opyright The British Standards Institution 2010 M5 Amendment of EU Directive 93/42/EEC - regulatory update Jan van Lochem – Gert Bos – Suzanne Halliday 5 November 2009 HKTDC Hong Kong International Medical Devices and Supplies Fair 2009 Seminar “Regulatory update on medical devices in Asia and EU” For more information on healthcare standards visit: http://shop.bsigroup.com/healthcarestandards

EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

Embed Size (px)

DESCRIPTION

BSI British Standards' presentation of EU Medical Devices Directive M5 Amendment 93 42 EEC regulatory updates. Presented at the HKTDC Hong Kong International Medical Devices and Supplies Fair 2009.By Jan van Lochem, Gert Bos and Suzanne Halliday. Seminar “Regulatory update on medical devices in Asia and EU”.It covers the revision of the EU medical device directive, implementation of the revised directive, the key changes, changes to clinical requirements, introduction of technical file sampling, and the impact revision on technical or design dossiers.

Citation preview

Page 1: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

M5 Amendment of EU Directive 93/42/EEC - regulatory update

Jan van Lochem – Gert Bos – Suzanne Halliday5 November 2009

HKTDC Hong Kong International Medical Devices and Supplies Fair 2009Seminar “Regulatory update on medical devices in Asia and EU”

For more information on healthcare standards visit: http://shop.bsigroup.com/healthcarestandards

Page 2: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

Content

• 1. Revision of the EU medical device directive

• 2. Implementation of the revised directive

• 3. Key changes

• 4. Changes to clinical requirements

• 5. Introduction of technical file sampling

• 6. Impact revision on technical or design dossiers

Page 3: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

1. Revision of the EU medical device directive

Page 4: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

History

• “The Commission shall, no later than 5 years from the date of implementation…..submit a report .....and if necessary make appropriate proposals” 93/42 Article 11.4

• It was decided that the review should:

Cover all aspects

Be open to all stakeholders

Page 5: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

Main Findings of Report

• Directives provide appropriate legal framework and, in general, work well

• Areas needing improvements: Conformity assessment

Designation & control NoBos

Clinical data & PMS

Member State resources devoted to implementation

• Most areas of improvement could be addressed by guidance

Page 6: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

2. Implementation of the revised directive

Page 7: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

Timescales for Implementing

• 2007/47/EC - 5th (M5) amendment of 93/42/EEC http://ec.europa.eu/enterprise/medical_devices/revision_mdd_en.htm

• MS required to transpose ‘M5’ into National Laws by 21 December 2008

Manufacturers and NoBo may start to implement

Directive fully enforced in all MS by 21 March 2010Manufacturers and Notified Bodies must be in compliance

Page 8: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

Q&A around amended MDD

• Do certificates change after 21 March 2010 2007/47 is amending directive, not stand-alone one Existing certificates remain valid; no need to re-issue

them on 21 March 2010 No need to refer to M5 or 2007/47 on certificates

• Do Notified Bodies audit to M5 before March 2010 Raise observations / OFIs rather than non-conformities to

cover the additional requirements Show M4 or M5 as audit basis in Notified Body

documents and reports

Page 9: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

Q&A around amended MDD

• Can Manufacturers use Essential Requirement Checklists including M5 requirements now?

Yes; but they must agree with their Notified Body which versions of their procedures are “official” up to 21 March 2010

• Is there a requirement to have had an M5 compliance audit before 21 March 2010?

No; except where a change in classification requires a different conformity route or additional review by NB

• After 21 March 2010 all audits must be to 93/42/EEC as amended by 2007/47/EC ie M5

Page 10: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

State of play certificate changes

• EC Certificates DO NOT need updating to refer to 2007/47/EC

• NO assessment required to 2007/47/EC before March 2010 to maintain EC QA Certificates (Annex II, V & VI)

• NO requirement to review and update EC Design Dossier Examination Certificates before March 2010

• NEW certificate needed in some reclassifications

Evolving guidance, further clarifications expected before March 2010

Page 11: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

3. Key changes

Page 12: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

Key MDD Changes

• Software in definition, single use devices defined

• Classification changes, PPE, MD

• Alternative labelling (draft directive)

• Clinical evaluation

• Single authorized rep

• Improved communication / Confidentiality

• Essential requirements

• Quality system & PMS

• Techfile sampling

Page 13: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

Essential requirements

• ER 6a: - always clinical evaluation

• ER 7.1: - biophysical modelling introduced

• ER 7.4: - major rewording combination products

• ER 7.4: - consultation on changes

• ER 7.5: - carcinogenic, mutagenic, toxic for reproduction, phthalates (67/548/EEC)

• ER 12.1a - Validation of software

• ER 13.3 - Authorized representative

• ER 13.3 - Single use consistency

• ER 13.6 - risks re-use IFU, latest revision IFU

Page 14: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

Impact on QA Assessments - General

• 2007/47 makes some changes to the assessment under the QA modules, eg:

Documentation of control of subcontractors

Procedure for clinical evaluation

Pro-active post market surveillance

• BSI already emphasises these points as they are both explicit in ISO 13485, and active PMS was already identified in ISO 14971 and NB-rec on PMS

Page 15: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

Impact on QA Assessments - General

• Surveillance audit focuses more on clinical follow-up data, PMS, update risk management file

• More audit time expected due to introduction of Technical File sampling class IIa and IIb devices

Page 16: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

All Annexes Document Retention Times

• What it Says The manufacturer or his authorised representative must,

for a period ending at least five years, and in the case of implantable devices at least 15 years, after the last product has been manufactured, keep (records) at the disposal of the national authorities

• What it Does Increases document retention time (in line with ISO

13485) Distinguishes implants from other products Appears to be unrelated to product lifetime

Page 17: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

Confidentiality

• Not treated as confidential: Name manufacturer / authorized rep

Certificate info including issuance, modifications, suspensions, withdrawals

Page 18: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

4. Changes to clinical requirements

Page 19: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

Key MDD Changes – clinical

• Clearer Definitions (Annex X) Clinical Data – safety / performance data (literature,

market information, clinical investigation)

Clinical Evaluation – review of any type of clinical data

Clinical Investigation – clinical study involving human subjects

Clinical Evaluation of Clinical Data is required for CE Marking

• Alignment with MEDDEV 2.7.1 and ISO 14155

Page 20: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

Key MDD changes – clinical data

Page 21: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

Annex X – Clinical Evaluation

• Clinical evaluation for all medical devices

• Where demonstration of conformity based on clinical data is not deemed appropriate, compliance to ERs must be justified through risk management output, performance evaluation, bench testing and preclinical evaluation. Reason duly substantiated.

• For implants and class III medical devices: clinical evaluation unless…….

Clinical evaluation in technical file (or full reference)

Clinical evaluation and documentation must be actively updated based on PMS data

Page 22: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

Clinical investigations

• Data relating to clinical investigation in Europe in European database

• Any interruption, modification, refusal of clinical investigation in EU member states communicated to other Member States in the EU

Page 23: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

5. Introduction of technical file sampling

Page 24: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

Key MDD changesTF sampling IIa and IIb Devices

• What it Says For devices in Class IIa the notified body shall assess

…… the technical documentation … for at least one representative sample for each device subcategory ….

For devices in Class IIb the notified body shall assess ….. The technical documentation … for at least one representative sample for each generic device group …

• What it Does Sets out different sampling regimes for Class IIa and

Class IIb Derived from GMDN terminology (BS EN ISO

15445:2000) GMDN definitions are reproduced in Article 1

Page 25: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

Key MDD Changes – Definitions

‘device subcategory’ means a set of devices having common areas of intended use or common technology (class IIa)

‘generic device group’ means a set of devices having the same or similar intended uses or commonality of technology allowing them to be classified in a generic manner not reflecting specific characteristics (class IIb) Critical for sampling of IIa and IIb technical documentation

Page 26: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

GMDN – a method of grouping devices

Page 27: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

Technical Documentation Sampling by Notified Bodies• Incorporated in audit plan Notified Bodies, extra

audit time Already a MEDDEV 2.10 “requirement”

• Adds similar sampling under Annex V and VI

• Defines sampling regimes

• NBOG Guidance 2009-4 (July 2009) applicable

Page 28: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

BSI approach to sampling

• Sampling regimes will be based on rationale agreed with manufacturer

BSI objective is to deliver required increased scrutiny in a pragmatic and practical way

• Class IIa Based on “subcategories” identified in manufacturer scope on

certificate

Use of NBOG guidance 2009-3 on product categories

• Class IIb Manufacturer starts from GMDN generic device group definitions

BSI, in agreement with manufacturer, rationalises down to representative samples

Page 29: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

Class IIa – Device subcategory

Page 30: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

Latest NBOG Guidance on TF sampling,reduced sampling numbers• For Class IIb devices: sample of generic device

groups, referenced by different GMDN preferred terms, to following plan:

Up to 2 groups: a sample from each group

Up to 10 groups: a sample from 3 (was 6) of these groups

Up to 20 groups: a sample from 5 (was 7) of these groups

Up to 30 groups: from 7 (was 8) groups a sample

N > 30 groups: from N/10 + 5 groups a sample

Page 31: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

Sampling –BSI approach

• Class IIa devices: a sample of each group over a 5 year period.

• Class IIb devices: determine sample plan for a 5 year period.

• The Sampling plan updated periodically, at least each 3 years.

Page 32: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

Depth of Assessment of Technical Files

• Not full Design Dossier review, but it will cover “essentials”

• No differentiation between Class IIa and IIb

• Guidance from NBOG indicates that assessment of each set of technical documentation should consist of review of:

classification validity of essential requirements checklist risk management pre-clinical data (studies in animal models, biocompatibility,

nanotechnology etc.) clinical data (Note NoBo review as per MEDDEV 2.7.1) declaration of conformity, and other technical documentation based on risk

Page 33: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

Sampling conclusions

• Major new requirement for Annex V: involvement NB in review of technical documentation

• Sampling regime

• Depth review based on NBOG guidance

• NBs need to agree sampling regimes with manufacturers as soon as possible

• Major new requirement for both NoBo and manufacturers

Page 34: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

6. Impact revision on technical or designdossiers

Page 35: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

2007/47/EC Impact on Technical Documentation• Notified Bodies, Manufacturers - Full Compliance

Required by 21 March 2010 Transition started December 2008

• New ER’s & Clinical Evaluation Requirements

• Proactive Review of Amendments by Manufacturer Impact on Technical Documentation & DoC

Confirmation Amendments Have Been Considered and if Appropriate Addressed

By Product or Manufacturers’ Full Scope, as Appropriate

Page 36: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

Design Examination or Type Examination Certificates

• Manufacturer updates DoC to reference 2007/47/EC amendments when satisfied that product complies

• NB will review compliance with modified Directive during Design Dossier renewals / changes and audits after 21 March 2010

• Irrespective manufacturer is responsible for ensuring full compliance with 2007/47/EC amendments from 21 March 2010

Page 37: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

2007/47/EC Clarifications on Technical Documentation• Annex II Technical Documentation

A statement indicating whether or not the device is manufactured utilizing tissues of animal origin as referred to in Commission Directive 2003/32/EC

• Annex II, V & VI Technical Documentation The solutions adopted as referred to in Annex I Section I (2),

• the preclinical evaluation,

• the clinical evaluation referred to in Annex X

• Annex X Confirm that plan for post market clinical follow-up exists and is

acted upon and justification for no post market clinical follow-up is documented

Page 38: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

Summary

• March 2010 full compliance required No effect on classification?

5 year sampling plans agreed with Notified Body

Labelling checked and updated

Devices already on the market

Devices in stock

Updated technical documentation

Update declarations of conformity

Page 39: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

What is next?

• In June 2008, EC started recast

• Significant revision of legal framework proposed Converge on global model

Harmonize on national differences, reduce fragmentation

Strengthen clinical requirements

Harmonization quality and oversight Notified Bodies

Adapt to new technologies

Involve possible EMEA for high risk devices

• EC preparing next steps, talking to stakeholders

Page 40: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

Relevant links

Website European Commission (new 28 October 2009):

http://ec.europa.eu/enterprise/sectors/medicaldevices/index_en.htm

Website NBOG:

http://www.nbog.eu/

Page 41: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards

For more information on healthcare standards visit: http://shop.bsigroup.com/healthcarestandards

Page 42: EU Medical Devices Directive M5 Amendment 93 42 EEC Regulatory Update - BSI British Standards presentation

© Copyright The British Standards Institution 2010

For more information visit http://shop.bsigroup.com/healthcarestandards