Formaldehyde Exposure Measurement at Capital Health
www.capitalhealth.ca 2
Formaldehyde - Introduction
Formaldehyde uses at Capital Health:
• Pathology
• Morgue
• Tissue preservation
www.capitalhealth.ca 3
Formaldehyde - Introduction
Tissue Preservation
• 10% buffered neutral formalin
• Volumes used 20 ml to 20 L
www.capitalhealth.ca 4
Formaldehyde-Introduction
Tissue Preservation - Pathology
www.capitalhealth.ca 5
Formaldehyde Exposure Limits
Alberta Occupational Health and Safety Code
• 8-hour TWA OEL- 0.75 ppm
• Protect against carcinogenicity
• Ceiling limit – 2 ppm
• Protect against irritation of eyes, nose and upper respiratory tract
www.capitalhealth.ca 6
Formaldehyde Exposure Limits
College of American Pathologists and College of Physicians and Surgeons of Alberta Accreditation Requirements:
• 8-hour TWA exposure limit- 0.75 ppm
• 8-hour TWA action level – 0.5 ppm
• 15 min STEL – 2 ppm
• Must be maintained below these levels in areas of Anatomic Pathology
• After initial monitoring, further monitoring is mandated if results equal or exceed 0.5 ppm ( action level) or 2 ppm STEL
www.capitalhealth.ca 7
Formaldehyde Exposure Limits
College of American Pathologists and College of Physicians and Surgeons – Cont’d
• Initial monitoring must be repeated anytime change in production, equipment, process, personnel, or control measures which may result in new or additional exposure to formaldehyde.
• Periodic monitoring is mandated only if the initial monitoring is at or exceeds 0.75 ppm (8 hour TWA) or 2 ppm ( STEL).
• If any personnel report signs or symptoms of respiratory or dermal conditions associated with formaldehyde exposure, the laboratory must promptly monitor the affected person’s exposure
• Requires that monitoring be performed in accordance with Alberta Occupational Health & Safety Act, code and Regulations.
www.capitalhealth.ca 8
Formaldehyde Exposure Limits
Alberta Occupational Health and Safety Code , Part 4 Chemical Hazards, Biological Hazards and Harmful Substances 20(1):
• ‘If a person measures the airborne concentration of a harmful substance for the purposes of complying with the occupational exposure limits as required by the Code, the person must make the measurement
• (a) in accordance with the NIOSH Manual of Analytical Methods, 4th Edition (August 1994), …. including the 2nd supplement (January 15, 1998), or
• (b) using methods or procedures that are approved by the Director of Occupational Hygiene….”
www.capitalhealth.ca 9
Formaldehyde- Potential Exposure Incidents
What Kind of Sampling / Measurement is Needed?
• TWA( 8 hour )
• Short term (STEL)
• Ceiling limit
www.capitalhealth.ca 10
Formaldehyde- Potential Exposure Incidents
www.capitalhealth.ca 11
Formaldehyde- Potential Exposure Incidents
www.capitalhealth.ca 12
Formaldehyde- Potential Exposure Incidents
www.capitalhealth.ca 13
Formaldehyde- Potential Exposure Incidents
www.capitalhealth.ca 14
Formaldehyde- Measurement
What measurement method is needed to measure Ceiling Limit exposures?
• What is available?
www.capitalhealth.ca 15
Formaldehyde Exposure Limits
Alberta Occupational Health and Safety Code , Part 4 Chemical Hazards, Biological Hazards and Harmful Substances 20(1):
• ‘If a person measures the airborne concentration of a harmful substance for the purposes of complying with the occupational exposure limits as required by the Code, the person must make the measurement
• (a) in accordance with the NIOSH Manual of Analytical Methods, 4th Edition (August 1994), …. including the 2nd supplement (January 15, 1998), or
• (b) using methods or procedures that are approved by the Director of Occupational Hygiene….”
www.capitalhealth.ca 16
Formaldehyde- NIOSH Methods
NIOSH
MethodDescription
Max.
Flow rate
(L/min.)
Min
Volume
(L)
Min TimeDetection
Limit
Minimum
Concentration
2016DNPH treated silica
gel tubes- HPLC1.5 1 40 sec 0.23 ug 0.2 ppm
2541HMP treated XAD-
2 tubes-GC0.1 1 10 min 3 ug 2.7 ppm
3500Impinger, bisulfite-
colorimetric1.0 1 1 min 2 ug 1.8 ppm
3800Extractive FTIR-
direct readingNA NA
1 min to fill
gas cell
(Miran
SapphIRe)
0.4 ppm
(10 meter
gascell)
0.4 ppm
www.capitalhealth.ca 17
Formaldehyde- Measurement
No evaluated NIOSH methods were found suitable for Ceiling Limit formaldehyde measurement.
• Direct reading instrument must be used
www.capitalhealth.ca 18
Formaldehyde- Measurement
Requirements of Direct Reading Instruments:
• Instrument method must be sensitive to 0.05 ppm
• Instrument must be accurate (±25% of true reading 95% of time-NIOSH/OSHA criteria)
• Instrument method must be able to take readings in as close to real time as possible (~ 10 second response time or less)
• Instrument method must be free of other interferences (formalin contains a high proportion of methanol, health care has many other background VOC’s)
www.capitalhealth.ca 19
Formaldehyde- Measurement
Requirements of Direct Reading Instrument – Cont’d:
• Data logging capability so that Ceiling Limit exceedances can be captured, then correlated to work log
• Method must be approved by Director of Occupational Hygiene-Alberta Occupational Health and Safety
www.capitalhealth.ca 20
Formaldehyde- Measurement
Direct Reading Instruments Available:
• Formaldemeter - poor response and interference from methanol and other oxidizable VOCs.
• CEA Instruments - very sensitive, too slow response to give real time, potential interferences.
• Miran sapphIRe - 1 minute cell fill time, not real time instrument, limited sensitivity, interferences.
• Detector tubes (Dragar, Gastec)- too slow response to give real time, poor sensitivity
www.capitalhealth.ca 21
Formaldehyde- Interscan
www.capitalhealth.ca 22
Formaldehyde- Measurement
Literature Review References for Interscan Instrument:
• Liu,F et al reported no statistical difference between Interscan 4160 and a phenol reagent method (J. Env. Health., Vol 20, No.2, 2003.)
• NCASI Technical Bulletin 579 Laboratory and Field Evaluation of a Portable Continuous Analyzer for Measurement of Formaldehyde in Workplace Atmospheres
• 0.3 - 8 ppm - mean difference was 3.5%, coeff. var. 0.01-0.06
www.capitalhealth.ca 23
Formaldehyde- Measurement
Literature Review References for Interscan Instrument Cont’d:
• NIOSH procedure RCT-APR-STP-0039, Determination of Formaldehyde Service Test Life, Air Purifying Respirators Standard Testing Procedure
• Mentions Interscan to give “real” time numbers in order to determine when formaldehyde breakthrough occurs in air purifying cartridges.
www.capitalhealth.ca 24
Formaldehyde- Measurement
Literature Review References for Interscan Instrument Cont’d:
• US EPA Final Report IE 2814 Residential Indoor Air Formaldehyde Testing Program: Pilot Study
• Used Interscan for real time formaldehyde measurements in conjunction with integrated sampling methods for building material off-gassing testing.
• US EPA Method 21 Determination of Volatile Organic Compound Leaks
• Gives specifications and performance criteria of direct reading instruments used in leak detection.
www.capitalhealth.ca 25
Formaldehyde- Interscan
Method of function
Electro-chemical voltametric sensor
Electro-catalytic sensing electrode
Sample pumped into instrument , then diffuses across medium to
electrode
Sensor electrolyte immobilized similar to electrolyte in alkaline batteries
No problem with leakage, orientation of sensor, sensor noise
Sealed reservoir eliminates reference electrode contamination
Very linear response allows single point calibration over wide range
www.capitalhealth.ca 26
Formaldehyde- Interscan
Sensitivity
• -0.1 ppm
Accuracy
• - 0.3 -8 ppm- mean difference was 3.5%, coeff. var. 0.01-0.06
Real Time Response?
• Almost immediate, <5 seconds
Interferences?
• Minimal to other VOC’s and alcohols, instrument drift reported
www.capitalhealth.ca 27
Formaldehyde- Interscan
Data Logging Capability?
• Yes - with external data logger (ACR Systems SmartReader7)
Approved by Director of Industrial Hygiene?
• No
• Request to Alberta Human Resources and Employment for approval of Interscan 4000 series for formaldehyde-Sep/06
• Dec / 06 approval granted with conditions:
www.capitalhealth.ca 28
Formaldehyde Interscan Approval
• Conditions:
• Specific to Capital Health facilities
• Normal calibration is done at least quarterly when performing according to specifications within 35-65% RH.
• Recalibration required on regular basis when RH outside 35-65%
• Regular replacement of sensor due to reports of drift (manufacturer recommends quarterly or calibration check).
www.capitalhealth.ca 29
Formaldehyde- Interscan Test Trial
• Calibration performed at ALS Laboratories.
• 1 m3 stainless steel calibration chamber
• Paraformaldehyde powder diluted with inert silica powder, preweighed amounts heated to vaporize formaldehyde in chamber.
• Formaldehyde concentrations measured by variety of methods.
www.capitalhealth.ca 30
Formaldehyde- Interscan Test Trial
• Formaldehyde concentrations measured by variety of methods:
• NIOSH 2016- DNPH treated silica gel tube--HPLC analysis
• NIOSH 3500- Impinger/bisulfite-chromotropic acid-colorimetry.
• Interscan 4000
• Miran SapphIRe
www.capitalhealth.ca 31
Formaldehyde-Interscan Test Trial
Form aldehyde Method Com parison
0
0.5
1
1.5
2
2.5
3
3.5
4
0 0.5 1 1.5 2 2.5 3 3.5 4 4.5
Obtained concentration (ppm)
NIOSH 2016NIOSH 3500Interscan-adjustedMiran
www.capitalhealth.ca 32
Formaldehyde- Survey Results
• Location- Anatomical Pathology Lab- specimen preparation/ grossing.
• Workers- 2 histology technicians, one pathology assistant
• Samples collected-
• Six 8 hour TWA samples on each worker –
• NIOSH 2016- DNPH treated silica gel/ HPLC analysis
• Random / as-needed Ceiling Limit samples using Interscan 4000
www.capitalhealth.ca 33
Formaldehyde- Survey Results
TWA Sample Results
Description Sample Time (min.) Air Concentration (ppm)
Pathology Assistant 367 0.051
Histology Technician 456 0.11
Histology Technician 414 0.030
Area- specimen receiving 473 / 474 0.017 / 0.02
Area- grossing 471 0.02
Area- specimen storage 471 0.11
www.capitalhealth.ca 34
Formaldehyde- Survey Results- GROSSING
Ceiling / Activity Specific Instantaneous Interscan Readings
Description Ventilation Type Air Concentration (ppm)
Grossing very large specimen Local hood 0.050, 0.070, 0.070
Placing large specimen back
into storage bucketLocal hood 0.100
Grossing smaller specimen Local hood 0.065
Returning specimen to
container on floorGeneral ventilation 0.080
Removing fixed specimen from
containerLocal hood 0.04
www.capitalhealth.ca 35
Formaldehyde- Survey Results- GROSSING
Ceiling / Activity Specific Instantaneous Interscan Readings
Description Ventilation Type Air Concentration (ppm)
Grossing medium sample Local hood0.045, 0.050, 0.060, 0.050,
0.070
Returning medium sample to
container-some splashingLocal hood 0.120
Grossing smaller specimen Local hood 0.040, 0.045
Dumpling formalin into waste
containerGeneral ventilation 0.065
Removing fixed specimen from
containerLocal hood 0.055
www.capitalhealth.ca 36
Formaldehyde- Survey Results- GROSSING
Ceiling / Activity Specific Instantaneous Interscan Readings
Description Ventilation Type Air Concentration (ppm)
Removing medium sample from
containerLocal hood 0.065
Grossing medium specimen Local hood 0.060
Opening specimen transport
box- some formalin spillageGeneral only 1.60
Removing endoscopy samples
from plastic bagsGeneral only 0.40, 0.65
Over garbage can (above
plastic bags disposed of here)General only 1.10, 0.45
www.capitalhealth.ca 37
Formaldehyde- Results- AUTOPSY SUITE
Ceiling / Activity Specific Instantaneous Interscan Readings
Description Ventilation Type Air Concentration (ppm)
Specimen disposal General only 0.80, 1.60, >1.99
Over open bucket of discarded
used specimensGeneral only 1.10
Room background- middle of
autopsy suite, door closedGeneral only 0.25
Addition of formaldehyde
neutralizer and stirring to neutralize
waste
General only 0.45
www.capitalhealth.ca 38
Formaldehyde- Results- AUTOPSY SUITE
Ceiling / Activity Specific Instantaneous Interscan Readings
Description Ventilation Type Air Concentration (ppm)
Over open bucket of waste formalin General only <1.99
Over open bucket of ‘deformalized’
formalinGeneral only 0.90
In specimen storage room-door
openGeneral only 0.20
In specimen storage room-after
door closed 1 minuteGeneral only 0.45
www.capitalhealth.ca 39
Formaldehyde- Interscan
Discussion:
• Interscan gave reliable, real-time information for formaldehyde concentrations
• TWA average samples indicated that average concentration was well below OEL’s
• However, Interscan identified activities/ tasks that exceeded or had potential to exceed Ceiling Limit OEL-
• very variable readings
• most likely log normally distributed with high standard deviation- highly skewed istribution
www.capitalhealth.ca 40
Formaldehyde- Interscan
Discussion Cont’d:
• Added benefit:
• worker is shown in real time the consequences of performing certain activities,
• can make practice improvements
• can immediately reassess changes in exposures from performing improved procedures
• this is important in reducing exposures through education on chemical handling with Ceiling Limit OELs.