43
Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical Center

Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Embed Size (px)

Citation preview

Page 1: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Medical Linear Accelerators in Radiation Therapy

Haijun Song, Ph.D.

Dept. of Radiation OncologyDuke University Medical Center

Page 2: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

MV Linear Accelerator

Page 3: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Anatomy of a Linac

Page 4: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Anatomy of a Linac

Page 5: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical
Page 6: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Beam MofiersCone

MLC (Multi-Leaf Collimator)

Page 7: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Radiosurgery with Linac

Page 8: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

MLC

Page 9: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Conformal RT (3DCRT) &Intensity Modulation RT (IMRT)

Page 10: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Conformal RT (3DCRT) &Intensity Modulation RT (IMRT)

Page 11: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Dose Map delivered by IMRT

Page 12: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Dose Map delivered by IMRT

Page 13: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Dose Map delivered by IMRT

Page 14: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Image Guided RT (IGRT)

Page 15: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Respiratory Gating RT

Page 16: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Tomotherapy and Robotic

Page 17: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Difference among Radiation Sources

Factors affecting radio-bio response• Gamma vs. Electron

Among Gamma sources• Photon Energy Spectrum• Dose Distribution• Dose Rate

Page 18: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Radioactive IsotopesExample: Cs-137

• single dominant energy, 0.66 MeV• Activity: 6000 Ci and 100 Ci, dual source• Dose rate range: 0.001 - 10 Gy/min

Page 19: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Radioactive Isotopes

Example: Ir-192• Complex energy spectrum: 30 gammas and

40 x rays.• Dose rate, 0.01 - 5 Gy/min being used RT.

Ir-192 photon spectrum

0.0001

0.001

0.01

0.1

1

10

100

1000

0.001 0.01 0.1 1 10

E (MeV)

Perc

ent p

er d

ecay

Page 20: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

From kV to MV

Commonly: 50 – 300 kVPBeam Quality: HVL a few mm Al to Cu.Dose Rate: around Gy/min

Page 21: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

From kV to MV

Commonly: 6 MV -25 MVBeam quality: PDD10 in water, 67% and up.Dose Rate: around Gy/min

Page 22: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

X-Ray tube vs. Linac

Page 23: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Acce. Tube: Travelling Wave

Page 24: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

X-Ray ProductionSide emission vs. straight through emmission

Page 25: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical
Page 26: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Photon energy spectrum from a 6 MV beam

Fl uence/ MeV/ I ndi cent el ect r on

0. 00E+00

2. 00E- 05

4. 00E- 05

6. 00E- 05

8. 00E- 05

1. 00E- 04

1. 20E- 04

1. 40E- 04

0 1 2 3 4 5 6

Fl uence/ MeV/ I ndi centel ect r on

Page 27: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Depth Dose Curves

Page 28: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Dose Distribution: One beam/parallel beams

Page 29: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Dose Distribution: One beam/parallel beams

Page 30: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

MV photon and electron beams

Page 31: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Depth Dose of e Beams

0 1 2 3 4 5 6 7 8 90

1 0

2 0

3 0

4 0

5 0

6 0

7 0

8 0

9 0

1 0 0

M e V1 81 51 21 086

Per

cent

Dep

th Io

niza

tion

D e p t h ( c m ) i n W a t e r

Page 32: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Dose Measurement

Ion chamber and electrometerIn water for MV photon or electron beamsIn air for KV beamsTracable to NIST Co-60 standard beam

CowDQ

Qw NkMD 60

,*=

Page 33: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Buildup region and skin sparingCompton e1 1 1 1

Compton e2 1 1 1

Compton e3 1 1

Compton e4 1

Depth in water

Total dose 1 2 3 3

Page 34: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Treating Superficial targetsBolus: tissue equivalent, soft to conform to skin

0 1 2 3 40

20

40

60

80

100

S

B dmax

Per

cent

Dep

th D

ose

Depth in Water (cm)

Page 35: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Dose Distribution: One beam/parallel beams

Page 36: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Buildup region and skin sparingcontinued

Skin sparing also means a target at shallow depth will not get adequate dose.

For small animals, “sparing” of a couple of centimeter from skin may “spare” the whole animal.

What do we do when we need to treat superficial target?

Page 37: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Bolus

Page 38: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Treating Superficial targetsBolus: tissue equivalent, soft to conform to skin

0 1 2 3 40

20

40

60

80

100

S

B dmax

Per

cent

Dep

th D

ose

Depth in Water (cm)

bolus

Page 39: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Treating Superficial targetsBolus: tissue equivalent, soft to conform to skin

0 1 2 3 40

20

40

60

80

100

S

B dmax

Per

cent

Dep

th D

ose

Depth in Water (cm)

bolus

Page 40: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Treating Superficial targets

MeV range electron beams /w bolus

kV (suferfical units, orthovoltage units)

MV photon beam with bolus

How useful are MeV electron beams for small animal irradiations?

Page 41: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

What can Rad Onc Offer Rad Biology

All offered to human can be offered to other species.

Imaging: CT, MRI

Precise and conformal dose coverage: 3DCRT, IMRT, IGRT, Respiratory Gating, radiosurgery, electron irradiation.

Precision and conformality means sparing healthy tissue, too.

And More: Brachytherapy, hyperthermia, proton, neutron.

Page 42: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Acce. Tube: Standing Wave

Page 43: Medical Linear Accelerators in Radiation Therapy - RadCCORE · Medical Linear Accelerators in Radiation Therapy Haijun Song, Ph.D. Dept. of Radiation Oncology Duke University Medical

Acce. Tube: Short Standing Wave