1
Micro-CT Imaging of an Insulin Bolus Infusion Through Continuous Subcutaneous Insulin Infusion (CSII) Sets in Live Swine: Quantitative Comparison of Distribution Patterns in Tissue Over Eight Days. David Diaz Ph.D. 1 , Aleksandr Dinesen M.S. 1 , Abdurizzagh Khalf 1 , Gabriella Eisler 1 , Channy Loeum 1 , Mathew Thakur Ph.D. 1 , Marc C Torjman Ph.D. 1 , Kenneth C Hsu 2 , Paul J Strasma 2 , Jeffrey I Joseph D.O. 1 1 Jefferson Artificial Pancreas Center, Department of Anesthesiology, Sidney Kimmel Medical College at Thomas Jefferson University, Philadelphia, PA; 2 Capillary Biomedical, Inc., Irvine, CA INTRODUCTION DISCUSSION RESULTS METHODS • Over 1 million people with diabetes use insulin pumps and insert a new continuous subcutaneous insulin infusion (CSII) set every 2–3 days for adequate blood glucose control. • Insulin absorption from the subcutaneous tissue is slow and varies over time leading to complications such as hyperglycemia, hypogly- cemia and glycemic variability. • Repeated trauma of insertion contributes to scar tissue formation and may result in infusion site loss. • Cannula kinking, cannula blockage, and leakage to the skin surface are reported as common infusion set problems. • Previous studies from this group identified a layer of inflammatory tissue surrounding implanted CSII cannulas which becomes thicker, denser, and more continuous as duration of implant increases. This layer may function as a mechanical barrier, slowing or inhibiting the movement of insulin into adjacent subcutaneous tissue containing capillary and lymph vessels. • A prototype kink-resistant infusion set cannula was developed using a soft polymer material with multiple ports and compared against commercially available single-port infusion sets in ambulatory large swine. • Using micro-computer-tomography (micro-CT) we tested the hypothesis that multiple-ports would increase fluid dispersion from a CSII cannula, potentially allowing the insulin to reach more functional capillary and lymph vessels. • Compared with the commercial CSII set, the investigational CSII set with a soft-flexible polymer cannula, wire-reinforced wall and multiple ports produced significantly greater volume and surface area of insulin distribution into adjacent vascular tissue. Research sponsored by Capillary Biomedical, Inc. and supported by JDRF award 2-IND-2016-232-M-X • Investigational multi-port cannulas demonstrated a larger surface area (p=0.018) and 3D volume (p=0.021) of contrast agent spread into the subcutaneous tissue compared to PTFE cannulas with a single distal orifice (see Figure 1). • No differences were observed in leak events (p=0.416) between commercial and investigational sets when evaluating cannulas indwelling for 2 to 8 days (see Figure 2). • Investigational cannulas had a significantly lower kinking rate (p=0.035) than commercial PTFE cannulas (0% vs 18% respectively). • None of the investigational CSII sets with the coil-reinforced polymer cannulas and multiple ports exhibited a kink. • The 70 µL bolus of insulin and x-ray contrast agent traveled from the commercial PTFE cannula’s single distal port and the investigational cannula’s multiple ports into adjacent connective tissue septa along the path of least resistance (see Figure 3). • The insulin/contrast bolus distended the connective tissue fibers to form a spherical distribution in commercial infusion sets and a cylindrical or cone distribution with a few “finger-like” projections for investigational catheters. • Many of the bolus infusions for PTFE cannulas were associated with high peak tubing pressures (2,000 to 5,000 mm Hg) that remained elevated after completion of the bolus, despite having no evidence of infusion set failure. • Commercial CSII infusion sets with a 6mm polytetrafluoroethylene (PTFE) cannula (Inset, Unomedical) and investigational sets with a 14 mm coil-reinforced polymer (CRP) cannula and 4 ports (one distal and 3 proximal ports set 2 mm apart in a helical pattern) were inserted in the soft abdominal tissue of ambulatory swine (n=8, 47 sets in total). Infusion sets were explanted at 10 minutes, 2 days, 4 days, 6 days, and 8 days post-implantation. • Insulin lispro (U-5) was continuously infused through each CSII sets (7 µL/hour) using an insulin pump with an additional 70 µL bolus one or two times per day with meals. • On the final study date, a 70 µL bolus of insulin lispro (U-100) and x-ray contrast agent (Isovue 300 Iopmidol) was infused through each CSII set. 0 Days 2 Days 6 Days 4 Days Capillary Biomedical Coil-Reinforced Soft Polymer Cannula Design Commercially Available CSII Set with PTFE Cannula 8 Days • PK/PD studies in swine and humans are currently underway to determine whether this greater volume and surface area of distribution is able to reach more functional capillary and lymph vessels leading to faster onset/offset and more consistent insulin absorption from dose-to dose. • The soft, flexible polymer wire-reinforced cannula eliminated kinking, a common early failure mode of commercial CSII sets with a PTFE cannula. • Insulin/x-ray contrast agent leaked onto the skin surface, along the path of least resistance in approximately 10-25% of the bolus infusions. • Insulin/x-ray contrast spreads efficiently into adjacent subcutaneous tissue immediately after CSII cannula insertion (0 days). Distribution into adjacent vascular tissue on days 2-8 may be slowed and inhibited by the surrounding layer of inflammatory tissue. Figure 3. Representative longitudinal sequence of three-dimensional micro-CT images illustrating the extent of insulin/contrast dispersion into the subcutaneous tissue surrounding investigational CSII sets with a coil-reinforced polymer cannula (top) and PTFE cannulas with a single orifice (bottom) indwelling for 10 minutes, 2 days, 4 days, 6 days, and 8 days (left to right). Figure 1. Evaluation of 3D surface area and volume dispersion measurements from a 70 µL bolus of insulin/x-ray contrast agent mixture infused through commercial CSII sets with a PTFE cannula and a single distal orifice versus investigational CSII sets with a coil-reinforced soft flexible polymer (CRP) and four ports along the cannula shaft. Significantly higher 3D surface areas and volumes were achieved with the CRP cannulas. Figure 2. Evaluation of infusion set leaks and cannula kinks from micro-CT analysis. The number of infusion sets with kinks was significantly lower (p=0.035) for the investigational CSII sets with a coil-reinforced polymer cannula compared to the commercial CSII sets with a PTFE cannula. The number of sets with leaks onto the skin surface was lower for investigational sets, although not statistically significantly (p=0.416). • Pressure was recorded during the bolus using an inline transducer (PendoTECH). • Tissue was excised 5 minutes after bolus infusion and frozen for micro-CT imaging. • Specimens were imaged using an Inveon micro-PET/CT imaging scanner (20 µm resolution) to determine the distribution pattern of insulin/x-ray contrast agent. • AMIRA 3D visualization software (FEI Company) was used to measure the surface area and volume of spread of insulin/contrast agent bolus into subcutaneous tissue. • Failure mode analysis was performed to evaluate the performance of each infusion set group related to cannula kinks and skin surface leakages using the micro-CT images. • Micro-CT imaging data were analyzed using ANOVA and categorical data using the Fisher ExactTest performed with the Systat software (ver. 13). 0 50 100 150 200 250 300 350 Mean ± SEM CRP PTFE * p=0.018 vs PTFE + p=0.021 vs PTFE * + Surface area (mm 2 ) Volume (mm 3 ) 0 10 20 30 40 50 60 70 80 90 100 No Leak Leak s i s y l a n A T C o r c i M r e p s k a e L Cannula (% of Cannulas) CRP PTFE p=0.416 (CRP vs PTFE) 0 20 40 60 80 100 No Kink Kinked s i s y l a n A T C o r c i M r e p s k n i K Cannula (% of Cannulas) CRP PTFE *p=0.035 * • Micro-CT images of infusions with high internal pressures often revealed less than 70 µL of insulin/x-ray contrast agent delivered into the subcutaneous tissue, despite having no insulin pump occlusion alarm. In clinical practice, a diabetic patient may believe a full bolus of insulin was delivered, when a significantly lower volume of insulin would have actually been delivered into the subcutaneous tissue. • This variability of actual insulin delivery into the subcutaneous tissue may be a major cause of variable insulin PK/PD in patients managing their diabetes in the real-world. Results from this research have guided the design and development of a next-generation insulin infusion set. • Clinical translation of this investigational CSII set could result in a more reliable insulin delivery system that significantly improves blood glucose control, simplifies/reduces the burden of diabetes management, improves the quality of life of people with diabetes and benefits clinical outcomes.

Micro-CT Imaging of an Insulin Bolus Infusion …...Micro-CT Imaging of an Insulin Bolus Infusion Through Continuous Subcutaneous Insulin Infusion (CSII) Sets in Live Swine: Quantitative

  • Upload
    others

  • View
    9

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Micro-CT Imaging of an Insulin Bolus Infusion …...Micro-CT Imaging of an Insulin Bolus Infusion Through Continuous Subcutaneous Insulin Infusion (CSII) Sets in Live Swine: Quantitative

Micro-CT Imaging of an Insulin Bolus Infusion Through Continuous Subcutaneous Insulin Infusion (CSII) Sets in Live Swine: Quantitative Comparison of Distribution Patterns in Tissue Over Eight Days.

David Diaz Ph.D.1, Aleksandr Dinesen M.S.1, Abdurizzagh Khalf 1, Gabriella Eisler1, Channy Loeum1, Mathew Thakur Ph.D.1, Marc C Torjman Ph.D.1,Kenneth C Hsu2, Paul J Strasma2, Jeffrey I Joseph D.O.1

1 Jefferson Artificial Pancreas Center, Department of Anesthesiology, Sidney Kimmel Medical College at Thomas Jefferson University, Philadelphia, PA; 2Capillary Biomedical, Inc., Irvine, CA

INTRODUCTION

DISCUSSION

RESULTS

METHODS

• Over 1 million people with diabetes use insulin pumps and insert a new continuous subcutaneous insulin infusion (CSII) set every 2–3 days for adequate blood glucose control.

• Insulin absorption from the subcutaneous tissue is slow and varies over time leading to complications such as hyperglycemia, hypogly-cemia and glycemic variability.

• Repeated trauma of insertion contributes to scar tissue formation and may result in infusion site loss.

• Cannula kinking, cannula blockage, and leakage to the skin surface are reported as common infusion set problems.

•Previousstudiesfromthisgroupidentifiedalayerofinflammatorytissue surrounding implanted CSII cannulas which becomes thicker, denser, and more continuous as duration of implant increases. This layer may function as a mechanical barrier, slowing or inhibiting the movement of insulin into adjacent subcutaneous tissue containing capillary and lymph vessels.

• A prototype kink-resistant infusion set cannula was developed using a soft polymer material with multiple ports and compared against commercially available single-port infusion sets in ambulatory large swine.

• Using micro-computer-tomography (micro-CT) we tested the hypothesisthatmultiple-portswouldincreasefluiddispersionfroma CSII cannula, potentially allowing the insulin to reach more functional capillary and lymph vessels.

• Compared with the commercial CSII set, the investigational CSII set withasoft-flexiblepolymercannula,wire-reinforcedwalland multipleportsproducedsignificantlygreatervolumeandsurfacearea of insulin distribution into adjacent vascular tissue.

Research sponsored by Capillary Biomedical, Inc. and supported by JDRF award 2-IND-2016-232-M-X

• Investigational multi-port cannulas demonstrated a larger surface area (p=0.018) and 3D volume (p=0.021) of contrast agent spread into the subcutaneous tissue compared to PTFE cannulas withasingledistalorifice(seeFigure1).

• No differences were observed in leak events (p=0.416) between commercial and investigational sets when evaluating cannulas indwelling for 2 to 8 days (see Figure 2).

•Investigationalcannulashadasignificantlylowerkinkingrate(p=0.035) than commercial PTFE cannulas (0% vs 18% respectively).

• None of the investigational CSII sets with the coil-reinforced polymercannulasandmultipleportsexhibitedakink.

•The70µLbolusofinsulinandx-raycontrastagenttraveledfrom the commercial PTFE cannula’s single distal port and the investigational cannula’s multiple ports into adjacent connective tissue septa along the path of least resistance (see Figure 3).

•Theinsulin/contrastbolusdistendedtheconnectivetissuefibersto form a spherical distribution in commercial infusion sets and a cylindricalorconedistributionwithafew“finger-like”projectionsfor investigational catheters.

• Many of the bolus infusions for PTFE cannulas were associated with high peak tubing pressures (2,000 to 5,000 mm Hg) that remained elevated after completion of the bolus, despite having no evidence of infusion set failure.

•CommercialCSIIinfusionsetswitha6mmpolytetrafluoroethylene(PTFE) cannula (Inset, Unomedical) and investigational sets with a 14 mm coil-reinforced polymer (CRP) cannula and 4 ports (one distaland3proximalportsset2mmapartinahelicalpattern)were inserted in the soft abdominal tissue of ambulatory swine (n=8,47setsintotal).Infusionsetswereexplantedat10minutes,2 days, 4 days, 6 days, and 8 days post-implantation.

• Insulin lispro (U-5) was continuously infused through each CSII sets (7 µL/hour) using an insulin pump with an additional 70 µL bolus one or two times per day with meals.

•Onthefinalstudydate,a70µLbolusofinsulinlispro(U-100)andx-raycontrastagent(Isovue300Iopmidol)wasinfusedthrougheach CSII set.

0 Days 2 Days 6 Days4 DaysCapillary Biomedical Coil-Reinforced Soft Polymer Cannula Design

Commercially Available CSII Set with PTFE Cannula

8 Days

• PK/PD studies in swine and humans are currently underway to determine whether this greater volume and surface area of distribution is able to reach more functional capillary and lymph vessels leading to faster onset/offset and more consistent insulin absorption from dose-to dose.

•Thesoft,flexiblepolymerwire-reinforcedcannulaeliminated kinking, a common early failure mode of commercial CSII sets with a PTFE cannula.

•Insulin/x-raycontrastagentleakedontotheskinsurface,alongthepathofleastresistanceinapproximately10-25%ofthebolusinfusions.

•Insulin/x-raycontrastspreadsefficientlyintoadjacentsubcutaneoustissue immediately after CSII cannula insertion (0 days). Distribution into adjacent vascular tissue on days 2-8 may be slowed and inhibited bythesurroundinglayerofinflammatorytissue.

Figure 3.Representativelongitudinalsequenceofthree-dimensionalmicro-CTimagesillustratingtheextentofinsulin/contrastdispersionintothesubcutaneoustissuesurroundinginvestigationalCSIIsetswithacoil-reinforcedpolymercannula(top)andPTFEcannulaswithasingleorifice(bottom)indwellingfor10minutes,2 days, 4 days, 6 days, and 8 days (left to right).

Figure 1. Evaluation of 3D surface area and volume dispersion measurements froma70µLbolusofinsulin/x-raycontrastagentmixtureinfusedthroughcommercialCSIIsetswithaPTFEcannulaandasingledistalorificeversusinvestigationalCSIIsetswithacoil-reinforcedsoftflexiblepolymer(CRP)andfourportsalongthecannulashaft.Significantlyhigher3Dsurfaceareasandvolumes were achieved with the CRP cannulas.

Figure 2. Evaluation of infusion set leaks and cannula kinks from micro-CT analysis.Thenumberofinfusionsetswithkinkswassignificantlylower(p=0.035) for the investigational CSII sets with a coil-reinforced polymer cannula compared to the commercial CSII sets with a PTFE cannula. The number of sets with leaks onto the skin surface was lower for investigational sets,althoughnotstatisticallysignificantly(p=0.416).

• Pressure was recorded during the bolus using an inline transducer (PendoTECH).

•Tissuewasexcised5minutesafterbolusinfusionandfrozenformicro-CT imaging.

• Specimens were imaged using an Inveon micro-PET/CT imaging scanner (20 µm resolution) to determine the distribution pattern ofinsulin/x-raycontrastagent.

•AMIRA3Dvisualizationsoftware(FEICompany)wasusedtomeasure the surface area and volume of spread of insulin/contrast agent bolus into subcutaneous tissue.

• Failure mode analysis was performed to evaluate the performance of each infusion set group related to cannula kinks and skin surface leakages using the micro-CT images.

•Micro-CTimagingdatawereanalyzedusingANOVAand categoricaldatausingtheFisherExactTestperformedwiththeSystat software (ver. 13).

0

50

100

150

200

250

300

350

Mea

n ±

SEM

CRP

PTFE

* p=0.018 vs PTFE+ p=0.021 vs PTFE

*

+

Surface area (mm2) Volume (mm3)

0

10

20

30

40

50

60

70

80

90

100

No Leak Leak

sisylanA TC orci

M rep skaeLC

annu

la(%

of C

annu

las)

CRP

PTFE

p=0.416 (CRP vs PTFE)

0

20

40

60

80

100

No Kink Kinked

sisylanA TC orci

M rep skniKC

annu

la(%

of C

annu

las)

CRP

PTFE

*p=0.035

*

• Micro-CT images of infusions with high internal pressures often revealedlessthan70µLofinsulin/x-raycontrastagentdeliveredintothe subcutaneous tissue, despite having no insulin pump occlusion alarm. In clinical practice, a diabetic patient may believe a full bolus ofinsulinwasdelivered,whenasignificantlylowervolumeofinsulinwould have actually been delivered into the subcutaneous tissue.

• This variability of actual insulin delivery into the subcutaneous tissue may be a major cause of variable insulin PK/PD in patients managing their diabetes in the real-world. Results from this research have guidedthedesignanddevelopmentofanext-generationinsulin infusion set.

• Clinical translation of this investigational CSII set could result in a morereliableinsulindeliverysystemthatsignificantlyimprovesbloodglucosecontrol,simplifies/reducestheburdenofdiabetesmanagement, improves the quality of life of people with diabetes andbenefitsclinicaloutcomes.