29
Chemwell 2902 / 2910 Instrument Application Phone: 734-487-8300 Toll Free: 800-757-5313 Fax: 734-483-1592 www.pointescientific.com Test: Uric Acid Catalog # : U7581 The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Uric Acid Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 Normal: 2.5 – 7.7 C Linear: <= 20, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 1 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours ******************************************************************************************************************** Add Reagent [Uric Acid Reagent], Volume [200.0 ul] ASSAY STEPS Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [505], Differential Filter [630] Add Sample, Volume [5.0 ul] Aspiration Speed=2, Air Gap=1ul, Dispense Speed=1, Dispense Height=High Incubate 00:10:00 [0h 10min 0sec] Read with Primary Filter [505], Differential Filter [630] Document number: APP-U7581-CW2910 Rev. 6-12

pointe chemwell 2910.pdf

Embed Size (px)

Citation preview

Page 1: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: Uric Acid Catalog # : U7581

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Uric Acid Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 2.5 – 7.7

C

Linear: <= 20, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 1 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours ********************************************************************************************************************

Add Reagent [Uric Acid Reagent], Volume [200.0 ul] ASSAY STEPS

Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [505], Differential Filter [630] Add Sample, Volume [5.0 ul] Aspiration Speed=2, Air Gap=1ul, Dispense Speed=1, Dispense Height=High Incubate 00:10:00 [0h 10min 0sec] Read with Primary Filter [505], Differential Filter [630] Document number: APP-U7581-CW2910 Rev. 6-12

Page 2: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: Albumin Catalog # : A7502

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Albumin Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 3.5 – 5.3

C

Linear: 0.5 – 8.0, Use Reflex Test, Dilution Ratio = 2 Units: g/dl # of decimals: 1 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************

Add Reagent [Albumin Reagent], Volume [200.0 ul] ASSAY STEPS

Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:00 [0h 1 min 0 sec] Read with Primary Filter [630], Differential Filter [700] Add Sample, Volume [2.0 ul] Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low Incubate 00:01:00 [0h 1 min 0 sec] Read with Primary Filter [630], Differential Filter [700] Document number: APP-A7502-CW2910 Rev. 6-12

Page 3: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: Alkaline Phosphatase Catalog # : A7516

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Alk Phos Assay Type: Chemistry Assay Mode: Factor Factor = 2187 (Adjusted 3988.1) Version: 0 Temperature: 37 ◦Normal: 35 – 123

C

Linear: <= 800, Use Reflex Test, Dilution Ratio = 2 Units: U/L # of decimals: 0 Assay Description: ******************************************************************************************************************** Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************

Add Sample, Volume [5.0 ul] ASSAY STEPS

Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low Add Reagent [Alk Phos working Reagent], Volume [200.0 ul] Aspiration Speed=2, Air Gap=50ul, Prewarm Time=15sec, Dispense Speed=1, Dispense Height=High Incubate 00:01:00 [0h 1 min 0 sec] Kinetic Read, Filters [450/630], Read Interval [60sec], 3 Times Initial Absorbance < 1.000 Delta Absorbance per Minute < 0.3657 Use Reflex Test. Dilution Ration = 2 Maximum SD of Delta Absorbance per Minute = 0.0300 Adjust absorbance ranges to path length. Document number: APP-A7516-CW2910 Rev. 6-12

Page 4: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: ALT (SGPT) Catalog # : A7526

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: ALT (SGPT) Assay Type: Chemistry Assay Mode: Factor Factor = -1768 (Adjusted -3004.2) Version: 0 Temperature: 37 ◦Normal: 4 – 36

C

Linear: <= 500, Use Reflex Test, Dilution Ratio = 2 Units: U/L # of decimals: 0 Assay Description: ******************************************************************************************************************** Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************

Add Sample, Volume [20.0 ul] ASSAY STEPS

Aspiration Speed=4, Air Gap=1ul, Dispense Speed=2, Dispense Height=Low Add Reagent [ALT working Reagent], Volume [200.0 ul] Aspiration Speed=1, Air Gap=50ul, Prewarm Time=15sec, Dispense Speed=1, Dispense Height=High Incubate 00:01:30 [0h 1min 30sec] Kinetic Read, Filters [340/405], Read Interval [60sec], 4 Times 0.8000 < Initial Absorbance Delta Absorbance per Minute < 0.1700 Use Reflex Test. Dilution Ration = 2 Maximum SD of Delta Absorbance per Minute = 0.0300 Adjust absorbance ranges to path length. Document number: APP-A7526-CW2910 Rev. 6-12

Page 5: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: Amylase Catalog # : A7564

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Amylase Assay Type: Chemistry Assay Mode: Factor Factor = 3178 (Adjusted 5795.3) Version: 0 Temperature: 37 ◦Normal: 25 – 125

C

Linear: <= 2000, Use Reflex Test, Dilution Ratio = 2 Units: U/L # of decimals: 0 Assay Description: ******************************************************************************************************************** Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************

Add Sample, Volume [5.0 ul] ASSAY STEPS

Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low Add Reagent [Amylase Reagent], Volume [200.0 ul] Aspiration Speed=2, Air Gap=50ul, Prewarm Time=15sec, Dispense Speed=1, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Kinetic Read, Filters [405/630], Read Interval [60sec], 3 Times Delta Absorbance per Minute < 0.6290 Use Reflex Test. Dilution Ration = 2 Maximum SD of Delta Absorbance per Minute = 0.0300 Adjust absorbance ranges to path length. Document number: APP-A7564-CW2910 Rev. 6-12

Page 6: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: AST (SGOT) Catalog # : A7561

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: AST (SGOT) Assay Type: Chemistry Assay Mode: Factor Factor = -1768 (Adjusted -3004.2) Version: 0 Temperature: 37 ◦Normal: 5 – 34

C

Linear: <= 500, Use Reflex Test, Dilution Ratio = 2 Units: U/L # of decimals: 0 Assay Description: ******************************************************************************************************************** Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************

Add Sample, Volume [20.0 ul] ASSAY STEPS

Aspiration Speed=4, Air Gap=1ul, Dispense Speed=2, Dispense Height=Low Add Reagent [AST working Reagent], Volume [200.0 ul] Aspiration Speed=1, Air Gap=50ul, Prewarm Time=15sec, Dispense Speed=1, Dispense Height=High Incubate 00:01:30 [0h 1min 30sec] Kinetic Read, Filters [340/405], Read Interval [60sec], 4 Times 0.8000 < Initial Absorbance Delta Absorbance per Minute < 0.1700 Use Reflex Test. Dilution Ration = 2 Maximum SD of Delta Absorbance per Minute = 0.0300 Adjust absorbance ranges to path length. Document number: APP-A7561-CW2910 Rev. 6-12

Page 7: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: auto HDL Catalog # : H7545

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: HDL Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 30 – 85

C

Linear: 2 – 150, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 0 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: autoHDL/LDL Calibrator, Curve valid interval: 7 days 0 hours ********************************************************************************************************************

Add Sample, Volume [3.0 ul] ASSAY STEPS

Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low Add Reagent [HDL R1 Reagent], Volume [180.0 ul] Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:05:00 [0h 5min 0sec] Read with Primary Filter [600], Differential Filter [700] Add Reagent [HDL R2 Reagent], Volume [60.0 ul] Aspiration Speed=0, Air Gap=25ul, Prewarm Time=0sec, Dispense Speed=1, Dispense Height=High Incubate 00:05:00 [0h 5min 0sec] Read with Primary Filter [600], Differential Filter [700] Document number: APP-H7545-CW2910 Rev. 6-12

Page 8: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: Direct Bilirubin Catalog # : B7538

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Direct Bilirubin Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 0.0 – 0.5

C

Linear: <= 20.0, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 1 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************

Add Sample, Volume [25.0 ul] ASSAY STEPS

Aspiration Speed=4, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low Add Reagent [Direct Bilirubin R1], Volume [180.0 ul] Aspiration Speed=1, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=1, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [545], Differential Filter [700] Add Reagent [Direct Bilirubin R2], Volume [20.0 ul] Aspiration Speed=1, Air Gap=25ul, Prewarm Time=0sec, Dispense Speed=1, Dispense Height=High Incubate 00:05:00 [0h 5min 0sec] Read with Primary Filter [545], Differential Filter [700] Document number: APP-B7538-CW2910 Rev. 6-12

Page 9: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: Total Bilirubin Catalog # : B7576

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Total Bilirubin Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 0.2 – 1.2

C

Linear: <= 20.0, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 1 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************

Add Reagent [Total Bilirubin working Reagent], Volume [200.0 ul] ASSAY STEPS

Aspiration Speed=1, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=1, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [545], Differential Filter [630] Add Sample, Volume [10.0 ul] Aspiration Speed=2, Air Gap=1ul, Dispense Speed=1, Dispense Height=High Incubate 00:05:00 [0h 5min 0sec] Read with Primary Filter [545], Differential Filter [630] Document number: APP-B7576-CW2910 Rev. 6-12

Page 10: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: BUN Catalog # : B7552

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: BUN Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 7 – 18

C

Linear: <= 150, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 0 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************

Add Sample, Volume [2.0 ul] ASSAY STEPS

Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low Add Reagent [BUN working Reagent], Volume [200.0 ul] Aspiration Speed=1, Air Gap=50ul, Prewarm Time=15sec, Dispense Speed=1, Dispense Height=High Incubate 00:01:15 [0h 1min 15sec] Kinetic Read, Filters [340/630], Fixed Time [2min 30sec] Document number: APP-B7552-CW2910 Rev. 6-12

Page 11: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: Calcium AZ II Catalog # : C7529

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Ca AZ II Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 8.5 – 10.4

C

Linear: <= 15.0, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 1 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours ********************************************************************************************************************

Add Reagent [Calcium AZ II Reagent], Volume [200.0 ul] ASSAY STEPS

Aspiration Speed=1, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [600], Differential Filter [700] Add Sample, Volume [2.0 ul] Aspiration Speed=1, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low Incubate 00:03:00 [0h 3min 0sec] Read with Primary Filter [600], Differential Filter [700] Document number: APP-C7529-CW2910 Rev. 6-12

Page 12: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: Calcium CPC Catalog # : C7503

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Calcium CPC Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 8.5 – 10.5

C

Linear: <= 20.0, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 1 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************

Add Reagent [Calcium Color Reagent], Volume [100.0 ul] ASSAY STEPS

Aspiration Speed=2, Air Gap=25ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Add Reagent [Calcium Buffer Reagent], Volume [100.0 ul] Aspiration Speed=2, Air Gap=25ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [545], Differential Filter [700] Add Sample, Volume [4.0 ul] Aspiration Speed=2, Air Gap=1ul, Dispense Speed=1, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [545], Differential Filter [700] Document number: APP-C7503-CW2910 Rev. 6-12

Page 13: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: CK Catalog # : C7522

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: CK Assay Type: Chemistry Assay Mode: Factor Factor = 6592 (Adjusted 12021.0) Version: 0 Temperature: 37 ◦Normal: <= 160

C

Linear: <= 1500, Use Reflex Test, Dilution Ratio = 2 Units: U/L # of decimals: 0 Assay Description: ******************************************************************************************************************** Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************

Add Sample, Volume [5.0 ul] ASSAY STEPS

Aspiration Speed=4, Air Gap=1ul, Dispense Speed=2, Dispense Height=Low Add Reagent [CK working Reagent], Volume [200.0 ul] Aspiration Speed=1, Air Gap=50ul, Prewarm Time=15sec, Dispense Speed=1, Dispense Height=High Incubate 00:02:00 [0h 2 min 0 sec] Kinetic Read, Filters [340/405], Read Interval [60sec], 2 Times Document number: APP-C7522-CW2910 Rev. 6-12

Page 14: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: Chloride Catalog # : C7501

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Chloride Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 98 – 106

C

Linear: 80 – 120, Use Reflex Test, Dilution Ratio = 2 Units: mEq/L # of decimals: 0 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************

Add Reagent [Chloride Reagent], Volume [200.0 ul] ASSAY STEPS

Aspiration Speed=1, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=1, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [505], Differential Filter [630] Add Sample, Volume [2.0 ul] Aspiration Speed=2, Air Gap=2ul, Dispense Speed=1, Dispense Height=High Incubate 00:05:00 [0h 5min 0sec] Read with Primary Filter [505], Differential Filter [630] Document number: APP-C7501-CW2910 Rev. 6-12

Page 15: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: CO2 Catalog # : C7502

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: CO2 Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 23 – 34

C

Linear: 2 - 40, Use Reflex Test, Dilution Ratio = 2 Units: mEq/L # of decimals: 0 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours ********************************************************************************************************************

ASSAY STEPS

Add Sample, Volume [2.0 ul] Aspiration Speed=1, Air Gap=2ul, Dispense Speed=1, Dispense Height=Low Add Reagent [CO2 Reagent], Volume [200.0 ul] Aspiration Speed=1, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:30 [0h 1min 30sec] Kinetic Read, Filters [405/505], Fixed Time [5min 0sec] Document number: APP-C7502-CW2910 Rev. 6-12

Page 16: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: Cholesterol Catalog # : C7510

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Cholesterol Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: < 200

C

Linear: <=700, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 0 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours ********************************************************************************************************************

Add Reagent [Cholesterol Reagent], Volume [240.0 ul] ASSAY STEPS

Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [505], Differential Filter [630] Add Sample, Volume [2.4 ul] Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low Incubate 00:10:00 [0h 10min 0sec] Read with Primary Filter [505], Differential Filter [630] Document number: APP-C7510-CW2910 Rev. 6-12

Page 17: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: Creatinine Catalog # : C7539

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Creatinine Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 0.4 – 1.4

C

Linear: <= 25.0, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 1 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours ********************************************************************************************************************

ASSAY STEPS

Add Sample, Volume [10.0 ul] Aspiration Speed=4, Air Gap=1ul, Dispense Speed=2, Dispense Height=Low Add Reagent [Creatinine working Reagent], Volume [200.0 ul] Aspiration Speed=1, Air Gap=50ul, Prewarm Time=15sec, Dispense Speed=1, Dispense Height=High Incubate 00:00:50 [0h 0min 50sec] Kinetic Read, Filters [505/630], Fixed Time [2min 30sec] Document number: APP-C7539-CW2910 Rev. 6-12

Page 18: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: Phosphorus Catalog # : P7516

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Phosphorus Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 2.5 – 4.8

C

Linear: <= 12.0, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 1 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours ********************************************************************************************************************

Add Reagent [Phosphorus Reagent], Volume [200.0 ul] ASSAY STEPS

Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [340], Differential Filter [630] Add Sample, Volume [4.0 ul] Aspiration Speed=2, Air Gap=1ul, Dispense Speed=1, Dispense Height=Low Incubate 00:05:00 [0h 5min 0sec] Read with Primary Filter [340], Differential Filter [405] Document number: APP-P7516-CW2910 Rev. 6-12

Page 19: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: Gamma GT Catalog # : G7571

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: GGT Assay Type: Chemistry Assay Mode: Factor Factor = 2211 (Adjusted 3935.9) Version: 0 Temperature: 37 ◦Normal: 9 – 54

C

Linear: <= 800, Use Reflex Test, Dilution Ratio = 2 Units: U/L # of decimals: 0 Assay Description: ******************************************************************************************************************** Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************

Add Sample, Volume [10.0 ul] ASSAY STEPS

Aspiration Speed=4, Air Gap=1ul, Dispense Speed=2, Dispense Height=Low Add Reagent [GGT working Reagent], Volume [200.0 ul] Aspiration Speed=1, Air Gap=50ul, Prewarm Time=15sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:30 [0h 1min 30sec] Kinetic Read, Filters [405/630], Read Interval [60sec], 3 Times Initial Absorbance < 0.8500 Delta Absorbance per Minute < 0.4520 Use Reflex Test. Dilution Ration = 2 Maximum SD of Delta Absorbance per Minute = 0.0300 Adjust absorbance ranges to path length. Document number: APP-G7571-CW2910 Rev. 6-12

Page 20: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: Glucose (Hex) Catalog # : G7517

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Glucose (Hex) Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 70 – 110

C

Linear: <= 500, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 0 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours ********************************************************************************************************************

Add Reagent [Glucose (Hex) Reagent], Volume [240.0 ul] ASSAY STEPS

Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [340], Differential Filter [405] Add Sample, Volume [2.4 ul] Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low Incubate 00:05:00 [0h 5min 0sec] Read with Primary Filter [340], Differential Filter [405] Document number: APP-G7517-CW2910 Rev. 6-12

Page 21: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: Glucose (Ox) Catalog # : G7521

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Glucose (Ox) Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 70 – 105

C

Linear: <= 500, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 0 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours ********************************************************************************************************************

Add Reagent [Glucose (Ox) Reagent], Volume [240.0 ul] ASSAY STEPS

Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [505], Differential Filter [630] Add Sample, Volume [2.4 ul] Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low Incubate 00:10:00 [0h 10min 0sec] Read with Primary Filter [505], Differential Filter [630] Document number: APP-G7521-CW2910 Rev. 6-12

Page 22: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: HbA1c Catalog # : H7541

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: HbA1c Assay Type: Chemistry Assay Mode: Cubic Spline Version: 1 Temperature: 37 ◦Normal: < 6%

C

Linear: 2 – 16 Units: % # of decimals: 1 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 14 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: autoHDL/LDL Calibrator, Curve valid interval: 7 days 0 hours ********************************************************************************************************************

Add Sample, Volume [5.0 ul] ASSAY STEPS

Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low Add Reagent [HBa1c R1 Reagent], Volume [150.0 ul] Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:05:00 [0h 5min 0sec] Add Reagent [HbA1c R2 Reagent], Volume [50.0 ul] Aspiration Speed=0, Air Gap=25ul, Prewarm Time=0sec, Dispense Speed=1, Dispense Height=High Incubate 00:05:00 [0h 5min 0sec] Read with Primary Filter [630], Differential Filter [None] Document number: APP-H7541-CW2910 Rev. 7-12

Page 23: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: Iron, Total Catalog # : I7505

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Iron, Total Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 60 – 150

C

Linear: <= 500, Use Reflex Test, Dilution Ratio = 2 Units: ug/dl # of decimals: 0 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours ********************************************************************************************************************

Add Sample, Volume [30.0 ul] ASSAY STEPS

Aspiration Speed=2, Air Gap=2ul, Dispense Speed=1, Dispense Height=Low Add Reagent [Iron Buffer Reagent], Volume [150.0 ul] Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:05:00 [0h 5min 0sec] Read with Primary Filter [545], Differential Filter [700] Add Reagent [Iron Color Reagent], Volume [3.0 ul] Aspiration Speed=2, Air Gap=2ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:10:00 [0h 10min 0sec] Read with Primary Filter [545], Differential Filter [700] Document number: APP-I7505-CW2910 Rev. 6-12

Page 24: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: LDH Catalog # : L7572

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: LDH Assay Type: Chemistry Assay Mode: Factor Factor = 3376 (Adjusted 6010.0) Version: 0 Temperature: 37 ◦Normal: 80 – 285

C

Linear: <= 1000, Use Reflex Test, Dilution Ratio = 2 Units: U/L # of decimals: 0 Assay Description: ******************************************************************************************************************** Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************

Add Sample, Volume [5.0 ul] ASSAY STEPS

Aspiration Speed=4, Air Gap=1ul, Dispense Speed=2, Dispense Height=Low Add Reagent [LDH working Reagent], Volume [200.0 ul] Aspiration Speed=1, Air Gap=50ul, Prewarm Time=15sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:30 [0h 1min 30sec] Kinetic Read, Filters [340/405], Read Interval [60sec], 4 Times 0.8000 < Initial Absorbance Delta Absorbance per Minute < 0.1210 Use Reflex Test. Dilution Ration = 2 Maximum SD of Delta Absorbance per Minute = 0.0300 Adjust absorbance ranges to path length. Document number: APP-L7572-CW2910 Rev. 6-12

Page 25: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: Lipase Catalog # : L7503

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Lipase Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 0 – 62

C

Linear: <= 600, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 0 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Lipase Standard, Curve valid interval: 7 days 0 hours ********************************************************************************************************************

Add Sample, Volume [2.5 ul] ASSAY STEPS

Aspiration Speed=4, Air Gap=1ul, Dispense Speed=2, Dispense Height=Low Add Reagent [Lipase Substrate (R1)], Volume [150.0 ul] Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:04:00 [0h 4min 0sec] Add Reagent [Lipase Activator (R2)], Volume [50.0 ul] Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:02:00 [0h 2min 0sec] Kinetic Read, Filters [545/700], Fixed Time [1min 0sec] Document number: APP-L7503-CW2910 Rev. 6-12

Page 26: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: Magnesium Catalog # : HM929

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Magnesium Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 1.5 – 2.6

C

Linear: <= 4.86, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 2 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours ********************************************************************************************************************

Add Reagent [Magnesium R1 Reagent], Volume [150.0 ul] ASSAY STEPS

Aspiration Speed=2, Air Gap=25ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Add Reagent [Magnesium R2 Reagent], Volume [150.0 ul] Aspiration Speed=2, Air Gap=25ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [545], Differential Filter [700] Add Sample, Volume [2.0 ul] Aspiration Speed=2, Air Gap=1ul, Dispense Speed=1, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [545], Differential Filter [700] Document number: APP-HM929-CW2910 Rev. 6-12

Page 27: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: Microprotein Catalog # : P7582

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Microprotein Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 28 – 141

C

Linear: 2 – 250, Use Reflex Test, Dilution Ratio = 2 Units: mg/24 hrs # of decimals: 0 Assay Description: for CSF and Urine ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Microprotein Standard, Curve valid interval: 7 days 0 hours ********************************************************************************************************************

Add Reagent [Microprotein Reagent], Volume [200.0 ul] ASSAY STEPS

Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [600], Differential Filter [700] Add Sample, Volume [4.0 ul] Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low Incubate 00:05:00 [0h 5min 0sec] Read with Primary Filter [600], Differential Filter [700] Document number: APP-P7582-CW2910 Rev. 6-12

Page 28: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: Total Protein Catalog # : T7528

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Total Protein Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 6.2 – 8.5

C

Linear: 1.0 – 15.0, Use Reflex Test, Dilution Ratio = 2 Units: g/dl # of decimals: 1 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours ********************************************************************************************************************

Add Reagent [Total Protein Reagent], Volume [200.0 ul] ASSAY STEPS

Aspiration Speed=2, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=2, Dispense Height=High Incubate 00:01:00 [0h 1min 0sec] Read with Primary Filter [545], Differential Filter [700] Add Sample, Volume [4.0 ul] Aspiration Speed=2, Air Gap=2ul, Dispense Speed=2, Dispense Height=Low Incubate 00:05:00 [0h 5min 0sec] Read with Primary Filter [545], Differential Filter [700] Document number: APP-T7528-CW2910 Rev. 6-12

Page 29: pointe chemwell 2910.pdf

Chemwell 2902 / 2910 Instrument Application

Phone: 734-487-8300 • Toll Free: 800-757-5313 • Fax: 734-483-1592 • www.pointescientific.com

Test: Triglycerides Catalog # : T7532

The following reagent parameters are provided by Pointe Scientific, Inc as a reference when Pointe Scientific reagents are used on this instrument. This application has not been validated by Pointe and should only be used in conjunction with your laboratory quality control validation program. Assay Name: Triglycerides Assay Type: Chemistry Assay Mode: Standard Version: 0 Temperature: 37 ◦Normal: 44 – 148

C

Linear: <= 1000, Use Reflex Test, Dilution Ratio = 2 Units: mg/dl # of decimals: 0 Assay Description: ******************************************************************************************************************** Blank used: Valid interval: 7 days 0 hours Controls: Level 1, Warn & Continue if out of range, Valid interval: 1 day 0 hours Level 2, Warn & Continue if out of range, Valid interval: 1 day 0 hours Standards: Chemistry Calibrator, Curve valid interval: 7 days 0 hours ********************************************************************************************************************

Add Reagent [Triglycerides Reagent], Volume [200.0 ul] ASSAY STEPS

Aspiration Speed=1, Air Gap=50ul, Prewarm Time=0sec, Dispense Speed=1, Dispense Height=High Incubate 00:02:00 [0h 2min 0sec] Read with Primary Filter [505], Differential Filter [630] Add Sample, Volume [2.5 ul] Aspiration Speed=2, Air Gap=1ul, Dispense Speed=1, Dispense Height=High Incubate 00:10:00 [0h 10min 0sec] Read with Primary Filter [505], Differential Filter [630] Document number: APP-T7532-CW2910 Rev. 6-12