Pre and Intraoperative 3D Image Resources for Planning and Treatment are Mandatory to Perform Complex EVAR Stéphan Haulon, J Sobocinski, B Maurel, T Martin Gonzalez, R Spear, A Hertault, R Azzaoui Aortic Centre, Vascular Surgery Lille, France

haulon site 2015

Embed Size (px)

Citation preview

FUSION

Pre and Intraoperative 3D Image Resources for Planning and Treatment are Mandatory to Perform Complex EVARStphan Haulon, J Sobocinski, B Maurel, T Martin Gonzalez, R Spear, A Hertault, R Azzaoui

Aortic Centre, Vascular SurgeryLille, France

DisclosuresResearch support, ConsultingCook Medical, GE Healthcare

Integrated EVAR workflow3 levels of simplificationPre-op In the ORPost-op 1-6 weeks before Procedure Following weekAssessmentFusionPlanning

Sizing and fusion preparation integrated into a planning softwareAssessmentFusionPlanning

BeforePre-op In the ORAfterSizingCTA overlay creation

Sizing & CTA overlay creation*Time saving in the ORPlanning overlay

Pre-op

Planning software RedosFragmented planning

Planning exportable to the C-armAssessmentFusionPlanning

3D overlayPlanned Landing zonesOstium outlineOptimized C-arm angulationsPlanning on CT AngioCT Angio fused in the C-arm

Recall of C-arm angulations from table-side

Overlay of Pre-Op CTA

Easy CT Angio registrationAssessmentFusionPlanning

2 Fluoro shots Table side alignment of bony structures from pre-op CT to intra-op fluoro

AssessmentFusionPlanning

Accurate fusionAssessmentFusionPlanning

Table side fine-tuning following stiff wire insertion

GE DISCOVERY IGS 730

GE DISCOVERY IGS 730

GE DISCOVERY IGS 730

Dose and contrast savingAssessmentFusionPlanningDose reduction by 40 to 59 % and contrast load by 24 to 26%

Non injected CT FusionAssessmentFusionPlanningTotal contrast throughout patients hospital stay was below 25 ml

Completion 3D angio + CE U/SAssessmentFusionPlanningIn the ORPost-op

CTA

USIn the ORPost-op

Completion DSAAssessmentBeforeAfterCompletion 3D AngioContrast savingImmediate revision Reduced costRepeated injectionsNo direct 3D asessment

Ease in 3D Angiography and good IQAssessmentFusionPlanning

24

25

Title or Job Number | XX Month 201X26

28

29

Significant reduction in overall dose and contrastAssessmentFusionPlanning-68%-50%

ConclusionEase of use and accuracy of sizing and fusion are crucial for routine useIntegrated workflow optimizes the use of pre-op dataLeveraging 3D capabilities from sizing to fusion to assessment with Cone beam CT help to reduce dose and contrast throughout patients hospital stay