27
Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

Embed Size (px)

Citation preview

Page 1: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

Xiao-Yan Zhao

Beam Instrumentation GroupAccelerator Center, IHEP

BEPCII Background Issues: Beam Loss Measurement

Page 2: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

BLM system Introduction Experiment Results Conclusion

Outline

Page 3: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

BLM System Structure Detector Detector Location

BLM System Introduction

Page 4: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

• The beam loss monitor (BLM) system is used for detecting or locating any possible excessive beam loss happened in the injection, or due to beam instabilities, or from bad vacuum, and so on. This system is a useful supplement to the beam diagnostic system for the troubleshooting.

Page 5: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

The system structure of BEPCII BLM

The system consists of detector, front-end electronics, CAN bus and PC. The pulse signals from detectors are fed to front-end electronics. The CAN bus and Ethernet are used to connect all front-end electronics to the PC and BLM data to the central Control Room, respectively.

Det

ecto

r

Det

ecto

r

Det

ecto

r

Det

ecto

r

Det

ecto

r

Det

ecto

r

Det

ecto

r

Det

ecto

r

CAN Bus

PC Computer(CAN Card)

Control Room

Ethernet

Front-end Electronics

Front-end Electronics

Front-end Electronics

Page 6: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

The detector is sensitive to MIPs (minimum ionizing particles) produced when an accelerated particle hits the wall of the vacuum chamber.

The detector consist of two PIN-diodes mounted face to face. An AND-gate detects the coincidence of pulses from the two PIN-diodes. MIPs cause ionizations in both PIN-diodes, a coincidence occurs and an output pulse is generated.

Photons do not cause ionization in both PIN-diodes. So, the detector is not sensitive to photons.

Maximum count rate: 10MHz

Detector (Bergoz Instrumentation Co.)

Page 7: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

The detectors are installed in the inner side and outer side of the vacuum pipe near the downstream of the most of quadrupole magnets.

Total of 113 pairs, namely 226 detectors are installed in the BEPCII double-ring. Among them, 75 pairs locate in the outer ring and 38 pairs locate in the inner ring, respectively.

Detector location

Page 8: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

The BLM layout in the IR

IP

e+ e-

56#: 6m from IP

57#: 2.1m from IP

58#: 0.26m from IP

63#: 6m from IP

62#: 2.1m from IP

61#: 0.26m from IP

Page 9: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

Ring Collimator Transport Line Collimator Beam Abort Beams Colliding

Experiment Results

Page 10: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

SR Injection, Current 0-257mA SR Run, E=2.5GeV

For Synchrotron Radiation mode, when the beam is injected or stored, the beam loss is small in the IR. So we only focus on the beam loss for the colliding mode.

Page 11: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

Ring Collimator Experiment (2007.12.28)

Page 12: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

The beam loss and lifetime are measured by adjusting the collimator with different aperture during the beam stored.

Page 13: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

TimeIb(mA,10bucke

t)(e- beam)

R4OCH02 (8m upstream the IP) Hori. Colli. position

Outer(mm) Inner(mm)

14:15 - 14:20 99.9 - 98.2 42 4214:23 - 14:28 93 - 91.3 32(14) 32(14)14:33 - 14:38 89.2 - 87.8 31 3114:43 - 14:48 84.9 - 83.6 30 3014:52 - 14:57 82 - 80.8 29 2915:02 - 15:07 79 - 77.8 28 2815:12 - 15:17 75.8 - 74.7 27 2715:25 - 15:30 72.3 - 71.2 26(11.3) 26(11.3)

1

10

100

1000

10000

100000

0

1

2

3

4

5

6

7

I nsi deOutsi deLi f et i me

Change the collimator aperture, the beam lifetime reduces and the beam loss increases (58# :0.26m West from the I

P)

42mm(R4OCH02:inside and outside)

32mm31mm30mm 29mm

27mm28mm 26mm

After the aperture adjustment, the lifetime and beam loss recover quickly.

Page 14: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

Similarly, the beam loss is measured by adjusting the collimator with different apertures during the beam injection.

Page 15: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

1

10

100

1000

10000

100000

1000000

15:41:01

15:41:31

15:42:01

15:42:30

15:43:00

15:43:30

15:44:00

15:44:30

15:45:00

15:45:30

15:46:00

15:46:30

15:47:00

15:47:30

15:48:00

15:48:30

15:49:00

15:49:30

15:50:00

15:50:30

I nsi deOut si de

During the e- beam injection, the beam loss is large.But the change of beam loss is not obvious, even if the collimator aperture is different.

e-:0~89mA R4OCH02:42mm

beam aborte-:107~155mA,R4OCH02:26mm

e-:0~40mA R4OCH02:26mm

e-:88~106mA R4OCH02:32mm

TimeIb(mA)

(e- beam, Injection)

R4OCH02 Collimator position

Outer (mm) Inner (mm)

15:42:06- 15:42:35 0 - 89.2 42 4215:43:26-15:43:35 88 - 106.8 32(14) 32(14)15:43:39-15:43:56 107 - 155.4 26(11.3) 26(11.3)15:45:36-15:45:49 0-40 26(11.3) 26(11.3)15:47:59-15:48:54 0 - 206 26(11.3) 26(11.3)

e-:0~206mA R4OCH02:26mm

Page 16: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

Transport line Collimator Experiment (2008.1.9)

The beam loss is measured for several times when beam is injected. We find the injection of e- beam brings more beam losses, so we focus on the injection of e- beam especially.

Page 17: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

In order to reduce the background during e- beam injection, The collimators located on the transport line are used for the control of beam energy spread and emittance by adjusting the collimators aperture.

Page 18: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

InjectionInjection

Run

In order to reduce the injection background, we change the aperture of energy spread collimator and emittance collimator.

Aperture 8x

Energy spread ±0.8%Aperture 3x

Energy spread ±0.3%

Page 19: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

This picture shows the beam loss along the ring when e- beam is injected.

Aperture 8x

Energy spread ±0.8%Aperture 3x

Energy spread ±0.3%

Page 20: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

BLM near IP

Energy spread ±0.8% aperture 8x

Energy spread ±0.3% aperture 3x

Inside Outside Inside Outside

6m West from IP(56#) 430,440 2,250,153 11,113 71,049

2.1m West from IP(57#) 464,906 39,238 20,144 3,329

0.26m West from IP(58#) 151,917 62,199 8,330 5,045

0.26m East from IP(61#) 116,090 112,953 6,235 7,706

2.1m East from IP(62#) 12,424 142,328 1,257 11,225

6m East from IP(63#) 69,774 87,094 12,525 11,656

This table shows the beam loss comparison when we inject e- beam with different collimators aperture.

The count of BLM in the IR reduces quickly after we set ±0.3% Energy spread and 3x aperture .

Page 21: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

Beam Abort Experiment

The study of beam abort method is very important because there isn’t a special abort system in BEPCII.

Page 22: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

beam abortbeam abort

The beam loss in the IR is large if we choose unfit methods when beam abort. So we optimize the method of beam abort by experiment

the old method of beam abort

the new method of beam abort

Page 23: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

Beams Colliding

In order to reduce the background of the second way, e-

and e+ beam are separated at the IP by local orbit bump.

For the beam injection on the colliding mode, there are two ways to inject beam. One is to separate beam firstly during the injection by tuning the RF phase. Second way is to inject beam with colliding. But we found the second way bring more beam loss at the IP.

Page 24: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

This picture shows the experiment result of injection with colliding e- and e+ beam at the IP.

e+:10mA, inject e-:0~10mA, Max beam loss≈5million/15s

Page 25: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

This picture shows the experiment result which e- and e+ beam are separated at the IP by local orbit bump.

Inject e-

e-:398mA, inject e+:0~400mA Max beam loss≈95,000/15s

Abort e-

e+:340mA, inject e-:0~330mA, Max beam loss≈160,000/15s

Page 26: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

1. The injection of e- beam brings more beam losses than the injection of e+ beam.

2. The ring collimator doesn’t show obvious effects to reduce the background. But the collimators on the transport line show clear effects to reduce the background during the electron beam injection.

3. The new beam abort method can obviously reduce the background.

4. If we use the injection way which separating the two beams by a local orbit bump, the background can be reduced.

5. For the SR mode, the beam loss in the IR is small.

Conclusion

Page 27: Xiao-Yan Zhao Beam Instrumentation Group Accelerator Center, IHEP BEPCII Background Issues: Beam Loss Measurement

Thanks!