Upload
syokri
View
219
Download
0
Embed Size (px)
Citation preview
8/10/2019 Sample Prep note
1/89
8/10/2019 Sample Prep note
2/89
2
Course Learning Outcomes
Students should be able to:
1. Discuss the importance of sample preparation prior
to sample analysis.
2. Explain the principles & procedures of separation
involve in the extraction method.
3. Describe the advantages & limitations of each
extraction method as compared to other extractionmethod.
4. Choose the right & suitable extraction method forsample analysis.
8/10/2019 Sample Prep note
3/89
STEPS IN MEASUREMENT
1. SAMPLING
Process of collecting a representativesample for analysis.
2. SAMPLE PREPARATION Extraction
Clean-up & preconcentration
3. SAMPLE ANALYSIS
Chromatographic / spectroscopic methods
3
8/10/2019 Sample Prep note
4/89
Despite the great technological advances in
the analytical field, most sophisticated
instruments cannot handle complex samplematrices directly &, as a result, a sample
preparation step is commonly involved in
an analytical procedure.
SAMPLE PREPARATION
4
8/10/2019 Sample Prep note
5/89
5
Also known as sample treatment/samplepretreatment.
Mean : Is a series of step required to bring thesample into the correct size & form that issuitable for analysis.
Often the bottleneck of analysis.
Time consuming (~ 61 %).
Laborious step.
Error prone.
SAMPLE PREPARATION
8/10/2019 Sample Prep note
6/89
Efficient sample preparation is therefore
important for a successful analysis.
Generally,
a clean sample helps to improve theseparation & detection,
while poorly treated sample may invalidatethe whole assay.
6
8/10/2019 Sample Prep note
7/89
To isolate & concentrate the analytes of interest
from interfering sample components.
To convert the analytes to a form that is
compatible with the instrument for the final
analysis.
To provide the analytes of interest at a
concentration appropriate for detection or
measurement.
Objective
7
8/10/2019 Sample Prep note
8/89
Extraction methods
Liquid samples1. Liquid-liquid extraction (LLE)
Liquid, gaseous samples
1. Solid phase extraction (SPE)
2. Solid phase microextraction (SPME)
8
8/10/2019 Sample Prep note
9/89
Solid, semi solid samples1. Soxhlet
2. Sonication extraction
3. Microwave assisted extraction (MAE)
4. Pressurized liquid extraction (PLE)
5. Supercritical fluid extraction (SFE)
Extraction methods
9
8/10/2019 Sample Prep note
10/8910
Liquid-Liquid Extraction (LLE)
A method to separatecompounds based on the their
relative solubility in 2 different
immiscible liquids, usuallywater & organic solvent.
An extraction of a substance
from one liquid phase into
another liquid phase.
8/10/2019 Sample Prep note
11/89
Commonly performed after a chemical
reaction as part of the work-up.
A distribution ratio (D) is often quoted as a
measure of how well-extracted a species is.
Concentration of a solute in the organic phase
Concentration in the aqueous phase
Distribution ratio (D) =
11
8/10/2019 Sample Prep note
12/89
Disadvantages :
Operate manually.
Large volume of organic solvents.
Formation of emulsion.
Time consuming.
Different polarity of samples result in lowrecoveries.
12
8/10/2019 Sample Prep note
13/8913
Solid Phase Extraction (SPE) An extraction method that uses a solid phase & a
liquid phase to isolate analyte from a solution.
Commercially available sorbents with differentphases, including non polar (C8 & C18), polar
(cyano & amino) & ion exchange sorbents.
Different interaction (Van der Waals, H bonding,
cation or anion exchange).
8/10/2019 Sample Prep note
14/89
A selection of SPE cartridges, available in
many sizes, shapes & types of stationaryphase.
14
http://en.wikipedia.org/wiki/File:SPE_Cartridges.jpg8/10/2019 Sample Prep note
15/89
15
Column
Cartridge
Disk
http://en.wikipedia.org/wiki/File:SPE_Manifold.jpg8/10/2019 Sample Prep note
16/89
How to use SPE ???
5 step processes to follow
16
8/10/2019 Sample Prep note
17/89
An SPE disk is recommended
for large volume samples,
samples containing high
amounts of particulates, orwhen a high flow rate is
required during sampling.
17
8/10/2019 Sample Prep note
18/89
Conditioned with a water-miscible organic solvent
such as methanol, followed
by water or an aqueousbuffer.
Methanol wets the surface
of the sorbent & penetrates
bonded alkyl phases,allowing water to wet the
silica surface efficiently.
18
8/10/2019 Sample Prep note
19/89
Accurately transfer the
sample to the tube or
reservoir, using avolumetric pipette or
micropipette.
The sample must be in a
form that is compatible withSPE.
19
8/10/2019 Sample Prep note
20/89
If compounds of interest are
retained on the packing, wash offunwanted, unretained materials
using the same solution in which
the sample was dissolved, oranother solution that will not
remove the desired compounds.
To remove unwanted, weakly
retained materials, wash the
packing with solutions that arestronger than the sample matrix,
but weaker than needed to remove
compounds of interest.
20
8/10/2019 Sample Prep note
21/89
Rinse the packing with a smallvolume (typically 200L to 2mL
depending on the tube size, or
5- 10mL depending on the disksize) of a solution that
removes compounds of
interest, but leaves behind anyimpurities not removed in the
wash step.
Collect the eluate.
21
8/10/2019 Sample Prep note
22/89
22
Relying on mainly non polar, hydrophobicinteractions, only non polar or very weakly polar
compounds will absorb to the surface.
When silica particles are bonded with a
hydrophobic phase, they become waterproof &
must be conditioned in order to interact with the
aqueous samples.
This is accomplished by passing methanol or a
similar solvent through the sorbent bed.
8/10/2019 Sample Prep note
23/89
23
This penetrates into the bonded layer & permitswater molecules & analytes to diffuse into the
bonded phase.
After conditioning, water is passed to remove the
excess solvent prior to adding the sample.
Following conditioning, the analyte & other
sample constituents are adsorbed on the sorbent
bed.
8/10/2019 Sample Prep note
24/89
24
A rinsing step removes some of the undesiredconstituents, while elution removes the desired
analytes, perhaps leaving other cosituent
behind, depending on the relative strength ofinteraction with the solid phase or the solubility
in the eluting solvent.
8/10/2019 Sample Prep note
25/89
Applications :
SPE is used most often to prepare liquid
samples & extract semivolatile or nonvolatile
analytes.
Also can be used with solids that are pre-
extracted into solvents.
25
8/10/2019 Sample Prep note
26/89
Advantages :
Simplicity.
High analyte recovery.
Extraction reproducibility. Lower solvent consumption.
Ability to automate the extraction process.
26
8/10/2019 Sample Prep note
27/89
Soxhlet Extraction
Typically, a Soxhlet
extraction is only required
where the desired
compound has a limitedsolubility in a solvent, &
the impurity is insoluble in
that solvent.
27
http://en.wikipedia.org/wiki/File:Soxhlet_Extractor.jpg8/10/2019 Sample Prep note
28/89
1: Stirrer bar2: Round bottom flask (the flask
should not be overfilled & the
volume of solvent should be 3 to 4times the volume of the soxhlet
chamber)
3: Distillation path4: Thimble 5: Solid
6: Siphon top 7: Siphon exit
8: Expansion adapter
9: Condenser
10: Cooling water in 11: Coolingwater out
A schematic representation of a Soxhlet extractor
28
http://en.wikipedia.org/wiki/File:Soxhlet_extractor.svg8/10/2019 Sample Prep note
29/89
The solid sample to be extracted is placed in athimble.
The solvent is added to the round bottom flask.
The solvent is heated to reflux.
Step in soxhlet extraction
29
8/10/2019 Sample Prep note
30/89
The solvent vapour travels up a distillation arm &floods into the chamber where it comes into
contact with the sample.
The condenser ensures that any solvent vapour
cools & drips back down into the chamber housingthe solid material.
Step in soxhlet extraction
30
8/10/2019 Sample Prep note
31/89
The chamber containing the solid material slowly
fills with warm solvent.
Some of the desired compound will then dissolve
in the warm solvent.
When the Soxhlet chamber is almost full, the
chamber is automatically emptied by a siphon
side arm, with the solvent running back down to
the distillation flask.
This cycle may be allowed to repeat many times,
over hours (12-18 hrs).31
8/10/2019 Sample Prep note
32/89
During each cycle, a portion of the non-volatile
compound dissolves in the solvent.
After many cycles the desired compound is
concentrated in the distillation flask.
The advantage of this system is that instead of
many portions of warm solvent being passed
through the sample, just one batch of solvent is
recycled.
32
8/10/2019 Sample Prep note
33/89
After extraction the solvent is removed, typically
by means of a rotary evaporator, yielding theextracted compound.
The non-soluble portion of the extracted solid
remains in the thimble & is usually discarded.
33
8/10/2019 Sample Prep note
34/89
Type of solvent used : less polar solvent(e.g. hexane, acetone).
Advantage of Soxhlet extraction :
Cheap apparatus.
34
8/10/2019 Sample Prep note
35/89
Disadvantages of Soxhlet extraction :
1. Large solvent consumption (50-200 mL for a
10g sample).
2. Long extraction time (16-24 hours).
3. Dirty extracts require tedious clean up
steps.
4. Not suitable for thermally labile compounds.
35
8/10/2019 Sample Prep note
36/89
Vibrational disruption of samples.
Sonication Extraction
36
8/10/2019 Sample Prep note
37/89
Advantages : Cold extraction - good for thermally labile
compounds.
Quick (a few hours).
A sequence of solvents may be used.
Disadvantages :
Physically violent process -may cause
breakdown of macromolecules & clays.
Labor intensive.
37
8/10/2019 Sample Prep note
38/89
Microwave radiation is utilized to heat up theextraction solvent & therefore also the sample
through ionic conduction & dipole rotation.
The solvent must be able to absorb the
microwave radiation & pass it to the samplemolecules in a form of heat.
Microwave Assisted Extraction (MAE)
38
8/10/2019 Sample Prep note
39/89
Suitable solvents : polar solvent with high dipole
moment (water, MeOH, acetone).
Step in MAE :
The solid samples & solvent is placed in a
vessel & heated using microwave radiation in
a closed system.
39
8/10/2019 Sample Prep note
40/89
Extraction perform in a closed vessel :
1. Temp. can be higher than the atmospheric b.p
of the solvent, thus extraction is faster due tothe increased diffusion.
2. Vessels are irradiated simultaneously (henceextraction) up to 12 vessels.
3. Volume of solvent : 20 50 mL.
4. Extraction time : 20 40 min.
40
8/10/2019 Sample Prep note
41/89
Microwave reaction system
41
8/10/2019 Sample Prep note
42/89
Advantages :
Wide applicability for fast extractions ofanalytes including some thermal instable
substances.
Enhance solvent extraction method.
Simultaneous & automatic extraction (up to
12 vessels).
Disadvantage :X Sample vessel has to been cooled before
the extract can be obtained.
42
8/10/2019 Sample Prep note
43/89
Pressurized Liquid Extraction (PLE)
PLE also known as pressurized solvent extraction
(PSE), accelerated solvent extraction (ASE),pressurized fluid extraction (PFE).
Utilized solvent at high temperature & pressure.
Higher pressure causes the solvent to remain liquefiedabove boiling point & allows solvent to penetrate the
sample matrix.
43
8/10/2019 Sample Prep note
44/89
High temperatures can disrupt the strong solute-
matrix interactions & decrease the viscosity of
organic solvents, allowing improved penetration ofthe matrix & thus improved extraction.
The solubility & diffusity of the analytes are
increased by the increased temperature, thus
making the extraction faster & more efficient.
44
8/10/2019 Sample Prep note
45/89
45
8/10/2019 Sample Prep note
46/89
5 steps in PLE :
1. Load sample into the extraction cell.
2. Fill the cell with solvent.
3. Increase temperature & pressure (for static
extraction, maintain temperature &pressure for a specific time).
4. Pump solvent into the cell.
5. Purge solvent from cell using N2.
46
8/10/2019 Sample Prep note
47/89
PLE can be performed :
1. Dynamic mode
The solvent continuously flows throughthe sample in the extraction vessel.
2. Static mode
The supercritical fluid circulates in the
extraction vessel for some period of timebefore being released to the trapping
vessel.
47
8/10/2019 Sample Prep note
48/89
Advantages :
Fast analysis (approximately 15 min/sample).
The high temperature & pressure increase the
capability of solvent to penetrate the sample
matrix.
The solubility & diffusity of the analytes are
increased by the increased temperature, thusmaking the extraction faster & less solvent is
required.
48
8/10/2019 Sample Prep note
49/89
Low solvent consumption (30-50 mL).
Use of less hazardous/environmental friendly
solvent such as acetone & hexane in the placeof methylene chloride.
Solvent systems similar to Soxhlet extraction,
so easy to optimize.
Automated (>20 samples can be processed as
a batch).
49
8/10/2019 Sample Prep note
50/89
Disadvantages :
May not be appropriate for thermally labilecompounds.
Expensive.
Produce dirty extracts which may need
clean-up.
50
8/10/2019 Sample Prep note
51/89
Introduced by Pawliszyns group (1990).
Was developed to address the need for a fast,
solvent-free, cost effective & field compatiblesample preparation method.
Is a microextraction technique, which means thatthe amount of extraction solvent is very small
compared to the sample volume.
Solid phase microextraction (SPME)
51
8/10/2019 Sample Prep note
52/89
All steps of the LLE such as extraction,concentration, (derivatization) & introduction to
the chromatograph are integrated into single
step & 1 device, considerably simplifying the
sample preparation procedure.
SPME reduces the time necessary for sample
preparation, decreases purchase & disposalcosts of solvents & can improve detection limits.
52
8/10/2019 Sample Prep note
53/89
The SPME technique enables the simultaneous
extraction & pre-concentration of volatile & non
volatile analytes from gaseous, liquid & solidsamples & even inorganic species.
SPME not suitable for sample with organic
solvent because organic solvent adsorb onto
the fiber as well & might damage it.
53
8/10/2019 Sample Prep note
54/89
The SPME apparatus is a very simple device.
The device consisting of a fiber holder & a fiber
assembly, with built-in fiber (12 cm long) inside
the needle which looks like a modified syringe.
The SPME fiber itself is a thin fused-silica optical
fiber, coated with a thin polymer film (such as
PDMS).
This film acts like a sponge, concentrating the
analytes on its surface during adsorption from
the sample matrix.54
8/10/2019 Sample Prep note
55/89
Design of the commercial SPME device
55
8/10/2019 Sample Prep note
56/89
56
8/10/2019 Sample Prep note
57/89
57
8/10/2019 Sample Prep note
58/89
When the coated fiber is placed into an aqueousmatrix for a pre-determined time, the analyte is
transferred from the matrix into the coating.
The extraction is considered to be complete when
the analyte has reached an equilibrium distributionbetween the matrix & fiber coating.
SPME Principle
58
8/10/2019 Sample Prep note
59/89
SPME relies upon the extraction of solutes from asample into the SPME absorptive layer.
After a sampling period - during which extractionhas ideally reached equilibrium - the absorbed
solutes are transferred into an inlet system that
desorbs the solutes into a gas (for GC) or liquid(for LC) mobile phase.
During desorption of the analyte, the polymericphase is cleaned & therefore ready for reuse.
59
8/10/2019 Sample Prep note
60/89
1. Direct Immersion (DI-SPME)
In DI-SPME, the fiber is directlyimmersed/inserted in liquid samples.
DI-SPME works best for low concentration
water based sample matrices.
Position the fiber just below the sample
surface & maintain this position consistently
for all extractions.
Two Basic Types of Extractions
60
8/10/2019 Sample Prep note
61/89
2. Head Space (HS-SPME)
In HS-SPME, the fiber is exposed in the vapor
phase above a gaseous, liquid or solid sample.
HS-SPME is suitable for the analysis of VOCs.
It is extremely important to keep the headspace
volume constant & keep the fiber position at thesame depth every time.
61
8/10/2019 Sample Prep note
62/89
62
8/10/2019 Sample Prep note
63/89
SPME procedure (extraction)
1. Pierce septum on samplecontainer;
2. Expose SPMEfiber/extract analytes;
3. Retract fiber/withdrawneedle.
63
8/10/2019 Sample Prep note
64/89
SPME procedure (desorption)
4. Pierce septum in GC inlet
(or introduce needle intoSPME/HPLC interface);
5. Expose fiber/desorbanalytes;
6. Retract fiber/withdraw
needle.
64
8/10/2019 Sample Prep note
65/89
Parameters affect the extraction efficiency :
Types of fiber coating
Thickness of the fiber coating
Extraction time
Extraction temperature
Sample agitation
Sample pH
Salt addition
65
8/10/2019 Sample Prep note
66/89
1. Type of fiber coating
The fiber is coated with a thin polymeric film,
which concentrates the analytes.
The type of fiber used affects the selectivity of
extraction (in general, polar fibers are used forpolar analytes & non-polar fibers for non-polar
analytes).
Optimizing extraction condition
66
8/10/2019 Sample Prep note
67/89
67
8/10/2019 Sample Prep note
68/89
2. Thickness of the fiber coating
The thickness affects both the equilibrium
time & sensitivity of the method.
The use of a thicker fiber requires a longer
extraction time but the recoveries are
generally higher.
68
8/10/2019 Sample Prep note
69/89
The thinnest film is employed to reduce
extraction times (faster the partition equilibriumcan be reached).
Volatile compounds require a thick film & a thin
film is effective for semivolatile compounds.
69
8/10/2019 Sample Prep note
70/89
3. Extraction time
The longer the extraction time, the more
solutes adsorbed onto the fiber until it reach
optimum extraction time whereby the amountof the adsorbed solutes is max.
Increasing the extraction time then will not
increase the amount anymore.
If the extraction is lower than the optimum
time, possibility only a fraction of solutes will
be adsorbed.70
8/10/2019 Sample Prep note
71/89
The extraction time depends on the size of the
compounds, fiber coating, type of extraction used &sample concentration.
Extraction times can be shorter when :
Analyzing small compounds (< 150 MW)
Using thinner fiber coatings. Using the headspace technique.
Working with high concentration samples (high
ppb or ppm range).
Extractions typically take 15-20 minutes.
71
8/10/2019 Sample Prep note
72/89
4. Extraction temperature
Optimization of extraction temperature is
generally more important in the headspace mode.
The use of heat during headspace SPME will help
release the analyte from the sample, improve
sensitivity & shorten the extraction time.
A constant temperature is advisable for all
extractions to obtain good precision.
72
8/10/2019 Sample Prep note
73/89
5. Sample agitation
Agitation accelerates the transfer of analytes
from the sample matrix to the fiber & reduce
the equilibrium time.
Agitation methods include magnetic stirring,sonication & vibration.
Vigorous or harsh agitation modes such as
sonication may affect the coating, thus they
should be used with caution.73
6 Sample pH
8/10/2019 Sample Prep note
74/89
6. Sample pH
Adjustment of sample pH may improve the
extraction yield for compound that can be
protonated.
pH is adjusted in order to obtain the analyte in
its neutral form to enhance the extraction
yield.
E.g., a sample may be acidified for extracting
acidic compounds or made alkaline for
extracting basic compounds.74
8/10/2019 Sample Prep note
75/89
f S
8/10/2019 Sample Prep note
76/89
Advantages of SPME
It is a rapid (analysis time very fast).
The configurations & operation of the SPMEdevices are very simple.
Reducing cost.
Solvent free & sensitive method for the
extraction of analytes.
Clean extracts & do not need any clean-up
or concentration.76
8/10/2019 Sample Prep note
77/89
It gives highly consistent, quantifiable results
from very low concentration of analytes.
SPME is becoming widely used as anextraction & concentration step prior to MS
analysis.
77
8/10/2019 Sample Prep note
78/89
It has small size, which is convenient for
designing portable devices for field or on-site
sampling.
On-site sampling can be done even by
nonscientists without the need to have
equipment at each location.
When properly stored, samples can beanalyzed days later in the laboratory without
significant loss of volatiles.
78
8/10/2019 Sample Prep note
79/89
Limitations of SPME
The volume of the polymer extraction phase isvery small & requires extreme precision during
manufacturing of the coating.
The quality of the fibers depends on the
manufacturer & sometimes the performance is
different from batch to batch.
79
8/10/2019 Sample Prep note
80/89
Some level of degradation of the fiber occurs
during repeated usage.
The carry-over of the fiber is also a problem
that in some cases is difficult to eliminate.
Fibers are fragile & can easily be broken.
80
8/10/2019 Sample Prep note
81/89
Application of SPME in various fields
of analytical chemistry
Environmental (water analysis).
Food samples (analysis of wines & other
alcoholic beverages).
Biological fluids (blood, urine).
Forensic, clinical & pharmaceutical (analysis
of drugs).
81
8/10/2019 Sample Prep note
82/89
Sample problem
Briefly discuss how the used of fiber give advantages to
SPME method when compared to SPE method.
82
8/10/2019 Sample Prep note
83/89
How if we use LLE ???(in term of extract quality, may or may
not require clean-up & concentration
step)
83
8/10/2019 Sample Prep note
84/89
84
Separation methods
Method Basic of method
Extraction Different in solubility in two immiscible
liquids
Ion exchange Difference in interaction of reactantswith ion exchange resin
Chromatography Difference in rate of movement of a
solute through a stationary phaseElectrophoresis Difference in migration rate of
charged species in an electric field
8/10/2019 Sample Prep note
85/89
Derivatization A process of chemically modifying a compound
to produce a new compound which has
properties that are suitable for analysis.
Organic compounds with readily exchangeable
H (acids, alcohols) generally exhibit poor
chromatographic behavior on a polar GCstationary phases.
85
8/10/2019 Sample Prep note
86/89
Advantages of derivatization :
Improved chromatography (better sensitivity
& quantification).
Prevents GC column degradation.
86
8/10/2019 Sample Prep note
87/89
Common derivatization methods:1. Fatty Acids
Methylation (BF3/MeOH or MeOH/HCl) to yield
Fatty Acid Methyl Esters (FAMES). Silylation (BSTFA) to yield trimethylsilyl (TMS)
esters.
2. Alcohols (n-alkanols, sterols)
Acylation (pyridine, acetic anhydride) to yield
acetates. Silylation (BSTFA) to yield trimethylsilyl (TMS)
ethers.
87
8/10/2019 Sample Prep note
88/89
Pyrolysis
Pyrolysis is a form of incineration that
chemically decomposes organic materials byheat in the absence of oxygen.
Pyrolysis typically occurs under pressure & atoperating temperatures above 430 C (800 F).
Organic materials are transformed into gases,small quantities of liquid & a solid residue
containing carbon & ash.
88
8/10/2019 Sample Prep note
89/89
Application
Pyrolysis treats & destroys :
Semi-volatile organic compounds (SVOCs),
fuels & pesticides in soil.
The process is applicable for the treatment of
organics from refinery wastes, coal tar wastes,
contaminated soils, hydrocarbons & volatileorganic compounds (VOCs).