ANALYTICAL BIOCOMPATIBILITY BIOSKILLS INTERVENTIONAL MED DEVICE PATHOLOGY • PHARMACOLOGY REGULATORY SURGICAL MED DEVICE TOXICOLOGY Thrombogenicity Testing for Blood-Contacting Medical Devices in an in vitro Human Blood-Loop Yan Chen, PhD American Preclinical Services [email protected] 7/11/2018 APS Confidential Information Advances in Cardiovascular Medical Devices

Thrombogenicity Testing for Blood-Contacting Medical ... · ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY

  • Upload
    others

  • View
    2

  • Download
    0

Embed Size (px)

Citation preview

Page 1: Thrombogenicity Testing for Blood-Contacting Medical ... · ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY

ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY • SURGICAL MED DEVICE • TOXICOLOGY

Thrombogenicity Testing for Blood-Contacting

Medical Devices in an in vitro Human Blood-Loop

Yan Chen, PhD

American Preclinical Services

[email protected]

7/11/2018 APS Confidential Information

Advances in Cardiovascular Medical Devices

Page 2: Thrombogenicity Testing for Blood-Contacting Medical ... · ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY

ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY • SURGICAL MED DEVICE • TOXICOLOGY

Thrombogenicity TestISO 10993-4 thrombogenicity testing is widely used for meeting regulatory requirements for approval of blood-contacting medical devices.

– Thrombosis deposition

– Thromboembolism

7/11/2018 APS Confidential Information

Page 3: Thrombogenicity Testing for Blood-Contacting Medical ... · ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY

ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY • SURGICAL MED DEVICE • TOXICOLOGY

In vivo Thrombogenicity • In vivo implant in clinical relevant location

• In vivo thrombogenicity study– NAVI (Non-anticoagulated venous implant)

– AVI (Anticoagulated venous implant)

– NAAI (Non-Anticoagulated arterial implant)

– AAI (Anticoagulated arterial implant)

7/11/2018 APS Confidential Information

Page 4: Thrombogenicity Testing for Blood-Contacting Medical ... · ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY

ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY • SURGICAL MED DEVICE • TOXICOLOGY

Limitations of in vivo Model

• Acknowledged by only a few regulatory bodies

• Controversy and caveats with the methodology

– The implant position

– The implant technique

– The extent of device-vessel wall contact (Tissue Factor)

– The explant technique

– Failure of LMCD

– Hydrophilic surface

– Statistical power

7/11/2018 APS Confidential Information

1Wolf, M. F., and Anderson, J. M., 2012, “Practical Approach to Blood Compatibility Assessments: General Considerations and Standards,” Biocompatibility and Performance of Medical Devices, J.-P. Boutrand, ed., Woodhead Publishing, Oxford, UK.

Page 5: Thrombogenicity Testing for Blood-Contacting Medical ... · ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY

ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY • SURGICAL MED DEVICE • TOXICOLOGY

In vitro Ovine Blood Loop

7/11/2018 APS Confidential Information

Journal of Medical Devices:

Special Issue on Cardiovascular Device

Development

Page 6: Thrombogenicity Testing for Blood-Contacting Medical ... · ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY

ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY • SURGICAL MED DEVICE • TOXICOLOGY

3RS Award for in vitro Alternative Assay

7/11/2018 APS Confidential Information

Page 7: Thrombogenicity Testing for Blood-Contacting Medical ... · ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY

ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY • SURGICAL MED DEVICE • TOXICOLOGY

In vitro Human Blood Loop

Enhanced potential for prediction of clinical risk

7/11/2018 APS Confidential Information

Page 8: Thrombogenicity Testing for Blood-Contacting Medical ... · ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY

ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY • SURGICAL MED DEVICE • TOXICOLOGY

In vitro Blood Loop Configuration

7/11/2018 APS Confidential Information

Page 9: Thrombogenicity Testing for Blood-Contacting Medical ... · ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY

ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY • SURGICAL MED DEVICE • TOXICOLOGY

Typical Assay Design

7/11/2018 APS Confidential Information

Page 10: Thrombogenicity Testing for Blood-Contacting Medical ... · ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY

ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY • SURGICAL MED DEVICE • TOXICOLOGY

Blood Collection/Qualification

• Drawn from healthy donors

• Final heparin 0.7 IU per mL

• Time from collection to loading of the loop (~3 hours)

• Baseline ACT (120 – 250 seconds)

• Minimal of three donors

7/11/2018 APS Confidential Information

Page 11: Thrombogenicity Testing for Blood-Contacting Medical ... · ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY

ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY • SURGICAL MED DEVICE • TOXICOLOGY

Temperature Control

7/11/2018 APS Confidential Information

Page 12: Thrombogenicity Testing for Blood-Contacting Medical ... · ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY

ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY • SURGICAL MED DEVICE • TOXICOLOGY

Study Assessment• Thrombus formation

– % surface area coverage

• Non-adherent thrombus– Weights of non-adherent thrombus

• Blood characterization – Complete blood counts (CBC)

– Activated clotting time (ACT)

– Platelet counts

7/11/2018 APS Confidential Information

Page 13: Thrombogenicity Testing for Blood-Contacting Medical ... · ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY

ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY • SURGICAL MED DEVICE • TOXICOLOGY

In situ Visual Observation

7/11/2018 APS Confidential Information

Page 14: Thrombogenicity Testing for Blood-Contacting Medical ... · ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY

ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY • SURGICAL MED DEVICE • TOXICOLOGY

Quantitative Thrombus Assessment

7/11/2018 APS Confidential Information

Insertion Site

= test device (Total Length = 10 cm)

= Thrombus formation

: Thrombus on segment: for 50% coverage

for 100% coverage

Cross section View - Example for Calculating Coverage

𝟑.𝟏𝒄𝒎 𝒙 𝟓𝟎% + 𝟏𝒄𝒎 𝒙 𝟏𝟎𝟎%

𝟏𝟎𝒄𝒎= 26% coverage

3.1cm 1.0 cm

Page 15: Thrombogenicity Testing for Blood-Contacting Medical ... · ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY

ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY • SURGICAL MED DEVICE • TOXICOLOGY

Positive and Negative Controls (N=13 studies)

7/11/2018 APS Confidential Information

Page 16: Thrombogenicity Testing for Blood-Contacting Medical ... · ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY

ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY • SURGICAL MED DEVICE • TOXICOLOGY

Non-Adherent Thrombus

7/11/2018 APS Confidential Information

Weights of recovered non-adherent thrombus fromseveral sets of assays were evaluated and in only oneinstance (a positive control loop from one individualassay) was any non-adherent thrombus visible. Inthis single instance, the observed weight ofthrombus was <10 mg.

Page 17: Thrombogenicity Testing for Blood-Contacting Medical ... · ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY

ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY • SURGICAL MED DEVICE • TOXICOLOGY

Complete Blood Count

7/11/2018 APS Confidential Information

Page 18: Thrombogenicity Testing for Blood-Contacting Medical ... · ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY

ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY • SURGICAL MED DEVICE • TOXICOLOGY

Blood ACT Values

7/11/2018 APS Confidential Information

B a s e lin e N e g a t iv e C o n tr o l P o s it iv e C o n tr o l

0

1 0 0

2 0 0

3 0 0

4 0 0

AC

T (

se

co

nd

s)

± s

tde

v

Page 19: Thrombogenicity Testing for Blood-Contacting Medical ... · ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY

ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY • SURGICAL MED DEVICE • TOXICOLOGY

Blood Platelet Counts

7/11/2018 APS Confidential Information

B a s e lin e N e g a tiv e C o n tro ls P o s it iv e C o n tro ls

0

2 0 0

4 0 0

6 0 0

Pla

tele

t C

ou

nt

(10

00

's p

er µ

L)

Page 20: Thrombogenicity Testing for Blood-Contacting Medical ... · ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY

ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY • SURGICAL MED DEVICE • TOXICOLOGY

In vivo – in vitro Comparison• The implant position

– Uniform and consistent blood flow, Controlled tubing diameter

• The implant technique – Controlled deployment procedure

• The extent of device-vessel wall contact (Tissue Factor)– Minimal tubing wall contacting, no TF components

• The explant technique– In situ thrombus observation, non-adherent thrombus quantification

• Failure of LMCD– ?– Positive and negative control

• Hydrophilic surface– Non-adherent thrombus quantification

• Statistical power– Minimal of there different donors, up to 3 replicates per loop

7/11/2018 APS Confidential Information

Page 21: Thrombogenicity Testing for Blood-Contacting Medical ... · ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY

ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY • SURGICAL MED DEVICE • TOXICOLOGY

Acknowledgement

• Dr. Mark E. Smith

• Dr. Michael L. Conforti

• Sarah E. Howard

• Matthew R. Cunningham

• Abigail C. Beltrame

7/11/2018 APS Confidential Information

Page 22: Thrombogenicity Testing for Blood-Contacting Medical ... · ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY

ANALYTICAL • BIOCOMPATIBILITY • BIOSKILLS • INTERVENTIONAL MED DEVICE • PATHOLOGY • PHARMACOLOGY • REGULATORY • SURGICAL MED DEVICE • TOXICOLOGY

7/11/2018 APS Confidential Information