking sample water

    5.1 Pools

    0.3m

    NELA0009

    Taking sample water from the pool

    5.2 Taking sample water from the pipe

    SEDIMENTSSEDIMENTS

    AIRAIR

    0.66d

    0.66 d

    d dNELA0022 NELA0023

    Taking sample water from the pipe Taking sample water from the pipe( RECOMENDED ! )

    5.3 Improper taking of sample water

    SEDIMENTS

    SEDIMENTS

    AIR

    AIR

    NELA0024 NELA0025

    CONTROLMATIK ABW d.o.o.Fajfarjeva 15, 1230 Domale, Slovenija, EvropaTel.: 01 721 35 52, Fax.: 01 721 93 60http://www.cm-abw.si, Email: [email protected]

    7

  • 7/30/2019 M1031C_Eng_24.12.02

    8/25

    MEASUREMENT EQUIPMENT

    5.4 Taking sample water

    ANALYSER

    d

    min 15 d NELA0027

    Taking sample water

    6. Connecting the inlet and outlet pipes of sample water

    Before connecting the analyser, you must wash out the inlet pipes.Connect the inlet pipe d10/d8 of the sample water to the inlet connector of theanalyser.The regulating valve serves as a regulating element (setting of water flow through theanalyser).Install an outlet pipe with a diameter of d20 mm on the outlet end of the outletarmature.The sample water flow through the analyser is set by means of the regulating valve.Sample water flow through both outlet pipes of the analyser must be even. Thesample water usage is 0.7 l per minute.

    7. Connecting to power supply

    Connect the M1031C analyser to a voltage of 24 VAC (+/- 10%), 50 - 60 Hz. Connectthe voltage through the 5-pole connector on ECP00266 cable to the connector on theinstrument (CONNECTION).All such connectors have a code indicator and can therefore only be connected ifturned the right way. The connectors on the cable are to be gently and correctly puton the connectors on the instrument and the fixing nut is to be gently turned.

    Sensor Motor

    Power

    NELA0001

    Connector CONNECTION serves toconnect voltage and signallingcommunication, varying from theexecution of the instrument. Accordingto the execution, a version with CANcommunication and current loop canbe supplied.

    CONTROLMATIK ABW d.o.o.Fajfarjeva 15, 1230 Domale, Slovenija, EvropaTel.: 01 721 35 52, Fax.: 01 721 93 60http://www.cm-abw.si, Email: [email protected]

    8

  • 7/30/2019 M1031C_Eng_24.12.02

    9/25

    MEASUREMENT EQUIPMENT

    NELA0002

    View of 5 pole connective connector oncable ECP00266 from connection clips

    Connecting the 5-pole connector in aversion with CAN communication.

    Pin: Description: Notes:

    1 CAN_SHLD CAN communication2 24 VAC Instrument supply3 0 VAC Instrument supply4 CAN_H CAN communication5 CAN_L CAN communication

    Connecting a 5-pole connector in aversion with passive current output.

    Pin: Description: Notes:

    1 - 4 - 20 mA Current loop ( - )2 24 VAC Instrument supply3 0 VAC Instrument supply4 + 4 - 20 mA Current loop ( + )5 - Free

    8. Current loop

    M1031C analyser (only in current loop execution!) has a passive current loop, whichis galvanically separated. Charge the current loop with direct voltage.

    The voltage drop on the current loop regulator is 8V. For inside protection of thecurrent loop regulator the maximal voltage must not exceed 30V.

    Connector CONNECTION

    L 230VAC L 230 VAC 24 VAC

    +24 VDC0 VDC

    0 VACN 230 VAC

    PeN 230VACPe

    ECN00120

    2

    3 4

    51

    Display,recorder unit,telemetry,...

    NELA0028

    The principle of connecting a rectifier with two voltage outputs to the transformer; 24VAC for charging the instrument and 24 VDC for charging the current loop.

    Connector CONNECTION

    L 230VACN 230VAC

    L 230 VAC 24 VAC0 VAC

    0 VAC19 L19 N

    +V-V

    19.2 VAC

    N 230 VAC

    ECN00120

    IZX00070

    Display,recorder unit,telemetry,...

    NELA0029

    543

    2 1

    The principle of connecting the instrument, ECN00120 transformer with twosecondary clips and a IZX00070 rectifying module for charging the instrument andcharging the current loop.

    9. Current loop table

    CONTROLMATIK ABW d.o.o.Fajfarjeva 15, 1230 Domale, Slovenija, EvropaTel.: 01 721 35 52, Fax.: 01 721 93 60http://www.cm-abw.si, Email: [email protected]

    9

  • 7/30/2019 M1031C_Eng_24.12.02

    10/25

    MEASUREMENT EQUIPMENT

    Current loop table helps us determine the free chlorine level by means of currentlevel of the current loop.

    Current loop QUANTITY OF FREE CHLORINE

    (mA) 0 - 1 mgCl2/l 0 - 2 mgCl2/l 0 - 5 mgCl2/l

    4,0 0,00 0,0 0,004,8 0,05 0,1 0,25

    5,6 0,10 0,2 0,50

    6,4 0,15 0,3 0,75

    7,2 0,20 0,4 1,00

    8,0 0,25 0,5 1,25

    8,8 0,30 0,6 1,50

    9,6 0,35 0,7 1,75

    10,4 0,40 0,8 2,00

    11,2 0,45 0,9 2,25

    12,0 0,50 1,0 2,50

    12,8 0,55 1,1 2,75

    13,6 0,60 1,2 3,0014,4 0,65 1,3 3,25

    15,2 0,70 1,4 3,50

    16,0 0,75 1,5 3,75

    16,8 0,80 1,6 4,00

    17,6 0,85 1,7 4,25

    18,4 0,90 1,8 4,50

    19,2 0,95 1,9 4,75

    20,0 1,00 2,0 5,00

    10. CAN network

    10.1 SeleCAN

    All CAN communication units (PLC control units, sensors, valves) in SELECANversions of communication must have parallel wiring. The maximum number of CANcommunication units connected in this way must not exceed 64.CAN network consists of three signals: CAN H (CAN +), CAN L (CAN -) and CANGND.

    10.2 Segmental circuit (RECOMMENDED!)

    The most recommended wiring is shown in the drawing below. The maincommunication line is divided into segments. CAN communication units A - I (PLCcontrol units, sensors, valves) are connected via CAN distributors 1 - 6.Terminators (T), which are connected to CAN H (CAN +) and CAN L (CAN -), mustbe placed at the beginning and at the end.

    CONTROLMATIK ABW d.o.o.Fajfarjeva 15, 1230 Domale, Slovenija, EvropaTel.: 01 721 35 52, Fax.: 01 721 93 60http://www.cm-abw.si, Email: [email protected]

    10

  • 7/30/2019 M1031C_Eng_24.12.02

    11/25

    MEASUREMENT EQUIPMENT

    Entire CAN network

    CAN NETWORK

    SEGMENT

    1 3 5

    642

    A

    T T

    B C D E F G H INELA0015

    The main CAN network is divided into more segmental networks. When dividing themain CAN network into segmental networks, you must pay attention to the length ofthe segment; the length of the segment depends on data flow velocity. One segmentis marked in the picture above and in the tables on the following page total cablelength can be determined. The tables are stated for cables with 5 ns/m and 10 ns/mtime delay.

    It is evident from the table below that when using a cable with 5ns/m time delay anddata flow velocity of 20Kbit/s, the total segment length can be up to 150 m; the lengthof the cable made of all in this way connected segments must not exceed 750 m.

    The length of the cable, which is used to connect CAN distributors 1 and 2 and allCAN communication units plugged to CAN distributor 2 (PLC control units, sensors,valves), is considered the total segment length.

    Cable with 5 ns/m time delay

    Bitrate Segment Total

    (KB/s) length (m) length (m)

    1000 3 15

    800 4 19

    500 6 30

    250 12 60

    125 24 120

    100 30 150

    50 60 300

    20 150 750

    10 300 1500

    Cable with 10 ns/m time delay

    Bitrate Segment Total

    (KB/s) length (m) length (m)

    1000 2 8

    800 2 9

    500 3 15

    250 6 30

    125 12 60

    100 15 75

    50 30 150

    20 75 375

    10 150 750

    10.3 Terminators

    Terminators are ohm resistors, which are connected to CAN H (CAN +) and CAN L

    CONTROLMATIK ABW d.o.o.Fajfarjeva 15, 1230 Domale, Slovenija, EvropaTel.: 01 721 35 52, Fax.: 01 721 93 60http://www.cm-abw.si, Email: [email protected]

    11

  • 7/30/2019 M1031C_Eng_24.12.02

    12/25

    MEASUREMENT EQUIPMENT

    (CAN -) at the beginning and the end of the CAN network.

    CAN network

    length

    Cable

    resistance

    Cable diameter Terminator

    resistance

    Max. transfer

    0 - 40 m 70 m/m 0.25 mm - 0.34

    mmAWG 23, AWG

    22

    124 1 Mbit pri 40 m

    40 - 300 m

  • 7/30/2019 M1031C_Eng_24.12.02

    13/25

    MEASUREMENT EQUIPMENT

    Points A - F represent CAN communication units (PLC control units, valves...).

    The basic circuit of CAN communication units (NELA0005), used byCONTROLMATIK ABW for connecting sensors (or rather other CAN communicationunits). The circuit that is printed in bold (continuous) represents the first sensor,which is linked to the PLC control unit; the circuit that is printed in dashes (interruptedline) represents the second sensor linked to the PLC control unit and in the sameway the circuit of all following sensors. (or rather other CAN communication units).Terminators (R*) are required at the beginning and the end of the CAN network.

    PLC control proces unit R*

    24 VAC

    0 VAC

    12

    35

    4

    2

    3 4

    NELA0005

    15

    R*

    CAN H ( CAN + )

    CAN L ( CAN - )

    CAN GND

    11. A description of keys and light warnings

    CONTROLMATIK ABW d.o.o.Fajfarjeva 15, 1230 Domale, Slovenija, EvropaTel.: 01 721 35 52, Fax.: 01 721 93 60http://www.cm-abw.si, Email: [email protected]

    13

  • 7/30/2019 M1031C_Eng_24.12.02

    14/25

    MEASUREMENT EQUIPMENT

    CAL DIAG

    ERR CAN

    EDIT ESC

    There are a display, four keys andthree light signals on the sensor, toenable the user easy communicationwith the unit, such as check

    measurements or check and changethe settings.

    11.1 When the EDIT light is OFF

    CALTo enter the CALIBRATION menu,press CAL key. This menu is used forcalibrations (settings).

    DIAGTo enter the DIAGNOSTICS menu,press DIAG key.

    By pressing the MENU key you canenter the menu.

    ESCTo exit the current menu or function,press the ESC key. The changes willnot be saved.

    11.2 When the EDIT light is ON

    CALPress this key to move up on the menuor to increase value.

    DIAGPress this key to move down on themenu or to decrease value.

    By pressing the CONFIRMATION keyyou can confirm the setting.

    ESC Press the ESC key to undo thesettings. The changes will not besaved.

    The EDIT, ERR and CAN light signals are hidden under a dark foil and are onlyvisible when active.

    EDIT The EDIT light signal is activated when we are using a settingmenu, where certain parameters can be changed.

    ERR The ERR light signal activates when a system error occurs.

    CONTROLMATIK ABW d.o.o.Fajfarjeva 15, 1230 Domale, Slovenija, EvropaTel.: 01 721 35 52, Fax.: 01 721 93 60http://www.cm-abw.si, Email: [email protected]

    14

  • 7/30/2019 M1031C_Eng_24.12.02

    15/25

    MEASUREMENT EQUIPMENT

    CAN The short signal flashes when the sensor is transmitting information(only if the sensor is connected to the CAN network).

    12. A description of the display and its functions

    At start-up, the display shows the basic data, type of measuring instrument,measuring value and unit, reference point value (if the instrument is operating as acontroller) and tempereture of the sample water.

    CHLOR (HOCl)

    REF: 0.35 10.5C0.00 mg/l

    Reference pointvalue

    Measuring valueand unit

    Type of measuringinstrument

    Temperature ofsample water

    13. Before starting

    User access is divided to different levels. Access to a certain level is only possiblewith a password. These passwords disable unauthorised person to change thesettings. There are three user levels in general:

    LOOK The first level is only intended for measurement readings. Nothing canbe changed at this level. No password is necessary.

    NORMAL The second level is intended for persons who are acquainted with theworking of the instrument and are qualified to operate it. Measurementscan be checked; primary settings (measuring, changing the reference,setting the current loop) can be changed at this level. The passwordis CCCC.

    ADMIN The third level also includes servicing diagnostics and is only intendedfor authorised specialist working at CONTROLMATIK ABW.

    14. Entering the password

    IMPORTANT:

    NEL00001

    CHLOR (HOCl)

    10.5C0.00 mg/l

    When connecting M1031C analyser to the feedingvoltage the basic screening shows up on the display.Access to all functions and settings is described at thislevel. When you are in one menu, you can alwaysreturn to the basic screening by pressing ESC key(when in one of the sub-menus by pressing ESC keymore times).

    CONTROLMATIK ABW d.o.o.Fajfarjeva 15, 1230 Domale, Slovenija, EvropaTel.: 01 721 35 52, Fax.: 01 721 93 60http://www.cm-abw.si, Email: [email protected]

    15

  • 7/30/2019 M1031C_Eng_24.12.02

    16/25

    MEASUREMENT EQUIPMENT

    NEL00002

    MAIN MENU

    -> PASSWORD

    To access the MAIN MENU, press CONFIRMATIONkey. The only alternative now is to enter the password.Confirm the password once more (pressCONFIRMATION key).

    NEL00003

    PASSWORD

    USER NAME

    PASSWORD

    :

    :

    LOOK

    AAAA

    Various levels were described in chapter XX BEFORE

    STARTING. The user can change the settings at levelNORMAL; the password is CCCC. The assumed levelis LOOK and is blinking in our case.

    NEL00004

    PASSWORDUSER NAMEPASSWORD

    ::

    NORMALAAAA

    To switch the level to NORMAL, press the CAL key. Toswitch the level to ADMIN, press the CAL key twice. Insuch case, we can press DIAG key to return to levelNORMAL. Press DIAG key once more and you come tothe third level option LOOK. Confirm the NORMALlevel (press CONFIRMATION key) and the first

    password sign starts blinking in the field for entering thepassword.

    NEL00006

    PASSWORDUSER NAMEPASSWORD

    ::

    NORMALCAAA

    If you press the CAL key, the first password signchanges into B. If you press the CAL key once more,the first sign changes into C (if you press DIAG key thesign changes into A again). Confirm the chosen sign C(press CONFIRMATION key).

    NEL00007

    PASSWORD

    USER NAME

    PASSWORD

    :

    :

    NORMAL

    CBAA

    15. Setting the free chlorine level

    NEL00001

    CHLOR (HOCl)

    10.5C0.00 mg/l

    NEL00021

    CALIBRATION

    -> SPAIN

    NEL00022

    SPAIN

    -018.23uA 10.5C0.55 mg/l

    The setting can only be

    changed if you have selected the NORMAL user levelbefore. Press the CAL key on the basic screening andenter the CALIBRATION menu.

    In the CALIBRATION menu you only have access toSPAIN function. The SPAIN function is only intended for

    setting the free chlorine level. Press CONFIRMATIONkey.

    In that case, another data shows up on the lower leftcorner of the display, telling the current level on themeasuring cell. The value of the current is given in uA(micro amperes).

    15.1 What you need to know:

    The zero is factory set and is not calibrated, but we recommend that you check the

    CONTROLMATIK ABW d.o.o.Fajfarjeva 15, 1230 Domale, Slovenija, EvropaTel.: 01 721 35 52, Fax.: 01 721 93 60http://www.cm-abw.si, Email: [email protected]

    16

  • 7/30/2019 M1031C_Eng_24.12.02

    17/25

    MEASUREMENT EQUIPMENT

    setting before the first calibration and also that you conduct a check-up from time totime.

    15.2 How to do it:

    You need the MXXX charcoal filter for this procedure. Unscrew the sample input atthe M1031C analyser and connect it through the MXXX charcoal filter. Pay attentionto the proper flow direction of the sample water running through the beforementioned filter. We need some additional pressure at the beginning, to keep thesample water running through the MXXX charcoal filter (the flow through the M1031Canalyser should be even). Keep the sample running through the MXXX charcoal filterfor at least ten minutes. The measured value should not exceed 0.05 mgCl2/l. In casethe value exceeds 0.05 mgCl2/l, clean the charcoal filter as well as the input filter onthe M1031C analyser.In case that the value is still above 0.05 mgCl2/l, please, consult with the authorisedservice of the CONTROLMATIK ABW.

    15.3 Setting the value:

    Before calibrating for the first time, we must keep the M1031C analyser running for atleast 24 hours. Because of the various water structures, the analyser works properlyonly when the measuring cell has oxidised correctly. Then we can set the analyserand keep it running for a week and then we calibrate it again. Keep the set analyserrunning and calibrate after a week. Always repeat the procedure after the measuringcell has been serviced and in case the M1031C analyser has not been running formore than a week.When you are calibrating the analyser this time, we recommend that you take notesof the values at specific levels. The data will be of great help when you are calibratingthe M1031C analyser later on.

    16. Setting free chlorine

    NEL00022

    SPAIN

    -018.23uA 10.5C0.63 mg/l

    NEL00023

    SPAIN

    -018.23uA 10.5C0.50 mg/l

    Press CONFIRMATION key in the SPAIN menu.

    The value of the measurement starts blinking and asignal appears in front of the value, telling you that youcan either increase or decrease the value. You can

    change the value by pressing CAL keys (up) or DIAGkeys (down). If you hold the key for a while, the valueincreases / decreases automatically. After you havechosen the appropriate value (0.50 mg/l in our case),press CONFIRMATION key.

    17. Setting the regulator

    CONTROLMATIK ABW d.o.o.Fajfarjeva 15, 1230 Domale, Slovenija, EvropaTel.: 01 721 35 52, Fax.: 01 721 93 60http://www.cm-abw.si, Email: [email protected]

    17

  • 7/30/2019 M1031C_Eng_24.12.02

    18/25

    MEASUREMENT EQUIPMENT

    NEL00025

    CHLOR (HOCl)

    REF: 0.35 10.5C0.00 mg/l

    NEL00010

    MAIN MENU

    -> PASSWORD

    REGULATOR

    TEMPERATUR

    NEL00027

    MAIN MENU

    -> REGULATOR

    TEMPERATUR

    AN.OUT 1

    NEL00028

    REGULATOR

    -> REF.

    Setting the regulator is

    only possible in executions that enable that. Thereference point of the regulator can be set in suchexecutions. In such cases, the reference point value isvisible in the left bottom corner in theMEASUREMENTscreening.

    Press CONFIRMATION key in the basic screeningMEASUREMENTS and access the MAIN MENU.

    NEL00027

    The sign ->, next to the description of the function(->PASSWORD), indicates, which function we willselect by CONFIRMATION (always the function that iswritten at the top).

    In this case you wish to enter the REGULATOR menu,therefore you do not confirm the PASSWORD function,

    but press the CAL key (up) to shift to other functions.Confirm the appropriate function ->REGULATOR (pressCONFIRMATION key).

    You can change the regulators reference point REF inthe REGULATOR menu. Confirm the selected menu(press the CONFIRMATION key).

    REGULATOR

    REF.

    -> 0.35 mg/l

    NEL00029

    The menu for manual setting of the regulatorsreference point shows up on the display. Press theCONFIRMATION key and the value will start blinking.

    You can change the value at this time by pressing theCAL key (up) or the DIAG key (down). If you wish toconfirm and save the new value, press theCONFIRMATION key, if not, press ESC. If you pressthe ESC key again, you exit the submenu. If you pressthe ESC key repeatedly, you exit all menus.

    18. Setting the temperature

    NEL00001

    CHLOR (HOCl)

    10.5C0.00 mg/l

    You can only set the temperature in executions that

    enable that. These are executions that come withoutautomatic temperature compensation and you have toset the temperature pattern yourself.

    NEL00011

    MAIN MENU

    -> TEMPERATUR

    -> AN. OUT 1

    -> SYSTEMNEL00012

    TEMPERATUR

    -> MANUAL VALUE

    CONTROLMATIK ABW d.o.o.Fajfarjeva 15, 1230 Domale, Slovenija, EvropaTel.: 01 721 35 52, Fax.: 01 721 93 60http://www.cm-abw.si, Email: [email protected]

    18

  • 7/30/2019 M1031C_Eng_24.12.02

    19/25

    MEASUREMENT EQUIPMENT

    NEL00013

    TEMPERATUR

    MANUAL VALUE

    -> 00.0 C

    Press CONFIRMATIONkey in the basic screeningmeasurements andaccess the MAIN MENU.

    Press the CAL key (up) so

    many times that theTEMPERATURE menureaches the top place of

    the MAIN MENU listing. At this point you also have asign next to the menu, indicating the selected menu ->TEMPERATURE. Then press CONFIRMATION key.

    You can only change the MANUAL VALUE in the

    selected menu. Confirm the selected menu (press theCONFIRMATION key).

    The menu for manual setting of the temperature showsup on the display. Press the CONFIRMATION key andthe value will start blinking. You can change the valueat this time by pressing the CAL key (up) or the DIAGkey (down). If you wish to confirm and save the newvalue, press the CONFIRMATION key, if not, press

    ESC. If you press the ESC key again, you exit thesubmenu. If you press the ESC key repeatedly, you exitall menus.

    19. Setting the current loop

    NEL00001

    CHLOR (HOCl)

    10.5C0.00 mg/l

    NEL00014

    MAIN MENU

    -> AN.OUT 1

    SYSTEM

    NEL00015

    AN.OUT 1

    -> MAXIMUM

    MINIMUM

    You can only set the current loop in executions thatenable that. You can set the minimal and maximal levelof the 4 - 20mA current loop.

    Press CONFIRMATION key in the basic screening

    MEASUREMENTS and access the main menu (MAINMENU).

    Press the CAL key (up) so many times that the AN.OUTmenu reaches the top place of the MAIN MENU listing.At this point you also have a sign next to the menu,indicating the selected menu -> AN.OUT 1. Then pressCONFIRMATION key.

    You can alter the MAXIMUM and MINIMUM in theselected menu. MAXIMUM (->MAXIMUM) is the firstselected menu. Press CONFIRMATION key.

    NEL00016

    AN.OUT 1

    MAXIMUM

    -> 1.00 mg/l

    NEL00017

    AN.OUT 1

    -> MINIMUM

    NEL00018

    AN.OUT 1

    MINIMUM

    -> 0.00 mg/l

    CONTROLMATIK ABW d.o.o.Fajfarjeva 15, 1230 Domale, Slovenija, EvropaTel.: 01 721 35 52, Fax.: 01 721 93 60http://www.cm-abw.si, Email: [email protected]

    19

  • 7/30/2019 M1031C_Eng_24.12.02

    20/25

    MEASUREMENT EQUIPMENT

    You can set the value offree chlorine, at which the

    current loop shall have amaximal deflection of 20mA. You can change thevalue by pressing theCONFIRMATION key.The value will startblinking and you canchange it at this time bypressing the CAL key (up)or the DIAG key (down). Ifyou wish to confirm and

    save the new value, pressthe CONFIRMATION key,

    if not, press ESC. Then press the ESC key again to exitthe menu.

    Press the CAL key (up) to access the selected menuMINIMUM (->MINIMUM) and set the minimal deflection

    of the current loop. Press CONFIRMATION key.

    You can set the value of free chlorine, at which thecurrent loop shall have a minimal deflection of 4 mA.You can change the value by pressing theCONFIRMATION key. The value will start blinking andyou can change it at this time by pressing the CAL key(up) or the DIAG key (down). If you wish to confirm andsave the new value, press the CONFIRMATION key, ifnot, press ESC. To exit the submenu, press the ESC

    key once; to exit all menus, press the ESC keyrepeatedly.

    20. Restart

    NEL00001

    CHLOR (HOCl)

    10.5C0.00 mg/l

    NEL00019

    MAIN MENU

    -> SYSTEM

    NEL00020

    SYSTEM

    -> RESTART

    In case you have any trouble, we recommend that yourestart the measuring instrument. You can do that if youaccess the special menu for restart RESTART(RESET).

    Press CONFIRMATION key in the basic screening

    MEASUREMENTS and access the main menu (MAINMENU).

    Press the CAL key (up) so many times that theSYSTEM menu reaches the top place of the MAINMENU listing. At this point you also have a sign next tothe menu, indicating the selected menu ->SYSTEM.Then press CONFIRMATION key.

    You can only choose RESTART in the selected menu.If you press the CONFIRMATION key, you restart the

    instrument (RESET).

    SW: 00.43

    M-1031

    CONTROLMATIK ABW

    NEL00051

    When you start or re-start the measuring instrument,the execution of the measuring instrument (M1031,M1121, M1321) and the software series as well asthe name of the manufacturer (SW:00.43) show up onthe display.

    21. Cleaning

    Analyser M1031C can be cleaned with a wet rag. Do not use any kind of chemicalsfor analyser cleaning .In case that water stone accumulates on the analyser, pleaseconsult the authorized service of Controlmatik ABW company.

    CONTROLMATIK ABW d.o.o.Fajfarjeva 15, 1230 Domale, Slovenija, EvropaTel.: 01 721 35 52, Fax.: 01 721 93 60http://www.cm-abw.si, Email: [email protected]

    20

  • 7/30/2019 M1031C_Eng_24.12.02

    21/25

    MEASUREMENT EQUIPMENT

    21.1 Cleaning of the measurement cell

    Cleaning of the measurement cell can be done only by authorized service ofControlmatik ABW company.

    21.2 Cleaning the inlet filter

    Clean the inlet filter when needed. Unscrew the screwof the inlet filter and pull it out together with the filternet.

    NELA0017

    Filter net can be cleaned mechanically or by usingchemicals. After chemical cleaning wash it verythoroughly with a streaming water.In case that the filter net is damaged it has to bereplaced with a new one.

    22. The menu strukture

    The Menu structure includes the main(MAIN MENU), where we set all systemsettings(regulator settings, temperature settings, current loop setting,...) and acalibration menu(CALIBRATION). Menus of all possible versions of the M1031CAnalyser are described in the structure.You enter the main menu (MAIN MENU) by pressing the MENU/CONFIRMATIONkey.

    You enter the calibration menu (CALIBRATION) by pressing the CAL key.

    MAIN MENU- PASSWORD

    CONTROLMATIK ABW d.o.o.Fajfarjeva 15, 1230 Domale, Slovenija, EvropaTel.: 01 721 35 52, Fax.: 01 721 93 60http://www.cm-abw.si, Email: [email protected]

    21

  • 7/30/2019 M1031C_Eng_24.12.02

    22/25

    MEASUREMENT EQUIPMENT

    - REGULATOR- REF.

    - TEMPERATUR- MANUAL VALUE

    - AN.OUT 1

    - MAXIMUM- MINIMUM

    - SYSTEM- RESTART

    CALIBRATION

    23. Diagnostics

    NEL00024

    001

    25.3C

    1904.9E

    018.23uA

    :

    :

    :

    :

    SENZOR-I

    SENZOR-R

    INT. TEMP.

    CAN ADDR.

    SENZOR-I:

    SENZOR-R:

    CAN ADDR.:

    INT. TEMP.:

    Press the DIAG key to enter the diagnostic menu. Youcan not change anything here, it is only meant to check

    the basic functions and settings.

    Current value, that is generated by the measuring celltelling the current level on the measuring cell is given inuA (micro amperes).

    Ohms resistance of the temperature sensor. Thissensor is needed in a version with automatictemperature compensation. The value is given in ohms

    (1902.3E).

    Electronic device adress, that is needed foridentification in CAN network.

    Measurement of sensors internal temperature. Itcontrols the temperature by which the device isoperating. The operating temperature has to bebetween 10C and 50C.

    24. Dimensions

    When choosing a place to mount the M1031C Analyser, besides the defineddimensions we have to consider the space needed needed for water sample inletpipe, analysers sink and power supply.

    M1031C analysers height is 11cm.

    CONTROLMATIK ABW d.o.o.Fajfarjeva 15, 1230 Domale, Slovenija, EvropaTel.: 01 721 35 52, Fax.: 01 721 93 60http://www.cm-abw.si, Email: [email protected]

    22

  • 7/30/2019 M1031C_Eng_24.12.02

    23/25

    MEASUREMENT EQUIPMENT

    450mm 500mm 540mm

    300 mm

    340 mm

    25. Technical data

    Measuring range: 0 - 1; 0 - 2 mgCl2/lDisplay: LCD grafic display 61mm * 22mm (120 * 32 points)Resolution: 0.01 mgCl2/lAccuracy: 5%Charging: 24 VAC (+/- 10%) 50 - 60HzConsumption: 5 VA (at 24 VAC)Sample water consumption: 0.7l/min

    CONTROLMATIK ABW d.o.o.Fajfarjeva 15, 1230 Domale, Slovenija, EvropaTel.: 01 721 35 52, Fax.: 01 721 93 60http://www.cm-abw.si, Email: [email protected]

    23

  • 7/30/2019 M1031C_Eng_24.12.02

    24/25

    MEASUREMENT EQUIPMENT

    Cell:Measuring electrode GoldAnode Copper

    Measurement response: 8 sec.Communication:

    Passive current output 4 - 20mA (DC voltage drop on current loop regulatoris 8V, current loop maximal voltage is 30V).SELECAN communication protocol

    Temperature sensor: KPYMeasuring sensor protection: IP65Weight: 4KgPACKAGING:Package weight: 4.5KgPackage dimensions:

    Width: 320 mmLength: 470 mm

    Height: 110 mm

    26. M1031C Analyser Codes

    M 1031 C / A C 1

    Measuring range: 1: 0 - 2 mg/Cl2/l

    Output: C: Passive current output 4 - 20 mA **D: Digital communication SELECAN

    CONTROLMATIK ABW d.o.o.Fajfarjeva 15, 1230 Domale, Slovenija, EvropaTel.: 01 721 35 52, Fax.: 01 721 93 60http://www.cm-abw.si, Email: [email protected]

    24

  • 7/30/2019 M1031C_Eng_24.12.02

    25/25

    MEASUREMENT EQUIPMENT

    Temp. comp. A: Automatic temp. compensation

    Standard versions of M1031C Analyser:

    M 1031C/ A C 1Automatic temperature compensationPassive current output 4 - 20mAMeasuring range 0 - 2mgCl2/lMeasuring cell

    M1031C/ A C 1 SAutomatic temperature compensationPassive current output 4 - 20mAMeasuring range 0 -2mgCl2/l

    Measuring cell with silver referential electrode Ag/AgCl

    M1031C/ A D 1Automatic temperature compensationDigital output - SELECAN communicationMeasuring range 0 - 2mgCl2/lMeasuring cell

    Note:

    ** Voltage drop on the current loop regulator is 8V. Because of the inner protectionin current loop regulator, maximal voltage can not exceed 30V.

    CONTROLMATIK ABW d o o 25