Citation preview

Page 1: Presentation: Thursday @ 2pm

Real-Time Motion Correction for High-Resolution Imaging of the Larynx: Implementation and Initial

Results

Presentation: Thursday @ 2pm#

5036

Electrical EngineeringStanford University

Joëlle K. Barral Juan M. Santos Dwight G. Nishimura

Page 2: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

2/38

In a Nutshell

We propose a real-time algorithm to combat the main types of motion that corrupt high-resolution larynx imaging.

Our algorithm combines navigator-based motion correction with a reacquisition strategy.

Page 3: Presentation: Thursday @ 2pm

MOTIVATION

Page 4: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

4/38

The Larynx

Thyroid cartilage

Sagittalhttp://www.antiquescientifica.com -- Drawing courtesy of Julie C. DiCarlo

Axial

Thyroid cartilage

Anterior commissure

Cricoid cartilage

Vocal cords

Page 5: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

5/38

Laryngeal Motion

Healthy volunteer

Real-time acquisition: 13 frames per secondNotice swallowing at time t = 18 s!

QuickTime™ and aYUV420 codec decompressor

are needed to see this picture.

Page 6: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

6/38

Laryngeal Motion

Cancer patient

: Outliers (Sporadic motion)

: Bulk motion (Drift)

High-frequencies: Respiration, 14 cycles per min

Motion detected by Cartesian navigators

Page 7: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

7/38

Laryngeal Motion Types How to mitigate their effects

Intermittent, sporadic motion:– Swallowing, coughing, jolting

Alternative ordering schemes

Continuous motion:–Flow (carotid arteries) Phase encodes L/R–Bulk motion (drift) Physical restraints; Coaching; Navigators–Respiration Diminishing Variance Algorithm (DVA)

If a continuous drift happens, DVA never converges.

Page 8: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

8/38

Diminishing Variance Algorithm (DVA)

Sachs, MRM 34: 412-422, 1995 -- Sachs, IEEE-TMI19: 73-79, 2000

Page 9: Presentation: Thursday @ 2pm

METHODS

Page 10: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

10/38

Proposed Approach

We propose to first correct the data based on the shift information. We then reacquire encodes whose projections could not be properly corrected.

Page 11: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

11/38

Implementation

1.5 TRTHawk

Santos, IEEE-EMBS 2: 1048-1051, 2004

Page 12: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

12/38

Pulse Sequence

Fast Large Angle Spin Echo = FLASE– Spin echo: immune against flow & off-

resonances– 3D: high-resolution

– T1-weighted contrast

Ma, MRM 35:903-910, 1996 -- Song, MRM 41:947-953, 1999

QuickTime™ and aYUV420 codec decompressor

are needed to see this picture.

Page 13: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

13/38

Encodes Ordering

Sequential

Square spiral

Pseudo-random

kz

ky

Examples with 32 phase encodes and 16 slice encodes

Elliptical (concentric)

Wilman, MRM 38: 793-802, 1997 -- Bernstein, MRM 50: 802-812, 2003

QuickTime™ and aYUV420 codec decompressor

are needed to see this picture.

QuickTime™ and aYUV420 codec decompressor

are needed to see this picture.

QuickTime™ and aYUV420 codec decompressor

are needed to see this picture.

QuickTime™ and aYUV420 codec decompressor

are needed to see this picture.

Page 14: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

14/38

Reconstruction Pipeline

Barral, ISMRM Motion Workshop 2010, p. 18

The user stops the scan when satisfactory image quality is obtained.

Page 15: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

15/38

GUI

X Y Z

S

S

Page 16: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

16/38

Experimental ParametersFOV 12 cm - Matrix size 256x128x32 - TR/TE = 80/10 ms

Sequential encodes order

Three-coil larynx dedicated array

First pass (full acquisition: 4096 encodes): 5 min 28 s

Each additional pass (64 encodes reacquired): 5 s

Phantom (orange) scans: coronal acquisitions

In vivo (larynx) scans: axial acquisitions

Barral, ISMRM 2009, p. 1318 -- Coil picture courtesy of Marta G. Zanchi

Page 17: Presentation: Thursday @ 2pm

PHANTOM EXPERIMENTS

Page 18: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

18/38

No Motion

An orange was scanned.

Phantom Experiment 1:

Page 19: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

19/38

No MotionOne pass = Full acquisition

As expected, image and corrected image are identical

Phantom Experiment 1:

Page 20: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

20/38

DVA

Non-rigid motion was simulated by switching from the coronal acquisition to an axial acquisition towards the middle of the scan, for several seconds.

Phantom Experiment 2:

Page 21: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

21/38

DVAPass # 1 = Full acquisition: 4096 encodes acquired As expected, motion correction fails

Motion detection successful Shift information meaningless

Phantom Experiment 2:

Page 22: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

22/38

DVAPass # 1

Pass # 6

When corrupted encodes are reacquired, a motion-free image is obtained.

Phantom Experiment 2:

Page 23: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

23/38

Motion Correction

Towards the middle of the scan, the table was manually translated. It was brought back to its original position several seconds later.

Phantom Experiment 3:

Page 24: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

24/38

Motion CorrectionPass # 1 = Full acquisition: 4096 encodes acquired

As expected, motion correction works

Phantom Experiment 3:

Page 25: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

25/38

Motion Correction

Blurry: the final position of the table did not perfectly match the original position.

Phantom Experiment 3:

Pass # 1

Pass # 4

Page 26: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

26/38

Combined Algorithm

Non-rigid motion was simulated by switching to an axial acquisition towards the middle of the scan, for several seconds. The table was then manually translated.

Phantom Experiment 4:

Page 27: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

27/38

Combined Algorithm

Pass # 1 = Full acquisition: 4096 encodes acquired

Motion correction successfully accounts for the translation

Phantom Experiment 4:

Page 28: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

28/38

Combined Algorithm

Pass # 1

Pass # 6

Reacquisition needed to correct for non-rigid motion

Phantom Experiment 4:

Page 29: Presentation: Thursday @ 2pm

IN VIVO EXPERIMENTS

Page 30: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

30/38

Without Instructions

A healthy volunteer was scanned.

In Vivo Experiment 1:

Page 31: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

31/38

Without Instructions One pass = Full acquisition

Slice 20/32

X

Y

In Vivo Experiment 1:

Slice 26/32

Page 32: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

32/38

Without Instructions

Sagittal reformat

In Vivo Experiment 1:

Page 33: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

33/38

With Instructions

A healthy volunteer was scanned. He was asked to swallow at will and to accentuate motion when the center of k-space was being acquired. For this experiment, 192 encodes were reacquired each additional pass.

In Vivo Experiment 2:

Page 34: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

34/38

With Instructions Pass # 1 = Full acquisition: 4096 encodes acquired

X

Y

In Vivo Experiment 2:

Swallowing properly detected

Only bulk motion corrected by motion-correction

Page 35: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

35/38

With Instructions

When corrupted encodes are reacquired, motion correction is needed to account for bulk shift (drift) that happened between passes.

In Vivo Experiment 2:

Pass # 1

Pass # 3

Page 36: Presentation: Thursday @ 2pm

WRAP-UP

Page 37: Presentation: Thursday @ 2pm

# 5036 Real-Time Motion Correction for Larynx Imaging -- J.K. Barral et al.

37/38

Conclusion & Future Work

Our real-time algorithm corrects for rigid-body motion and reacquires encodes that could not be corrected.

Additional scans are needed to validate the robustness of the method in vivo.

Future work will improve the flexibility of the algorithm and improve the user interface.