Upload
others
View
23
Download
0
Embed Size (px)
Citation preview
CLINICAL APPLICATIONS LESSON PLAN
THE PURITAN BENNETT™ 980 VENTILATOR
OVERVIEW
This guide is provided as a convenience companion document to the Operator’s Manual. It is not intended to replace the Operator’s Manual, which should always be available while using the ventilator. It is important to familiarize yourself with all information in the Operator’s Manual relevant to your institution’s use of the ventilator, including on-screen help, instructions, warnings and cautions.
Puritan Bennett™ Leak Sync Software
Noninvasive Ventilation
Volume Control Plus
Pressure Control
Puritan Bennett™ BiLevel Software
Puritan Bennett™ PAV™ *+ Software
Tube Compensation Software
Volume Support
Puritan Bennett™ NeoMode 2.0 Software
Puritan Bennett™ Proximal Flow Sensor
CONTENTS
3 3 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
PURITAN BENNETT™ LEAK SYNC SOFTWARE
Patient interface
ETT (cuffed, deflated cuff or cuffless)
Trach tube (cuffed, deflated cuff or cuffless)
Mask or other external interface
Breathing circuit system
PROBLEM: WHERE DO LEAKS COME FROM?
5 5 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Auto-triggering
Late cycling-off of pressure support and volume support breaths
Overestimation of delivered tidal volume/flow
The possible need for frequent manual adjustments
PROBLEM: COMMON LEAK CHALLENGES
6 6 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Year Author Key Results
2010 Mahmoud1 An air leak of 40% indicated the displayed VT was underestimated by 1.2 mL/kg, thus by about 24% of target VT (generally 5 mL/kg).
2001 Main2 Leaks larger than 20% resulted in inconsistent tidal volume delivery and gross overestimation of compliance (C) and resistance (R) regardless of ventilator mode.
1. Mahmoud RA, Proquitté H, Fawzy N, Bührer C, Schmalisch G. Tracheal tube airleak in clinical practice and impact on tidal volume measurement in ventilated neonates. Pediatr Crit Care Med. 2011;12(2):197-202.
2. Main E, Castle R, Stocks J, James I, Hatch D. The influence of endotracheal tube leak on the assessment of respiratory function in ventilated children. Intensive Care Med. 2001;27(11):1788-1797.
Studies show that air leaks occur in up to 70% of children and can lead to significant clinical consequences, including1,2:
Inconsistent ventilatory delivery1
Unreliable lung function testing2
Inappropriate clinical assessment due to error in monitored volume1
Increased work of breathing due to trigger asynchronies2
PROBLEM: LEAKS
7 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Leak Sync software may reduce leak-related ventilation and monitoring problems* during both inspiration and exhalation.1
1. Oto J et al. A Comparison of Leak Compensation in Acute Care Ventilators During Noninvasive and Invasive Ventilation: A Lung Model Study Respir Care 2013;58(12):2027–2037 (Puritan Bennett™ 840 Leak comp software was in the study).
*Compared to when Leak Sync is off
SOLUTION: PURITAN BENNETT™ LEAK SYNC SOFTWARE
8 8 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Reduces auto-triggering and delayed cycling1
Leak compensates the data values for VTI (displayed as VTL) and VTE
Servo-controls breath delivery based on leak-compensated values
Works in VC+ and VS
SOLUTION: PURITAN BENNETT™ LEAK SYNC SOFTWARE
9 9 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
1. Oto J et al. A Comparison of Leak Compensation in Acute Care Ventilators During Noninvasive and Invasive Ventilation: A Lung Model Study Respir Care 2013;58(12):2027–2037 (Puritan Bennett™ 840 Leak comp software was in the study).
Name: “Leak Compensation” changed to “Leak Sync”
VC+ and VS added
Defaults to ON with NIV and neonatal
PURITAN BENNETT™ LEAK SYNC SOFTWARE UPDATES
10 10 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
May reduce manual adjustments
of sensitivity settings1
Improves triggering and cycling2
Leak compensates the data values for inspiratory and expiratory volumes to
make them more relevant to the clinical
situation1
Allows for VC+ and VS
Enables avoidance of nuisance alarms
May provide reassurance of appropriate
level of ventilatory support by servo-
controlling volume-targeted breath types
from a leak-compensated VTI (displayed at VTL) data value1
POTENTIAL BENEFITS OF PURITAN BENNETT™ LEAK SYNC SOFTWARE
11 11 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
1. Puritan BennettTM 980 Series Ventilator Operator’s Manual Leak Sync Addendum Page B-2
2. Oto J et al. A Comparison of Leak Compensation in Acute Care Ventilators During Noninvasive and Invasive Ventilation: A Lung Model Study Respir Care 2013;58(12):2027–2037 (Puritan Bennett™ 840 Leak comp software was in the study).
Menu tab/setup/more settings
Enabled or disabled
Enabled by default with neonatal patient type and also when NIV is selected
Set DSENS in L/min
PURITAN BENNETT™ LEAK SYNC SOFTWARE SETUP
12 12 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
LS on vent setup button
DSENS = L/min (not %)
VTL and VTE = estimated patient values
Graphic displays estimated lung flows
Data displayed:
VLEAK
%LEAK
LEAK (leak rate@PEEP)
DISPLAY CHANGES WITH LEAK SYNC SOFTWARE ENABLED
13 13 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
The Puritan Bennett™ 840 ventilator with Leak Compensation software outperformed all other ventilators included in the studies.1
Invasive ventilation The Puritan Bennett™ 840 ventilator outperformed all other
ventilators in this study by requiring fewer breaths to achieve
synchronization during increasing and decreasing leaks.1
Noninvasive ventilation
The study demonstrated wide variations between the ability
of ventilators to compensate for an air leak during noninvasive
ventilation.1
The Puritan Bennett™ 840 and Philips RespironicsTM* V60
ventilators were the only ventilators in this study that adapted
well to increasing or decreasing leaks.1
1. Oto J, Marchese A, Kacmarek M. A comparison of leak compensation in acute care ventilators during invasive ventilation: a lung model study. Respir Care. 2013;58:2027-2037.
PERFORMANCE
14 14 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Defaults on: NIV or Neonatal
Defaults off: INV/Pediatric or Adult
Adjusts within ~ 3 breaths
VTL and VTE compensated display
Flow/volume graphics leak compensated
Bias flow automatically adjusted
HOW PURITAN BENNETT™ LEAK SYNC SOFTWARE WORKS
15 15 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Capabilities
Neonate, pediatric and adult invasive and non-invasively
Dynamic compensation of leaks during invasive and noninvasive ventilation
Limitations
Leak Sync software not active with:
Tube compensation (TC)
Proportional Assist™* Ventilation (PAV™ *+ software)
CAPABILITIES AND LIMITATIONS
16 16 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
NONINVASIVE VENTILATION
Mechanical ventilation may lead to significant asynchrony.1
Leaks may impact patient tolerance to NIV.2
Leaks may result in monitoring errors, breath management issues, and trigger and cycle asynchrony.2
1. Mahmoud RA, Proquitté H, Fawzy N, Bührer C, Schmalisch G. Tracheal tube airleak in clinical practice and impact on tidal volume measurement in ventilated neonates. Pediatr Crit Care Med. 2011;12(2):197-202.
2. Vignaux L, Vargas F, Roeseler J, et al. Patient-ventilator asynchrony during non-invasive ventilation for acute respiratory failure: a multicenter study. Intensive Care Med. 2009;35(5):840-846.
PROBLEM: INVASIVE AND NONINVASIVE VENTILATION PRESENT CHALLENGES
18 18 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
NIV ventilation selection
Use with mask or other leak-prone patient interface
Works with various modes/breath types
Leak Sync software enabled
SOLUTION: NON-INVASIVE VENTILATION (NIV) WITH PURITAN BENNETT™ LEAK SYNC SOFTWARE
19 19 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Help patients tolerate NIV so they can continue using NIV
Help avoid complications associated with intubation/artificial airways1
1. Epstein S. Should a Patient Be Extubated and Placed on Noninvasive Ventilation After Failing a Spontaneous Breathing Trial? Respir Care 2010;55(2):198–206
POTENTIAL BENEFITS OF NIV WITH PURITAN BENNETT™ LEAK SYNC SOFTWARE
20 20 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Setup button
Menu tab/vent
NIV
Shows applicable settings
CPAP mode is available
Flow trigger
↑TI SPONT limit
DSENS defaults to OFF
NIV SETUP
21 21 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
NIV-compatible selection of mode and breath types (NEO CPAP added) is available
Volume alarms are disabled to reduce nuisance alarms
Leak Sync software automatically enabled
HOW NIV SOFTWARE WORKS
22 22 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Capabilities
Stable respiratory drive
Neonatal
Pediatric
Adult
Limitations
Not available with:
Tube compensation (TC)
BiLevel
VC+ and VS
Proportional Assist™* Ventilation Plus (PAV™ *+ software) breath types
Only flow triggering is available.
NIV CAPABILITIES AND LIMITATIONS
23 23 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
VOLUME CONTROL PLUS
PROBLEM: FIXED FLOW PATTERNS MAY RESULT IN DISCOMFORT AND ASYNCHRONY
25 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
VC+ breaths deliver a variable flow.
The ventilator manages inspiratory pressure to achieve the desired tidal volume target.
SOLUTION: VOLUME CONTROL PLUS (VC+)
26 26 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Unlike standard VC, VC+:
Provides variable flow delivery and volume
Uses the lowest pressure necessary for volume delivery
Escalates pressure if volume delivery decreases
Active valve accommodates breathing variations.
POTENTIAL BENEFITS OF VC+
27 27 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Select:
Gender and height or predicted body weight (PBW)
Invasive vent type
AC or SIMV mode
VC+ mandatory breath type
Adjust all available settings.
Set all appropriate alarms.
VC+ SETUP
28 28 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Variable flow with active exhalation valve
Titration of pressure
1 – 3 cmH2O/breath
Alarm strategy
↑PPEAK limit (determines max target pressure)
↓PPEAK limit
Volume not delivered
↑ VTI limit
HOW VC+ SOFTWARE WORKS AND ALARM STRATEGY
29 29 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Check that TI is long enough.
Check that Rise Time is fast enough.
Check that the set tidal volume is not too high in comparison with the ↑PPEAK limit.
Check for auto-PEEP causing ↑PPEAK limit and resolve.
TROUBLESHOOTING INADEQUATE VT DELIVERY
30 30 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Capabilities
Neonate, pediatric and adult
Leak Sync software compatible
Limitations
Invasive only
High effort may result in regulating of pressure down to undesirable levels.
Leaks without Leak Sync software enabled may result in regulating of pressure down to undesirable levels.
CAPABILITIES AND LIMITATIONS
31 31 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
PRESSURE CONTROL
Fixed flow patterns in VC may result in discomfort and asynchrony.
VC+ may not always provide a consistent delivered pressure.
TWO PROBLEMS
33 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
PC breaths deliver variable flow and volume.
The inspiratory pressure target remains constant.
SOLUTION: PRESSURE CONTROL (PC)
34 34 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Mean airway pressure more consistent while allowing more natural breathing.
POTENTIAL BENEFITS OF PC
35 35 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Enter the patient’s gender and height or predicted body weight (PBW).
Select INV or NIV.
Touch AC or SIMV mode.
Touch the PC mandatory breath type.
Adjust all available settings.
Set all appropriate alarms.
PC SETUP
36 36 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
The user determines whether to set Inspiratory Time, I:E Ratio or Expiratory Time as the primary setting.
Changes in the preset rate will not alter the primary variable set in the timing padlocked area.
HOW TO USE THE TIMING PADLOCKS
37 37 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Mandatory, assisted/controlled pressure-based breath
INV or NIV
The user sets inspiratory time, inspiratory pressure/PEEP and pressure rise.
The ventilator manages flow to meet the patient’s needs and control pressure delivery.
HOW PC SOFTWARE WORKS
38 38 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Capabilities
May be used with neonate, pediatric and adult patients.
May be combined with Leak Sync.
Limitations
Tidal volume is variable.
CAPABILITIES AND LIMITATIONS
39 39 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
PURITAN BENNETT™ BILEVEL SOFTWARE
The limitations of volume control may lead to additional sedation.1
Inverse ratios and monitoring are limitations in traditional PCV.
1. Pohlman M. Excessive tidal volume from breath stacking during lung-protective ventilation for acute lung injury. Crit Care Med. 2008;36:3019–3023.
TWO PROBLEMS
41 41 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Mandatory-assisted and spontaneous-assisted pressure-based breath types
Active valve
Strategies for boosting mean airway pressure
Normal I:E or extended inverse available (>4:1)
SOLUTION: PURITAN BENNETT™ BILEVEL SOFTWARE
42 42 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Flow and volume variable
Active valve
Conventional or APRV style of breath timing
Potential for improved oxygenation1
Potential for reduction in sedation1
1. Putensen C. et al. Long-Term Effects of Spontaneous Breathing During Ventilatory Support in Patients with Acute Lung Injury. Am J Respir Crit Care Med Vol 164. pp 43–49, 2001
POTENTIAL CLINICAL BENEFITS OF PURITAN BENNETT™ BILEVEL SOFTWARE
43 43 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Enter gender and height or predicted body weight (PBW).
Touch Invasive.
Touch BiLevel mode.
Adjust available settings. Note:
PH: Above ambient, not above PEEP; must be set at least 5 above PL
Spontaneous Type: TC or PS
Set all appropriate alarms.
PURITAN BENNETT™ BILEVEL SOFTWARE SETUP
44 44 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
User determines the primary setting. Options:
TH (Time High)
TH ratio to TL (Time High to Time Low ratio)
TL (Time Low)
Changes in set frequency will not alter the primary variable set in the timing padlocked area.
PURITAN BENNETT™ BILEVEL SOFTWARE SETUP: TIMING PADLOCK TO LOCK ‘RELEASE TIME’
45 45 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
PH: Pressure during the TH period
PL: Pressure during the TL period
When TH:TL > 4:1, with the TL period locked, changes in set f will change the TH, but the TL will remain constant regardless of patient effort period to accommodate the new f setting while maintaining the set TL period.
PURITAN BENNETT™ BILEVEL SOFTWARE SETUP
46 46 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Two pressure levels, breath availability and active valve
Conventional or APRV breath timing
Alarm strategy
PPEAK (peak pressure) may be 1 to 2 cmH2O higher than set pressure
HOW PURITAN BENNETT™ BILEVEL SOFTWARE WORKS
47 47 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Pressure support
Works on PL and PH
PL + PS = target pressure
May be left at “0” cmH20
Tube Compensation
PURITAN BENNETT™ BILEVEL SOFTWARE: SPONTANEOUS SUPPORT OPTIONS
48 48 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Capabilities
Neonate, pediatric, adult
APRV inverse ratio settings
Support Spont breaths with PS, TC
Limitations
Only invasive
Limited respiratory mechanics
CAPABILITIES AND LIMITATIONS
49 49 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
PH – High Pressure
TH – High Pressure Time
Habashi NM. Other approaches to open-lung ventilation: airway pressure release ventilation. Crit Care Med. 2005;33(3 Suppl):S228-40.
DESCRIBE CHANGES THAT MAY IMPACT OXYGENATION
50 50 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
PURITAN BENNETT™ PAV™*+ SOFTWARE
Levine et al. Rapid disuse atrophy of diaphragm fibers in mechanically ventilated humans. NEJM. 2008;358:1327-1335.
PROBLEM: WHEN THE VENTILATOR CONTROLS BREATHS INSTEAD OF THE PATIENT, ATROPHY OF DIAPHRAGM FIBERS HAPPENS RAPIDLY DUE TO DISUSE
52 52 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
24% of mechanically ventilated patients exhibit patient-ventilator asynchrony in >10% of their respiratory efforts.
Thille A. Patient-ventilator asynchrony during assisted mechanical ventilation. Intensive Care Med. 2006;32:1512.
PROBLEM: PATIENT-VENTILATOR ASYNCHRONY IS PREVALENT
53 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Asynchrony
Sedation
Prolonged ventilation time1
Possible muscle atrophy2 and VAP3
Weaning is delayed
1. Kollef M et al. The use of continuous intravenous sedation is associated with prolongation of mechanical ventilation. Chest. 1998;114:541–548.
2. Levine S et al. Rapid Disuse Atrophy of Diaphragm Fibers in Mechanically Ventilated Humans. NEJM. 2008;358:1327- 1335.
3. Rello J et al. Epidemiology and outcomes of ventilator-associated pneumonia in a large US database. Chest. 2002;122:2115-2121.
PROBLEM: ASYNCHRONY MAY LEAD TO THE USE OF SEDATION
54 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
24% of ICU patients exhibit asynchrony in greater than 10% of their breaths.1
Asynchrony is associated with 18 extra days on the ventilator: duration of mechanical ventilation 7 (3-20) days vs. 25 (9-42) days p=0.005 ventilation (days; IQR).1
Average cost per day in the ICU for a patient receiving mechanical ventilation is $4,000.2
1. Thille A et al. Patient-ventilator asynchrony during assisted mechanical ventilation. Intensive Care Med. 2006;32:1512.
2. Dasta J et al. Daily cost of an intensive care unit day: the contribution of mechanical ventilation. Crit Care Med. 2005;33:1266-1271.
24% [ ]
18 days [ ]
$72,000 [ ]
PROBLEM: ASSOCIATED ASYNCHRONY COSTS
55 55 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Monitors patient’s demand breath by breath
Monitors compliance and resistance
Allows variable volume, flow, inspiratory time and pressure
Promotes natural breathing*
*Compared to VC, VC+ and PS
SOLUTION: PURITAN BENNETT™ PAV™*+ SOFTWARE
56 56 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Encourages use of the diaphragm1
Increases support as the patient’s demand increases
Improves patient-ventilator synchrony1,2
May help the clinician better manage a patient’s work of breathing3
1. Xirouchaki N. et al. Proportional assist ventilation with load-adjustable gain factors in critically ill patients: comparison with pressure support. Int Care Med. 2008;34(11):2026-2034.
2. Costa R. A physiologic comparison of proportional assist ventilation with load-adjustable gain factors (PAV+) versus pressure support ventilation (PSV) Intensive Care Med (2011) 37:1494–1500
3. Eumorfia Kondili, George Prinianakis, Christina Alexopoulou, Eleftheria Vakouti, Maria Klimathianaki, Dimitris Georgopoulos Respiratory Load compensation during mechanical ventilation-proportional assist ventilation with load-adjustable gain factors versus pressure support Intensive Care Med DOI 10.1007/s00134-006-0110-0
POTENTIAL BENEFITS OF PURITAN BENNETT™ PAV™*+ SOFTWARE
57 57 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Enter gender and height or predicted body weight (≥25 kg).
Touch:
Invasive vent type
SPONT mode
PAV™*+
Select % Support, PEEP and FiO2.
Select tube type.
Adjust the tube ID (6-10 mm ID).
Adjust the alarms.
PURITAN BENNETT™ PAV™*+ SOFTWARE SETUP
58 58 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
↑PPEAK
↑VTI limit
↓VE or ↓VT
PURITAN BENNETT™ PAV™*+ SOFTWARE KEY ALARMS
59 59 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
PA to PAV™*+ description
Default support starts at 70% Support
Two graphs and WOB bar
Enhanced calculation of compliance and resistance
PURITAN BENNETT™ PAV™*+ SOFTWARE CHANGES
60 60 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Measures values that create work
Assesses compliance and resistance (4 - 10 breaths with an automatic plateau)
Measures flow and volume every 5 msec
Uses the “equation of motion”
HOW PURITAN BENNETT™ PAV™*+ SOFTWARE WORKS: MEASUREMENTS
61 61 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
% Support, PEEP and O2%
Tube size and type
Monitors:
Compliance, resistance and PEEPI
Work of breathing bar
WOBTOT and WOBPT
Elastic and resistance work
Estimated lung pressure shadow trace
Flow and volume waveforms
HOW PURITAN BENNETT™ PAV™*+ SOFTWARE WORKS: SETTINGS AND MONITORED PARAMETERS
62 62 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
High inspired tidal volume (↑VTI)
High circuit pressure (↑PPEAK)
PAV STARTUP TOO LONG
PAV R&C NOT ASSESSED
Set all other traditional alarms
PURITAN BENNETT™ PAV™*+ SOFTWARE ALARM STRATEGY
63 63 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
PAV™*+ software measures work variables.
The primary settings are % Support, PEEP and O2%.
The software amplifies patient efforts to offset the total work of breathing.
The software adjusts delivered pressure within the same breath.
The breath stops when patient inspiratory flow ceases.
Delivered pressure adjusts dynamically to patient effort.
SUMMARY OF HOW PURITAN BENNETT™ PAV™*+ SOFTWARE WORKS
64 64 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Capabilities
Intact respiratory drive
25 kg (55 lb.) or greater
6.0 I.D. tube or greater
Limitations
Apnea
Oversedation
Abnormal respiratory drive
Leaks
<25 kg (55 lb.)
Severe air trapping
Pneumatic nebulizer
Not available with NIV
CAPABILITIES AND LIMITATIONS
65 65 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
TUBE COMPENSATION
An endotracheal or tracheostomy tube can impose an increased work of breathing.1
Pressure Support limitations:
Fixed delivery pressure approach
T-Piece trials reduce the monitoring capabilities.
1. Maeda Y. Does the tube-compensation function of two modern mechanical ventilators provide effective work of breathing relief? Critical Care 2003, 7:R92-R97
PROBLEM: IMPOSED WOB FROM A TRACH OR AN ET TUBE
67 67 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Tube Compensation (TC) assists the patient in overcoming the flow resistance of the artificial airway.
TC software supports spontaneous breaths.
Only to the point of overcoming the artificial airways
TC may reduce the work of breathing.1
1. Maeda Y. Does the tube-compensation function of two modern mechanical ventilators provide effective work of breathing relief? Critical Care 2003, 7:R92-R97
SOLUTION: TUBE COMPENSATION (TC)
68 68 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Provide variable pressure with variable effort.
Reduce the imposed work associated with artificial airways.1
Allow for enhanced monitoring compared to the T-Piece approach.
1. Maeda Y. Does the tube-compensation function of two modern mechanical ventilators provide effective work of breathing relief? Critical Care 2003, 7:R92-R97
POTENTIAL BENEFITS OF TUBE COMPENSATION
69 69 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Select TC for spontaneous type.
In particular, mention:
% Support: 10% to 100%
Tube I.D.
Tube type
↑PPEAK
↑VTI
TUBE COMPENSATION SETUP
70 70 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Supports spontaneous breathing in SPONT, BiLevel and SIMV.
TC checks the flow rate every 5 ms.
TC uses flow, the % Support setting and a look-up table to determine delivered pressure.
% Support settings are from 10% to 100% (in 5% increments).
HOW TUBE COMPENSATION SOFTWARE WORKS
71 71 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
↑PPEAK
↑Inspired tidal volume
All of typical alarms
TC-SPECIFIC ALARM STRATEGY
72 72 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Capabilities
Intact respiratory drive
7.0 kg (15.4 lb.) or greater
4.5 I.D. tube or bigger
Limitations
Oversedation
NeoMode <7.0 kg (15.4 lb.)
<4.5 I.D. tube
Apnea
CAPABILITIES AND LIMITATIONS
73 73 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
VOLUME SUPPORT
Pressure support does not provide a target tidal volume.
Pressure support does not down-regulate pressure automatically.
PROBLEM: LIMITATIONS OF PRESSURE SUPPORT
75 75 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Clinician can set a target tidal volume.
Automated pressure delivery is based on inspired volume.
VS provides variable flow.
SOLUTION: VOLUME SUPPORT (VS)
76 76 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Encourages the use of the diaphragm
Automatically decreases support pressure if target tidal volume is exceeded (1 – 3 cmH2O)
Increases pressure if the patient’s ability to take an adequate breath is compromised (1 – 3 cmH2O)
1. Sassoon C. Assist–Control Mechanical Ventilation Attenuates Ventilator-induced Diaphragmatic Dysfunction. Am J Respir Crit Care Med Vol 170. pp 626–632, 2004
POTENTIAL BENEFITS OF VOLUME SUPPORT
77 77 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Feature Volume Support Pressure Support
Lowest pressure for the set volume Yes No
Escalates if needed Yes No
Adjusts for changes in Raw and C Yes No
COMPARING VOLUME SUPPORT TO PRESSURE SUPPORT
78 78 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Select SPONT mode
Select VS for spontaneous type
Select:
VT SUPP
Trigger setting
Oxygen %
Rise %
ESENS
PEEP
↑VTI
↑PPEAK
VOLUME SUPPORT SETUP
79 79 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
VS varies the inspiratory pressure to deliver the operator-set target tidal volume.
During VS, the inspiratory pressure target cannot be lower than PEEP + 1.5 cmH2O and cannot exceed the ↑PPEAK limit - 3 cmH2O.
Pressure adjust a max of +/- 3 cmH2O/breath if needed.
Test breath at startup. “VS startup” is displayed in the GUI’s prompt area.
HOW VOLUME SUPPORT SOFTWARE WORKS
80 80 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
↑PPEAK limit (determines max target pressure)
Volume not delivered
↑VTI limit
VS-SPECIFIC ALARM STRATEGY
81 81 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Capabilities
Neonatal to adult
Spontaneously breathing patients
Invasive ventilation only
SPONT mode only
Limitations
Oversedation
Apnea
High effort may result in regulating of pressure down to undesirable levels.
Leaks without Leak Sync software enabled may result in regulating of pressure down.
NIV
CAPABILITIES AND LIMITATIONS
82 82 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
PURITAN BENNETT™ NEOMODE 2.0 SOFTWARE
Small tidal volume requirements
Ability to use AC and SIMV to ventilate with NIV
Auto-triggering and delayed cycling1
Management of nuisance alarms with CPAP
Reducing excessive oxygen exposure
Excessive tidal volume delivery
1. Oto J et al. A Comparison of Leak Compensation in Acute Care Ventilators During Noninvasive and Invasive Ventilation: A Lung Model Study Respir Care 013;58(12):2027–2037.
GENERAL PROBLEM
84 84 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Puritan Bennett™ Leak Sync software
A wide spectrum of modes both INV and NIV
Volume targeted availability down to 2.0 mL
Assist Control (VC, VC+ and PC)
SIMV (VC, VC+ - Dual Mode and PC)
BiLevel software
Spontaneous
Volume Support – Dual Mode
Pressure Support
Assist Control (VC, PC)
SIMV (VC, PC)
Spontaneous (PS)
CPAP
POTENTIAL SOLUTION: PURITAN BENNETT™ NEOMODE 2.0 SOFTWARE
85 85 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
A NIV CPAP-specific mode with nuisance alarm management
A configurable temporary elevate oxygen feature
Enhanced monitoring of C20/C and mL/kg
POTENTIAL SOLUTION: PURITAN BENNETT™ NEOMODE 2.0 SOFTWARE
86 86 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
One machine for all patient sizes
Possible reduced use of invasive ventilation
Automatic management of leaks
Reduced alarms in the NICU with NCPAP
Better management of oxygen exposure
Intended delivered volume managed by monitoring
C20/C
PBW/kg
POTENTIAL SOLUTION/BENEFITS OF PURITAN BENNETT™ NEOMODE 2.0 SOFTWARE
87 87 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Install the neonatal adapter door and the single patient use neonatal expiratory filter.
Connect a neonatal breathing circuit to the ventilator.
Select the neonatal breathing circuit type in Short Self-Test (SST).
Neonatal-specific settings and safety boundaries are activated.
PURITAN BENNETT™ NEOMODE 2.0 SOFTWARE SETUP
88 88 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Select actual patient weight
Range for neonates is 0.3 kg to 7.0 kg (0.66 lb. to 15 lb.)
Select INV or NIV and modes
Includes A/C, SIMV, Spont, BiLevel and CPAP
Breath type VC, VC+ or PC
All other specific settings
Configurable elevated FiO2
Leak Sync software automatically enabled
PURITAN BENNETT™ NEOMODE 2.0 SOFTWARE SETUP
89 89 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
When in NeoMode 2.0 mode, if NIV CPAP is used:
VE TOT, VTE SPONT and ↑VTI alarms are not available.
VE and VTE displays are not available.
TA can be adjusted but defaults to off to avoid inadvertent alarms.
Adjust the DSENS alarm.
HOW NEOMODE 2.0 SOFTWARE WORKS: ALARM STRATEGIES IN CPAP
90 90 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Indications:
0.3 kg (0.66 lb.) and above
VT 2 mL and above
Neonatal breathing circuit
Neonatal adapter door and neonatal expiratory filter
Limitations:
Pediatric or adult breathing circuit
>7.0 kg (15 lb.) PBW
CAPABILITIES AND LIMITATIONS OF PURITAN BENNETT™ NEOMODE 2.0 SOFTWARE
91 91 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
PURITAN BENNETT™ PROXIMAL FLOW SENSOR
Circuit volume affecting graphs
Calibrations during ventilation
Water contamination of Proximal Flow Sensor
PROBLEMS: ACCURATE GRAPHICS AND PROXIMAL FLOW SENSOR ISSUES
93 93 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Measures flow and volume reflected at the patient wye
Displays both proximal flow and proximal pressure
Uses proximal flow values in the display of graphics.
Requires no manual calibration
Weight 6.6 g; dead space <1 mL
SOLUTION: PURITAN BENNETT™ PROXIMAL FLOW SENSOR AND MONITORING SOFTWARE
94 94 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Improved graphics for patient management
Improved patient data
POTENTIAL BENEFITS OF THE PURITAN BENNETT™ PROXIMAL FLOW SENSOR
95 95 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Proximal Flow Hardware
Proximal Flow Sensor
PURITAN BENNETT™ PROXIMAL FLOW OPTION COMPONENTS
96 96 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
To disable or enable the Proximal Flow Option:
1. Touch the Configure wrench icon
2. Touch the Options tab
3. Touch the Prox tab
4. Touch the Enabled or Disabled button
5. Touch Accept
Periodic expiratory phase purging occurs regularly for accurate measurements.
PURITAN BENNETT™ PROXIMAL FLOW SENSOR SETUP: ENABLING/DISABLING THE SENSOR
97 97 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Installed during the SST (prompts are given)
Two connections:
Install monitoring end at the patient circuit wye
Install connector end on the ventilator’s front panel behind a clear door
Neonatal invasive ventilation only
PROXIMAL FLOW SENSOR SETUP WITH PURITAN BENNETT™ NEOMODE 2.0 SOFTWARE
98 98 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Alarms respond to values monitored with the flow sensor
Breaths are controlled from the internal flow sensors (prox flow values are for monitoring only)
Prox inoperative
PURITAN BENNETT™ PROXIMAL FLOW SENSOR SETUP: ALARM STRATEGY
99 99 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Measures flows, pressures and tidal volumes
Requires NeoMode 2.0 software option
Purge and auto-zero
Waveform Monitoring: PCIRCY
VY
VTY
Monitored data: VTIY
VTLY
VTEY
VTE SPONTY
VTE MANDY
VETOTY
LEAKY
HOW THE PURITAN BENNETT™ NEONATAL PROXIMAL FLOW SENSOR WORKS
100 100 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Patient + Circuit Patient
COMPARING THE VALUES FOR LEAK AND LEAKY
101 101 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
Capabilities
Measuring flows, pressures and tidal volumes of invasively ventilated neonatal patients
PBW of 0.3 kg (0.66 lb) to 7.0 kg (15.4 lb) using ET tube sizes from 2.5 mm to 4.0 mm
Limitations
Non-invasive (NIV)
Pediatric and adult
ET >4.0 mm
PBW >7.0 kg (15.4 lb)
Nebulized medications
CAPABILITIES AND LIMITATIONS
102 102 Puritan BennettTM 980 Ventilator Clinical Applications Lesson Plan | June 1, 2015
© 2016 Medtronic. All rights reserved. Medtronic, Medtronic logo and Further, Together are trademarks of Medtronic. Proportional Assist and PAV are registered trademarks of The University of Manitoba, Canada. Used under license. TM* Third party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company.