PulmicortRespule VsBeclomethasone 2005-Vaghi

  • Upload
    mapuxd

  • View
    218

  • Download
    0

Embed Size (px)

Citation preview

  • 8/10/2019 PulmicortRespule VsBeclomethasone 2005-Vaghi

    1/3

    Short communication

    In vitro comparison of nebulised budesonide (Pulmicort Respulesw)and beclomethasone dipropionate (Clenilw per Aerosol)*

    A. Vaghia, E. Bergb,*, S. Liljedahlb, J.O. Svenssonb

    aAzienda Ospedaliera G. Salvini, Garbagnate, Italy

    bAstraZeneca R&D, Scheelvagen 2, Lund S-221 87, Sweden

    Received 3 August 2004; revised 15 October 2004; accepted 23 October 2004

    Abstract

    This study compared the in vitro performance of two inhaled corticosteroid products for nebulisation, Pulmicort Respulesw (budesonide

    0.5 mg/mL) and Clenilw per Aerosol (beclomethasone dipropionate (BDP) 0.4 mg/mL). Each product was used in combination with three

    different nebulisers (2 mL/test, 5 min run time) and the dose to the lungs was determined according to standard methods. The shape of the

    suspended particles in each product was studied using scanning electron microscopy (SEM). Overall, a higher fine particle dose was achieved

    with Pulmicort Respulesw versus Clenilw per Aerosol, with estimated dose to the lungs of 814 and 36% of nominal dose, respectively.

    SEM showed that budesonide particles were small, typically w23 mm in diameter, whereas those of BDP were needle-shaped and up to

    w10 mm long. The more favourable particle shape and size of suspended budesonide may explain the higher fine particle dose with Pulmicort

    Respulesw versus Clenilw per Aerosol.

    q 2004 Elsevier Ltd. All rights reserved.

    1. Introduction

    The use of inhaled corticosteroids such as budesonide

    and beclomethasone dipropionate (BDP) represents the

    cornerstone of asthma management [1], enabling many

    asthmatic patients who are otherwise limited by their

    disease to enjoy a near normal or normal lifestyle. These

    agents are available in different formulations, from metered-

    dose and dry powder inhalers to products for nebulisation, to

    meet the diverse requirements of the heterogeneous

    population of asthmatic patients. However, one potential

    limitation of such product diversity is that differences in

    formulation technology can impact upon drug delivery tothe lungs [2], which may lead to potential differences in

    clinical response. Comparative studies are therefore import-

    ant to determine the performance characteristics of available

    products for inhalation treatment of asthma, as any

    differences may have implications for clinical efficacy.

    The aim of the present study, therefore, was to comparethe in vitro performance of two inhaled corticosteroid

    products for nebulisation, namely Pulmicort Respulesw

    (budesonide 0.5 mg/mL; AstraZeneca) and Clenilw per

    Aerosol (BDP 0.4 mg/mL; Chiesi).

    2. Methods

    Different conventional jet nebuliser devices were used in

    this study to simulate the types of conditions under which

    the study products may be administered in routine practice.

    Three nebulisers were used (experiments performed intriplicate): Cirrusw (relatively small droplet size 23 mm;

    Intersurgical), Pari LC Plusw (medium droplet size 45 mm;

    Pari), and Omron COMP Air Elite NE-C21w (large droplet

    size 68 mm; Medel; in Italy marketed as Clenny). The

    Cirrusw and Pari LC Plusw nebulisers were used with a

    compressor (Pari Masterw; Pari), whereas the Omron

    nebuliser has a built-in compressor. Each nebuliser was

    connected to a breathing simulator (Pari Compasw; Pari);

    experimental conditions were: tidal volume, 500 mL;

    1094-5539/$ - see front matter q 2004 Elsevier Ltd. All rights reserved.

    doi:10.1016/j.pupt.2004.10.004

    Pulmonary Pharmacology & Therapeutics 18 (2005) 151153

    www.elsevier.com/locate/ypupt

    *This study was supported by AstraZeneca R&D, Lund, Sweden.

    * Corresponding author. Tel.:C46 46 336162; fax: C46 46 337168.

    E-mail address: [email protected] (E. Berg).

    http://www.elsevier.com/locate/ypupthttp://www.elsevier.com/locate/ypupt
  • 8/10/2019 PulmicortRespule VsBeclomethasone 2005-Vaghi

    2/3

    frequency, 15 breaths/min; and 1:1 inhalation to exhalation

    ratio (i.e. 2 s inhalation, 2 s exhalation). These settings were

    in accordance with the CEN-standard for nebuliser testing

    [3]. Each nebuliser was charged with 2 mL product and run

    for 5 min, after which the amount of drug on the inhalation/

    exhalation filters and contained within the nebuliser residue

    was quantified using established methods.

    The test was performed as described in the European

    Standard, except that clinical formulations were used

    instead of sodium fluoride [3]. Budesonide and BDP were

    quantified by liquid chromatography (LC) method, using

    internal standard (fluocinolone acetonide), ethanol/water(43:57) mobile phase, a Supelcosil LC-18 column (5 mm

    particles, 5!0.46 cm), and UV detection at 254 nm.

    Particle size distribution of the nebulised suspensions

    was evaluated in tandem experiments using an Andersen

    impactor with the nebuliser outlets centred into the USP-

    throat (23 mm overlap).

    The shape and size of the suspended particles in the two

    products was investigated by scanning electron microscopy

    (SEM). In brief, a container with suspension was shaken,

    opened and transferred to a glass beaker. While stirring,

    0.2 mL of Clenilw or 0.4 mLw of Pulmicort suspension was

    filtered through an isopore polycarbonate membrane filter,

    pore size 0.2 mm. A piece of the filter with residue was cut

    out, fixed to a metal holder, sputtered with gold and studied

    in a SEM, JEOL JSM-5200 scanning microscope, operated

    at 15 kV. Representative micrographs of the drug particles,

    the residue on the filter, were recorded digitally.

    3. Results and discussion

    In this study we sought to compare the in vitro

    performance characteristics of two available corticoster-

    oid-containing products for nebulisation, namely Pulmicort

    Respulesw (budesonide) and Clenilw per Aerosol (BDP).

    Analysis of inhalation filters showed that for both products,

    drug deposition (mean (SD)) was lowest for Cirrusw

    (budesonide, 12 (1)% of nominal dose; BDP, 10 (2)% of

    nominal dose) and highest for the Omron nebuliser

    (budesonide, 29 (2)% of nominal dose; BDP, 37 (1)% of

    nominal dose). The corresponding values for the Pari LCPlusw was 27 (3)% (BDP) and 27 (1)% (budesonide) of

    nominal dose. These findings are consistent with the droplet

    size produced by each nebuliser, from relatively small with

    Cirrusw to large with the Omron nebuliser. For both

    products, the exhaled amount as a percentage of nominal

    dose was comparable for the Cirrusw and Pari LC Plusw

    nebulisers (1113%). Data are not available for the Omron

    nebulizer as it cannot be connected to an exhalation filter.

    As expected, the major proportion of added drug was

    retained in the nebuliser device and was comparable for the

    two products (Cirrusw,w75%; Pari LC Plusw,w55%; and

    Omron,w

    50%). The remainder was accounted for by lossto the environment as a result of nebuliser leakage.

    In vitro findings for particle size distribution for the two

    products evaluated with the three nebuliser devices are

    summarised inFig. 1. Overall, dose quality (% of dose in the

    respirable range i.e. !5 and!3 mm) was highest for

    Cirrusw and lowest for the Omron nebuliser, irrespective

    of administered product. In all cases the fine particle dose

    (!5 mm) was higher for Pulmicort Respulesw (budesonide)

    than for Clenilw per Aerosol (BDP), the difference being

    even more pronounced for the fraction !3 mm(Table 1).

    Combining these results with those for the dose on the

    inhalation filter it was possible to estimate comparative

    Table 1

    Summary of in vitro findings (mean [SD]) for particle size distribution of Pulmicort Respulesw (budesonide) and Clenilw per Aerosol (beclomethasone

    dipropionate (BDP)) following administration by three nebuliser devices

    Nebuliser

    Cirrusw Pari LC Plusw Omron COMP Air Elite NE-C21w

    BDP Budesonide BDP Budesonide BDP Budesonide

    Particles !5 mm (%)a 54 (6.3) 71 (4.4) 23 (3.9) 51 (4.9) 8 (1.0) 26 (1.7)

    Particles !3 mm (%)a 23 (3.7) 47 (4.3) 7 (1.1) 32 (4.9) 2 (0.3) 14 (2.0)

    MMAD (mm) 4.8 (0.5) 3.3 (0.2) 7.5 (0.5) 4.5 (0.6) 10.0 (0.1) 7.4 (0.1)

    MMAD, mass median aerodynamic diameter.a Expressed as a percentage of dose to impactor.

    Fig. 1. In vitro particle size distribution (meanGSD) for Pulmicort

    Respulesw (budesonide) and Clenilw per Aerosol (beclomethasone

    dipropionate (BDP)) expressed as a percentage of the nominal dose when

    administered using three different nebuliser devices.

    A. Vaghi et al. / Pulmonary Pharmacology & Therapeutics 18 (2005) 151153152

  • 8/10/2019 PulmicortRespule VsBeclomethasone 2005-Vaghi

    3/3

    doses to the lungsthis was the amount of drug on the

    inhalation filter in percentage of nominal dose (reported in

    first paragraph under Section 3)!the fraction !5 mm (or

    3 mm) fromTable 1.Across the three nebulisers, Pulmicort

    Respulesw (budesonide) achieved an estimated dose to the

    lungs (!5 mm) of 814% compared with 36% of nominal

    dose for Clenilw

    per Aerosol (BDP). Thus, PulmicortRespulesw will deliver 23 times more drug to the lungs

    than Clenilw per Aerosol. Looking at the dose in droplets !

    3 mm, this difference was 27-fold in favour of Pulmicort

    Respulesw (49 versus 12% for Clenilw per Aerosol).

    Given that there is a clear relationship between particle size

    and lung deposition and, in turn, clinical effect [4], such

    findings indicate that the two products may differ in terms of

    clinical efficacy and comparative studies, using dose-

    response or dose-reduction designs in symptomatic asthma

    patients, are warranted. Such studies should be of highquality and involve a dose-response or dose-titration phase

    rather than attempt to draw conclusions from comparing

    individual doses[5].

    Scanning electron micrograph analysis showed that

    budesonide particles were small, typically about 23 mm

    in diameter, whereas those of BDP were needle-shaped and

    up to about 10 mm long (Fig. 2). This difference in particle

    size is the most probable explanation for the difference in

    particle size distribution and drug disposition between the

    two products, as outlined above.

    In summary, the findings of this study with two

    corticosteroid-containing productions for nebulisation treat-ment of asthma show that there are differences in the fine

    particle dose produced by different nebulisers. However, a

    higher fine particle dose is consistently achieved with

    Pulmicort Respulesw (budesonide) versus Clenilw per

    Aerosol (BDP), contributing to a higher estimated dose to

    the lungs. This difference is most likely due to a more

    favourable particle shape and size of suspended budesonide,

    and further studies are warranted to determine whether this

    may contribute to improved clinical efficacy.

    References

    [1] Global Initiative for Asthma. Workshop report: strategy for asthma

    management and prevention (updated 2002). Scientific information and

    recommendations for asthma programs. NIH publication No. 02-3659.

    Bethesda, MD: National Institutes for Health; 2002.

    [2] Thorsson L, Edsbacker S, Conradson TB. Lung deposition of

    budesonide from Turbuhaler is twice that from a pressurized metered

    dose inhaler (pMDI). Eur Respir J 1994;7:183944.

    [3] European Standard, E N 135441. Respiratory therapy equipment, part

    1: nebulizing systems and their components. Brussels, Belgium:

    European Committee for Standardization; 2001.

    [4] Pauwels R, Newman S, Borgstrom L. Airway deposition and airway

    effects of antiasthma drugs delivered from metered-dose inhalers. Eur

    Respir J 1997;10:212738.

    [5] Beasley R, Sterk PJ, Kerstjens HAM, Decramer M. Comparative

    studies of inhaled corticosteroids in asthma. Eur Respir J 2001;17:

    57980.

    Fig. 2. Scanning electron micrographs of suspended budesonide and

    beclomethasone dipropionate particles in the Pulmicort Respulesw and

    Clenilw per Aerosol products (micrographs taken August 2003).

    A. Vaghi et al. / Pulmonary Pharmacology & Therapeutics 18 (2005) 151153 153