840 VentilatorVolume Ventilation Plus Software Option

Page 2: Puritan Bennett™ 840 Ventilator Volume Ventilation Plus Software

2

Healthcare providers and patients throughout the world depend on Covidien for state-of-the-art ventilation therapy. Whether your needs include acute care for critically ill patients with chronic respiratory failure or a solution to transition patients to home care, we have the right system for the task at hand.

Healthcare professionals know all too well the range of issues that impact ventilation outcomes today. At Covidien, our innovations are systematically tackling the issues that truly matter—patient safety, medical efficacy and healthcare efficiency. The Volume Ventilation Plus software option is another example of how we’re helping medical professionals improve ventilation outcomes and quality of life for their patients.

A Breakthrough in Ventilatory Carefor Patients and Clinicians

Page 3: Puritan Bennett™ 840 Ventilator Volume Ventilation Plus Software

Three DimenSiOnS Of exCellenCe BuilT inTO The VOlume VenTilATiOn PluS SySTem

• Patient Safety—Three specific alarms come with the VC+ and VS options that alert the clinician of changes in patient condition and provide lung protection.

• Clinician Support—The VV+ option offers the patient enhanced comfort and synchrony (clinician support) while offering clinicians a way to maintain targeted tidal volume.

• Healthcare Efficiency—The VV+ system expands the therapeutic range options for clinicians. It is also effective in weaning patients from anesthesia, resulting in less time and money spent on postoperative recovery.

A Brief OVerView Of VOlume VenTilATiOn

Standard volume ventilation has been associated with increased work of breathing, flow starvation and preven-tion of spontaneous breathing or cough-ing during the inspiratory phase.1, 2 If higher sedation is used to treat ventila-tor asynchrony, the increased sedation or paralysis could lead to increased time on the ventilator.3-7 In contrast, recent research has indicated that limiting damage believed to be caused by over distention of the lungs in acute respira-tory distress syndrome (ARDS) may play an important role in improving survival. Lower tidal volume and lower lung pressure have been associated with improved survival.8-10 Whether volume or pressure strategies are best for achiev-ing this lung protection goal remains undetermined.

This clinical paper describes two new breath delivery strategies with sig-nificant improvements in spontaneous breathing and preset volume delivery.

3

Page 4: Puritan Bennett™ 840 Ventilator Volume Ventilation Plus Software

PreSSure-BASeD AnD VOlume-BASeD BreAThS

The phrase “dual modes” of ventilation has caused some confusion regarding when to employ one kind of breath type or another kind of breath type. Dual modes are not actually modes but, rather, breath types. A breath type is defined as the means by which either a mandatory or spontaneous breath is de-livered (i.e., Pressure Support, Pressure Control, Volume Control).

A Pressure breath type (whether man-datory or spontaneous) is determined by presetting the delivered peak pres-sure, while allowing flow and or tidal volume to vary.

A VOlume BreATh TyPe

A Volume breath type is determined by presetting a tidal volume, allowing pressure to rise and fall as lung imped-ance changes. Volume breath goals include maintaining consistent ventila-tion, maintaining low volumes for lung protection and eliminating CO2.

VOlume VenTilATiOn PluS

Volume Ventilation Plus (VV+) is an option that combines two different dual mode volume-targeted breath types–Volume Control Plus (VC+) for delivery of mandatory breaths in A/C and SIMV, and Volume Support (VS) for delivery of spontaneous breaths in SPONT only.

PreSSure AnD VOlume BreATh TyPeS

PreSSure COnSTAnT VOlume COnSTAnT

PC(In A/C or SIMV)

Volume Ventilation(in CMV or SIMV)

PS(In SIMV for Spont)

Mandatory†VC+/PRVC/AutoFlow

VAPS/Pres Au

BiPAP Bi-Level(Conv I:E or APRV)

Spontaneous †VS

†Dual Modes

Tube Compensation (TC) or Proportional Assist™* Ventilation (PAV™*)

4

Page 5: Puritan Bennett™ 840 Ventilator Volume Ventilation Plus Software

VOlume COnTrOl PluS

Volume Control Plus (VC+) utilizes a clinician-set inspiratory time and clini-cian-set target tidal volume. The ventila-tor initially delivers a single standard volume test breath with a decelerating flow pattern and plateau to determine the relative lung compliance. If the deliv-ered tidal volume is either greater or less than the preset value, the target pres-sures for subsequent breaths are adjusted to correct any discrepancies.

The volume-targeted mandatory breath delivered in VC+ is designed to address problems sometimes encountered with standard volume ventilation. When vol-ume ventilation is desired, VC+ can help assure a high level of synchrony.

VC+ CliniCAl ADVAnTAgeS

Common uses of VC+: • The clinician sets only the target volume

and inspiratory time. The flow rate is then delivered in a descending ramp pattern, with a maximized initial inspi-ratory flow. VC+ reduces the potential for inappropriately low flow settings and inadvertent flow starvation.1-13

• A pressure control style of breath allows active spontaneous breathing during the inspiratory phase of the mandatory breath. Breathing, cough-ing or splinting is then allowed by venting excess pressure.

Note: The software program for VV+ includes the VC+ and VS options (part #4-078126-00).

Volume Test Breathdetermines PC target

P Pt t

t tf f

Pressure target is adjusted to maintain target tidal volume

5

Page 6: Puritan Bennett™ 840 Ventilator Volume Ventilation Plus Software

SPOnTAneOuS BreAThing During VC+

• Spontaneous breathing is allowed during the inspiratory phase through the use of the active exhalation valve.

• Flow is determined automatically, thereby reducing the potential for inap-propriately low flow rates.

• A higher or lower initial flow rate is allowed, increasing the potential for patient synchrony during splinting or aggressive flow demands.

VC+ OPerATing feATureS

The Puritan Bennett™ 840 Ventilator’s VC+ option allows easy substitution for standard volume ventilation breaths in A/C or SIMV modes, since VC+ is a volume breath type.

Operating features:• The ventilator quickly attains target

tidal volume through its short series of test breaths. Intelligent recognition of adequate plateau helps protect against incorrect target values.

• Pressure is incrementally adjusted during start up. Pressure adjustment minimizes risk of large pressure and tidal volume swings caused by sudden and momentary changes in compliance.

• Tidal volume limits can be set to termi-nate breaths, should the tidal volume rise above desired levels before subse-quent pressure targets are automati-cally lowered to deliver breath at preset volumes.

• Any disconnect is recognized and alarmed specifically as a circuit discon-nect. Reconnection results in immediate resumption of previous target pressure and volume levels, assur-ing immediate resumption of mean airway pressure.

Spontaneous efforts

Spontaneous efforts

Peak cm h2O

inSP V lpm

exP

Target Volume(500ml) VT (400ml) VT (500ml)

P target

6

Page 7: Puritan Bennett™ 840 Ventilator Volume Ventilation Plus Software

VOlume SuPPOrT

Volume Support (VS) utilizes a breath delivery control algorithm similar to the one used in the VC+ option. However, VS utilizes pressure support, instead of pressure control, to adjust flow. The clinician still sets the target volume but does not set the inspiratory time or ven-tilator rate. The ventilator then delivers a spontaneous pressure support style of breath and varies the pressure up or down to guarantee the preset tidal vol-ume. As the patient assumes more of the work of breathing, the ventilator senses increased tidal volumes and decreases pressure. Should the tidal volume de-crease, the ventilator will automatically increase VS to protect the patient from hypoventilation.

Patient triggering determines the venti-latory rate; patient demand determines the inspiratory time.

VS CliniCAl ADVAnTAgeS

The VS option is a spontaneous breath type utilizing a varying pressure support strategy to maintain a target tidal volume.

Common uses for VS:Weaning from anesthesia. Clinicians set a target level of volume. As patients breathe above or below the preset volume, the support is increased or decreased to restore set volume. When patients begin to awaken and take larger and more frequent spontaneous breaths, the ventilator decreases its VS. How-ever, if the patient’s level of conscious-ness and respiratory drive decrease, VS increases. The VS option includes an apnea ventilation safety back up, should the patient cease breathing altogether.

The problems associated with target-ing minute volume arise when patients satisfy the minute volume criteria by breathing rapidly and shallowly. VS focuses on tidal volume, instead of minute volume.

Control tidal volume and increase pa-tient comfort. Although the literature about the efficacy of VS as a primary mode is scant, the VS option has also been used successfully as a primary mode of ventilation, especially in infants and pediatric patients. The clinician sets the patient’s tidal volume. The ventilator will limit pressure without terminat-ing breaths, if pressure rises to within 5 cm H2O of the upper pressure limit. As compliance improves and the pressure falls, the patient can be ventilated with a minimal tidal volume at the lowest pos-sible pressure. The higher setting allows patients more control over the support they are receiving (control of inspiratory time, control of respiratory rate and con-trol of flow rate).

Two different breath types for preset targeted tidal volumes and spontaneous breathing

7

Page 8: Puritan Bennett™ 840 Ventilator Volume Ventilation Plus Software

1. Alonso JA, Kallet R, Siobal M, Kraemer RH, Marks JD. Does autoflow optimize inspiratory flow (Vi)? A lung model study. Critical Care Med. 1999;27(1):93A.

3. Leatherman JW, Fluegel WL, David WS, Davies SF, Iber C. Muscle weakness in mechanically ventilated patients with severe asthma. Am J Respir Crit Care Med. 1996;153(5):1686-1690.

4. Rudis MI, Guslits BJ, Peterson EL, et al. Economic impact of prolonged motor weakness complicating neuromuscular blockade in the intensive care unit. Crit Care Med. 1996;(10):1749-1756.

5. Kollef M, Levy N, Ahrens T, Schaif R. Prentice D, Sherman D. The use of continuous I.V. sedation is associated with prolongation of mechanical ventilation. Chest 1998; 114(2):541-8.

6. Behbehani NA, Al-Mane F, D’yachkova Y, Paré P, FitzGerald JM. Myopathy following mechanical ventilation for acute severe asthma: the role of muscle relaxants and corticosteroids. Chest. 1999;115(6):1627-1631.

7. Brook AD, Ahrens TS, Schaiff R, et al. Effect of a nursing-implemented sedation protocol on the duration of mechanical ventilation. Crit Care Med. 1999;27(12):2609-2615.

8. Amato MB, Barbas CS, Medeiros DM, et al. Effect of a protective-ventilation strategy on mortality in the acute respiratory distress syndrome. N Engl J Med. 1998;5;338(6):347-354.

9. Artigas A, Bernard GR, Carlet J, et al. The American-European Consensus Conference on ARDS, part 2. Ventilatory, pharmacologic, supportive therapy, study design strategies and issues related to recovery and remodeling. Intensive Care Med. 1998;24(4):378-398.

10. Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. The Acute Respiratory Distress Syndrome Network. N Engl J Med. 2000;4;342(18):1301-1308.

11. Amato MB, Barbas CS, Bonassa J, Saldiva PH, Zin WA, de Carvalho CR. Volume-assured pressure support ventilation (VAPSV). A new approach for reducing muscle workload during acute respiratory failure. Chest. 1992;102(4):1225-1234.

12. MacIntyre N, Gropper C, Westfall T. Combining pressure-limiting and volume-cycling features in a patient-interactive mechanical breath. Critical Care Med. 1994; 22:353-357.

13. Haas CD, Branson RD, Folk LM, et al. Patient-determined inspiratory flow during assisted mechanical ventilation. Respir Care. 1995; 40(7):716-721.

COVIDIEN, COVIDIEN with logo, Covidien logo and positive results for life are U.S. and internationally registered trademarks of Covidien AG. Other brands are trademarks of a Covidien company. Proportional Assist and PAV are registered trademarks of

The University of Manitoba, Canada. Used under license. ©2009, 2010 Covidien. All rights reserved.

10-VE-0204 VE09902

6135 Gunbarrel AvenueBoulder, CO 80301

800-635-5267 www.covidien.com

VOlume SuPPOrT AnD PreSSure SuPPOrT

The VS option has advantages and limitations compared to Pressure Sup-port (PS). The VS option may help keep patients in less danger of over- or under-ventilation, unless the patient’s respira-tory drive and effort exceed the set tidal volume. PS will not decrease support, even if patient demand increases. The choice of VS or PS depends on the insti-tution’s protocols, the disease entity, the acuity of the patient being treated and the clinician’s therapeutic goals.

SummAry

New flexibility in breath delivery can increase comfort for patients breathing spontaneously on mechanical ventila-tion. The Puritan Bennett™ 840 Ven-tilator, with its VC+ and VS options, expands the therapeutic range for clini-cians and patients alike. The VC+ option offers clinicians the opportunity to take advantage of these new capabilities while maintaining control of tidal volume. VS may be useful in improving patient com-fort and/or in weaning from anesthesia. Both options represent a breakthrough in ventilatory care of patients.