Joint AESAN/EFSA Workshop Science Supporting Risk Surveillance of Imports’ - 10 February 2010, Seville (Spain) The importance of harmonised analytical methods to provide reliable results Robouch, Piotr European Commission Joint Research Centre (JRC) Institute of Reference Materials and Measurements (IRMM), Geel (Belgium) [email protected] At the previous EFSA Joint Event in Paris, several speakers mentioned the importance of harmonised analytical methods to provide reliable data in the frame of official control or for customs inspection of imported goods. How to guarantee reliable results? A possible demonstration may reside in the participation to "inter-laboratory comparisons" more specifically proficiency testing (PT) exercises meant to determine the performance of laboratories for specific test/measurements and to monitor laboratories'' continuing performance. Hence, it is expected from a laboratory having reported satisfactory results in a PT exercise for a given measurand (i.e. a given analyte in a particular matrix determined by a specific analytical technique) to report reliable results for similar samples. Such quality results may either reduce the number of unjustified disputes based on discrepant analytical results; or properly identify a serious risk to be notified. The IUPAC "Harmonised protocol for the proficiency testing of analytical chemistry laboratory" (Thomson et al. Pure Appl. Chem., Vol. 78, No. 1, pp. 145196, 2006) sets the basic principles on how to organise a PT and to evaluate PTs. These principles have taken onboard in two complementary ISO standards: - ISO 17043:2010 "Conformity assessment General requirements for proficiency testing" (recently published) and ISO 13528:2005 "Statistical methods for use in proficiency testing by inter-laboratory comparisons". Many inter-laboratory comparisons are organised by the Institute of Reference Materials and Measurements (IRMM). Several PTs are run in the frame of the European Measurement Evaluation Program (IMEP) or in the frame of the Community Reference Laboratories (CRLs) mandates to monitor the reliability of the measurement capability throughout the Member States.

The importance of harmonised analytical methods to provide ... · Statistical methods for use in proficiency testing by inter-laboratory comparisons IUPAC-The international harmonized

  • Upload
    others

  • View
    5

  • Download
    0

Embed Size (px)

Citation preview

Joint AESAN/EFSA Workshop ‘Science Supporting Risk Surveillance of Imports’ - 10 February 2010, Seville (Spain)

The importance of harmonised analytical methods

to provide reliable results

Robouch, Piotr

European Commission – Joint Research Centre (JRC) – Institute of Reference Materials

and Measurements (IRMM), Geel (Belgium)

[email protected]

At the previous EFSA Joint Event in Paris, several speakers mentioned the importance of

harmonised analytical methods to provide reliable data in the frame of official control or for

customs inspection of imported goods.

How to guarantee reliable results? A possible demonstration may reside in the participation to

"inter-laboratory comparisons" – more specifically proficiency testing (PT) exercises – meant to

determine the performance of laboratories for specific test/measurements and to monitor

laboratories'' continuing performance. Hence, it is expected from a laboratory having reported

satisfactory results in a PT exercise for a given measurand (i.e. a given analyte in a particular

matrix determined by a specific analytical technique) to report reliable results for similar

samples. Such quality results may either reduce the number of unjustified disputes based on

discrepant analytical results; or properly identify a serious risk to be notified.

The IUPAC "Harmonised protocol for the proficiency testing of analytical chemistry laboratory"

(Thomson et al. Pure Appl. Chem., Vol. 78, No. 1, pp. 145–196, 2006) sets the basic principles

on how to organise a PT and to evaluate PTs. These principles have taken onboard in two

complementary ISO standards: - ISO 17043:2010 "Conformity assessment General

requirements for proficiency testing" (recently published) and – ISO 13528:2005 "Statistical

methods for use in proficiency testing by inter-laboratory comparisons".

Many inter-laboratory comparisons are organised by the Institute of Reference Materials and

Measurements (IRMM). Several PTs are run in the frame of the European Measurement

Evaluation Program (IMEP) or in the frame of the Community Reference Laboratories (CRLs)

mandates to monitor the reliability of the measurement capability throughout the Member States.

Piotr Robouch – EFSA WS Sevilla 10/02/2009 piotr robouch@ec europa eu 1Piotr Robouch EFSA WS, Sevilla, 10/02/2009 [email protected] 1

Joint Research Centre (JRC)

The importance of harmonised harmonised

analytical methods to provide to provide

reliable resultsPiotr Robouch

IRMM - Institute for Reference Materials and MeasurementsGeel - Belgium

http://irmm.jrc.ec.europa.eu/http://www.jrc.ec.europa.eu/

2Piotr Robouch – EFSA WS Sevilla 10/02/2009 piotr robouch@ec europa eu 2Piotr Robouch EFSA WS, Sevilla, 10/02/2009 [email protected]

What did we hear in Paris?

Harmonisation/improvement of analytical methods

Experience knowledge & confidenceExperience, knowledge & confidence

Setting appropriate contaminants limits

Fair trade

Effective official control & border inspections

Proper monitoring

Appropriate risk analysis/assessment

Rapid and transparent notification

mutual recognition

Piotr Robouch – EFSA WS Sevilla 10/02/2009 piotr robouch@ec europa eu 3Piotr Robouch EFSA WS, Sevilla, 10/02/2009 [email protected] 3

Joint Research Centre (JRC)

6 custom borders

IRMM - Institute for Reference Materials and MeasurementsGeel - Belgium

Taxation = f(sugar content)

http://irmm.jrc.ec.europa.eu/http://www.jrc.ec.europa.eu/

4Piotr Robouch – EFSA WS Sevilla 10/02/2009 piotr robouch@ec europa eu

The challenge4Piotr Robouch EFSA WS, Sevilla, 10/02/2009 [email protected]

5Piotr Robouch – EFSA WS Sevilla 10/02/2009 piotr robouch@ec europa eu

Inter-Laboratory Comparison5Piotr Robouch EFSA WS, Sevilla, 10/02/2009 [email protected]

(ISO 17043:2010)[…]

6Piotr Robouch – EFSA WS Sevilla 10/02/2009 piotr robouch@ec europa eu 6Piotr Robouch EFSA WS, Sevilla, 10/02/2009 [email protected]

7Piotr Robouch – EFSA WS Sevilla 10/02/2009 piotr robouch@ec europa eu 7Piotr Robouch EFSA WS, Sevilla, 10/02/2009 [email protected]

8Piotr Robouch – EFSA WS Sevilla 10/02/2009 piotr robouch@ec europa eu

Relevant documents8Piotr Robouch EFSA WS, Sevilla, 10/02/2009 [email protected]

ISO 17043:2010Conformity assessment Conformity assessment General requirements for proficiency testing

ISO 13528:2006Statistical methods for use in proficiency testing by inter-laboratory comparisons

IUPAC- The international harmonized protocol for the proficiency testing

of analytical chemistry laboratories

- Harmonized guidelines for single-laboratory Validation of methods of Analysis

COMMISSION DECISION 657/2002/EC … concerning the performance of analytical methods and the interpretation of results (cf Food)interpretation of results (cf. Food)

9Piotr Robouch – EFSA WS Sevilla 10/02/2009 piotr robouch@ec europa eu 9Piotr Robouch EFSA WS, Sevilla, 10/02/2009 [email protected]

What can we learn?

10Piotr Robouch – EFSA WS Sevilla 10/02/2009 piotr robouch@ec europa eu 10Piotr Robouch EFSA WS, Sevilla, 10/02/2009 [email protected]

Mandate:

− Coordinate network of national reference laboratoriesCoordinate network of national reference laboratories

− Organising interlaboratory comparisons

− Setting up EU-wide standards for testingg p g

− Providing scientific and technical assistance to EU Commission

− Training analysts from national laboratories

http://www.irmm.jrc.be/html/CRLs/index.htm

11Piotr Robouch – EFSA WS Sevilla 10/02/2009 piotr robouch@ec europa eu 11Piotr Robouch EFSA WS, Sevilla, 10/02/2009 [email protected]

Some PTs organised by IRMM

12Piotr Robouch – EFSA WS Sevilla 10/02/2009 piotr robouch@ec europa eu

Initial assumptions…12Piotr Robouch EFSA WS, Sevilla, 10/02/2009 [email protected]

Same sample treatment

Validated standard operating procedure (same?)p g p ( )

Qualified staff

Same interpretation of the legislation

comparable measurement results

13Piotr Robouch – EFSA WS Sevilla 10/02/2009 piotr robouch@ec europa eu

Cd in Crab13Piotr Robouch EFSA WS, Sevilla, 10/02/2009 [email protected]

(EC) 1881/2006( )

Many people consider as a delicacy:- Hepatopancreas (midgut gland)

… human consumption

Hepatopancreas (midgut gland) Extremely large organ with several functions, including the secretion of digestive enzymes and absorption and storage of digested food. It fills most of storage of digested food. It fills most of the area around the stomach, depending on its contents of food reserves and water.

14Piotr Robouch – EFSA WS Sevilla 10/02/2009 piotr robouch@ec europa eu 14Piotr Robouch EFSA WS, Sevilla, 10/02/2009 [email protected]

The directors of the national metrology institutes(NMIs) of 38 Member States of the BIPM and(NMIs) of 38 Member States of the BIPM andrepresentatives of two international organizationssigned a Mutual Recognition Arrangement(CIPM MRA) for national measurement standardsand for calibration and measurement certificatesissued by NMIs. A number of other institutes haveissued by NMIs. A number of other institutes have

signed since then.

MRA i t i d fMRA is a response to a growing need for an open,transparent and comprehensive scheme to giveusers reliable quantitative information on theqcomparability of national metrology servicesand to provide the technical basis for wideragreements negotiated for international tradeagreements negotiated for international trade,commerce and regulatory affairs.

15Piotr Robouch – EFSA WS Sevilla 10/02/2009 piotr robouch@ec europa eu 15Piotr Robouch EFSA WS, Sevilla, 10/02/2009 [email protected]

Where to find the PT you need?

The European Proficiency Testing Information System helps finding a suitable proficiency testing (PT) scheme in your region or worldwide. The database lists more than 1000 schemes, mainly in the fields of chemical and mechanical testing.

htt // ti b d / /i d hthttp://www.eptis.bam.de/en/index.htm

16Piotr Robouch – EFSA WS Sevilla 10/02/2009 piotr robouch@ec europa eu 16Piotr Robouch EFSA WS, Sevilla, 10/02/2009 [email protected]

Thank youThank you