Non-contact free-fall concept

Water passes through the AquaScat without touching the optics:

» No window fouling. Hence, the measured values are not falsified and there is no drift.

» Very Low to very high turbidity values can be measured with no contamination of the optics

» Dissolved metals like iron or manganese do not contaminate the optics

» Extremely low maintenance is the result

Measuring light passes through the entire sample beam:

» There is no selection of the sample at the surface. The entire water beam is measured and hence, the result is representative.

Page 4: Sigrist AquaScat. AquaScat WTM A Page 2Doc. No. 11892 E/1

AquaScat WTM A

Page 4 Doc. No. 11892 E/1

Page 5: Sigrist AquaScat. AquaScat WTM A Page 2Doc. No. 11892 E/1

AquaScat WTM A

Page 5 Doc. No. 11892 E/1

Re-calibration with secondarystandard (High stability optical glass)

» Precise re-calibration is possible without the use of Formazine

» Allows a stable measurement over a long time period

» Allows a fully automatic operation » Purchase and storage of Formazine

is not needed

WTMA: Automatic calibration check

WTM and HT: Manual calibration check

Page 6: Sigrist AquaScat. AquaScat WTM A Page 2Doc. No. 11892 E/1

AquaScat WTM A

Page 6 Doc. No. 11892 E/1

Manual calibration method for AquaScat WTM and HT

Page 7: Sigrist AquaScat. AquaScat WTM A Page 2Doc. No. 11892 E/1

AquaScat WTM A

Page 7 Doc. No. 11892 E/1

Automatic calibration method for AquaScat WTMA

Page 8: Sigrist AquaScat. AquaScat WTM A Page 2Doc. No. 11892 E/1

AquaScat WTM A

Page 8 Doc. No. 11892 E/1

Low levels of stray light

The design of the AquaScat in combination with

high quality optical components minimizes the

quantity of stray light inside the instrument:

Page 9: Sigrist AquaScat. AquaScat WTM A Page 2Doc. No. 11892 E/1

AquaScat WTM A

Page 9 Doc. No. 11892 E/1

Control Unit

The control unit of all AquaScat is a colour touch screen:

» Values, graphs, alarm and status messages can be selected by the operator.

» An internal data logger allows the recall of measured data for the last 32 days

» Password protection avoids unauthorized access

» Various means of communication including an integrated web server

Page 10: Sigrist AquaScat. AquaScat WTM A Page 2Doc. No. 11892 E/1

AquaScat WTM A

Page 10 Doc. No. 11892 E/1

Cost of ownership

In addition to its daily dependability, the product design was focused on long useable life and low maintenance costs:

» The maintenance work is very low» The cost for consumables during 10

years of operation are at AUD 150

Page 11: Sigrist AquaScat. AquaScat WTM A Page 2Doc. No. 11892 E/1

AquaScat WTM A

Page 11 Doc. No. 11892 E/1

Turbidity measurement application

» The installation is simple» An optional de-aeration system can be

installed if required.» 2 Analogue outputs can transfer two

measuring ranges to the PLC

Page 12: Sigrist AquaScat. AquaScat WTM A Page 2Doc. No. 11892 E/1

AquaScat WTM A

Page 12 Doc. No. 11892 E/1

AquaScat Wall Mounting Set

Level Control overflow

Inlet to AquaScat

Flow Control Valve

Docking Station

Level Control

De-aeration tube

Water Inlet

Water Outlet