Upload
andyyangyanan
View
143
Download
8
Embed Size (px)
DESCRIPTION
Wing Phase Array test
Citation preview
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
Procedures
** On A/C ALL
Task 57-29-07-270-801-A-01Inspection of Upper Wing Skin between Trailing Edge and the Forward Fastener Row of the Rear and False Rear Spar
1. Task Supporting Data
A. Area of ApplicabilityWing Top Skin panel Trailing Edge in the following areas:- Area A (See FIGURE 57-29-07-991-001-A, Sheet 1)
Between the Trailing edge and the forward fastener row of the Rear Spar in Rib Bay 3-4 and Rib Bays9-10 thru 26-27
- Area B (See FIGURE 57-29-07-991-001-A, Sheet 2)Between the Trailing edge and the forward fastener row of the False Rear Spar
- Area C (See FIGURE 57-29-07-991-001-A, Sheet 3)Between the forward fastener row and aft fastener row of the Rear Spar in Rib Bays 4-5 and 5-6.
B. Description of Possible DamageCorrosion, that initiates at the edge of the Wing Top Skin and/or around the Rear Spar bolting, whichpenetrates the lower surface 10% of the material thickness with a 10 mm (0.394 in) diameter.
C. Reason for the JobISB A320-57-1154ISB A320-57-1155
D. Job Set-Up Information
(1) Tooling :
REFERENCE DESIGNATION
99D57204164000 REFERENCE BLOCK
AWP05-64-08-050 PROBE
OMNIPPA16128 ULTRASONIC EQUIPMENT OMNISCAN MX PAUT16-128/U8100029
(2) References :
SRM 51-75-11PB001SRM 57-21-11PB001ISB A320-57-1154ISB A320-57-1155FIGURE 57-29-07-991-001-A - Inspection Area A and Possible DamageFIGURE 57-29-07-991-002-A - Phased Array Rapid Scan AWP05-64-08-050FIGURE 57-29-07-991-003-A - Reference Block 99D57204164000FIGURE 57-29-07-991-004-A - Inspection Area A Scan PathsFIGURE 57-29-07-991-005-A - Inspection Area B Scan PathsFIGURE 57-29-07-991-006-A - Inspection Area C Scan PathsFIGURE 57-29-07-991-007-A - Screen Display Gate LocationsFIGURE 57-29-07-991-008-A - Calibration Screen Display for A-Scan, C-Scan A% (Amplitude Scan)and C-Scan A-I (Thickness Scan)FIGURE 57-29-07-991-009-A - Typical Tapered Skin Display ScreenFIGURE 57-29-07-991-010-A - Typical Skin Pocket Display Screen - Rib Bay 19-20 and 20-21FIGURE 57-29-07-991-011-A - Examples of Signal Loss caused by Skin Pockets, Tapered Skins andSteep Ramped Skin AreasFIGURE 57-29-07-991-012-A - Typical Skin Corrosion Display Screen
Page 1Nov 01/1057-29-07-PB4
Revision n: 86
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
2. Procedure
Subtask 57-29-07-271-001-A
A. Preparation for Inspection
(1) Any paint layer on the scan surfaces need not be removed provided it is in good condition with nobubbles or flakes. If there is any doubt about the condition of a paint layer in the scan areas, the paintlayer (excluding primer) must be removed using an approved method. (See SRM 51-75-11PB001)
(2) Make sure that all decals or other adhered substances are removed from the inspection area.
(3) Make sure that the inspection area is free from any visible damage or discontinuities.
(4) Use an approved marker to mark the centre line of Wing Ribs and Rear Spar.
(5) Identify the interface of Wing Ribs and Rear Spar to allow Rib Bay identification.
(6) Inspection Area A. (See FIGURE 57-29-07-991-004-A)
(a) Identify the Rib Bay to be inspected.
(b) Locate the first row of fasteners forward of the Rear Spar interface and identify the first and lastfastener in the row.
(c) Measure 20 mm (0.787 in) from the forward edge of the first fastener head of the fastener rowand mark the position with an approved marker.
(d) Measure 20 mm (0.787 in) from the forward edge of the last fastener head of the fastener rowand mark the position with an approved marker.
(e) Using a straight edge and an approved marker, draw a line between the marked positions Rib toRib.
(f) Measure 40 mm (1.575 in) scan pitch intervals aft from the marked line in LIST STRUCTUREtowards trailing edge of the skin. Using an approved marker mark the remaining scan paths.Make sure that the final scan path overlaps the trailing skin edge.
(g) Repeat steps LIST STRUCTURE thru or LIST STRUCTURE for each Rib Bay to be inspectedwithin Area A.
(7) Inspection Area B. (See FIGURE 57-29-07-991-005-A)
(a) Identify the False Rear Spar.
(b) Locate the first row of fasteners forward of the False Rear Spar and identify the first and lastfastener in the row.
(c) Measure 20 mm (0.787 in) from the forward edge of the first fastener head of the fastener rowand mark the position with an approved marker.
(d) Measure 20 mm (0.787 in) from the forward edge of the last fastener head of the fastener rowand mark the position with an approved marker.
(e) Using a straight edge and an approved marker, draw a line between the marked positions.
(f) Measure 40 mm (1.575 in) scan pitch intervals aft from the marked line in LIST STRUCTURE tothe trailing edge of the skin. Using an approved marker mark the remaining scan paths. Makesure that the final scan path overlaps the trailing skin edge.
(g) Repeat steps LIST STRUCTURE thru or LIST STRUCTURE for the complete False Rear Sparinterface.
(8) Identify Inspection Area C. (See FIGURE 57-29-07-991-006-A)
(a) Identify the Rib Bay to be inspected.
Page 2Nov 01/1057-29-07-PB4
Revision n: 86
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
(b) Locate the first row of fasteners forward of the Rear Spar interface and identify the first and lastfastener in the row.
(c) Measure 20 mm (0.787 in) from the forward edge of the first fastener head of the fastener rowand mark the position with an approved marker.
(d) Measure 20 mm (0.787 in) from the forward edge of the last fastener head of the fastener rowand mark the position with an approved marker.
(e) Using a straight edge and an approved marker, draw a line between the marked positions.
(f) Measure 40 mm (1.575 in) scan pitch intervals aft from the marked line in LIST STRUCTUREuntil the last scan path is at least 20 mm (0.787 in) aft from the aft row of fasteners on the RearSpar. Using an approved marker mark the remaining scan paths.
(g) Repeat steps LIST STRUCTURE thru or LIST STRUCTURE for each Rib Bay to be inspectedwithin Area C.
Subtask 57-29-07-272-001-A
B. Instrument Adjustment
(1) The equipment used in the development of this procedure is shown in TABLE 1.
Item Equipment
Instrument ULTRASONIC EQUIPMENT OMNISCAN MXPAUT16-128/U8100029 (OMNIPPA16128) with softwareversion MXU 2.0 R7 or higher
Probe PROBE (AWP05-64-08-050), 5MHz Phased Array RapidScan, 50 mm (1.969 in) wheel probe, 64 elements, 0.8element pitch, 5 m (16.40 ft) connection cable (SeeFIGURE 57-29-07-991-002-A)
Reference Block REFERENCE BLOCK (99D57204164000) (See FIGURE57-29-07-991-003-A)
Inspection Software CD Rom CD Rom NTMSA572907WPXXX supplied withAWP05-64-08-050 contains the OMNIPPA16128 softwarefile
Probe Guide Spirit Level 1 m (3.28 ft)
Couplant Water Spray
Measure Aid Tape Measure
Approved Marker EDDING 8404
Equipment and Materials
TABLE 1
NOTE : Any comparable ultrasonic instrument may be used provided that it satisfies the requirements ofthis procedure and is capable of resolving the reference defect in the reference block at therequired level of sensitivity.
NOTE : A PC that has a CD reader and a USB1 port will be required to load the files from PROBE(AWP05-64-08-050) CD Rom. ULTRASONIC EQUIPMENT OMNISCAN MXPAUT16-128/U8100029 (OMNIPPA16128) is fitted with a SD Card reader with USB1connector. The SD Card is located in ULTRASONIC EQUIPMENT OMNISCAN MXPAUT16-128/U8100029 (OMNIPPA16128).
Page 3Nov 01/1057-29-07-PB4
Revision n: 86
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
NOTE : The use of USB mouse improves the navigation of the ULTRASONIC EQUIPMENTOMNISCAN MX PAUT16-128/U8100029 (OMNIPPA16128) menu.
NOTE : CD Rom NTMSA572907WPXXX is supplied with PROBE (AWP05-64-08-050) when loanedfrom AIRBUS. If you do not have the CD Rom all of the required parameters contained withinthe procedure can be manually inputted. (See TABLE 3)
NOTE : Acquisition Module is only required if the operators own OMNISCAN has a different modulefitted.
(2) Connect the PROBE (AWP05-64-08-050) and Encoder connector to the ULTRASONIC EQUIPMENTOMNISCAN MX PAUT16-128/U8100029 (OMNIPPA16128).
(3) Connect the USB mouse if available.
(4) Preload SD Card with the files from CD ROM NTMSA572907WPXXX. Using ULTRASONICEQUIPMENT OMNISCAN MX PAUT16-128/U8100029 (OMNIPPA16128) SD Card Reader connectedvia a USB1 port to a PC with CD reader.
(5) Insert the flash card into the ULTRASONIC EQUIPMENT OMNISCAN MX PAUT16-128/U8100029(OMNIPPA16128).
(6) Make sure the PROBE (AWP05-64-08-050) wheel is full of water.
(7) Make sure that no air bubbles are present within the wheel water path.
(8) Tighten the adjusting nut on the PROBE (AWP05-64-08-050) wheel until the nut is approximately 3mm (0.118 in) from the end of the thread. (See FIGURE 57-29-07-991-002-A)
(9) Calibrate the Instrument in accordance manufacturers instructions and TABLE 2.
DESCRIPTION F1 KEY MENU F2 -- F6 KEYS SUBMENU
F7 -- F12 KEYSPARAMETER
INPUT VALUE
File File Open Setup >NTMSA572907.ops
Preferences Pref Units Millimeters
Display Selection Display A-C-[C]
Display Selection C-Scan 1 A %
Display Selection C-Scan 2 Thickness
Display Color Select Amplitude
Display Color Load Listed Files >ONDT_RFTOFD.pal
Display Color Start (%) 0
Display Color End (%) 100
Display Color Select Depth
Display Color Load PALSA572907.pal
Measurements Reading List 1
Measurements Reading Group Custom
Measurements Reading Field 1 A %
Measurements Reading Field 2 T (A - I)
Measurements Reading Field 3 None
General Settings
Measurements Reading Field 4 None
Page 4Nov 01/1057-29-07-PB4
Revision n: 86
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
Probe/ part Select Select Select Tx/Rx
Probe/ part Select Probe User>NDTS_5MHZ_WHEEL
Probe/ part Select Wedge User > Wheel
Probe/ part Select Auto Detect Off
Probe/ part Parts Geometry Plate
Probe/ part Parts Thickness (mm) 20
Probe/ part Parts Material Aluminium
Probe/ part Position Scan Offset (mm) 0.00
Probe / WedgeSettings
Probe/ part Position Skew (deg) 90.0
Focal Law Configuration Law Configuration Linear
Focal Law Aperture Elements Qty 8
Focal Law Aperture First Element 1
Focal Law Aperture Last Element 64
Focal Law Aperture Element Step 1
Focal Law Aperture Wave Type LW
Focal Law Beam Min Angle 0.0
Sector Scan settings
Focal Law Beam Focus Depth 50
UT Settings General Gain (dB) 15.0
UT Settings General Wedge Delay (s) 0.00
UT Settings General Velocity (m/s) 6300
UT Settings Pulser Tx/Rx Mode PE
UT Settings Pulser Voltage (V) 40
UT Settings Pulser PW (ns) Auto
UT Settings Pulser Max PRF Optimum
UT Settings Pulser Pulser 1
UT Settings Receiver Filter None
UT Settings Receiver Rectifier FW
UT Settings Receiver Video Filter Off
UT Settings Receiver Averaging 1
UT Settings Receiver Reject (%) 0
UT Settings
UT Settings Beam Beam Delay (s) 25.91 (orapproximate)
Initial Setup Settings
TABLE 2
(10) Set the instrument in accordance with CD Rom NTMSA572907WPXXX or TABLE 3.
DESCRIPTION F1 KEY MENU F2 -- F6 KEYS SUBMENU
F7 -- F12 KEYSPARAMETER
INPUT VALUE
Page 5Nov 01/1057-29-07-PB4
Revision n: 86
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
Scan Encoder Type Quad
Scan Encoder Origin 0.0
Scan Inspection Type One-Line Scan
Scan Inspection Scan Encoder 1
Scan Inspection Max. Scan Speed(mm/s)
84.00 (orapproximate)
Scan Area Scan Start (mm) 0.00
Scan Area Scan End (mm) 200
Scan Area Scan Resolution (mm) 0.80 (or approximate)
Encoder Settings
Scan Area Encoder Tolerance(mm)
0.00
Encoder Settings
TABLE 3
(11) Calibrate the Encoder.
(a) Open the WIZARD function and select an Encoder Type Calibration. (See TABLE 4)
DESCRIPTION F1 KEY MENU F2 -- F6 KEYS SUBMENU
F7 -- F12 KEYSPARAMETER
INPUT VALUE
Wizard Calibration Type EncoderEncoder Calibration
Wizard Calibration Select Calibration Start
Encoder Calibration 1
TABLE 4
(b) Select a INPUT VALUE start position Origin of 0.00 mm. (See TABLE 5)
DESCRIPTION F1 KEY MENU F2 -- F6 KEYS SUBMENU
F7 -- F12 KEYSPARAMETER
INPUT VALUE
N/A N/A Origin (mm) 0.00Encoder Calibration
N/A N/A Set Origin Next
Encoder Calibration 2
TABLE 5
(c) Select a INPUT VALUE of 50.00 mm. Rotate the wheel 50 mm (1.969 in) in the scanningdirection with the aid of a tape measure and select Calibrate. (See TABLE 6)
DESCRIPTION F1 KEY MENU F2 -- F6 KEYS SUBMENU
F7 -- F12 KEYSPARAMETER
INPUT VALUE
N/A N/A Distance (mm) 50.00Encoder Calibration
N/A N/A Set Distance Calibrate
Encoder Calibration 3
TABLE 6
(d) Rotate the wheel 50 mm (1.969 in) back to the start position of the scan and observe the upper
Page 6Nov 01/1057-29-07-PB4
Revision n: 86
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
left of the screen and make sure the Calibrated Axis (X) returns to 0.00 mm. (See TABLE 7)
DESCRIPTION F1 KEY MENU F2 -- F6 KEYS SUBMENU
F7 -- F12 KEYSPARAMETER
INPUT VALUE
Encoder Calibration N/A N/A Accept Accept
Encoder Calibration 4
TABLE 7
(e) If INPUT VALUE is:- Acceptable select ACCEPT- Unacceptable repeat steps LIST STRUCTURE thru or LIST STRUCTURE.
(12) Set the Interface Gate location.
(a) Put the PROBE (AWP05-64-08-050) on a flat surface.
(b) Set the Range to 31.5 mm (or approximate). (See TABLE 8)
DESCRIPTION F1 KEY MENU F2 -- F6 KEYS SUBMENU
F7 -- F12 KEYSPARAMETER
INPUT VALUE
Adjust A-Scan Display UT Settings General Range 31.5 mm (orapproximate)
A-Scan Range Settings
TABLE 8
(c) Apply hand pressure on the PROBE (AWP05-64-08-050) handle until the spring resistance is metand the PROBE (AWP05-64-08-050) wheel is in contact with the flat surface.
(d) Set the A-Scan Interface Signal start position in accordance with TABLE 9 and FIGURE57-29-07-991-007-A, Screen Display A.
DESCRIPTION F1 KEY MENU F2 -- F6 KEYS SUBMENU
F7 -- F12 KEYSPARAMETER
INPUT VALUE
Adjust A-Scan Display UT Settings General Start
A-Scan Start Settings
TABLE 9
(e) Set the Interface Gate start position in accordance with TABLE 10 and FIGURE57-29-07-991-007-A, Screen Display A.
DESCRIPTION F1 KEY MENU F2 -- F6 KEYS SUBMENU
F7 -- F12 KEYSPARAMETER
INPUT VALUE
Gate/Alarm Gate Gate I
Gate/Alarm Gate Parameters Position
Alter Interface Gate
Gate/Alarm Gate Start
Interface Gate Start Settings
TABLE 10
(f) Release hand pressure from the PROBE (AWP05-64-08-050) handle.
(g) Adjust the Interface Gate width to cover this signal in accordance with TABLE 11 and FIGURE
Page 7Nov 01/1057-29-07-PB4
Revision n: 86
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
57-29-07-991-007-A, Screen Display B.
DESCRIPTION F1 KEY MENU F2 -- F6 KEYS SUBMENU
F7 -- F12 KEYSPARAMETER
INPUT VALUE
Gate/Alarm Gate Gate I
Gate/Alarm Gate Parameters Position
Alter Interface Gate
Gate/Alarm Gate Width
Interface Gate Width Settings
TABLE 11
(h) Adjust the Interface Gate threshold to 20% Full Screen Height (FSH). (Refer to TABLE 12)
DESCRIPTION F1 KEY MENU F2 -- F6 KEYS SUBMENU
F7 -- F12 KEYSPARAMETER
INPUT VALUE
Gate/Alarm Gate Gate I
Gate/Alarm Gate Parameters Position
Alter Interface Gate
Gate/Alarm Gate Threshold 20
Interface Gate Threshold Settings
TABLE 12
(13) Select REFERENCE BLOCK (99D57204164000).
(14) Do a Zero-Degree Sensitivity Calibration away from spotfaces in accordance with the manufacturersinstructions with a Reference Amplitude of 80% FSH.
NOTE : During the Zero-Degree Sensitivity Calibration it is possible to identify any failing elementswithin the array.
(15) Do a TCG Calibration away from the spotfaces in accordance with manufacturers instructions on steps1, 2 and 3 of the REFERENCE BLOCK (99D57204164000) with a Reference Amplitude of 80% 5%.
(16) Set Gate A to a start position of 2 mm (or approximate) with a width of 20 mm and a threshold of25% FSH. (See TABLE 13)
DESCRIPTION F1 KEY MENU F2 -- F6 KEYS SUBMENU
F7 -- F12 KEYSPARAMETER
INPUT VALUE
Gate/Alarm Gate Gate A
Gate/Alarm Gate Parameters Position
Gate/Alarm Gate Start 2 (or approximate)
Gate/Alarm Gate Width 20
Gate Settings
Gate/Alarm Gate Threshold (%) 25
Gate A Settings
TABLE 13
(17) Put the PROBE (AWP05-64-08-050) away from spotfaces on Step 1 of REFERENCE BLOCK(99D57204164000) and apply hand pressure on the PROBE (AWP05-64-08-050) handle until the springresistance is met and establish a Back Wall Echo (BWE). Adjust the Gain to 80% FSH. (See TABLE
Page 8Nov 01/1057-29-07-PB4
Revision n: 86
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
14)
DESCRIPTION F1 KEY MENU F2 -- F6 KEYS SUBMENU
F7 -- F12 KEYSPARAMETER
INPUT VALUE
UT Settings UT Settings General Gain (dB)
Gain Settings
TABLE 14
(18) Do a scan over the spotfaces on the REFERENCE BLOCK (99D57204164000). When the scan iscomplete Pause the Screen Display. (See TABLE 15)
DESCRIPTION F1 KEY MENU F2 -- F6 KEYS SUBMENU
F7 -- F12 KEYSPARAMETER
INPUT VALUE
Starting the Scan Scan Start Start N/A
Pausing the Scan Scan Start Pause N/A
Starting and Pausing the Scan
TABLE 15
(19) Enter the thickness of the Step. (See TABLE 16)
DESCRIPTION F1 KEY MENU KEYS F7 -- F12 KEYSPARAMETER
INPUT VALUE
Measurements Thickness Source A - I
Measurements Thickness Min (mm) 0.05
Measurements Thickness Max (mm) (Enter ReferenceStandard StepThickness)
General Settings
Measurements Thickness Echo Qty 1
Thickness Measurement Settings
TABLE 16
(20) Make sure that both spotfaces are clearly displayed on both uncorrected C-Scan displays. (See FIGURE57-29-07-991-008-A)
(a) If spotfaces:- Cannot be displayed clearly check that max thickness reading is correct in accordance with
TABLE 16- Still cannot be clearly displayed repeat steps LIST STRUCTURE thru or LIST STRUCTURE.
(21) Turn off the Screen Display Pause. (See TABLE 17)
DESCRIPTION F1 KEY MENU KEYS F7 -- F12 KEYSPARAMETER
INPUT VALUE
Turn Pause Off Scan Start Pause N/A
Pause Settings
TABLE 17
(22) Put the PROBE (AWP05-64-08-050) away from spotfaces on Steps 2 and 3 of REFERENCE BLOCK
Page 9Nov 01/1057-29-07-PB4
Revision n: 86
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
(99D57204164000) and repeat steps LIST STRUCTURE thru or LIST STRUCTURE.
(23) Save the calibration file on the unit with a relevant filename. (See TABLE 18)
DESCRIPTION F1 KEY MENU KEYS F7 -- F12 KEYSPARAMETER
INPUT VALUE
Save Calibration Setupfile
File File Save Setup As File Name:Calibration_Setup
Save Calibration Setup File
TABLE 18
Subtask 57-29-07-273-001-A
C. Inspection
CAUTION : YOU MUST BE AWARE OF THE STRUCTURE CONFIGURATION IN THE INSPECTIONAREAS, INSPECTION AREAS WILL CONTAIN SKIN THICKNESS CHANGES, SKIN POCKETS,TAPERED SKINS AND RADII. STRUCTURAL CHANGES WILL AFFECT THE INSPECTIONAND SCREEN DISPLAYS. (SEE SRM 57-21-11PB001, FIGURE 57-29-09-010, FIGURE57-29-09-011 AND FIGURE 57-29-09-012).
(1) Open the Calibration Setup File. (See TABLE 19)
DESCRIPTION F1 KEY MENU F2 -- F6 KEYS SUBMENU
F7 -- F12 KEYSPARAMETER
INPUT VALUE
Save Calibration Setupfile
File File Open Open:Calibration_Setup
Open Calibration Setup File
TABLE 19
(2) Identify the inspection area.
(3) Measure the scan path distance of the inspection area and enter the scan path length + 50 mm inaccordance with TABLE 20.
DESCRIPTION F1 KEY MENU F2 -- F6 KEYS SUBMENU
F7 -- F12 KEYSPARAMETER
INPUT VALUE
Enter Scan PathLength
Scan Area Scan End (mm) (Scan Path Distance+ 50 mm)
Enter Scan Path Length
TABLE 20
(4) Put the Spirit Level on the Wing Skin as a PROBE (AWP05-64-08-050) wheel guide for the scanpaths.
(5) Spray Water onto the inspection area to aid coupling.
(6) Put the PROBE (AWP05-64-08-050) wheel at the start position of the scan and make sure that theindex of the PROBE (AWP05-64-08-050) wheel lines up with the marked scan path.
(7) Apply hand pressure to the PROBE (AWP05-64-08-050) wheel until spring resistance is met and adjustthe gain to place the BWE at 80% FSH.
Page 10Nov 01/1057-29-07-PB4
Revision n: 86
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
(8) Start the scan and rotate the PROBE (AWP05-64-08-050) wheel along the marked scan path, usingthe Spirit Level as a guide.
(9) Make sure that the index of the PROBE (AWP05-64-08-050) wheel is following the scan path.
(10) Complete the scan path and Pause the Screen Display. (See TABLE 21)
DESCRIPTION F1 KEY MENU F2 -- F6 KEYS SUBMENU
F7 -- F12 KEYSPARAMETER
INPUT VALUE
Starting the Scan Scan Start Start N/A
Pausing the Scan Scan Start Pause N/A
Starting and Pausing the Scan
TABLE 21
(11) Thickness readings must not be taken directly from the A-Scan Screen Display. Readings must be takenfrom the number in the field labelled as T(A - I) located in the top portion of the Screen Display.
(12) Using the USB mouse double click on the C-Scan A - I (Thickness Scan) display in ten evenly spacedrandom areas along the complete scan. Read off the obtained thickness readings and input themaximum thickness reading obtained in accordance with TABLE 22.
NOTE : Rib Bays and False Rear Spar Inspection Areas containing Tapered Skin thickness will beidentified on the C-Scan A - I (Thickness Scan) Screen Display as a gradual thickness change.(See FIGURE 57-29-07-991-009-A)
NOTE : Skin Pockets are located in Rib Bays 19-20 thru 22-23 and False Rear Spar Inspection Areas.Pockets will be displayed as a distinct shape on the C-Scan A% (Amplitude Scan) and athickness change and distinct shape on the C-Scan A - I (Thickness Scan) Screen Display.(See SRM 57-21-11PB001 and FIGURE 57-29-07-991-010-A, Sheets 1 and 2)
NOTE : False Rear Spar Inspection area contain Steep Ramped skin thickness. The Steep Ramped areascan cause a loss of signal to occur. (See FIGURE 57-29-07-991-011-A)
DESCRIPTION F1 KEY MENU F2 -- F6 KEYS SUBMENU
F7 -- F12 KEYSPARAMETER
INPUT VALUE
Inputting maximumpart thickness
Measurements Thickness Max. (mm) (Enter MaximumThickness across thescan)
Inputting Maximum Thickness
TABLE 22
(13) Analysis the C-Scan A% (Amplitude Scan) and C-Scan A - I (Thickness Scan) displays using thezoom function in accordance with manufacturers instructions and SRM 57-21-11PB001.
(14) Corrosion will be indicated by a reduction in the BWE reduction of 4dB (or 30% FSH) with acorresponding thickness change of 10% of the material thickness. (See FIGURE 57-29-07-991-012-A)
(15) Repeat steps LIST STRUCTURE thru or LIST STRUCTURE for all scan paths and all areas to beinspected.
Page 11Nov 01/1057-29-07-PB4
Revision n: 86
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
Subtask 57-29-07-971-001-A
D. Acceptance Criteria
(1) All indications that show a BWE reduction of 4dB (or 30% FSH) with a corresponding thicknessreduction 10% of the material thickness, which do not correspond the structural configuration givenin SRM 57-21-11PB001, must be classified as corrosion and recorded then reported to AIRBUS stating:(See FIGURE 57-29-07-991-012-A)- The location of the corrosion within the Rib Bay or False Rear Spar- The depth range of the corrosion- The size of the corrosion area- The scan must be saved as a DATA file for later referral. (See TABLE 23)
DESCRIPTION F1 KEY MENU F2 -- F6 KEYS SUBMENU
F7 -- F12 KEYSPARAMETER
INPUT VALUE
File File Save Mode Inspection Data
File File File Name Enter: File Name
Saving a DATA file
File File Save Data N/A
Save DATA Files
TABLE 23
Subtask 57-29-07-943-001-A
E. Final NDT Requirements
(1) Remove all inspection equipment, couplant and markings from the inspection area.
Page 12Nov 01/1057-29-07-PB4
Revision n: 86
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
** On A/C ALL
INSPECTION AREA ARIB BAY 3 4
INSPECTION AREA ARIB BAY 9 10 THRU 26 27
INSPECTIONAREA
POSSIBLECORROSION
TOP SKIN
REAR SPARTYPICALRECESSRADIUS
TYPICALN_NT_572907_1_0010101_01_00
A A
RIB27
RIB9
RIB3
RIB4
AA
Page 13Nov 01/10
FIGURE 57-29-07-991-001-A SHEET 01Inspection Area A and Possible Damage 57-29-07-PB4
Revision n: 86
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
** On A/C ALL
INSPECTION AREA BFALSE REAR SPAR
INSPECTIONAREA
POSSIBLECORROSION
TOP SKIN
FALSE REAR SPARTYPICALRECESSRADIUS
RIB9
RIB3
RIB4
N_NT_572907_1_0010102_01_00
B B
B B
Page 14Nov 01/10
FIGURE 57-29-07-991-001-A SHEET 02Inspection Area B and Possible Damage 57-29-07-PB4
Revision n: 86
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
** On A/C ALL
INSPECTION AREA CRIB BAY 4 5 AND 5 6
INSPECTIONAREA
POSSIBLECORROSION
REAR SPAR
TYPICALRECESSRADIUS
RIB4
RIB6 CC
C C
TOP SKIN
N_NT_572907_1_0010103_01_01
Page 15Nov 01/10
FIGURE 57-29-07-991-001-A SHEET 03Inspection Area C and Possible Damage 57-29-07-PB4
Revision n: 86
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
** On A/C ALL
NOTE: DIMENSIONS IN MILLIMETERS (INCHES IN BRACKETS)
140.0(5.51)
PROBE INDEX PROBE INDEX
SPRING ADJUSTMENTNUT
173.0(6.81)
151.0(5.95)
112.0(4.41)
PROBE INDEX
N_NT_572907_1_0020101_01_01
Page 16Nov 01/10
FIGURE 57-29-07-991-002-A SHEET 01Phased Array Rapid Scan AWP05-64-08-050 57-29-07-PB4
Revision n: 86
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
** On A/C ALL
15.0(0.59)20.0
(0.79)15.0
(0.59)15.0
(0.59)20.0
(0.79)
20.0(0.79)
15.0(0.59)15.0
(0.59)
15.0(0.59)
50.0(1.97)
4.0(0.16)
0.50(0.02)
1.0(0.04)
1.0(0.04)
2.00(0.08)
1.5(0.06)
3.0(0.12)
5.0(0.20)
10.0(0.40)
15.0(0.60)
200.0(7.87)
50.0(1.97)
50.0(1.97)
50.0(1.97)
150.0(5.91)
A
A
A A
NOTE:
ITEM NOMENCLATURE1 REFERENCE BLOCK
2 SIMULATEDCORROSION3 LABEL
QTY1
6
4 USE 3.0 (0.12) HIGH CHARACTERS10.0 (0.39) DIA SPOTFACES
AI 7150T651 OR SIMILAR
TO MINIMUM DEPTH
MATERIAL / REMARKS
PAINT ALL SURFACES WITH PRIMER (MATERIAL NO 16006)AND FINISH NONSPOTFACE SURFACES
AFTER PAINTING ENGRAVE PART NUMBER AND STEPIDENTIFICATIONS. USE 3.0 (0.12) HIGH CHARACTERS TOMINIMUM DEPTH. FILL IN TOP COAT (MATERIAL NO. 16018).
DIMENSIONS IN MILLIMETERS (INCHES IN BRACKETS)SPOTFACE MILLED TO A TOLERANCE OF +/ 0.05 (0.002)ALL OTHER TOLERANCES +/ 0.20 (0.008).
STEP 3
STEP 2
STEP 1
99D57204164000
ITEM 3
ITEM 1
ITEM 2
WITH TOP COAT (MATERIAL NO 16018).
N_NT_572907_1_0030101_01_02
Page 17Nov 01/10
FIGURE 57-29-07-991-003-A SHEET 01Reference Block 99D57204164000 57-29-07-PB4
Revision n: 86
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
** On A/C ALL
40.0
(1.57
)
20.0
(0.79
) 40.
0(1.
57)
SCAN
DIR
ECTI
ON
PRO
BE PO
SITI
ON
AT
ST
ART
OF
SCAN
, IND
EX P
OIN
T O
F PR
OBE
ON
THE
INTE
RFA
CE O
F TH
E RI
BCE
NTER
LINE
AND
THE
SCA
N PA
TH
RIB
CE
NTE
RLI
NE
RIB
CE
NTE
RLI
NE
FIR
ST S
CAN
PATH
20.
0 (0.
79)
FORW
ARD
FRO
M F
IRST
RO
WO
F BO
LTS
FORW
ARD
OF
REA
R S
PAR
REA
R S
PAR
BOLT
ING
FIN
AL SC
AN PA
THO
VERL
APPI
NG S
KIN
EDG
E
F W DIN
BD
NO
TE:
THE
DR
AWIN
G IS
NO
T TO
SCA
LE. T
HE A
CTUA
L NU
MBE
R O
F SC
AN P
ATHS
WIT
HIN
TH
E IN
SPEC
TIO
N AR
EA W
ILL
VARY
.D
IMEN
SIO
NS IN
MIL
LIM
ETER
S (IN
CHES
IN B
RACK
ETS)
.SC
AN P
ATH
(PRO
BE C
ENTE
RLIN
E)
EDG
E O
FSK
IN P
ANEL
X
Y
N_N
T_57
2907
_1_0
0401
01_0
1_01
Page 18Nov 01/10
FIGURE 57-29-07-991-004-A SHEET 01Inspection Area A Scan Paths 57-29-07-PB4
Revision n: 86
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
** On A/C ALL
40.0
(1.57
)
20.0
(0.79
) 40.
0(1.
57)
SCAN
DIR
ECTI
ON
PRO
BE PO
SITI
ON
AT
ST
ART
OF
SCAN
, IND
EX P
OIN
T O
F PR
OBE
ON
THE
INTE
RFA
CE O
F TH
E RI
BCE
NTER
LINE
AND
THE
SCA
N PA
TH
FIR
ST S
CAN
PATH
20.
0 (0.
79)
FORW
ARD
FRO
M F
IRST
RO
WO
F BO
LTS
FORW
ARD
OF
FALS
E RE
AR S
PAR
SPAR
BOLT
ING
NO
TE:
THE
DR
AWIN
G IS
NO
T TO
SCA
LE. T
HE A
CTUA
L NU
MBE
R O
F SC
AN P
ATHS
WIT
HIN
TH
E IN
SPEC
TIO
N AR
EA W
ILL
VARY
.D
IMEN
SIO
NS IN
MIL
LIM
ETER
S (IN
CHES
IN B
RACK
ETS)
.F W DIN
BD
FIN
AL SC
AN PA
THO
VERL
APPI
NG S
KIN
EDG
E
SCAN
PAT
H (P
ROBE
CEN
TERL
INE)
EDG
E O
F SK
IN PA
NEL
SCAN
PAT
H TO
CO
NTIN
UEUN
TIL
THE
LAST
BO
LTIN
G IN
THE
FALS
E R
EAR
SP
AR
FALS
E RE
AR
N_N
T_57
2907
_1_0
0501
01_0
1_01
Page 19Nov 01/10
FIGURE 57-29-07-991-005-A SHEET 01Inspection Area B Scan Paths 57-29-07-PB4
Revision n: 86
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
** On A/C ALL
40.0
(1.57
)
20.0
(0.79
) 40.
0(1.
57)
SCAN
DIR
ECTI
ON
PRO
BE PO
SITI
ON
AT
ST
ART
OF
SCAN
, IND
EX P
OIN
T O
F PR
OBE
ON
THE
INTE
RFA
CE O
F TH
E RI
BCE
NTER
LINE
AND
THE
SCA
N PA
TH
RIB
CE
NTE
RLI
NE
RIB
CE
NTE
RLI
NE
FIR
ST S
CAN
PATH
20.
0 (0.
79)
FORW
ARD
FRO
M F
IRST
RO
WO
F BO
LTS
FORW
ARD
OF
REA
R S
PAR
REA
R S
PAR
BOLT
ING
NO
TE:
THE
DR
AWIN
G IS
NO
T TO
SCA
LE. T
HE A
CTUA
L NU
MBE
R O
F SC
AN P
ATHS
WIT
HIN
TH
E IN
SPEC
TIO
N AR
EA W
ILL
VARY
.D
IMEN
SIO
NS IN
MIL
LIM
ETER
S (IN
CHES
IN B
RACK
ETS)
.
F W DIN
BD
FIN
AL S
CAN
PATH
AT
LEAS
T20
.0
(0.79
) AFT
O
F TH
E AF
T R
OW
OF
FAST
ENER
S O
N TH
ER
EAR
SPA
R
SCAN
PAT
H (P
ROBE
CEN
TERL
INE)
X
Y
N_N
T_57
2907
_1_0
0601
01_0
1_01
Page 20Nov 01/10
FIGURE 57-29-07-991-006-A SHEET 01Inspection Area C Scan Paths 57-29-07-PB4
Revision n: 86
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
** On A/C ALL
N_NT_572907_1_0070101_01_00
GATE B
INTERFACE SIGNAL GATE A
SCREEN DISPLAY A
INTERFACE GATE
SCREEN DISPLAY BGATE B
INTERFACE SIGNAL GATE A
INTERFACE GATE
NO PRESSURE
PRESSURE
Page 21Nov 01/10
FIGURE 57-29-07-991-007-A SHEET 01Screen Display Gate Locations 57-29-07-PB4
Revision n: 86
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
** On A/C ALL
APPROXIMATLEY 4dB ATTENUATION DIFFERENCE
20% THICKNESS CHANGE
10% THICKNESS CHANGE
N_NT_572907_1_0080101_01_00
Page 22Nov 01/10
FIGURE 57-29-07-991-008-A SHEET 01Calibration Screen Display for A-Scan, C-Scan A%
(Amplitude Scan) and C-Scan A-I (Thickness Scan)
57-29-07-PB4Revision n: 86
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
** On A/C ALL
SKIN THICKNESS TAPER
N_NT_572907_1_0090101_01_01
Page 23Nov 01/10
FIGURE 57-29-07-991-009-A SHEET 01Typical Tapered Skin Display Screen 57-29-07-PB4
Revision n: 86
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
** On A/C ALL
RIB BAY 2021 DISPLAY SCREEN
RIB BAY 1920 DISPLAY SCREEN
SKIN POCKETS
SKIN POCKETS
N_NT_572907_1_0100101_01_01
Page 24Nov 01/10
FIGURE 57-29-07-991-010-A SHEET 01Typical Skin Pocket Display Screen - Rib Bay 19-20 and
20-21
57-29-07-PB4Revision n: 86
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
** On A/C ALL
RIB BAY 2122 SCREEN DISPLAY
RIB BAY 2223 SCREEN DISPLAY
SKIN POCKETS
SKIN POCKETS
N_NT_572907_1_0100102_01_00
Page 25Nov 01/10
FIGURE 57-29-07-991-010-A SHEET 02Typical Skin Pocket Display Screen - Rib Bay 21-22 and
22-23
57-29-07-PB4Revision n: 86
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
** On A/C ALL
AREAS OF SIGNAL LOSSCAUSED BY STEEP
RAMPED AREAS
TAPERED SKINS SKIN POCKETS
N_NT_572907_1_0110101_01_01
Page 26Nov 01/10
FIGURE 57-29-07-991-011-A SHEET 01Examples of Signal Loss caused by Skin Pockets, Tapered
Skins and Steep Ramped Skin Areas
57-29-07-PB4Revision n: 86
@A318/A319/A320/A321NONDESTRUCTIVE TESTING MANUAL
** On A/C ALL
CORROSIONINDICATION CAUSING
AN ATTENUATION LOSS> 4dB
CORROSIONINDICATION CAUSINGMATERIAL LOSS > 10%
N_NT_572907_1_0120101_01_01
Page 27Nov 01/10
FIGURE 57-29-07-991-012-A SHEET 01Typical Skin Corrosion Display Screen 57-29-07-PB4
Revision n: 86