2
PUPILLOGRAPHY Completely integrated with the topography of the ante- rior surface of the cornea, it performs the pupil measu- rement in scotopic (0.04 lux), mesopic (4 lux), photopic (50 lux) condions and in dynamic mode. Knowing the center and the diameter of the pupil, is essenal for many clinical procedures which seek to opmize vision quality. KERATOCONOUS SCREENING Keratoconous screening soſtware, provides the clinician with imporatant informaon about the paents cornea. Understanding this can help prevent complicaons asso- ciated with ectasia before corneal surgery is undertaken. ADVANCED ANALYSIS OF THE TEAR FILM Placido disk technology allows for the advanced analysis of the tear film such as NI-BUT (Non Invasive Break-up Time). OCULAR WAVEFRONT Phoenix soſtware enables the analysis of the paent’s ocular wavefront, both as a deviaon from the wavefront and as a refracve dfference from the ideal. DENSITOMETR For an objecve assessment of cataract and opcal me- dia opacity evaluaon, Osiris-T can acquire backlit ima- ges without reflecons. DYNAMIC ACCOMODATION Real-me measurement of the ocular wavefront is indi- spensable during the evaluaon of the accommodave phases. Customizable exam modes (ramps or square wa- ves) are available to evaluate the paent’s ability to focus at near. Osiris-T ABERROMETER | TOPOGRAPHER The informaon provided by Osiris-T, a corneal topo- grapher combined with a total ocular aberrometer, is in- dispensable for the correct evaluaon of paents who have, in addion to tradional low-order defects, even more complex ocular aberraons, either corneal or in- ternal. ABERROMETER: Osiris-T has a unique pyramidal sensor design that allows the aberrometer to measure aberraons with a resolu- on of 45,000 points (at the maximum pupil diameter), with a wide dynamic range. It is also able to measure the ocular wavefront in real me with a frame rate of up to 33 images per second: this makes it possible to measu- re and view changes in power and aberraons while the paent is accomodang. Phoenix soſtware offers a wide range of analysis opons, such as refracve error maps and visual simulaons (PSF, MTF and convoluon with optotype), which helps the clinician to understand and explain the paent’s visual problems. TORIC LENS ASSISTANT For the evaluaon of the performances of a toric system the combinaon of corneal topography imported from CSO topographers and ocular aberraon makes it possi- ble to disnguish whether any asgmac residue is due to a rotaon of the lens or to an incorrect calculaon. ANALYSIS SOFTWARE FOR ABERRATIONS The tool integrates with the Phoenix soſtware, offe- ring a wide range of analysis opons, such as refracve error maps and visual simulaons (PSF, MTF and con- voluon with optotype), which helps the clinician to understand and explain the paent’s visual problems. TOPOGRAPH: Using a reflecon topography system based on a 22 ring Placido disk, Osiris-T is able to measure the morphology and the corneal refracve component by means of the sa- gial curvature, tangenal curvature, elevaon and power maps. The availabilty of consolidated synthesis parame- ters makes the follow-up and diagnosis of keratoconus simple and intuive. The integraon of topographry and aberrometery enables the calculaon of the internal com- ponent of the wavefront. FEATURES OF THE PHOENIX SOFTWARE FEATURES OF THE PHOENIX SOFTWARE Osiris-T uses the Phoenix soſtware plaorm allowing paent data to be saved for future review and analysis, shared by all CSO devices.

Osiris-T TOPOGRAPH: FEATURES OF THE PHOENIX SOFTWARE

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

PUPILLOGRAPHYCompletely integrated with the topography of the ante- rior surface of the cornea, it performs the pupil measu- rement in scotopic (0.04 lux), mesopic (4 lux), photopic (50 lux) conditi ons and in dynamic mode. Knowing the center and the diameter of the pupil, is essenti al for many clinical procedures which seek to opti mize vision quality.

KERATOCONOUS SCREENINGKeratoconous screening soft ware, provides the clinician with imporatant informati on about the pati ents cornea. Understanding this can help prevent complicati ons asso-ciated with ectasia before corneal surgery is undertaken.

ADVANCED ANALYSIS OF THE TEAR FILMPlacido disk technology allows for the advanced analysis of the tear film such as NI-BUT (Non Invasive Break-up Time).

OCULAR WAVEFRONTPhoenix soft ware enables the analysis of the pati ent’s ocular wavefront, both as a deviati on from the wavefront and as a refracti ve dff erence from the ideal.

DENSITOMETRFor an objecti ve assessment of cataract and opti cal me-dia opacity evaluati on, Osiris-T can acquire backlit ima-ges without reflecti ons.

DYNAMIC ACCOMODATIONReal-ti me measurement of the ocular wavefront is indi-spensable during the evaluati on of the accommodati ve phases. Customizable exam modes (ramps or square wa-ves) are available to evaluate the pati ent’s ability to focus at near.

Osiris-TABERROMETER | TOPOGRAPHER The informati on provided by Osiris-T, a corneal topo-grapher combined with a total ocular aberrometer, is in-dispensable for the correct evaluati on of pati ents who have, in additi on to traditi onal low-order defects, even more complex ocular aberrati ons, either corneal or in-ternal.

ABERROMETER: Osiris-T has a unique pyramidal sensor design that allows the aberrometer to measure aberrati ons with a resolu-ti on of 45,000 points (at the maximum pupil diameter), with a wide dynamic range. It is also able to measure the

ocular wavefront in real ti me with a frame rate of up to 33 images per second: this makes it possible to measu-re and view changes in power and aberrati ons while the pati ent is accomodati ng.Phoenix soft ware off ers a wide range of analysis opti ons, such as refracti ve error maps and visual simulati ons (PSF, MTF and convoluti on with optotype), which helps the clinician to understand and explain the pati ent’s visual problems.

TORIC LENS ASSISTANTFor the evaluati on of the performances of a toric system the combinati on of corneal topography imported from CSO topographers and ocular aberrati on makes it possi-ble to disti nguish whether any asti gmati c residue is due to a rotati on of the lens or to an incorrect calculati on.

ANALYSIS SOFTWARE FOR ABERRATIONSThe tool integrates with the Phoenix soft ware, off e-ring a wide range of analysis opti ons, such as refracti ve error maps and visual simulati ons (PSF, MTF and con-voluti on with optotype), which helps the clinician to understand and explain the pati ent’s visual problems.

TOPOGRAPH:Using a reflecti on topography system based on a 22 ring Placido disk, Osiris-T is able to measure the morphology and the corneal refracti ve component by means of the sa-gitt al curvature, tangenti al curvature, elevati on and power maps. The availabilty of consolidated synthesis parame-ters makes the follow-up and diagnosis of keratoconus simple and intuiti ve. The integrati on of topographry and aberrometery enables the calculati on of the internal com-ponent of the wavefront.

FEATURES OF THE PHOENIX SOFTWARE FEATURES OF THE PHOENIX SOFTWARE Osiris-T uses the Phoenix soft ware platf orm allowing pati ent data to be saved for future review and analysis, shared by all CSO devices.

Osiris-TABERROMETER | TOPOGRAPHER

TECHNICAL DATA

Data transfer USB 3.0

Power supply external power source 24 VCCIn: 100-240Vac - 50/60Hz - 0.9-05A - Out: 24Vdc - 40W

Power net cable with plug C14

Dimensions (HxWxD) 515 x 315 x 255mm

Weight 6.9Kg

Chin rest movement 70mm ± 1mm

Minimum height of the chin cup from the table 24cm

Base movement (xyz) 105 x 110 x 30mm

Working distance 74mm

LIGHT SOURCES

Aberrometer Led @850nm

Fixati on Led @450-650nm

Placido Led @635nm

Pupillometry and auxiliary Led @780nm

ABERROMETRY

Points measured at maximum pupil 45000

Spati al resoluti on 41μm

Pupil size range 2-9mm

Dioptric range Sph from -25D to +15D; Cyl up to 10DRepeatability 0.05D on test eyes

TOPOGRAPHY

Placido rings 22Measured points 5632Topographic covering (in 43D) 10mmDioptric measurement range from 1D to 100DMeasurement accuracy Class A according to UNI EN ISO 19980-2012Compati bility with standard DICOM v3 (IHE integrati on profi le EYECARE Workfl ow)

MINIMUM SYSTEM REQUIREMENT

PC: 4 GB RAM - Video Card 1 GB RAM (not shared) resoluti on 1024 x 768 pixels - USB 3.0 type AOperati ng system: Windows XP, Windows 7 and Windows 10 (32/64 bit).

*The specifi cati ons and the images are not contractually binding and can be modifi ed without noti ce. Windows® is a Microsoft Corporati on trade mark.

CO127 | Rev. 00 del 01/2018

Osiris-TABERROMETER | TOPOGRAPHER

0051

YOUR PROFESSIONAL PARTNER SINCE 1967

0051

Via degli Stagnacci 12/E50018 - Scandicci - FI - Italytel +39 055 72219 | fax +39 055 7215557email. [email protected] | web. www.csoitalia.it

EN