View
216
Download
0
Embed Size (px)
Citation preview
Slide 1 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Lecture 1
The Principles of Microscopy
BME 695Y / BMS 634
Confocal Microscopy: Techniques and Application Module
Purdue University Department of Basic Medical Sciences, School of Veterinary Medicine
& Department of Biomedical Engineering, Schools of Engineering
J.Paul Robinson, Ph.D.
Professor of Immunopharmacology & Biomedical EngineeringDirector, Purdue University Cytometry Laboratories
These slides are intended for use in a lecture series. Copies of the graphics are distributed and students encouraged to take their notes on these graphics. The intent is to have the student NOT try to reproduce the figures, but to LISTEN and UNDERSTAND the
material. All material copyright J.Paul Robinson unless stated. A useful textbook for this lecture series is Jim Pawley’s “Handbook of Confocal Microscopy” Plenum Press which has been used extensively for material and ideas to support the class.
UPDATED February 2004
http://www.cyto.purdue.edu/
Slide 2 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Evaluation• Required 100% attendance at all lectures and Practicals
• Completion of all practical material
• Demonstrated competence in practical classes• Complete laboratory notebook at end of class (these will be
graded)
Overview of the Course
• Basic Image Analysis• 3D image analysis• Live Cell Studies• Advanced Applications
• Microscopy• Fluorescence• Basic Optics• Confocal Microscopes
Slide 3 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Learning Goals of this Course
At the end of this course you will:
• Have a good background in 2 D image analysis
• Understand the basics of image structure
• Know the basics of 3D image analysis
• Understand the operation and function of a transmitted light microscope, fluorescence microscope and confocal microscope
• Learn about preparation techniques and assay systems
• Learn about many applications of the technologies of confocal imaging
Practical Aspects
• Learn to use a microscope, a fluorescence microscope, a confocal microscope
• Learn to do basic digital imaging, image manipulation, 3D analysis
• Learn to use several image analysis packages
Slide 4 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Introduction to Lecture 11. Introduction to Microscopy2. History of Microscopy3. Köhler Illumination4. Basic optical terms5. Light absorption6. Magnification7. Optical properties of microscopes8. Components of the microscope9. Numerical aperture, resolution and refractive index10. Aberrations11. Interference and optical filters
Slide 5 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Microscopes
• Upright
• Inverted
• Köhler Illumination
• Fluorescence Illumination
"Microscope" was first coined by members of the first "Academia dei Lincei" (Academy of the Lynx} scientific society which included Galileo. It was not Galileo tho’ who came up with the word, it was Johannes Faber, an entomologist and member of the same society that gave the magnifying instrument the name “microscope”
Slide 6 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Earliest Microscopes• 1590 - Hans & Zacharias Janssen of Middleburg, Holland manufactured the first compound microscopes• 1590 – 1609 - Galileo – one of the earliest microscopists (naming of term “microscope” • 1660 - Marcello Malpighi circa 1660, was one of the first great microscopists, considered the father
embryology and early histology - observed capillaries in 1660• 1665 - Robert Hooke (1635-1703)- book Micrographia, published in 1665, devised the compound
microscope most famous “microscopical” observation was his study of thin slices of cork. He wrote:
“. . . I could exceedingly plainly perceive it to be all perforated and porous. . . these pores, or cells, . . . were indeed the first microscopical pores I ever saw, and perhaps, that were ever seen, for I had not met with any Writer or Person, that had made any mention of them before this.”
Slide 7 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Earliest Microscopes•1673 - Antioni van Leeuwenhoek (1632-1723) Delft, Holland, worked as a draper (a fabric merchant); he is also known to have worked as a surveyor, a wine assayer, and as a minor city official.
•Leeuwenhoek is incorrectly called "the inventor of the microscope" •Created a “simple” microscope that could magnify to about 275x, and
published drawings of microorganisms in 1683
•Could reach magnifications of over 200x with simple ground lenses - however compound microscopes were mostly of poor quality and could only magnify up to 20-30 times. Hooke claimed they were too difficult to use - his eyesight was poor.
•Discovered bacteria, free-living and parasitic microscopic protists, sperm cells, blood cells, microscopic nematodes •In 1673, Leeuwenhoek began writing letters to the Royal Society of London - published in Philosophical Transactions of the Royal Society•In 1680 he was elected a full member of the Royal Society, joining Robert Hooke, Henry Oldenburg, Robert Boyle, Christopher Wren
lens.exe
Slide 8 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Early Microscopes (Hooke)
1665
Slide 9 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
1670-1690
• Back: Italian compound microscopes - 1670
• Italian Compound microscopes
• Back: 1670 (probably Campani)
• This microscope was formerly at the University of Bologna - it contains a field lens which was the first optical advance about 1660. Only opaque objects can be viewed.
• Front: Guiseppe Campani, Rome - 1690 - Campani was the leading Italian telescope and microscope maker in the late `17th century - he probably invented the screw focusing mechanism shown on this scope - the slide holder in the base allows transparent and opaque objects to be viewed
Slide 10 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Screwbarrel Microscope - 1720
• Made by Charles Culpeper
Slide 11 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Secondary Microscopes
• George Adams Sr. made many microscopes from about 1740-1772 but he was predominantly just a good manufacturer not inventor (in fact it is thought he was more than a copier!)
• Simple microscopes could attain around 2 micron resolution, while the best compound microscopes were limited to around 5 microns because of chromatic aberration
• In the 1730s a barrister names Chester More Hall observed that flint glass (newly made glass) dispersed colors much more than “crown glass” (older glass). He designed a system that used a concave lens next to a convex lens which could realign all the colors. This was the first achromatic lens. George Bass was the lens-maker that actually made the lenses, but he did not divulge the secret until over 20 years later to John Dollond who copied the idea in 1759 and patented the achromatic lens.
Slide 12 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
George Adams Toymaker to Kings
• This microscope made by George Adams, Mathematical Instrument maker to King George III around 1763, It was probably intended for the Prince of Wales, the future King George IV. The instrument is based on the design of the “Universal Double Microscope" (London Museum of
Science)
Slide 13 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
The famous patent of 1758• George Bass was the
lens-maker that actually made the lenses, but he did not divulge the secret until over 20 years later to John Dollond who copied the idea in 1757 and patented the achromatic lens in 1758.
Slide 14 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Secondary Microscopes
• George Adams Sr. made many microscopes from about 1740-1772 but he was predominantly just a good manufacturer not inventor (in fact it is thought he was more than a copier!)
“New Improved Compound Microscope, George Adams, 1790Adams described this instrument in his “Essays on the Microscope” in 1787. The mechanism allowed freedom of movement. The specimen could be viewed in direct light or in light reflected from a large mirror.
© J.Paul Robinson
Slide 15 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Giovanni Battista Amici• In 1827 Giovanni Battista Amici, built high quality microscopes and
introduced the first matched achromatic microscope in 1827. He had previously (1813) designed “reflecting microscopes” using curved mirrors rather than lenses. He recognized the importance of coverslip thickness and developed the concept of “water immersion”
© J.Paul Robinson© J.Paul Robinson
Slide 16 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Joseph Lister• In 1830, by Joseph Jackson Lister (father of Lord Joseph Lister) solved the
problem of Spherical Aberration - caused by light passing through different parts of the same lens. He solved it mathematically and published this in the Philosophical Transactions in 1830
© J.Paul Robinson
Slide 17 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Pasteur - 1860
Louis Pasteur – his microscope was made in Paris by Nachet in about 1860 and was brass
Slide 18 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Abbe & Zeiss • Ernst Abbe together with Carl Zeiss published a paper in 1877 defining the
physical laws that determined resolving distance of an objective. Known as Abbe’s Law
“minimum resolving distance (d) is related to the wavelength of light (lambda) divided by the Numeric Aperture, which is proportional to the angle of the light cone (theta) formed by a point on the object, to the
objective”.
Slide 19 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Abbe & Zeiss
• Abbe and Zeiss developed oil immersion systems by making oils that matched the refractive index of glass. Thus they were able to make the a Numeric Aperture (N.A.) to the maximum of 1.4 allowing light microscopes to resolve two points distanced only 0.2 microns apart (the theoretical maximum resolution of visible light microscopes). Leitz was also making microscope at this time.
Zeiss student microscope 1880
Slide 20 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Abbe, Zeiss & Schott
• Abbe and Zeiss developed oil immersion systems by making oils that matched the refractive index of glass. Thus they were able to make the a Numeric Aperture (N.A.) to the maximum of 1.4 allowing light microscopes to resolve two points distanced only 0.2 microns apart (the theoretical maximum resolution of visible light microscopes). Leitz was also making microscope at this time.
• Dr Otto Schott formulated glass lenses that color-corrected objectives and produced the first “apochromatic” objectives in 1886.
Slide 21 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Modern Microscopes
• Early 20th Century Professor Köhler developed the method of illumination still called “Köhler Illumination”
• Köhler recognized that using shorter wavelength light (UV) could improve resolution
Slide 22 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Köhler Illumination
Specimen Field stopField iris
Conjugate planes for illuminating rays
Specimen Field stopField iris
Conjugate planes for image-forming rays
condenser eyepiece
retina
Slide 23 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Some Definitions• Absorption
– When light passes through an object the intensity is reduced depending upon the color absorbed. Thus the selective absorption of white light produces colored light.
• Refraction– Direction change of a ray of light passing from one transparent medium to
another with different optical density. A ray from less to more dense medium is bent perpendicular to the surface, with greater deviation for shorter wavelengths
• Diffraction– Light rays bend around edges - new wavefronts are generated at sharp
edges - the smaller the aperture the lower the definition
• Dispersion– Separation of light into its constituent wavelengths when entering a
transparent medium - the change of refractive index with wavelength, such as the spectrum produced by a prism or a rainbow
Slide 24 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Properties of Light
• Refraction• A Lens• Refractive Index• Numerical Aperture• Resolution• Aberrations• Fluorescence
Slide 25 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Reflection and Refraction• Snell’s Law: The angle of
reflection (Ør) is equal to the angle of incidence (Øi) regardless of the surface material
• The angle of the transmitted beam (Øt) is dependent upon the composition of the material
t
i
r
Incident Beam
Reflected BeamTransmitted
(refracted)Beam
n1 sin Øi = n2 sin Øt
The velocity of light in a material of refractive index n is c/n
Slide 26 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Properties of thin Lensesf
1
p+
1
q=
1
f
f
p q
Resolution (R) = 0.61 xNA
Magnification = q
p(lateral)(Rayleigh criterion)
Slide 27 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Refraction & Dispersion
Light is “bent” and the resultant colors separate (dispersion).Red is least refracted, violet most refracted.
dispersion
Short wavelengths are “bent” more than long wavelengths
refraction
Slide 28 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Refraction
But it is really here!!
He sees the fish here….
Slide 29 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Absorption
Control
No blue/green light red filter
Slide 30 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Absorption Chart
Color in white lightColor in white light Color of light absorbedColor of light absorbed
red
blue
green
magenta
cyan
yellow
blue
blue
blue
blue
green
green
green
green
red
red
red
redblack
gray green bluepink
Slide 31 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
The light spectrumWavelength ---- Frequency
Blue light
488 nm
short wavelength
high frequency
high energy (2 times the red)
Red light
650 nm
long wavelength
low frequency
low energy
Photon as a wave packet of energy
Slide 32 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Magnification
• An object can be focussed generally no closer than 250 mm from the eye (depending upon how old you are!)
• this is considered to be the normal viewing distance for 1x magnification
• Young people may be able to focus as close as 125 mm so they can magnify as much as 2x because the image covers a larger part of the retina - that is it is “magnified” at the place where the image is formed
Slide 33 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Magnification1000mm
35 mm slide24x35 mm
M = 1000 mm35 mm
= 28
p The projected image is 28 times larger than we would see it at 250 mm from our eyes.
If we used a 10x magnifier we would have a magnification of 280x, but we would reduce the field of view by a factor of 10x.
Slide 34 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Some Principles
• Rule of thumb is is not to exceed 1,000 times the NA of the objective
• Modern microscopes magnify both in the objective and the ocular and thus are called “compound microscopes” - Simple microscopes have only a single lens
Slide 35 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Basic Microscopy
• Bright field illumination does not reveal differences in brightness between structural details - i.e. no contrast
• Structural details emerge via phase differences and by staining of components
• The edge effects (diffraction, refraction, reflection) produce contrast and detail
Slide 36 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Microscope Basics
• Originally conformed to the German DIN standard
• Standard required the following– real image formed at a tube length
of 160mm
– the parfocal distance set to 45 mm
– object to image distance set to 195 mm
• Currently we use the ISO standard
Focal lengthFocal lengthof objectiveof objective= 45 mm= 45 mm
MechanicalMechanicaltube lengthtube length= 160 mm= 160 mm
Object toImage Distance = 195 mm
Slide 37 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
The Conventional Microscope
Focal lengthof objective= 45 mm
Object toImage Distance = 195 mm
Mechanicaltube length= 160 mm
Modified from “Pawley “Handbook of Confocal Microscopy”, Plenum Press
Slide 38 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Upright Scope
BrightfieldSource
Epi-illuminationSource
Slide 39 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Inverted Microscope
BrightfieldSource
Epi-illuminationSource
Slide 40 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Slide 41 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Conventional Finite Opticswith Telan system
Sample being imaged
Intermediate Image
Telan Optics
Objective
Other optics
Ocular
45 mm
160 mm195 mm
Modified from “Pawley “Handbook of Confocal Microscopy”, Plenum Press
Slide 42 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Infinity Optics
Sample being imaged
Primary Image Plane
Objective
Other optics
Ocular
Other optics
Tube Lens
InfiniteImageDistance
The main advantage of infinity corrected lens systems is the relative insensitivity to additional optics within the tube length. Secondly one can focus by moving the objective and not the specimen (stage)
Modified from “Pawley “Handbook of Confocal Microscopy”, Plenum Press
Slide 43 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Images reproduced from:
http://micro.magnet.fsu.edu/
Slide 44 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
• Microscope Basics, Magnification, Optical systems
Images reproduced from:
http://micro.magnet.fsu.edu/
Slide 45 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Microscope Components• Ocular • Objectives• Condensor • Numerical Aperture• Refractive Index• Aberrations• Optical Filters
Slide 46 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Ocular - Eyepiece• Essentially a projection lens (5x to 15x
magnification) Note: there is usually an adjustment call the inter-pupillary distance on eyepieces for personal focusing
• Huygenian– Projects the image onto the retina of the eye– your eye should not be right on the lens, but
back from it (eyecups create this space)
• Compensating– designed to work with specific apochromatic or flat
field objectives - it is color compensated and cannot be mixed with other objectives (or microscopes)
• Photo-adapter– designed to project the image on the film in the
camera - usually a longer distance and lower magnification from 0.5x to 5x
Immediate above Images reproduced from:
http://micro.magnet.fsu.edu/
Slide 47 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Condensor• Has several purposes
– must focus the light onto the specimen
– fill the entire numerical aperture of the objective (i.e. it must match the NA of the objective)
• Most microscopes will have what is termed an “Abbe” condenser (not corrected for aberrations)
• Note: If you exceed 1.0 NA objective, you probably will need to use oil on the condensor as well (except in inverted scopes)
Slide 48 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Microscope Objectives
Immediate above Images reproduced from:
http://micro.magnet.fsu.edu/
Slide 49 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Objectives
PLAN-APO-40X 1.30 N.A. 160/0.22
Flat field Apochromat Magnification Numerical Tube CoverglassFactor Aperture Length Thickness
- Infinity corrected
Slide 50 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Objectives
Limit for smallest resolvable distance d between 2 points is (Rayleigh criterion):
d = 1.22
Thus high NUMERICAL APERTURE is critical for high magnification
In a medium of refractive index n the wavelength gets shorter:n
This defines a “resel” or “resolution element”
Slide 51 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Numerical Aperture• Resolving power is directly related to numerical aperture.
• The higher the NA the greater the resolution
• Resolving power:The ability of an objective to resolve two distinct lines very close
together
NA = n sin u
– (n=the lowest refractive index between the object and first objective element) (hopefully 1)
– u is 1/2 the angular aperture of the objective
Slide 52 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
A
NA=n(sin )
Light cone
Slide 53 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Numerical Aperture• The wider the angle the lens is capable of receiving light at, the greater its
resolving power
• The higher the NA, the shorter the working distance
Images reproduced from:
http://micro.magnet.fsu.edu/
Slide 54 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Numerical Aperture• For a narrow light beam (i.e. closed illumination aperture diaphragm) the
finest resolution is (at the brightest point of the visible spectrum i.e. 530 nm)…(closed condenser).
NA
2 x NA
.000532 x 1.00= 0.265 m
.000531.00 = 0.53 m
• With a cone of light filling the entire aperture the theoretical resolution is…(fully open condenser)..
=
=
Slide 55 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Object Resolution• Example:
40 x 1.3 N.A. objective at 530 nm light
2 x NA
.000532 x 1.3
= 0.20 m=
40 x 0.65 N.A. objective at 530 nm light
2 x NA
.000532 x .65
= 0.405 m=
R=/(2NA) 1R=0.61 /NA 2R=1.22 /(NA(obj) + NA(cond)) 3
Slide 56 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Images reproduced from:
http://micro.magnet.fsu.edu/
Slide 57 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Microscope Objectives
SpecimenCoverslip
Oil
MicroscopeObjective
Stage
60x 1.4 NAPlanApo
Slide 58 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Refractive Index
Objective
n=1.52
n = 1.52
n = 1.52
Specimen
Coverslip
Oil
n=1.33
n = 1.52
n = 1.0
n = 1.5
Water
n=1.52
Air
Slide 59 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Numerical Aperture• Resolving power is directly related to numerical aperture.
• The higher the NA the greater the resolution
• Resolving power:The ability of an objective to resolve two distinct lines very close
together
NA = n sin u
– (n=the lowest refractive index between the object and first objective element) (hopefully 1)
– u is 1/2 the angular aperture of the objective
Slide 60 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
A
NA=n(sin )
Light cone
Slide 61 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Numerical Aperture• The wider the angle the lens is capable of receiving light at, the greater its
resolving power
• The higher the NA, the shorter the working distance
Images reproduced from:
http://micro.magnet.fsu.edu/
Slide 62 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Numerical Aperture• For a narrow light beam (i.e. closed illumination aperture diaphragm) the
finest resolution is (at the brightest point of the visible spectrum i.e. 530 nm)…(closed condenser).
NA
2 x NA
.000532 x 1.00= 0.265 m
.000531.00 = 0.53 m
• With a cone of light filling the entire aperture the theoretical resolution is…(fully open condenser)..
=
=
Slide 63 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Object Resolution• Example:
40 x 1.3 N.A. objective at 530 nm light
2 x NA
.000532 x 1.3
= 0.20 m=
40 x 0.65 N.A. objective at 530 nm light
2 x NA
.000532 x .65
= 0.405 m=
R=/(2NA) 1R=0.61 /NA 2R=1.22 /(NA(obj) + NA(cond)) 3
Slide 64 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Images reproduced from:
http://micro.magnet.fsu.edu/
Slide 65 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Microscope Objectives
SpecimenCoverslip
Oil
MicroscopeObjective
Stage
60x 1.4 NAPlanApo
Slide 66 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Refractive Index
Objective
n=1.52
n = 1.52
n = 1.52
Specimen
Coverslip
Oil
n=1.33
n = 1.52
n = 1.0
n = 1.5
Water
n=1.52
Air
Slide 67 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
• Monochromatic Aberrations– Spherical aberration
– Coma
– Astigmatism
– Flatness of field
– Distortion
• Chromatic Aberrations– Longitudinal aberration
– Lateral aberration
Sources of Aberrations
Images reproduced from:
http://micro.magnet.fsu.edu/
Slide 68 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Monochromatic Aberration - Spherical aberration
Generated by nonspherical wavefronts produced by the objective, and increased tube length, or inserted objects such as coverslips, immersion oil, etc. Essentially, it is desirable only to use the center part of a lens to avoid this problem.
F1 F2
F1
Corrected lens
Immediate left Image reproduced from:
http://micro.magnet.fsu.edu/
Slide 69 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Monochromatic Aberrations - Coma
Coma is when a streaking radial distortion occurs for object points away from the optical axis. It should be noted that most coma is experienced “off axis” and therefore, should be less of a problem in confocal systems.
1
23
Images reproduced from:
http://micro.magnet.fsu.edu/
Slide 70 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Monochromatic Aberrations - Astigmatism
If a perfectly symmetrical image field is moved off axis, it becomes either radially or tangentially elongated.
Images reproduced from:
http://micro.magnet.fsu.edu/
Slide 71 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
• Monochromatic Aberrations– Flatness of Field– Distortion
Lenses are spherical and since points of a flat image are focused onto a spherical dish, the central and peripheral zones will not be in focus. Complex Achromat and PLANAPOCHROMAT lenses partially solve this problem but at reduced transmission.
DISTORTION occurs for objects components out of axis. Most objectives correct to reduce distortion to less than 2% of the radial distance from the axis.
Slide 72 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Useful Factoids
The intensity of light collected The intensity of light collected decreasesdecreases as the square of the magnificationas the square of the magnification
The intensity of light The intensity of light increases increases as the as the square of the numerical aperturesquare of the numerical aperture
Slide 73 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Fluorescence Microscopes• Cannot view fluorescence emission in a single optical
plane
• Generally use light sources of
much lower flux than confocal systems
• Are cheaper than confocal systems
• Give high quality photographic images
(actual photographs) whereas confocal
systems are restricted to small resolution images
Slide 74 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Fluorescent Microscope
Dichroic Filter
Objective
Arc Lamp
Emission Filter
Excitation Diaphragm
Ocular
Excitation Filter
EPI-Illumination
Slide 75 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Interference in Thin Films• Small amounts of incident light are reflected at the
interface between two material of different RI
• Thickness of the material will alter the constructive or destructive interference patterns - increasing or decreasing certain wavelengths
• Optical filters can thus be created that “interfere” with the normal transmission of light
Interference and Diffraction: Gratings• Diffraction essentially describes a departure from
theoretical geometric optics
• Thus a sharp objet casts an alternating shadow of light and dark “patterns” because of interference
• Diffraction is the component that limits resolution
Slide 76 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Polarization and Phase: Interference
• Electric and magnetic fields are vectors - i.e. they have both magnitude and direction
• The inverse of the period (wavelength) is the frequency in Hz
Wavelength (period T)
Axis of
Magnetic F
ield
Axis of Propagation
Axi
s of
Ele
ctri
c F
ield
Modified from Shapiro “Practical Flow Cytometry” Wiley-Liss, p78
Slide 77 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Interference
ConstructiveInterference
DestructiveInterference
A
B
C
D
A+B
C+D
Am
plitude
0o 90o 180o 270o 360o Wavelength
Figure modified from Shapiro “Practical Flow Cytometry” Wiley-Liss, p79
Here we have a phase difference of 180o (2 radians) so the waves cancel each other out
The frequency does not change, but the amplitude is doubled
Slide 78 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Construction of Filters
Dielectric filtercomponents
Single Opticalfilter
“glue”
Multiple elements
Coatings are often magnesium fluoride
Anti-Reflection Coatings
Slide 79 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Anti-Reflection Coatings
Optical Filter
MultipleElements
Coatings are often magnesium fluoride
Dielectric filtercomponents
Slide 80 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Band Pass Filters
Transmitted LightWhite Light Source
630 nm BandPass Filter
620 -640 nm Light
Transmitted LightLight Source520 nm Long Pass Filter
>520 nm Light
Transmitted LightLight Source575 nm Short Pass Filter
<575 nm Light
Long Pass Filters
Short Pass Filters
Slide 81 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Optical Filters
Dichroic Filter/Mirror at 45 deg
Reflected light
Transmitted LightLight Source
510 LP dichroic Mirror
Slide 82 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Filter Properties Light Transmission
%T
Wavelength
100
0
50
Notch
Ban
dp
ass
Slide 83 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
The intensity of the radiation is inversely proportional to the square of the distance traveled
Slide 84 t:/classes/BMS602B/lecture 1 602_B.ppt© J.Paul Robinson - Purdue University Cytometry Laboratories
Summary Lecture 1
• History, simple versus compound microscopes
• Köhler illumination
• Refraction, Absorption, dispersion, diffraction, Magnification
• Upright and inverted microscopes
• Optical Designs - 160 mm and Infinity optics
• Components of the microscope
• Numerical Aperture (NA)
• Refractive Index/refraction (RI), Aberrations
• Fluorescence microscope
• Properties of optical filters