VENTURA CPAP USER GUIDE ventura EMERGENCY USE ONLY not FOR ROUTINE CLINICAL USE WITHOUT FURTHER APPROVAL ALL HOSPITALS MUST WORK WITH THEIR OXYGEN engineering TEAMS TO ASCERTAIN THEIR VIE OUTFLOW, AND DOWNSTREAM FLOWS AND PRESSURES TO SPECIFIC WARD AREAS, BEFORE DEPLOYING THIS DEVICE OXYGEN INLET flow adjustment oxygen adjustment FLOW ON/OFF TO patient high low high low ON OFF AIR INTAKE ventura ventura A B C v2.0

VENTURA FLOW ON/OFF OXYGEN INLET CPAP INTAKE USER … · 2020-04-10 · ventura cpap user guide ventura emergency use only not for routine clinical use without further approval all

  • Upload
    others

  • View
    7

  • Download
    0

Embed Size (px)

Citation preview

Page 1: VENTURA FLOW ON/OFF OXYGEN INLET CPAP INTAKE USER … · 2020-04-10 · ventura cpap user guide ventura emergency use only not for routine clinical use without further approval all

VENTURACPAPUSER GUIDE

vent

ura

E M E R G E N C Y U S E O N LYn o t F O R R O U T I N E C L I N I C A L U S E W I T H O U T F U R T H E R A P P R O V A L

A L L H O S P I T A L S M U S T W O R K W I T H T H E I R O X Y G E N e n g i n e e r i n g T E A M S T O A S C E R T A I N T H E I R V I E O U T F L O W , A N D D O W N S T R E A M F L O W S A N D P R E S S U R E S T O S P E C I F I C W A R D A R E A S , B E F O R E D E P L O Y I N G T H I S D E V I C E

OXYGENINLET

flowadjustment

oxygenadjustment

FLOW ON/OFF

TOpatient

high

low

high

low

ON

OFF

AIR INTAKE

v e n t u r a

vent

ura

A

B

C

v 2 . 0

Page 2: VENTURA FLOW ON/OFF OXYGEN INLET CPAP INTAKE USER … · 2020-04-10 · ventura cpap user guide ventura emergency use only not for routine clinical use without further approval all

ONLY TO BE USED WITH THE PATIENT CIRCUIT PROVIDED

flow generator

oxygen analyser(MODELS VARY)

safety peepvalve

Patient Interface

Ab

c

flow flagindicator

(20 cmh 0)2

PEEP VALVE(cpap: 5-10 cmh 0)2

INCREASEFLOW

CORRECTFLOW

REDUCEFLOW

POSITIONDURING

INSPIRATION

1. Turn all controls (A, B, C) OFF (clockwise) and connect the UCL-Ventura to the 4-bar oxygen supply directly or by using an extension hose.

2. Connect the flow generator to the patient circuit shown.

3. Turn the FLOW ON/OFF control (A) fully ON (anticlockwise) and open the FLOW ADJUSTMENT control (B) (3-4 complete rotations anticlockwise).

4. Increase the oxygen concentration by turning the OXYGEN ADJUSTMENT control (C) (anticlockwise) to the desired level.

5. Apply mask to the patient with appropriate straps and ensure a good fit.

6. Alter the FLOW ADJUSTMENT control (B) to ensure optimal flow to:

• Flutter the mask outlet valve (larger port) throughout the respiratory cycle (the valve will flutter less during inspiration).

• Hold the blue flow indicator flag (where fitted) horizontal or just below horizontal on inspiration (the flag will remain below horizontal on expiration).

• A virtually continuous flow at the outlet filter should be felt on the back of a gloved hand throughout the respiratory cycle.

7. Alter the OXYGEN ADJUSTMENT control (C) if the oxygen concentration has changed.

If the device becomes damaged it should be withdrawn from use.

USING THE VENTURA CPAP