Upload
magnus-abner-adams
View
227
Download
0
Tags:
Embed Size (px)
Citation preview
Oliver Findl 1,2, Nino Hirnschall 1,2, Alja Crnej 2
1 VIROS Vienna Institute for Research in Ocular Surgery, Hanusch Hospital, Vienna, Austria2 Moorfields Eye Hospital, London, UK
Effect of Fluorescein dye staining of the tear film on Scheimpflug measurements of
central corneal thickness
Financial interest: noneOF is scientific advisor to Alcon, AMO, Croma & Carl Zeiss Meditec AG
Background
• Fluorescein dye is commonly used in routine ophthalmic examination
– staining epithelial defects of the cornea
– break up time of the tear film
– Goldmann applanation tonometry
• Correct CCT measurement needed for:
– corneal refractive surgery
– diagnosis of corneal diseases (keratoconus)
– to correct for systematic errors of IOP measurements using Goldmann applanation.
CCT measurement devices
• Pentacam: rotating Scheimpflug photographic
– measures the thickness from the image, which is created from the back-scattered light from the corneal tissue
=> Influenced by fluorescein eye drops?
• ACMaster: partial coherence interferometry (PCI)
– gold standard for CCT measurement
– measures distances between surface interfaces with a different refractive index
Methods
• Prospective randomised controlled study
• 30 eyes of 15 patients (8 female, 7 male)
Methods3 x rotating Scheimpflug imaging + 3 x PCI measurement
Instilling 0.25% fluorescein + 0.5% proxymetacaine hydrochloride
Rotating Scheimpflug imaging + PCI measurement
Irrigation with 2ml BSS
Rotating Scheimpflug imaging + PCI
Rotating Scheimpflug imaging + PCIafter 5min, 10min, 20min, 40min, 60min, (120min)
Without Fluorescein With Fluorescein
Results
Continuous measurement group
CCTScheimpflug (SD) in µm
CCTPCI (SD) in µm
Baseline 539.1 (32.2) 519.0 (33.1)
1min 585.7 (29.7) 520.6 (33.2)
5min 570.1 (31.6) 519.7 (33.0)
10min 560.8 (35.0) 519.8 (32.8)
20min 550.1 (33.4) 520.0 (32.7)
40min (10 patients)
545.0 (33.8) 523.2 (38.7)
Irrigation group
CCTScheimpflug (SD) in µm
CCTPCI (SD) in µm
Baseline 536.5 (34.7) 519.4 (33.9)
1min 586.5 (44.3) 521.7 (33.8)
After irrigating 561.7 (39.5) 520.7 (34.0)
Rotating Scheimpflug: 46.6µm difference
PCI: 1.6µm difference
Results
CC
T d
iffer
ence
(P
CI)
in µ
m
CCT difference (Scheimpflug) in µm
Results
time after applying Fluorescein (min)
Sub
ject
s w
ith F
luor
esce
in la
yer
(%)
Fluorescein layer -> more than 5µm difference to the baseline measurement
Number of patients (%) with a fluorescein layer +5µm compared to baseline
Results
time after applying Fluorescein (min)
Sch
irmer
2 r
esul
ts (
sec)
Conclusion
• Scheimpflug imaging
• Fluorescein dye causes intense back-scatter of light resulting in a measurement error
• Irrigating the eye reduces this effect only insignificantly
• PCI: is not influenced by altered or higher reflectivity
=> Fluorescein eye drops MUST NOT be used prior to Scheimpflug imaging measurements