Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: Uric Acid Catalog # : U7581
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Uric Acid Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 2.5 – 7.7
C
Linear: <= 20, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 1 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours ********************************************************************************************************************
Add Reagent [Uric Acid Reagent], Volume [200.0 ul] ASSAY STEPS
Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [505], Differential Filter [630] Add Sample, Volume [5.0 ul] Aspiration Speed=2, Air Gap=1ul, Dispense Speed=1, Dispense Height=High Incubate 00:10:00 [0h 10min 0sec] Read with Primary Filter [505], Differential Filter [630] Document number: APP-U7581-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: Albumin Catalog # : A7502
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Albumin Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 3.5 – 5.3
C
Linear: 0.5 – 8.0, Use Reflex Test, Dilution Ratio = 2 Units: g/dl # of decimals: 1 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************
Add Reagent [Albumin Reagent], Volume [200.0 ul] ASSAY STEPS
Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:00 [0h 1 min 0 sec] Read with Primary Filter [630], Differential Filter [700] Add Sample, Volume [2.0 ul] Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low Incubate 00:01:00 [0h 1 min 0 sec] Read with Primary Filter [630], Differential Filter [700] Document number: APP-A7502-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: Alkaline Phosphatase Catalog # : A7516
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Alk Phos Assay Type: Chemistry Assay Mode: Factor Factor = 2187 (Adjusted 3988.1) Version: 0 Temperature: 37 ◦Normal: 35 – 123
C
Linear: <= 800, Use Reflex Test, Dilution Ratio = 2 Units: U/L # of decimals: 0 Assay Description: ******************************************************************************************************************** Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************
Add Sample, Volume [5.0 ul] ASSAY STEPS
Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low Add Reagent [Alk Phos working Reagent], Volume [200.0 ul] Aspiration Speed=2, Air Gap=50ul, Prewarm Time=15sec, Dispense Speed=1, Dispense Height=High Incubate 00:01:00 [0h 1 min 0 sec] Kinetic Read, Filters [450/630], Read Interval [60sec], 3 Times Initial Absorbance < 1.000 Delta Absorbance per Minute < 0.3657 Use Reflex Test. Dilution Ration = 2 Maximum SD of Delta Absorbance per Minute = 0.0300 Adjust absorbance ranges to path length. Document number: APP-A7516-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: ALT (SGPT) Catalog # : A7526
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: ALT (SGPT) Assay Type: Chemistry Assay Mode: Factor Factor = -1768 (Adjusted -3004.2) Version: 0 Temperature: 37 ◦Normal: 4 – 36
C
Linear: <= 500, Use Reflex Test, Dilution Ratio = 2 Units: U/L # of decimals: 0 Assay Description: ******************************************************************************************************************** Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************
Add Sample, Volume [20.0 ul] ASSAY STEPS
Aspiration Speed=4, Air Gap=1ul, Dispense Speed=2, Dispense Height=Low Add Reagent [ALT working Reagent], Volume [200.0 ul] Aspiration Speed=1, Air Gap=50ul, Prewarm Time=15sec, Dispense Speed=1, Dispense Height=High Incubate 00:01:30 [0h 1min 30sec] Kinetic Read, Filters [340/405], Read Interval [60sec], 4 Times 0.8000 < Initial Absorbance Delta Absorbance per Minute < 0.1700 Use Reflex Test. Dilution Ration = 2 Maximum SD of Delta Absorbance per Minute = 0.0300 Adjust absorbance ranges to path length. Document number: APP-A7526-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: Amylase Catalog # : A7564
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Amylase Assay Type: Chemistry Assay Mode: Factor Factor = 3178 (Adjusted 5795.3) Version: 0 Temperature: 37 ◦Normal: 25 – 125
C
Linear: <= 2000, Use Reflex Test, Dilution Ratio = 2 Units: U/L # of decimals: 0 Assay Description: ******************************************************************************************************************** Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************
Add Sample, Volume [5.0 ul] ASSAY STEPS
Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low Add Reagent [Amylase Reagent], Volume [200.0 ul] Aspiration Speed=2, Air Gap=50ul, Prewarm Time=15sec, Dispense Speed=1, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Kinetic Read, Filters [405/630], Read Interval [60sec], 3 Times Delta Absorbance per Minute < 0.6290 Use Reflex Test. Dilution Ration = 2 Maximum SD of Delta Absorbance per Minute = 0.0300 Adjust absorbance ranges to path length. Document number: APP-A7564-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: AST (SGOT) Catalog # : A7561
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: AST (SGOT) Assay Type: Chemistry Assay Mode: Factor Factor = -1768 (Adjusted -3004.2) Version: 0 Temperature: 37 ◦Normal: 5 – 34
C
Linear: <= 500, Use Reflex Test, Dilution Ratio = 2 Units: U/L # of decimals: 0 Assay Description: ******************************************************************************************************************** Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************
Add Sample, Volume [20.0 ul] ASSAY STEPS
Aspiration Speed=4, Air Gap=1ul, Dispense Speed=2, Dispense Height=Low Add Reagent [AST working Reagent], Volume [200.0 ul] Aspiration Speed=1, Air Gap=50ul, Prewarm Time=15sec, Dispense Speed=1, Dispense Height=High Incubate 00:01:30 [0h 1min 30sec] Kinetic Read, Filters [340/405], Read Interval [60sec], 4 Times 0.8000 < Initial Absorbance Delta Absorbance per Minute < 0.1700 Use Reflex Test. Dilution Ration = 2 Maximum SD of Delta Absorbance per Minute = 0.0300 Adjust absorbance ranges to path length. Document number: APP-A7561-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: auto HDL Catalog # : H7545
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: HDL Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 30 – 85
C
Linear: 2 – 150, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 0 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: autoHDL/LDL Calibrator, Curve valid interval: 7 days 0 hours ********************************************************************************************************************
Add Sample, Volume [3.0 ul] ASSAY STEPS
Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low Add Reagent [HDL R1 Reagent], Volume [180.0 ul] Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:05:00 [0h 5min 0sec] Read with Primary Filter [600], Differential Filter [700] Add Reagent [HDL R2 Reagent], Volume [60.0 ul] Aspiration Speed=0, Air Gap=25ul, Prewarm Time=0sec, Dispense Speed=1, Dispense Height=High Incubate 00:05:00 [0h 5min 0sec] Read with Primary Filter [600], Differential Filter [700] Document number: APP-H7545-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: Direct Bilirubin Catalog # : B7538
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Direct Bilirubin Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 0.0 – 0.5
C
Linear: <= 20.0, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 1 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************
Add Sample, Volume [25.0 ul] ASSAY STEPS
Aspiration Speed=4, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low Add Reagent [Direct Bilirubin R1], Volume [180.0 ul] Aspiration Speed=1, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=1, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [545], Differential Filter [700] Add Reagent [Direct Bilirubin R2], Volume [20.0 ul] Aspiration Speed=1, Air Gap=25ul, Prewarm Time=0sec, Dispense Speed=1, Dispense Height=High Incubate 00:05:00 [0h 5min 0sec] Read with Primary Filter [545], Differential Filter [700] Document number: APP-B7538-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: Total Bilirubin Catalog # : B7576
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Total Bilirubin Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 0.2 – 1.2
C
Linear: <= 20.0, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 1 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************
Add Reagent [Total Bilirubin working Reagent], Volume [200.0 ul] ASSAY STEPS
Aspiration Speed=1, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=1, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [545], Differential Filter [630] Add Sample, Volume [10.0 ul] Aspiration Speed=2, Air Gap=1ul, Dispense Speed=1, Dispense Height=High Incubate 00:05:00 [0h 5min 0sec] Read with Primary Filter [545], Differential Filter [630] Document number: APP-B7576-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: BUN Catalog # : B7552
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: BUN Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 7 – 18
C
Linear: <= 150, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 0 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************
Add Sample, Volume [2.0 ul] ASSAY STEPS
Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low Add Reagent [BUN working Reagent], Volume [200.0 ul] Aspiration Speed=1, Air Gap=50ul, Prewarm Time=15sec, Dispense Speed=1, Dispense Height=High Incubate 00:01:15 [0h 1min 15sec] Kinetic Read, Filters [340/630], Fixed Time [2min 30sec] Document number: APP-B7552-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: Calcium AZ II Catalog # : C7529
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Ca AZ II Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 8.5 – 10.4
C
Linear: <= 15.0, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 1 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours ********************************************************************************************************************
Add Reagent [Calcium AZ II Reagent], Volume [200.0 ul] ASSAY STEPS
Aspiration Speed=1, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [600], Differential Filter [700] Add Sample, Volume [2.0 ul] Aspiration Speed=1, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low Incubate 00:03:00 [0h 3min 0sec] Read with Primary Filter [600], Differential Filter [700] Document number: APP-C7529-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: Calcium CPC Catalog # : C7503
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Calcium CPC Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 8.5 – 10.5
C
Linear: <= 20.0, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 1 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************
Add Reagent [Calcium Color Reagent], Volume [100.0 ul] ASSAY STEPS
Aspiration Speed=2, Air Gap=25ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Add Reagent [Calcium Buffer Reagent], Volume [100.0 ul] Aspiration Speed=2, Air Gap=25ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [545], Differential Filter [700] Add Sample, Volume [4.0 ul] Aspiration Speed=2, Air Gap=1ul, Dispense Speed=1, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [545], Differential Filter [700] Document number: APP-C7503-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: CK Catalog # : C7522
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: CK Assay Type: Chemistry Assay Mode: Factor Factor = 6592 (Adjusted 12021.0) Version: 0 Temperature: 37 ◦Normal: <= 160
C
Linear: <= 1500, Use Reflex Test, Dilution Ratio = 2 Units: U/L # of decimals: 0 Assay Description: ******************************************************************************************************************** Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************
Add Sample, Volume [5.0 ul] ASSAY STEPS
Aspiration Speed=4, Air Gap=1ul, Dispense Speed=2, Dispense Height=Low Add Reagent [CK working Reagent], Volume [200.0 ul] Aspiration Speed=1, Air Gap=50ul, Prewarm Time=15sec, Dispense Speed=1, Dispense Height=High Incubate 00:02:00 [0h 2 min 0 sec] Kinetic Read, Filters [340/405], Read Interval [60sec], 2 Times Document number: APP-C7522-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: Chloride Catalog # : C7501
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Chloride Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 98 – 106
C
Linear: 80 – 120, Use Reflex Test, Dilution Ratio = 2 Units: mEq/L # of decimals: 0 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************
Add Reagent [Chloride Reagent], Volume [200.0 ul] ASSAY STEPS
Aspiration Speed=1, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=1, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [505], Differential Filter [630] Add Sample, Volume [2.0 ul] Aspiration Speed=2, Air Gap=2ul, Dispense Speed=1, Dispense Height=High Incubate 00:05:00 [0h 5min 0sec] Read with Primary Filter [505], Differential Filter [630] Document number: APP-C7501-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: CO2 Catalog # : C7502
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: CO2 Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 23 – 34
C
Linear: 2 - 40, Use Reflex Test, Dilution Ratio = 2 Units: mEq/L # of decimals: 0 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours ********************************************************************************************************************
ASSAY STEPS
Add Sample, Volume [2.0 ul] Aspiration Speed=1, Air Gap=2ul, Dispense Speed=1, Dispense Height=Low Add Reagent [CO2 Reagent], Volume [200.0 ul] Aspiration Speed=1, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:30 [0h 1min 30sec] Kinetic Read, Filters [405/505], Fixed Time [5min 0sec] Document number: APP-C7502-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: Cholesterol Catalog # : C7510
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Cholesterol Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: < 200
C
Linear: <=700, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 0 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours ********************************************************************************************************************
Add Reagent [Cholesterol Reagent], Volume [240.0 ul] ASSAY STEPS
Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [505], Differential Filter [630] Add Sample, Volume [2.4 ul] Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low Incubate 00:10:00 [0h 10min 0sec] Read with Primary Filter [505], Differential Filter [630] Document number: APP-C7510-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: Creatinine Catalog # : C7539
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Creatinine Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 0.4 – 1.4
C
Linear: <= 25.0, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 1 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours ********************************************************************************************************************
ASSAY STEPS
Add Sample, Volume [10.0 ul] Aspiration Speed=4, Air Gap=1ul, Dispense Speed=2, Dispense Height=Low Add Reagent [Creatinine working Reagent], Volume [200.0 ul] Aspiration Speed=1, Air Gap=50ul, Prewarm Time=15sec, Dispense Speed=1, Dispense Height=High Incubate 00:00:50 [0h 0min 50sec] Kinetic Read, Filters [505/630], Fixed Time [2min 30sec] Document number: APP-C7539-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: Phosphorus Catalog # : P7516
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Phosphorus Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 2.5 – 4.8
C
Linear: <= 12.0, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 1 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours ********************************************************************************************************************
Add Reagent [Phosphorus Reagent], Volume [200.0 ul] ASSAY STEPS
Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [340], Differential Filter [630] Add Sample, Volume [4.0 ul] Aspiration Speed=2, Air Gap=1ul, Dispense Speed=1, Dispense Height=Low Incubate 00:05:00 [0h 5min 0sec] Read with Primary Filter [340], Differential Filter [405] Document number: APP-P7516-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: Gamma GT Catalog # : G7571
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: GGT Assay Type: Chemistry Assay Mode: Factor Factor = 2211 (Adjusted 3935.9) Version: 0 Temperature: 37 ◦Normal: 9 – 54
C
Linear: <= 800, Use Reflex Test, Dilution Ratio = 2 Units: U/L # of decimals: 0 Assay Description: ******************************************************************************************************************** Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************
Add Sample, Volume [10.0 ul] ASSAY STEPS
Aspiration Speed=4, Air Gap=1ul, Dispense Speed=2, Dispense Height=Low Add Reagent [GGT working Reagent], Volume [200.0 ul] Aspiration Speed=1, Air Gap=50ul, Prewarm Time=15sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:30 [0h 1min 30sec] Kinetic Read, Filters [405/630], Read Interval [60sec], 3 Times Initial Absorbance < 0.8500 Delta Absorbance per Minute < 0.4520 Use Reflex Test. Dilution Ration = 2 Maximum SD of Delta Absorbance per Minute = 0.0300 Adjust absorbance ranges to path length. Document number: APP-G7571-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: Glucose (Hex) Catalog # : G7517
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Glucose (Hex) Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 70 – 110
C
Linear: <= 500, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 0 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours ********************************************************************************************************************
Add Reagent [Glucose (Hex) Reagent], Volume [240.0 ul] ASSAY STEPS
Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [340], Differential Filter [405] Add Sample, Volume [2.4 ul] Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low Incubate 00:05:00 [0h 5min 0sec] Read with Primary Filter [340], Differential Filter [405] Document number: APP-G7517-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: Glucose (Ox) Catalog # : G7521
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Glucose (Ox) Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 70 – 105
C
Linear: <= 500, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 0 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours ********************************************************************************************************************
Add Reagent [Glucose (Ox) Reagent], Volume [240.0 ul] ASSAY STEPS
Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [505], Differential Filter [630] Add Sample, Volume [2.4 ul] Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low Incubate 00:10:00 [0h 10min 0sec] Read with Primary Filter [505], Differential Filter [630] Document number: APP-G7521-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: HbA1c Catalog # : H7541
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: HbA1c Assay Type: Chemistry Assay Mode: Cubic Spline Version: 1 Temperature: 37 ◦Normal: < 6%
C
Linear: 2 – 16 Units: % # of decimals: 1 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 14 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: autoHDL/LDL Calibrator, Curve valid interval: 7 days 0 hours ********************************************************************************************************************
Add Sample, Volume [5.0 ul] ASSAY STEPS
Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low Add Reagent [HBa1c R1 Reagent], Volume [150.0 ul] Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:05:00 [0h 5min 0sec] Add Reagent [HbA1c R2 Reagent], Volume [50.0 ul] Aspiration Speed=0, Air Gap=25ul, Prewarm Time=0sec, Dispense Speed=1, Dispense Height=High Incubate 00:05:00 [0h 5min 0sec] Read with Primary Filter [630], Differential Filter [None] Document number: APP-H7541-CW2910 Rev. 7-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: Iron, Total Catalog # : I7505
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Iron, Total Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 60 – 150
C
Linear: <= 500, Use Reflex Test, Dilution Ratio = 2 Units: ug/dl # of decimals: 0 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours ********************************************************************************************************************
Add Sample, Volume [30.0 ul] ASSAY STEPS
Aspiration Speed=2, Air Gap=2ul, Dispense Speed=1, Dispense Height=Low Add Reagent [Iron Buffer Reagent], Volume [150.0 ul] Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:05:00 [0h 5min 0sec] Read with Primary Filter [545], Differential Filter [700] Add Reagent [Iron Color Reagent], Volume [3.0 ul] Aspiration Speed=2, Air Gap=2ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:10:00 [0h 10min 0sec] Read with Primary Filter [545], Differential Filter [700] Document number: APP-I7505-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: LDH Catalog # : L7572
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: LDH Assay Type: Chemistry Assay Mode: Factor Factor = 3376 (Adjusted 6010.0) Version: 0 Temperature: 37 ◦Normal: 80 – 285
C
Linear: <= 1000, Use Reflex Test, Dilution Ratio = 2 Units: U/L # of decimals: 0 Assay Description: ******************************************************************************************************************** Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************
Add Sample, Volume [5.0 ul] ASSAY STEPS
Aspiration Speed=4, Air Gap=1ul, Dispense Speed=2, Dispense Height=Low Add Reagent [LDH working Reagent], Volume [200.0 ul] Aspiration Speed=1, Air Gap=50ul, Prewarm Time=15sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:30 [0h 1min 30sec] Kinetic Read, Filters [340/405], Read Interval [60sec], 4 Times 0.8000 < Initial Absorbance Delta Absorbance per Minute < 0.1210 Use Reflex Test. Dilution Ration = 2 Maximum SD of Delta Absorbance per Minute = 0.0300 Adjust absorbance ranges to path length. Document number: APP-L7572-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: Lipase Catalog # : L7503
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Lipase Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 0 – 62
C
Linear: <= 600, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 0 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Lipase Standard, Curve valid interval: 7 days 0 hours ********************************************************************************************************************
Add Sample, Volume [2.5 ul] ASSAY STEPS
Aspiration Speed=4, Air Gap=1ul, Dispense Speed=2, Dispense Height=Low Add Reagent [Lipase Substrate (R1)], Volume [150.0 ul] Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:04:00 [0h 4min 0sec] Add Reagent [Lipase Activator (R2)], Volume [50.0 ul] Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:02:00 [0h 2min 0sec] Kinetic Read, Filters [545/700], Fixed Time [1min 0sec] Document number: APP-L7503-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: Magnesium Catalog # : HM929
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Magnesium Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 1.5 – 2.6
C
Linear: <= 4.86, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 2 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************
Add Reagent [Magnesium R1 Reagent], Volume [150.0 ul] ASSAY STEPS
Aspiration Speed=2, Air Gap=25ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Add Reagent [Magnesium R2 Reagent], Volume [150.0 ul] Aspiration Speed=2, Air Gap=25ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [545], Differential Filter [700] Add Sample, Volume [2.0 ul] Aspiration Speed=2, Air Gap=1ul, Dispense Speed=1, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [545], Differential Filter [700] Document number: APP-HM929-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: Microprotein Catalog # : P7582
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Microprotein Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 28 – 141
C
Linear: 2 – 250, Use Reflex Test, Dilution Ratio = 2 Units: mg/24 hrs # of decimals: 0 Assay Description: for CSF and Urine ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Microprotein Standard, Curve valid interval: 7 days 0 hours ********************************************************************************************************************
Add Reagent [Microprotein Reagent], Volume [200.0 ul] ASSAY STEPS
Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [600], Differential Filter [700] Add Sample, Volume [4.0 ul] Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low Incubate 00:05:00 [0h 5min 0sec] Read with Primary Filter [600], Differential Filter [700] Document number: APP-P7582-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: Total Protein Catalog # : T7528
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Total Protein Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 6.2 – 8.5
C
Linear: 1.0 – 15.0, Use Reflex Test, Dilution Ratio = 2 Units: g/dl # of decimals: 1 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours ********************************************************************************************************************
Add Reagent [Total Protein Reagent], Volume [200.0 ul] ASSAY STEPS
Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [545], Differential Filter [700] Add Sample, Volume [4.0 ul] Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low Incubate 00:05:00 [0h 5min 0sec] Read with Primary Filter [545], Differential Filter [700] Document number: APP-T7528-CW2910 Rev. 6-12
Chemwell 2902 / 2910 Instrument Application
Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com
Test: Triglycerides Catalog # : T7532
The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Triglycerides Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 44 – 148
C
Linear: <= 1000, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 0 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours ********************************************************************************************************************
Add Reagent [Triglycerides Reagent], Volume [200.0 ul] ASSAY STEPS
Aspiration Speed=1, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=1, Dispense Height=High Incubate 00:02:00 [0h 2min 0sec] Read with Primary Filter [505], Differential Filter [630] Add Sample, Volume [2.5 ul] Aspiration Speed=2, Air Gap=1ul, Dispense Speed=1, Dispense Height=High Incubate 00:10:00 [0h 10min 0sec] Read with Primary Filter [505], Differential Filter [630] Document number: APP-T7532-CW2910 Rev. 6-12