7/28/2019 PID+ Model 102 Tech Training 613.pdf
1/51
PID Analyzers, LLC copyright2003-2013
Portable Analyzers 102 &
102+
PID (VOC)+, CG, TCD, IR
& EC Sensors
With this kind of Flexibilityin an Analyzer, You CanDesign Your Own IH/Safety
AnalyzerPID Analyzers, LLC
7/28/2019 PID+ Model 102 Tech Training 613.pdf
2/51
PID Analyzers, LLC copyright2003-2013
Portable Analyzer: Readout
& Head
7/28/2019 PID+ Model 102 Tech Training 613.pdf
3/51
PID Analyzers, LLC copyright2003-2013
102+
In the 102+ amp board, there are 5amplifiers:
PID- high input impedence- namps
CG, TCD, IR- Wheatstone bridge-measure diff. T for CG, TCD or microbolometer for IR
EC1- Biased or UnbiasedEC2- O2 or Unbiased
EC3- Unbiased
7/28/2019 PID+ Model 102 Tech Training 613.pdf
4/51
PID Analyzers, LLC copyright2003-2013
Schematic of PID
7/28/2019 PID+ Model 102 Tech Training 613.pdf
5/51
PID Analyzers, LLC copyright2003-2013
Photoionization Process
R + h = R+ + e Where:
R= a molecule with an IP = IP of
the molecule
R+ is the ion formed by the process
that is collected in the ion chamber-concentration is proportional to thecurrent produced
7/28/2019 PID+ Model 102 Tech Training 613.pdf
6/51
PID Analyzers, LLC copyright2003-2013
Model 102 & 102+ PIDs
Two Analyzers102 PID only
Can be upgraded
Optional 9.5 & 11.7 Snap on Heads102+ PID
Optional ppb mode (1-50,000 ppb)
1-3 additional sensors can be added for cost
of sensor at time of purchase or laterSensors: CG or TCD, EC (select from > 20)
Optional 9.5 & 11.7 Snap on Heads
7/28/2019 PID+ Model 102 Tech Training 613.pdf
7/51
PID Analyzers, LLC copyright2003-2013
IP (eV) Response Response Response
Detector 10.6 11.7 FID
Methanol 10.85 0 2.3 0.6
Acetoni tr ile 11.72 0 1.3 1.6
Toluene 8.83 9.3 11.4 7
THF 9.45 6.2 6.3 4.5
Acetic Anhydr ide 10.35 0.9 4 2.5
Dichloromethane 11.35 0.2 6.8 0.6
Pyridine 9.32 6.1 7 5
Benzene 9.25 10.0 10.0 6
7/28/2019 PID+ Model 102 Tech Training 613.pdf
8/51
PID Analyzers, LLC copyright2003-2013
PID Assembly
7/28/2019 PID+ Model 102 Tech Training 613.pdf
9/51
PID Analyzers, LLC copyright2003-2013
7/28/2019 PID+ Model 102 Tech Training 613.pdf
10/51
7/28/2019 PID+ Model 102 Tech Training 613.pdf
11/51
PID Analyzers, LLC copyright2003-2013
102 & 102+ Snap on Heads
102 102+ note additional board
7/28/2019 PID+ Model 102 Tech Training 613.pdf
12/51
PID Analyzers, LLC copyright2003-2013
Snap-On Modules
It is possible to have multiple heads forthe 102+ or other models- (electronics forhead are $350 plus sensors)
CG, TCD & IR use the same circuit but aresetup differently (current)-can haveseparate heads for each
Unbiased EC sensors can be interchanged
by customer, recalibrated then run Readout unit will recognize type of PID or
type of WB/sensor
7/28/2019 PID+ Model 102 Tech Training 613.pdf
13/51
PID Analyzers, LLC copyright2003-2013
Exploded View of 102
7/28/2019 PID+ Model 102 Tech Training 613.pdf
14/51
PID Analyzers, LLC copyright2003-2013
Amp & Sensor Boards
7/28/2019 PID+ Model 102 Tech Training 613.pdf
15/51
PID Analyzers, LLC copyright2003-2013
Sensor Board
7/28/2019 PID+ Model 102 Tech Training 613.pdf
16/51
PID Analyzers, LLC copyright2003-2013
Electronic Schematic
CPURS232
Keyboard
Digital- Switch
DetectorsReal Time ClockDiagnostics
DAC to 0-1V
ADC
Display
Data Storage in RAM
1-4
Sensors
AMP
7/28/2019 PID+ Model 102 Tech Training 613.pdf
17/51
PID Analyzers, LLC copyright2003-2013
Portable Analyzer Boards
Display board-Provides a display of the signal, errors, andfunctions
Processor board-A 32 bit processor is used to control theA/D, cal, other functions
Amplifier board-Signal from the ion chamber is amplifiedand sent to the processor board
Connector board-Power in from charger, RS232 out and PIDoutput connections
Interface board-Provided the input/output from theprocessor board
Low Voltage board-Power supply for HV board, pump,
electronics High Voltage board-for PID-Supplies the voltage for the
lamp and the bias for the ion chamber
7/28/2019 PID+ Model 102 Tech Training 613.pdf
18/51
PID Analyzers, LLC copyright2003-2013
7/28/2019 PID+ Model 102 Tech Training 613.pdf
19/51
PID Analyzers, LLC copyright2003-2013
7/28/2019 PID+ Model 102 Tech Training 613.pdf
20/51
PID Analyzers, LLC copyright2003-2013
Keypad for portableanalyzers
7/28/2019 PID+ Model 102 Tech Training 613.pdf
21/51
PID Analyzers, LLC copyright2003-2013
Keypad Functions for Portable
Analyzers
7/28/2019 PID+ Model 102 Tech Training 613.pdf
22/51
PID Analyzers, LLC copyright2003-2013
Cal Menu
7/28/2019 PID+ Model 102 Tech Training 613.pdf
23/51
PID Analyzers, LLC copyright2003-2013
RS232
The digital input/output is the telephone jack inthe base of the 102
A bluetooth option is available to connectwirelessly to a PC that is < 200 away
The RS232 is used to connect to the PC todownload data or display a continuous graph ofthe concentration of up to 4 sensorssimultaneously the software on the PC should
be the Grapher software that is included with the102
7/28/2019 PID+ Model 102 Tech Training 613.pdf
24/51
PID Analyzers, LLC copyright2003-2013
Grapher Software
Included with portable analyzer Need RS232 Cable & Software Can graph ,print & save data from
Analyzer Connect RS232 cable to PC Turn logging on Analyzer & setup Open Grapher and connect
Setup grapher: logging, channels, rangeand components Save data as csv file, save configuration
7/28/2019 PID+ Model 102 Tech Training 613.pdf
25/51
PID Analyzers, LLC copyright2003-2013
VOC, NH3, CO, O2
7/28/2019 PID+ Model 102 Tech Training 613.pdf
26/51
PID Analyzers, LLC copyright2003-2013
ppm & ppb Vocs
7/28/2019 PID+ Model 102 Tech Training 613.pdf
27/51
PID Analyzers, LLC copyright2003-2013
Log Menu
7/28/2019 PID+ Model 102 Tech Training 613.pdf
28/51
PID Analyzers, LLC copyright2003-2013
Applications for PID
Safety/IH Monitoring Emergency Response
Area Monitoring VOCs+ LEL +Toxics
Leak Detection for VOCs in
chemical/petrochemical plants Transportation- Loading/Unloading Chemicals
Monitoring Pilot Plants
Laboratory Monitoring
Short term; medium term monitoring- using thesleep mode
102+
7/28/2019 PID+ Model 102 Tech Training 613.pdf
29/51
PID Analyzers, LLC copyright2003-2013
102+
Features/Specifications
Detector 1 for VOCs: PID 10.2, PID 11.7Detector 2-4: CG, TCD, IR, electrochemistry
LEL, TCD, Electrochemical ( select from >20)
NEMA 2 Enclosure
10.75 L x 3 W x 2.25 D Weight 1.9 pounds
Linear to 0.1 to 3,000 ppm (10.2); 1-3,000ppm for (11.7)
LEL- 0-100%; O2 0-30%; Toxics- ppb-ppm LCD digital display
Fast response-
7/28/2019 PID+ Model 102 Tech Training 613.pdf
30/51
PID Analyzers, LLC copyright2003-2013
Key Benefits of Model 102
Most sensitive PID available
Only PID with 11.7 lamp that lasts > 30days
Electronic zero for PID
Easy to clean ion chamber & lamp
Most Sensitive PID; wide linear range to
>5,000 ppm
Easy to use software in 102
7/28/2019 PID+ Model 102 Tech Training 613.pdf
31/51
PID Analyzers, LLC copyright2003-2013
Sensors for Amplifier Channels
CH0 PID- ppb
CH1 PID- ppm or EC1CH2 CG or TCD or IR or EC1
(Wheatstone bridge or voltage)
CH3 O2 or EC2 (current or voltage)CH4 EC3 (biased or unbiased)
7/28/2019 PID+ Model 102 Tech Training 613.pdf
32/51
PID Analyzers, LLC copyright2003-2013
Electrochemical Reactions
Electrochemicalsensors generate asignal from
oxidation/reductionreaction that isproportional toconcentration
Oxidation:
CO = CO2 + 2H++ 2e-
Reduction:
Cl2 + 2e- = 2Cl-
7/28/2019 PID+ Model 102 Tech Training 613.pdf
33/51
PID Analyzers, LLC copyright2003-2013
Electrochemical Sensors
7/28/2019 PID+ Model 102 Tech Training 613.pdf
34/51
PID Analyzers, LLC copyright2003-2013
7/28/2019 PID+ Model 102 Tech Training 613.pdf
35/51
PID Analyzers, LLC copyright2003-2013
Uses for Electrochemical
Sensors
Used in portable instruments (Model 102,)for leak detection (measurement in ppmrange for toxics & % for O2); confined
space (CO, H2S), compliments PID thatmeasures HC from 0.1 to 3000 ppm
Other uses for EC (Model 1000)are forcontinuous monitoring of toxics in
industrial environments
7/28/2019 PID+ Model 102 Tech Training 613.pdf
36/51
PID Analyzers, LLC copyright2003-2013
Features of Portable ECS
Instant on response Long (2 year) lifetime sensors Robust instrument Rapid response with pump Can draw sample up to 25 away- you
dont have to go inside the area or tunnelto test the atmosphere
Fewer interferences Easy to change 7,000 logging capability; easy to download
data; no special software required
7/28/2019 PID+ Model 102 Tech Training 613.pdf
37/51
PID Analyzers, LLC copyright2003-2013
Addnal Features of EC sensors
Wide range of sensors (20) andgrowing
Instant on for portables
Long lifetime sensors (typically 2yrs.; some available up to 5 yrs.)
Fast response
Easy to interchange sensors or heads
Up to 3 EC sensors in Model 1004
7/28/2019 PID+ Model 102 Tech Training 613.pdf
38/51
PID Analyzers, LLC copyright2003-2013
Features of Combustible Gas
Sensor (CGS) Easy to calibrate
Cal values stored in RAM
The combustible gas sensor has a
library of 20 compounds. This allowsthe user to calibrate with one gas andvia the library function, correct forresponse and LEL value for any of the> 20 gases in the library
7/28/2019 PID+ Model 102 Tech Training 613.pdf
39/51
PID Analyzers, LLC copyright2003-2013
Types of Sensors-LEL
LEL Two heated catalytic
elements are set in ablock; one is sealed; theother is in contact with thesample. Each is in a leg of
a Wheatstone Bridgecircuit. If the samplecontains hydrocarbons orcombustible inorganiccompounds, they will burnincreasing the temperatureof the sensors andchanging the resistance.The change in resistance isproportional to theconcentration.
7/28/2019 PID+ Model 102 Tech Training 613.pdf
40/51
PID Analyzers, LLC copyright2003-2013
Wheatstone Bridge Circuitry
Wheatstonebridge circuitry
is used for a
variety of sensorssuch as:Combustible Gas,Thermal
Conductivity,Semiconductorand IR sensors
7/28/2019 PID+ Model 102 Tech Training 613.pdf
41/51
PID Analyzers, LLC copyright2003-2013
Combustible Gas Sensor
Sensitivities
7/28/2019 PID+ Model 102 Tech Training 613.pdf
42/51
PID Analyzers, LLC copyright2003-2013
Applications for TCD
Used in portable instruments (Model102 & 102+) for leak detection(measurement in % range to 100%);
compliments PID that measures from0.1 to 3000 ppm
Other uses for TCD (Model 204)are
for continuous monitoring of H2, inlarge power plant transformers
7/28/2019 PID+ Model 102 Tech Training 613.pdf
43/51
7/28/2019 PID+ Model 102 Tech Training 613.pdf
44/51
PID Analyzers, LLC copyright2003-2013
TCD Sensitivities
7/28/2019 PID+ Model 102 Tech Training 613.pdf
45/51
PID Analyzers, LLC copyright2003-2013
Features of TCD
Useful for measurement of compounds withthermal conductivity different from air (1.0)
Easy to calibrate
Can be turned off to maximize batterylifetime
Cal values stored in RAM
The thermal conductivity sensor has a libraryof 15 compounds. This allows the user tocalibrate with one gas and via the libraryfunction, correct for response and LEL valuefor any of the > 15 gases in the library
7/28/2019 PID+ Model 102 Tech Training 613.pdf
46/51
IR Sensor-CO2 , CH4 & N2O
Dual beam IR sensor Spectra for IR Sensors
PID Analyzers, LLC copyright2003-2013
D t L i S ft 1 4
7/28/2019 PID+ Model 102 Tech Training 613.pdf
47/51
PID Analyzers, LLC copyright2003-2013
Data Logging Software 1-4
Channels
7/28/2019 PID+ Model 102 Tech Training 613.pdf
48/51
PID Analyzers, LLC copyright
2003-2013
Sensor Configurations
Sensor 1 PID or EC1 Sensor 2 LEL,TCD or EC1 Sensor 3 EC2- O2 Sensor 4 EC2- other
The uP recognizes the proper sensor
configuration and sets up theappropriate library for the sensor ineach configuration
7/28/2019 PID+ Model 102 Tech Training 613.pdf
49/51
PID Analyzers, LLC copyright
2003-2013
Typical Users
Health & Safety- Director of Safety,Director of environmental health & safety,Safety officer, safety technician, industrial
hygienist Plant- Instrumentation technician,
Maintenance supervisor, Plant engineer
Environmental consultants- Project
manager, project engineer, field manager
HVAC- Engineer or Maintenance Manager
7/28/2019 PID+ Model 102 Tech Training 613.pdf
50/51
PID Analyzers, LLC copyright
2003-2013
Summary Model 102+
This Model 102+ has:PID sensor for VOCs
Ability to have 4 sensors in a single
analyzer LEL sensor for high levels
>20 electrochemical sensors tochoose from
Ability to build your own AnalyzerCompact size
PC software for downloading
7/28/2019 PID+ Model 102 Tech Training 613.pdf
51/51
PID Analyzers LLC copyright
Summary (cont.)
PID Analyzers has a very wide range ofportable analyzers that can be used tomeasure:
VOCs, O2, LEL, H2S, HCl, ETO, NO, NO2,SO2, CO, CO2, H2, PH3, SiH4, B2H6,Formaldehyde, HCN, NH3, AsH3, H2O, T
The Technologies include: PID, CG, TCD,IR, Electrochemistry (4 types- 30 differentsensors ), Capacitance, RTD