Upload
amie-norma-hood
View
226
Download
0
Tags:
Embed Size (px)
Citation preview
Xiao-Yan Zhao
Beam Instrumentation GroupAccelerator Center, IHEP
BEPCII Background Issues: Beam Loss Measurement
BLM system Introduction Experiment Results Conclusion
Outline
BLM System Structure Detector Detector Location
BLM System Introduction
• The beam loss monitor (BLM) system is used for detecting or locating any possible excessive beam loss happened in the injection, or due to beam instabilities, or from bad vacuum, and so on. This system is a useful supplement to the beam diagnostic system for the troubleshooting.
The system structure of BEPCII BLM
The system consists of detector, front-end electronics, CAN bus and PC. The pulse signals from detectors are fed to front-end electronics. The CAN bus and Ethernet are used to connect all front-end electronics to the PC and BLM data to the central Control Room, respectively.
Det
ecto
r
Det
ecto
r
Det
ecto
r
Det
ecto
r
Det
ecto
r
Det
ecto
r
Det
ecto
r
Det
ecto
r
CAN Bus
PC Computer(CAN Card)
Control Room
Ethernet
Front-end Electronics
Front-end Electronics
Front-end Electronics
The detector is sensitive to MIPs (minimum ionizing particles) produced when an accelerated particle hits the wall of the vacuum chamber.
The detector consist of two PIN-diodes mounted face to face. An AND-gate detects the coincidence of pulses from the two PIN-diodes. MIPs cause ionizations in both PIN-diodes, a coincidence occurs and an output pulse is generated.
Photons do not cause ionization in both PIN-diodes. So, the detector is not sensitive to photons.
Maximum count rate: 10MHz
Detector (Bergoz Instrumentation Co.)
The detectors are installed in the inner side and outer side of the vacuum pipe near the downstream of the most of quadrupole magnets.
Total of 113 pairs, namely 226 detectors are installed in the BEPCII double-ring. Among them, 75 pairs locate in the outer ring and 38 pairs locate in the inner ring, respectively.
Detector location
The BLM layout in the IR
IP
e+ e-
56#: 6m from IP
57#: 2.1m from IP
58#: 0.26m from IP
63#: 6m from IP
62#: 2.1m from IP
61#: 0.26m from IP
Ring Collimator Transport Line Collimator Beam Abort Beams Colliding
Experiment Results
SR Injection, Current 0-257mA SR Run, E=2.5GeV
For Synchrotron Radiation mode, when the beam is injected or stored, the beam loss is small in the IR. So we only focus on the beam loss for the colliding mode.
Ring Collimator Experiment (2007.12.28)
The beam loss and lifetime are measured by adjusting the collimator with different aperture during the beam stored.
TimeIb(mA,10bucke
t)(e- beam)
R4OCH02 (8m upstream the IP) Hori. Colli. position
Outer(mm) Inner(mm)
14:15 - 14:20 99.9 - 98.2 42 4214:23 - 14:28 93 - 91.3 32(14) 32(14)14:33 - 14:38 89.2 - 87.8 31 3114:43 - 14:48 84.9 - 83.6 30 3014:52 - 14:57 82 - 80.8 29 2915:02 - 15:07 79 - 77.8 28 2815:12 - 15:17 75.8 - 74.7 27 2715:25 - 15:30 72.3 - 71.2 26(11.3) 26(11.3)
1
10
100
1000
10000
100000
0
1
2
3
4
5
6
7
I nsi deOutsi deLi f et i me
Change the collimator aperture, the beam lifetime reduces and the beam loss increases (58# :0.26m West from the I
P)
42mm(R4OCH02:inside and outside)
32mm31mm30mm 29mm
27mm28mm 26mm
After the aperture adjustment, the lifetime and beam loss recover quickly.
Similarly, the beam loss is measured by adjusting the collimator with different apertures during the beam injection.
1
10
100
1000
10000
100000
1000000
15:41:01
15:41:31
15:42:01
15:42:30
15:43:00
15:43:30
15:44:00
15:44:30
15:45:00
15:45:30
15:46:00
15:46:30
15:47:00
15:47:30
15:48:00
15:48:30
15:49:00
15:49:30
15:50:00
15:50:30
I nsi deOut si de
During the e- beam injection, the beam loss is large.But the change of beam loss is not obvious, even if the collimator aperture is different.
e-:0~89mA R4OCH02:42mm
beam aborte-:107~155mA,R4OCH02:26mm
e-:0~40mA R4OCH02:26mm
e-:88~106mA R4OCH02:32mm
TimeIb(mA)
(e- beam, Injection)
R4OCH02 Collimator position
Outer (mm) Inner (mm)
15:42:06- 15:42:35 0 - 89.2 42 4215:43:26-15:43:35 88 - 106.8 32(14) 32(14)15:43:39-15:43:56 107 - 155.4 26(11.3) 26(11.3)15:45:36-15:45:49 0-40 26(11.3) 26(11.3)15:47:59-15:48:54 0 - 206 26(11.3) 26(11.3)
e-:0~206mA R4OCH02:26mm
Transport line Collimator Experiment (2008.1.9)
The beam loss is measured for several times when beam is injected. We find the injection of e- beam brings more beam losses, so we focus on the injection of e- beam especially.
In order to reduce the background during e- beam injection, The collimators located on the transport line are used for the control of beam energy spread and emittance by adjusting the collimators aperture.
InjectionInjection
Run
In order to reduce the injection background, we change the aperture of energy spread collimator and emittance collimator.
Aperture 8x
Energy spread ±0.8%Aperture 3x
Energy spread ±0.3%
This picture shows the beam loss along the ring when e- beam is injected.
Aperture 8x
Energy spread ±0.8%Aperture 3x
Energy spread ±0.3%
BLM near IP
Energy spread ±0.8% aperture 8x
Energy spread ±0.3% aperture 3x
Inside Outside Inside Outside
6m West from IP(56#) 430,440 2,250,153 11,113 71,049
2.1m West from IP(57#) 464,906 39,238 20,144 3,329
0.26m West from IP(58#) 151,917 62,199 8,330 5,045
0.26m East from IP(61#) 116,090 112,953 6,235 7,706
2.1m East from IP(62#) 12,424 142,328 1,257 11,225
6m East from IP(63#) 69,774 87,094 12,525 11,656
This table shows the beam loss comparison when we inject e- beam with different collimators aperture.
The count of BLM in the IR reduces quickly after we set ±0.3% Energy spread and 3x aperture .
Beam Abort Experiment
The study of beam abort method is very important because there isn’t a special abort system in BEPCII.
beam abortbeam abort
The beam loss in the IR is large if we choose unfit methods when beam abort. So we optimize the method of beam abort by experiment
the old method of beam abort
the new method of beam abort
Beams Colliding
In order to reduce the background of the second way, e-
and e+ beam are separated at the IP by local orbit bump.
For the beam injection on the colliding mode, there are two ways to inject beam. One is to separate beam firstly during the injection by tuning the RF phase. Second way is to inject beam with colliding. But we found the second way bring more beam loss at the IP.
This picture shows the experiment result of injection with colliding e- and e+ beam at the IP.
e+:10mA, inject e-:0~10mA, Max beam loss≈5million/15s
This picture shows the experiment result which e- and e+ beam are separated at the IP by local orbit bump.
Inject e-
e-:398mA, inject e+:0~400mA Max beam loss≈95,000/15s
Abort e-
e+:340mA, inject e-:0~330mA, Max beam loss≈160,000/15s
1. The injection of e- beam brings more beam losses than the injection of e+ beam.
2. The ring collimator doesn’t show obvious effects to reduce the background. But the collimators on the transport line show clear effects to reduce the background during the electron beam injection.
3. The new beam abort method can obviously reduce the background.
4. If we use the injection way which separating the two beams by a local orbit bump, the background can be reduced.
5. For the SR mode, the beam loss in the IR is small.
Conclusion
Thanks!