X-Ray & Computed Tomography (CT)

X-Ray & Computed Tomography (CT)

Embed Size (px)

DESCRIPTION

X-Ray & Computed Tomography (CT). Diagnostic X-Ray Imaging. BASIC PRINCIPLES OF X-RAY IMAGING In the simplest case, an X-ray imaging system requires • X-Ray Source • A patient to image • Film (image receptor) • Radiologist/Diagnostician. IDEAL DESCRIPTION OF IMAGING PROCESS - PowerPoint PPT Presentation

Citation preview

Page 1: X-Ray & Computed Tomography (CT)

X-Ray & Computed Tomography (CT)

Page 2: X-Ray & Computed Tomography (CT)
Page 3: X-Ray & Computed Tomography (CT)

Diagnostic X-Ray Imaging

BASIC PRINCIPLES OF X-RAY IMAGINGIn the simplest case, an X-ray imaging system

requires• X-Ray Source• A patient to image• Film (image receptor)• Radiologist/Diagnostician

Page 4: X-Ray & Computed Tomography (CT)

IDEAL DESCRIPTION OF IMAGING PROCESS• X-Rays are generated within the tube, and they are directed towards the

patient.• As the x-ray photons pass through the patient, some are absorbed, others

scattered, and some pass through the patient with no interaction. • The transmitted photons i.e., those which do not interact with the patient)

are detected (received) by the photon receptor, usually based around film.• The formation of an image on the film is dependent on the number of

photons which are captured (detected) by the receptor• Areas of the film which are dark have received a large number of photons;

brighter areas have received fewer. • The distribution of the light and dark areas on film is approximately a

projection onto a two dimensional map of the three-dimensional distribution of attenuating structures within the patient.

Page 5: X-Ray & Computed Tomography (CT)

There are many aspects which complicate the simplistic, ideal situation:

• Statistical arrival of photons (Poisson process).• Photon scatter.• Lines of projection are not parallel i.e., one has beam

divergence).• Photon detection is inefficient.• Beam hardening.• X-rays represent a form of ionizing radiation - there are

health risks associated with prolonged or repetitive exposures.

Page 6: X-Ray & Computed Tomography (CT)

NATURE OF X-RAYSX-rays represent electromagnetic radiation in the frequency range of about 1018-1020

Hz. At this frequency range, the free space wavelength is of the order ofλ = c/f = 3 × 108/1018 = 3 × 1010 m

At such dimensions of wavelength, the quantum nature of electromagnetic phenomena becomes significant and, it sets fundamental limits on imaging. The primary significance of this quantum nature is that the electromagnetic radiation is delivered in discrete lumps of energy, known as quanta, or photons.

The quantity of energy in each photon is related to the wavelength of the radiation according toE = hf = hc/λ

where h is Planck’s constant, 6.626 × 10−34 Js. Thus, the energy of 1 quantum of x-ray radiation (i.e., 1 X-ray photon) at 1020 Hz is

about6.626 × 10−34 × 1020J = 6.626 × 10−14J .

Page 7: X-Ray & Computed Tomography (CT)

NATURE OF X-RAYSInstead of quoting the photon energy in Joules, it is more common to use the

relationship1eV= 1.6 × 10−19 J to express the energy in eV’s.

Using this relation yields an energy of approximately 414keV.

Any description of photon arrival has to be treated in statistical terms. The arrival of X-ray photons at a detector may be treated as obeying Poisson statistics.

Page 8: X-Ray & Computed Tomography (CT)

X-Ray generation

“Braking” Radiation• The “Bremsstrahlung,” or “braking”

mechanism, is the principal mechanism of x-ray production in diagnostic radiology. Braking radiation is the radiation released as an electron is rapidly decelerated by a nucleus in a target material.

• It arises through bombarding a material of high atomic number nucleus with fast-moving electrons.

X-ray

Page 9: X-Ray & Computed Tomography (CT)

X-ray generation• In order to bombard a target of atoms with a stream of high energy

electrons, the arrangement shown in the figure below is usually used.

Page 10: X-Ray & Computed Tomography (CT)

Characteristic X-RaysIf we bombard a suitable target material with a stream of high energy

electrons, as we do in the generation of Bremsstrahlung radiation described above, we also experience another mechanism of x-ray production, arising from energy level transitions. The x-rays arising from this mechanism are known as characteristic x-rays. Characteristic x-rays are so called because they are very much dependent on the target material, i.e., the electronic configuration of the atom. For example, Tungsten anodes have characteristic lines at 69.5 and 59.3 keVenergies, Molybdenum anodes have lines at 20.0 and 17.3 keV.

• Characteristic x-rays are formed by electrons from the L and higher bands reverting to the K-shell, after a K shell electron has been ejected by the arrival of an electron of high kinetic energy.

• The wavelength of the photon produced by the L−→K or M−→K transition is dependent on the precise energy levels between the bands.

Page 11: X-Ray & Computed Tomography (CT)
Page 12: X-Ray & Computed Tomography (CT)
Page 13: X-Ray & Computed Tomography (CT)

X-RAY INTERACTIONS WITH MATTER

When a thin beam of X-ray photons passes through matter, it becomes weaker and is said to be attenuated as photons are removed from the forward direction of propagation. Attenuation takes place through the action of two processes:

• Scattering• Absorption

Page 14: X-Ray & Computed Tomography (CT)

Planar Xray imaging or radiography: a 2D projection• (shadow or silhouette) of a 3D body is produced on film by irradiating the body

with Xray photons.• Each ray of Xray photons is attenuated by a factor depending upon the integral of

the linear attenuation coefficient along the path of the ray, and produces a corresponding gray level (signal)at the point hit on the film or the detecting device used.

Page 15: X-Ray & Computed Tomography (CT)

Beer’s law or Beer–Lambert law

• Consider the ray path marked as AB in Figure. • Let Ni denote the number of X-ray photons incident

upon the body being imaged, within a specified time interval.

• Let No be the number of photons exiting the body. • The mutually parallel rays within the plane PQRS are

represented by the coordinates (t, s) that are at an angle θ with respect to the (x, y) coordinates, with the s axis being parallel to the rays.

• s= −x sin θ + ycosθ.

Page 16: X-Ray & Computed Tomography (CT)

• The use of monochromatic or monoenergetic X rays is assumed.• μ(x, y): linear attenuation coefficient at (x, y) in the sectional plane

PQRS.• μ(x, y) depends upon the density of the object and the frequency

(wavelength or energy) of the radiation used.

Page 17: X-Ray & Computed Tomography (CT)

When the rays are parallel to the x axis,we have θ= 90◦, s = −x, ds = −dx,and the planar image is given by

Page 18: X-Ray & Computed Tomography (CT)

2D Planar X-ray• A measurement of the exiting X rays (that is, No, and Ni for reference) thus

gives us only an integral of μ(x, y) over the ray.• The internal details of the body along the ray path are compressed onto a

single point on the film or a single measurement.• The radiographic image produced is a 2D planar image of the 3D object,

where the internal details are superimposed.• When the rays are parallel to the x axis, we haveθ = 90◦, s = −x, ds = −dx,and the planar image is given byg(y, z) =Z−μ(x, y, z) dx.equation

Page 19: X-Ray & Computed Tomography (CT)

Screen-film detector

• The X rays exiting from the body strike a fluorescent (phosphor) screen made of compounds of rare earth elements such as lanthanum oxybromide or gadolinium oxysulfide.

• The Xray photons are converted into visible light photons.• A light sensitive film in contact with the screen (in a light tight cassette)

records the result.• The film contains a layer of silverhalide emulsion with a thickness of about

10 μm.• The exposure or blackening of the film depends upon the number of light

photons that reach the film.• A thick screen provides a high efficiency of conversion of X rays to light, but

causes loss of resolution due to blurring.• The typical thickness of the phosphor layer in screens is in the range 40 −

100 μm.

Page 20: X-Ray & Computed Tomography (CT)
Page 21: X-Ray & Computed Tomography (CT)

• A fluoroscopy system uses an image intensifier and a video camera in place of the film to capture the image and display it on a monitor as a movie or video.

• Images are acquired at a rate of 2 − 8 frames/s (fps), with the Xray beam pulsed at 30 − 100 ms per frame.

• In computed radiography (CR), a photo-stimulable phosphor plate (made of europiumactivated barium fluorohalide) is used instead of film to capture and temporarily hold the image.

• The latent image is scanned using a laser and digitized.• In digital radiography (DR), the film or the entire screen-film combination

is replaced with solid state electronic detectors.

Page 22: X-Ray & Computed Tomography (CT)

• Examples: Figures show the posterioranterior(PA, that is, backtofront) and lateral (sidetoside) Xray images of the chest of a patient.

Page 23: X-Ray & Computed Tomography (CT)

Physical and technical considerations

• Target and focal spot: An electron beam with energy in the range of 20 − 140 keV is used to produce X rays for diagnostic imaging.

• Typical target materials used: tungsten and molybdenum.• Focal spot: area of the target struck by the electron beam to generate X

rays.• Nominal focal spot: diameter in mm as observed in the imaging plane (on

the film).• A small focal spot is desired in order to obtain a sharp image, especially in

magnification imaging.• Typical focal spot sizes in radiography: 0.1 − 2 mm.• A focal spot size of 0.1 − 0.3 mm is desired in mammography.

Page 24: X-Ray & Computed Tomography (CT)

Energy: • The penetrating capability of an Xray beam is mainly determined by the accelerating

voltage applied to the electron beam that impinges the target in the Xray generator.• Indicator of penetrating capability (the “energy” of the Xray beam): kV p, kilovolt

peak.• Higher kV p: more penetrating Xray beam.• The actual unit of energy of an Xray photon is the electron volt or eV , which is the

energy gained by an electron when a potential of 1 V is applied to it.• The kV p measure relates to the highest possible Xray photon energy that may be

achieved at the voltage used.• Lower energy Xray photons are absorbed at or near the skin surface, and do not

contribute to the image.• In order to prevent unwanted radiation, a filter is used at the Xray source to absorb

low-energy X rays.• Typical filter materials: aluminium and molybdenum.

Page 25: X-Ray & Computed Tomography (CT)

• Imaging of soft tissue organs such as the abdomen is performed with low-energy X rays in the range of 60 − 100 kV p.

• The use of a higher kV p would result in low differential attenuation and poor tissue detail visibility or contrast.

• A few other energy levels used in projection radiography are:– abdomen: 60 − 100 kV p;– chest: 80 − 120 kV p;– skull: 70 − 90 kV p.• The kV p to be used depends upon the distance between the Xray source

and the patient, the size (thickness) of the patient, the type of grid used, and several other factors.

Page 26: X-Ray & Computed Tomography (CT)

Exposure• For a given tube voltage (kV p), the total number of Xray photons released

at the source is related to the product of the tube current (mA) and the exposure time (s), together expressed as the product mAs.

• For a given body being imaged, the number of photons that arrive at the film is related to the mAs quantity.

• A low mAs results in an underexposed film (faint or light image), whereas a high mAs results in an overexposed or dark image (as well as increased Xray dose to the patient).

• Typical exposure values: 2 − 120 mAs.• Most imaging systems determine automatically the required exposure for a

given mode of imaging, patient size, and kV p.• Some systems use an initial exposure of the order of 5 ms to estimate the

penetration of the X rays through the body being imaged, and then determine the required exposure.

Page 27: X-Ray & Computed Tomography (CT)

BEAM HARDENING• Remember that a true x-ray beam will typically contain a range of photon energies

(range of wavelengths). • If one passes an x-ray beam through a slab of attenuating material (such as tissue),

then the photons of lower energy will experience a higher attenuation than the photons of higher energy. This means that, relative to the incident x-ray beam, there will be a greater proportion of photons with higher photon energies (higher frequencies, lower wavelengths) in the exit beam.

• This does not mean that the overall beam energy has increased: it just means that the distribution of photon energies is shifted towards higher energy photons.

• Beam hardening can cause a loss of image contrast if there is a highly attenuating material in the path of the beam, near to the entrance point into the body i.e., “early” in the photon path.

• The effect of beam hardening may be reduced by pre-filtering or pre-hardening the Xray beam and narrowing its spectrum.

• The use of monoenergetic X rays from a synchrotron or a laser obviates this problem.

Page 28: X-Ray & Computed Tomography (CT)

Scatter and the use of grids• As an Xray beam propagates through a body, photons are lost due to

absorption and scattering at each point in the body.• The angle of the scattered photon is a random variable.• A scattered photon contributes to noise at the point where it strikes the

detector.• Scattering results in the loss of contrast.• The noise effect of the scattered radiation is significant in gamma ray

emission imaging.• The effect of scatter may be reduced by the use of grids, collimation, or

energy discrimination.• Scattered (secondary) photons usually have lower energy levels than the

primary photons.

Page 29: X-Ray & Computed Tomography (CT)

Grid• Array of mutually parallel Xray absorbing strips if the X rays are in a

parallel beam (as in chest imaging), or converging toward the Xray source in the case of a diverging beam (as in breast imaging).

• Lattice or honeycomb grids with parallel strips in crisscross patterns are also used in mammography.

• Xray photons that arrive via a path that is not aligned with the grids will be stopped from reaching the detector.

Page 30: X-Ray & Computed Tomography (CT)
Page 31: X-Ray & Computed Tomography (CT)

• A typical grid contains thin strips of lead or aluminium with a strip density of 25 − 80 lines/cm and a grid height to strip width ratio in the range of 5 : 1 to 12 : 1.

• The space between the grids is filled with low attenuation material such as wood.

• A stationary grid produces a line pattern that is superimposed upon the image.

• Grid artifact is prevented in a reciprocating grid, where the grid is moved about 20 grid spacing during exposure: the movement smears the grid shadow and renders it invisible.

• Low levels of grid artifact may appear if the bucky does not move at a uniform pace or starts moving late or ends movement early with respect to the Xray exposure interval.

• Disadvantages: double radiation dose, reduced contrast.

Page 32: X-Ray & Computed Tomography (CT)
Page 33: X-Ray & Computed Tomography (CT)

Photon detection noiseInteraction between an Xray beam and a detector:• Photons lost due to scatter and absorption.• Some photons may pass through unaffected (or undetected).• The small size of the detectors in DR and CT imaging reduces their

detection efficiency.• Scattered and undetected photons cause noise.

Page 34: X-Ray & Computed Tomography (CT)

Ray stopping by heavy implants• Extremely heavy parts or components, such as metal screws or pins in

bones and surgical clips that are nearly Xray opaque and entirely stop the incoming Xray photons, can completely block an Xray beam.

• No photons would be detected at the corresponding point of exit from the body.

• The attenuation coefficient for the corresponding path would be indefinite, or within the computational context, infinity.

• A reconstruction algorithm would not be able to redistribute the attenuation values over the points along the corresponding ray path in the reconstructed image.

• This leads to streaking artifacts in CT images.

Page 35: X-Ray & Computed Tomography (CT)

Special techniques for enhanced Xray imaging:• digital subtraction angiography (DSA).• Dual energy imaging.

Page 36: X-Ray & Computed Tomography (CT)

Tomography

• Problem in X-ray: visualizing the details of the interior of the human body or other objects.

• Laminagraphy, planigraphy, or “classical” tomography used synchronous movement of the Xray source and film in such a way as to produce a relatively sharp image of a single focal plane of the object, with the images of all other planes being blurred.

• The smearing of information from the other planes of the object causes loss of contrast in the plane of interest.

• CT imaging made film based tomography obsolete.

Page 37: X-Ray & Computed Tomography (CT)
Page 38: X-Ray & Computed Tomography (CT)
Page 39: X-Ray & Computed Tomography (CT)

Computed tomography• The technique of CT imaging was developed during the late 1960s and the

early 1970s, producing images of cross sections of the human head and body as never seen before (noninvasively and non destructively!).

• In the simplest form of CT imaging, only the desired cross sectional plane of the body is irradiated using a finely collimated ray of Xray photons.

• Ray integrals are measured at many positions and angles around the body, scanning the body in the process.

• The principle of image reconstruction from projections, is then used to compute an image of a section of the body: hence the name computed tomography.

Page 40: X-Ray & Computed Tomography (CT)
Page 41: X-Ray & Computed Tomography (CT)
Page 42: X-Ray & Computed Tomography (CT)

• figure

Page 43: X-Ray & Computed Tomography (CT)

1st generation

• 300 projections: 3-4 minutes

Page 44: X-Ray & Computed Tomography (CT)

2nd generation

• 300 views: 20 seconds

Page 45: X-Ray & Computed Tomography (CT)

3rd generation

• 5 seconds per slice

Page 46: X-Ray & Computed Tomography (CT)

4th generation

• 4-5 seconds: improved image quality

Page 47: X-Ray & Computed Tomography (CT)

Spiral CT

Page 48: X-Ray & Computed Tomography (CT)
Page 49: X-Ray & Computed Tomography (CT)
Page 50: X-Ray & Computed Tomography (CT)

Single slice CT

Page 51: X-Ray & Computed Tomography (CT)

Single slice CT

Page 52: X-Ray & Computed Tomography (CT)

Single slice CT

Page 53: X-Ray & Computed Tomography (CT)

Back projection

Page 54: X-Ray & Computed Tomography (CT)
Page 55: X-Ray & Computed Tomography (CT)
Page 56: X-Ray & Computed Tomography (CT)
Page 57: X-Ray & Computed Tomography (CT)

Problems

• If a patient is exposed to x-ray radiation of • Compute the energy consumed by the patient.• If a rectangular object is exposed to planar CT

with parallel X-ray sources, draw the expected projections each 90 apart.

• If a circular object is exposed to planar CT with fan X-ray sources, draw the expected projections each 90 apart. Show the point exposed to maximum dosage.