Reagecon Diagnostics Ltd.
Shannon Free Zone, Shannon, Clare, V14 X073
Calibration Laboratory
Registration number: 265C
is accredited by the Irish National Accreditation Board (INAB) to undertake calibration as detailed in the scope bearing the registration number detailed
above, in conformity with ISO/IEC 17025:2017“General requirements for the competence of testing and calibration laboratories”
(This certificate must be read in conjunction with the publicly available scope of accreditation)
Date of award of accreditation: 12/07/2010Date of last renewal of accreditation: 14/07/2016
Expiry date of this certificate of accreditation:14/07/2021
This accreditation shall remain in force until further notice subject to continuing conformity with the above standard, applicable EA/ILAC requirements and any further requirements
specified by the Irish National Accreditation Board.
INAB Registration No. 265C Issue 2, 24/02/2020
Manager: Chairperson: : Dr Adrienne Duff Ms Ita Kinahan
Organisations are subject to annual surveillance and are re-assessed every five years. The renewal date on this certificate confirms the latest date of renewal of accreditation. To confirm the validity of this certificate, please
contact the Irish National Accreditation Board.INAB is a signatory of the European co-operation for Accreditation (EA) Multilateral Agreement (MLA) and the International Laboratory Accreditation Cooperation (ILAC) Mutual Recognition Arrangement for Calibration.
The Metropolitan Building, James Joyce Street, Dublin 1, Ireland Tel: 1890 289 389. Int Tel: +353 1 614 7000. Email: [email protected] Web: www.inab.ie
Edition 3, 24/02/2020
Page 1 of 6
INAB Registration No. 265C
Schedule of Accreditation
Organisation Name
Reagecon Diagnostics Ltd.
Trading As
INAB Reg No
265C
Contact Name
Dean Meagher
Address
Shannon Free Zone, Shannon, Clare, V14 X073
Contact Phone No
Website
http://www.reagecon.com
Accreditation Standard
ISO 17025 C
Date Initially Awarded
12/07/2010
Scope Classification Metrology
Services available to the public¹
Yes
¹ Refer to document on interpreting INAB Scopes of Accreditation
Sites from which accredited services are delivered (the detail of the accredited services delivered at each site are on the Scope of Accreditation)
Name Address
1 Head Office Shannon Free Zone, Shannon, Clare, V14 X073
Edition 3, 24/02/2020
Page 2 of 6
INAB Registration No. 265C
Scope of Accreditation
Head Office
Metrology
Category: A
Metrology field - Calibrated Device Type
Measured quantity Calibration range Calibration and measurement capability (CMC)
Std. ref/SOP Products Remarks
104 Volume - .02 Special laboratory volumetric apparatus
N/A Single Channel: 5 to 10 μL 10 to 20 μL 20 to 100 μL 100 to 200 μL 200 to 500 μL 500 to 1000 μL 1000 to 2,000 μL 2,000 to 5,000 μL 5,000 to 10,000 μL
Single Channel: ±0.15 μL ±0.15 μL ±0.45 μL ±0.70 μL ±2.0 μL ±4.0 μL ±10 μL ±30 μL ±50 μL
CAL-TS 5,8,9,11,13, 14,15,20,22,23,31, 37 (A)(B), 39,40,41, 42,46,80,85,104
For water delivered from piston and/or plunger operated volumetric apparatus
107 Temperature measuring equipment - .09 Digital temperature indicator systems
- 45 °C to +140 °C 140 °C to 400 °C 400 °C to 650 °C
± 0.050 °C ± 0.13 °C ± 0.32 °C
CAL-TS 5,49,50,52, 53,62,63,64, 66,67,68,69 ,72,74,85,101, 104,112
113 Time - .99 Other Time interval 0 minutes to 24 hours
±0.60 seconds CAL-SW
* Notes: 1. In accordance with INAB policy, uncertainties are calculated for an estimated confidence level of not less than 95%. 2. Calibration and measurement capability expressed as an uncertainty (±) to be
Edition 3, 24/02/2020
Page 3 of 6
INAB Registration No. 265C
reported in compliance with EA-4/02, “Expression of the Uncertainty of Measurement in Calibration”. Calibration Measurement Capability (CMC) is expressed in terms of the following parameters: - Measurand or reference material - Calibration or measurement method or procedure and type of instrument or material calibrated/measured - Measurement range and additional parameters where applicable - Measurement uncertainty
Edition 3, 24/02/2020
Page 4 of 6
INAB Registration No. 265C
Head Office
Metrology
Category: B
Metrology field - Calibrated Device Type
Measured quantity Calibration range Calibration and measurement capability (CMC)
Std. ref/SOP Products Remarks
101 Mass - .01 Precision laboratory balances
N/A 1 mg to 5 g 5 g to 20 g 20 g to 100 g
±0.031 mg ±0.048 mg ±0.083 mg
Documented in-house methods: CAL-TS 8,9,11,14,15, 22,26,31, 32,37 (a) (b), 38,41, 85, 104, 108 Using OIML Class E2 from 1mg to 600g.
101 Mass - .02 Industrial balances
100 g to 200 g 200 g to 600 g
±0.16 mg ±0.26 mg
Documented in-house methods: CAL-TS 8,9,11,14,15, 22,26,31, 32,37 (a) (b), 38,41, 85, 104, 108 Using OIML Class E2 from 1mg to 600g.
101 Mass - .03 Industrial weighing appliances
100 g to 1 kg 1 kg to 2 kg 2 kg to 5 kg 5 kg to 10 kg 1 kg to 10 kg 10 kg to 20 kg 20 kg to 50 kg 50 kg to 100 kg 100 kg to 160 kg
±1.6 mg ±3.1 mg ±7.7 mg ±11 mg ±0.16 g ±0.31 g ±0.47 g ±0.69 g ±0.80 g
Documented in-house methods: CAL-TS 8,9,11,14,15,22,26, 31, 32,37 (a) (b), 38,41, 85, 104, 108. Using Class F1 from 100g to 5kg Using Class M1: 5 x 20kg and 6 x 10kg
Edition 3, 24/02/2020
Page 5 of 6
INAB Registration No. 265C
104 Volume - .02 Special laboratory volumetric apparatus
Single Channel: 5 to 10 μL 10 to 20 μL 20 to 100 μL 100 to 200 μL 200 to 300 μL 300 to 1,000 μL 1,000 to 2,000 μL 2,000 to 5,000 μL 5,000 to 10,000 μL
Single Channel: ±0.15 μL ±0.45 μL ±0.60 μL ±0.70 μL ±2.0 μL ±4.0 μL ±10 μL ±30 μL ±50 μL
CAL-TS 5,8,9,11,13,14, 15,20,22,23,31, 37 (A)(B), 39,40, 41,42,43, 46,80, 85,86,104
For water delivered from piston and/or plunger operated volumetric apparatus
108 Temperature controlled enclosures - .01 Ovens, furnaces, baths
-45 °C to + 140 °C ±0.17 °C CAL-TS 49,50,52,54,56, 57,58,59,60,61, 62,65,66,67,68, 69,72,74,75,76, 77,85,100,104
108 Temperature controlled enclosures - .02 Incubators
-45 °C to +140 °C ±0.17 °C
CAL-TS 49,50,52,54,56, 57,58,59,60,61, 62,65,66,67,68, 69,72,74,75,76, 77,85,100,104
108 Temperature controlled enclosures - .03 Autoclaves and sterilising ovens
80 °C to 140 °C ±0.32 °C CAL-TS 49,52,62,65, 66,67,68,69,72, 74,85,87a,87b, 101,102,104 CAL-SW
108 Temperature controlled enclosures - .04 Industrial freezers
-90 °C to -45 °C -45 °C to + 140 °C
±0.40 °C ±0.17 °C
CAL-TS 49,50,52,54,56, 57,58,59,60,61, 62,65,66,67,68, 69,72,74,75,76, 77,85,100,104
* Notes: 1. In accordance with INAB policy, uncertainties are calculated for an estimated confidence level of not less than 95%. 2. Calibration and measurement capability expressed as an uncertainty (±) to be reported in compliance with EA-4/02, “Expression of the Uncertainty of Measurement in Calibration”.
Edition 3, 24/02/2020
Page 6 of 6
INAB Registration No. 265C
Calibration Measurement Capability (CMC) is expressed in terms of the following parameters: - Measurand or reference material - Calibration or measurement method or procedure and type of instrument or material calibrated/measured - Measurement range and additional parameters where applicable - Measurement uncertainty