Broken Austin Moore Prosthesis Fatigue or Failure? Vinod Naneria Choithram Hospital & Research centre, Indore, India

Broken austin moore prosthesis fatigue or failure

Embed Size (px)

Citation preview

Page 1: Broken austin moore prosthesis  fatigue or failure

Broken Austin Moore ProsthesisFatigue or Failure?

Vinod Naneria

Choithram Hospital & Research centre, Indore, India

Page 2: Broken austin moore prosthesis  fatigue or failure

A broken stem of Austin Moore’s hip prosthesis

Page 3: Broken austin moore prosthesis  fatigue or failure

Purpose of presentation

Fatigue Failure• It is sure implant failure.

• It requires revision.

• There is significant anterior thigh pain.

• There are loose, hence stem failure occur.

• It has worked for so many years, that’s the reason it has failed.

• Yes…………………………………!

• Just a metal fatigue

• It does not requires any treatment.

• Hip movements are normal.

• Proximal part – femoral head – shaft are rigidly fixed.

• It has worked for so many years – how can it fail?

• No……………………………..!

Page 4: Broken austin moore prosthesis  fatigue or failure

The debate?

• Broken stem of AMP –

– Is it loose?

– Has it failed after serving for many years?

– Why it broke?

– Is it same as broken stem of a cemented THR?

– Does it need revision?

– Is revision mendatory?

Not loose – Not failed – No revision

Page 5: Broken austin moore prosthesis  fatigue or failure

Observations

• All AMP were done in young age lasted long.

• Good quality bone support AMP well.

• Bone growth is noted in fenestrations of AMP.

• Hypertrophy of the calcar region noted in all cases.

• There was no sign of loosening in any case.

• There was no sinking of proximal prosthesis in femoral canal.

Page 6: Broken austin moore prosthesis  fatigue or failure

Observations - THR

• In THR stem brake due to fixed distal part and loose proximal part.

• Since proximal part is loose, it needs revision.

• In contrast to THR, breakage of AMP stem is due to loose distal part and a Fixed proximal part.

Page 7: Broken austin moore prosthesis  fatigue or failure

Feb 2012

Page 8: Broken austin moore prosthesis  fatigue or failure

Oct 2013 January 2014

Loose proximally Loose & broken stem

Page 9: Broken austin moore prosthesis  fatigue or failure

Well fixed stem Loose stem

Broken stem

Page 10: Broken austin moore prosthesis  fatigue or failure

Observations

• Even in a well fixed prosthesis, there are constant bending strain of cyclic loading leading to remodeling of the proximal femur and to varus setting of the head-neck of prosthesis.

• A broken stem of AMP is a metal fatigue failure.

Page 11: Broken austin moore prosthesis  fatigue or failure

Observations

• Since proximal part is well fixed, revision is not required.

• The initial pain following breakage of stem is due to change in the distribution of mechanical forces, as the lateral femoral cortex start bearing more load.

Page 12: Broken austin moore prosthesis  fatigue or failure

Bone in fenestration Bone hypertrophy at the tip Bone hypertrophy lower lateral cortex

Page 13: Broken austin moore prosthesis  fatigue or failure

Case one

• 18 years old lady had a puerperal sepsis - 1982

• Acute renal shut down ended up in CRF.

• Had fracture neck of femur -1985.

• AMP was done considering failures of internal fixation in CRF patients(Past experience of failure in two cases) and limited life expectancy.

• Had Renal transplant in 1987

• She was 7th case of renal transplant at Choithram.

Page 14: Broken austin moore prosthesis  fatigue or failure

Case history continue…

• On regular follow up.

• Delivered a baby girl 1991.

• Severe obesity.

• Stem broken noted in 1995 (x-rays)

• Last follow up 2006.

• Had Ca Breast with multiple mets – died 2007.

• No x-ray changes noted between 1995 – 2006.

Page 15: Broken austin moore prosthesis  fatigue or failure

Case one1995 2006

Page 16: Broken austin moore prosthesis  fatigue or failure

2006

Page 17: Broken austin moore prosthesis  fatigue or failure

Case Two

• 54 years old Female

• AMP done out side.

• Had acute onset of anterior thigh pain May 2012.

• A suggestion was given to for Total Replacement out side.

• Seen May 2012 with Broken Stem

• Full range free movements at hip

• Last seen Dec 2012.

Page 18: Broken austin moore prosthesis  fatigue or failure

Feb 2010

May 2012

Dec 2012

Lateral corticalHypertrophy

Case Two

Page 19: Broken austin moore prosthesis  fatigue or failure
Page 20: Broken austin moore prosthesis  fatigue or failure

Case Three

Lateral cortical hypertrophy and migration of loose distal fragment medially

Page 21: Broken austin moore prosthesis  fatigue or failure

Case Three

• This old man had AMP left hip before 1992 (X-rays)

• He had fracture trochanter on right side –2000.

• DHS was done in 2000.

• Last follow up 2006.

• Rigid fixation by bone in fenestrations.

• Medial migration of distal loose piece.

Page 22: Broken austin moore prosthesis  fatigue or failure

Case Four

1992

Page 23: Broken austin moore prosthesis  fatigue or failure

Case four

2000

Page 24: Broken austin moore prosthesis  fatigue or failure

Case four

2006

Page 25: Broken austin moore prosthesis  fatigue or failure

Case Five

• 22 years old male had RTA – fracture neck femur. 1986

• Operated by multiple Knowles Pins.

• Broken pins noted – 1988 – fracture united.

• AVN – 1990 – AMP done.

• Pain in anterior thigh 2002 due to broken stem.

• Last follow up 2006.

Page 26: Broken austin moore prosthesis  fatigue or failure

1986 Knowles pinning Pins breakage 1988

By – Late Prof. V. S. Inamdar

Page 27: Broken austin moore prosthesis  fatigue or failure

AVN 1990 AMP 1990

By – Late Prof. V. S. Inamdar

Page 28: Broken austin moore prosthesis  fatigue or failure

2002 2006

Page 29: Broken austin moore prosthesis  fatigue or failure
Page 30: Broken austin moore prosthesis  fatigue or failure
Page 31: Broken austin moore prosthesis  fatigue or failure
Page 32: Broken austin moore prosthesis  fatigue or failure

Case sixth

• This 82 years old male had an AMP – 1993

• Complains of acute anterior thigh pain since last one month.

• A broken stem noted Sept, 2013.

• Was suggested Total Joint Replacement out side.

• There is no difficulty in ADL.

• Left as such under observation.

Page 33: Broken austin moore prosthesis  fatigue or failure
Page 34: Broken austin moore prosthesis  fatigue or failure

Case six

Page 35: Broken austin moore prosthesis  fatigue or failure

Case seventh

Page 36: Broken austin moore prosthesis  fatigue or failure

Case eighth

• 36 yrs Male• Bilateral AVN hips• AMP done 1988 rt side at the age of 36 yrs• THR cemented done on left 1990 at the age of 38.• Revision THR for shattered femur 2007• Sciatic palsy• Removal of Memmen plate and wires April 2009.• Removal of THR Sept 2009.• Metal fatigue 2012• Last follow up Nov 2014.

Page 37: Broken austin moore prosthesis  fatigue or failure

Case Eighth

• Acetabular poly wear 2007.

• Sustained periprosthetic fracture with shattered femur.

• Non cemented “Solution Hip” with Mammamplate and circlage wires + bone grafting of acetabulum.

• Sciatic nerve palsy.

Page 38: Broken austin moore prosthesis  fatigue or failure

Case Eighth

• Persistent pain in hip

• Plate and wires removed in April 2009

• Pain in knee and hip persisting

• Solution hip system removed for suspected low grade infection Sept. 2009.

• Serial x-rays showing broken amp in Jan 2012.

• Last follow up 2014.

• Sciatic nerve has partially improved.

Page 39: Broken austin moore prosthesis  fatigue or failure

2000

10 years post THR12 years post AMP

Page 40: Broken austin moore prosthesis  fatigue or failure

June 2004

Signs of poly wear

Page 41: Broken austin moore prosthesis  fatigue or failure

Dec 2005

Well fixedExtensive poly wearWaiting for revision

Page 42: Broken austin moore prosthesis  fatigue or failure

Severe poly wearBadly comminutedVancouver B3Fracture of shaft +Proximal loosening Loose cement Loose prosthesis

June 2007

Page 43: Broken austin moore prosthesis  fatigue or failure

June 2007

Page 44: Broken austin moore prosthesis  fatigue or failure

June 2008 April 2009

Page 45: Broken austin moore prosthesis  fatigue or failure

April 2009

Persistent pain at hip and kneeSonography hip ? Small collectionAspiration – culture negative

Page 46: Broken austin moore prosthesis  fatigue or failure

Sept 2009

Removal of prosthesis due to pain

Page 47: Broken austin moore prosthesis  fatigue or failure

Sept 2009

21 years post implantationPain freeWell fixed AMPHypertrophy of CalcarBone in the fenestrationBone at the tip of AMPAmbulation with cane support.Lt. Sciatic nerve improving

Page 48: Broken austin moore prosthesis  fatigue or failure

Dec 2010

Page 49: Broken austin moore prosthesis  fatigue or failure

Incidental findingSigns of metal fatigue

2012

Page 50: Broken austin moore prosthesis  fatigue or failure

Nov 2014

No painNot aware of broken prosthesis

AMP1988 - 2014

Page 51: Broken austin moore prosthesis  fatigue or failure

Summary & Conclusion

• Broken stem is not a implant failure.

• Broken stem is not a sign of loosening of implant.

• Cyclic loading of bending forces are probably the cause.

• A follow up of cases of broken stem for more than 14 years, did not show any deterioration in fixation of proximal implant.

• There was hypertrophy of the lateral cortex of femur at the site of broken stem.

Page 52: Broken austin moore prosthesis  fatigue or failure

DISCLAIMER

Information contained and transmitted by this presentation is based on personal experience and collection of cases at Choithram Hospital & Research centre, Indore, India. It is intended for use only by the students of orthopaedic surgery. Views and opinion expressed in this presentation are personal. Depending upon the x-rays and clinical presentations viewers can make their own opinion. For any confusion please contact the sole author for clarification. Every body is allowed to copy or download and use the material best suited to him. I am not responsible for any controversies arise out of this presentation. For any correction or suggestion please contact [email protected]