MEMO SENS

www.knick.deLatest Product Information:

Stratos®Pro A2... MSOXYInstruction Manual

Page 2: Analizator ph, o2 knicks

2

WarrantyWarrantyDefects occurring within 3 years from delivery date shall be remedied free of charge at our plant (carriage and insurance paid by sender).Sensors, fittings, and accessories: 1 year.

Subject to change without notice.

Return of Products Under WarrantyPlease contact our Service Team before returning a defective device. Ship the cleaned device to the address you have been given. If the device has been in contact with process fluids, it must be decontaminated/disinfected before shipment. In that case, please attach a corresponding certificate, for the health and safety of our service personnel.

DisposalPlease observe the applicable local or national regulations concerning the disposal of “waste electrical and electronic equipment”.

Page 3: Analizator ph, o2 knicks

3

Documents Supplied

CD-ROMComplete documentation:• Instructionmanuals• Safetyinstructions• Shortinstructions

Safety InformationIn official EU languages and others.• ATEX/IECEX/FM/CSA• ECDeclarationsofConformity

Short InstructionsInGerman,English,French,Russian,Spanish, Portuguese, Swedish, and Dutch. MorelanguagesonCD-ROMand on our website: www.knick.de• Installationandcommissioning• Operation• Menustructure• Calibration• Errormessagesandrecommendedactions

Specific Test Report

Stratos® Pro A211 / A411

Other languages: www.knick.de

Kurzübersicht .............................. 15

QuickStart .................................... 27

Быстрый старт ........................... 39

Inicio rápido ................................ 51

Início rápido ................................ 63

Guida rapida ................................ 75

Short Instructions .........................3

Snabbstart.................................... 87

QuickStart .................................... 99

Stratos® Pro SeriesSafety Instructions

www.knick.de

Page 4: Analizator ph, o2 knicks

4

ContentsDocuments Supplied ..................................................................... 3Introduction .................................................................................... 7

Intended Use ....................................................................................................7

Safety Information ......................................................................... 8Overview of Stratos Pro A2... MSOXY .......................................10Assembly ........................................................................................11

Package Contents .........................................................................................11MountingPlan,Dimensions .....................................................................12PipeMounting,ProtectiveHood ............................................................13PanelMounting .............................................................................................14

Installation .....................................................................................15Installation Instructions .............................................................................15RatingPlates/TerminalAssignments...................................................16Wiring of Stratos Pro A2...MSOXY ..........................................................17MemosensCable ..........................................................................................18Wiring Examples ...........................................................................................19

User Interface, Keypad ................................................................20Display ............................................................................................21

Signal Colors (Display Backlighting) ......................................................21

Measuring Mode...........................................................................22Selecting the Mode / Entering Values ......................................23Operating Modes ..........................................................................24

MenuStructureofModesandFunctions ............................................25HOLDMode ....................................................................................................26Alarm .................................................................................................................27

Configuration ................................................................................28MenuStructureofConfiguration ...........................................................28ParameterSetA/B ........................................................................................30Configuration(OriginalforCopy) ...........................................................36

Page 5: Analizator ph, o2 knicks

5

ContentsSensor ...............................................................................................................38CurrentOutput1 ..........................................................................................42CurrentOutput2 ..........................................................................................48Correction .......................................................................................................50Alarm .................................................................................................................52Time and Date ...............................................................................................54Tag Number ...................................................................................................54

Digital Sensors ..............................................................................56Operation ........................................................................................................56Connecting a Digital Sensor .....................................................................57SensorReplacement ...................................................................................58

Calibration .....................................................................................60SelectingaCalibrationMode ..................................................................61Zero Calibration ............................................................................................62Product Calibration ......................................................................................64SlopeCalibration(Medium:Air) ..............................................................66Temp Probe Adjustment ............................................................................68

Measurement ................................................................................69Diagnostics ....................................................................................70Service ............................................................................................75Operating States ...........................................................................78Product Line and Accessories ....................................................79A2...X: Supply Units and Connection ........................................80Specifications ................................................................................81Error Handling ...............................................................................86Error Messages ..............................................................................87EC Declaration of Conformity ....................................................90Stratos Pro A2...X MSOXY: Control Drawings ..........................92

Page 6: Analizator ph, o2 knicks

6

ContentsFM Control Drawing .....................................................................93CSA Control Drawing ...................................................................94FDA 21 CFR Part 11 ......................................................................95

Electronic Signature – Passcodes ...........................................................95AuditTrail ........................................................................................................95

Index ...............................................................................................96Trademarks ..................................................................................................103

Passcodes .....................................................................................104

Page 7: Analizator ph, o2 knicks

7

IntroductionIntended UseTheStratosProA2...MSOXYisusedfordissolvedoxygenandtem-perature measurement in biotechnology, pharmaceutical industry, as well as in the field of industry, environment, food processing, and sewage treatment.The sturdy molded enclosure can be fixed into a control panel or mounted on a wall or at a post. The protective hood, which is avail-able as accessory, provides additional protection against direct weather exposure and mechanical damage.The device has been designed for digital standard sensors as well as fordigitaltracesensors(Option).Plain-textmessagesinalarge,backlitdisplayallowintuitiveopera-tion. The colored display backlighting signals alarm messages (red) or HOLDmode(orange).Dianostics functions are provided by “Sensocheck“ and “Sensoface“ monitoring and indication of the sensor condition. The internal log-bookcanhandleupto100entries–upto200withAuditTrail(TAN).The device provides two parameter sets which can be switched manually or via a control input for different process adaptations or different process conditions (e.g. beer and CIP).Password protection for granting access rights during operation can be configured.Twofloating,digitalcontrolinputs(“Hold“and“Control“)areavailablefor external control.The device provides two current outputs (for transmission of mea-sured value and temperature, for example).

Approvals for Measurement in Hazardous Locations:

Stratos Pro A2...N MSOXY: General Safety.

Stratos Pro A2...X MSOXY:ApprovedforoperationinhazardouslocationsaccordingtoIECEx/ATEX/FM/CSA.

Page 8: Analizator ph, o2 knicks

8

Safety InformationSafety information –Be sure to read and observe the following instructions!The device has been manufactured using state of the art technology and it complies with applicable safety regulations.When operating the device, certain conditions may nevertheless lead to danger for the operator or damage to the device.

See also separate document:“Safety Instructions“. •(ECDeclarationofConformity,FM,CSA,ATEX(ifapplicable) Certificates)

CAUTION!

Please note:Before commissioning it must be proved that the device may be connected with other equipment.

Commissioning must only be performed by trained personnel autho-rizedbytheoperatingcompany!Wheneveritislikelythatprotectionhas been impaired, the device shall be made inoperative and secured against unintended operation. The protection is likely to be impaired if, for example:

the device shows visible damage•the device fails to perform the intended measurements•after prolonged storage at temperatures above 70°C•after severe transport stresses•

Before recommissioning the device, a professional routine test must be performed. This test must be carried out at the manufacturer's factory.

Page 9: Analizator ph, o2 knicks

9

Safety Information

Terminals:Screw terminal, suitable for single wires / flexible leads up to 2.5 mm2 (AWG14).Recommendedtorquefortheterminalscrews:0.5...0.6Nm.

StratosProA2...Xratingplateatoutsidebottomoffrontwithcheckboxes for marking the respective application after installation

Information for Installation in Hazardous Locations(Stratos Pro A2...X MSOXY)

BesuretoobservethestipulationsofEN60079-10/EN60079-14•or the corresponding local regulations during installation and commissioning. See also separate “Safety Instructions“ document.

Approvals for application in hazardous locations(Stratos Pro A2...X MSOXY)

according to IECEx in Zone 0, 1, 20, 21•accordingtoATEXinZone0,1,2,20,21•according to • cCSAus in Class I Div 1, 2 / Zone 1accordingtoFMinClassI,Div1,2/Zone1•

When the device provides different types of protection, the operator must specify the applied type of protection during installation. To do so, use the checkboxes on the rating plate:

Important Notice: The operator must indicate the type of protection!

Memosens Ex wiring Uo (V)

Io (mA)

Po (mW)

Ci (μH)

Li (μH)

StratosProA2...X/A2...B 5 124 155 10.6 0

Page 10: Analizator ph, o2 knicks

10

OverviewOverview of Stratos Pro A2... MSOXY

17

9

8

5

6

13

14

Output1

Input +

Input –

CONTROL

CONTROL

+Output1,2/HART

–Output1/HART

–Output2

input

Control

10

11

HOLD

HOLD

input

Current

input

HOLD

Output2

1

2

3

4

RS485Power supply

RS485A

RS485B

GND/Shield

Page 11: Analizator ph, o2 knicks

11

AssemblyPackage ContentsChecktheshipmentfortransportdamageandcompleteness!The package should contain:Frontunit,rearunit,bagcontainingsmallparts•Specific test report•Documentation (cf Pg 3)•CD-ROM•

Fig.:Assemblingtheenclosure

Jumper (3 x)1) Washer (1 x), for conduit 2) mounting: Place washer between enclosure and nutCable tie (3 x)3) Hingepin(1x),insertable4) from either sideEnclosure screw (4 x)5)

Sealing insert (1 x)6) Rubberreducer(1x)7) Cable gland (3 x)8) Fillerplug(3x)9) Hexagonnut(5x)10) Sealing plug (2 x), for sealing 11) in case of wall mounting

11

10

9

8

7 6 5 4

1

2

3

Page 12: Analizator ph, o2 knicks

12

AssemblyMounting Plan, Dimensions

Fig.:Mountingplan(Alldimensionsinmm!)

Cable gland (3 x)1) Knockouts for cable gland or 2) ½" conduit, 21.5 mm dia. (2 knockouts) Conduitsnotincluded!Knockout for pipe mounting 3) (4 x)Knockout for wall mounting 4) (2 x)

148

148

42 4242

14

11741

21431

2

34 80

746.

2

4

3

Page 13: Analizator ph, o2 knicks

13

AssemblyPipe Mounting, Protective Hood

Fig.:ZU0737protectivehoodforwallandpipemounting (Alldimensionsinmm!)

Fig.:ZU0274pipe-mountkit(Alldimensionsinmm!)

147199

91

185

1

2

3

4

Hoseclampwithwormgear1) drive to DIN 3017 (2 x)Pipe-mountplate(1x)2) Forverticalorhorizontalposts3) or pipesSelf-tappingscrew(4x)4)

ø40...ø60

Page 14: Analizator ph, o2 knicks

14

AssemblyPanel Mounting

Fig.:ZU0738panel-mountkit(Alldimensionsinmm!)

Circumferential sealing 1) (1 x)Screw (4 x)2) Position of control panel3) Span piece (4 x)4) Threaded sleeve (4 x)5)

Cutout 138 x 138 mm (DIN 43700)

<30 76 31

1...221

2345

Page 15: Analizator ph, o2 knicks

15

InstallationInstallation Instructions

Installation of the device must be carried out by trained experts in •accordance with this instruction manual and as per applicable local and national codes.Be sure to observe the technical specifications and input ratings •duringinstallation!Besurenottonotchtheconductorwhenstrippingtheinsulation!•The supplied current must be galvanically isolated. If not, connect •an isolator module.Allparametersmustbesetbyasystemadministratorpriorto•commissioning!

Terminals: suitable for single wires / flexible leads up to 2.5 mm2 (AWG14)

Additionalsafetyprecautionshavetobetakenforapplications in hazardouslocationsaccordingtoIECEx,ATEX,FM,CSA! (See also separate “Safety Instructions“ document.)

Page 16: Analizator ph, o2 knicks

16

InstallationRating Plates / Terminal Assignments

Fig.:StratosProA2...Nouterratingplateatbottomoffront

Fig.:TerminalassignmentsofStratosProA2...

Fig.:StratosProA2...Xratingplateatoutsidebottomoffront

Important Notice: The operator must indicate the type of protection!When the device provides different types of protection, the operator must specify the applied type of protection during installation. To do so, use the checkboxes on the rating plate.See also “Safety Information“ chapter.

Fig.:Exampleofadditionalapprovalplate(cCSAus,FM)The specifications refer to the respective device.

Page 17: Analizator ph, o2 knicks

17

InstallationWiring of Stratos Pro A2... MSOXY

Fig.:Terminals, device opened, back of front unit

Terminal row 11 (BN) Power supply

2 (GN) RS485A

3(YE) RS485B

4(WH) GND/shield

5 + input

6 – input

7 PA(equip.bonding)

8 +out1,2/HART

9 –out1/HARTIn addition: 2HARTpins(betweenterminalrow1and2

Terminal row 210 hold

11 hold

12 n.c.

13 contr

14 contr

15 n.c.

16 n.c.

17 – out 2

18 n.c.

Areasforplacingthescrewdriver to pull out the terminals

1 9 10 18HART

SensorconnectionRS-485

Page 18: Analizator ph, o2 knicks

18

Memosens Cable

Power supply +

RS485(A)

RS485(B)

Power supply – (GND)

Shield

brown

green

yellow

white

transparent

Page 19: Analizator ph, o2 knicks

19

Wiring ExamplesExample 1:Measuringtask: Oxygen(standard)Sensors (example): SE706-NMSN(Knick)Cable (example): CA/MS-003NAA(Knick)

Sens

or(s

)Ca

ble

Dev

ice

1 2 3 4

3,3

V

RS 4

85 (A

)

RS 4

85 (B

)

GN

D/s

hiel

d

Yello

w

Whi

te

Shie

ld

(tra

nspa

rent

)

Brow

n

Gre

en

Page 20: Analizator ph, o2 knicks

20

User Interface, Keypad

Key Functionmeas Returntolastmenulevel•

Directly to measuring mode (press > 2 s)•info Retrieveinformation•

Show error messages•enter Configuration: Confirm entries, •

next configuration stepCalibration: •Continue program flowMeasuringmode:Displayoutputcurrent•

Arrow keysup / down

Measuringmode:Callmenu•Menu:Increase/decreaseanumeral•Menu:Selection•

Arrow keysleft / right

Measuringmode:Callmenu•Menu:Previous/nextmenugroup•Numberentry:Movebetweendigits•

1 IrDAtransmitter/receiver2 Display3 Keypad4 Ratingplate(bottom)

1

3

4

2

MEMO SENS

Page 21: Analizator ph, o2 knicks

21

Display

MEMO SENS

1 Temperature2 Sensocheck3 Interval/response time4 Sensor data5 Digital sensor devaluated6 Limit values7 Alarm8 Service9 ParametersetsA/B10 Calibration11 Digital sensor12 Waiting time running

13 Info available14HOLDmodeactive15Maindisplay16 Secondary display17 Proceed with enter18 Digital sensor19 Diagnostics20 Configuration mode21 Calibration mode22Measuringmode23 Sensoface24Measurementsymbol

1 2 3 4 5 6 7 8 9 10 11

171819202122

121314

15

16

2324

Signal Colors (Display Backlighting)

Red AlarmOrange HOLDmode(Calibration,Configuration,Service)Turquoise DiagnosticsGreen InfoPurple Sensoface message

Page 22: Analizator ph, o2 knicks

22

The device must be configured for the respective measurement task!

Measuring ModeAftertheoperatingvoltagehasbeenconnected,thedeviceautomati-callygoesto“Measuring“mode.Tocallthemeasuringmodefromanotheroperatingmode(e.g.Diagnostics,Service):Holdmeas key depressed (> 2 s).

Process variableTemperature

Sensoface indicator(sensor status)

Time

Modeindicator(measuring)

Holdmeas key depressed to call the measuring mode(pressing once more switches the display)

Active parameter set

enter key

In measuring mode the display indicates:Measuredvalueandtime(24/12hAM/PM)aswellastemperaturein•°Cor°F(formatsselectedduringconfiguration)

By pressing the meas key in measuring mode you can view the following displays (for approx. 60 sec):

MeasuredvalueandselectionofparametersetA/B •(ifsetto“Manual“)Measuredvalueandtag(pointofmeasurementdesignation •– entered during configuration)Time and date•

Pressing the enter key shows the output currents. They are displayed as long as enterishelddepressed,thenthemeasured-valuedisplaywill return after 3 sec.

Page 23: Analizator ph, o2 knicks

23

Selecting the Mode / Entering Values

To enter a value:Select numeral: left / right arrow key5) Change numeral: up / down arrow key6) Confirm entry with 7) enter

To select the operating mode:Hold1) meas key depressed (> 2 s) (measuring mode)Press any arrow key: the selection menu appears2) Select operating mode using left / right arrow key3) Press 4) enter to confirm the selected mode

Selection menu

Selected mode (blinks)

1 2

3

4

5 6

7

Page 24: Analizator ph, o2 knicks

24

Operating ModesDiagnosticsDisplay of calibration data, display of sensor data, performing a device self-test,viewingthelogbookentries,displayofhardware/softwarever-sions of the individual components. The logbook can store 100 events (00...99). They can be displayed directly on the device. The logbook can beextendedto200entriesusingaTAN(Option).

HOLDManualactivationofHOLDmode,e.g.forreplacingadigitalsensor. The signal outputs adopt a defined state.

CalibrationEvery sensor has typical characteristic values, which change in the course of the operating time. Calibration is required to supply a cor-rect measured value. The device checks which value the sensor delivers when measuring in a known solution. When there is a deviation, the de-vice can be “adjusted“. In that case, the device displays the “actual“ value and internally corrects the measurement error of the sensor. Calibration must be repeated at regular intervals. The time between the calibration cycles depends on the load on the sensor. During calibration the device isinHOLDmode.During calibration the analyzer remains in the HOLD mode until it is stopped by the operator.

ConfigurationTheanalyzermustbeconfiguredfortherespectivemeasurementtask.In the “Configuration“ mode you select the connected sensor, the mea-surement range to be transmitted, and the conditions for warning and alarmmessages.DuringconfigurationthedeviceisinHOLDmode.Configuration mode is automatically exited 20 minutes after the last keystroke. The device returns to measuring mode.

ServiceMaintenancefunctions(monitor,currentsource),IrDAoperation,pass-codeassignment,resettofactorysettings,enablingofoptions(TAN).

Page 25: Analizator ph, o2 knicks

25

Menu Structure of Modes and Functions

Measuringmode

TAGdisplay

Display of calibration data

Display of sensor data

Selftest:RAM,ROM,EEPROM,module

Logbook: 100 events with date and time

Display of direct, uncorrected sensor signals

Display of software version, model designation, serial number

ManualactivationofHOLDmode,e.g.forsensorreplacement.Thesignaloutputsbehaveasconfigured(e.g.lastmeasuredvalue,21mA)

Calibration in water/air (as configured)

Zero calibration

Product calibration

Adjustmentoftemperatureprobe

ConfiguringparametersetA

Configuring parameter set B

Display of measured values for validation (simulators)

Current source, output 1

Current source, output 2

ActivatingtheIrDAinterface

Specifying access codes for operating modes

Resettofactorysetting

EnablinganoptionviaTAN

Pressing any arrow key opens the selection menu. Select the menu group using the left/right arrow keys. Press enter to open a menu. Press meas to return.

(Accessviacode, factory setting:5555)

measCLK display

meas meas

after 60 s after 60 s

Page 26: Analizator ph, o2 knicks

26

HOLD ModeTheHOLDmodeisasafetystateduringconfigurationandcalibration.Outputcurrentisfrozen(Last)orsettoafixedvalue(Fix).TheHOLDmodeisindicatedbyorangedisplaybacklighting.

Terminating the HOLD Mode TheHOLDmodeisendedbyswitchingtomeasuringmode(holdmeas key depressed). The display reads “Good Bye“, after that, the HOLDmodeisexited.When the calibration mode is exited, a confirmation prompt ensures that the installation is ready for operation (e.g.: sensor reinstalled, located in process).

HOLD mode, display icon:

HOLD active

21

4HOLD active

Outputcurrent[mA]

OutputsignalHOLDSettingFIX=21.0mA

OutputsignalHOLDSettingLAST

Output Signal During HOLD:

Output Signal ResponseLast:• Theoutputcurrentisfrozenatitslastvalue.Recommendedfor short configuration procedures. The process should not change decisively during configuration. Changes are not noticed with this setting!Fix:• The output current is set to a value that is noticeably different from the process value to signal the control system that the device is being worked at.

Page 27: Analizator ph, o2 knicks

27

Alarm

Manual Activation of HOLDTheHOLDmodecanbeactivatedmanuallyfromtheHOLDmenu.Thisallows checking or replacing a sensor, for example, without provoking unintended reactions of outputs or contacts.Press meas key to return to selection menu.

AlarmWhen an error has occurred, Err xx is displayed immediately.Onlyafterexpiryofauser-defineddelaytimewillthealarmberegis-tered and entered in the logbook.During an alarm the display blinks, the display backlighting turns red.

Errormessagescanalsobesignaledbya22mAoutputcurrent(seeConfiguration).2 sec after the failure event is corrected, the alarm status will be deleted.

External Activation of HOLDTheHOLDmodecanbeactivatedfromoutsidebysendingasignaltotheHoldinput(e.g.fromtheprocesscontrolsystem).

10

11

Power supply12...24VAC/DC

StratosProA2...

input

HOLD

HOLDinactive 0...2VAC/DCHOLDactive 10...30VAC/DC

Process control system (PCS)

Page 28: Analizator ph, o2 knicks

28

ConfigurationMenu Structure of ConfigurationThedeviceprovides2parametersets“A“and“B“.Byswitching between the parameter sets you can adapt the device to different measurement situations, for example. Parameterset“B“onlypermitssettingofprocess-relatedparameters.The configuration steps are assigned to different menu groups. Using and , you can jump between the individual menu groups.Each menu group contains menu items for setting the parameters. Pressing enter opens a menu item. The values are edited using and . Pressing enter confirms/stores the settings. Returntomeasurement:Pressmeas.

Select menu group Menu group Code Display Select

menu item

Sensor selection SNS:

Menuitem1

...

Menuitem...

Current output 1 OT1:

Current output 2 OT2:

Compensation COR:

Alarmmode ALA:

Setting the clock CLK:

Tag number TAG:

enter

enter

enter

enter

Page 29: Analizator ph, o2 knicks

29

ConfigurationParameter Set A/B: Configurable Menu Groups

(SomeparametersareidenticalinAandB.TheyareconfiguredinparametersetAonly.)

Menu group Parameter set A Parameter set BSENSOR Sensor selection ---OUT1 Current output 1 Current output 1OUT2 Current output 2 Current output 2CORRECTION Compensation CompensationALARM Alarm mode Alarm modePARSET Parameter set

selection---

CLOCK Setting the clock ---TAG Tag number ---

External Switchover of Parameter Sets A/BYou can switch between parameter sets A and B by applying a signal to the CONTROL input. For confi guration, see Pg 35.

13

14

e.g. process control system

Max. 30 V AC/DC

Stratos Pro A2...

input

Control

Parameter set A active 0...2 V AC/DCParameter set B active 10...30 V AC/DC

Page 30: Analizator ph, o2 knicks

30

Parameter Set A/BManual selection

Display Action RemarkTo switch between parameter sets:Press meas

Manualselectionofparameter sets must have beenpresetinCONFIGmode. Default setting is a fixedparametersetA.Wrong settings change the measurement proper-ties!

MEMO SENS

PARSETblinksinthelower line.Select parameter set using and keys

MEMO SENS

Select PARSETA/PARSETB

Confirm with enterCancel with meas

Configuration

Page 31: Analizator ph, o2 knicks

31

Configuration

Configuration Choices DefaultSensor (SENSOR)SNS: (Select text line) MEMOSENS

STANDARD*)

TRACES*)(TAN)

MEMOSENS

*)Thesesensorsappearinthemenuselectionbutcanonlybeusedwithameasuringmoduleinstalled.StratosProA2...MSOXYisintendedforconnectingaMemosenssensorviaRS-485interface. It does not provide a measuring module.Forinformationonretrofittingandtherespectivecosts,please contact the manufacturer (see back of this manual).

MEASMODE dO%dOmg/ldOppmGAS%

dO%

TEMPUNIT °C/°F °C

CALMODE CALAIRCALWTR

CALAIR

CALTIMER ON/OFF OFF

ON CAL-CYCLE 0...9999 h 0168 h

Page 32: Analizator ph, o2 knicks

32

Configuration

Output 1 (OUT1, no trace measurement)OT1: CHANNEL OXY/TMP OXY

OXYdO%

BEGIN4mA (0mA)

000.0...600.0% 000.0%

END20mA 0.000...600.0% 600.0%

OXYdOmg/l

BEGIN4mA (0mA)

00.00...99.99 mg/l

END20mA 00.00...99.99 mg/l

OXYdOppm

BEGIN4mA (0mA)

00.00...99.99 ppm

END20mA 00.00...99.99 ppm

OXYGAS%

BEGIN4mA (0mA)

00.00...99.99%

END20mA 00.00...99.99%

TMP°C BEGIN4mA (0mA)

–20...150 °C

END20mA –20...150 °C

TMP°F BEGIN4mA (0mA)

–4...302°F

END20mA –4...302°F

FILTERTIME 0...120 SEC 0000 SEC

22mA-FAIL ON/OFF OFF

HOLDMODE LAST/FIX LAST

FIX HOLD-FIX 4...22mA 021.0mA

Configuration Choices Default

Page 33: Analizator ph, o2 knicks

33

Output 1 (OUT1, trace measurement)OT1: CHANNEL OXY/TMP OXY

OXYdO%

BEGIN4mA (0mA)

000.0...600.0% 000.0%

END20mA 000.0...600.0% 600.0%

OXYdOmg/l

BEGIN4mA (0mA)

000.0...99.00 mg/l

END20mA 000.0...99.00 mg/l

OXYdOppm

BEGIN4mA (0mA)

00.00...99.99 ppm

END20mA 00.00...99.99 ppm

OXYGAS%

BEGIN4mA (0mA)

0000 ppm... 50.00%

END20mA 0000 ppm... 50.00%

TMP°C BEGIN4mA (0mA)

–20...150 °C

END20mA –20...150 °C

TMP°F BEGIN4mA (0mA)

–4...302°F

END20mA –4...302°F

FILTERTIME 0...120 SEC 0000 SEC

22mA-FAIL ON/OFF OFF

HOLDMODE LAST/FIX LAST

FIX HOLD-FIX 4...22mA 021.0mA

Configuration

Configuration Choices Default

Page 34: Analizator ph, o2 knicks

34

Configuration

Configuration Choices DefaultOutput 2 (OUT2)OT2: CHANNEL OXY/TMP TMP

... other steps like output 1

Temperature compensation (CORRECTION)COR: SALINITY 00.00...45.00 ppt 00.00 ppt

PRESSUREUNIT BAR/KPA/PSI BAR

PRESSURE MAN/EXT*

MAN BAR 0.000...9.999BAR 1.013BAR

KPA 000.0...999.9KPA

PSI 000.0...145.0 PSI

EXT I-Input 4(0)...20mA 4...20mA

BAR BEGIN4mA (0mA)

0.000...9.999BAR 0.000BAR

END20mA 0.000...9.999BAR 9.999BAR

KPA BEGIN4mA (0mA)

000.0...999.9KPA

END20mA 000.0...999.9KPA

PSI BEGIN4mA (0mA)

000.0...145.0 PSI

END20mA 000.0...145.0 PSI

Alarm (ALARM)ALA: DELAYTIME 0...600 SEC 0010 SEC

SENSOCHECK ON/OFF OFF

*EXTwithexternalIinputoptiononly

Page 35: Analizator ph, o2 knicks

35

Configuration

Configuration Choices DefaultParameter set (PARSET)PAR: Select fixed parameter set

(A)orswitchbetweenA/Bvia control input or manu-ally in measuring mode

PARSETFIX/CNTRINPUT/ MANUAL

PARSETFIXA(fixed parameter setA)

Real-time clock (CLOCK)CLK: FORMAT 24 h / 12 h

24 h TIMEhh/mm 00..23:00...59

12 h TIMEhh/mm 00...11AM/PM:00...59

DAY/MONTH 01...31/01...12

YEAR 2000...2099

Tag number (TAG)TAG: (Input in text line) XXXXXXXXXX

Page 36: Analizator ph, o2 knicks

36

Configuration (Original for Copy)Default Settings of Parameter Sets

TwocompleteparametersetsarestoredintheEEPROM. Asdelivered,thetwosetsareidenticalbutcanbeedited.

Please note:Fillinyourconfiguration data on the following pages or use them as original for copy.

*)TheseparameterscannotbeadjustedinparametersetB,thevaluesarethesameasinparametersetA.

Parameter Parameter set A Parameter set BSNS:Measuringmode ---*)

SNS: Sensor type ---*)

SNS: Temperature unit ---*)

SNS: Calibration mode ---*)

SNS: Calibration timer ---*)

SNS: Calibration cycle ---*)

OT1:Processvariable

OT1:Currentstart

OT1:Currentend

OT1:Filtertime

OT1:22mAerrorcurrent

OT1:HOLDmode

OT1:HOLD-FIXcurrent

Page 37: Analizator ph, o2 knicks

37

(Original for Copy) ConfigurationParameter Parameter set A Parameter set BOT2:Processvariable

OT2:Currentstart

OT2:Currentend

OT2:Filtertime

OT2:22mAerrorcurrent

OT2:HOLDmode

OT2:HOLD-FIXcurrent

COR:Salinity(ppt)

COR:Pressureunit (BAR,KPA,PSI)

COR:Pressure(MAN/EXT)

COR:Ext.currentinput(Option)

ALA:Delay

ALA:Sensocheckon/off

PAR: Parameter set selection

---*)

CLK: Time format ---*)

CLK: Time hh/mm ---*)

CLK: Day/month ---*)

CLK:Year ---*)

TAG:Tagnumber ---*)

*)TheseparameterscannotbeadjustedinparametersetB,thevaluesarethesameasinparametersetA.

Page 38: Analizator ph, o2 knicks

38

1

ConfigurationSensorSelect: Sensor type, measuring mode, temperature unit

Press any arrow key.1 Select 2 CONF using keys,

press enter. Select parameter set using 3 ,

press enter. Select 4 SENSOR menu using keys,

press enter. Allitemsofthismenugroupareindicatedby5

the “SNS:” code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.

End: Press 6 meas keyuntilthe[meas]modeindicator is displayed.

2

3

4

Select sensor type

Select measuring mode

Temperature unit

Calibration medium: water/air

Calibration timer

5enter

enter

6

enter

enter

meas

5

Page 39: Analizator ph, o2 knicks

39

ConfigurationMenu item Action ChoicesSelect sensor type Select sensor type using

keys.

Confirm with enter

MEMOSENS STANDARDTRACES(withTAN)

Select measuring mode

Select measuring mode using keys.DO:MeasurementinliquidsGAS:MeasurementingasesConfirm with enter

dO %, dOmg/ldOppmGAS%

Temperature unit Select temperature unit using keys.

Confirm with enter

°C°F

5

Page 40: Analizator ph, o2 knicks

40

ConfigurationSensorSelect: Calibration medium: water/air, calibration timer

1 Press any arrow key.1 Select 2 CONF using keys,

press enter. Select parameter set using 3 ,

press enter. Select 4 SENSOR menu using keys,

press enter. Allitemsofthismenugroupareindicatedby5

the “SNS:” code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.

End: Press 6 meas keyuntilthe[meas]modeindicator is displayed.

2

3

4

Select sensor type

Select measuring mode

Temperature unit

Calibration medium: water/air

Calibration timer

5enter

enter

6

enter

enter

meas

5

Page 41: Analizator ph, o2 knicks

41

ConfigurationMenu item Action ChoicesCalibration medium: air/water

Select calibration medium using keys.AIR:Cal medium airWTR:Cal medium water

Confirm with enter

CAL_AIRCAL_WTR

Calibration timer Switch calibration timer on/off using keys.

Confirm with enter

OFFON

(ON:Calibration cycle)

Enter calibration cycle in hours using keys.

Confirm with enter

0...9999 h0168 h

5

Note for the calibration timer:WhenSensocheckhasbeenactivatedintheConfiguration>Alarmmenu, the expiration of the calibration interval is indicated by Sensoface:

Display Status

+

+

Over80%ofthecalibrationintervalhas already past.

The calibration interval has been exceeded.

ThecalibrationtimersettingsapplytobothparametersetsAandB.The time remaining until the next due calibration can be seen in the diagnostics menu (see “Diagnostics“).

Page 42: Analizator ph, o2 knicks

42

Process variableCurrent startCurrent endTime averaging filterOutputcurrentduringerrormessageOutputcurrentduringHOLDOutputcurrentforHOLDFIX

ConfigurationCurrent Output 1Process variable, current start, current end

1 Press any arrow key.1 Select 2 CONF using keys,

press enter. Select parameter set using 3 ,

press enter. Select 4 OUT1 menu using keys,

press enter. Allitemsofthismenugroupareindicatedby5 the“OT1:”code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.

End: Press 6 meas keyuntilthe[meas]modeindicator is displayed.

2

3

4

5enter

enter

6

enter

5

enter

meas

Page 43: Analizator ph, o2 knicks

43

ConfigurationMenu item Action ChoicesProcess variable Select using keys:

OXY:O2 valueTMP:Temperature

Confirm with enter

OXY/TMP

Current start Modifydigitusing, select next digit using keys.

Confirm with enter

000.0...0600%(OXY)0.000...0150%(OXY, tracesOption) –20...150 °C / –4...302°F(TMP)

Current end Enter value using keys.

Confirm with enter

000.0...0600%(OXY)0.000...0150%(OXY, tracesOption) –20...150 °C / –4...302°F(TMP)

Forprocess variables comprising several decades, decimal point and dimension can be shifted using the cursor keys. Then, the desired number is entered using and .Formeasurementingases(GAS),thismethodisusedtoswitchbetweenppmand%forvolumeconcentration(10000ppm=1%).

5

Example2:Range50...70% Advantage:Higherresolutionin range of interest

Assignment of Measured Values: Current Start and Current End

Outputcurrent

[%]

20 4

100

0

[%]

20 4

70

50[mA]

Example 1: Range0...100%

[mA]

Outputcurrent

5070

Oxygen saturation Oxygen

saturation

Page 44: Analizator ph, o2 knicks

44

ConfigurationCurrent Output 1Adjust time interval of output filter

Process variableCurrent startCurrent endTime averaging filterOutputcurrentduringerrormessageOutputcurrentduringHOLDOutputcurrentforHOLDFIX

1 Press any arrow key.1 Select 2 CONF using keys,

press enter. Select parameter set using 3 ,

press enter. Select 4 OUT1 menu using keys,

press enter. Allitemsofthismenugroupareindicatedby5 the“OT1:”code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.

End: Press 6 meas keyuntilthe[meas]modeindicator is displayed.

2

3

4

5enter

enter

6

enter

5

enter

meas

Page 45: Analizator ph, o2 knicks

45

ConfigurationMenu item Action ChoicesTime averaging filter

Enter value using keys.

Confirm with enter

0...120 SEC (0000 SEC)

5

Time Averaging FilterTosmoothenthecurrentoutput,alow-passfilterwithadjustablefiltertime constant can be switched on. When there is a jump at the input (100%),theoutputlevelisat63%afterthetimeintervalhasbeenreached. The time interval can be set from 0 to 120 sec. If the time interval is set to 0 sec, the current output directly follows the input.Please note: The filter only acts on the current output, not on the display, thelimitvalues,orthecontroller!DuringHOLDthefilterisnotapplied.Thispreventsajumpatthe output.

Time interval 0...120 s

Oxy

0/4-20mA

Time interval 0...120 s

Temp

Display

Page 46: Analizator ph, o2 knicks

46

ConfigurationCurrent Output 1Output current during Error and HOLD

Process variableCurrent startCurrent endTime averaging filterOutputcurrentduringerrormessageOutputcurrentduringHOLDOutputcurrentforHOLDFIX

1 Press any arrow key.1 Select 2 CONF using keys,

press enter. Select parameter set using 3 ,

press enter. Select 4 OUT1 menu using keys,

press enter. Allitemsofthismenugroupareindicatedby5 the“OT1:”code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.

End: Press 6 meas keyuntilthe[meas]modeindicator is displayed.

2

3

4

5enter

enter

6

enter

5

enter

meas

Page 47: Analizator ph, o2 knicks

47

ConfigurationMenu item Action ChoicesOutputcurrentdur-ing error message

SelectON(22mAforerrormessage)orOFFusing keys.Confirm with enter

ON/OFF

OutputcurrentduringHOLD

LAST:DuringHOLDthelast measured value is maintained at the output.FIX:DuringHOLDavalue(to be entered) is main-tained at the output.Select using Confirm with enter

LAST/FIX

OutputcurrentforHOLDFIX

OnlywithFIXselected:Enter current which is to flow at the output during HOLDEnter value using keys.

Confirm with enter

00.00...22.00mA21.00 mA

HOLD active

21

4HOLD active

Outputcurrent[mA]

OutputsignalHOLDSettingFIX=21.0mA

OutputsignalHOLDSettingLAST

Output Signal During HOLD:

5

Page 48: Analizator ph, o2 knicks

48

ConfigurationCurrent Output 2Current start. Current end.

Process variableCurrent startCurrent endTime averaging filterOutputcurrentduringerrormessageOutputcurrentduringHOLDOutputcurrentforHOLDFIX

1 Press any arrow key.1 Select 2 CONF using keys,

press enter. Select parameter set using 3 ,

press enter. Select 4 OUT2 menu using keys,

press enter. Allitemsofthismenugroupareindicatedby5 the“OT2:”code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.

End: Press 6 meas keyuntilthe[meas]modeindicator is displayed.

2

3

4

5enter

enter

6

enter

5

enter

meas

Page 49: Analizator ph, o2 knicks

49

ConfigurationMenu item Action ChoicesProcess variable Select using keys:

OXY:O2 valueTMP:TemperatureConfirm with enter

OXY/TMP

.

.

.

All the following adjustments are made as for current output 1 (see there)!

5

Page 50: Analizator ph, o2 knicks

50

ConfigurationCorrectionSalinity correction. Pressure correction. Current input

1 Press any arrow key.1 Select 2 CONF using keys,

press enter. Select parameter set using 3 ,

press enter. Select 4 CORRECTION menu using keys,

press enter. Allitemsofthismenugroupareindicatedby5 the“COR:”code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.

End: Press 6 meas keyuntilthe[meas]modeindicator is displayed.

2

3

4

enter

enter

6

enter

5

enter

meas

SalinityPressure unitPressure correctionManualpressureinputCurrent inputCurrent range/Pressure range

5

Page 51: Analizator ph, o2 knicks

51

ConfigurationMenu item Action ChoicesEnter salinity Enter salinity of the

process medium.Enter value using keys.

Confirm with enter

00.00...45.00 ppt

Enter pressure unit Select desired pressure unit using keys.

Confirm with enter

Bar/kPa/PSI

Enter pressure correction

Select desired procedure for pressure correction using keys:MAN: ManualspecificationEXT: Ext. pressure correctionvia current inputConfirm with enter

MAN/EXT

(Manualpressureinput)

Enter value using keys.

Confirm with enter

Input range:0.000...9.999BAR/000.0...999.9KPA/000.0...145.0 PSI

1.013 BAR

Current input/ Pressure range

Forexternalpressuredetection, enter 0(4) ... 20mAcurrentinputandcurrent start / end values for pressure parameter using keys.

0(4)...20mA

0.000...9.999 Bar /000.0...999.9KPA/000.0...9990.9 PSI

5

Page 52: Analizator ph, o2 knicks

52

ConfigurationAlarmAlarm delay. Sensocheck

1 Press any arrow key.1 Select 2 CONF using keys,

press enter. Select parameter set using 3 ,

press enter. Select 4 ALARM menu using keys,

press enter. Allitemsofthismenugroupareindicatedby5 the“ALA:”code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.

End: Press 6 meas keyuntilthe[meas]modeindicator is displayed.

2

4

enter

enter

enter

3

6

meas

enter5 Delay

Sensocheck

5

Page 53: Analizator ph, o2 knicks

53

Errormessagescanbesignaledbya22mAoutputcurrent(seeErrorMessagesandConfigurationofOutput1/Output2).

The alarm delay time delays the color change of the display back-lightingtoredandthe22mAsignal(ifconfigured).

ConfigurationMenu item Action ChoicesAlarmdelay Enter alarm delay using

keys.Confirm with enter

0...600 SEC(010 SEC)

Sensocheck Select Sensocheck / SensofaceSelectONorOFFusing keys.Confirm with enter.

ON/OFF

5

Page 54: Analizator ph, o2 knicks

54

ConfigurationTime and DateTag Number

1 Press any arrow key.1 Select 2 CONF using keys,

press enter. SelectparametersetAusing3 keys,

press enter. 4 Press enter. Select 5 CLOCK or TAG using keys, press

enter. Allitemsofthismenugroupareindicatedby6 the“CLK:”or“TAG”code. Press enter to select menu, edit with arrow keys (see next page). Confirm (and proceed) with enter.

End: Press 7 meas keyuntilthe[meas]modeindicator is displayed.

2

3

4

enter

MEMO SENS

6

enter

5

enter

meas

Time formatTimeDay and monthYear

Tag number

enter

5

Page 55: Analizator ph, o2 knicks

55

Time and DateControl of the calibration and cleaning cycles is based on the time and dateoftheintegratedreal-timeclock. In measuring mode the time is shown in the lower display. When using digital sensors, the calibration data is written in the sensor head.In addition, the logbook entries (cf Diagnostics) are provided with a time stamp.

Please note:Afterprolongedpoweroutage(>5days)thetimedisplayis• replaced by dashes and cannot be used for processing. Enter the correct time.Thereisnoautomaticswitchoverfromwintertosummertime! •Besuretomanuallyadjustthetime!

Configuration

Menu item Action ChoicesTag number Select character using

keys,select next digit using keys.

Confirm with enter

A...Z,0...9,–+<>?/@

The first 10 characters are seen in the display with-out scrolling.

Tag Number (“ TAG“)Youcanenteradesignationforthepointofmeasurement(tag number) in the lower display line. Up to 32 digits are possible.Pressing meas (repeatedly) in the measuring mode indicates the tag number.Beingpartofthedeviceconfiguration,the“TAG“canbereadoutviaIrDA.Astandardizedtagnumberhelps,forexample,tocorrectlyre-installadevice after repair.

5

Page 56: Analizator ph, o2 knicks

56

Digital SensorsOperation

The device only switches to measuring mode when the connected sensor corresponds to the type configured (Sensoface is happy):

MEMO SENS

Otherwise,anerrormessageisreleased.Theinfo icon is displayed. Youcandisplaytheerrortextinthebottomlineusingthekeys. Sensoface is sad (see table of error messages and Sensoface in the Appendix):

Page 57: Analizator ph, o2 knicks

57

Connecting a Digital SensorStep Action/Display Remark

Connect sensor

MEMO SENS

BeforeaMemosens sensor is connected, theerrormessage“NOSENSOR“isdisplayed.

Wait until the sensor data are displayed.

MEMO SENS

The hourglass in the display blinks.

Check sensor dataMEMO SENS

View sensor informa-tion using keys, confirm with enter.

The display color changes to green.

Sensoface is friendly when the sensor data are okay.

Go to measuring mode

Press meas, info, or enter

After60secthedeviceautomatically returns to measuring mode (timeout).

Possible error messagesSensor devaluatedReplacesensor

When this error message appears, the sensor can-not be used any more. Sensoface is sad.

Sensor defectiveReplacesensor

When this error message appears, the sensor cannot be used.Sensoface is sad.

Digital Sensors

Page 58: Analizator ph, o2 knicks

58

Digital SensorsSensor Replacement

AdigitalsensorshouldonlybereplacedduringHOLDmode to pre-vent unintended reactions of the outputs or contacts. When you first want to calibrate the new sensor, it can also be replaced in calibration mode.

Step Action/Display RemarkSelectHOLDmode Press any key to call

the selection menu, selectHOLDusingthe keys, press enter to confirm.

Now the device is in HOLDmode.TheHOLDmode can also be acti-vated externally via the HOLDinput.DuringHOLDtheoutputcurrentisfrozenatitslast value or set to a fixed value.

Disconnect and remove old sensorInstall and connect new sensor.

Temporary messages which are activated dur-ing the replacement are indicated but not output to the alarm contact and not entered in the log-book.

Wait until the sensor data are displayed.

Page 59: Analizator ph, o2 knicks

59

Digital Sensors

Step Action/Display RemarkCheck sensor data

View sensor informa-tion using keys, confirm with enter.

Youcanviewthesensormanufacturer and type, serial number, and last calibration date.

Check measured valuesExitHOLD Hit meas key:

Returntoselectionmenu. Holdmeas key depressed: Device switches to measuring mode

The sensor replacement is entered in the extended logbook.

Page 60: Analizator ph, o2 knicks

60

CalibrationCalibration adapts the device to the individual sensor characteristics. It is always recommended to calibrate in air. Compared to water, air is a calibration medium which is easy to handle, stable, and thus safe. In the most cases, however, the sensor must be dismounted for a calibration in air.When dealing with biotechnological processes which require sterile conditions,thesensorcannotberemovedforcalibration.Here, calibration must be performed with aeration directly in the process medium(e.g.aftersterilization).In the field of biotechnology, for example, often saturation is measured and calibration is performed in the medium for reasons of sterility. Forotherapplicationswhereconcentrationismeasured(watercontroletc.), calibration in air has proved to be useful.

Please note:Allcalibrationproceduresmustbeperformedbytrainedperson-•nel. Incorrectly set parameters may go unnoticed, but change the measuring properties.

Ifa2-pointcalibrationisprescribed,thezerocalibrationshouldbe•performed prior to slope calibration.

Common Combination: Process Variable / Calibration Mode / Calibration MediumProcess variable

Cal mode

Calibration Default rel. humidity

Default cal pressure

Saturation(%) SAT Water 100% Process pressure

Concentration (mg/l, ppm)

Conc Air 50% 1.013 bar

The calibration procedures for these two common applications are describedonthefollowingpages.Ofcourse,othercombinationsofprocess variable and calibration mode are possible.

Page 61: Analizator ph, o2 knicks

61

Selecting a Calibration ModeCalibration is used to adapt the device to the individual sensor characteristics,namelyzeroandslope.Accesstocalibrationcanbeprotectedwithapasscode (SERVICEmenu).First,youopenthecalibrationmenuandselectthecalibration mode:

CAL_WTR/CAL_AIR

Calibration in water/air (as configured)

CAL_ZERO Zero calibrationP_CAL Product calibration (calibration with sampling)CAL-RTD Temperature probe adjustment

Page 62: Analizator ph, o2 knicks

62

Zero CalibrationWhenazerocalibrationisperformed,thesensorshouldremainforat least 10 to 30 minutes in the calibration medium in order to obtain stable,non-driftingvalues.Duringzerocalibration,adriftcheckisnotperformed.Zerocurrentofaproperlyfunctioningsensorisnotablylessthan0.5%ofaircurrent.The display (bottom: measured value, top: entered value) does not changeuntilaninputcurrentisenteredforthezeropoint.Whenmeasuringinanoxygen-freemedium,thedisplayedcurrentcanbe taken directly.

Zero calibration

Page 63: Analizator ph, o2 knicks

63

Zero CalibrationDisplay Action Remark

Select calibration, proceed with enter

Readyforcalibration. Hourglassblinks.Placesensorinoxygen-free medium

Display (3 sec)Now the device is in HOLDmode.

Maindisplay:Zero current. Press enter to save this value or correct using arrow keys and then save with enter.Secondary display: Sensor current measuredDisplay of slopeDisplayofnewzerocurrent.End calibration with enter key, place sensor in process

Sensoface display

The oxygen value is shown in the main dis-play, “enter“ blinks. StopHoldwithenter.

New calibration: SelectREPEAT,press enter key.

Quit with enter. Afterendofcalibra-tion, the outputs re-maininHOLDmodefor a short time.

Page 64: Analizator ph, o2 knicks

64

Product CalibrationCalibration with SamplingDuring product calibration the sensor remains in the process. The measurement process is only interrupted briefly.

Procedure: During sampling the currently measured value is stored in the device. The device immediately returns to measuring mode.The cal mode indicator blinks and reminds you that calibration has not been termi-nated.Thecomparisonvalueismeasuredonthesite,e.g.usingaportableDOmeterin a bypass.Thisvalueisthenenteredinthedevice.Thenewvalueforslopeorzeroiscalculatedfromthestoredvalueandthecomparisonvalue.Fromthemeasuredvalue,thedeviceautomaticallyrecognizeswhetheranewslopeorzeromustbecalculated(aboveapprox.5%saturation:slope,below:zero).If the sample is invalid, you can take over the measured value stored during sam-pling instead of the comparison value. In that case the old calibration values remain stored.Afterwards,youcanstartanewproductcalibration.Thefollowingdescribesaproductcalibrationwithslopecorrection–aproductcalibrationwithzerocorrec-tion is performed correspondingly.

Display Action RemarkSelect calibration,then product calibra-tionP_CAL.Proceed with enter

Readyforcalibration. Hourglassblinks.

Display (3 sec)Now the device is in HOLDmode.

Take sample and save value. Proceed with enter

Now the sample can be measured. If the value is already available, press info+enter to proceed to step 2.

Product calibration

Page 65: Analizator ph, o2 knicks

65

Display Action RemarkThe device returns to measuring mode.

FromtheblinkingCALmodeindicatoryou see that product calibration has not been terminated.

Product calibration step 2:When the sample value has been determined, open the product cali-bration once more.

Display (3 sec)Now the device is in HOLDmode.

The stored value is displayed (blinking) and can be overwritten with the lab value.Proceed with enter

Display of new slope andzero.Sensoface is active.

Proceed with enter

Relatedto25°Cand1013 mbars

Display of new oxy value.Sensoface is active.To end calibration:SelectMEAS,thenenter

To repeat calibra-tion:SelectREPEAT,then enter

End of calibration Afterendofcalibra-tion, the outputs re-maininHOLDmodefor a short time.

Product Calibration

Page 66: Analizator ph, o2 knicks

66

Display Action RemarkSelect calibration (SLOPE).Immerse sensor in cal medium, start with enter

“Mediumwater“or“Mediumair“is selected in the configuration.

Enter cal pressure

Proceed with enter

Default: 1.000 barUnit bar/kpa/PSI

Drift check: Display of:Sensorcurrent(nA)Responsetime(s)Temperature(°C/°F)

Device goes to HOLDmode.

The drift check might take some time.

Display of calibration data(slopeandzero)and Sensoface

Proceed with enter

Relatedto25°Cand1013 mbars

Display of selected process value.To end calibration:SelectMEASusing, then enter

To repeat calibra-tion:SelectREPEATusing , then enter

Place sensor in process.End of calibration

Afterendofcalibra-tion, the outputs re-maininHOLDmodefor a short time.

Slope Calibration (Medium: Water)(air-saturated)

Page 67: Analizator ph, o2 knicks

67

Display Action RemarkSelect calibration. Place sensor in air, start with enterDevicegoestoHOLDmode.

“Mediumwater“or“Mediumair“isselected in the con-figuration.

Enter relative humidity using arrow keys

Proceed with enter

Default for relative humidity in air: rH=50%

Enter cal pressure using arrow keysProceed with enter

Default: 1.000 barUnit bar/kpa/PSI

Drift check:Display of:Sensorcurrent(nA)Responsetime(s)Temperature(°C/°F)Proceed with enter

The drift check can take some minutes.

Display of calibration data(slopeandzero). Proceed with enter.

Display of selected process variable (here: %vol).NowthedeviceisinHOLDmode:Reinstallthe sensor and check whether the message isOK.MEASendscalibration,REPEATpermits repetition.

Afterendofcalibra-tion, the outputs re-maininHOLDmodefor a short time.

Slope Calibration (Medium: Air)

Page 68: Analizator ph, o2 knicks

68

Temp Probe AdjustmentDisplay Action Remark

Select temp adjust-ment. Proceed with enter

Wrong settings change the measurement properties!

Measurethetem-perature of the process medium using an exter-nal thermometer.

Display (3 sec)Now the device is in HOLDmode.

Enter the measured temperature value.Maximumdifference:10 K.Proceed with enter

Display of actual temperature (un-compensated) in the lower display.

The corrected tempera-ture value is displayed.Sensoface is active.To end calibration: SelectMEAS,thenenterTo repeat calibration: SelectREPEAT,thenenter

Afterendofcalibra-tion, the outputs re-maininHOLDmodefor a short time.

Aftercalibrationisended, the device will switch to measuring mode.

Temp probe adjustment

Page 69: Analizator ph, o2 knicks

69

MeasurementDisplay Remark

orAM/PMand°F:

Fromtheconfigurationorcalibrationmenus,you can switch the device to measuring mode by pressing the meas key.(Waitingtimeforsignalstabilizationapprox.20sec).In the measuring mode the main display showstheconfiguredprocessvariable(Oxy[%]ortemperature),thesecondarydisplayshows the time and the second configured processvariable(Oxy[%]ortemperature).The[meas]modeindicatorlightsandtheactiveparameterset(A/B)isindicated.

Pressing the enter key briefly shows the output currents. By pressing the meas key you can step through the following dis-plays. When no key has been pressed for 60 sec, the device returns to the standard display.

Selecting the1) parameter set (if set to “manual“ in the configuration). Select the desired parameter set using thearrowkeys(PARSETAorPARSETBblinks in the lower display line). Confirm with enter.

Furtherdisplays(each with meas).

Displayoftagnumber(“TAG“)2) Display of 3) time and date

Page 70: Analizator ph, o2 knicks

70

DiagnosticsIn the Diagnostics mode you can access the following menus without interrupting the measurement:CALDATA Viewing the calibration dataSENSOR Viewing the sensor dataSELFTEST Startingadeviceself-testLOGBOOK Viewing the logbook entriesMONITOR Displaying currently measured valuesVERSION Displaying device type, software version, serial number

Accesstodiagnosticscanbeprotectedwithapasscode (SERVICEmenu).

Please note:HOLDisnotactiveduringDiagnosticsmode!

Action Key RemarkActivate diagnostics

Press any arrow key to call the selection menu.(Display color changes to turquoise.)SelectDIAGusing keys, confirm with enter

Select diagnostics option

Use keys to select from:CALDATASENSORSELFTESTLOGBOOKMONITORVERSIONSee next pages for further proceeding.

End meas End with meas.

Diagnostics(DIAG)

Page 71: Analizator ph, o2 knicks

71

DiagnosticsDisplay Menu item

Display of calibration dataSelectCALDATAusing, confirm with enter.Use thekeys to select the desired parameter from thebottomlineofthedisplay(LAST_CALZEROSLOPENEXT_CAL).The selected parameter is shown in the main display.

Press meas to return to measurement.

Display of sensor dataForanalogsensors,thetypeisdisplayed,for digital sensors, the manufacturer, type, serial number, and last calibration date.In each case Sensoface is active.

Display data using keys,return with enter or meas.

Page 72: Analizator ph, o2 knicks

72

DiagnosticsDisplay Menu item

Device self-test(To abort, you can press meas.)

1 Display test: Display of all segments with changing background colors white/green/red. Proceed with enter

2 RAM test:Hourglassblinks, thendisplayof--PASS--or--FAIL-- Proceed with enter

3 EEPROM test:Hourglassblinks, thendisplayof--PASS--or--FAIL-- Proceed with enter

4 FLASH test: Hourglassblinks, thendisplayof--PASS--or--FAIL-- Proceed with enter

5 Module test: Hourglassblinks, thendisplayof--PASS--or--FAIL-- Press enter or meas to return to measuring mode.

Page 73: Analizator ph, o2 knicks

73

DiagnosticsDisplay Menu item

Display of logbook entriesSelectLOGBOOKusing, confirm with enter.

With the keys, you can scroll backwards and forwardsthroughthelogbook(entries-00-...-99-),-00-being the last entry.

If the display is set to date/time, you can search for a particular date using the keys.Pressto view the corresponding message text.

If the display is set to the message text, you can search for a particular message using the keys.Pressto display the date and time.

Press meas to return to measurement.

Extended logbook / Audit Trail (via TAN)With the keys, you can scroll backwards and forwards through the extended logbook (entries-000-...-199-),-000-beingthelastentry.Display: CFRAuditTrailalsorecordsfunctionactivations (CALCONFIGSERVICE),someSensofacemessages (cal timer, wear), and opening of the enclosure.

Page 74: Analizator ph, o2 knicks

74

DiagnosticsDisplay Menu item

Display examples:

Display of currently measured values (sensor monitor)SelectMONITORusing, confirm with enter.Use the keys to select the desired parameter fromthebottomlineofthedisplay:I-OXY,I-INPUT,OPERATIONTIME.Theselectedparameterisshowninthe main display.Press meas to return to measurement.

Display of directly measured value(for validation, sensor can be immersed in a calibra-tion solution, for example, or the device is checked by using a simulator)

Display of sensor operating time

VersionDisplay of device type, software/hardware version, and serial number for all device components.

Use the keys to switch between software and hardware version. Press enter to proceed to next device component.

Page 75: Analizator ph, o2 knicks

75

ServiceService(SERVICE)A

In the Service mode you can access the following menus:MONITOR displaying currently measured values OUT1 testing current output 1OUT2 testing current output 2IRDA activatingandcommunicatingviatheIrDAinterfaceCODES assigning and editing passcodesDEFAULT resetting the device to factory settingsOPTION enablingoptionsviaTAN.

Please note:HOLDisactiveduringServicemode!

Action Key/Display RemarkActivate Service

Press any arrow key to call the selection menu.SelectSERVICEusing keys, press enter to confirm.

Passcode Enter passcode “5555“ for service mode using the keys.

Press enter to confirm.

Display In service mode the following icons are displayed:

[diag]modeindicator•HOLDtriangle•Service (wrench)•

End meas End by pressing meas.

Page 76: Analizator ph, o2 knicks

76

ServiceMenu item Remark

Display of currently measured values (sensor monitor) with HOLD mode activated:SelectMONITORusing, press enter to confirm.Select variable in the bottom text line using.

The selected parameter is shown in the main display.AsthedeviceisinHOLDmode,youcanperformvalidations using simulators without influencing the signal outputs.

Press meas to return to the service menu.Returntomeasurement:Pressmeas once more.

Specify current at outputs 1 and 2:SelectOUT1orOUT2usingthekeys, press enter to confirm.Enter a valid current value for the respective output using keys.Press enter to confirm.Forcheckingpurposes,theactualoutputcurrentisshown in the bottom right corner of the display.End by pressing enter or meas.

IrDA communication:SelectIRDAusing, press enter to confirm.

When IrDAcommunicationisactive,thedeviceremainsintheHOLDmodeforreasonsofsafety.FurtheroperationisperformedviaIrDA.

End communication by pressing meas.

Exception: Firmware update (must not be interrupted!)

Page 77: Analizator ph, o2 knicks

77

ServiceMenu item Remark

Assigning passcodes:Inthe“SERVICE-CODES“menuyoucanassignpass-codestoDIAG,HOLD,CAL,CONF,andSERVICEmodes(Service preset to 5555).When you have lost the Service passcode, you have to request an “AmbulanceTAN“fromthemanufac-turer specifying the serial number of your device. Toenterthe“AmbulanceTAN“,calltheServicefunc-tionandenterpasscode7321.AftercorrectinputoftheambulanceTANthedevicesignals“PASS“for4secand resets the Service passcode to 5555.

Reset to factory settings:Inthe“SERVICE-DEFAULT“menuyoucanresetthedevice to factory settings.Caution!Afteraresettofactorysettingthedevicemustbe reconfigured completely, including the sensor parameters!

Option request:Communicate the serial number and hardware/ software version of your device to the manufacturer. These data can be viewed in the Diagnostics/Version menu.The“transactionnumber“(TAN)youwillthenreceiveis only valid for the device with the corresponding serial number.Release of options:Optionscomewitha“transactionnumber“(TAN). This TANmustbeenteredandconfirmedusingenter to release the option.

Page 78: Analizator ph, o2 knicks

78

Operating StatesOperatingstatesA

Operating status O

UT1

OUT2

Tim

e ou

t

Measuring -

DIAG 60 s

CAL_ZEROZero point

No

CAL_SLOPE Slope

No

P_CAL Product calibration S1

No

P_CAL Product calibration S2

No

CAL_RTD Temp adjustment

No

CONF_A ParSetA

20 min

CONF_B ParSet B

20 min

SERVICEMONITOR 20 min

SERVICEOUT1 20 min

SERVICEOUT2 20 min

SERVICEIRDA 20 min

SERVICECODES 20 min

SERVICEDEFAULT 20 min

SERVICEOPTION 20 min

HOLDinput No

Explanation: asconfigured(Last/FixorLast/Off )

active manual

Page 79: Analizator ph, o2 knicks

79

Product Line and Accessories Order Code Stratos Pro A 2...

TANExample A 2 1 1 X - PH - 1

2-wire / 4-20 mA A 2 B,C,E

CommunicationWithout (HART retrofittable via TAN) 0 A

Version numberVersion 1

ApprovalsGeneral Safety NATEX / IECEX Zone 2 BATEX / IECEX / FM / CSA Zone 1 / Cl 1 Div 1 XOther approvals Z

Measuring channelMemosens pH / Redox digital MSPHMemosens Cond digital MSCONDMemosens Oxy digital MSOXYDual COND (2x2-electrode sensors, analog) N CCpH / ORP value(ISM digital per TAN)

Measuring module PH F

Cond, 2-/4-electrode Measuring module CONDConductivity, electrodeless Measuring module CONDIOxygen (ISM digital and traces per TAN)

Measuring module OXY D, F

OptionsWithout 2nd current output 0With 2nd current output 1

TAN optionsHART SW-A001 (A)Logbook SW-A002 (B)Extended logbook (Audit Trail) SW-A003 (C)Trace oxygen measurement SW-A004 (D)Current input + 2 digital inputs SW-A005 (E)ISM digital SW-A006 (F)

Mounting accessoriesPipe-mount kit ZU 0274Protective hood ZU 0737Panel-mount kit ZU 0738

Page 80: Analizator ph, o2 knicks

80

Recommended Power Supply Units: Order No.:

Repeaterpowersupply,IS,24VAC/DC, output0/4...20mA

WG20A2

Repeaterpowersupply,IS,90...253VAC, output4...20mA

WG21A7

Repeaterpowersupply,IS,90...253VAC,HART,output4...20mA

WG21A7Opt.470

Repeaterpowersupply,IS,24VAC/DC, output4...20mA

WG21A7Opt.336

Repeaterpowersupply,IS,24VAC/DC,HART,output4...20mA

WG21A7Opt.336,470

Repeaterpowersupply,non-IS,24VDC, output4...20mA

IsoAmpPWRB10116

Repeaterpowersupply,non-IS,24VDC,HART,output0/4...20mA/0...10V

IsoAmpPWRA20100

A2...X: Supply Units and Connection

Connection to Supply Units

8

9

17

+ out

-out1

-out2

+

-

+

-

Supply unit 1

Supply unit 2

Stratos ProA2...X

Hazardous-area requirements for cable:Between the wires:Solid insulation ≥ 0.5 mmTest voltage ≥ 500 V

Page 81: Analizator ph, o2 knicks

81

OXY input InputforMemosenssensors

Operatingmodes GAS Measurementingases

DO Measurementinliquids

Display ranges Standard

Saturation(-10...80°C) 0.0...600.0%

Concentration(-10...80°C) 0.00 ... 99.99 mg/l

(Dissolved oxygen) 0.00 ... 99.99 ppm

Volume concentration in gas 0.00...99.99%vol

Display ranges Traces

Saturation(-10...80°C) 0.000...150.0%

Concentration(-10...80°C) 0000 ... 9999 µg/l / 10.00 ... 20.00 mg/l

(Dissolved oxygen) 0000 ... 9999 ppb / 10.00 ... 20.00 ppm

Volume concentration in gas 0000...9999ppm/1.000...50.00%vol

Input correction Pressure correction * 0,000 ... 9.999 bars / 999.9 kPa / 145.0 psi

manually or through current input 0(4)...20mA

Salinity correction 0.0 ... 45.0 g/kg

Sensor standardization *

Operatingmodes* AIRAutomaticcalibrationinair

WTRAutomaticcalibrationinair-saturatedwater

Product calibration

Zero calibration

Calibration range Standard

Zero point ±2nA

Slope 25...130nA(at25°C,1013mbars)

Calibration range Traces

Zero point ±2nA

Slope 200...550nA(at25°C,1013mbars)

Calibration timer * Interval 0000 ... 9999 h

Pressure correction * Manual0.000...9.999bars/999.9kPa/145.0psi

Sensocheck / Sensoface Provides information on the sensor condition, evaluationofzero/slope,responsetime,calibrationinterval, wear, can be switched off

Specifications

Page 82: Analizator ph, o2 knicks

82

Specifications

I input Currentinput0/4...20mA/50Ωforexternalpressurecompensation

Start/end of scale Configurable 0 … 9.999 bars

Characteristic Linear

Measurementerror2.4) <1%currentvalue+0.1mA

HOLD input Galvanicallyseparated(OPTOcoupler)

Function SwitchesdevicetoHOLDmode

Switching voltage 0...2V(AC/DC) HOLDinactive

10...30V(AC/DC) HOLDactive

CONTROL input Galvanicallyseparated(OPTOcoupler)

Function SelectingparametersetA/B

Switching voltage 0...2V(AC/DC) ParametersetA

10...30V(AC/DC) Parameter set B

Output 1 Currentloop4...20mA,floating,protectedagainstinversepolarity, HARTcommunication,supplyvoltage14...30V

Process variable* DOsaturation/DOconcentration/Temperature

Characteristic Linear

Overrange* 22mAinthecaseoferrormessages

Outputfilter* PT1 filter, time constant 0 ... 120 s

Measurementerror3) <0,25%currentvalue+0,025mA

Start/end of scale * Configurable within selected range

Minimumspan Standard:5%/0.5mg/l(ppm)/2%vol

Traces:0.2%/20µg/l(ppb)/100ppm

Temperature:20K/36°F

Output 2 Currentloop4...20mA,floating,protectedagainstinversepolarity

Process variable* DOsaturation/DOconcentration/Temperature

Characteristic Linear

Overrange* 22mAinthecaseoferrormessages

Page 83: Analizator ph, o2 knicks

83

Specifications

Outputfilter* PT1 filter, time constant 0 ... 120 s

Measurementerror2,3,4) <0,25%currentvalue+0,05mA

Start/end of scale * Configurable within selected range

Minimumspan Standard:5%/0.5mg/l(ppm)/2%vol

Traces:0.2%/20µg/l(ppb)/100ppm

Temperature:20K/36°F

Real-time clock Different time and date formats selectable

Power reserve > 5 days

Display LCdisplay,7-segmentwithicons

Maindisplay Character height approx. 22 mm, unit symbols approx. 14 mm

Secondary display Character height approx. 10 mm

Text line 14 characters, 14 segments

Sensoface 3 status indicators (friendly, neutral, sad face)

Modeindicators meas, cal, conf, diag

Furthericonsforconfigurationandmessages

Alarmindication Redbacklightingincaseofalarm

Keypad Keys: meas, info, 4 cursor keys, enter

HART communication HARTVersion6 DigitalcommunicationbyFSKmodulationofoutputcurrent1

Device identification, measured values, status and messages, parameter setting, calibration, records

IrDA interface Infrared interface for transmission of records and logbook, parameter setting, calibration, firmware update

FDA 21 CFR Part 11 Accesscontrolbyeditablepasscodes

LogbookentryandflagviaHARTinthecaseofconfigurationchanges

Messageandlogbookentrywhenenclosureisopened

Page 84: Analizator ph, o2 knicks

84

Specifications

Diagnostics Functions

Calibration data Calibrationdate,zero,slope,responsetime

Deviceself-test Displaytest,automaticmemorytest(RAM,FLASH,EEPROM), module test

Logbook 100 events with date and time

Extendedlogbook(TAN) AuditTrail:200eventswithdateandtime

Service functions

Sensor monitor Display of direct, uncorrected sensor signal

Current source Currentspecifiableforoutput1and2(00.00...22,00mA)

IrDA ActivatingtheIrDAfunction

Passcodes Assigningpasscodesformenuaccess

Factorysetting ResettingallparameterstofactorysettingException: calibration data

TAN Enabling optionally available additional functions

Data retention Parameters,calibrationdata,logbook>10years(EEPROM)

EMC EN61326-1(GeneralRequirements)

Emitted interference Class B (residential area)

Immunity to interference IndustryEN61326-2-3(ParticularRequirementsforTransmitters)

Explosion protectionStratos Pro A2...XMSOXY

Europe: ATEXZone0,1,2

USA: FMClIDiv1,2/Zone0,1,2 (pending)

Canada: cCSAusClIDiv1,2/Zone0,1,2

International: IECEx

Memosens Ex wiring Uo (V)

Io (mA)

Po (mW)

Ci (μH)

Li (μH)

StratosProA2...X/A2...B 5 124 155 10.6 0

Page 85: Analizator ph, o2 knicks

85

Specifications

Nominal operating conditions

Ambienttemperature -20...+65°C

Transport/Storage temperature

-20...+70°C

Relativehumidity 10...95%notcondensing

Supply voltage 14 ... 30 V

Enclosure MoldedenclosuremadeofPBT,glassreinforced

Fastening Wall, pipe/post, or panel mounting

Color Gray,RAL7001

Ingress protection IP 67

Flammability UL94V-0

Dimensions 148 mm x 148 mm

Control panel cutout 138 mm x 138 mm to DIN 43 700

Weight 1.2 kg (1.6 kg incl. accessories and packaging)

Cable glands 3knockoutsforM20x1.5cableglands

2 knockouts for NPT ½” or rigid metallic conduit

Connections Terminals, conductor cross section max. 2.5 mm2

*User-defined

1)Automaticrangeselection

2) To IEC 746 Part 1, at nominal operating conditions

3) ± 1 count

4) Plus sensor error

Page 86: Analizator ph, o2 knicks

86

Error HandlingAlarm condition:

The display backlighting turns • redThe alarm icon • is displayedThecompletemeasured-valuedisplayblinks•“• ERR xxx“ is displayed in the lower menu line

Press the [info]keytoviewashorterrortext:The error text appears in the lower menu line•The main display reads “• InFo“.

Parameter errors:Configuration data such as current range, limit values, etc are checked during the input.If they are out of range,

“• ERR xxx“ is displayed for 3 sec,the display backlighting flashes red,•the respective maximum or minimum value is shown,•input must be repeated.•

Ifafaultyparameterarrivesthroughtheinterface(IrDA,HART),an error message will be displayed: “• ERR 100...199“thefaultyparametercanbelocalizedbypressingthe[• info]key

Calibration errors:If errors occur during calibration,

an error message will be displayed•calibration will be restarted•

Sensoface:If the Sensoface becomes sad,

the display backlighting will turn purple•the cause can be seen by pressing the • info keythe calibration data can be seen in the Diagnostics menu•

Page 87: Analizator ph, o2 knicks

87

Error codes

Error Messages

ErrorInfo text(is displayed in case of fault when the Info key is pressed)

Problem Possible causes

ERR 99 DEVICEFAILURE Error in factory settingsEEPROMorRAMdefectiveThis error message only occurs in the case of a total defect. The device must be repaired and recalibrated at the factory.

ERR 98 CONFIGURATIONERROR Error in configuration or calibration dataMemoryerrorindeviceprogramConfiguration or calibration data defective; completely reconfig-ure and recalibrate the device.

ERR 97 NOMODULEINSTALLED No modulePlease have the module replaced in the factory.

ERR 96 WRONGMODULE Wrong modulePlease have the module replaced in the factory.

ERR 95 SYSTEMERROR System errorRestartrequired. If error still persists, send in the device for repair.

ERR 01 NOSENSOR O2 sensor *Sensor defectiveSensor not connectedBreak in sensor cable

ERR 02 WRONGSENSOR Wrong sensor *

ERR 03 CANCELEDSENSOR Sensor devaluated *

Page 88: Analizator ph, o2 knicks

88

Error Messages

ErrorInfo text(is displayed in case of fault when the Info key is pressed)

Problem Possible causes

ERR 04 SENSORFAILURE Failure in sensor *ERR 05 CALDATA Error in cal data *ERR 11 OXYRANGE Display range violation

SATsaturationCONCconcentratonorGASvolumeconcentration

ERR 12 SENSORCURRENTRANGE Measuring range of sensor exceeded

ERR 13 TEMPERATURERANGE Temperature range violation

ERR 15 SENSOCHECK SensocheckERR 60 OUTPUTLOAD Load errorERR 61 OUTPUT1TOOLOW Output current 1

<0(3.8)mAERR 62 OUTPUT1TOOHIGH Output current 1

>20.5mAERR 63 OUTPUT2TOOLOW Output current 2

<0(3.8)mAERR 64 OUTPUT2TOOHIGH Output current 2

>20.5mAERR 69 TEMP.OUTSIDETABLE Temperature value outside

tableERR 100...255

VOIDPARAMETER Invalid parameter

*Digitalsensors

Page 89: Analizator ph, o2 knicks

89

Page 90: Analizator ph, o2 knicks

90

EC Declaration of Conformity

Page 91: Analizator ph, o2 knicks

91

EC Declaration of Conformity

Page 92: Analizator ph, o2 knicks

92

Stratos Pro A211X MSOXY: Control Drawings

Page 93: Analizator ph, o2 knicks

93

FM Control Drawing

Page 94: Analizator ph, o2 knicks

94

CSA Control Drawing

Page 95: Analizator ph, o2 knicks

95

FDA 21 CFR Part 11Conformity with FDA 21 CFR Part 11Intheirdirective“Title21CodeofFederalRegulations,21CFRPart11,ElectronicRecords;ElectronicSignatures“theUSAmericanhealthagencyFDA(FoodandDrugAdministration)regulatestheproductionand processing of electronic documents for pharmaceutical devel-opment and production. This results in requirements for measuring devices used for corresponding applications. The following features ensure that the measuring devices of this Series meet the demands of FDA21CFRPart11:

Electronic Signature – PasscodesAccesstothedevicefunctionsisregulatedandlimitedbyindividuallyadjustable codes–“Passcodes“(seeSERVICE).Thispreventsunauthor-izedmodificationofdevicesettingsormanipulationofthemeasure-mentresults.Appropriateuseofthesepasscodesmakesthemsuitableas electronic signature.

Audit TrailEvery (manual) change of device settings can be automatically docu-mented.Eachchangeistaggedwitha“ConfigurationChangeFlag“,whichcanbeinterrogatedanddocumentedusingHARTcommunica-tion.AltereddevicesettingsorparameterscanalsoberetrievedanddocumentedusingHARTcommunication.

Extended logbookAuditTrailalsorecordsfunctionactivations(CAL,CONFIG,SERVICE),some Sensoface messages (cal timer, wear), and opening of the enclosure.

Page 96: Analizator ph, o2 knicks

96

AAccesscodes95, 104Accessories79Alarm27

Delay 52AmbulanceTAN77Applicationinhazardouslocations15Approvalsforapplicationinhazardouslocations9, 84Assembly11AuditTrail95

BBacklighting 21

CCalibration 24, 60

Calibration errors 86Calibration data 71Calibration errors 86Calibration mode 61Calibration timer 41CD-ROM3Compulsory marking 9, 16Configuration 24

Alarm52Correction 50Current output 1 42Current output 2 48Individual configuration data 36Menugroups29Menustructure28OutputcurrentduringErrorandHOLD46Sensocheck 52Sensor 38Tag number 54Time and date 54Time averaging filter 44

Connection examples 19

Index

Page 97: Analizator ph, o2 knicks

97

Connection to supply units 80Control Drawings 92CSAControlDrawings94

DDate 55

Display 69Deviceself-test72Device type, display 74Diagnostics 24, 70

Calibration data 71Deviceself-test72Sensor data 71Sensor monitor 74Version 74

Digital sensors 56Dimensions 12Display 21

Display test 72Display backlighting 21Disposal 2Documentation 3

EEC Declaration of Conformity 90EEPROMtest72Electronic Signature 95Enclosure 12Enclosure components 11Entering values 23Error handling 86Error messages 87Explosion protection 84Extended logbook 73, 95

FFDA21CFRPart1195FLASHtest72FMControlDrawings93

Index

Page 98: Analizator ph, o2 knicks

98

IndexHHOLD24, 26

End 26ExternalactivationofHOLD27ManualactivationofHOLD27OutputsignalduringHOLD26, 47Outputsignalresponse26

IInfo text 87Installation 15

Hazardouslocations9Intended use 7IrDAcommunication76

KKeypad 20

LLogbook 73

MMeasuredvalues,display74Measurement69Measuring22Measuringmodeselection38Measuringrange43Memosens56

Connection 57Ex wiring 9, 84Sensor replacement 58

Memosenscable18Menustructure25

Configuration 28Moduletest72Mountingplan12

Page 99: Analizator ph, o2 knicks

99

IndexOOperatingmodes24Operatingmode,selection23Operatingstates78Options77, 79

TAN77Ordercode79Outputcurrent,fixedvalue76Outputfilter44OutputsignalduringHOLD26, 47Overview10

PPackage contents 3, 11Panel mounting 14Parameter error 86ParametersetA/B29

Display 69Manualselection30

Passcodes 95, 104Assigningapasscode77

Pipe mounting 13Pointofmeasurement(TAG)55Power supply units 80Pressure correction 50Product calibration 64Product line 79Protective hood 13

RRAMtest72Ratingplates16Registeredtrademarks103Releaseofoptions77Resettofactorysettings77Returnofproductsunderwarranty2

Page 100: Analizator ph, o2 knicks

100

IndexSSafety information 7, 8Safety instructions 3Salinity 50Selection menu 23Sensocheck

Configuration 53Sensoface 86Sensor connection 19Sensor data, display 71Sensor monitor 74, 76Serial number, display 74Service 24, 75

Factorysetting77IrDAcommunication76Passcodes 77Releasingoptions77Sensor monitor 76Specifying current outputs 76

Service passcode lost 77Signal colors 21Signal lines 17Slope calibration 66

Calibration medium 41Software version, display 74Specifications 81Start-Up8Supply units 80

TTagnumber(TAG)55TANoptions77, 79

EnteringaTAN77Temperature probe adjustment 68Terminal assignments 16Terminals 9, 15, 16Time 55

Display 69

Page 101: Analizator ph, o2 knicks

101

Time averaging filter 45Trademarks 103

UUser Interface 20

WWarranty 2Wiring 17

Examples 19Power supply units 80

ZZero calibration 62

Index

Page 102: Analizator ph, o2 knicks

102

Index

Page 103: Analizator ph, o2 knicks

103

Trademarks

Thefollowingnamesareregisteredtrademarks.Forpracticalreasonsthey are shown without trademark symbol in this manual.Stratos®Sensocheck®Sensoface®Calimatic®GainCheck®InPro®isaregisteredtrademarkofMettler-Toledo.Memosens® isaregisteredtrademarkofEndress+HauserConductaGmbHandKnickElektronischeMessgeräteGmbH&Co.KG.HART®isaregisteredtrademarkoftheHARTCommunicationFoundation.

Page 104: Analizator ph, o2 knicks

TA-212.145-MS-KNE01 20091218 Softwareversion:1.x

KnickElektronische Messgeräte GmbH & Co. KGP.O.Box370415D-14134Berlin

Phone:+49(0)30-80191-0Fax: +49(0)30-80191-200Internet: http://[email protected]

PasscodesIntheSERVICE–CODESmenuyoucanassignpasscodestoprotecttheaccess to certain functions.

Mode of operation Passcode

Service (SERVICE) 5555

Diagnostics (DIAG)

HOLD mode

Calibration (CAL)

Configuration (CONF)