European Committee for Medical Ultrasound Safety (ECMUS)

Ultrasound Imaging

How do you know it’s SAFE?

EFSUMB Safety Committee

Page 2: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

Why are we concerned about safety?

• The range of clinical applications is becoming wider

• The number of patients undergoing ultrasound examinations is increasing

• New techniques with higher acoustic output levels are being introduced.

Page 3: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

What are bio-effects?

The result of ultrasound

interacting with biological

molecules as it passes

through tissue

Page 4: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

Safety

Is a bioeffect likely to cause a

problematic change (e.g. cell,

gene or DNA damage) ?

Page 5: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

Mechanisms of Action

Heat (Ispta)

Cavitation (P-)

Page 6: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

Attenuation =

Absorption +

Scatter

INCIDENT BEAM

IMAGE

Absorption 60-80%

HEAT

Page 7: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

HEATING (Absorption)

Increases with: frequency, exposure duration,

pulse repetition frequency

Page 8: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

2

Pulsed

Doppler Colour

Doppler

Harmonic

imaging B-mode

2D,3D

M-mode

Page 9: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

ATL UM 9

(Linear-Array L 10-5)

Transducer Self Heating :

International limits for probe surface temperature due to self-heating : T < 41 °C (in a water bath 2 l / 10 min) IEC 601-2-5 T < 43 °C (coupled to skin) T < 50 °C (emitting into air) IEC 60601-2-37

© CKollmann, Euroson 2010

Temperature distribution due to probe self-heating for diagnostic devices (maximum) :

B-Mode Pulsed Doppler Colour Doppler

Ispta = 11 mW cm2, MI = 0,5 Ispta = 533 mW cm2, MI = 0,9 Ispta = 606 mW cm2, MI = 0,3

Page 10: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

Biological consequences of heat depend on temperature rise

and duration.

Thermal Effects

Page 11: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

Thermal Effects

Tissues containing a large component of actively dividing cells are most

sensitive to the effects of heat.

Page 12: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

ACOUSTIC CAVITATION

Page 14: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

• Formation/activity of gas filled bubbles in an ultrasound exposed medium

• At MHz frequencies bubble radius ~1 µm

• Stable cavitation – bubbles oscillate

• Inertial cavitation – bubbles expand too far then collapse very rapidly, releasing enough energy to damage tissue

Acoustic Cavitation

Page 15: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

Inertial Cavitation

Page 16: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

How does the risk of heating & cavitation change

with imaging conditions?

Page 17: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS) All modes

Co

nta

ct tim

e

Ca

vitation

2

He

atin

g

Ca

vita

tio

n

2

He

atin

g

Ou

tpu

t P

ow

er

Fre

quency

Ca

vita

tion

He

atin

g

Increase Increasing importance Decreasing importance

Page 18: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS) Imaging & M-mode

M-m

ode

Cavitation

2

Heating

Secto

r fo

rmat

Heati

ng

Deep

er/

mo

re

foc

al

zo

ne

s

2

He

atin

g

Na

rro

w S

ecto

r fo

rma

t

2

He

atin

g

Page 19: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

2

Pulsed Doppler

Other Factors which may influence Heating and cavitation

Range Gate Width

(pulse length may vary with gate width)

Range Gate Depth

(power may increase with depth)

Doppler Velocity Range (pulse repetition frequency may increase)

Page 20: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

2

Receiver Gain has NO effect on heating or cavitation

So …..

Maximise it!

Page 21: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

A diagnostic exposure that produces a maximum in situ

temperature rise of no more than 1.5C above physiological levels

(37C) may be used clinically without reservation on thermal

grounds

WFUMB’98 - 01

Thermal Effects

WFUMB 1992

Page 22: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

A diagnostic exposure that elevates embryonic and fetal in

situ temperature above 41oC (by 4oC) for 5 min should be

considered to be potentially hazardous.

WFUMB’98 - 01

Thermal Effects

WFUMB 1992

Page 23: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

“On Screen” Labelling

Designed to provide safety related information

AIUM/NEMA Output Display Standard

Page 24: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

The Thermal index (TI) is an on-screen guide to the user of the potential for tissue heating.

Estimate of the tissue temperature rise in ºC which might be possible under "reasonable worst-case conditions”

THERMAL INDEX (TI)

Page 25: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

TI = Acoustic Power Output

(Acoustic Power to produce a 1oC rise)

THERMAL INDEX

Page 26: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

Soft tissue index

TIS

Bone

TIB

TIC – bone near transducer

THERMAL INDEX

Page 27: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

The Mechanical index (MI) is an on-screen guide of the

likelihood and magnitude of nonthermal effects.

MECHANICAL INDEX

Page 28: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

MI = p/f P: in situ pressure

f: frequency

MECHANICAL INDEX (MI)

Page 29: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

Power setting

High power low gain

Lower power high gain

Page 30: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

Power setting

High power low gain

Lower power high gain

Page 31: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

From scientific evidence of ultrasound-induced biological effects to date, there is

NO REASON to withhold diagnostic scanning during

pregnancy,

provided it is

1. medically indicated

2. is used prudently by fully trained operators.

Page 32: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

Ultrasound exposure during pregnancy

With increasing mineralisation of fetal bones, the possibility of

heating sensitive tissues such as brain and spinal cord increases.

So …. extra vigilance is advised

Page 33: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

3D imaging

No additional safety considerations

(particularly if there are significant pauses during scanning

to study or manipulate the reconstructed images)

Page 34: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

4D imaging (real-time 3D)

Involves continuous exposure

Guard against prolonging examination times unduly to improve the recorded

image sequence beyond that necessary for diagnostic purposes.

Page 35: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

Ultrasound Contrast Agents (UCAs)

• UCAs are not licensed for pregnancy

• Caution should be exercised when using in tissues for

which damage to microvasculature may be important (eg.

eye, brain, neonate)

•Exercise caution when using UCAs in patients with

severe coronary artery disease and pulmonary

hypertension.

Keep MI low, and avoid long exposure times

Refer to EFSUMB CEUS guidelines (2011)

Page 36: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

http://www.bmus.org/policies-guides

0 0.5 1.0 1.5 2.0 2.5 3.0

0.7

RECOMMENDED

RANGE

PROVIDED

ADEQUATE IMAGES

CAN BE OBTAINED (especially in 1st trimester)

Unlimited time

Observe ALARA

< 60

mins

< 30

mins

< 15

mins

< 4

mins

< 1

min NOT

RECOMMENDED

for

OB

scanning

Recommended scanning time limits for these TIs

(observe ALARA)

THERMAL INDEX

OBSTETRIC SCANNING

Monitor TIS up to 10 weeks post-LMP, TIB thereafter.

Page 37: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

Epidemiological safety studies Recent systematic reviews:

• Torloni MR. WHO systematic review of the literature

and meta-analysis.

UOG 2009;33:599-608

• Whitworth M. Ultrasound for fetal assessment in early

pregnancy.

Cochrane Database Syst Rev 2010

• Salvesen KÅ. Ultrasound in pregnancy and non-right

handedness: meta-analysis of randomized trials.

UOG 2011;38:267-71

Page 38: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

Only one controversial issue Prenatal ultrasound is associated with

left-handedness Published studies

Year Journal 1. author Type of study

1993 BMJ Salvesen RCT

1998 Early Hum Dev Kieler RCT

2001 Epidemiology Kieler Cohort

2002 Epidemiology Kieler Cohort

2011 Ultrasound Obstet Gynecol Heikkilä RCT

2011 Ultrasound Obstet Gynecol Salvesen Meta-analysis

Page 39: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

Meta-analysis Forrest plot from Salvesen UOG 2011;38:267-71

Page 40: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

We should worry – why? • 5 epidemiological studies demonstrate an

increased risk of left-handedness of 15-30%

• No other epidemiological study on ultrasound

and handedness has been published

• Experimental studies indicate effects on the

brain in some animal models

• Modern scanners produce higher outputs

Page 41: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

We should not worry – why? •Being left-handed is normal

(10-15% of population)

•A statistical association does

not imply a causal relationship

•The biological plausibility of this association is questionable

Page 42: Ultrasound Imaging How do you know it s SAFE? · Heating and cavitation Range Gate Width (pulse length may vary with gate width) ... worst-case conditions” ... induced biological

European Committee for Medical Ultrasound Safety (ECMUS)

European Committee for Medical Ultrasound Safety

www.efsumb.org/ecmus

British Medical Ultrasound Society

www.bmus.org

World Federation for Ultrasound in Medicine & Biology

www.wfumb.org

Safety Statements