Upload
others
View
6
Download
0
Embed Size (px)
Citation preview
VMAX LFT USER MANUAL
Aaron - 25/09/2018
User Manual for
VMAX Lung Function
Machine
1
At Heartscope Victoria, prevention of infection transmission to technicians exposed to contaminated
spirometer surfaces is accomplished through proper use of anti-bactericidal sanitizer and use of barrier
devices, such as suitable gloves and masks.
The Mass Flow Sensor will need to be cleaned at least monthly or at an interval of every 50 test subjects
(500 blows). These instructions explain the proper cleaning and decontamination of the Detachable Mass
Flow Sensor.
1. Disconnect Tubing A and Tubing B from the Mass Flow Sensor
2. Detach the Mass Flow Sensor from the Breathing Valve Body
3. Under the Mass Flow Sensor, squeeze the ribbed sides of the connector, and then pull downwards
Infection Control Policy
Mass Flow Sensor
Tubing B
Tubing A
2
4. Place the Mass Flow Sensor into the disinfectant solution
The entire Mass Flow Sensor can be safely immersed in liquid disinfectant. The Sensor Cable, however,
cannot be immersed, because the connectors on each end can be damaged by liquids.
5. Allow the Mass Flow Sensor to soak in the disinfectant solution for 5 minutes
6. After disinfecting, rinse and thoroughly dry the sensor.
Make sure that the Mass Flow Sensor is rinsed and dried thoroughly before reconnecting the cable
The newly installed Mass Flow Sensor must successfully pass a Mass Flow Sensor Calibration before being
used for patient testing.
7. Note down the date in the maintenance database
If unaware of the maintenance database, please contact IT for assistance
3
This is the timetable for annual, monthly and daily user test hygiene:
Every Test
A new mouthpiece should be used for each test subject.
Daily
If the patient has come into contact with any parts of the spirometer then clean and disinfect accordingly
using the Clinell Disinfectant wipes. This includes the plastic gripper on the Spirometer which the patient
grips when doing the test. Ensure this is cleaned and disinfected at the end of every session.
Perform a visual inspection at the end of testing. If there is visible contamination to the flow head or
elsewhere on the device, clean and disinfect accordingly.
Monthly
The Mass Flow Sensor is disinfected regularly. The frequency of this is dependent on the Facilities’ Risk
Assessment, usage and test environment, but should be cleaned at least monthly or every 50 subjects (500
blows). They should also be replaced in the event of damage, or if visibly contaminated.
Annually
Annual inspection, maintenance and calibration certification should be obtained from the manufacturer or a
qualified service supplier. All these procedures are a part of the manufacturer’s recommended PPM (Planned
Preventative Maintenance) service procedure.
Timetable of Hygiene Procedures
4
The following Calibrations must be performed before the first test of every day.
It can also be performed at any time for verification.
1. Turn on the 3 cylinders (always check the level of the gauges)
2. Turn on the Vmax Machine and wait for 5 minutes for the Machine to warm up
3. Select “Flow Sensor Calibration | 1” to open the Flow Volume Calibration screen
4. The Flow Volume Calibration screen will be displayed below.
Calibration
Calibration Procedure
5
5. Attach the Calibration Syringe to the Mass Flow Sensor
6. Select “F1” on the Flow Volume Calibration screen to begin the calibration
The Mass Flow Sensor Zero dialog box will be displayed
7. Hold the Calibration Syringe still for approximately 10-15 seconds and then select “Space Continue”
8. A timer will count down to zero seconds before continuing to the Vmax Flow Volume Calibration screen
6
9. Perform 5 inspiratory and expiratory strokes within the target ranges shown in yellow on the graph
The bar graph segments on the right turn to green when the mean flow rate of a stroke falls within the
adjacent range, which is what we are aiming for.
* to ensure that the strokes are within the yellow target ranges, turn the knob of the lever first before pulling
in really slow motion IN and OUT making a nice loop. Increase the speed and circular motion each time. Once
the bar graph segments on the right turn green, the screen should look like this:
10. Once the bar graph segments on the right turn to green, the next calibration graph will be displayed:
7
11. Perform 5 inspiratory and expiratory strokes within the target ranges shown by the red lines:
Four of the five strokes are displayed and should appear as follows:
• One stroke should reach the lowest red dotted line (0.5 LPS).
• One stroke should reach the highest dotted line (3.0 LPS).
• One stroke should be halfway between the dotted lines (1.5 LPS)
• One stroke (fourth stroke) should include a peak-flow rate near 12 LPS (8 LPS minimum)
This stroke should be created without ‘banging’ the piston at either end of the stroke
12. The Calibration is successful when the time and date are present on the top table
Note that none of the values on the top table should be in red. Click “F3” to store the calibration.
8
1. Select “New Study | 2” in the Vmax Program Manager Window
2. The Vmax Demographic Input dialog box will be displayed
To enter the patient demographics, click in each text box and type or select the information.
Press ENTER to move the insertion point to the next text box.
Vmax Overview Create New Patient
Will accept letters, numbers &
symbols
Computer
automatically
enters today’s date
Computer automatically
calculates age from the
date of birth
Can be used if you are unable to measure the patient’s height
Pakenham, St Albans etc.
Reporting Physician
Technician Initials
Clinical Details (SOB etc.)
9
3. Select “Other” from the top menu ribbon then select “Additional Parameters”
Type the referring doctor’s details as demonstrated above.
01: Referring Doctor’s Name
02: Clinic Name
03: Clinic Address
04: CC Doctor Name (if applicable)
05: CC Clinic Name (if applicable)
06: CC Clinic Address (if applicable)
Then select “F3” to save the Referring Doctor’s details
3. Select F3 to store the new patient information and return to the Vmax Program Manager
10
1. Select “Pulmonary Function | 5” on the Vmax Program Manager
2. The Pulmonary Function Program window will be displayed.
Select Test Level: “Pre” then Select “1 Flow Volume Loop”
Pre-Spirometry
Flow Volume Loop
11
Once the 3rd test is done, choose the best result (number), click F4 and then F3 (to store)
3. The Flow Volume Loop Test screen will be displayed
4. Explain the procedure to the patient, and the purpose of the mouthpiece and nose clip
5. Attach the mouthpiece and instruct the patient to sit in front of the Spirometer
Adjust the height of the spirometer accordingly to ensure the patient is comfortable
6. Attach the nose clip onto the patient and instruct them to place their mouth onto the mouthpiece
Ensure that the patient’s mouth is tightly sealed around the mouthpiece
7. Select “F1” on the Flow Volume Loop Test screen to begin the test procedure
8. Normal Breathing: Instruct the patient to do Normal Breathing first (coaching patient repeatedly)
Ensure that you record at least 3 stable tidal breaths
9. Maximal Inspiration/Expiration: Now tell the patient: “on your next breath, [press F1] take a deep breath
and then forcefully exhale and completely empty the lungs.” Coach the patient to “keep exhaling” until the
smaller box on the right side turns green. Then instruct the patient to take a deep breath in and then come
off the mouthpiece.
10. Select “End Test” to check the reference values / outcome
REPEAT STEPS 7 TO 10 AT LEAST 3 TIMES TO OBTAIN REPRODUCIBLE RESULTS
12
1. Prepare Ventolin, Spacer and a Timer
2. Set the Timer to 5 minutes and press Start
3. Give the patient the 1st puff
Instruct them to take a deep breath in, *hold for 2 seconds*, then breathe out. Repeat 3 times.
4. After 1 minute, give the patient the 2nd puff
Instruct them to take a deep breath in, *hold for 2 seconds*, then breathe out. Repeat 3 times.
5. After another minute, give the patient the 3rd puff
Instruct them to take a deep breath in, *hold for 2 seconds*, then breathe out. Repeat 3 times.
6. Let the patient rest for a period of 2 minutes to let the Ventolin administer
1. Select Test Level: “Post” then Select “1 Flow Volume Loop” on the Pulmonary Function Program window
Post-Spirometry
Flow Volume Loop
Administering Ventolin
Begin Post-Spirometry Test
13
Once the 3rd test is done, choose the best result (number), click F4 and then F3 (to store)
2. The Flow Volume Loop Test screen will be displayed
3. Select “F1” on the Flow Volume Loop Test screen to begin the test procedure
4. Normal Breathing: Instruct the patient to do Normal Breathing first (coaching patient repeatedly)
Ensure that you record at least 3 stable tidal breaths
5. Maximal Inspiration/Expiration: Now tell the patient: “on your next breath, [press F1 during inhalation]
take a deep breath and then forcefully exhale and completely empty the lungs.”
Coach the patient to “keep exhaling” until the smaller box on the right side turns green.
Then instruct the patient to take a deep breath in and then come off the mouthpiece.
6. Select “End Test” to check the reference values / outcome
REPEAT STEPS 3 TO 6 AT LEAST 3 TIMES TO OBTAIN REPRODUCIBLE RESULTS
ENSURE ALL THREE LIGHTS ON THE RIGHT SIDE OF THE SCREEN ARE IN GREEN
ALL THREE MUST BE GREEN (PRE AND POST SPIROMETRY)
14
1. Select Test Level: “Post” then Select “Single Breath DLCO | C” on the Vmax Program Manager
2. The Single Breath DLCO Test screen will be displayed
DLCO
Single Breath Diffusing Capacity
15
Once the 2 trials are done, choose the best result (number)
3. Select “F1” on the Single Breath DLCO Test screen to begin the test procedure
Face the Spirometer away from the patient’s face
4. Select “Space to Continue” to flush the breathing valve with 100% DLCO gas
Note that you will hear a loud cleaning sound
5. After the calibration is complete, explain the test procedure instructions to the patient
6. Instruct the patient to attach the nose clip and put in the mouthpiece
Ensure the patient’s mouth is tightly sealed around the mouthpiece
Normal Breathing:
7. Instruct the patient to begin with Normal Breathing first (coaching patient)
Maximal Expiration/Inspiration:
8. Instruct the patient to exhale completely; press F1 during the maximal exhalation
9. When “Full Inhalation Then Hold” is displayed on the screen, instruct the patient to inhale completely
Breathe Hold:
10. Coach the patient to hold their breath until the exhalation valve opens
The patient will need to hold their breath for at least 10 seconds (until 2nd vertical green line is reached)
Maximal Expiration:
11. Coach the patient to exhale completely (empty lungs) until the computer ends the test
12. The patient can now come off the mouth piece
REPEAT STEPS 3 TO 12 ONLY 2 TIMES TO OBTAIN REPRODUCIBLE RESULTS
13. Select “Exit” on the top of the Single Breath DLCO Test screen
16
1. Select “Reports | F8” on the Vmax Program Manager
2. The Reports screen will be displayed. Select “Heartscope – PW (JT)” from the Reports List drop down box
Reports
Obtaining the Report
17
3. Select “F7” to access PFT Comments and Interpretation Edit Box
4. The PFT Comments and Interpretation Edit Box will be displayed
18
5. Select “F5 Comments”. Then select one of the “A” to “I” options accordingly
Double click on one of the options to insert the sentence into the comments section
6. Select “F8 Narrative”. Select Category “A – Indications for Evaluation”
The Narrative section should address the following points:
• What was the reason for the test?
• Is the patient a smoker?
• Did the patient have any difficulties with the technique?
Select one of the options accordingly if required, or type inside the Narrative section
Double click on one of the options to insert the sentence into the narrative section
19
7. Select “F7 LOGIC Tree”. Then select “F2” to access User LOGIC Paths Manager
8. The LOGIC Pathways Manager window will be displayed
Select “F2” to Generate text using Selected LOGIC Pathway
Copy all of the generated text and select “ESC” to return to the PFT Comments Edit box
20
9. Paste the generated text into the LOGIC Tree section
Select “F3” to Store Results and Exit
10. Select “Print PDF” on the top bar
Ensure that all information is correct and all relevant
graphs are displayed in the Report
21
The fo
The following is a list of potential problems you may eventually encounter using the Vmax Lung Function
Machine. For problems not covered by this list, or for any questions, contact technical support.
Condition Possible Causes Possible Remedies No power to system (computer, monitor, Vmax modules)
Loose power cable Check all power cable connections
Defective power outlet Try a known, good power outlet
Flow Volume Calibration Failure
Defective calibration syringe Try an alternate calibration syringe Verify the syringe volume setting in Calibration Setup
Defective breathing circuit Check the breathing circuit assembly for incorrect connection, incorrect assembly, or damage
Balloon valve tubing connection Switch tubing connectors
Dirty or defective mass-flow sensor Mass Flow Sensor Zero Check and Mass Flow Sensor Cleaning
Leak in spirometer dry rolling seal Replace seal
Incorrectly assembled spirometer Disassemble and re-assemble
Resistance to breathing during any test
Breathing circuit defective or incorrectly assembled
Check breathing circuit for damage or incorrect assembly
Spirometer incorrectly assembled Disassemble and re-assemble spirometer
Balloon valve tubing incorrectly assembled Switch tubing connectors
Inadequate oxygen or diffusion mix gas pressure
Make sure gas cylinders are completely turned on, contain at least 200 PSI internal pressure, and are set to 50-60 PSI delivered pressure (10-20 PSI above oxygen pressure for Vmax diffusion mix)
Condition Possible Causes Possible Remedies Hardware Check Warning. Check Power, Cables.
Analyzer Module turned off Verify that the Analyzer Module is turned on
Loose or disconnected interface cable between the computer and the Analyzer Module or the spirometer
Verify that the interface cable is securely attached
Hardware lock-up Turn the entire system off and then turn it back on
The Mass Flow Sensor does not respond. Check the Sensor Cable or Substitute Another Sensor
The mass-flow sensor cable is loose or disconnected from the flow signal port on the rear of the Analyzer Module
Verify that the cable is securely attached
Defective mass flow sensor Replace the entire Mass Flow Sensor Assembly
The Sensors are Responding Incorrectly to Calibration Gas. Check Calibration Gas Tank Pressures and Connections
Calibration tank turned off or only slightly opened
Make sure calibration gas is completely turned on
Inadequate gas pressure in calibration tank
Make sure there is at least 200 PSI remaining in the calibration gas cylinders
Hoses from gas tanks are switched or disconnected at the rear of the instrument
Make sure gas hoses are attached correctly
Ensure that the Sample Line is Connected to the Calibration Fitting
The Permapure sample tubing is not connected to the inlet port on the front of the Analyzer Module
Connect the sample tubing to the inlet port
General Problems
Troubleshooting
Warning Messages