Upload
others
View
1
Download
0
Embed Size (px)
Citation preview
Faculty of Bioscience Engineering
Academic year 2013 – 2014
Model-based analysis of aeration in lab and full-scale
activated sludge systems
Giacomo Bellandi
Promoter: Prof. dr. ir. Ingmar Nopens
Tutor: Ing. Youri Amerlinck
Master’s dissertation submitted in partial fulfillment of the requirements for
the degree of
Master in Environmental Sanitation
Faculty of Bioscience Engineering
Academic year 2013 – 2014
Model-based analysis of aeration in lab and full-scale
activated sludge systems
Giacomo Bellandi
Promoter: Prof. dr. ir. Ingmar Nopens
Tutor: Ing. Youri Amerlinck
Master’s dissertation submitted in partial fulfillment of the requirements for
the degree of
Master in Environmental Sanitation
ad Ivaldo
04-06-2014
De auteur en de promotor geven de toelating dit afstudeerwerk voor consultatie beschikbaar te
stellen en delen ervan te kopiëren voor persoonlijk gebruik. Elk ander gebruik valt onder de
beperkingen van het auteursrecht, in het bijzonder met betrekking tot de verplichting uitdrukkelijk
de bron te vermelden bij het aanhalen van resultaten uit dit afstudeerwerk.
The author and the promoter give the permission to use this thesis for consultation and to copy
parts of it for personal use. Every other use is subject to the copyright laws, more specifically the
source must be extensively specified when using results from this thesis.
The promoter, The tutor, The author,
Prof. dr. Ir. Ingmar Nopens Youri Amerlinck Giacomo Bellandi
i
Contents
Contents ................................................................................................................................................ i
Acknowledgments............................................................................................................................... iii
Abstract ............................................................................................................................................... iv
Summary .............................................................................................................................................. v
1. Introduction and literature review ................................................................................................ 1
1.1. Brief history of wastewater treatment.............................................................................................................. 1
1.2. Wastewater treatment plants ................................................................................................................................ 3
1.2.1. The UCT (University of Cape Town) configuration .......................................................................... 4
1.2.2. Submerged aeration systems ....................................................................................................................... 6
1.3. Modelling of wastewater treatment plants..................................................................................................... 6
1.3.1. Mathematical models ....................................................................................................................................... 6
1.3.2. Activated sludge model ................................................................................................................................... 7
1.3.3. Modelling Oxygen transfer ............................................................................................................................ 7
1.3.4. Predictions of oxygen transfer in presence of surfactants ......................................................... 10
1.3.5. Bubble column tests ....................................................................................................................................... 12
2. Materials and methods ................................................................................................................ 16
2.1. The Eindhoven WWTP............................................................................................................................................. 16
2.2. The Eindhoven wastewater treatment plant model in WEST® .......................................................... 17
2.3. Experimental data collection................................................................................................................................ 19
2.3.1. Off-gas measurements ................................................................................................................................... 19
2.3.2. Bubble column ................................................................................................................................................... 21
2.3.3. Imaging and bubble size measurements .............................................................................................. 23
3. Results ........................................................................................................................................ 27
3.1. Bubble column ............................................................................................................................................................. 27
3.1.1. Rheological measurements ......................................................................................................................... 27
3.1.2. Bubble size measurements ......................................................................................................................... 28
3.1.2.1 Clean water bubble size measurements ............................................................................................... 28
3.1.2.2 Effects of salt addition (50 mg/l) ............................................................................................................. 30
3.1.2.3 Effects of salt addition (100 mg/l) .......................................................................................................... 33
ii
3.1.2.4 Effects of viscosity variation (0.2 g/l Xanthan TER) ...................................................................... 36
3.1.2.5 Effects of viscosity variation (0.8 g/l Xanthan TER) ...................................................................... 39
3.1.3. Lab scale oxygen transfer measurements ........................................................................................... 42
3.2. Full-scale aeration efficiency ................................................................................................................................ 45
3.2.1. Off-gas measurements ................................................................................................................................... 45
3.2.2. The Eindhoven aeration model performances on αSOTE prediction ................................... 50
3.2.2.1 Predictions of DO and NH4 using full scale αSOTE measurements ........................................ 53
4. Discussion ................................................................................................................................... 59
4.1. Bubble column tests .................................................................................................................................................. 59
4.2. Off-gas measurements ............................................................................................................................................. 60
4.3. Aeration model of Eindhoven WWTP .............................................................................................................. 60
5. Conclusions and perspectives ..................................................................................................... 62
References .......................................................................................................................................... 64
iii
Acknowledgments
First and foremost, I would like to thank my academic promoter Ingmar Nopens for his always kind
availability, contagious positive energy and strength which always characterize his person and
motivate who is in contact with him. I have a secret question to ask myself in the hard times, “What
would Ingmar do?!”, and that’s the answer too.
In the same warm way I would like to thank very much my tutor Youri who sustained me through
the long forest of the bad results, through which every student has to pass sooner or later (in my
case quite some time later), with stimulating ideas, precious organization and motivation.
There are so many people that have been important in this period and I would like to thank all and
each one of them. First of all Tinne (before I forgot like the last time!) for the precious help and
source of noise in the lab, Thomas who’s my very good friend and excellent colleague, Stijn who has
always been keeping my back, Andreia who came directly from Portugal to help me, Daan for all the
loughs and the motivational playlist, Sophie for her kind attitude and surprising skills with the
ADCP, Michael for the excellent learning curve on Italian statements to catch girls, Robin for the
memorable time in the BIOMATH weekend, and last but not least Chaim and Stijn for all the
respirometer parties in the lab. Nonetheless, all the BIOMATH family to which I’ll always owe a lot
of gratitude. Thanks Elena, Ivo, Ashish, Usman, Wouter, Séverine, Timothy, Stijn, Wim, Andres,
Katrijn and everyone that passed through the BIOMATH walls.
Ora, dopo tutti i discorsi di rito, vorrei però arrivare al dunque. Voglio ringraziare la mia famiglia, la
forza che mi ha sostenuto e mi sostiene tutti i giorni. Babbo e mamma, vi voglio un bene che scoppio
e vi voglio ringraziare per tutto il sostegno e la fiducia che mi avete dato, spero siate fieri di me
almeno un millesimo di quanto lo sono io di voi. Gianluca e Elena, che sono tutto il mondo per me, li
voglio ringraziare per essere cosi bellini che li mangerei, ma soprattutto per essere sempre nei miei
pensieri vicini vicini. Ci sono stati e ci saranno sempre momenti duri, e questa tesi non è uno di
quelli, e se c’è qualcosa che mi da la forza per tirare avanti anche un carro armato in quei momenti,
quella cosa e il pensiero di Gianluca, Elena e Annalisa. Grazie Na’, che ci sei sempre come un pilastro
in ogni cosa che faccio, e ci portiamo sempre insieme.
Linda, semplicemente la persona piu bella del mondo. La persona che mi fa ridere e innamorare,
arrabbiare e scompisciare. Mi regali ogni giorno quella scintilla che rende le giornate piene di vita.
Grazie, senza di te questa tesi non l’avrei mai fatta.
Sicuramente prima di finire voglio ringraziare la mia altra parte della famiglia, Massimo, la Tiziana,
Simone (Toxy) e Simone (Gloxy), che per me sono importanti come parti del corpo, e che per me
son tutti fratelli (Tiziana compresa), vi voglio bene. E grazie a tutti i letterati che allietano le serate
al barre con poemi e sonetti, ma soprattutto grazie a Grazia, Graziella e Grazialcazzo (che tanto i
ringraziamenti non li legge mai nessuno ☺).
iv
Abstract
Aeration in activated sludge treatment is used to supply the amount of oxygen needed in order to
accomplish the oxidation of part of the nutrients and pollutants unwanted in our river systems. As
the most energy intensive step in wastewater treatment, oxygen supply needs to be performed
efficiently, facilitating the transport of oxygen from the gas to the liquid phase. To do so, the
available surface for exchange needs to be maximized and the coalescence process hindered.
Moreover, the chemical and physical properties of the liquid play an important role affecting the
surface tension and ultimately the diffusion coefficient of oxygen. In the framework of the
development of a sound aeration model, this work studied the (i) effects of salt addition and
viscosity variation on oxygen transfer in a lab scale bubble column, along with (ii) aeration
efficiency measurements performed at a full scale wastewater treatment plant and (iii) the
evaluation and analysis of the prediction performances of one available aeration model. The bubble
column tests provided precious informations on the role singularly played by typical wastewater
characteristics (i.e. low electrolyte concentrations help the oxygen transfer process while
increasing viscosity not necessarily hinders it) and were a valuable tool for the interpretation of the
off-gas measurements performed at the full scale municipal facility of Eindhoven. However, further
experiments should be carried out to rigorously answer some unsolved questions. The model
results confirm the need of a better understating and finer representation of those dynamics
important in the oxygen transfer process, and of the definition of new dependencies for the
accomplishment of robust predictions.
v
Summary
A first introduction and presentation of the concept of wastewater treatment along with its
development throughout history is provided in this thesis in order to stress the importance and
impact that this process, in its different applications, has on our society and on the environment
that hosts it. A brief overview of the different wastewater treatment technologies available is given
with particular attention to conventional activated sludge systems and their most energy intensive
step, i.e. aeration. The focus is then concentrated on submerged aeration systems, on the processes
characterizing oxygen transfer and on the available models that allow its description and
prediction. The materials and methods section provides a thorough description on the practical and
theoretical tools used and on their application. The evolution of bubble sizes and oxygen dissolution
dynamics, studied in lab scale bubble column, are presented in the results along with full scale
measurements of oxygen transfer at the Eindhoven wastewater treatment facility and the outputs
of the currently available model for the same plant. Interpretation of the results is finally discussed
interconnecting the findings for a complete interpretation and understanding of important
dynamics occurring during the aeration process and important clues are provided on the
possibilities for further development of aeration models.
1
1. Introduction and literature review
1.1. Brief history of wastewater treatment
Improved sanitation and adequate water quality availability are the basis for ensuring a healthy
population and primary tools to counteract a vast number of diseases. This was evident already
since the prehistoric period but disposal problems were primarily limited because of the very small
nomadic communities. The first stable civilizations of the Mesopotamian Empire (3500-2500 B.C.)
showed the need to address the problem of sanitation within the growing communities using
connected drainage systems (Lofrano and Brown, 2010). Nonetheless, the Indus civilization left
proves of extensive knowledge in the urban planning with high priority to hygiene. A sewage
network under paved streets dated 14th century B.C., directly connected to sanitary facilities, was
discovered in the Greek island of Santorini (Angelakis et al., 2005) and is only one of the many
examples proving the organized sanitation of the Ancient Greeks. Later the Romans showed an
outstanding improvement in engineering developing these systems further. The construction of a
colossal aqueduct net for irrigation and drinking water supply brought fresh drinking water from
the surrounding hills in the whole city center, and a massive sewage system serving the whole
metropolis ensured a regular cleaning of the streets and waste removal (Henze et al., 2008). The
Cloaca Maxima (6th century B.C.) represents the greatest expression of this milestone of sewage
network.
Figure 1 - View of a branch of the Cloaca Maxima in Rome
Thanks primarily to this very efficient water management system and a well-organized water
supply and waste collection, ancient Rome reached the population of 1,200,000-1,700,000
inhabitants around half of the 2nd century A.D. gaining the title of “Regina Aquorum” (queen of
waters) (Henze et al., 2008). Interestingly, parts of the ancient water supply net and of the Cloaca
Maxima are nowadays still functional (Lofrano and Brown, 2010). However, the collected sewage
did not undergo any treatment and the wastewater was usually discharged directly into the Tiber.
After the Roman empire collapsed, the entire sanitary concept was abandoned (at least in Europe)
and a long period (also known as the sanitary dark age) had to pass before the first collection
Introduction and literature review
2
systems appeared (1800 A.D.). In the early 19th century in Europe a special cart travelling around
the city, gathering the so called “humanure” appeared. The latter was used as fertilizer in the
agricultural fields without any treatment. However, the presence of this cart was not very welcome
in the cities as its inevitable smell bothered the population. Only at the beginning of the 20th century
the first use of microorganisms and the first implementations of the activated sludge (AS) process
were seen.
Firstly observed in the UK, the sludge was believed to be activated similarly to activated carbon and
as such named “activated sludge” by Ardern and Locket in their publication dated 1914.
Approaching the mid-20th century research started to increase significantly and the understanding
of the treatment process gained depth with the characterization of the substrate used by the
microbial communities. The concepts of dissolved oxygen (DO) and biochemical oxygen demand
(BOD) were used in the first river system models to understand the maximum loading a river
stream could handle preventing the DO to fall below a threshold limit (Phelps, 1944). In the second
half of the 20th century the problem of eutrophication began to rise due to the rapid growth of the
population and the increasing discharge of nitrogen (N) and phosphorous (P) in the natural river
streams from both growing cities and enhanced usage of fertilizers. With the general awareness
and agreement that nitrogen and phosphate also needed to be removed from the waste stream, the
application of a mathematical model, developed for understanding the kinetics of continuous
cultures (Monod, 1950), helped to apprehend that nitrification was dependent on the maximum
specific growth rate of autotrophic nitrifying organisms. In order to achieve nitrification, the
retention time of the sludge had to be sufficiently long to ensure the growth of these communities.
However, for the sake of eutrophication prevention, nitrogen had to be completely removed from
the water stream and the discovery of some heterotrophic bacteria capable of converting nitrate to
nitrogen gas (McCarty, 1964) led to the nitrification-denitrification concept of the AS process. Right
after this, the implementation of the pre-denitrification step followed and the combination of the
two processes with the introduction of the recycle flow was a successful improvement also
regarding biological phosphorous removal (Barnard, 1973). The strong increase in population of
the 1970s was the cause of two main issues for the sanitation field. Firstly, the necessity of city
areal expansion caused the incorporation of many treatment plants, initially built outside the urban
area, inside the residential space. Therefore, when the consequential need of increasing the
treatment capacity of these plants became tangible, space efficient technologies such as membrane
bioreactors (MBRs) successfully entered the market. Secondly, the energy crisis pushed the
attention of the research towards the development and improvement of anaerobic processes (e.g.
upflow anaerobic sludge blanket reactors (UASB)), and towards the optimization and further
understanding of the most energy demanding steps in the WWT such as aeration and pumping.
The evolution of sanitation was recently elected from the readers of the British Medical Journal as
the greatest medical advance since 1840 preceding closely the discovery of antibiotics and the
development of anesthesia (Ferriman, 2007). Water and sanitation were defined as two key
components of the right to an adequate standard of living (Amnesty Intenational, 2010). From an
economical point of view, investments for improved sanitation in developing countries are
estimated to generate a revenue between 5 and 29 US$ for each US$ invested (Hutton and Haller,
2004). Although this seems a rather well established concept for most of the industrialized
Introduction and literature review
3
countries, in 2011 just 64% of the world population relied on improved sanitation and the United
Nations Millennium development Goal of 75% by 2015 seems still far (WHO and UNICEF, 2013).
Despite the evidences brought up from numerous scientific publications and the efforts of several
international institutions to rise general awareness on this fundamental issue, proper sewage
treatment not always takes place and progresses in meeting sanitation targets are slow. In this
framework, it is however not only important to provide suitable technologies but also to
sustainably develop long lasting and low energy demanding solutions.
1.2. Wastewater treatment plants
The wastewater treatment plant (WWTP) is currently conceived as an industrial structure in which
one or multiple technologies are applied with the purpose of removing a number of (bio-)chemical
compounds from a water stream. Generally a WWTP is composed of an initial step called primary
treatment where the coarse and easily settable solids can be removed from the wastewater stream
entering the facility. For this purpose, sand grids and primary clarifiers are normally the default
devices used in the primary treatment of a WWTP. Then follows the biological treatment, in which
AS removes the biologically degradable fraction from the wastewater stream. In this step, at least
one tank needs to be provided with dissolved oxygen (DO) from aeration devices in order to ensure
the aerobic metabolism of the microbial communities responsible for the oxidation of organic
matter (OM). The mixed liquor (AS plus wastewater) is retained for a sufficiently long time in the
bioreactor in order to ensure adequate removal of the compounds which are unwanted in the
effluent. As a final step of the treatment, the separation between AS flocs and the treated water is
performed by means of a gravity separator (secondary clarifier). The overflow of the secondary
settler is therefore the purified water that can be discharged or reused, while the underflow is a
concentrated stream of AS. The underflow of the secondary clarifier is recycled to the bioreactor
with the exception of a small fraction called waste sludge. This wasted AS fraction is the extra
sludge generated from the biomass growth and can be collected along with the underflow of the
primary clarifier for anaerobic digestion (energy recovery) and eventual disposal (Tchobanoglous
et al., 2003).
To the present date a large number of different technologies and configurations exist for the
treatment of wastewater streams. The choice of the application of a specific technology depends
mostly on the environmental constraints (temperature, light availability, water scarcity…), type of
wastewater, purpose of the treatment. However, the local, national or international legislation on
discharge limits can impose stringent standards for the water composition exiting the plant. The
European Council Directive 91/271/EEC aims to protect the environment from urban and
industrial wastewater discharges. Therefore, the Directive sets limits for effluent concentration of
organics, nutrients and suspended solids, and defines the required methods for collection,
treatment and discharge of wastewater. To a more general level, the European Water Framework
Directive 2000/60/EC extends the water management to all surface and ground water.
Nonetheless, the selection of a suitable technology often primarily depends on economic aspects.
From here the need of economically suitable solutions which are environmentally sustainable also
from an energetic point of view.
Introduction and literature review
4
1.2.1. The UCT (University of Cape Town) configuration
This configuration was developed in South Africa where groundwater overexploitation, seawater
intrusion and water scarcity were the main driving forces stimulating innovation in the water
sector. Therefore, from the main necessities of water re-use and nutrient removal, a WWTP
configuration capable of removing not only nitrogen but also phosphorous along with providing a
high quality effluent was conceived. The UCT-configuration (Tchobanoglous et al., 2003; Wentzel et
al., 2008) is composed of at least three bioreactors in which the biomass undergoes a sequence of
anaerobic, anoxic and aerobic conditions (Figure 2).
Figure 2 - Schematic representation of the UCT-layout. Blue lines represent the mainstream
of wastewater and mixed liquor with the two internal recirculations, recycle A is a fundamental step for P removal while recycle B plays a crucial role for N removal. Purple
lines represent underflow streams from the two settlers for AS recirculation and waste.
The anaerobic reactor is, for this layout, the first step of the biological treatment and provides an
environment in which neither DO nor nitrogen oxides (NOx) are present. In the anoxic tank that
follows oxygen is absent in the form of DO but there is presence of nitrate (NO3-) and nitrite (NO2
-).
In the last step of the biological treatment DO is present ensuring aerobic conditions for the
oxidation of organic compounds. This sequential alternation is the basis for the accomplishment of
the three main objectives of the system which are carbon, nitrogen and phosphorous removal.
The oxidation of OM is mainly accomplished in the aerobic reactor by heterotrophic bacteria which
oxidize carbonaceous compounds to CO2, water and minerals in order to get energy for their
growth. The fraction of carbon in the wastewater that can be oxidized biologically is usually
identified as BOD5 and can be measured via a closed bottle test (APHA et al., 2012). The chemically
oxidizable carbon (COD) includes both biochemically removable and inert carbon sources. In this
process the availability of a substantial amount of oxygen in solution is of crucial importance for
ensuring the presence and activity of the heterotrophic biomass responsible for the oxidation of
ammonium and OM. At the same time the dissolution of oxygen requires an energy investment
which needs to be carefully balanced with a control strategy, being the most energy demanding step
of the treatment sequence (Reardon, 1995).
The removal of nitrogen is accomplished via the intercurrence of two processes, referred to as
nitrification and denitrification. The nitrogen, usually present in the wastewater under the form of
Introduction and literature review
5
ammonium (NH4+) and organic nitrogen (nitrogen contained in OM), is firstly oxidized to NO3
-
under aerobic conditions and then reduced to nitrogen gas (N2) in anoxic conditions. The chemical
reactions can be summarized as follows for NH4+:
NH4+ + 3 O2 � NO3
- + 2 H+ + H2O (1)
NO3- + C-source � CO2 + N2 + H2O + OH- (2)
In the aerobic tank the oxidation of NH4+ to NO2
- is accomplished by ammonia oxidizing bacteria
(AOB) and subsequently nitrite oxidizing bacteria (NOB) complete the reaction to NO3-. These two
bacterial consortia are autotrophic since they use CO2 as carbon source. The recirculation of the AS
from the aerobic reactor to the anoxic reactor ensures the denitrification step (Equation 2) where
heterotrophs use the oxidation potential of NO3- to oxidize the incoming organic carbon, thus
reducing NO3- to NO2
- and finally to N2 which is free to escape as gas into the atmosphere. The
application of a denitrification step before the nitrification has been of particular importance for
energy recovery since the oxygen contained in the nitrogen oxides can be reused to oxidize part of
the incoming organic matter.
A more complex mechanism, from a conceptual and technical point of view, is the removal of
phosphorus which can be accomplished by chemically provoking its precipitation or favoring the
growth of polyphosphate accumulating organisms (PAOs). The latter process is referred to as
enhanced biological phosphorus removal (EBPR). Providing an anaerobic phase, a consortia of
facultative heterotrophs are able to ferment readily biodegradable OM into volatile fatty acids
(VFA) which are then stored by PAOs in their biomass under the form of polyhydroxyalkanoates
(PHAs). PHAs are energy dense storage compounds and are used by PAOs as their energy source in
anaerobic conditions by hydrolyzing internal polyphosphate reserves and consuming glycogen.
However, this latter process causes release of phosphorus into solution. When PAOs are brought
under anoxic or aerobic conditions they start oxidizing PHA as carbon sources for their growth and
for glycogen regeneration. Since PAOs incorporate more phosphorus in the aerobic/anoxic phase
(up to 24% of their dry weight) than what they release in the anaerobic phase, a net P-removal can
be achieved. Therefore, fundamental steps to be ensured in a P-removal oriented AS process are an
anaerobic phase to advantage PAOs growth and sufficient concentration of readily biodegradable
OM to be converted into VFA (Wentzel et al., 2008).
As described so far, besides ensuring appropriate environmental physical-chemical conditions in
the bioreactors, the recirculation of a well-defined fraction of mixed liquor between the tanks is of
crucial importance for stimulating the growth of specific microbial consortia in a certain section of
the WWTP. Recycle stream flow rate and placement are thus to be carefully planned in the design
step. As schematized in Figure 2, recycle B ensures a constant flow of nitrates to the anoxic tank so
that denitrification can be accomplished. Additionally, recycle A guarantees the return of PAOs into
the anaerobic selector and is located usually at the end of the anoxic tank in order to minimize the
amount of nitrates to be sent in the anaerobic tank. Finally, the flow of return sludge from the
secondary sedimentation tank is maneuvered so to maintain the needed amount of biomass in the
AS system for a successful biological removal of carbon and nutrients.
Introduction and literature review
6
1.2.2. Submerged aeration systems
A number of solutions exist to provide the necessary oxygen to an AS system (Figure 3). Surface
aerators increase the contact area at the gas-liquid interface by agitation of the water surface,
shearing the liquid into small droplets under very turbulent conditions. Although these devices
have easier maintenance (primarily in terms of accessibility), submerged solutions are currently
the most widely used applications due to their higher efficiency and, hence, lower energy
requirements.
Figure 3 – Examples of three different aeration strategies. Surface aeration (a), coarse
bubble aeration (b) and fine bubble aeration (c).
In submerged systems, oxygen is brought into solution by bubbling air or pure oxygen into the
liquid through nozzles or porous material placed at the tank bottom. These devices are called
diffusers and can be divided in two classes, coarse bubble and fine bubble diffusers, depending on
the size of the generated bubbles. Coarse bubble systems generally have orifices larger than 6 mm
and release bubbles up to 50 mm in diameter, whereas a bubble is considered fine when having a
diameter less than 5 mm (WEF, 2010). Rising coarse bubbles are classified as high-flow regime
interfaces due to their high interfacial velocity at the gas-liquid interface, while fine bubbles present
lower interfacial velocity and are therefore grouped as low-flow regime interfaces (Rosso and
Stenstrom, 2006a; Wagner et al., 2002). This is of particular importance for the renewal of the
oxygen concentration in the gas phase at the interfacial layer.
1.3. Modelling of wastewater treatment plants
1.3.1. Mathematical models
A mathematical model is the description of a real-life situation (natural process or behavior) using
mathematical concepts and language. As a theoretical representation of the reality it can only
describe a process to some extent of accuracy but the evolving knowledge and available
computational power make it nowadays possible to solve very complex problems in biological
systems like a WWTP. Especially useful for including the available knowledge, unravelling
underlying processes and making predictions of processes, mathematical models can be used for a
variety of applications such as system optimization, design, development of control strategies,
scenario simulation, etc. (De Pauw, 2005). In general a model transforms a certain input supplied
Introduction and literature review
7
by the user into an output by means of one or multiple (coupled) equations. Therefore, the output
represents the answer the user is interested in.
Mathematical models can be firstly divided in deterministic or stochastic (also called non-
deterministic). The output variable of a stochastic model is obtained probabilistically and is
therefore affected by a certain variability generated by a probability function. On the other hand, a
model is defined deterministic when the selected set of assumptions and equations “determine” a
specific output or response. A deterministic model is composed of a set of algebraic and/or
differential equations containing variables, constants and parameters. A constant always has a fixed
value (e.g. π). On the other hand, a parameter value can vary between simulations but does not
change during a simulation. A variable is a model constituent that can change based on e.g. time or
space. Other than input and output variables, the state variable is the third type of model
constituent and contains the informations used to evaluate the future state of the system. Models
with input variables evolving over time are called dynamic models, otherwise are called static or
steady-state. The linearity or non-linearity of a model depends on how their variables and
parameters appear in the model structure, but more generally a linear model can be defined as
solvable analytically and therefore without the use of numerical techniques (De Pauw, 2005;
Dochain and Vanrolleghem, 2001).
1.3.2. Activated sludge model
In 1982 the International Association on Water Pollution Research and Control (IAWPRC), now
International Water Association (IWA), created a task group on mathematical modelling for design
and operation of activated sludge processes and assigned them the objective of combining the
available knowledge on modeling nitrogen-removal AS plants for developing a model with
minimum complexity. This collaboration resulted in the activated sludge model No. 1 (ASM1). The
ASM is composed of a series of differential equations describing the behavior of AS and is currently
the basis of a common working platform for wastewater treatment modelers. Several extensions
were developed in the following years including more details of the AS process such as storage
properties of the biomass (i.e. ASM3). Furthermore, ASM2 and ASM2d were developed to include
respectively EBPR and the denitrifying capacity of PAOs (Henze et al., 2000).
1.3.3. Modelling Oxygen transfer
For any AS process aeration is a fundamental step since it provides the biomass with the necessary
oxygen in the aerobic section. The oxygen is transferred by shearing the water surface or bubbling
air through macroscopic pores or porous material, always trying to create the maximum oxygen
exchange rate between the gas and liquid phase. Obviously this represents a major energy
demanding process for a WWTP and was estimated to be ranging between 45 to 75 % of the plant’s
energy expenditure (Reardon, 1995). The increasing cost of energy since 1970 awoke the interest
of researchers towards a more in depth understanding of aeration design, specification and
operation, and hence modelling and optimization.
Generally, for a gas bubble immersed into a liquid, the exchange of oxygen per unit of time between
the gas phase and the liquid phase can be described by Equation 3.
Introduction and literature review
8
����� = �� �
(��∗ − ��) (3)
�� = 2 ∙ � ��∙�� (4)
where kL (m/s) is the oxygen transfer coefficient, Cs* (mg/l) is the dissolved oxygen in clean water
at saturation, V (m3) the volume of the liquid phase, A (m2) the surface of the gas-liquid interface
and Ci (mg/l) the dissolved oxygen concentration in the bulk liquid at time t. The oxygen transfer
coefficient is function of the diffusion coefficient D (m2/s) and the mean bubble residence time te
(s). Being D, te and A not easily measurable, an overall oxygen transfer coefficient can be defined as
kLa (h-1).
��� = �� � (5)
The speed with which the oxygen is dissolved in water is function of the difference between the
actual DO concentration and the saturation concentration, but is also strongly dependent of the
physical and geometrical properties of the control volume.
Standardized testing methods for measuring full-scale oxygen transfer were developed by the U.S.
Environmental Protection Agency and the American Society of Civil Engineers (Stenstrom et al.,
2006). The results of about three years of testing lead to the selection of the most influencing
parameters affecting oxygen transfer and therefore aeration systems performance, combined in the
alpha (-), beta (-) and theta (-) factors (Stenstrom and Gilbert, 1981).
The alpha factor is defined as the ratio between the overall oxygen transfer coefficient in
wastewater (kLa) and the one in clean water (kLa*) (Equation 6).
� = ������∗ (6)
This difference between the two kLa is influenced by the presence of surface active agents
(surfactants) and others contaminants affecting the shape of the bubbles and the gas flow at the
interface (Hebrard et al., 2000; Rosso and Stenstrom, 2006b; Stenstrom and Gilbert, 1981).
Ultimately, also the physical properties of the liquid together with its flowing regime can influence
the kLa as coalescence or breakage can vary the bubble sizes and thus the available area for gas
transfer (a coefficient). Viscosity in particular was observed to affect the shape of a bubble plume
(Figure 4) and thus increasing the chances that a bubble has to collide with a neighboring one
(Fabiyi and Novak, 2008; Ratkovich et al., 2013).
Introduction and literature review
9
Figure 4 – Differences in bubble characteristics in a low (a) and high (b) viscosity system.
The mean bubble size in the high viscosity system is much larger (Fabiyi and Novak, 2008).
The beta factor, β, is defined as the ratio between the saturation DO concentration in wastewater
(Cs) and the saturation concentration in clean water (Cs*) (Equation 7).
� = ����∗ (7)
Therefore, beta is affected by several environmental and process conditions having an effect on the
maximum saturation level of DO, among which salinity, temperature, pressure, suspended and
dissolved matter (Stenstrom and Gilbert, 1981; Vogelaar et al., 2000).
The theta factor, also known as geometric temperature correction coefficient, is used to relate mass
transfer coefficients to a standard temperature (Equation 8). Generally a value of 1.024 should be
used unless differently specified and strongly supported by consultants and manufacturers.
Moreover, it is also recommended to limit the temperature correction lower than 10°C, although it
is well known that greater corrections are often needed (Stenstrom and Gilbert, 1981).
���(�) = ���(20°) ∙ "#$%& (8)
Alpha, beta and theta factors are important parameters to standardize aeration performances and
avoid bias due to site-specific environmental and process conditions. In general, for submerged
aeration devices, the oxygen transfer efficiency (OTE, %) is defined by Equation 9 and can be
measured with the off-gas method (Iranpour et al., 2000; Redmon et al., 1983).
'�( = ()*,�,$)*,-./))*,�, (9)
Introduction and literature review
10
where O2,in and O2,out are respectively the mole fractions (%) of oxygen in the gas stream going in
and out of the aerated tank. Typically, for clean water applications, results are reported as standard
oxygen transfer efficiency (SOTE, %), referring to zero DO, zero salinity, 20°C and 1 atm. In order to
correct for process water conditions the alpha factor is used and results are normally shown as
αSOTE (Equation 10).
�0'�( = '�( ∙ �1*2∗(3∙�14∗ $��) ∙ "(%&$#) (10)
This method allows to standardize results of OTE calculating the oxygen saturation concentration in
clean water at 20°C (Cs*20) and the saturation concentration for clean water at half depth (Capela et
al., 2004) in process temperature conditions (Cs*
T).
Additionally, it has to be considered that also the air flow rate has an effect on oxygen transfer
(Figure 5). In fact, an increase in air flow rate is observed to initially decrease steeply the SOTE until
reaching a minimum plateau (Gillot and Heduit, 2000).
Figure 5 - Standard oxygen transfer efficiency (SOTE) as function of the collected air flow rate (Gillot and Héduit, 2008).
1.3.4. Predictions of oxygen transfer in presence of surfactants
Surfactants in wastewater are typically oil, soaps and detergents. These compounds are generally
composed of a hydrophobic tail and an hydrophilic head, and have three peculiar behaviors in
water: the formation of micelles (the hydrophobic tails of multiple molecules attract to each other
forming an agglomerate), the accumulation at the water surface (bringing the hydrophobic tail out
of the liquid) and the accumulation on the bubble surface. This accumulation on the gas-liquid
interface induces two major problems: the increase in rigidity of the interface and a decrease in the
internal gas circulation of the bubble which have a direct effect on the diffusion coefficient and
therefore on the kLa (Ferri and Stebe, 2000; Rosso and Stenstrom, 2006a).
Introduction and literature review
11
Figure 6 - α factors at different flow regimes (defined by the Reynolds (Re) number) for
different aerator types. Adapted from (Rosso and Stenstrom, 2006b).
Figure 6 shows values of α factors measured for different aeration devices with regard to the
respective interfacial flow regimes (expressed by means of the Reynolds (Re) number). In the
region of fine bubble aerators operation, diffusional transport is the driving force for mass
exchange and the gas transfer is controlled by surfactant interfacial migration. In this range of flow
an increase in Re leads to increased surfactants transport to the interface which decreases the α
factor. With regard to coarse bubble diffusers and high shear aerators (surface aerators and
turbines), operating in the turbulent flow domain, an increase in Re results in an enhanced surface
renewal rate and therefore in higher α values (Garner and Hammerton, 1954; Rosso and Stenstrom,
2006a). However, it must be pointed out that the variability in α factor for a given Re value is
considerably high, meaning that other mechanisms are in play.
The mean cell retention time, or sludge retention time (SRT), was observed to be related to the
evolution of the α value. In fact, SRT comprehends in some way the degree of degradation of
contaminants in the wastewater, and therefore also of surfactants. The α value was observed to
increase with increasing SRT showing that, allowing the biomass for a higher contaminants
degradation, ameliorated the oxygen transfer. However, some discrepancies were observed
between the two parameters for plants working with comparable SRT (Groves et al., 1992; Rieth et
al., 1995; Rosso et al., 2005; Wagner, 1999). Interestingly, the parameter χ (s2) and the regression
coefficients for the α factor and OTE prediction (Equation 11, 13 and 14) were defined after a
regression analyses of a large dataset of OTE collected with the off-gas technique over a period of
fifteen years (Rosso et al., 2005).
5 = 67#89�: (11)
;��< = �=7��>�?∙@A∙B (12)
� = 0.172 ∙ FGH5 − 0.131 (13)
Introduction and literature review
12
�0'�( = 5.717 ∙ FGH5 − 6.815 (14)
where AFR (m3/s) is the air flow rate, aspec (m2) is the diffuser specific area, Nd is the total number of
diffusers, Z (m) is the diffusers submergence and Qair (s-1) is the resulting normalized air flux. Figure
7 shows the efficiency parameters α and αSOTE (reported per meter of tank depth) in function of
Qair and SRT (reported as MCRT) for different aerator types, a subset of data used in the design of
the aeration model just described (Equation 11-14).
Figure 7 - Efficiency parameters in function of the normalized air flow rate and mean cell retention time (MCRT or SRT). CDi: ceramic discs; CDo: ceramic domes; CP: ceramic plates;
MD: membrane discs; Tu: ceramic, plastic and membrane tubes; MP: membrane panels (Rosso et al., 2005).
Apparently, up to 30 % of the variability in the α value cannot be explained due to the several
interactions taking place in the mass transfer process and to the lack of knowledge regarding the
effect of aerator submergence (Gillot and Héduit, 2008). In order to take into account the effect of
diffuser submergence the equivalent contact time (ECT) was included in the prediction of α along
with SRT and airflow rate (Gillot and Héduit, 2008). Although ECT seems to combine most of the
generally known factors having an effect on mass transfer, for this method a calculation or
estimation of the kLa*(20°) is necessary a priori complicating the application in predictive aeration
models for WWTP.
1.3.5. Bubble column tests
Bubble columns have been extensively used in the evaluation of aerator’s performance and for the
definition of useful design parameters. Their major advantages are the ease in controlling the
Introduction and literature review
13
process conditions and the relatively simple construction. In bubble columns specific process
conditions, such as temperature or pressure, can be finely controlled and the effect of specific
parameters can be isolated and characterized. Also for wastewater applications bubble columns
represent an important source of informations. Predictive models for oxygen transfer have been
developed using dimensional analysis methods to define the correlation coefficients linking a
selected number of operational parameters influencing kLa and therefore mass transfer (Gillot et al.,
2005; Pittoors et al., 2014; Rosso and Stenstrom, 2006a). However, these studies focus more on the
effect of contaminants on the mass exchange process not fully considering the physical properties
of a liquid and its effects on bubble size. Coalescence and breakup can play in fact an important role
on the bubble size distribution and therefore on the total available surface for mass exchange.
Bubble column tests combined with image analysis showed that presence and concentration of
electrolytes promote to some extent bubble coalescence when salinity is not sufficiently high to
have an influence on viscosity (Lessard and Zieminski, 1971; Ruen-ngam et al., 2008). Moreover,
similar tests revealed that bubble coalescence is also affected by temperature (Figure 8) and a
critical velocity exists, specific for a given liquid at a given temperature but not depending on the
bubble diameter, from which bubbles approaching with a higher speed will bounce back rather
than coalesce (Ribeiro and Mewes, 2006).
Figure 8 - Relative velocity (urel) and the results of individual collisions as a function of
bubble diameter (Ribeiro and Mewes, 2006).
Interestingly, as temperature increases, also the critical velocity increases indicating an
enhancement in bubble coalescence. These latter findings were correctly predicted only by
coalescence models assuming immobile or partially immobile interfaces (also in pure liquids)
suggesting that the liquid thinning process leading to coalescence follows a viscosity dependent
mechanism (Ribeiro and Mewes, 2006). Finally, also the presence of solids, known to affect the
viscosity of the medium (Rosenberger et al., 2002), was observed to influence hydrodynamics along
with mass transfer. In particular, for tests performed in an air-lift with AS sludge at different
concentrations, as the solids content of the mixed liquor increased both the gas holdup and the
mass transfer coefficient decreased (Jin et al., 2006).
Computational fluid dynamics (CFD) simulations are a useful tool to understand the hydrodynamics
taking place in a process using one or multiple fluids and have been extensively used to study the
Introduction and literature review
14
events occurring in bubble columns. The two fluids model, considering both the liquid and gaseous
phase, has been of crucial importance for the study of aerated systems. However, this model was
observed to perform well only in the homogeneous flow domain where the narrow bubble size
distribution can be sufficiently well described by the assumption of a constant bubble diameter. In
order to be able to describe the hydrodynamics in the heterogeneous flow, where bubbles often
show a bimodal distribution (Ribeiro and Lage, 2004; Ruen-ngam et al., 2008), the assumption of a
constant bubble diameter is no longer sufficient and the use of a population balance model (PBM) is
required. In this model, population balance equations (PBEs) are used to describe the evolution of a
certain population as a result of external forces and interactions creating a pressure on their
distribution. The two models (CFD and PBM) are coupled using the gas holdup and the kinetic
energy dissipation calculated with the CFD to solve the PBM equations and get the bubble size
distribution necessary to calculate interphase forces and turbulence modifications of the CFD
model (Wang and Wang, 2007). The use of an integrated CFD-PBM model significantly improved
the insights on the processes driving the coalescence and breakup of bubbles by coupling the
dynamic interactions between the physical properties and the hydrodynamic behavior of a fluid
with the size distribution of the dispersed gas phase. In particular, it was observed that increasing
both the gas velocity and the liquid viscosity resulted in a change in bubble size distribution from
unimodal to bimodal and an increase in viscosity accelerated this process significantly increasing
the average bubble size (Xing et al., 2013). Interestingly, for the range of viscosity usually occurring
in AS processes, i.e. in the range of 10 and 100 mPa s-1 (Ratkovich et al., 2013; Rosenberger et al.,
2002), increasing viscosity was observed to decrease the total gas holdup and the volume fraction
of small bubbles in the CFD-PBM model (Xing et al., 2013). The use of a coupled model led also to
the definition of five different mechanisms for bubble breakup and coalescence (Figure 9). The
breakage can be procured by turbulence and collision, by shear and by interfacial instability.
Concerning coalescence, the effect of turbulent eddies was observed to be much larger as compared
to the other coalescing processes.
Figure 9 - Different mechanisms leading to either bubble coalescence or breakup (Wang, 2011; Xing et al., 2013).
Recent progresses showed that important factors to be considered in the CFD-PBM coupled model
are the effects of liquid viscosity, surface tension (and therefore surfactants and electrolytes
presence), solids concentration, temperature and pressure, and therefore further development is
needed (Wang, 2011).
Introduction and literature review
15
Based on these observations it is evident that the importance of an accurate description and
understanding of both chemical and physical processes occurring during the oxygen transfer is of
crucial importance for the correct quantification of its efficiency. The evidences reported in
literature describing the processes affecting oxygen transfer must not be ignored. The inclusion of
the knowledge coming from the CFD-PBM model could substantially reduce the observed
variability in the prediction of the efficiency parameters based on SRT and airflow since the bubble
size has a direct relation with the kLa coefficient.
16
2. Materials and methods
2.1. The Eindhoven WWTP
The WWTP of Eindhoven (The Netherlands) is the third largest in the country and is operated by
Waterboard De Dommel. Designed to treat the wastewater of 750,000 inhabitant equivalents (IE)
with a load of 136 gCOD/d/IE, the plant is composed of three parallel treatment lines equipped
with one primary settler, one bioreactor and four secondary sedimentation tanks (Figure 10). The
treated effluent is then discharged in the river Dommel.
Figure 10 - Aerial view of the WWTP of Eindhoven and its main process units.
Each bioreactor (Figure 11) is designed according to the UCT layout and consists of one anaerobic
tank (inner ring), one anoxic tank (middle ring) and one aerobic/anoxic tank (outer ring), all
operating in plug-flow configuration. The pre-settled wastewater enters the inner (anaerobic) ring
of the bioreactor and is directed around four sub-divisions ensuring its plug-flow operation. After
the fourth compartment of the inner ring, the mixed liquor is directed to the middle (anoxic) ring
through an opening at the bottom of the tank. At this point the AS is circulated, with a retention
time of 3.5 h, by means of impellers. An overflow located at the outer wall of the middle ring is
feeding the outer (aerobic/anoxic) ring of the bioreactor, while a recirculation pump returns a
fraction of the mixed liquor (recycle A) to the inner ring for P removal. In the outer ring, alternated
aerobic and anoxic zones are maintained. Three pairs of impellers located on three bridges around
the outer ring ensure a minimum of 0.25 m/s mixed liquor flow velocity in order to prevent settling
of the AS flocs (Bosma et al., 2007). The AS exits the outer ring via an underflow located at its outer
wall after the summer package (cascade outflow) while a fraction of the mixed liquor is recycled
back into the middle ring for denitrification.
Materials and methods
17
Figure 11 - Scheme of a bioreactor. The full black arrows show the mixed liquor direction
and the dotted arrows show the recirculation flows throughout the different compartments.
Aeration to the biomass is provided in the outer ring by plate aerators divided in two sections, a
continuously active summer package and a winter package. The winter package is used only
occasionally to increase the aerated volume in the tank (e.g. when low temperatures decrease the
bacterial activity or during rain events when the influent load increases). On the other hand, the
summer package is always active and its airflow is controlled by an ammonia-DO feedback cascade
control which reduces the airflow when the effluent ammonia from the bioreactor is below 1 mg/l.
Additionally, a feedforward control takes action when the incoming flow rate to the plant is above
11,000 m3/h. When this happens, the DO set point is increased to 6 mg/l and both summer and
winter packages are used in order to ensure nitrification.
Thanks to the very advanced Supervisory Control and Data Acquisition (SCADA) system the WWTP
of Eindhoven disposes of high quality dataset of influent, effluent and process data. Therefore, time
series with the resolution of a minute were used in this work for the model simulations.
2.2. The Eindhoven wastewater treatment plant model in WEST®
The multi-platform modelling and experimentation system WEST® (World-wide Engine for
Simulation, Training and automation, MIKEbyDHI, USA) was developed in BIOMATH and used also
for the implementation of the Eindhoven WWTP model (Amerlinck et al., 2013). This software was
developed for incorporating simulation in the design and optimization of WWTPs and provides an
extensive model library. More generally, it allows the construction of models of any kind of system
representable with differential-algebraic equations (Benedetti et al., 2008). The first model of the
Eindhoven WWTP was composed of a single treatment line and used one Continuous-flow Stirred-
Tank Reactor (CSTR) for each of the biological treatments. Over the years several improvements
were included in the model structure in order to most accurately describe the behavior of the
treatment line in all (bio)physical-chemical aspects. Particularly for the biological treatment, the
use of multiple tanks in series was adopted to describe zones of the same ring with different
Materials and methods
18
characteristics and the presence of recirculation flows. Still at the present, the latest version of the
model configuration still considers the three parallel lines of the Eindhoven WWTP as one single
line having the total volume of the plant (Figure 12).
Figure 12 - Layout of the Eindhoven WWTP model in WEST14. The black boxes indicate the
subdivision of the three main compartments of the bioreactors.
However, the three rings of the biological treatment are modelled separately using a number of
tanks in series, each one of them being a CSTR. This configuration enables to mimic the plug flow
character of the three rings of the bioreactors and to separate areas of the same tank having
different conditions (e.g. aerobic/anoxic tank). The inner ring is composed of four tanks resembling
the different compartments of the anaerobic treatment. Two tanks represent the middle ring where
only anoxic conditions are maintained. Because of the higher complexity and heterogeneity of the
outer ring in terms of flow and aerated zones, a more detailed subdivision is needed separating its
volume in seven tanks. Two of these tanks are aerated and represent the summer and winter
package respectively. The remaining five tanks composing the outer ring resemble the four main
anoxic areas in the biological tank and a cascade collecting the outflow towards the secondary
clarifier. In order to resemble the carrousel type character of the middle and outer ring the
recirculation flows are also included. The recycle A and recycle B, necessary for a robust
representation of the UCT layout are implemented by controllers resembling very closely the
actions described in the WWTP process manual. Finally, the return flow of AS (RAS) from the
secondary settler and the wasted fraction are determined by their online full-scale measurements.
Two controllers are implemented in the model for the summer and winter package. The aeration
model used for these controllers (see section 1.3.4) assumes a constant SOTE and calculates the kLa
Materials and methods
19
needed in the ASM resulting in variable α with time. The value of α varies depending on a dynamic
relation with SRT and AFR. The SRT is calculated from the predicted mass of AS in the bioreactor
and the wasted fraction from the secondary settler.
For the biological process, the use of the ASM2d modified model (Gernaey and Jørgensen, 2004)
guarantees the inclusion of both nitrogen and biological phosphorous removal. This model requires
the fractionation of the influent total COD and soluble COD.
The influent data used in the model consists of influent water flow rate (m3/d) and concentrations
(mg/l) of total and soluble COD, total suspended solids (TSS), total phosphorous (TP) and ammonia
(NH). Also necessary as input to the model are the data of AFR (m3/d) to the summer and winter
packages, air and water temperature (°C), pressure (Pa), along with flow rates (m3/d) of aluminum
salts dosing, RAS and waste AS. Despite an excellent data quality and availability, sensor failures are
always a present threat and during the month of July and the first eight days of August very few
data are available. For this reason, for the simulations of the month of August, only the data from
the period 8th – 31st of August were used.
2.3. Experimental data collection
2.3.1. Off-gas measurements
Redmon et al. (1983) stated “The off-gas measurement technique may be a tool for obtaining more
useful design data for aeration systems”. A decade later, an official protocol for process water
testing was developed based on this technique (ASCE, 1997). Today this technique is an established
and reliable measurement for the evaluation of the efficiency performance of submerged aeration
systems and was used in this work to monitor the aerated zone of the biological tank number two
(ATII) of the Eindhoven WWTP. The off-gas equipment was composed of a reinforced polyethylene
hood floating on the wastewater surface (1.5 x 1.5 x 0.3 m, LxWxH). The hood was connected to an
off-gas analyser (evolution of the analyser in Leu et al., 2009) through a flexible hose of 40 mm in
diameter (Figure 13).
Figure 13 - Off-gas analyser (right) and floating hood (left).
Materials and methods
20
In the off-gas analyser, a vacuum pump diverges a small fraction of the off-gas from the main hose
to a desiccator unit in order to remove water vapour. The spilled air flow is then circulated inside a
zirconium oxide fuel cell (AMI Model 65, Advanced Micro Instruments, USA) to measure oxygen
partial pressure. Ambient air was sampled by means of a three-way valve at the start and end of
each experiment as reference for the efficiency evaluation. A schematic representation of the off-gas
analyser is given in Figure 14.
Figure 14 - Off-gas analysis system scheme
When the humidity is stripped out of the gas stream, only the knowledge of the CO2 content is
necessary in order to calculate the actual mass fraction of oxygen (Redmon et al., 1983). With this
purpose, the CO2 content of both the ambient air and the off-gas stream was measured with a
photo-acoustic infrared gas analyser (X-Stream, Emerson). Knowing the CO2 content of the gas
stream, the partial pressure of oxygen and its ratio with inerts were calculated using the following
equations.
MNO/� = QR1−QR−Q�'2R
(15)
MNOS/� = T-UV$T-U$TWX*-U (16)
where MNO/� and MNOS/� represent the molar ratio of oxygen to inerts in the inlet and off-gas
respectively. Q< and QOS are the mole fractions of water vapor in the inlet and off-gas, while Q�)*<
and Q�)*OS are the mole fractions of CO2. Finally, OTE can be calculated with Equation 17 similarly
to Equation 9 considering the dynamic CO2 content in the off-gas.
'�( = Y7-/�$Y7-U/�Y7-/� (17)
The floating hood was equipped with an LDO probe (Hach-Lange) and DO data were acquired in
order to correct for variable DO gradients during the oxygen transfer process and relate the
efficiency results to standard conditions with Equation 10. Data of DO and oxygen content in the
off-gas were acquired with a data acquisition card (DAQ-card USB-6341, National Instruments)
using a graphical user interface developed in LabView (National Instruments, USA). Adjustments
Materials and methods
21
for CO2 content in the off-gas were performed in a post processing step when both the data from the
off-gas analyser and the X-Stream were available.
The off-gas measurements were performed in the month of August 2012 during an extensive
measurement campaign in three locations of ATII (Figure 15), namely the beginning, the middle
and the end of the summer package according to the flow direction.
Figure 15 - Picture of the hood placed at the beginning of the summer package (s.p.) of the ATII (right) and schematic view of the three hood locations (right, red dots).
Dynamics of OTE were monitored during daylight from the 3rd of August until the 24th with the
exception of Saturdays and Sundays when the plant was not accessible. From the morning of the
25th until the early morning of the 27th of August the data were recorded continuously leaving the
instrumentation operating on the site during night hours. The location at the beginning of the
summer package was monitored from August 3rd until August 10th, while the location in the middle
was monitored between August 13th and 17th and the last location hosted the hood from August 20th
to the 27th . The monitoring of the three locations did not have the same time duration due to the
fact that aeration efficiency monitoring was not the sole purpose of the measurement campaign.
2.3.2. Bubble column
A Plexiglas cylindrical column (Figure 16) was used to monitor the fate of air bubbles generated
from a fine pore diffuser in different conditions of salinity and viscosity mimicking typical
characteristics of AS. The column is 160 cm high with an inner diameter of 38 cm. The water level
was maintained in all the experiments at 150 cm resulting in a total liquid volume inside the
column of 170.1 litres. The diffuser used was a disc aerator OXYFLEX®-MT 300 (Supratec,
Germany) with a diameter of 30 cm mounted on a metal plate at the bottom of the column. The
membrane of the aerator was located 20 cm from the bottom of the column. The column was
equipped with a membrane DO probe (Mettler Toledo) to monitor DO dynamics for the evaluation
of the kLa.
Materials and methods
22
Figure 16 - An image of the bubble column (left) and a scheme of the setup used to monitor
bubble sizes and oxygen transfer (right).
For the estimation of the kLa the DO present was stripped by means of N2 gas and, when the DO was
as low as 0.6 mg/l, compressed air was injected until the saturation level was almost reached. In
this way the kLa could be calculated by integrating the re-aeration curve as follows:
Z V��∗$�� [��
�\�2 = ��� Z []�\
�2 (18)
ln(��∗ − ��) = −��� ∙ ] + ln(��∗ − �&) (19)
where �& is the DO concentration at the beginning of the curve, �a is the DO concentration at the
end of the curve, ]& is the initial time and ]a is the final instant of the re-aeration curve. A plot of
ln(��∗ − ��) over time returns a straight line whose slope is -kLa.
Additionally, the off-gas of the column was captured and the oxygen content monitored with the
same analyser used in the full-scale measurements. Softened water was used as the reference clean
water solution. Sodium chloride (NaCl) and Xanthan TER (Colltec GmbH) were used to vary salinity
and viscosity of the medium respectively in order to approach the AS sludge characteristic as much
as possible. The salts concentrations were chosen based on typical wastewater characteristics
(Tchobanoglous et al., 2003). For mimicking the rheological behaviour of AS with a concentration of
10 gMLSS/l, the Xanthan TER concentration was obtained from published results (Rosenberger et
al., 2011). On the other hand, for the AS concentration of 5 gMLSS/l the same literature reported
only the viscosity profile. Therefore, by means of a rotational rheometer (conical concentric
cylinder AR 2000, TA Instruments US) similar in the working principle to the one used by
Rosenberger et al. (2011), the wanted concentrations of Xanthan TER were found fitting the
published results.
Materials and methods
23
Table 1 summarizes the NaCl, Xanthan TER concentrations and the air flows tested in this
experiment. Knowing the value of the kLa in clean water at the different air flows, the relative
effects of salinity and viscosity variations on the α value were evaluated (Equation 6).
Table 1 List of parameters tested and respective values used in the experiments. Viscosity
values report the Xanthan TER concentration and the corresponding AS concentration.
Parameter Values tested
Air flow rate (l/min) 2 – 4 – 6 – 8
Xanthan TER (g/l [gMLSSS/l]) 0 [0] – 0.2 [5] – 0.8 [10]
NaCl (mg/l) 0 – 50 – 100
Additionally, at the different air flows and for all the solutions tested, the gas holdup of the column
was measured manually so that the amount of air trapped inside the liquid from the various
solutions could be monitored.
2.3.3. Imaging and bubble size measurements
A high speed camera (i-SPEEDLT, Olympus) was used to capture high frequency images in order to
evaluate the bubble size distribution (BSD) along the height of the column at the different
conditions of air flow, NaCl and Xanthan TER additions. The camera was equipped with a zoom lens
of focal length 12.5 – 75 mm and aperture f/1.8 to f/22 (Pentax). After a trial period the most
suitable settings were found using a focal length of 75 mm and aperture of f/8, placing the camera
88 cm from the centre of the column, and using an outdoor light of 500W located at the opposite
side of the column (5cm from the outer edge). In order to provide a diffused source of light for a
homogeneous image background, a paper foil was placed on the column surface in front of the light.
These settings provided a reasonably narrow focal plane and a sufficiently bright image to ease the
image processing in the recognition of the bubbles edges. Seven filming locations along the height
of the column were established, namely at 5 cm above the aerator, 20, 40, 60, 80, 100 and 120 cm.
This resolution allowed a very thorough monitoring of the dynamics along the height. For each
location, and for each measurement, three movies were recorded and at least the first 1000 images
of each movie were sampled for the image processing making each BSD the result of the analysis of
about 3000 RBG images in jpeg format. All the settings described for the use of the high speed
camera were used in all the filming with the exception of the paper foil which was not used in
presence of the Xanthan solution. Due to the turbidity of the Xanthan solution the paper foil was not
needed to obtain diffused light.
The image obtained from the high speed camera with the settings described above, allowed to have
a resolution of 0.095 (+/- 0.005) mm/pixel with image dimensions of 600x800 pixels. Having the
resolution far below the millimetre is an important feature to thoroughly detect size variations in
fine bubble systems.
The image analysis was implemented in Matlab® (MathWorks) during this thesis. This code is the
result of an empirical selection of processing methods which were the best performing for the
Materials and methods
24
purpose of an accurate bubble detection. Hereunder, the main steps characterizing the image
analysis are listed and an accurate description of each step follows.
• Image loading and selection of the red channel
• Edge detection and objects filling
• Removal of open edges and of objects touching the image border
• Objects labelling and measurement
• Filtering for circularity reciprocal and convexity
• Final bubble size distribution
The script selects only the red channel of the image as it provides the best contrast in general
(Figure 17, left). Firstly, the Canny method (Canny, 1986) is applied on the black and white image,
for which it calculates the Gaussian derivative on the pixels space detecting the magnitude and the
orientation of the black and white gradients. This allows the enhancement of local maxima of the
black and white gradients and the suppression of the remaining values. Finally, high and low
gradient thresholds refine the bubble detection and the identified edges are composing the output
binary gradient mask (Figure 17, right).
Figure 17 - Red channel of the original image (left) and the binary gradient mask (right) resulting from the edge detection function
After this step the dilate function helps enclosing the edges and the filling of closed objects is
performed while open lines are removed from the image. The borders of the images are cleared
from any object touching the four pixels around the frame perimeter so that only entire objects are
analysed. The rough edges of the filled objects are then eroded of one pixel to smooth the surface.
Consequently, a recursive identification number is given to every object present on the binary
image (i.e. to every “island” of ones) and for each object the values of perimeter, area and convex
area (this latter is the area of the convex regions around the object) are provided. With these shape
parameters, the circularity reciprocal (1/C, -) (Equation 20) and the convexity (-) of each object are
calculated. The circularity reciprocal returns value one for a circle, decreases to zero for ellipsoidal
shapes and increases above one when the perimeter of the object is irregular. Unlike circularity,
convexity is more strictly a measure of roughness and is calculated as the ratio between the internal
area of an object and the area of the imaginary elastic band around it. Therefore, convexity returns
Materials and methods
25
a value of one for e.g. a circle or an ellipse and its value decreases towards zero for more irregular
shapes. Circularity and convexity are thus filtering criteria especially useful to eliminate those blobs
formed by multiple bubbles so that only single bubbles are considered in the final size distribution
(Figure 18). For this last step to work properly, the cut-off criteria were carefully defined with some
preliminary experiments. A maximum of 1.6 and a minimum of 0.2 were optimal values for
circularity, while a maximum of 0.92 was optimal for convexity.
1/� = bc<�dc�c<*e∙�∙�<c� (20)
Figure 18 - Outlined original image before (left) and after (right) the application of the filter
based on circularity and convexity.
The equivalent diameter of each of the remaining bubbles is then stored in memory. The equivalent
diameter, or equivalent projected area diameter, is the size of a circle with the same area as the
detected bubble and is calculated as:
[cf = g�∙e� hi* (21)
where A is the area of the bubble projected on the image. At the end of the image analysis the
number based distribution of the selected bubbles is built dividing the sizes in ten classes with
range of 0.2 mm from 0 to 20 mm. The frequency of occurrence in each of the size classes is then
expressed in number percentage and results in the final number based distribution.
The number based distributions obtained from the image analysis were used, along with the
column gas holdup, to separately calculate the factor kL and a (Baz-Rodríguez et al., 2014; Ruen-
ngam et al., 2008). The value of a can be estimated as:
� = j∙kU�l*(V$kU) (22)
were εg (-) is the fractional gas holdup and d32 (mm) is the Sauter mean diameter. From the gas
holdup (ΔV) and the height of the liquid (VL), it is possible to calculate εg as:
Materials and methods
26
mS = nno� (23)
Concerning the d32, the number based distributions obtained from the image analysis can be used
as follows:
[p% = ∑r���l∑r���* (24)
where ni is the occurrence frequency number of the equivalent area diameter di. Finally, the value
of kL can be calculated as:
�� = ���� (25)
The separate evaluation of kL and a can help in understanding whether a change in kLa is mainly
due to bubble size and amount of surface for exchange or to physical-chemical properties at the
gas-liquid interface.
27
3. Results
3.1. Bubble column
In this section the results are presented of the experimental work performed with the bubble
column in order to understand the effect of a change in electrolyte concentration and viscosity on
the oxygen transfer mechanism. The measurements were first performed in clean water (softened
tap water) and then repeated using two concentrations of NaCl (i.e. 0.05 g/l and 0.1g/l) and two
concentrations of Xanthan TER (i.e. 0.2 g/l and 0.8 g/l). Image analysis tools were used for bubble
size measurements and the gassing out method was used to evaluate the kLa of the system.
3.1.1. Rheological measurements
The concentrations of Xanthan TER were chosen in order to mimic the rheological behaviour of AS
at the concentrations of 10 gMLSS/l (i.e. typical value used in MBR systems) and of 5 gMLSS/l (i.e.
closer to the conditions of conventional AS plant) according to the measurements of Rosenberger et
al. (2011). The profiles of apparent viscosity measured for a set of different concentrations of
Xanthan TER were plotted on the published results (Figure 19). The apparent viscosity profile for
the concentration of 0.8 g/l of Xanthan TER was successfully reproduced (Figure 19, left, red
triangles versus white squares) despite the few differences in the instrumentation used. The major
difference was that Rosenberger et al. (2011) used a double gap cylinder while in this work a single
gap cylinder was used. As expected, decreasing the concentration of Xanthan TER, the apparent
viscosity also decreased showing more and more a similar behaviour to the one of Newtonian fluids
(i.e. constant apparent viscosity over shear rate).
The measurements were performed only once due to time constraints and limited availability of the
instrument. However, for the concentration of 0.4 g/l of Xanthan TER, two measurements were
performed (Figure 19, 0.4 and 0.4b) and the two apparent viscosity profiles matched very closely.
Interestingly, all the curves measured present a rising tail when approaching the highest values of
shear rate differing from the measurements of Rosenberger et al. (2011). Also, the length of the
rising tail, is increasing with decreasing Xanthan TER concentration. A possible explanation for this
observation can lay in the differences between the equipment. The rising tail of a viscosity profile is
usually observed when turbulent flow conditions start to form due to the high speed of the
cylindrical rotor. Since the viscosity measurements have as first assumption that laminar flow
conditions prevail, the part of the measurements in which the viscosity is increasing cannot be
considered reliable.
The Xanthan TER concentration chosen by Rosenberger et al. (2011) matches closely the relative
AS profile in the shear rate around 100 s-1 (Figure 19, left), which is the range of shear occurring in
aerated AS systems (Rosenberger et al., 2011, 2002). Therefore, throughout all the concentrations
tested in this work, the 0.2 g/l Xanthan TER seems to reproduce better than the others the AS
concentration of 5 gMLSS/l of the published results (Figure 19, right).
Results
28
Figure 19 – Measurements of apparent viscosity from Rosenberger et al. (2011) (background black and white graphs) and the rheological measurements performed in this work for
different Xanthan TER concentrations (coloured triangles, no difference between right and left)
3.1.2. Bubble size measurements
In order to have an estimate of the accuracy of the image analysis tool, three sub-samples of a
regular set of 3000 images, used for one of the measurements, were analysed separately and
reported in Figure 20. The use of a sub-sample of 1000 images acquired in sequence resulted in
about 9000 measured diameters composing the final distribution. As expected, the number based
distributions show the highest sensitivity at smaller diameters, while they appear not so sensitive
to variations in the number of large particles. At the value of 50% the distributions differ for
0.08mm, while for lower percentages the gap increases to 0.3mm. The use of the full set of 3000
images is therefore recommended to increase the robustness of the analysis method.
Figure 20 – Number-based cumulative distribution plot of three sub-samples (1000 images each) of the same set of images (3000).
3.1.2.1 Clean water bubble size measurements
The bubble size analysis in clean water shows the effect of height and air flow on the equivalent
diameter of the bubbles (Figure 21 and Figure 22). The bubbles dimension increases with height
due to the (hydrostatic) pressure drop they experience on their surface, causing the internal gas
volume to expand. The effect of pressure was detectable from the image analysis at all air flow rates
(Figure 21 a, b, c and d) showing very similar smooth, continuous shifts in the distribution of the
diameters..
0 1 2 3 40
20
40
60
80
100
deq
(mm)
Num
ber
%
123
Results
29
Figure 21 – Number-based cumulative bubble size distributions in clean water at different heights. The imposed air flow is reported in the title of the graphs.
Beside the pressure effect, also bubble coalescence and breakup can potentially take place during
the rising path along the column height. These effects can be better observed in Figure 22 where
the distributions are plotted with respect to the airflow rate. Close to the aerator (Figure 22, a, b
and c) all the air flows studied produced very similar distributions with the majority of the bubbles
below 1 mm. Further away from the aerator, at 60 cm (Figure 22, d), the difference between the
lowest and highest air flow becomes considerably larger suggesting the occurrence of a higher
coalescence with increasing air flow rate. However, when the bubbles reach 80 and 100 cm from
the aerator (Figure 22 e and f respectively) the difference between the lowest and the highest air
flow seems to slightly decrease. Reaching the highest observation point, at 120 cm from the air
diffuser (Figure 22, g), the difference between the 2 and 8 l/min air flows increases again, i.e. at the
highest flow rate 50% of the bubbles are 0.4 mm bigger than at the lowest air flow rate.
0 0.5 1 1.5 2 2.5 3 3.5 40
10
20
30
40
50
60
70
80
90
100
a
2 l/min
deq
(mm)
Num
ber
%
5 cm20
40
60
80
100120
0 0.5 1 1.5 2 2.5 3 3.5 40
10
20
30
40
50
60
70
80
90
100
b
4 l/min
deq
(mm)
Num
ber
%
0 0.5 1 1.5 2 2.5 3 3.5 40
10
20
30
40
50
60
70
80
90
100
c
6 l/min
deq
(mm)
Num
ber
%
0 0.5 1 1.5 2 2.5 3 3.5 40
10
20
30
40
50
60
70
80
90
100
d
8 l/min
deq
(mm)
Num
ber
%
Results
30
Figure 22 – Number-based cumulative bubble size distributions for clean water at different air flows. The height (in cm) above the aerator is given in the figure title.
3.1.2.2 Effects of salt addition (50 mg/l)
The addition of 50 mg/l of NaCl resulted in very similar bubble size distributions, along the height
and among the air flow rates, to what was observed in clean water (Figure 23 and Figure 24). The
four tested flow rates generated bubbles below the millimeter and crossed the 1.5 mm size when
reaching 120 cm from the aerator.
0 1 2 3 40
20
40
60
80
100
a
5 cm from aerator
deq
(mm)
Num
ber
%
2 l/min
46
8
0 1 2 3 40
20
40
60
80
100
b
20 cm from aerator
deq
(mm)
Num
ber
%
0 1 2 3 40
20
40
60
80
100
c
40 cm from aerator
deq
(mm)
Num
ber
%
0 1 2 3 40
20
40
60
80
100
d
60 cm from aerator
deq
(mm)
Num
ber
%
0 1 2 3 40
20
40
60
80
100
e
80 cm from aerator
deq
(mm)
Num
ber
%
0 1 2 3 40
20
40
60
80
100
f
100 cm from aerator
deq
(mm)
Num
ber
%
0 1 2 3 40
20
40
60
80
100
g
120 cm from aerator
deq
(mm)
Num
ber
%
Results
31
Figure 23 – Number-based cumulative bubble size distributions for 50 mg/l NaCl solution at
different heights. The respective air flow is reported in the title of the graphs.
Close to the aerator (Figure 24, a) the distributions at all air flow rates are very close and below the
millimeter, suggesting that, regardless of the air flow, the bubbles have very similar dimensions.
Already at 20 cm from the aerator (Figure 24, b) the discrepancy between the lowest and the
highest air flow increases, and this is particularly visible in the larger diameter range where a
difference of more than 0.2 mm can be found at 80% of the cumulative distribution. This
discrepancy increases with increasing height, at 40 cm from the aerator (Figure 24, c), suggesting a
predominant coalescence process in this first section of the bubbles rising path with increasing air
flow rate. This coalescence zone seems however to have an earlier end in the NaCl solution
compared to the clean water case since, at 60 cm from the aerator (Figure 24, d), the discrepancy
between distributions appears to stop or even reduce. However, all the distributions are still
shifting to the right side of the graphs with increasing height, and in this movement both the effect
of pressure and coalescence overlap. For this reason, in order to make a good comparison with the
clean water case at the net of the pressure effect, the distributions of the NaCl solution are shown in
terms of deviation from the respective clean water distributions (Figure 25).
0 0.5 1 1.5 2 2.5 3 3.5 40
10
20
30
40
50
60
70
80
90
100
a
2 l/min
deq
(mm)
Num
ber
%
5 cm20
40
60
80
100120
0 0.5 1 1.5 2 2.5 3 3.5 40
10
20
30
40
50
60
70
80
90
100
b
4 l/min
deq
(mm)
Num
ber
%
0 0.5 1 1.5 2 2.5 3 3.5 40
10
20
30
40
50
60
70
80
90
100
c
6 l/min
deq
(mm)
Num
ber
%
0 0.5 1 1.5 2 2.5 3 3.5 40
10
20
30
40
50
60
70
80
90
100
d
8 l/min
deq
(mm)
Num
ber
%
Results
32
Figure 24 – Number-based cumulative bubble size distributions for 50 mg/l NaCl solution at
different air flows. The height (in cm) from the aerator is given in the figure title.
The distributions in the NaCl solution do not differ substantially from the clean water
measurements (Figure 25, a) for the air flow rate of 2 l/min, and maximum deviations, or
oscillations, are in the range of +/-3%. Increasing the air flow rate to 4 l/min (Figure 25, b), the
differences become more pronounced and close to the aerator (5 cm) the NaCl solution shows a
drop of more than 8% for the size of 0.8 mm and an increase of about 4% for the diameters around
1.5 mm with respect to the clean water sample. However, this difference fades out with height, and
at 60 cm the NaCl solution shows a smaller bubble size than the clean water sample with a 4%
increase of the sizes closer to 1 mm and a small drop in the 2 mm range (Figure 25, b). Increasing
the air flow to 6 l/min (Figure 25, c), a similar drop in the size distribution below 1 mm can be
noticed but to a smaller extent (i.e. -6%), which again fades out along the height of the column
resulting in more similar bubble size distributions of the NaCl solution compared to the clean water
case. Close to the aerator, an inversion of this trend can be observed when reaching the 8 l/min air
flow (Figure 25, d), where the diameters around 0.5 mm show an increase of 5.5% coming from a
reduction of 6% in the number of bubbles of 1 mm. Although to a lower extent, this can partially be
visible already at 6 l/min (Figure 25, c). This increase in small diameters seems to disappear
already at 20 cm from the aerator where an increase is shown in the sizes around 1.5 mm. This
latter observation suggests that, even though smaller bubbles are generated by the diffuser in the
0 1 2 3 40
20
40
60
80
100
a
5 cm from aerator
deq
(mm)
Num
ber
%
2 l/min
46
8
0 1 2 3 40
20
40
60
80
100
b
20 cm from aerator
deq
(mm)
Num
ber
%
0 1 2 3 40
20
40
60
80
100
c
40 cm from aerator
deq
(mm)
Num
ber
%
0 1 2 3 40
20
40
60
80
100
d
60 cm from aerator
deq
(mm)
Num
ber
%
0 1 2 3 40
20
40
60
80
100
e
80 cm from aerator
deq
(mm)
Num
ber
%
0 1 2 3 40
20
40
60
80
100
f
100 cm from aerator
deq
(mm)
Num
ber
%
0 1 2 3 40
20
40
60
80
100
g
120 cm from aerator
deq
(mm)
Num
ber
%
Results
33
NaCl solution at 8 l/min of air flow, a strong coalescence takes place already before 20 cm and
continues along the height. However, in general, the distributions at 120 cm for the NaCl seem
always not to deviate more than 3% from the respective clean water case, suggesting that, even
though differences are present close to the diffuser, , the dimensions tend to find a common
equilibrium in both solutions.
Figure 25 – Number-based graphs of the 50 mg/l NaCl solution in percentage of deviation
from the clean water solution. The respective air flow is given in the figure title.
3.1.2.3 Effects of salt addition (100 mg/l)
The increase of NaCl concentration to 100 mg/l presents similar trends over height in the results of
the bubble size analysis (Figure 26) suggesting that the effects of pressure on the bubble size in
both cases are comparable. However, at 2, 4 and 6 l/min of air flow rate (Figure 26, a, b and c) the
difference from the relative change in size due to the height is lower than what is observable at 8
l/min (Figure 26, d), i.e. for 2, 4 and 6 l/min at 50% of the distribution there is a difference of 0.6
mm between the location at 5 cm and the one at 120 cm from the diffuser, while this difference
increases to 0.9 mm for 8 l/min.
0 0.5 1 1.5 2 2.5 3 3.5 4
-8
-6
-4
-2
0
2
4a
2 l/min
deq
(mm)
Num
ber
%
5 cm20
40
60
80
100120
0 0.5 1 1.5 2 2.5 3 3.5 4
-8
-6
-4
-2
0
2
4b
4 l/min
deq
(mm)
Num
ber
%
0 0.5 1 1.5 2 2.5 3 3.5 4
-8
-6
-4
-2
0
2
4c
6 l/min
deq
(mm)
Num
ber
%
0 0.5 1 1.5 2 2.5 3 3.5 4
-8
-6
-4
-2
0
2
4d
8 l/min
deq
(mm)
Num
ber
%
Results
34
Figure 26 – Number-based cumulative bubble size distributions for 100 mg/l NaCl solution
at different heights. The respective air flow is reported in the title of the graphs.
Grouping the distributions according to the air flow rate (Figure 27) it can be firstly noticed how
the increase in air flow rate, in the range investigated, does not influence the bubble size at 5 cm
from the aerator in the same solution (Figure 27, a). The large majority of the generated bubbles
are below 1 mm at the point of generation. However, already at 20 cm above the aerator, the effect
of air flow rate becomes visible in the distributions and at 50% of the cumulative distribution a shift
of 0.15 mm can be observed from the lowest to the highest air flow (Figure 27, b). This discrepancy
increases with height until 80 cm from the diffuser (Figure 27, c) where the gap between the
distributions seems to find an equilibrium range of variation up to the highest location (Figure 27,
e, f and g).
0 0.5 1 1.5 2 2.5 3 3.5 40
10
20
30
40
50
60
70
80
90
100
a
2 l/min
deq
(mm)
Num
ber
%
5 cm20
40
60
80
100120
0 0.5 1 1.5 2 2.5 3 3.5 40
10
20
30
40
50
60
70
80
90
100
b
4 l/min
deq
(mm)
Num
ber
%
0 0.5 1 1.5 2 2.5 3 3.5 40
10
20
30
40
50
60
70
80
90
100
c
6 l/min
deq
(mm)
Num
ber
%
0 0.5 1 1.5 2 2.5 3 3.5 40
10
20
30
40
50
60
70
80
90
100
d
8 l/min
deq
(mm)
Num
ber
%
Results
35
Figure 27 – Number-based cumulative bubble size distributions for 100 mg/l NaCl solution
at different air flows. The height (in cm) from the aerator is given in the figure title.
Figure 28 allows to visualize the effect of 100 mg/l NaCl addition on the bubble size distributions
with respect to the clean water measurements. What can be noticed at first sight, is the peak that
recursively occurs in the distributions for 2, 6 and 8 l/min at the location 5 cm far from the aerator
(Figure 28 a, c and d). However, this does not appear for the air flow of 4 l/min (Figure 28, b) which
instead shows very low deviations from the clean water measurements. In Figure 28 (a), all
locations show a similar trend of increased percentages in the diameter range below 1 mm caused
by a decrease in the bigger sizes. However, apart from the locations at 5, 60 and 80 cm above the
diffuser, the deviations are close to 2% or even lower. In general, it seems that NaCl addition of 100
mg/l promoted the formation of considerably smaller bubbles compared to clean water but along
the bubble rising path this difference gradually disappears due to the coalescence effect.
0 1 2 3 40
20
40
60
80
100
a
5 cm from aerator
deq
(mm)
Num
ber
%
2 l/min
46
8
0 1 2 3 40
20
40
60
80
100
b
20 cm from aerator
deq
(mm)
Num
ber
%
0 1 2 3 40
20
40
60
80
100
c
40 cm from aerator
deq
(mm)
Num
ber
%
0 1 2 3 40
20
40
60
80
100
d
60 cm from aerator
deq
(mm)
Num
ber
%
0 1 2 3 40
20
40
60
80
100
e
80 cm from aerator
deq
(mm)
Num
ber
%
0 1 2 3 40
20
40
60
80
100
f
100 cm from aerator
deq
(mm)
Num
ber
%
0 1 2 3 40
20
40
60
80
100
g
120 cm from aerator
deq
(mm)
Num
ber
%
Results
36
Figure 28 – Number-based graphs of the 100 mg/l NaCl solution in percentage of deviation
from the clean water solution. The respective air flow is given in the figure title.
3.1.2.4 Effects of viscosity variation (0.2 g/l Xanthan TER)
The concentration of 0.2 g/l Xanthan TER was used to mimic the rheological conditions of AS at the
concentration of 5 g MLSS/l.
The combined effect of pressure and coalescence on the bubble size along the height of the column
in presence of the 0.2 g/l Xanthan TER solution is still measurable although the deviations from the
5 to the 120 cm locations are considerably smaller than what was observed in clean water (Figure
29). Interestingly, the smaller bubbles recorded are rarely below 1 mm and almost never smaller
than 0.5 mm. .
0 0.5 1 1.5 2 2.5 3 3.5 4
-10
-8
-6
-4
-2
0
2
4
6
8
10 a
2 l/min
deq
(mm)
Num
ber
%
5 cm20
40
60
80
100120
0 0.5 1 1.5 2 2.5 3 3.5 4
-10
-8
-6
-4
-2
0
2
4
6
8
10 b
4 l/min
deq
(mm)
Num
ber
%
0 0.5 1 1.5 2 2.5 3 3.5 4
-10
-8
-6
-4
-2
0
2
4
6
8
10 c
6 l/min
deq
(mm)
Num
ber
%
0 0.5 1 1.5 2 2.5 3 3.5 4
-10
-8
-6
-4
-2
0
2
4
6
8
10 d
8 l/min
deq
(mm)
Num
ber
%
Results
37
Figure 29 – Number-based cumulative bubble size distributions for 0.2 g/l Xanthan TER
solution at different heights. The respective air flow is reported in the title of the graphs.
Also in the 0.2 g/l Xanthan TER solution the sizes of the generated bubbles do not substantially vary
with the air flow rate (Figure 30, a) but show a maximum difference of 0.25 mm at 50% of the
distribution. Increasing the distance to 20 and even 40 cm from the diffuser gap among the
distributions at the different air flows increases almost exclusively for the bigger diameters (Figure
30, b and c). At 80 cm above the diffuser the difference from the distribution curve obtained with 2
l/min and the distribution obtained with 8 l/min of air flow seems to reach its maximum extent
(Figure 30, d) and this gap appears to be maintained, with some fluctuation, throughout all the
successive locations towards the water surface (Figure 30, e, f and g).
0 0.5 1 1.5 2 2.5 3 3.5 40
10
20
30
40
50
60
70
80
90
100
a
2 l/min
deq
(mm)
Num
ber
%
5 cm20
40
60
80
100120
0 0.5 1 1.5 2 2.5 3 3.5 40
10
20
30
40
50
60
70
80
90
100
b
4 l/min
deq
(mm)
Num
ber
%
0 0.5 1 1.5 2 2.5 3 3.5 40
10
20
30
40
50
60
70
80
90
100
c
6 l/min
deq
(mm)
Num
ber
%
0 0.5 1 1.5 2 2.5 3 3.5 40
10
20
30
40
50
60
70
80
90
100
d
8 l/min
deq
(mm)
Num
ber
%
Results
38
Figure 30 – Number-based cumulative bubble size distributions for 0.2 g/l Xanthan TER
solution at different air flows. The height (in cm) from the aerator is given in the figure title.
The deviations of the distributions in Xanthan TER solution from the distributions in clean water
show very similar trends for all the air flows (Figure 31). In all the measurements of this
experiment the Xanthan TER solution resulted in sensibly bigger bubbles with respect to the clean
water case. The air flow of 2 l/min (Figure 31, a) shows the most gradual trend along the height, but
also for higher flow rates the same can be observed. At 5 cm from the aerator, more than 20% less
bubbles below 1mm in diameter were registered and a 20% increase was observed around 1.5 mm
(Figure 31, a). Increasing the distance from the diffuser, the range in which these changes occur
remains similar but the magnitude changes. Approaching the water surface, the negative difference
shown around 1 mm levels off to 10%, while the positive peak at 1.5 mm reaches 30% at 80 cm
from the aerator and decline again towards the outlet of the column (Figure 31, a). Increasing the
air flow rate to 4, 6 and 8 l/min, the highest (positive and negative) peaks are always registered at 5
cm from the diffuser and the deviation lines smoothen down in the +/-10% range progressively
towards the surface.
0 1 2 3 40
20
40
60
80
100
a
5 cm from aerator
deq
(mm)
Num
ber
%
2 l/min
46
8
0 1 2 3 40
20
40
60
80
100
b
20 cm from aerator
deq
(mm)
Num
ber
%
0 1 2 3 40
20
40
60
80
100
c
40 cm from aerator
deq
(mm)
Num
ber
%
0 1 2 3 40
20
40
60
80
100
d
60 cm from aerator
deq
(mm)
Num
ber
%
0 1 2 3 40
20
40
60
80
100
e
80 cm from aerator
deq
(mm)
Num
ber
%
0 1 2 3 40
20
40
60
80
100
f
100 cm from aerator
deq
(mm)
Num
ber
%
0 1 2 3 40
20
40
60
80
100
g
120 cm from aerator
deq
(mm)
Num
ber
%
Results
39
Figure 31 – Number-based graphs of the 0.2 g/l Xanthan TER solution in percentage of
deviation from the clean water solution. The respective air flow is given in the figure title.
3.1.2.5 Effects of viscosity variation (0.8 g/l Xanthan TER)
The solution of 0.8 g/l of Xanthan TER increased the viscosity of the liquid to that similar conditions
found in MBR systems operated at a concentration of about 10 gMLSS/l of AS. The image analysis
revealed very homogeneous bubble size distributions throughout the height of the column as
compared to the rest of the experiments performed, i.e., in Figure 32, all the distributions within the
same graph are laying in the same size range showing no particular variation with increasing
distance from the diffuser. ). The effects of pressure and coalescence along the height of the column
are much less apparent from the distributions with a small exception for the set of curves obtained
with 2 l/min of air flow (Figure 32, a). In the same graph, the curves relative to the distances of 5
and 80 cm from the diffuser, show a rather high number percentage of bubbles in the range of 0.5 to
1 mm (10 to 20%) which differs from what is observed from the rest of the curves. In order to
verify the reliability of these curves, the images relative to the locations at 5 and 80 cm from the
aerator were visually checked and the image analysis was further tested on these images. No failure
was found in the software and bubbles in the range of 0.5 to 1 mm were actually observed in both
sets. The shape of the initial part of the distribution at 5 cm from the diffuser can be explained by
the presence of few pores generating very small bubbles, and the shape of the initial part of the
0 0.5 1 1.5 2 2.5 3 3.5 4
-20
-10
0
10
20
30a
2 l/min
deq
(mm)
Num
ber
%
5 cm20
40
60
80
100120
0 0.5 1 1.5 2 2.5 3 3.5 4
-20
-10
0
10
20
30b
4 l/min
deq
(mm)
Num
ber
%
0 0.5 1 1.5 2 2.5 3 3.5 4
-20
-10
0
10
20
30c
6 l/min
deq
(mm)
Num
ber
%
0 0.5 1 1.5 2 2.5 3 3.5 4
-20
-10
0
10
20
30d
8 l/min
deq
(mm)
Num
ber
%
Results
40
distribution at 80 cm from the aerator could be the result of breakage occurring along the height.
However, the absence of this initial shape in the rest of the distributions, also for the experiment
with the lower concentration of Xanthan TER, is not clearly explainable and needs further
confirmation through repetitive measurements (outside scope of this work). At higher flow rates (4,
6 and 8 l/min) the distributions are very close and similar to each other in shape (Figure 32, b, c
and d).
Figure 32 – Number-based cumulative bubble size distributions for 0.8 g/l Xanthan TER solution at different heights. The respective air flow is reported in the title of the graphs.
Grouping the distributions according to the air flow rates tested, the diameters of the bubbles at 5
cm from the aerator do show significant variations, suggesting again that, regardless of the air flow
tested, the generated bubbles have similar initial size (Figure 33, a). Getting further in height from
the diffuser, it can be noticed that increases in the air flow rate produced a slightly higher
percentage of smaller bubbles (Figure 33, b). At 40 cm from the aerator (Figure 33, c) this
discrepancy between the air flows increases and reaches its maximum, showing a gap of 0.3 mm, at
50% of the cumulative distribution (i.e. at the air flow of 8 l/min, 50% of the bubbles are at least 0.3
mm smaller than those generated at 2 l/min). This trend is maintained at all heights further than 5
cm from the aerator suggesting that coalescence is counteracted by the increased flow rate.
However, as observed from Figure 32, nor the effect of coalescence neither the effect of pressure
0 0.5 1 1.5 2 2.5 3 3.5 40
10
20
30
40
50
60
70
80
90
100
a
2 l/min
deq
(mm)
Num
ber
%
5 cm20
40
60
80
100120
0 0.5 1 1.5 2 2.5 3 3.5 40
10
20
30
40
50
60
70
80
90
100
b
4 l/min
deq
(mm)
Num
ber
%
0 0.5 1 1.5 2 2.5 3 3.5 40
10
20
30
40
50
60
70
80
90
100
c
6 l/min
deq
(mm)
Num
ber
%
0 0.5 1 1.5 2 2.5 3 3.5 40
10
20
30
40
50
60
70
80
90
100
d
8 l/min
deq
(mm)
Num
ber
%
Results
41
can be detected along the height. Interestingly, at the highest viscosity more variation can be
observed in the smaller sizes than in the bigger sizes, the opposite of what was observed in the
other experiments.
Figure 33 – Number-based cumulative bubble size distributions for 0.8 g/l Xanthan TER
solution at different air flows. The height (in cm) from the aerator is given in the figure title.
When comparing the 0.8 g/l Xanthan TER solution to the clean water results, all the air flow rates
tested produce significantly bigger bubbles in the viscous system than in clean water (Figure 34).
The bubbles with equivalent diameter in the range of 0.7 to 1.5 mm are, along the height, 30 to 15%
less in the viscous system, while the bubbles with size of 1.5 to 2 mm considerably increased (40%)
(Figure 34, a). The air flow of 4 l/min (Figure 34, b) resulted in a similar deviation profile
concerning the location at 5 cm from the aerator, while the amount of bigger bubbles diminishes
gradually with the height but remaining above 10%. The percentage of deviation for the location at
5 cm from the aerator maintains a similar extent and shape also at 6 and 8 l/min of air flow, while
the rest of the locations result in a similarly shaped curve but with a lower maximum and minimum.
Hence, the impact of viscosity, is especially important at low flow rates while higher flows
counteract its effect.
0 1 2 3 40
20
40
60
80
100
a
5 cm from aerator
deq
(mm)
Num
ber
%
2 l/min
46
8
0 1 2 3 40
20
40
60
80
100
b
20 cm from aerator
deq
(mm)
Num
ber
%
0 1 2 3 40
20
40
60
80
100
c
40 cm from aerator
deq
(mm)
Num
ber
%
0 1 2 3 40
20
40
60
80
100
d
60 cm from aerator
deq
(mm)
Num
ber
%
0 1 2 3 40
20
40
60
80
100
e
80 cm from aerator
deq
(mm)
Num
ber
%
0 1 2 3 40
20
40
60
80
100
f
100 cm from aerator
deq
(mm)N
umbe
r %
0 1 2 3 40
20
40
60
80
100
g
120 cm from aerator
deq
(mm)
Num
ber
%
Results
42
Figure 34 – Number-based graphs of the 0.8 g/l Xanthan TER solution in percentage of
deviation from the clean water solution. The respective air flow is given in the figure title.
3.1.3. Lab scale oxygen transfer measurements
Oxygen transfer measurements were performed in all the solutions and the different air flow rates
considered in this study. The kLa of the column in clean water varies from 4.36 (+/-0.09) 1/h at the
lowest air flow rate, up to 12.55 (+/- 0.05) 1/h at 8 l/min (Figure 35). The dissolution of 0.05 g/l
NaCl is initially not having a relevant effect on the kLa for the air flows of 2 and 4 l/min.
Interestingly, when further increasing the air flow rate, the kLa of the 0.05 g/l NaCl solution
becomes about 9.9% and 4.9% higher than in clean water respectively under 6 and 8 l/min air
flows. A further improvement is provided by the dissolution of 0.1 g/l NaCl which enhances the kLa
in the entire range of air flows giving a maximum increment with respect to clean water of 23.4% at
an air flow of 6 l/min. Regardless of the air flow, the α factor in the 0.1 g/l solution is always greater
than 1 (Figure 36). On the other hand, the addition of 0.2 g/l of Xanthan TER, caused a general
decrease in the oxygen transfer by to 20.29% in kLa with respect to clean water and an α value of
about 0.8 for all the air flows. A further increase in viscosity (0.8 g/l Xanthan TER solution) caused
the kLa to drop even further (-32.4%) at 2 l/min of air flow, registering the lowest kLa measured
(2.95 +/-0.22 h-1) . However, increasing the air flow to 4 l/min under the highest viscosity, resulted
in a steeper increase in kLa than in all the other solutions tested, making the kLa to jump over the
0 0.5 1 1.5 2 2.5 3 3.5 4-30
-20
-10
0
10
20
30
40 a
2 l/min
deq
(mm)
Num
ber
%
5 cm20
40
60
80
100120
0 0.5 1 1.5 2 2.5 3 3.5 4-30
-20
-10
0
10
20
30
40 b
4 l/min
deq
(mm)
Num
ber
%
0 0.5 1 1.5 2 2.5 3 3.5 4-30
-20
-10
0
10
20
30
40 c
6 l/min
deq
(mm)
Num
ber
%
0 0.5 1 1.5 2 2.5 3 3.5 4-30
-20
-10
0
10
20
30
40 d
8 l/min
deq
(mm)
Num
ber
%
Results
43
value of 6 1/h. The kLa of the 0.8 g/l Xanthan solution increases with the air flow more steeply than
for the rest of the solutions, and at 6 l/min the α factor crosses the value of 1 (Figure 36). At the
highest air flow, the kLa of the 0.8 Xanthan TER solution is 25.1% bigger than the respective clean
water value.
Figure 35 – Measured kLa of the system at the different airflows and liquid compositions.
Figure 36 – Effect of air flow and solution characteristics on the measured α. The red
horizontal line is the reference for α=1.
The gas holdup was measured in all the solutions and for each air flow rate in order to have an
estimate of the volume of air trapped in the liquid phase. As expected, the values of the gas holdup
for the clean water and the two solutions with NaCl presented no differences (Figure 37, right). On
the other hand, the solution of 0.2 g/l Xanthan TER retained on average 0.16 (+/-0.03) liters more
of air than the clean water. The highest viscosity retained the same amount of air at the air flow of 2
l/min compared to the 0.2 g/l Xanthan TER (i.e. 0.31 l). However, increasing the flow rate in the
higher viscosity medium, also increased the retained fraction of air recursively arriving to a
maximum of 0.91 l of gas holdup. The Sauter diameter of the measured bubbles in the different
Results
44
solutions revealed similar values between the NaCl solutions and the clean water test (Figure 37,
right). On the other hand, the lowest Xanthan TER concentration showed a 10% increase in the
Sauter diameter as compared to the clean water sample for 6 and 8 l/min of air flow rate. The
Sauter diameter for the highest Xanthan TER concentration presented a different profile than for
the other solutions, showing a constant, or slightly decreasing, value over the air flow rate range
(ca. 0.04 mm drop). This observation can explain, along with the increase in gas hold up, the steeply
increasing kLa profile with air flow rate for the Xanthan TER solution. However, the separated
values of kL and a in Figure 38, give a more detailed picture of what process takes place in more
extent (i.e. if either the available surface area for exchange or the diffusion properties at the
interface play a more important role).
Figure 37 – Measured gas holdup (left) and Sauter diameter (right) in function of the air flow
for the different solutions.
Among the clean water and NaCl solutions the calculated a factors show no relevant difference
throughout the air flow rate range (Figure 38, right), while deriving kL puts in evidence the
remaining effects on the oxygen transfer due to gas-liquid interface characteristics (i.e. residence
time and diffusion coefficient). For the 0.05 g/l NaCl solution the kL is lower than the one in clean
water at the air flow of 2 l/min but increasing the air flow to 4 l/min the kL value reaches the one of
the respective clean water sample. (Figure 38, left). Further increasing the airflow the kL reaches its
maximum (and with it also the kLa) at 6 l/min.
Figure 38 – Effect of air flow rate on the calculated a factor (left) and on kL (right) in the different solutions.
Results
45
For the highest NaCl concentration, the kL is constantly bigger than the relative clean water value.
Moreover, although the clean water kL seems to have a descending trend (from 1.42 m/h at 2 l/min
to 1.22 m/h at 8l/min), the kL at the highest NaCl concentration seems not to be influenced by the
air flow rate and oscillates around 1.45 m/h.
The Xanthan TER solutions show higher a and lower kL relatively to the clean water and NaCl
solutions (Figure 38). The a factor for the 0.2 g/l Xanthan TER solution increases less steeply
though with respect to air flow rate than the clean water sample (Figure 38, left), but at the same
time, its kL increases with the air flow (Figure 38, right) making the resulting α value rather
constant. Increasing the viscosity to 0.8 g/l of Xanthan TER, at the air flow of 2 l/min the a factor
results lower than the relative value at lower viscosity, but still higher than the clean water sample.
Increasing the air flow the a value increases with a higher rate than for all other solutions reaching
almost the value of 17 m2/m3 of available area for mass transfer (Figure 38, left). On the other hand,
the kL also increases but is, at all the air flows, sensibly lower than the clean water sample, and
lower than the unity (Figure 38, right). This latter observation makes the a factor, i.e. the available
area for oxygen transfer, the major responsible for the steep increase in the α factor.
3.2. Full-scale aeration efficiency
In this section the results of the aeration efficiency measurements performed in the WWTP of
Eindhoven during the month of August 2012 are reported. Additionally, the off-gas measurements
were used in the plant model in WEST to monitor and improve the performances of the currently
used aeration model.
3.2.1. Off-gas measurements
The OTE performances of the summer package aeration system used in ATII of the Eindhoven
WWTP were monitored using a floating hood and an oxygen analyzer. The monitoring took place
during the month of August 2012, from the 3rd until of the 27th. The 3rd of August the hood was
placed at the beginning of the summer package. However, this was the start-up of the campaign
and, due to the testing of the equipment, the data availability is limited and not of relevant
importance. The measurements officially started the 6th of August, which is taken as day 1 of the
campaign. Table 2 summarizes the subdivision of the days along the summer package during the
measurement campaign.
Table 2 – Locations of the hood on the summer package of ATII during the off-gas
measurement campaign
Beginning Middle End
Day 1 to 5 Day 6 to 10 Day 11 to 17
During day 1 and day 2, the αSOTE fluctuated around 25% in a range of 7% maximum variation
giving an overall constant profile from the morning till the evening of both days (Figure 39). The DO
profile was also rather stable showing a slight increase towards 1 mg/l when approaching 5PM. The
same initial profile can be observed in Day 3 (Figure 39) where, before reaching 5PM the αSOTE
fluctuated around 25%. The NH4 sensor located at the end of the summer package in ATII was
Results
46
under maintenance from day 1 until day 4 and the data are only available from day 5. The increase
in air flow rate during day 3 was also observable at the plant and resulted in both the increase in
DO above 1 mg/l (around 6PM) and the relative drop in αSOTE.
Figure 39 – αSOTE and DO dynamics measured with the hood at the beginning of the summer
package. Data of NH4 (green dashed line) and air flow are obtained from the SCADA system and relative to ATII.
Higher fluctuations in both αSOTE and DO can be observed during day 4 where, for part of the
morning and the afternoon, the efficiency fluctuated around the value of 20% and even lower.
During day 5, the αSOTE profile was again rather stable, around 25%, similarly to day 1, 2 and 3.
The fluctuations observable in all the air flow profiles, with the exception of day 2, are due to a yet
not solved issue with the controllers of the aeration packages.
12PM 1PM 2PM 3PM 4PM 5PM0
5
10
15
20
25
30
35
40
αS
OT
E (%
)
time (hh:mm)
0
1000
2000
3000
4000
5000
6000
7000
8000
Air
flow
rate
(m3/h
)0
1
2
3
4
5
DO
and
NH 4
(mg/
l)
Day 1
12PM 3PM 6PM0
5
10
15
20
25
30
35
40
αS
OT
E (%
)
time (hh:mm)
0
1000
2000
3000
4000
5000
6000
7000
8000
Air
flow
rate
(m3/h
)
0
1
2
3
4
5
DO
and
NH 4
(mg/
l)
Day 2
12PM 3PM 6PM0
5
10
15
20
25
30
35
40
αS
OT
E (%
)
time (hh:mm)
0
1000
2000
3000
4000
5000
6000
7000
8000
Air
flow
rate
(m3/h
)
0
1
2
3
4
5
DO
and
NH 4
(mg/
l)
Day 3
9AM 12PM 3PM 6PM0
5
10
15
20
25
30
35
40
αS
OT
E (%
)
time (hh:mm)
0
1000
2000
3000
4000
5000
6000
7000
8000
Air
flow
rate
(m3/h
)
0
1
2
3
4
5
DO
and
NH 4
(mg/
l)
Day 4
12PM 3PM0
5
10
15
20
25
30
35
40
αS
OT
E (%
)
time (hh:mm)
0
1000
2000
3000
4000
5000
6000
7000
8000
Air
flow
rate
(m3/h
)
0
1
2
3
4
5
DO
and
NH 4
(mg/
l)
Day 5
Results
47
Moving the hood to the middle of the summer package (Figure 40), clearly higher DO values can be
noticed. At the same time, although to a lower extent, also an general increase in the αSOTE profiles
of about 2 to 3% is observable. Day 6 was characterized by a slightly decreasing αSOTE due to the
gradual increase in air flow rate caused by the high NH4 levels. During day 7, the αSOTE showed a
decrease only in the second part of the afternoon when the increasing NH4 caused the controller to
increase the flow rate of air.
Figure 40 – αSOTE and DO dynamics measured with the hood in the middle of the summer
package. Data of NH4 (green dashed line) and air flow are obtained from the SCADA system and relative to ATII.
More constant, although presenting slight variations, is the αSOTE profile that was recorded on day
8. Despite the NH4 increase after 3PM, the increase in air flow rate was not yet considerable to
influence the αSOTE. A different dynamic was observed on day 9, when instead of the usual
constant or slightly decreasing αSOTE, the low NH4 allowed the controller to lower the air flow in
12PM 3PM 6PM0
5
10
15
20
25
30
35
40
αS
OT
E (%
)
time (hh:mm)
0
1000
2000
3000
4000
5000
6000
7000
8000
Air
flow
rate
(m3/h
)
0
1
2
3
4
5
DO
and
NH 4
(mg/
l)
Day 6
9AM 12PM 3PM0
5
10
15
20
25
30
35
40
αS
OT
E (%
)
time (hh:mm)
0
1000
2000
3000
4000
5000
6000
7000
8000
Air
flow
rate
(m3/h
)
0
1
2
3
4
5
DO
and
NH 4
(mg/
l)
Day 7
12PM 3PM0
5
10
15
20
25
30
35
40
αS
OT
E (%
)
time (hh:mm)
0
1000
2000
3000
4000
5000
6000
7000
8000
Air
flow
rate
(m3/h
)
0
1
2
3
4
5
DO
and
NH 4
(mg/
l)
Day 8
9AM 12PM 3PM 6PM0
5
10
15
20
25
30
35
40
αS
OT
E (%
)
time (hh:mm)
0
1000
2000
3000
4000
5000
6000
7000
8000
Air
flow
rate
(m3/h
)0
1
2
3
4
5
DO
and
NH 4
(mg/
l)
Day 9
12PM 3PM0
5
10
15
20
25
30
35
40
αS
OT
E (%
)
time (hh:mm)
0
1000
2000
3000
4000
5000
6000
7000
8000
Air
flow
rate
(m3/h
)
0
1
2
3
4
5
DO
and
NH 4
(mg/
l)
Day 10
Results
48
the late afternoon in order to save energy. As a consequence, an increase in αSOTE up to 30% was
recorded. During the last day in which the middle of the summer package was monitored (day 10),
the αSOTE returned to fluctuate around the value of 25% showing a gentle decrease in the
afternoon due to the increased air flow rate generated by the rising NH4.
Figure 41 - αSOTE and DO dynamics measured with the hood at the end of the summer
package. Data of NH4 (green dashed line) and air flow are obtained from the SCADA system
and relative to ATII.
At the last location of the summer package (Figure 41), the efficiencies observed are often above
30%. During day 11, NH4 increased again in the late afternoon which increased the air flow rate and
decreased the αSOTE slightly below 30%. However, on day 11, the NH4 peak was a bit delayed as
compared to day 12 and 13 and further decreases in αSOTE might have happened. In fact, from day
11 to 15, all the variables show very similar relative and absolute behavior. Interestingly, when the
NH4 peak starts to cross the value of 1 mg/l (around 3PM), the controller of the air flow begins to
12PM 3PM 6PM0
5
10
15
20
25
30
35
40
αS
OT
E (%
)
time (hh:mm)
0
1000
2000
3000
4000
5000
6000
7000
8000
Air
flow
rate
(m3/h
)
0
1
2
3
4
5
DO
and
NH 4
(mg/
l)
Day 11
12PM 3PM 6PM0
5
10
15
20
25
30
35
40
αS
OT
E (%
)time (hh:mm)
0
1000
2000
3000
4000
5000
6000
7000
8000
Air
flow
rate
(m3/h
)
0
1
2
3
4
5
DO
and
NH 4
(mg/
l)
Day 12
12PM 3PM 6PM0
5
10
15
20
25
30
35
40
αS
OT
E (%
)
time (hh:mm)
0
1000
2000
3000
4000
5000
6000
7000
8000
Air
flow
rate
(m3/h
)
0
1
2
3
4
5
DO
and
NH 4
(mg/
l)
Day 13
12PM 3PM 6PM0
5
10
15
20
25
30
35
40
αS
OT
E (%
)
time (hh:mm)
0
1000
2000
3000
4000
5000
6000
7000
8000
Air
flow
rate
(m3/h
)
0
1
2
3
4
5
DO
and
NH 4
(mg/
l)
Day 14
12PM 3PM 6PM0
5
10
15
20
25
30
35
40
αS
OT
E (%
)
time (hh:mm)
0
1000
2000
3000
4000
5000
6000
7000
8000
Air
flow
rate
(m3/h
)
0
1
2
3
4
5
DO
and
NH 4
(mg/
l)
Day 15
Results
49
not properly respond, and gives the high fluctuations with the extent or 2000 or 3000 m3/h in a
very short time frame. This particular behavior was observed each day from day 11 to 15 (Figure
41) and in some of the days when the hood was in the middle location (Figure 40).This behavior
might have occurred as well when the hood was located at the beginning of the summer package
but it is not possible to state this with certainty due to the lack of NH4 data.
The last two days of the campaign (day 16 and 17) the αSOTE and DO profile were monitored
continuously. Similarly to what was observed in the previous figures, the NH4 first rose in in the
afternoon of day 16 provoking once more the fluctuations in the air flow due to the controller.
However, day 16 differs from the previous days in the last location due to the higher NH4 peak (i.e.
4 mg/l) which caused the air flow to jump to 8000 m3/h. This sudden increase in air supply caused
the DO to reach 5 mg/l and the αSOTE dropped about 5%. Around 7PM of day 16 the NH4 was
completely depleted and the controller could lower the air flow rate. Interestingly, another case of
high fluctuations was registered around 9PM when the NH4 was probably not the cause. After
midnight of day 16 a long steady rain event started (2.87 cm of rain fell in the area of Eindhoven in
about 12 hours) that lasted until the end of the measurement campaign. Around 7AM of day 17 the
feedforward control of the plant detected the increase in influent flow and enhanced the air flow in
order to have enough nitrification capacity in the bioreactor. As a consequence, the relative
increase in air flow and DO and the drop in αSOTE (of about 10%) are observable in Figure 42.
Interestingly, when the peak in NH4 was almost depleted, the fluctuation in the air flow occurred
again (3PM of day 17).
Figure 42 – αSOTE and DO dynamics measured continuously at the end of the summer
package during the last two days of the campaign (i.e. from 9AM of day 16 until late night of day 17).
The αSOTE monitored along the length of the summer package showed some variability especially
due to the periods of high fluctuation caused by the controller. It can be noticed that these
fluctuations in αSOTE were more visible towards the end of the summer package. The reason for
this trend is that the fluctuations in the DO, due to the jumping air flow, are not pronounced at the
beginning of the summer package as they are towards the end. However, despite the variability in
the dataset of αSOTE, an increasing efficiency in oxygen transfer can be observed from the
12PM 3PM 6PM 9PM 12AM 3AM 6AM 9AM 12PM 3PM 6PM 9PM 12AM 3AM0
5
10
15
20
25
30
35
40
45
50
αSO
TE
(%
)
time (hh:mm)
0
1
2
3
4
5
6
DO
and
NH 4 (
mg/
l)
Day 16 and 17
0
1000
2000
3000
4000
5000
6000
7000
8000
9000
10000
Air
flow
rat
e (m
3 /h)
Results
50
beginning towards the end of the aeration package of ATII (Figure 43). In particular, although the
standard deviation is rather wide, the beginning of the summer package shows on average about
2% and 6% less efficiency than the middle and end locations respectively.
Figure 43 – Average values of αSOTE at the three locations monitored on the summer
package of ATII
3.2.2. The Eindhoven aeration model performances on αSOTE prediction
In this section the data collected during the off-gas measurements are compared with the modeled
results of αSOTE in order to evaluate the prediction performances of the aeration model.
Additionally, an evaluation of the relative effects on the NH4 predictions is given. Figures are
presented according to the time reference used for the off-gas measurement campaign.
The model was run with the data collected by the SCADA system of the WWTP of Eindhoven during
the month of August so that the results could be compared with the off-gas measurements
performed in the same period. Unfortunately, due to a multiple sensor failure that lasted for the
whole month of July and part of August, it was only possible to run the dynamic simulations starting
from day 6 of the off-gas measurement campaign. Therefore, comparisons with the measured
αSOTE from the beginning of the summer package were not possible. As input for the steady state
simulations, average data from the months of May and June were used, hereby only selecting dry
periods of normal operation.
The modeled results are 4 to 6% lower than the measured ones when the hood was placed in the
middle of the summer package (Figure 44). During day 6, the modeled αSOTE seems to follow with
a similar slope the descending measured profile. The following day the measured αSOTE decreases
more sharply towards its end but the modeled curve seems not to follow the same behavior
although it is decreasing. An increasing measured αSOTE profile during day 8 is initially followed by
the modeled results which, however, starts to decrease earlier. The high presence of noise in the
measured αSOTE as compared to the model profile is due to the high variability of the air flow rate
which, in the case of the input data used for the model, was first filtered.
Beginning Middle End0
0.05
0.1
0.15
0.2
0.25
0.3
0.35
αS
OT
E (
%)
Results
51
Figure 44 – Comparison between the modeled αSOTE and the αSOTE measured in the middle
of the summer package.
The profiles of measured DO from the sensor placed on the hood and from the SCADA system are
about 0.5 mg/l off (Figure 45) due to their location, the hood was placed directly on the aerated
zone while the DO sensor of ATII is located 20 m after the end of the summer package. The modeled
results follow sufficiently the profiles of measured DO only for some of the descending parts. When
the DO is increasing the predictions tend to underestimate with the exception of the last peak at the
end of day 8. The persistent difference between the modeled data and the DO from the hood is
deriving from the gap in αSOTE observed in Figure 44.
Figure 45 – Comparison between the modeled DO, the DO logged by the SCADA system and the DO sensor on the floating hood (middle of summer package).
The NH4 predictions are underestimating the actual values in almost all the cases (Figure 46).
However, the peaks in the simulation seem to gradually increase, whereas the measured NH4 peaks
maintain at a rather constant level. The sudden increase in both measured and modeled NH4 is due
to the arrival of a peak in the influent from a rain event during the night between day 8 and 9.
6 6.5 7 7.5 8 8.5 916
18
20
22
24
26
28
30
αS
OT
E (%
)
time (d)
αSOTE Modeled
αSOTE from hood
6 6.5 7 7.5 8 8.5 90
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
DO
(mg/
l)
time (d)
Modeled DODO from SCADADO from hood
Results
52
Figure 46 – Comparison between the NH4 profile modeled and the NH4 data logged by the SCADA system.
Unfortunately, day 8 was the last possible comparison with the measured data from the middle of
the summer package due to a rain event that changed the influent conditions. The model used is
only calibrated for dry weather (i.e. influent flow of about 85000 m3/d) and the rain that fell in the
night between day 8 and 9 brought the influent flow above 120000 m3/d which heavily disturbs the
predictions.
In the last location, at the end of the summer package, the recorded efficiency was the highest of the
aeration package. Therefore, when comparing the αSOTE predictions of this period with the off-gas
measurements, the gap observed reaches 8 to 12% (Figure 47). However, the profiles of both data
sets show similar trends within the days with the exception of day 12 which reports a decrease of
10% in the measured αSOTE. Looking at the general trend, the prediction of αSOTE slightly
decreases from day 10 to 15 while the measured profiles of αSOTE maintain a rather constant
trend.
Figure 47 - Comparison between the modeled αSOTE and the αSOTE measured at the end of the summer package.
Similarly to what was observed for the location in the middle of the summer package, the modeled
DO profile resembles rather well the behavior of the measured DO (Figure 48). However, the gap
between the modeled profile and the measured DO at the hood location increased with respect to
that observed in the middle of the summer package. This is due to the higher DO values at the end
6 6.5 7 7.5 8 8.5 90
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
NH
4 (m
g/l)
time (d)
Modeled NH4
NH4 from SCADA
10 11 12 13 14 15 1616
18
20
22
24
26
28
30
32
34
36
αS
OT
E (%
)
time (d)
αSOTE Modeled
αSOTE from hood
Results
53
of the aeration package, where the oxygen dissolution reaches its maximum in the tank. The peaks
of the predicted DO seem to gradually decrease while this is not the case for the measured DO.
Figure 48 - Comparison between the modeled DO, the DO logged by the SCADA system and
the DO sensor on the floating hood (end of summer package).
Also for the end of the summer package the predicted NH4 profile is initially underestimating the
real value and presents an increasing trend which was not observed in the measured NH4 profile
(Figure 49). The reason for this response of the model might be linked to the decreasing DO
predicted which is lowering the oxidation capacity of the volume.
Figure 49 - Comparison between the NH4 profile modeled and the NH4 data logged by the
SCADA system.
3.2.2.1 Predictions of DO and NH4 using full scale αSOTE measurements
The measured αSOTE during the off-gas campaign was now used as input for the CSTR representing
the summer package, to run the new predictions of DO and NH4. The simulations were initiated for
steady state with the same dataset used in the previous section. The dynamic simulations were run
in the same way as in the previous section until the measurements of αSOTE were available. From
that moment direct measurements of αSOTE could be used as input for the model. The dynamic
simulation was re-initiated with the last dynamic results and the new prediction of DO and NH4 was
performed with the measured αSOTE.
10 11 12 13 14 15 160
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
DO
(m
g/l)
time (d)
Modeled DODO from SCADADO from hood
10 11 12 13 14 15 160
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
NH
4 (m
g/l)
time (d)
Modeled NH4
NH4 from SCADA
Results
54
Running the simulation for day 6, the new DO profile, resulting from the use of the measured
αSOTE, gave an initially better prediction of the DO from the SCADA system than for the previous
simulation (Figure 50, left). However, the value of the new DO increases much faster and reaches
the values of DO measured from the hood. This higher DO concentration in the new profile results
in even lower concentrations of NH4, far below the SCADA measurements (Figure 50, right).
Figure 50 – Model performances for DO (left) and NH4 (right) predictions using αSOTE measurements as model input. Middle of summer package for day 6.
From the simulation of day 7, the new DO profile suddenly detaches from the old simulated DO
values and matches almost perfectly the measurements of the hood (Figure 51, left). Also in this
case, the high new value of DO results in a much more underestimated NH4 concentration as
compared to the old prediction (Figure 51, right).
Figure 51 - Model performances for DO (left) and NH4 (right) predictions using αSOTE
measurements as model input. Middle of summer package for day 7.
The last day of measurements on the middle location of the summer package does not show a very
close fit of the new DO profile with any of the measurements on the site (Figure 52, left). The
prediction of the new DO based on measured αSOTE falls initially in between the measurements of
6 6.1 6.2 6.3 6.4 6.5 6.6 6.7 6.8 6.9 70
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
DO
(mg/
l)
time (d)
New DO prediction
Old DO prediction
DO from SCADA
DO from hood
6 6.1 6.2 6.3 6.4 6.5 6.6 6.7 6.8 6.9 70
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
NH
4 (m
g/l)
time (d)
New NH4 prediction
Old NH4 prediction
NH4 from SCADA
7 7.1 7.2 7.3 7.4 7.5 7.6 7.7 7.8 7.9 80
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
DO
(mg/
l)
time (d)
New DO prediction
Old DO prediction
DO from SCADA
DO from hood
7 7.1 7.2 7.3 7.4 7.5 7.6 7.7 7.8 7.9 80
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
NH
4 (m
g/l)
time (d)
New NH4 prediction
Old NH4 prediction
NH4 from SCADA
Results
55
the SCADA system and the measurement from the hood, and finishes above all. Once more, given
the elevated DO, the new prediction of the NH4 profile results much lower than the measured value.
Figure 52 - Model performances for DO (left) and NH4 (right) predictions using αSOTE measurements as model input. Middle of summer package for day 8.
The simulations ran using the measured αSOTE from the end of the summer package were
performed with the same methodology. The measured αSOTE of day 11 caused the new simulated
DO to reach and surpass the DO concentrations measured with the hood, and therefore reach a
much higher value than the DO from the SCADA system (Figure 53, left). As a consequence, the new
predicted NH4 concentration is almost not visible in the graph (Figure 53, right).
Figure 53 - Model performances for DO (left) and NH4 (right) predictions using αSOTE measurements as model input. End of summer package for day 11.
Passing to the simulation of day 12, the new DO is also up to the level, or even higher than the
measurements from the hood location (Figure 54, left) and, as a result, the NH4 concentration is
also in this case extremely low (Figure 54, right).
8 8.1 8.2 8.3 8.4 8.5 8.6 8.7 8.8 8.9 90
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
DO
(mg/
l)
time (d)
New DO prediction
Old DO prediction
DO from SCADA
DO from hood
8 8.1 8.2 8.3 8.4 8.5 8.6 8.7 8.8 8.9 90
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
NH
4 (m
g/l)
time (d)
New NH4 prediction
Old NH4 prediction
NH4 from SCADA
11 11.1 11.2 11.3 11.4 11.5 11.6 11.7 11.8 11.9 120
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
DO
(mg/
l)
time (d)
New DO prediction
Old DO prediction
DO from SCADA
DO from hood
11 11.1 11.2 11.3 11.4 11.5 11.6 11.7 11.8 11.9 120
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
NH
4 (m
g/l)
time (d)
New NH4 prediction
Old NH4 prediction
NH4 from SCADA
Results
56
Figure 54 - Model performances for DO (left) and NH4 (right) predictions using αSOTE
measurements as model input. End of summer package for day 12.
Similarly as for the latter simulation, the new DO prediction is far above the old simulation and
even higher than the measurements at the hood location (Figure 55, left). Moreover, the new NH4
predicted profile is again merely visible (Figure 55, right)
Figure 55 - Model performances for DO (left) and NH4 (right) predictions using αSOTE measurements as model input. End of summer package for day 13.
The measured αSOTE used for the new simulation of Day 14 resulted in a considerably high initial
value of the new DO, which then descended again to meet the DO hood measurements (Figure 56,
left). The new NH4 simulation remains noticeably lower than the old NH4 simulation and the NH4
data from the SCADA system (Figure 56, right).
12 12.1 12.2 12.3 12.4 12.5 12.6 12.7 12.8 12.9 130
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
DO
(mg/
l)
time (d)
New DO prediction
Old DO prediction
DO from SCADA
DO from hood
12 12.1 12.2 12.3 12.4 12.5 12.6 12.7 12.8 12.9 130
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
NH
4 (mg/
l)
time (d)
New NH4 prediction
Old NH4 prediction
NH4 from SCADA
13 13.1 13.2 13.3 13.4 13.5 13.6 13.7 13.8 13.9 140
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
DO
(mg/
l)
time (d)
New DO prediction
Old DO prediction
DO from SCADA
DO from hood
13 13.1 13.2 13.3 13.4 13.5 13.6 13.7 13.8 13.9 140
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
NH
4 (m
g/l)
time (d)
New NH4 prediction
Old NH4 prediction
NH4 from SCADA
Results
57
Figure 56 - Model performances for DO (left) and NH4 (right) predictions using αSOTE
measurements as model input. End of summer package for day 14.
On day 15, although it experienced a relatively high αSOTE as compared to the other days in the
location at the end of the summer package, the new DO prediction results lower than the value
measured with the hood (Figure 57, left). At the end of the new simulation the DO profile matches
the DO data from the SCADA system. Despite the lower new predicted DO with respect to the one
measured at the hood location, the NH4 profile still tends to very low concentrations far from the
SCADA measurements.
Figure 57 - Model performances for DO (left) and NH4 (right) predictions using αSOTE measurements as model input. End of summer package for day 15.
In order to have a better fit of the results concerning NH4 prediction the relative half saturation
constant was increased from the previously calibrated value of 0.05 mg/l to the ASM2d default
value of 1 mg/l reporting the results in Figure 58.
14 14.1 14.2 14.3 14.4 14.5 14.6 14.7 14.8 14.9 150
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
DO
(mg/
l)
time (d)
New DO prediction
Old DO prediction
DO from SCADA
DO from hood
14 14.1 14.2 14.3 14.4 14.5 14.6 14.7 14.8 14.9 150
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
NH
4 (mg/
l)
time (d)
New NH4 prediction
Old NH4 prediction
NH4 from SCADA
15 15.1 15.2 15.3 15.4 15.5 15.6 15.7 15.8 15.9 160
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
DO
(mg/
l)
time (d)
New DO prediction
Old DO prediction
DO from SCADA
DO from hood
15 15.1 15.2 15.3 15.4 15.5 15.6 15.7 15.8 15.9 160
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
NH
4 (mg/
l)
time (d)
New NH4 prediction
Old NH4 prediction
NH4 from SCADA
Results
58
Figure 58 - Model performances for DO (left) and NH4 (right) with the change in half
saturation constant for NH4
6 6.1 6.2 6.3 6.4 6.5 6.6 6.7 6.8 6.9 70
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
DO
(mg/
l)
time (d)
New DO prediction
Old DO prediction
DO from SCADA
DO from hood
6 6.1 6.2 6.3 6.4 6.5 6.6 6.7 6.8 6.9 70
0.5
1
1.5
2
2.5
3
3.5
4
4.5
5
NH
4 (mg/
l)
time (d)
New NH4 prediction
Old NH4 prediction
NH4 from SCADA
59
4. Discussion
4.1. Bubble column tests
The use of a bubble column allowed a better understanding of fine bubble aerator performance in
different conditions of salinity and viscosity. Four air flow rates were tested using clean water, with
the addition of NaCl and with Xanthan TER. Two concentrations were used for each of the two
compounds, i.e. 0.05 and 0.1 g/l for NaCl, 0.2 and 0.8 g/l for Xanthan TER. The concentrations of
both compounds were chosen based on the available literature (Rosenberger et al., 2011) in order
to recreate, at least in part, typical conditions of wastewater and AS. The NaCl concentrations were
chosen in order to resemble typical variations of electrolyte content in wastewater with respect to
clean water, and hence their effect on the oxygen transfer. On the other hand, two Xanthan TER
concentrations were chosen to mimic the rheological features of two typical AS concentrations, i.e.
5 and 10 g MLSS/l. The separate effects of the change in electrolyte concentration and viscosity on
the bubble size and oxygen transfer could be successfully evaluated and quantified.
The use of a high speed camera made it possible to acquire a large quantity of images with a pixel
resolution of 0.095 mm. The image analysis tool developed was able to detect variations in bubble
sizes due to the effect of hydrostatic pressure and medium composition.
The addition of the higher concentration of NaCl resulted in smaller bubbles generated close to the
aerator. This was also visible for the lower concentration of NaCl but only from the 6 l/min of air
flow. However, the coalescence process seems to take place in the first 20 to 40 cm from the
diffuser and at lower depths the bubbles in the NaCl solution had a similar size distribution as to
what was observed in clean water. The effect on kLa was visible just on the kL factor and the
observations were in good agreement with the literature since the effect of electrolyte
concentrations has been seen to enhance the exchange of oxygen by lowering the surface tension
(Baz-Rodríguez et al., 2014; Ruen-ngam et al., 2008). However, for the lower NaCl solution, also the
air flow seems to increase the kLa (higher than the clean water kLa only at 6 and 8 l/min) which can
be due to the gradual decrease in bubble size, but also to enhanced turbulence (and thus increased
surface renewal) and lowered surface tension.
All the bubble size measurements in Xanthan TER solutions resulted in significantly larger bubbles
compared to the clean water case. However, the highest differences in bubble size were observed
close to the aerator for all the air flow rates tested. Values of α always lower than 1 were observed
consistently for the lowest Xanthan TER concentration which, despite the higher gas holdup with
respect to clean water, presented a low kL. This leads to the observation that, at low viscosity,
diffusion limitations are prevailing on the higher surface available for exchange and, even for higher
air flows (with which kL increases due to higher turbulence and surface renewal), the kLa remains
constantly lower than in clean water. The increase in viscosity with the highest concentration of
Xanthan TER presented a steeply increasing kLa profile over the air flow rate
Discussion
60
range which started below the clean water value and ended with the highest kLa registered in this
work. This increase can be addressed to the relative augmentation of a, and therefore in the
available surface for exchange with increasing air flow. Also the kL was observed to increase with
air flow, supposedly due to enhanced surface renewal, but remained significantly lower than the
respective clean water value suggesting that turbulence did not play a major role in the increase in
kLa. In fact, the higher concentration of Xanthan TER, despite the sensibly bigger size of the bubble
with respect to clean water, showed α values greater than one for a sufficiently high air flow but
mainly due to the considerably higher capacity in trapping the gas bubbles, i.e. increased gas
holdup.
4.2. Off-gas measurements
Measurements of oxygen transfer efficiency were performed on the ATII of the WWTP of Eindhoven
at the beginning, in the middle and at the end of the summer package using a floating hood
equipped with a DO probe and an off-gas oxygen analyser. The main findings of the campaign were
the observation of an increasing αSOTE from the beginning towards the end of the summer package
(following the flow direction) and the relatively high efficiency of this system as compared to other
similar applications. This difference in αSOTE between the locations, considering that the entire
aeration package provides homogeneously the air flow, is addressable to the DO gradient along the
aeration package (i.e. the water entering the aerated section has DO concentrations close to zero
since it is coming from the anoxic zone) and to the gradual contaminant oxidation occurring (Rosso
and Stenstrom, 2006a). Fluctuations in αSOTE due to the air flow rate were observed almost every
day of measurements in concomitance with the rise of NH4 above 1 mg/l indicating an imperfect
operation of the controller. These variations could not be really explained from the results of the
bubble column, however, tests in controlled conditions at higher flow rates than the ones tested in
the lab, could give an answer to the occurrence and extent of these changes in efficiency. These
fluctuations were more and more visible on the αSOTE profile towards the end of the summer
package since the effect of air flow on DO was more pronounced. A longer period of continuous
measurements in the last two days of the campaign, revealed that these fluctuations occurred also
in moments in which the air flow was being decreased after the occurrence of the NH4 peak. A
possible explanation could be still a problem with the air flow controller linked to a certain range of
air flow.
4.3. Aeration model of Eindhoven WWTP
The model of the WWTP of Eindhoven was used to run simulations with the data made available
from the SCADA system of the plant for the same period in which the off-gas measurements
campaign was performed. In this way the performances of the aeration model were compared with
the measurements of off-gas in terms of αSOTE. The aeration model for both middle and summer
package returned lower αSOTE and DO values than those measured from the hood. However, there
was relatively higher accuracy in terms of αSOTE in the prediction for the middle of the summer
package rather than for the end location. For this reason, the aeration model is potentially better
predicting the aeration efficiencies and DO values from the location at the beginning of the summer
Discussion
61
package. However, the lack of influent data needed to run the model for the initial period of the off-
gas measurement campaign is still an obstacle.
A reasonably accurate prediction resulted when the modeled DO was compared with the one
measured by the SCADA system. However, the modeled DO profile for the end location showed a
decreasing trend in the DO peaks and therefore a constant loss of nitrification capacity. Regarding
the precision in the NH4 prediction, the model almost always underestimated the actual value
measured by the plant sensor and the NH4 profile showed, in both middle and end locations, an
increasing trend of the peaks in time.
The αSOTE measurements represented an important source of informations on the performances
of the aeration model and were also tested as direct model input for DO and NH4 prediction. As a
consequence of the use of a higher αSOTE with respect to the previous simulation results, the DO
predictions often overestimated the SCADA measurements and went to match the measurements
from the hood. As expected the NH4 was even further depleted as compared to the previous
simulations and therefore more distant from the measured NH4. These observations might lead to
consider a re-evaluation of the model parameters linked to NH4 depletion as was illustrated by
lowering the NH4 affinity. Doing so, as observed in the last simulation shown, is probably going to
improve the predictions of the modeled DO profile in matching the measured values, since a higher
amount of DO will be used for nitrification.
62
5. Conclusions and perspectives
From the bubble column tests one important observation can be drawn which is the importance of
the presence of a dynamic α factor in an aeration model. Different compositions of the liquid
medium respond in a different manner in terms of α and therefore on the ultimate aeration
efficiency. Therefore, it is of crucial importance that a dynamic α is considered in an aeration model.
However, by nature kL and a, are influenced by both physical and chemical properties of the system,
nonetheless by hydrodynamic conditions which are heavily depending on air flow rate in the
aerated zones. Therefore, the speed with which oxygen is dissolved is depending not only on the
amount and type of electrolytes and surfactants in the liquid phase, but also on the physical
properties of the liquid, such as viscosity, and their interplay with a variable air flow rate on the
speed of oxygen dissolution.
The bubble column tests showed how both bubble size and kLa measurements are important and
complementary for a satisfying description of the oxygen transfer process. Further studies should
be performed in order to understand how the interaction between electrolyte dissolution and
viscosity resolves in the final kLa measurement. This is of crucial importance, as the establishment
of new dependencies and the development of more detailed relations can help filling the need for
better predictive models for aeration performances (and from here more precise energy demand
and carbon footprint predictions). The coalescence process was observed and monitored in the
different medium tested and its extent of occurrence could be measured revealing that viscous
mediums at low air flow present less coalescence than clean water. However, for higher flow rates
than the ones tested in this work, this latter conclusion might not be valid anymore since stronger
hydrodynamic forces can play a more important role.
The off-gas measurements were a valuable tool to gather informations on the performances of the
aeration system used in the Eindhoven WWTP. A more in depth understanding of the dynamics
occurring in the different sections of the summer package was possible and provided good insights
on the capabilities of this systems. The aeration efficiency increases along the length of the summer
package in the direction of the flow due to the gradient in DO from the beginning to the end location
but probably also to the change in wastewater characteristics (i.e. the gradual oxidation of
contaminants affecting the surface tension) (Rosso and Stenstrom, 2006a).
The αSOTE measurements from the off-gas campaign were used to evaluate the performance of the
aeration model used in the Eindhoven WWTP model. The predictions ran with the influent data
acquired by the SCADA system of the plant, resembled reasonably well the DO concentrations
measured in the aeration tank. However, comparisons with data measured from the hood (αSOTE
and DO) revealed very similar trends but with a consistent gap. The predictions of NH4
concentration in the bioreactor as compared to the actual measurements were not accurate, which
suggests the need of a revaluation of the parameters responsible for NH4 dynamics. Further off-gas
measurements should be performed again in the location at the beginning of the summer package
which might show more similar results to the aeration model output.
Conclusions and perspectives
63
Possible future improvement of the aeration model could consider the subdivision of the summer
aeration package in three tanks in series resembling more accurately the three locations monitored
at the Eindhoven WWTP and their local aeration efficiency. Moreover, a step forward in the model
development, would be represented by the inclusion of a PBM considering the evolution of bubble
size and relative available surface for exchange along the height, throughout the air flow rate range
and the influent flow rates. However, as hydrodynamics play a crucial role in the oxygen transfer
process, CFD modeling is a necessary and crucial tool to be coupled with PBM. Additionally, a more
consistent link between the influent composition and the relative effect on oxygen transfer could
also represent a turning point in the development of a sound aeration model. This work is a first
step towards this direction.
64
References
Amerlinck, Y., Flameling, T., Maere, T., Weijers, S., Nopens, I., 2013. Practical application of dynamic process models for wastewater treatment plant optimization: work in progress, in: WEFTEC. p. 5.
Amnesty Intenational, 2010. Haki zetu, ESC rights in Practice. Amnesty International Netherlands, Amsterdam.
Angelakis, A.N., Koutsoyiannis, D., Tchobanoglous, G., 2005. Urban wastewater and stormwater technologies in ancient Greece. Water Res. 39, 210–20.
APHA, AWWA, WEF, 2012. Standard Methods for the Examination of Water & Wastewater. American Public Health Association, New York.
ASCE, 1997. ASCE Standard: Standard Guidelines for In-Process Oxygen Transfer Testing (18-96). ASCE, New York.
Barnard, J.L., 1973. Biological nutrient removal without the addition of chemicals. Water Res. 9, 485–490.
Baz-Rodríguez, S. a., Botello-Alvarez, J.E., Estrada-Baltazar, A., Vilchiz-Bravo, L.E., Padilla-Medina, J. a., Miranda-López, R., 2014. Effect of electrolytes in aqueous solutions on oxygen transfer in gas–liquid bubble columns. Chem. Eng. Res. Des. In press.
Benedetti, L., Bixio, D., Claeys, F., Vanrolleghem, P. a., 2008. Tools to support a model-based methodology for emission/immission and benefit/cost/risk analysis of wastewater systems that considers uncertainty. Environ. Model. Softw. 23, 1082–1091.
Bosma, A., Dalstra, G., Verhoeven, M., Reitsma, B., 2007. Meten en modelleren van het stromingsgedrag in de actief slibtanks van de rwzi Eindhoven met CFD. Afvalwaterwetenschap 6, 115–128.
Canny, J., 1986. A computational approach to edge detection. IEEE Trans. Pattern Anal. Mach. Intell. 8, 679–98.
Capela, S., Heduit, A., Roustan, M., 2004. Influence of water depth on fine bubble aeration efficiency in the presence of surfactants, in: Preprints of the 3rd IWA World Water Congress.
De Pauw, D., 2005. Optimal experimental design for calibration of bioprocess models: a validated software toolbox. Ghent University.
Dochain, D., Vanrolleghem, P.A., 2001. Dynamical modelling and estimation in wastewater treatment processes. IWA Publishing.
References
65
Fabiyi, M.E., Novak, R., 2008. Evaluation of the factors that impact succesful membrane biological reactor operation at high solids concentration. Membr. Technol. 2008, 503–512.
Ferri, J., Stebe, K., 2000. Which surfactants reduce surface tension faster? A scaling argument for diffusion-controlled adsorption, Advances in colloid and interface science.
Ferriman, A., 2007. BMJ readers choose sanitation as greatest medical advance since 1840. Br. Med. J. 334, 111.
Garner, F.H., Hammerton, D., 1954. Circulation inside gas bubbles. Chem. Eng. Sci. 3, 1–11.
Gernaey, K. V., Jørgensen, S.B., 2004. Benchmarking combined biological phosphorus and nitrogen removal wastewater treatment processes. Control Eng. Pract. 12, 357–373.
Gillot, S., Capela-Marsal, S., Roustan, M., Héduit, a, 2005. Predicting oxygen transfer of fine bubble diffused aeration systems--model issued from dimensional analysis. Water Res. 39, 1379–87.
Gillot, S., Héduit, a, 2008. Prediction of alpha factor values for fine pore aeration systems. Water Sci. Technol. 57, 1265–9.
Gillot, S., Heduit, A., 2000. Effect of air flow rate on oxygen transfer in an oxidation ditch equipped with fine bubble diffusers and slow speed mixers. Water Res. 34, 1756–1762.
Groves, K.P., Daigger, G.T., Simpkin, T.J., Redmon, D.T., Ewing, L., 1992. Evaluation of oxygen transfer efficiency and apha-factor on a variety of diffused aeration systems. Water Environ. Res. 64, 691–698.
Hebrard, G., Destrac, P., Roustan, M., Huyard, A., Audic, J.M., 2000. Determination of the water quality correction factor α using a tracer gas method. Water Res. 34, 684–689.
Henze, M., Gujer, W., Mino, T., van Loosdrecht, M., 2000. Activated sludge models ASM1, ASM2, ASM2d and ASM3. IWA Publishing.
Henze, M., van Loosdrecht, M.C.M., Ekama, G.A., Brdjanovic, D., 2008. Biological Wastewater Treatment. IWA Publishing.
Hutton, G. (World H.O., Haller, L. (World H.O., 2004. Evaluation of the Costs and Benefits of Water and Sanitation Improvements at the Global Level.
Iranpour, R., Magallanes, A., Zermen, M., Varsh, V., Abrishamchi, A., Stenstrom, M.K., 2000. Assessment of aeration basin performance efficiency: sampling methods and tank coverage. Water Res. 34, 3137–3152.
Jin, B., Yin, P., Lant, P., 2006. Hydrodynamics and mass transfer coefficient in three-phase air-lift reactors containing activated sludge. Chem. Eng. Process. Process Intensif. 45, 608–617.
References
66
Lessard, R.R., Zieminski, S.A., 1971. Bubble coalescence and gas transfer in aqueous electrolytic solutions. Ind. Eng. Chem. Fundam. 10, 260–269.
Leu, S.-Y., Rosso, D., Larson, L.E., Stenstrom, M.K., 2009. Real-Time aeration efficiency monitoring in the activated sludge process and methods to reduce energy consumption and operating costs. Water Environ. Res. 81, 2471–2481.
Lofrano, G., Brown, J., 2010. Wastewater management through the ages: a history of mankind. Sci. Total Environ. 408, 5254–64.
McCarty, P.L., 1964. Thermodynamics of biological synthesis and growth, in: Procs. 2nd Int. Conf. on Water Pollution Control. pp. 169–199.
Monod, J., 1950. Technique for continuous culture - theory and application, Ann. Inst. Pasteur.
Phelps, E.B., 1944. Stream sanitation, John Wiley and Sons Inc.
Pittoors, E., Guo, Y., Van Hulle, S.W.H., 2014. Oxygen transfer model development based on activated sludge and clean water in diffused aerated cylindrical tanks. Chem. Eng. J. 243, 51–59.
Ratkovich, N., Horn, W., Helmus, F.P., Rosenberger, S., Naessens, W., Nopens, I., Bentzen, T.R., 2013. Activated sludge rheology: a critical review on data collection and modelling. Water Res. 47, 463–82.
Reardon, D.J., 1995. Turning down the power. Civ. Eng. 65, 54–56.
Redmon, D.T., Boyle, W.C., Ewing, L., 1983. Oxygen transfer efficiency measurements in mixed liquor using off-gas techniques. Water Pollut. Control Fed. 55, 1338–1347.
Ribeiro, C.P., Lage, P.L.C., 2004. Experimental study on bubble size distributions in a direct-contact evaporator. Brazilian J. Chem. Eng. 21, 69–81.
Ribeiro, C.P., Mewes, D., 2006. On the effect of liquid temperature upon bubble coalescence. Chem. Eng. Sci. 61, 5704–5716.
Rieth, M.G., Chiesa, S.C., Polta, R.C., 1995. Effects of operational variables on the oxygen transfer performance of ceramic diffusers. Water Environ. Res. 67, 781–787.
Rosenberger, S., Helmus, F.P., Krause, S., Bareth, a, Meyer-Blumenroth, U., 2011. Principles of an enhanced MBR-process with mechanical cleaning. Water Sci. Technol. 64, 1951–8.
Rosenberger, S., Kubin, K., Kraume, M., 2002. Rheology of Activated Sludge in Membrane Bioreactors. Eng. Life Sci. 2, 269–275.
Rosso, D., Iranpour, R., Stenstrom, M.K., 2005. Fifteen years of off-gas tranfer efficiency measurements on fine pore aerators: key role of sludge age and normalized air flux. Water Environ. Res. 77, 266–273.
References
67
Rosso, D., Stenstrom, M.K., 2006a. Surfactant effects on alpha-factors in aeration systems. Water Res. 40, 1397–404.
Rosso, D., Stenstrom, M.K., 2006b. Economic Implications of Fine-Pore Diffuser Aging. Water Environ. Res. 78, 810–815.
Ruen-ngam, D., Wongsuchoto, P., Limpanuphap, A., Charinpanitkul, T., Pavasant, P., 2008. Influence of salinity on bubble size distribution and gas–liquid mass transfer in airlift contactors. Chem. Eng. J. 141, 222–232.
Stenstrom, M.K., Gilbert, R.G., 1981. Effects of alpha, beta and theta factor upon the design, specification and operation of aeration systems. Water Res. 15, 643–654.
Stenstrom, M.K., Leu, S.-Y. (Ben), Jiang, P., 2006. Theory to Practice: Oxygen Transfer and the New ASCE Standard, in: WEFTEC. pp. 4838–4852.
Tchobanoglous, G., Burton, F.L., Stensel, H.D., 2003. Wastewater Engineering: Treatment and Reuse, Engineering, McGraw-Hill series in civil and environmental engineering. McGraw-Hill.
Vogelaar, J.C.T., Klapwijk, A.M., Van Lier, J.B., Rulkens, W.H., 2000. Temperature effects on the oxygen transfer rate between 20 and 55°C. Water Res. 34, 1037 – 1041.
Wagner, M., 1999. Factor influencing the magnitude of alpha-values of fine bubble aeration systems, in: WEFTEC.
Wagner, M., Cornel, P., Krause, S., 2002. Efficiency of different aeration systems in full scale membrane bioreactors, in: WEFTEC. pp. 434–443.
Wang, T., 2011. Simulation of bubble column reactors using CFD coupled with a population balance model. Front. Chem. Sci. Eng. 5, 162–172.
Wang, T., Wang, J., 2007. Numerical simulations of gas–liquid mass transfer in bubble columns with a CFD–PBM coupled model. Chem. Eng. Sci. 62, 7107–7118.
WEF, 2010. Design of Municipal Wastewater Treatment Plants. WEF Press, McGraw-Hill Professional, and ASCE.
Wentzel, M.C., Comeau, Y., Ekama, G.A., van Loosdrecht, M.C.M., Brdjanovic, D., 2008. Enhanced biological phosphorus removal, Biological waste water treatment. Principles, modelling and design. IWA Publ. 155–170.
WHO and UNICEF, 2013. Progress on sanitation and drinking-water.
Xing, C., Wang, T., Wang, J., 2013. Experimental study and numerical simulation with a coupled CFD–PBM model of the effect of liquid viscosity in a bubble column. Chem. Eng. Sci. 95, 313–322.