Upload
gus-umar
View
907
Download
51
Tags:
Embed Size (px)
Citation preview
7/15/2019 X-MET5000+5100 Quick Manual en Ver2
1/19
X-MET5000 /X-MET5100
Oxford Instruments Industrial Analysis
Email: [email protected]
www.oxford-instruments.com
1/19
Ver 2
October 2009
SW version 1.1 / 2.0 / 3.0
Quick Start Guide
X-MET5000 / X-MET5100
7/15/2019 X-MET5000+5100 Quick Manual en Ver2
2/19
X-MET5000 /X-MET5100
Oxford Instruments Industrial Analysis
Email: [email protected]
www.oxford-instruments.com
2/19
X-MET Radiation Safety
Please read the Users Manual carefully before operating the
X-MET analyzer. This quick start manual is a summary of the
actual users manual.
Radiation safety and user training is required before operating
the analyzer.
Follow all the safety instructions in Users Manual, such as:
Do not let any person closer than one meter from the
primary X-ray beam
Never point the radiation beam towards yourself or anyone
else.
Ensure that the analyzer is in touch with the sample all the
time while measuring.
Keep the instrument at a right-angle (90) against the
sample to minimize scattering radiation
While measuring small samples, use bench top stand,
background plate or radiation shield to minimize the
scattering radiation.
Protect the unit with password to prevent unauthorized use.
7/15/2019 X-MET5000+5100 Quick Manual en Ver2
3/19
X-MET5000 /X-MET5100
Oxford Instruments Industrial Analysis
Email: [email protected]
www.oxford-instruments.com
3/19
Radiation Protection
- Examples of potential misuse
Do not let any person closer than
one meter from the primary X-ray
beam
Do not point the analyzer at any
body part
Do not measure against any body
part
7/15/2019 X-MET5000+5100 Quick Manual en Ver2
4/19
X-MET5000 /X-MET5100
Oxford Instruments Industrial Analysis
Email: [email protected]
www.oxford-instruments.com
4/19
Do not use fingers or other body
parts to block the IR sensor. Use
safety shield when measuring
small samples.
Keep the instrument in touch with
the sample during the whole
measurement.
Keep the instrument at correct
angle (90) against the sample to
minimize scattering radiation.
When measuring repeatedly low
density samples such as plastic,
wood, paper, soil or minerals
safety shield should be used.
Radiation Protection
- Examples of potential misuse
7/15/2019 X-MET5000+5100 Quick Manual en Ver2
5/19
X-MET5000 /X-MET5100
Oxford Instruments Industrial Analysis
Email: [email protected]
www.oxford-instruments.com
5/19
About the battery
The battery is located inside the handle of the
X-MET unit. To replace the battery, push the switch and pull
the battery out of the handle.
Each fully charged battery gives
approximately 4 - 5 hours of operating time.
Change the battery when battery indicator on
the PDA screen turns to red (only one bar left).
There is also a battery indicator on the side of
the battery. Push and hold the button to check
battery status. If only one led will ignite, charge
the battery before starting the measurements.
Recharging the battery
Before removing the battery, exit X-MET
program and turn off the analyzer. You can
also charge without disconnecting it from the
X-MET.
Attach the battery to the charger. When the
Power LED is green, charger is operational.
Blinking green Charger LED indicates that the
battery is charging.
When the battery is fully charged, the Charger
LED turns to green.
Blinking red Charger LED indicates that either
the battery or the charger is faulty.
7/15/2019 X-MET5000+5100 Quick Manual en Ver2
6/19
X-MET5000 /X-MET5100
Oxford Instruments Industrial Analysis
Email: [email protected]
www.oxford-instruments.com
6/19
Basic Use of the Analyzer
Indicator lightsThe green light indicates that the power is on. The
green light turns to yellow when IR proximity
sensor sees sample in front of the analyzer nose.
When the red light is lit, x-rays are being
generated. There are three x-ray warning lights:
one on each side of the analyzer nose and one on
the back below PDA.
IR proximity sensor
To start the measurement the IR sensor must be
covered by the sample. This will prevent
accidental exposure to radiation. The green status
light will turn to orange when sample covers the
IR sensor.
Disconnecting PDA
Before disconnecting PDA exit X-MET programand turn off the unit from main power switch. To
disconnect PDA first remove the rubber frame and
possible rain cover. Then slide out PDA carefully.
Be careful and do not use force when connecting
PDA back to X-MET to prevent damage to the
connectors.
7/15/2019 X-MET5000+5100 Quick Manual en Ver2
7/19
X-MET5000 /X-MET5100
Oxford Instruments Industrial Analysis
Email: [email protected]
www.oxford-instruments.com
7/19
Measuring with optional light travel stand
1
2
3
4
5
ChargerConnect cables 3 and 5 directly to the charger,
then connect mains cable to AC outlet.
Light travel stand (optional accessory).
Remove the PDA and place X-MET firmly to the stand.
Data/power cable Connect one end to the charger1 and
other end to the X-MET battery
PDA chargerConnect one end to the mains outlet
and the other end to the small connector in cable 5.
PDA connection cable Connect one end to the charger
1 and other end to the PDA. The third connector (USB) will
not be used.
7/15/2019 X-MET5000+5100 Quick Manual en Ver2
8/19
X-MET5000 /X-MET5100
Oxford Instruments Industrial Analysis
Email: [email protected]
www.oxford-instruments.com
8/19
Switching the analyzer on and off
To switch on analyzer:
1. Push the power button until the green standby-light is lit. Notice that
the standby light turns to yellow when the proximity sensor detects a
sample in front of the analyzer nose.
2. Push the calendar button and the X-MET software will start. If the
software fails to initialize tap the Windows flag icon in top left cornerof the PDA screen and select X-MET
To switch off analyzer:
Push and hold the power button until you see that the progress-bar
on PDA screen is full. PDA will switch itself off automatically.
Power button
Calendar button
7/15/2019 X-MET5000+5100 Quick Manual en Ver2
9/19
X-MET5000 /X-MET5100
Oxford Instruments Industrial Analysis
Email: [email protected]
www.oxford-instruments.com
9/19
Starting up the Analyzer
Start up the Analyzer by pushing the power buttonuntil the status-light turn to green.
Start the X-MET operating software by pressing
the PDA Calendar button or:
Push the power key in top right corner of the
PDA.
Tap the START icon on the top left corner of
the PDA screen. (Figure 1)
Choose X-MET program from the menu.
X-MET start up screen appears.
Tap logon when the screen with warning sign
appears. If your unit is protected with password,
you must give it in the following screen. You can
change the password settings in Settings ->
User Setup.
The unit will automatically go to locked mode if it
havent been used for 5 minutes. To continue
measuring, password is required.
The measurement screen will appear. In this
mode the analyzer is ready to use. Tapping the
screen will open the main menu (figure 4).
1
2
3
4
7/15/2019 X-MET5000+5100 Quick Manual en Ver2
10/19
X-MET5000 /X-MET5100
Oxford Instruments Industrial Analysis
Email: [email protected]
www.oxford-instruments.com
10/19
Main menu Status bar
explanation of the symbols
1 2 3 4 5 6
1. Active Calibration method
2. If Result Format is activated, it is shown here
3. Log file is activated when the notepad symbol is shown.
4. This figure shows the current user level. Two human figures
mean normal user, one human figure means supervisor and
tool symbol means service level. The user level can be
changed in settings -> user setup menu.The yellow background indicates that the proximity sensor
detects a sample in front of the analyzer nose and the unit is
ready to measure.
5. The lightning sign sybolizes that instrument is in mains
connection.
6. Battery indicator shows the current charge status of the battery.
When only one red bar is left, it is recommended to shut down
the instruments and change the battery as soon as possible.
7/15/2019 X-MET5000+5100 Quick Manual en Ver2
11/19
X-MET5000 /X-MET5100
Oxford Instruments Industrial Analysis
Email: [email protected]
www.oxford-instruments.com
11/19
Selecting the Method
Screening Method
(Alloy mode / RoHS mode etc.)
Methods which are named as mode
are screening methods. Screening
method detects automatically thesample type and selects the correct
Empirical or FP method to calculate
analysis result.
Empirical Methods (Copper, Titanium etc.)
Empirical calibration method, which is
specific for certain sample type or alloy
group. Empirical calibration works
properly only with the specific sampletype it is calibrated.
FP Methods (Alloy_fp, Plastic_fp etc)
General standardless calibration
method, which works for wide range of
materials, like metal alloys or plastics.
The calibration methods in X-MET are customer and application specific so it is
possible that your analyzer does not have all the methods listed below. Proper
method is chosen from Select Method -menu before starting the measurement. Someof the methods are visible only in Supervisor User level. In most applications
Screening method should be used.
7/15/2019 X-MET5000+5100 Quick Manual en Ver2
12/19
X-MET5000 /X-MET5100
Oxford Instruments Industrial Analysis
Email: [email protected]
www.oxford-instruments.com
12/19
To change the measurement time, choose Set
Measurement Time from main menu.
When the measurement time is set to 0, the
measurement can be stopped any time by
releasing the trigger. The results will beconstantly updated on the result screen during
the measurement.
When value is given (other than 0) for
measurement time, the measurement will stop
automatically when the time is elapsed.
For metal alloys, the typical measurement
time is 2-20 seconds, depending on thedesired accuracy. Soil and plastic samples
usually require at least 60 seconds
measurement time for accurate results.
Setting the measurement time
7/15/2019 X-MET5000+5100 Quick Manual en Ver2
13/19
X-MET5000 /X-MET5100
Oxford Instruments Industrial Analysis
Email: [email protected]
www.oxford-instruments.com
13/19
Name the sample before starting the
measurement.
To name the sample, choose Name
Sample from the main menu.
The sample name (max. 20 characters) will
be saved in log file, which can be defined in
Output Settings section.
Pre-defined sample name can be selected
from drop down menu. 20 previous sample
names can be found from the list.
If space and number are placed after the
sample name, the program will start to index
the results automatically. (e.g. Test 1, Test 2,Test 3. etc.)
Maximum 80 characters of additional
information can be added on Additional
Information field. The information will also
be saved in the log file.
Up to 8 additional information fields can be
added.
The optional barcode scanne save the
barcode information in either name or
additional information field.
Naming the Sample
1
2
7/15/2019 X-MET5000+5100 Quick Manual en Ver2
14/19
X-MET5000 /X-MET5100
Oxford Instruments Industrial Analysis
Email: [email protected]
www.oxford-instruments.com
14/19
The analysis results can be stored
automatically to PDA main memory or
memory card.
Analysis results are saved to LOG file,
which is a text file, that can be opened
with almost any text editor (e.g. Microsoft
Excel, Notepad or optional Oxford
Instruments Report Generator software).
When LOG file option is activated, the file
name and all the results will be saved in
this file, until the file name will be changed.
Sample spectra can be automatically
saved by activating the Spectra directory
option and defining folder for spectra files.
This must be done before measuring the
sample.
Separate average log file is for averageresults. This log file will not include
individual results but only averages.
Average will also saved normal log file
after the individual results.
GPS log file will create a separate file
including GPS coordinates and the
analysis results. This file (.kml) can be
opened in a map program (e.g. Google
Earth)
Easiest way to transfer LOG file to the
external PC is to use memory card and
memory card reader which are included
with the X-MET5000 instrument.
Output settings
7/15/2019 X-MET5000+5100 Quick Manual en Ver2
15/19
X-MET5000 /X-MET5100
Oxford Instruments Industrial Analysis
Email: [email protected]
www.oxford-instruments.com
15/19
Result Screens
Information which is displayed on result screen
depend on various user interface settings and used
calibration method:
1 Default Result Screen
Analysis results in % or ppm
Optional standard deviation (STD) display
Grade information, grade limits shown on the right.
Hign and low alert colors are user defineable.
2 Average Result
When Average Mode is activated from the Settings
menu, average can be calculated to 2-50
measurement results.
(ReferX-MET5000 Users Manual for more
information)
3 Result Format -function
When customer or application specific Result Format
is activated from the Main Menu, color coded alarms
and messages can be displayed for fast decision
making. (ReferX-MET5000 Users manual for more
information).
1
2
3
7/15/2019 X-MET5000+5100 Quick Manual en Ver2
16/19
X-MET5000 /X-MET5100
Oxford Instruments Industrial Analysis
Email: [email protected]
www.oxford-instruments.com
16/19
Select appropriate method and measurement time.
Place the measurement window in direct angle against the sample.
Make sure that the sample covers the IR proximity sensor, otherwise
the measurement will not start.
Keep the trigger pulled during the whole measurement time.
If the measurement do not start, center the measurement
window and pull the trigger again.
In Timed Assay mode (see Users Manual) press and release
the trigger to start measurement. Measurement can be
stopped any time by quickly pressing and releasing the trigger
again.
Measuring the samples
7/15/2019 X-MET5000+5100 Quick Manual en Ver2
17/19
X-MET5000 /X-MET5100
Oxford Instruments Industrial Analysis
Email: [email protected]
www.oxford-instruments.com
17/19
Analyzing Metal Samples with X-MET
For metal alloys, the typical measurement time is 2-20 seconds, depending on
the desired accuracy. If necessary, check that the standard deviation (STD) is
adequate small.
Make sure that the sample is in the middle of the measurement window.
Cover the whole window with the sample when possible.
When measuring small samples, use longer measuring time than usual.
Pay attention to the possible coating of the sample. The penetration of the
radiation is typically less than 0.5 mm.
Rusty / unclean or painted surface may distort the analysis results. Clean the
surface before analyzing.
Ensure that the measurement window is clean. Metal dust in the window
might have a significant effect on the results.
7/15/2019 X-MET5000+5100 Quick Manual en Ver2
18/19
X-MET5000 /X-MET5100
Oxford Instruments Industrial Analysis
Email: [email protected]
www.oxford-instruments.com
18/19
Analyzing plastic, soil and other low density samples
with X-MET
For plastics and soil samples it is recommended to use at least 60 secondmeasurement time to ensure accurate results. Indicative screening resultscan be obtained with shorter measurement times. Timed Assaymode isrecommended in long measurement times, see Users manual for moreinformation.
Ensure that the sample covers the whole measurement window. In casethat sample is not uniform, notice that analysis result is average of themeasurement spot. It is recommended to use small sample cover,bench-top stand or background plate while analyzing plastic or other lowdensity samples.
It is recommended that the plastic samples should be at least 3 mm thickto reach the optimal accuracy. It is possible to measure thinner samples
(1 mm and up) for indicative screening analysis. It is recommended to pileup thinner samples if possible.
Make sure that both the measurement window and the sample are clean.Dirt might have a significant effect on the analysis results. Themeasurement window can be cleaned with small soft brush, cloth or bygently blowing the dirt away with pressurized air
7/15/2019 X-MET5000+5100 Quick Manual en Ver2
19/19
X-MET5000 /X-MET5100
Oxford Instruments Industrial Analysis
Email: [email protected]
www.oxford-instruments.com
Analyzing light elements with X-MET5100
Light element (Mg, Al, Si, P, S) analysis is much more sensitive to sample
preparation and conditions than heavier elements.
The following actions ensures quality analysis on light elements:
The measurement window must be clean and intact When measuring soil or mining samples the sample must be dry, ground to
small particle size and homogenized
Sample cup and thin film (e.g. 4 m Prolene) must be used. Plastic sample
bag is too thick and will block the weak signal coming from light elements.
Typical sample bag will block up to 97 % of incoming aluminum signal!
Metal samples must have clean and even surface. Jagged and dirty
surfaces will cause error on the results.
Use long enough measurement time (minimum 60 seconds)
Error in l ight element analysis wi l l af fect also th e heavy element resul ts.
Therefore i t is cr i t ical to p repare the samp le properly.