16

MT-1732products.moravek.com/DataSheets/MT1732DS.pdf · MT-1732 Hypoxanthine, [8-3H]- Lot 212-131-009-A-20091216-DG A) All chromatograms were run using the HPLC method described on

  • Upload
    others

  • View
    0

  • Download
    0

Embed Size (px)

Citation preview

Page 1: MT-1732products.moravek.com/DataSheets/MT1732DS.pdf · MT-1732 Hypoxanthine, [8-3H]- Lot 212-131-009-A-20091216-DG A) All chromatograms were run using the HPLC method described on
Page 2: MT-1732products.moravek.com/DataSheets/MT1732DS.pdf · MT-1732 Hypoxanthine, [8-3H]- Lot 212-131-009-A-20091216-DG A) All chromatograms were run using the HPLC method described on

MT-1732

Hypoxanthine, [8-3H]-

Lot 212-131-009-A-20091216-DG

A) All chromatograms were run using the HPLC method described on the Product Data Sheet.

Concentrations and volumes:

Standard solution concentration was 1.0 mg/ml.

Hypoxanthine, [8-3H]- concentration was 0.2 mCi/ml. Volume of standard alone injection was 1.0 µl.

Volume of Hypoxanthine, [8-3H]- alone injection was 1.0 µl.

Co-injection solution consisted of 1.0 µl Hypoxanthine, [8-3H]- + 1.0 µl standard.

Volume of co-injection was 2.0 µl.

Volume of blank injection was 1.0 µl.

B) Mass spectrometry – Positive mode C) NMR

Page 3: MT-1732products.moravek.com/DataSheets/MT1732DS.pdf · MT-1732 Hypoxanthine, [8-3H]- Lot 212-131-009-A-20091216-DG A) All chromatograms were run using the HPLC method described on

MT-1732Hypoxanthine, [8-3H]-

Lot 212-131-009-A-20091216-DG

Chromatogram Info:

File Name : L:\int21298 File Created : 4/1/2014 1:49:01 PM

Origin : Acquired, Acquisition started 9/12/2011 11:56:43 AM Acquired Date : 9/12/2011 12:36:42 PM

Project : Test By : Administrator

Method : Unit2-40minrun By : Administrator

Description : Radiochemical trace of 3H material alone

Created : 6/16/2007 8:19 AM Modified : 4/1/2014 2:06 PM

Column : Detection : Radiochemical

Mobile Phase : Temperature :

Flow Rate : Pressure :

Note :

[min.]Time

10 20 30 40

[mV]

Volta

ge

0

200

400

600

3.6

1

8.7

2

9.7

3

10.2

4

10.8

5

11.1

6

L:\ int 21298 - Unit 2 Radio

4/1/2014 2:07 PM Page 1 of 2Chromatogram L:\int21298.PRM

Page 4: MT-1732products.moravek.com/DataSheets/MT1732DS.pdf · MT-1732 Hypoxanthine, [8-3H]- Lot 212-131-009-A-20091216-DG A) All chromatograms were run using the HPLC method described on

Result Table (Uncal - L:\int21298 - Unit 2 Radio)

CompoundName

Reten. Time[min]

Area[mV.s]

Height[mV]

Area[%]

Height[%]

Efficiency[th.pl]

Eff/l[t.p./m]

Symmetry/Tailing[-]

Response Factor

Resolution[-]

1 3.64 179.923 10.30 0.99 1.51 1348.21 26964.29 1.67

2 8.71 17966.207 664.54 98.54 97.63 3369.06 67381.27 3.27 10.2

3 9.75 16.774 1.86 0.09 0.27 82232.22 1644644.41 2.06 2.8

4 10.25 19.266 1.56 0.11 0.23 66773.08 1335461.66 4.06 3.4

5 10.77 10.096 1.22 0.06 0.18 44652.67 893053.47 1.86 2.9

6 11.10 40.863 1.21 0.22 0.18 6396.55 127931.08 2.94 0.9

Total 18233.130 680.69 100.00 100.00

4/1/2014 2:07 PM Page 2 of 2Chromatogram L:\int21298.PRM

Page 5: MT-1732products.moravek.com/DataSheets/MT1732DS.pdf · MT-1732 Hypoxanthine, [8-3H]- Lot 212-131-009-A-20091216-DG A) All chromatograms were run using the HPLC method described on

MT-1732Hypoxanthine, [8-3H]-

Lot 212-131-009-A-20091216-DG

Chromatogram Info:

File Name : L:\int21298 File Created : 4/1/2014 1:49:01 PM

Origin : Acquired, Acquisition started 9/12/2011 11:56:43 AM Acquired Date : 9/12/2011 12:36:42 PM

Project : Test By : Administrator

Method : Unit2-40minrun By : Administrator

Description : UV trace of 3H material alone

Created : 6/16/2007 8:19 AM Modified : 4/1/2014 2:05 PM

Column : Detection : UV 250nm

Mobile Phase : Temperature :

Flow Rate : Pressure :

Note :

[min.]Time

10 20 30 40

[mV]

Volta

ge

-1.0

-0.5

0.0

0.5

1.0

1.5

2.0

8.5

1

L:\ int 21298 - Unit 2 UV

4/1/2014 2:05 PM Page 1 of 2Chromatogram L:\int21298.PRM

Page 6: MT-1732products.moravek.com/DataSheets/MT1732DS.pdf · MT-1732 Hypoxanthine, [8-3H]- Lot 212-131-009-A-20091216-DG A) All chromatograms were run using the HPLC method described on

Result Table (Uncal - L:\int21298 - Unit 2 UV)

CompoundName

Reten. Time[min]

Area[mV.s]

Height[mV]

Area[%]

Height[%]

Efficiency[th.pl]

Eff/l[t.p./m]

Symmetry/Tailing[-]

Response Factor

Resolution[-]

1 8.46 23.885 1.48 100.00 100.00 7084.65 141693.02 0.85

Total 23.885 1.48 100.00 100.00

4/1/2014 2:05 PM Page 2 of 2Chromatogram L:\int21298.PRM

Page 7: MT-1732products.moravek.com/DataSheets/MT1732DS.pdf · MT-1732 Hypoxanthine, [8-3H]- Lot 212-131-009-A-20091216-DG A) All chromatograms were run using the HPLC method described on

MT-1732Hypoxanthine, [8-3H]-

Lot 212-131-009-A-20091216-DG

Chromatogram Info:

File Name : L:\int21299 File Created : 4/1/2014 1:49:01 PM

Origin : Acquired, Acquisition started 9/12/2011 1:29:33 PM Acquired Date : 9/12/2011 2:09:32 PM

Project : Test By : Administrator

Method : Unit2-40minrun By : Administrator

Description : UV trace of standard material alone

Created : 6/16/2007 8:19 AM Modified : 4/1/2014 2:07 PM

Column : Detection : UV 250nm

Mobile Phase : Temperature :

Flow Rate : Pressure :

Note :

[min.]Time

10 20 30 40

[mV]

Volta

ge

0

50

100

150

200

250

8.1

1

L:\ int 21299 - Unit 2 UV

4/1/2014 2:07 PM Page 1 of 2Chromatogram L:\int21299.PRM

Page 8: MT-1732products.moravek.com/DataSheets/MT1732DS.pdf · MT-1732 Hypoxanthine, [8-3H]- Lot 212-131-009-A-20091216-DG A) All chromatograms were run using the HPLC method described on

Result Table (Uncal - L:\int21299 - Unit 2 UV)

CompoundName

Reten. Time[min]

Area[mV.s]

Height[mV]

Area[%]

Height[%]

Efficiency[th.pl]

Eff/l[t.p./m]

Symmetry/Tailing[-]

Response Factor

Resolution[-]

1 8.13 4292.539 257.76 100.00 100.00 6023.18 120463.64 0.84

Total 4292.539 257.76 100.00 100.00

4/1/2014 2:07 PM Page 2 of 2Chromatogram L:\int21299.PRM

Page 9: MT-1732products.moravek.com/DataSheets/MT1732DS.pdf · MT-1732 Hypoxanthine, [8-3H]- Lot 212-131-009-A-20091216-DG A) All chromatograms were run using the HPLC method described on

MT-1732Hypoxanthine, [8-3H]-

Lot 212-131-009-A-20091216-DG

Chromatogram Info:

File Name : L:\int21300 File Created : 4/1/2014 1:49:01 PM

Origin : Acquired, Acquisition started 9/12/2011 2:57:53 PM Acquired Date : 9/12/2011 3:37:52 PM

Project : Test By : Administrator

Method : Unit2-40minrun By : Administrator

Description : Radiochemical trace of 3H material co-injected with standard

Created : 6/16/2007 8:19 AM Modified : 4/1/2014 2:08 PM

Column : Detection : Radiochemical

Mobile Phase : Temperature :

Flow Rate : Pressure :

Note :

[min.]Time

10 20 30 40

[mV]

Volta

ge

0

100

200

300

400

500

600

700

3.7

1

8.6

2

10.1

3

11.0

4

L:\ int 21300 - Unit 2 Radio

4/1/2014 2:09 PM Page 1 of 2Chromatogram L:\int21300.PRM

Page 10: MT-1732products.moravek.com/DataSheets/MT1732DS.pdf · MT-1732 Hypoxanthine, [8-3H]- Lot 212-131-009-A-20091216-DG A) All chromatograms were run using the HPLC method described on

Result Table (Uncal - L:\int21300 - Unit 2 Radio)

CompoundName

Reten. Time[min]

Area[mV.s]

Height[mV]

Area[%]

Height[%]

Efficiency[th.pl]

Eff/l[t.p./m]

Symmetry/Tailing[-]

Response Factor

Resolution[-]

1 3.69 250.583 13.01 1.25 1.77 1311.97 26239.42 2.11

2 8.59 19706.939 720.60 97.93 97.76 3151.77 63035.32 3.63 9.6

3 10.12 63.497 1.39 0.32 0.19 31914.89 638297.75 3.88 3.7

4 11.01 101.819 2.12 0.51 0.29 19284.74 385694.80 5.85 3.3

Total 20122.838 737.12 100.00 100.00

4/1/2014 2:09 PM Page 2 of 2Chromatogram L:\int21300.PRM

Page 11: MT-1732products.moravek.com/DataSheets/MT1732DS.pdf · MT-1732 Hypoxanthine, [8-3H]- Lot 212-131-009-A-20091216-DG A) All chromatograms were run using the HPLC method described on

MT-1732Hypoxanthine, [8-3H]-

Lot 212-131-009-A-20091216-DG

Chromatogram Info:

File Name : L:\int21300 File Created : 4/1/2014 1:49:01 PM

Origin : Acquired, Acquisition started 9/12/2011 2:57:53 PM Acquired Date : 9/12/2011 3:37:52 PM

Project : Test By : Administrator

Method : Unit2-40minrun By : Administrator

Description : UV trace of 3H material co-injected with standard

Created : 6/16/2007 8:19 AM Modified : 4/1/2014 2:09 PM

Column : Detection : UV 250nm

Mobile Phase : Temperature :

Flow Rate : Pressure :

Note :

[min.]Time

10 20 30 40

[mV]

Volta

ge

0

50

100

150

200

250 8.3

1

L:\ int 21300 - Unit 2 UV

4/1/2014 2:09 PM Page 1 of 2Chromatogram L:\int21300.PRM

Page 12: MT-1732products.moravek.com/DataSheets/MT1732DS.pdf · MT-1732 Hypoxanthine, [8-3H]- Lot 212-131-009-A-20091216-DG A) All chromatograms were run using the HPLC method described on

Result Table (Uncal - L:\int21300 - Unit 2 UV)

CompoundName

Reten. Time[min]

Area[mV.s]

Height[mV]

Area[%]

Height[%]

Efficiency[th.pl]

Eff/l[t.p./m]

Symmetry/Tailing[-]

Response Factor

Resolution[-]

1 8.33 4146.290 241.26 100.00 100.00 5835.27 116705.37 0.84

Total 4146.290 241.26 100.00 100.00

4/1/2014 2:09 PM Page 2 of 2Chromatogram L:\int21300.PRM

Page 13: MT-1732products.moravek.com/DataSheets/MT1732DS.pdf · MT-1732 Hypoxanthine, [8-3H]- Lot 212-131-009-A-20091216-DG A) All chromatograms were run using the HPLC method described on

MT-1732Hypoxanthine, [8-3H]-

Lot 212-131-009-A-20091216-DG

Chromatogram Info:

File Name : L:\int21297 File Created : 4/1/2014 1:49:01 PM

Origin : Acquired, Acquisition started 9/12/2011 10:38:16 AM Acquired Date : 9/12/2011 11:18:15 AM

Project : Test By : Administrator

Method : Unit2-40minrun By : Administrator

Description : UV trace of blank injection

Created : 6/16/2007 8:19 AM Modified : 4/1/2014 2:10 PM

Column : Detection : UV 250nm

Mobile Phase : Temperature :

Flow Rate : Pressure :

Note :

[min.]Time

10 20 30 40

[mV]

Volta

ge

-1

0

1

2

3

4

5

L:\ int 21297 - Unit 2 UV

4/1/2014 2:10 PM Page 1 of 2Chromatogram L:\int21297.PRM

Page 14: MT-1732products.moravek.com/DataSheets/MT1732DS.pdf · MT-1732 Hypoxanthine, [8-3H]- Lot 212-131-009-A-20091216-DG A) All chromatograms were run using the HPLC method described on

Result Table (Uncal - L:\int21297 - Unit 2 UV)

CompoundName

Reten. Time[min]

Area[mV.s]

Height[mV]

Area[%]

Height[%]

Efficiency[th.pl]

Eff/l[t.p./m]

Symmetry/Tailing[-]

Response Factor

Resolution[-]

No peak to report

4/1/2014 2:10 PM Page 2 of 2Chromatogram L:\int21297.PRM

Page 15: MT-1732products.moravek.com/DataSheets/MT1732DS.pdf · MT-1732 Hypoxanthine, [8-3H]- Lot 212-131-009-A-20091216-DG A) All chromatograms were run using the HPLC method described on

C:\Xcalibur\data\MT1732-20091216-DG 11/15/2010 10:48:14 AM MT1732-20091216-DG

RT: 0.00 - 2.00

0.0 0.5 1.0 1.5 2.0

Time (min)

0

5

10

15

20

25

30

35

40

45

50

55

60

65

70

75

80

85

90

95

100

Relative Abundance

1.531.51

1.90

0.27

0.42 1.36

1.16

0.47

1.141.120.51

0.69

0.771.66

1.09

NL:7.62E6

TIC MS MT1732-20091216-DG

MT1732-20091216-DG#1-344 RT: 0.00-2.00T: + c NSI Full ms [77.00-197.00]m/z= 133.32-145.83

m/z Intensity Relativem/z Intensity Relative

134.96 14666.2 1.43 136.30 7778.7 0.76 137.18 1027820.6 100.00 138.27 33784.5 3.29 139.17 446286.2 43.42 140.27 11161.2 1.09 141.16 6240.9 0.61 142.27 3421.3 0.33 143.11 6115.5 0.60 144.21 4604.5 0.45 145.19 6131.7 0.60

MT1732-20091216-DG #1-344 RT: 0.00-2.00 AV: 344 NL: 1.03E6T: + c NSI Full ms [77.00-197.00]

80 100 120 140 160 180

m/z

0

10

20

30

40

50

60

70

80

90

100

Relative Abundance

137.18

139.17

159.08149.09 183.02114.23 168.08 194.98134.96105.0192.1282.12

Page 16: MT-1732products.moravek.com/DataSheets/MT1732DS.pdf · MT-1732 Hypoxanthine, [8-3H]- Lot 212-131-009-A-20091216-DG A) All chromatograms were run using the HPLC method described on

98

76

54

32

1pp

m

8.0668.0698.1248.168

1.00

PC 1.00

GB 0

LB 0.00 Hz

SSB 0

WDW no

SF 320.1305850 MHz

SI 32768

SFO1 320.1321857 MHz

PL1 2.00 dB

P1 10.00 usec

NUC1 3H

======== CHANNEL f1 ========

TD0 1

D1 2.00000000 sec

TE 300.0 K

DE 6.00 usec

DW 81.000 usec

RG 46341

AQ 1.3271540 sec

FIDRES 0.376760 Hz

SWH 6172.839 Hz

DS 2

NS 5000

SOLVENT D2O

TD 16384

PULPROG zg

PROBHD 5 mm DUX 3H−1H

INSTRUM spect

Time 15.21

Date_ 20100210

PROCNO 1

EXPNO 1

NAME MT1732−20091216−DG

Batch 20091216−DG

MT1732 3H NMR in D2O