Microwave near-field scanning microscope Abstract We present the development of a novel near-field scanning microwave microscope based on a dielectric resonator. This probe allows local characterization of conducting and insulating films. Our probe has proved the capability of contactless mapping of local thickness, conductivity, and spatially- resolved Electron Spin Resonance (ESR). As the probe is also polarization sensitive, it allows locally mapping the Hall Effect of semiconductors and the magnetization of ferromagnetic thin films. Recently, we designed various probes operating in the frequency range 4-26 GHz with high quality factor, very good sensitivity and high spatial resolution of micron and sub-micron. Our probe also allows the integration of an optical path to the sample deeming it suitable for optically detected magnetic resonance and other optical-microwave measurements. F. Sakran, M. Abu-Teir, A. Copty, M.Golosovsky and D. Davidov Racah Institute of Physics, The Hebrew University of Jerusalem, Jerusalem, Israel References : [1] C. Gao and X. –D. Xiang, Rev. Sci. Instrum. 69, 3846 (1998) [2] M. Golosovsky and D. Davidov, Appl. Phys. Lett. 68, 1579 (1996) [3] J. Gallop, L. Hao, and F. Abbas, Physica C 282-287, 1579 (1997) [4] Abu-Teir M, Golosovsky M, Davidov D, Near-field scanning microwave probe based on a dielectric resonator ,REV. SCI. INSTRUM. 72 (4): 2073-2079 APR 2001 [5] Abu-Teir M, Sakran F, Golosovsky M, Local contactless measurement of the ordinary and extraordinary Hal l effect using near-field microwave microscopy APPL PHYS LETT 80 (10): 1776-1778 MAR 11 2002 [6] Sakran F, Copty A, Golosovsky M, Electron spin resonance microscopic surface imaging using a microwa a) Conductivity/Thickness of thin conducting films: Measurement Setup 0 0 0 s 2 1 Z ty Reflectivi Z Z Z Z Z s s y resistivit the is - ty permeabili space free the is - frequency t measuremen the is - f f Z 0 s 0 thickness. film the is - d d Z s Z0 - is the impedance of free space; Zs - is the effective surface impedance; Thick films: Thin films: Near-field reflectivity (S 11 ) of thin silver films of different thickness d. Note the increase of S 11 upon increasing film thickness. The reflectivity Vs. film sheet resistance. The inset shows the a Q factor Vs. sheet resistance. Probe spatial resolution: X-scan over 0.1 mm chromium strips. The resonator includes slot width of 60 µm. Higher resolution can be obtained by narrowing the slit width. b) Hall effect : M R H R S e xy 0 12 -1 0 1 2 3 0 0.1 0.2 0.3 n-Si p-Si H (T S 12 ( x10 4 ) ) 0 1 2 3 4 5 6 7 0 0.1 0.2 0.3 0.4 80 nm Fig.5 M. Abu-Teir et al. S 12 (x10 4 ) 30 nm 10 nm Ni H(T) -0.5 0 0.5 1 1.5 0 0.1 0.2 0.3 increasing field decreasing field H(T) S 12 (x10 4 ) Ni d=30nm -1 -0.5 0 0.5 1 -0.15 -0.1 -0.05 0 0.05 0.1 0.15 -15 -10 -5 0 5 10 15 X (m m) N dFeB m agnet South pole up N orth pole up S 12 (x 10 4 ) μ 0 H (T) Contactless measurement of the Hall effect in Si wafers on a metal substrate. c) Electron Spin Resonance (ESR) : Measurement Setup H g h 10 mm Spatially-resolved ESR signal from four DPPH grains. An XY- scan, at fixed magnetic field and at fixed frequency, over four small grains of DPPH. We observe four clearly defined ‘‘hills,’’ corresponding to the four DPPH grains. the inset shows the optical image of the sample. -200 -100 0 100 200 300 2990 3000 3010 3020 M agnetic Field (G ) -100 -50 0 50 100 150 8.43 8.44 8.45 8.46 8.47 8.48 Frequency (G H z) -200 -100 0 100 200 300 2990 3000 3010 3020 M agnetic Field (G ) -100 -50 0 50 100 150 8.43 8.44 8.45 8.46 8.47 8.48 Frequency (G H z) Local ESR signal from a 120- μm-thick DPPH layer measured by a 9 GHz probe. The inset shows the ESR signal (using a different DPPH sample) obtained via a frequency sweep and a field modulation. (a) Probe design and (b) measurement setup. Microwave Hall effect in ferromagnetic Ni films. (Extraordinary Hall effect) Mapping of the perpendicular magnetic field of a NdFeB permanent magnet. The solid curve yields the calculated field of the magnet. XYZ stage Prob e 2” Si wafer Permane nt Magnet N S Introduction Results Our Microwave Near-Field Scanning Microscope based on dielectric resonator and narrow slit. (a) The 9 GHz probe design, (b) The actual probe, (c) the probe’s resonance as measured by a Network Vector Analyzer. The probe transmit more than 90% of the incident microwave power. 8.865 8.870 8.875 8.880 8.885 8.890 -60 -50 -40 -30 -20 -10 0 R eflection (d B ) (F req u en cy (G H z)) slit coaxial connector sapphire resonator tuning screw air-gap sapphire transducer coax-to waveguide adaptor SMA connector (a) (b) (c) Probe design g – is the g- factor μ B - is the Bohr magneton H – is the magnetic field ν – is the resonant frequency of the probe h – is plank constant S 12 is the microwave reflectivity measured at port 2. ρ xy is the Hall resistivity H is the magnetic field M is the magnetization. The first term here, R o H, represents the ordinary Hall effect. The second term R e H represaents the Extra ordinary Hall effect. Microwave scanning probes for local characterization of conducting and insulating films attract considerable interest since they are contactless, versatile, and provide high spatial resolution. Recently several microwave scanning probes have been developed, namely coaxial tip [1], slot aperture [2], and dielectric resonator [3]. In our work, there are two important requirements: spatial resolution and sensitivity. In order to achieve high spatial resolution, one needs to use small aperture radiators for near-field scanning. The combination of the dielectric resonator and small aperture, which we report here, provide a highly efficient and sensitive microwave scanning probe. We have developed a variety of Near-Field Scanning Microwave Microscopes working in the 4-26 GHz frequency range [4]. Our probes already proved their efficiency of measuring thickness of conducting layers in the range of 0.1-1 μm mostly applicable to the semiconductor industry. Our near-field microscope allows measurement of the (a) Hall effect [5] (b) Ferromagnetic Resonance (FMR) and (c) Magnetoresistance of magnetic thin films with micron spatial resolution. Moreover, our probe can perform localized electron spin resonance (ESR) measurements [6]. Conclusions : 1. Contactless measurements of conductivity/thickness of thin conducting films in the range 0.1- 1μm. 2. Local, sensitive and contactless Hall effect in semiconductors and magnetization of thin ferromagnetic films. Possibility for low temperature measurements. 3. Our ESR spectrometer is local, sensitive, contactless and non limited to

Microwave near-field scanning microscope Abstract We present the development of a novel near-field scanning microwave microscope based on a dielectric

Embed Size (px)

Citation preview

Page 1: Microwave near-field scanning microscope Abstract We present the development of a novel near-field scanning microwave microscope based on a dielectric

Microwave near-field scanning microscope

Abstract We present the development of a novel near-field scanning

microwave microscope based on a dielectric resonator. This probe allows local characterization of conducting and insulating films. Our probe has proved the capability of contactless mapping of local thickness, conductivity, and spatially-resolved Electron Spin Resonance (ESR). As the probe is also polarization sensitive, it allows locally mapping the Hall Effect of semiconductors and the magnetization of ferromagnetic thin films. Recently, we designed various probes operating in the frequency range 4-26 GHz with high quality factor, very good sensitivity and high spatial resolution of micron and sub-micron. Our probe also allows the integration of an optical path to the sample deeming it suitable for optically detected magnetic resonance and other optical-microwave measurements.

F. Sakran, M. Abu-Teir, A. Copty, M.Golosovsky and D. DavidovRacah Institute of Physics, The Hebrew University of Jerusalem, Jerusalem, Israel

References :[1] C. Gao and X. –D. Xiang, Rev. Sci. Instrum. 69, 3846 (1998)

[2] M. Golosovsky and D. Davidov, Appl. Phys. Lett. 68, 1579 (1996)

[3] J. Gallop, L. Hao, and F. Abbas, Physica C 282-287, 1579 (1997)

[4] Abu-Teir M, Golosovsky M, Davidov D, Near-field scanning microwave probe based on a dielectric resonator,REV. SCI. INSTRUM. 72 (4): 2073-2079 APR 2001

[5] Abu-Teir M, Sakran F, Golosovsky M, Local contactless measurement of the ordinary and extraordinary Hall effect using near-field microwave microscopy APPL PHYS LETT 80 (10): 1776-1778 MAR 11 2002

[6] Sakran F, Copty A, Golosovsky M, Electron spin resonance microscopic surface imaging using a microwave scanning probe APPL PHYS LETT 82 (9): 1479-1481 MAR 3 2003

a) Conductivity/Thickness of thin conducting films:

Measurement Setup

00

0s 21

ZtyReflectivi

Z

Z

ZZ

Z s

s

yresistivit the is -

typermeabili spacefree the is -

frequency tmeasuremen the is - f

fZ

0

s

0

thickness. film the is - dd

Z s

Z0 - is the impedance of free space; Zs - is the effective surface impedance;

Thick films:

Thin films:

Near-field reflectivity (S11) of thin silver films of different thickness d.Note the increase of S11 upon increasing film thickness.

The reflectivity Vs. film sheet resistance.The inset shows the a Q factor Vs. sheet resistance.

Probe spatial resolution: X-scan over 0.1 mm chromium strips. The resonator includes slot width of 60 µm. Higher resolution can be obtained by narrowing the slit width.

b) Hall effect:

MRHRS exy 012

-1

0

1

2

3

0 0.1 0.2 0.3

n-Si

p-Si

H(T)

S 12 (

x 1

04 )

)

0

1

2

3

4

5

6

7

0 0.1 0.2 0.3 0.4

80 nm

Fig.5 M. Abu-Teir et al.

S12 (

x10

4)

30 nm

10 nm

Ni

H(T)

-0.5

0

0.5

1

1.5

0 0.1 0.2 0.3

increasing fielddecreasing field

H(T)

S12 (

x1

04)

Ni d=30nm

-1

-0.5

0

0.5

1

-0.15

-0.1

-0.05

0

0.05

0.1

0.15

-15 -10 -5 0 5 10 15

X (mm)

NdFeB magnet

South pole up

North pole up

S12

(x

104 )

μ 0

H (

T)

Contactless measurement of the Hall effect in Si wafers on a metal substrate.

c) Electron Spin Resonance (ESR) :

Measurement Setup

Hgh

10 mm

Spatially-resolved ESR signal from four DPPH grains. An XY-scan, at fixed magnetic field and at fixed frequency, over four small grains of DPPH. We observe four clearly defined ‘‘hills,’’ corresponding to the four DPPH grains. the inset shows the optical image of the sample.

-200

-100

0

100

200

300

2990 3000 3010 3020

Magnetic Field (G)

-100

-50

0

50

100

150

8.43 8.44 8.45 8.46 8.47 8.48

Frequency (GHz)

-200

-100

0

100

200

300

2990 3000 3010 3020

Magnetic Field (G)

-100

-50

0

50

100

150

8.43 8.44 8.45 8.46 8.47 8.48

Frequency (GHz)

Local ESR signal from a 120- μm-thick DPPH layer measured by a 9 GHz probe. The inset shows the ESR signal (using a different DPPH sample) obtained via a frequency sweep and a field modulation.

(a) Probe design and (b) measurement setup.

Microwave Hall effect in ferromagnetic Ni films. (Extraordinary Hall effect)

Mapping of the perpendicular magnetic field of a NdFeB permanent magnet. The solid curve yields the calculated field of the magnet.

XYZ stage

Probe

2” Si wafer

Permanent Magnet

NS

Introduction

Results

Our Microwave Near-Field Scanning Microscope based on dielectric resonator and narrow slit. (a) The 9 GHz probe design, (b) The actual probe, (c) the probe’s resonance as measured by a Network Vector Analyzer. The probe transmit more than 90% of the incident microwave power.

8.865 8.870 8.875 8.880 8.885 8.890-60

-50

-40

-30

-20

-10

0

Ref

lect

ion

(dB

)

(Frequency (GHz))

slit

coaxial connector

sapphire resonator

tuning screwair-gap

sapphiretransducer

coax-to waveguide adaptor

SMA connector

(a)

(b)

(c)

Probe design

g – is the g-factor

μB - is the Bohr magneton

H – is the magnetic field

ν – is the resonant frequency of the probe

h – is plank constant

S12 is the microwave reflectivity measured at port 2.

ρxy is the Hall resistivity H is the magnetic field

M is the magnetization.

The first term here, RoH, represents the ordinary Hall effect.

The second term ReH represaents the Extra ordinary Hall effect.

Microwave scanning probes for local characterization of conducting and insulating films attract considerable interest since they are contactless, versatile, and provide high spatial resolution. Recently several microwave scanning probes have been developed, namely coaxial tip [1], slot aperture [2], and dielectric resonator [3]. In our work, there are two important requirements: spatial resolution and sensitivity. In order to achieve high spatial resolution, one needs to use small aperture radiators for near-field scanning. The combination of the dielectric resonator and small aperture, which we report here, provide a highly efficient and sensitive microwave scanning probe.We have developed a variety of Near-Field Scanning Microwave Microscopes working in the 4-26 GHz frequency range [4]. Our probes already proved their efficiency of measuring thickness of conducting layers in the range of 0.1-1 μm mostly applicable to the semiconductor industry.Our near-field microscope allows measurement of the (a) Hall effect [5] (b) Ferromagnetic Resonance (FMR) and (c) Magnetoresistance of magnetic thin films with micron spatial resolution. Moreover, our probe can perform localized electron spin resonance (ESR) measurements [6].

Conclusions :

1. Contactless measurements of conductivity/thickness of thin conducting films in the range 0.1- 1μm.

2. Local, sensitive and contactless Hall effect in semiconductors and magnetization of thin ferromagnetic films. Possibility for low temperature measurements.

3. Our ESR spectrometer is local, sensitive, contactless and non limited to sample size. The sensitivity of our present probe is already better than the sensitivity of a conventional ESR spectrometer.