Upload
solene-barbotin
View
215
Download
0
Embed Size (px)
Citation preview
7/28/2019 Wetting and El. Properties of Biomed. Alloys
1/19
Wetting and electrochemical properties of biomedical
alloys
Solne Barbotin
Abstract
Wettability of Stainless Steel and Cobalt-Chromium-Molybdenum alloy in two differentsolutions which are Phosphate Buffer Solution (PBS) and PBS plus bovine serum albumin is
examined under application of a potential, using the captive bubble method. It is found thatwettability of both alloys increases with potential decrease, in the two different solutions, thusshowing wettability has a link with potential.
Keywords : wettability, potential, captive bubble method, stainless steel, CoCrMo alloy
I. Introduction
Wettability allows to character-
ize the surface of a material, and
can be linked to the surface en-
ergy, which is an important parameter. In-
deed, it was found that it can be related tofriction and wear phenomena [1]. When
two materials are sliding on each other,
which is the case for hip prosthesis, parti-
cles can detached from one of the material
thus leading to formation of third body
particles. Their presence can lead to in-
flammation of the neighboring tissues, and
thus cause pain to the patient. It is stated
that the better the wettability, the better the
materials sliding properties thus influenc-
ing the wear phenomena.The main goal of this project was then
to study the influence of potential on the
materials wettability. In order to perform
MX/TIC EPFL, semester project. Supervised byAnna Igual and Stefano Mischler.
the tests, the first objective was to design
an electrochemical cell, and then to state
if applying a potential the cathodic way
or anodic way has an effect on the mate-
rials sample wettability. The influence of
the type of alloy on the wettability using
CoCrMo alloy and Stainless Steel (SS) isalso investigated, as well as the influence
of the electrolyte solution on the samples
wettability, using Phosphate Buffer Solu-
tion (PBS) and PBS plus bovine serum al-
bumin.
Then, if it occurs that the potential seems
to have an effect, it will be stated which
effect has potential change on the surface
wettability characteristics.
II. Preliminaries
Wettability characterizes the ease with
which a liquid drop spreads on a surface
solid. It is characterized by the contact
angle theta which depends on the three
1
7/28/2019 Wetting and El. Properties of Biomed. Alloys
2/19
Semester project Autumn 2012 S.Barbotin
surface tensions sg , sl and lg , acting on
the contact line between the solid, liquid
and gazeous phases. At equilibrium, the
sum of the parallele forces at the surface of
the solid is null. Follows the relationship,based on Youngs equation [2]:
cos lg + sl = sg (1)
cos =sg sl
lg(2)
A number of friction and surface phenom-
ena are explicable in terms of the surface
energy of adhesion of the contacting mate-
rial.It is interesting to know the wettability of
a material, because it can be related to the
materials surface energy, through Youngs
equation, as stated before.
In order to increase of an increment dA
a solid or liquids surface, it is necessary
to move a certain quantity of atoms from
the mass to the surface. Since the surface
atoms have less bonds than those inside
the mass, the solid or liquid free energy
increases of a quantity dG. dG = dA.The definition of surface energy comes
from a more general formulation of free
energy change or free enthalpy of a ther-
modynamic system, before surface effects
[2]:
dF = PdV SdT+i
idni + dA (3)
dG = VdP SdT+i
idni + dA (4)
=
F
A
V,T,ni
; =
G
A
P,T,ni
(5)
In a cubic face centered metal, an atom has
12 close neighbors and so, as many bonds.
The energy to provide to break these bonds
is equivalent to the enthalpy of sublimation.
The atoms possess 9 close neighbors if they
belong to a surface which orientation cor-
responds to the Miller indices (111). Thus,according to the simple model exposed be-
fore, the metals surface energy depends on
the crystalline orientation. For a cfc metal,
atoms belonging to the orientation surface
(111), (100) and (110) have respectively 9, 8
and 6 close neighbors. The surface energy
then increase in the order (111) < (100)
< (110). Of maximum compactness, the
(111) orientation surfaces are, thermody-
namically speaking, the most stable.
III. Materials and Methods
III.1 Materials
III.1.1 Electrochemical cell
The electrochemical cell built to host the
tests was made of PMMA for the cell itself
Fig. 1(a) and Fig. 1(b), size 74 74 65 mm
(length width height) + lid, a rubber seal
was added in the lid to hold the sample,
size 20 3, 0 mm (rubber ring diameter
ring diameter thickness). The syringes
needle diameter was 0,8mm.
The cell was chosen to be geometrically
squared and not cylindrical, because if
cylindrical, it might would have led to ge-
ometrical deformations of the bubble on
the pictures that were taken with a camera.
The size of the cell was chosen based on
the following concerns : the cell needed tobe big enough to allow good visual appre-
ciation and to allow the reference electrode
placement (height) but also small enough
in order not to waste solution that was at
disposal.
A reference electrode calomel (SCE)
2
7/28/2019 Wetting and El. Properties of Biomed. Alloys
3/19
Semester project Autumn 2012 S.Barbotin
(a) Closed electrochemical cell (b) Open elctrochemical cell
Figure 1: Electrochemical cell
(a) Reference
electrode
(b) Counter electrode (c) Potentio-stat.
Figure 2: Materials used for the electrochemical part of the experiment
and a counter electrode made of plat-
inum wire were used for the experi-
ment and connected to a potentiostat
Fig. 2(a)Fig. 2(b)Fig. 2(c). The results pre-sented in section IV, for the potential values
are all vs SCE.
III.1.2 Stainless Steel (SS)
The SS sample Fig. 3(a) was of grade 316L,
and having the following dimensions :
- 20 mm diameter
- 5 mm height
The surface of the sample in contact with
the electrolyte during experiment was of
r2 = 3,14cm2.
3
7/28/2019 Wetting and El. Properties of Biomed. Alloys
4/19
Semester project Autumn 2012 S.Barbotin
(a) Stainless steel (b) CoCrMo
Figure 3: Samples
III.1.3 Cobalt-Chromium-Molybdenum
alloy (CoCrMo)
The CoCrMo sample Fig. 3(b) is of grade
LC ASTM F 1537-00. It was characterized
with X-Ray Fluorescence analysis, giving
the following composition :
- Fe 0.24 %
- Co 64.95 %
- Cr 28.80 %
- Mo 5.68 %
- Ni 0.32 %
It has the following dimensions :
- 19mm diameter
- 6 mm height
The surface of the sample in contact with
the electrolyte during experiment was then
of r2 = 3, 0cm2.
III.1.4 Phosphate Buffer Solution (PBS)
The phosphate buffer solution was pre-
pared with the following quantities :
- 8 g/L NaCl
- 0.2 g/L KCl
- 1.44 g/L Na2HPO4- 0.25 g/L KH2PO4
III.1.5 PBS + bovine serum albumin
The PBS+albumin solution was prepared
with the following quantities :
- 8 g/L NaCl
- 0.2 g/L KCl
- 1.44 g/L Na2HPO4- 0.25 g/L KH2PO4- 0.5 g/L BSA (albumine)
III.2 Methods
III.2.1 Samples polishing
The first step before starting the experi-ment is to mirror polish the samples, using
silicon carbide sandpaper. The polishing
buffers granulometry used were the follow-
ings, with their corresponding grain size,
using water as a lubricant :
- Grade 500 (29-33m)
4
7/28/2019 Wetting and El. Properties of Biomed. Alloys
5/19
Semester project Autumn 2012 S.Barbotin
- Grade 1200 (13-15m)
- Grade 2400 (6-8m)
- Grade 4000 (4-5m)
Then, the last polishing step was per-
formed on a velvet polishing buffer, us-
ing red lubricant (no water) with diamond
sprayed particles with a size of 3m. Red
lubricant cools down the sample and helps
removing the debris, it is alcohol based.
III.2.2 The captive bubble method
(CBM)
The captive bubble method is used for mak-
ing the experiment in a retrospective way.It is possible to conduct the experiment un-
der condition mimicking the one in situ,
this condition being total immersion in a
fluid : the prosthesis are indeed immersed
in the body synovial fluid. Another rea-
son for using the captive bubble method,
is that because it is immersed in a liquid, it
is possible to apply an electrochemical po-
tential and study the wettability evolution
in function of this potential.
The methods procedure is as follow.
An air bubble is injected from a syringe
into the PMMA cell chamber containing
the solution Fig. 4(b). The bubble is re-
leased from the tip of a needle and floated
3-4 mm to the sample surface held by the
rubber seal in the cell lid, at the top of the
liquid. The bubble size is in the range of 8
to 12 L. Fig. 4(a)
III.2.3 Instruments
For the contact angle measurement a sim-
ple method was used based on the photog-
raphy of finite dimension bubbles Fig. 6(a)
Fig. 6(b). For this, an instrument consisting
of two parts was used :
(a) Image of the bubble obtained with the CBM
(b) Scheme of the bubble introduction in the system
Figure 4: CBM
- a microscope with numerical camera
connected to a computer Fig. 5(a)
- a translation table XYZ allowing sam-
ples positioning in front of the camera
A bubble of 102 L of air is injected
on the downwards surface of the sample,
with a syringe. The bubble image is cap-
tured by the video camera connected to the
computer. Fig. 5(b)
III.2.4 Cathodic and Anodic Steps, Lin-
ear Sweep Voltammetry (LSV)
and Cyclic Voltammetry (CV)
The cathodic and anodic steps consist of
diminishing for the cathodic steps and in-
creasing for the anodic steps the potential,
5
7/28/2019 Wetting and El. Properties of Biomed. Alloys
6/19
Semester project Autumn 2012 S.Barbotin
(a) Positionning table and camera
(b) System with cell, camera and computer
Figure 5: Operating system
by steps of 100mV, starting from the Open
Circuit Potential (OCP) of the sample im-
mersed in the electrolyte. The OCP is the
potential of the working electrode relative
to the reference electrode when no poten-
tial or current is being applied to the cell.
The bubble is changed in between each po-
tential step, by stirring the electrolyte with
the syringe at the samples surface.
LSV consists of applying a potential
that constantly diminishes or constantly in-creases over time, by steps of 0,002V, with-
out changing the bubble, and taking pic-
tures every 50mV or every 100mV, starting
at the OCP. For CV, it is the same, the only
difference is that the potential starts di-
minishing so going in the cathodic domain
until it reaches the limit potential that was
fixed, and then it starts increasing, going
to the anodic domain. These two methods
for measuring were used to globally see
if there was a change in the contact angle,and in which domain. It was an approach
to apprehend the behavior of the bubble.
III.2.5 Contact angle measurements
This step is also very important for repro-
ducibility. Indeed, the method used for
measuring such a precise angle, is very
important because variations of angles are
sometimes very small.
The drop analysis plugin [3] created by
the Laboratory BIG in EPFL was created
for drops contact angle measurements, but
can also apply to the CBM since it is partic-
ularly accurate for unsharp or noisy bound-
aries, which corresponds to the images ob-
tained with the CBM.
The drop analysis plugin is using a new
approach to measure contact angle and sur-
face tension : the Low Bond Axisymmetric
Drop Shape Analysis (LB-ADSA)In LB-ADSA, the theoretical profile is
not fitted to a discretized drop contour but
is optimized based on an image energy
approach. In this approach, segmentation
and fitting are combined in what can be
seen as a model-based segmentation. The
complete pixel information is used during
the fitting process. This approach is par-
ticularly advantageous when a clear accu-
rate contour detection is difficult because
of unsharp or noisy boundaries. Appli-cation of image energies to segmentation
tasks is an active research domain. Fol-
lowing current proposals, a gradient im-
age energy comprising gradient direction
is used. This energy term has the bene-
fit of being invariant to parametrization.
6
7/28/2019 Wetting and El. Properties of Biomed. Alloys
7/19
Semester project Autumn 2012 S.Barbotin
(a) Image before fitting (b) Final fitting (c) Fitting with angles
(d) Measurements table
Figure 7: Fitting steps
In addition, it is suggested to account for
pixel value information by a region energy.
Region energies are statistically based and
offer the advantage of having a wide at-
traction range. The image energy approach
has already been applied to the domain of
contact angle measurements (DropSnake
method) and demonstrated its potential for
noisy and/or low contrast data.
Finally, the presented method is applied
to a continuous image of the drop by us-
ing cubic B-spline interpolation. Then, the
evolution procedure takes place in this con-
tinuous domain to avoid inaccuracies in-
troduced by pixelization and discretization.
Fig. 7(a) show the approximation curve. It
(the green curve) can be fitted to the bub-ble contour by modifying width, heigth
and other parameters, and once a close
fitting is reached manually, an automatic
fitting was performed using the gradient
energy approximation, leading to the fit-
ting in Fig. 7(b).
The angle measured by the program is
the one called on Fig. 7(c) and presented
in Fig. 7(d) as CA[], which is the angle
formed by the air bubble. The angle of
interest for us, is the one formed by the
water, that is to say the angle on the
figure, which is the one calculated as 1802and presented in the results.
For the program to be able to analyze
the bubble, the images must be black and
white and the bubble must stand upwards
like in Fig. 8(b). Since most of the images
taken were in color and with the bubble
downwards, a plugin flipping the image
and changing color to black and white was
applied. For the other images, because
there was some problems concerning thepixels colorization as seen on Fig. 8(a), a
moving average of 2 was vertically and
horizontally applied, which means a color
mix is done between two adjacent pixels,
in order to get rid of this problem. It was
primordial to get rid of it because it would
7
7/28/2019 Wetting and El. Properties of Biomed. Alloys
8/19
Semester project Autumn 2012 S.Barbotin
(a) Color picture of the bubble
(b) Black and white picture of the bubble
Figure 6: Exemple of the two image types taken
have interfered with the energy gradient
fitting.
IV. Results
IV.1 Electrochemical cell conception
A functional cell was obtained, allowing
good visibility of the samples surface, andso of the bubble evolution. The pictures
taken were of good quality considering the
electrolyte in which the pictures were taken
was a liquid, rendering good quality more
difficult to obtain than with the sessile drop
method, where the electrolyte is air.
(a) Problem encountered with black and white pic-
tures
(b) Final image
Figure 8: Black and white image steps
IV.2 SS in PBS solution
Fig. 9(a) with CV shows a drop in contact
angle when decreasing the potential. The
contact angle starts diminishing at -0,6 V,
going from 30 up to 10 when reaching
-0,95 V, at a constant rate. Then, the poten-
tial evolution was reversed and increased
8
7/28/2019 Wetting and El. Properties of Biomed. Alloys
9/19
Semester project Autumn 2012 S.Barbotin
0
5
10
15
20
25
30
35
-1,
2
-1
-0,
8
-0,
6
-0,
4
-0,
2
0
0,
2
0,
4
0,
6
0,
8
Watercontactang
le[]
Poten0al[V]
CVSSPBS
CVSSPBS
(a) Contact angle in function of potential
-8
-7
-6
-5
-4
-3
-2
-1
0
-1,3 -0,8 -0,3 0,2 0,7
log(|i|)[A]
Poten0al[V]
CV
(b) Current curve
Figure 9: SS PBS CV
anodically, reaching 0,6 V with no more
evolution of the contact angle, thus being
stationnaire at 101.
For the cathodic steps, Fig. 10(a), the con-
tact angle starts decreasing at 0,35 V until
0,5 V for the first test and until 0,6 V for
the 2nd test. At 0,5 V, for the first test, the
contact angle stops decreasing and is sta-
tionary, whereas for the 2nd test the contactangle keeps decreasing at a constant rate.
Test 2 had to be stop at 0,6 V. Indeed, final
values of contact angles are so low when
decreasing potential, that for cathodic steps
test 2, the contact angle of bubbles at poten-
tials lower than -0,6 V were not measurable.
0
2
4
6
8
10
12
14
-0,8 -0,6 -0,4 -0,2 0
Watercontactangle[]
Poten0al[V]
SSPBSCathodic
SSPBS
cathodic2
SSPBS
cathodic1
(a) Contact angle in function of potential
-0,9
-0,8
-0,7
-0,6
-0,5
-0,4
-0,3
-0,2
-0,1
0
-2,E-04
-2,E-04
-1,E-04
-1,E-04
-9,E-05
-7,E-05
-5,E-05
-3,E-05
-1,E-05
0 500 1000 1500 2000 2500 3000 3500
Poten&al[V]
Currenti[A]
Time[s]
SSPBScathodicsteps1current
SSPBScathodicsteps2current
SSPBScathodicsteps1poten?al
SSPBScathodicsteps2poten?al
(b) Current and potential curves
Figure 10: SS PBS cathodic steps
The sample surface was so hydrophilic thatthe air bubble didnt stick to it anymore,
thus rendering impossible further contact
angle measurements.
Concerning the anodic steps curves in
Fig. 11(a), the contact angles are fluctuating
with the potential between 25 and 12, the
9
7/28/2019 Wetting and El. Properties of Biomed. Alloys
10/19
Semester project Autumn 2012 S.Barbotin
0
5
10
15
20
25
30
-0,3 -0,2 -0,1 0 0,1 0,2 0,3
Watercontactangle[]
Poten0al[V]
SSPBSAnodicsteps
SSPBS
anodic1
SSPBS
anodic2
SSPBS
anodic3
(a) Contact angle in function of potential
-0,3
-0,2
-0,1
0
0,1
0,2
0,3
0,4
0,0E+00
2,0E-06
4,0E-06
6,0E-06
8,0E-06
1,0E-05
1,2E-05
1,4E-05
1,6E-05
1,8E-05
0 1000 2000 3000
Poten&al[V]
Currenti[A]
Time[s]
SSPBSanodicsteps1current
SSPBSanodicsteps2current
SSPBSanodicsteps3current
SSPBSanodicsteps1poten=al
SSPBSanodicsteps2poten=al
SSPBSanodicsteps3poten=al
(b) Current and potential curves
Figure 11: SS PBS anodic steps
contact angle globally diminishing with
increasing potential.
0
2
4
6
8
10
12
-1,5 -1 -0,5 0
Watercontactan
gle[]
Poten0al[V]
SSPBS+albuminCathodic
LSVSSalbumin
cathodic1
SSalbumincathodic2
SSalbumin
cathodic3
(a) Contact angle in function of potential
-1,8E-03
-1,6E-03
-1,4E-03
-1,2E-03
-1,0E-03
-8,0E-04
-6,0E-04
-4,0E-04
-2,0E-04
0,0E+00
-1,4 -0,9 -0,4 0,1
Currenti[A]
Poten.al[V]
LSVSS
PBS
+albumin
cathodic1
(b) LSV
-1,2
-1
-0,8
-0,6
-0,4
-0,2
0
-2,5E-03
-2,0E-03
-1,5E-03
-1,0E-03
-5,0E-04
0,0E+00
0 500 1000 1500 2000 2500
Poten&al[V]
Currenti[A]
Time[s]
SSPBS+albumincathodic2current
SSPBS+albumincathodic3current
SSPBS+albumincathodic2poten@al
SSPBS+albumincathodic3poten@al
(c) Cathodic steps
Figure 12: SS PBS+albumin cathodic
10
7/28/2019 Wetting and El. Properties of Biomed. Alloys
11/19
Semester project Autumn 2012 S.Barbotin
0
5
10
15
20
25
-1,7 -1,2 -0,7 -0,2 0,3
Watercontactangle[]
Poten0al[V]
LSVCoCrMoCathodic
LSVCoCrMo
cathodic1
LSVCoCrMo
cathodic2
LSVCoCrMo
cathodic3
(a) Contact angle in function of potential
-5,0E-02
-4,0E-02
-3,0E-02
-2,0E-02
-1,0E-02
0,0E+00
1,0E-02
-2 -1,5 -1 -0,5 0
Currenti[A]
Poten.al[V]
LSVCoCrMoPBScathodic1
LSVCoCrMoPBScathodic2
LSVCoCrMoPBScathodic3
(b) Current in function of potential
Figure 13: LSV CoCrMo PBS cathodic
IV.3 SS in PBS+albumin solution
For SS in PBS+albumin solution only the
cathodic behavior presented in Fig. 12(a)
was studied, because it appears to be
the most interesting one. The tests werestopped when contact angle measurements
were not possible anymore because of hy-
drogen evolution.
With albumin, the initial wettability of
the sample is already very good. At OCP,
the contact angle is of only 11, 8 and 6
0
5
10
15
20
25
30
35
-1,3 -0,8 -0,3 0,2
Watercontactang
le[]
Poten0al[V]
CoCrMoCathodicSteps
CoCrMo
cathodic1
CoCrMo
cathodic2
(a) Contact angle in function of potential
-1,3
-1,1
-0,9
-0,7
-0,5
-0,3
-0,1
0,1
-4,0E-03
-3,5E-03
-3,0E-03
-2,5E-03
-2,0E-03
-1,5E-03
-1,0E-03
-5,0E-04
0,0E+00
0 1000 2000
Poten&al[V]
Currenti[A]
Time[s]
CoCrMoPBScathodic1current
CoCrMoPBScathodic2current
CoCrMoPBScathodic1poten?al
CoCrMoPBScathodic2poten?al
(b) Current and potential curves
Figure 14: CoCrMo PBS cathodic steps
respectively for the first cathodic test, 2nd
and then LSV test. Eventhough there weresome small contact angle fluctuations, the
contact angle decreases with decreasing
potential, until it reaches a very low value
of 1-3. The LSV test shows a later decrease
in contact angle, and also a later hydrogen
evolution, thus reaching lower potential
11
7/28/2019 Wetting and El. Properties of Biomed. Alloys
12/19
Semester project Autumn 2012 S.Barbotin
0
5
10
15
20
25
-1 -0,5 0 0,5 1
Watercontactan
gle[]
Poten0al[V]
LSVCoCrMoAnodic
LSVCoCrMo
anodic1
LSVCoCrMo
anodic2
(a) Contact angle in function of potential
-1,0E-04
0,0E+00
1,0E-04
2,0E-04
3,0E-04
4,0E-04
5,0E-04
-0,6 -0,4 -0,2 0 0,2 0,4 0,6
Currenti[A]
Poten.al[V]
LSVCoCrMo
PBSanodic2
(b) Current in function of potential
Figure 15: CoCrMo PBS anodic
than the other two tests.
As for SS in PBS, values reached contact
angles as low as 11, thus presenting a
very good wettability.
IV.4 CoCrMo in PBS solution
For the CoCrMo in PBS solution, concern-
ing the anodic behavior, one can see in
Fig. 15(a) the contact angle remains con-
0
2
4
6
8
10
12
-1,5 -1 -0,5 0
Watercontactang
le[]
Poten0al[V]
CoCrMoPBS+albuminCathodic
LSVCoCrMo
albumin
cathodic1
CoCrMo
albumin
cathodic1
CoCrMo
albumin
cathodic2
(a) Contact angle in function of potential
-1,4
-1,2
-1
-0,8
-0,6
-0,4
-0,2
0
-4,5E-03
-4,0E-03
-3,5E-03
-3,0E-03
-2,5E-03
-2,0E-03
-1,5E-03
-1,0E-03
-5,0E-04
0,0E+00
0 500 1000 1500 2000
Poten&al[V]
Currenti[A]
Time[s]
CoCrMoPBS+albumincathodicsteps1current
CoCrMoPBS+albumincathodic2current
CoCrMoPBS+albumincathodic1potenal
CoCrMoPBS+albumincathodic2potenal
(b) Current and potential curves
-7,E-04
-6,E-04
-5,E-04
-4,E-04
-3,E-04
-2,E-04
-1,E-04
0,E+00
-1,4 -1,2 -1 -0,8 -0,6 -0,4
Currenti[A]
Poten.al[V]
LSVCoCrMoPBS+albumincathodic1
(c) Current in function of potential
Figure 16: CoCrMo PBS+albumin cathodic
12
7/28/2019 Wetting and El. Properties of Biomed. Alloys
13/19
Semester project Autumn 2012 S.Barbotin
0
2
4
6
8
10
12
-1 -0,5 0 0,5
Watercontactangle[
]
Poten0al[V]
CoCrMoPBS+albuminAnodic
LSVCoCrMo
albumin
(a) Contact angle in function of potential
0,0E+00
5,0E-06
1,0E-05
1,5E-05
2,0E-05
2,5E-05
3,0E-05
-0,7 -0,2 0,3
Currenti[A]
Poten.al[V]
LSVCoCrMoPBS+albuminanodic1
(b) Current in function of potential
Figure 17: CoCrMo PBS+albumin anodic
stant while increasing the potential. For
the first test, the contact angle is of 210,5,
and for the 2nd is of 10,5. Some prob-
lems occured with the current file of the
first anodic test, thus it was not possible to
0
5
10
15
20
25
-1 -0,5 0 0,5
Watercontactangle[]
Poten0al[V]
SSPBSaverage
SSPBS
Cathodicaverage
SSPBS
anodic
average
(a) SS contact angle in function of potential
0
5
10
15
20
25
30
-1,80 -0,80 0,20
Watercontactangle[]
Poten0al[V]
CoCrMoPBSaverage
LSVCoCrMo
PBScathodic
average
CoCrMoPBS
cathodic
stepsaverage
LSVCoCrMo
PBSanodic
average
(b) CoCrMo contact angle in function of potential
Figure 18: SS and CoCrMo PBS cathodic and
anodic average
present it in Fig. 15(b)
When looking at the cathodic curves,
both LSV and cathodic steps curves in
Fig. 13(a) and Fig. 14(a) respectively, it is
clear the global behavior is a diminutionof contact angle when there is diminution
of potential. For the LSV curves, the 2nd
test shows a different behavior than the
two other ones, because this test was per-
formed after an anodic test, without any
polishing in between the tests. The first
13
7/28/2019 Wetting and El. Properties of Biomed. Alloys
14/19
Semester project Autumn 2012 S.Barbotin
and 3rd tests show a decrease in contact an-
gle when reaching -1,0 V. If compared with
the cathodic steps curve, it can be stated
that the decrease in contact angle starts
later when performing the LSV measure-ments than when performing the cathodic
steps, for which it starts at around -0,7 V.
What needs to be said concerning the
first and 2nd curves of the cathodic steps,
is that the first test was performed chang-
ing the bubble between each steps, as writ-
ten in section 3.2.3, whereas the second test
was performed letting the same bubble in
place between each potential steps. The
two curves have however the same shape,contact angle decreasing at the same poten-
tial.
For both LSV and cathodic steps, the
range of contact angle evolution is of
102, except for test 2 in LSV for which
the range is of 40,5.
Thus, CoCrMo presents a good wettabil-
ity, final values ranging between 20-10.
Graphs showing together cathodic and
anodic contact angle behavior, using an
average of the different tests values forCoCrMo and SS in PBS are shown in
Fig. 18(a) and Fig. 18(b), so as to have a
global view.
IV.5 CoCrMo in PBS+albumin solu-tion
As well as for SS in PBS+albumin, the ini-
tial wettability of CoCrMo in PBS+albumin
is much higher than without albumin.
When going to the anodic domain inFig. 17(a), as for CoCrMo in PBS and as
expected, the contact angle doesnt change
and stays at 10,50,5. The test had to be
stopped at 0,4 V due to corrosion of the
sample.
Concerning the behavior when going
to the cathodic domain in Fig. 16(a), the
contact angle is again globally decreasing,
eventhough the three presented curves are
quite different. The first test shows a very
slight contact angle decrease, and had tobe stopped at -1,2 V because of hydrogen
evolution. A bigger decrease might have
been observed otherwise. For the 2nd test,
the contact angle starts increasing (1) then
decreases continuously when reaching -0,8
V up until -1,2 V. The 3rd curve presents
some fluctuations, but a final contact angle,
when reaching -1,2 V, much lower than at
OCP.
The range of contact angle variation hereis of 4,50,5 for cathodic steps, and of
only 10,5 for LSV. The contact angle
which is finally reached is very low : 3-
5, thus indicating a very good wettability
of the sample.
V. Discussion
CoCrMo has a contact angle range of vari-
ation smaller than SS, and this might affect
future possible applications in link with
potential (amplitude between 7 to 20 and
4 to 11 for SS in PBS and PBS+albumin
respectively, then 5 to 11 and 1 to 5 for
CoCrMo in PBS and PBS+albumin respec-
tively).
Concerning the addition of albumin, the
results in Fig.12, 16 and 17 state that the ini-
tial wettability is much higher than when
only immersed in PBS. It is because the
albumin molecules are adsorbed at the sur-face, thus modifying the surface chemistry
and so increasing the wettability.
About the anodic steps, for SS, the con-
tact angle seems to be fluctuating while
increasing the potential. This can be due to
two different reasons. The first one would
14
7/28/2019 Wetting and El. Properties of Biomed. Alloys
15/19
Semester project Autumn 2012 S.Barbotin
be that while increasing the potential in the
anodic domain, there is oxidation of the al-
loy, which change of oxidation state with
the increasing potential, thus leading to dif-
ferent wettability properties. The secondone would be that, because the bubble is
changed between each measurements, its
volume might varies of few L due to the
pipettes uncertainty, and so leading to con-
tact angle variations. Also, when injecting
the bubble, it happens the syringe nearly
touches the sample, thus maybe leading to
small imprecisions. Knowing that for the
CoCrMo sample the contact angle is very
stable along with the potential increase Fig.
15 and 17, the first hypothesis can seem
doubtfull. Since when increasing poten-
tial in the anodic domain for CoCrMo for-
mation of different chromium oxides take
place, it appears unusual that iron oxides
formation would influence the wettability
and not the one of different chromium ox-
ides. Otherwise, the different chromium
oxides have the same wettability. The rea-
son that for increases of potential to the
anodic domain the contact angle doesntchange, is that there is already formation
of a passive film at the sample surface be-
fore starting the measurements. Indeed,
the sample in contact with air already oxi-
dizes, and then when immersed in the elec-
trolyte, there is even more oxidation, thus
no change in the contact angle is observed.
The CV curve in Fig. 9(a) shows an in-
teresting behavior. When the potential de-
creases the contact angle decreases as well,but once the potential change direction and
starts increasing, the contact angle doesnt
increase up until it reaches its initial value.
It says constant, and this is because the
surface chemistry changed with the poten-
tial decrease, and thus wont be the same
anymore.
It is difficult to predict quantitatively the
contact angle behavior. On one hand, it
seems like the decrease in contact angle is
quantified (see V 1rst paragraph) but on
the other hand, the starting contact angle
is never the same, eventhough the OCP is
the same.
Also, it is interesting to remark that for
some tests, the wettability was so good
that the bubble wouldnt stick to the sam-
ple anymore. This case happened mostly
when albumin was added to the elec-
trolyte.
Finally, the obtained results show a qual-itative repeatability for a decrease in po-
tential for both SS and CoCrMo, with or
without albumin. For the increase of po-
tential, it would be necessary to perform
more tests, since some differences were ob-
served.
VI. Conclusion
An electrochemical cell was successfullydesigned, and allowed to perform good
contact angle measurements via the captive
bubble method. This experimental method
used to perform the tests, revealed itself to
be very good. Indeed, the obtained photos
are of nice quality, the contrast between the
bubble and the fluid being very clear.
Concerning the contact angle results, it
appears the same behaviour is observed
for SS and CoCrMo alloy. Increasing the
potential in the cathodic domain increaseswettability and so decreases contact angle.
The contact angle decrease can vary from
2 up to 20 in the PBS solution, and from
1 to 8 in the PBS+albumin solution, show-
ing that the quantitative behaviour cannot
be clearly defined. Since it is clear decreas-
15
7/28/2019 Wetting and El. Properties of Biomed. Alloys
16/19
Semester project Autumn 2012 S.Barbotin
ing the potential increases wettability, it
could lead to improvement of alloys sur-
face wettability in metallic hip prosthesis.
Future work should try to investigate
more precisely the influence of surface en-ergy on wear and friction behaviour and if
small increase in wettability (e.g. 51 con-
tact angle decrease in PBS+albumin) has or
not a significant effect on surface energy, as
well as investigating more the influence of
an increase in potential on wettability and
the quantitative behaviour between poten-
tial and wettability.
Acknowledgements
I would like to thank my supervisors Anna
Igual and Stefano Mischler for their contin-
uous help and patience. I also thank the
Tribology and Interfacial Chemistry Group
(TIC) and lATELIER IMX of EPFL for their
collaboration. I address a special thank to
Yann Barbotin and Pierre Mettraux who
took on their time to help me resolve all
the encountered problems.
References
[1] E. Rabinowicz, Influence of surface en-
ergy on friction and wear phenomena,
Journal of Applied Physics, vol. 32, no. 8,
pp. 14401444, 1961.
[2] D. Landolt, Corrosion et chimie de sur-
faces des mtaux. PPUR, 1997, vol. 12.
[3] A. Stalder, T. Melchior, M. Mller,
D. Sage, T. Blu, and M. Unser, Low-bond axisymmetric drop shape anal-
ysis for surface tension and contact
angle measurements of sessile drops,
Colloids and Surfaces A: Physicochemical
and Engineering Aspects, vol. 364, no. 1,
pp. 7281, 2010.
[4] A. Igual-Munoz and S. Mischler, Inter-
laboratory study on electrochemical
methods for the characterization of
cocrmo biomedical alloys in simulated
body fluids (efc 61), European Federa-tion of Corrosion, 2010.
16
7/28/2019 Wetting and El. Properties of Biomed. Alloys
17/19
Semester project Autumn 2012 S.Barbotin
Appendix
17
7/28/2019 Wetting and El. Properties of Biomed. Alloys
18/19
Semester project Autumn 2012 S.Barbotin
(a) Cell
18
7/28/2019 Wetting and El. Properties of Biomed. Alloys
19/19
Semester project Autumn 2012 S.Barbotin
(b) Lid
Figure 19: Technical drawings19