LEADLESS PACEMAKER Dr Abhishek Rathore MD SJICR, Bangalore

Leadless pacemaker

Embed Size (px)

Citation preview

Page 1: Leadless pacemaker

LEADLESS PACEMAKER

Dr Abhishek Rathore MD SJICR, Bangalore

Page 2: Leadless pacemaker

• 1932 1st external cardiac pacemaker system by Alber S. Hyman

• 1958 1st complete epicardial pacing system by Rune Elmquist and A ̊ke Senning

• 1968 1st transvenous temporary pacing lead the same year by Seymour Furman

Sperzel J et al, Europace  2015 Oct;17(10):1508-13

Page 3: Leadless pacemaker

What is Leadless Pacemaker ?

• A leadless pacemaker is a small implantable device that sends electrical pulses to the heart whenever it senses that the heartbeat is too slow. 

Page 4: Leadless pacemaker

• Advantages:

• Less invasive • Less implantation time (25 to 40mins vs 60 mins)• Less recovery time • Eliminates complications related to the transvenous leads eg.

venous obstruction, insulation breaks, lead dislodgement, electrical malfunction, lead fractures, and infection.

• No visible lump or scar• shoulder mobility is maintained(6 out of 10 patients experience

reduced mobility in the shoulder)• No Pocket hematoma/infection ( Vs 0.5-1.5% incidence)• No Hemothorax/pneumothorax• Less radiation• better quality of life.

Page 5: Leadless pacemaker

History of Leadless Pacing

Vardas P E et al.EUR J CPE 1991

8 Dogs

Page 6: Leadless pacemaker

What's Needed for Leadless Pacemaker ?

• Catheter based delivery system• High density energy source• Low power electronics• Novel communication scheme• Biocompatible materials• Dependable fixation design• Retrievability capability

Page 7: Leadless pacemaker

Leadless Pacemakers

Page 8: Leadless pacemaker

Devices Availaible:• Nanostim leadless pacemaker (St.

Jude Medical, St. Paul, Minnesota)

• 2012

• < 10% of the size of a traditional pacemaker,

• Battery life - 8.4 years to 12.4 years

• A screw-in helix fixes it to RV

• Micra Transcatheter Pacing System (TPS) (Medtronic, Minneapolis, Minnesota)

• 2013

• < 30% smaller than the Nanostim

• 10 to 15 years

• Four tines (prongs) anchor to RV

Page 9: Leadless pacemaker

Device features  MicraTM  NanotismTM

Volume (cm3) 0.8 1.0

Length (mm) 26 42

Diameter (mm) 6.7 5.99

Weight (gr) 2 2

Introducer (Fr) 24 18

Fixation Self-expanding nitinol tines

Helical wire screw (+nylon tines)

Pacing mode VVI(R) VVI(R)

Rate Response Sensor Accelerometer Temperature

Telemetry Radiofrequency Conductive

Battery Lithium silver Vanadium oxide/Carbon Monoflouride

Lithium Carbon Flouride

Page 10: Leadless pacemaker

Micra vs Nanotism

Page 11: Leadless pacemaker

Periprocedural data    MicraTM    NanotismTM

Number of patients   725   526

Successful implantation (%)   99.2   95.8

Final device position (%)

ApexSeptumMid-septumOutflow tract                 

65.23.87.80.6

ApexApical septum Outflow/septum/other   Missing data                        

38.119.042.70.2

Procedure duration (min)   34.8±24.1   46.5±25.3

Fluoroscopy duration (min)   8.9±16.6   13.9±9.1

Major complications (%)   4.0   6.5

Device dislodgement (%)   0   1.1

Page 12: Leadless pacemaker

Implantation System

Page 13: Leadless pacemaker

•  

Implantation steps. (A) A venogram (optional); (B) LCP positioned into the RV by deflecting the catheter and placed 0.5–1 cm from the RV apex; (∼ C and D) Protective cover is pulled back to expose the flexible part of the catheter; (E) The pacemaker is undocked from the delivery catheter while a tethered connection is maintained. In case the position is suboptimal, the LCP can be reengaged, unscrewed, and repositioned. (F) The LCP is released by rotating the release knob of the catheter.

Page 14: Leadless pacemaker
Page 15: Leadless pacemaker

Contraindications for Leadless pacemaker

• Pacemaker, ICD, CRT, neurostimulator or any other implanted device which uses electrical current

• Unstable angina pectoris, recent myocardial infarction• Mechanical tricuspid valve• Implanted vena cava filter• Left ventricular assist device• Morbidly obese where telemetric programmer communication can be

limited• Limited or missing femoral venous access or stenosis• Unable to tolerate emergency sternotomy• Intolerance to titanium or nickel-titanium (nitinol) alloy

Page 16: Leadless pacemaker
Page 17: Leadless pacemaker

Leadless Trial Permanent Leadless Cardiac Pacing Results of the LEADLESS Trial

• Background—Conventional cardiac pacemakers are associated with several potential short- and long-term complications related to either the transvenous lead or subcutaneous pulse generator. We tested the safety and clinical performance of a novel completely self-contained leadless cardiac pacemaker (LCP) ie,Nanotism.

• Methods and Results—The primary safety endpoint was freedom from complications at 90 days. Secondary performance endpoints included implant success rate, implant time and measures of device performance (pacing/sensing thresholds and rate-responsive performance). The mean age of the patient cohort (n=33) was 77 ± 8 years and 67% of the patients were male (n= 22/33). The most common indication for cardiac pacing was permanent AF with atrioventricular block (n = 22, 67%). The implant success rate was 97% (n= 32). Five patients (15%) required the use of >1 LCP during the procedure. One patient developed RV perforation and cardiac tamponade during the implant procedure, and eventually died as a result of stroke. The overall complication-free rate was 94% (31/33). After 3 months of follow-up, the measures of pacing performance (sensing, impedance and pacing threshold) either improved or were stably within the accepted range.

• Conclusions—In a prospective non-randomized study, a completely self-contained single-chamber leadless cardiac pacemaker has shown to be safe and feasible. The absence of a transvenous lead and subcutaneous pulse generator could represent a paradigm shift in cardiac pacing. Reddy VY et al. Circulation. 2014;129:1466-71

Page 18: Leadless pacemaker
Page 19: Leadless pacemaker
Page 20: Leadless pacemaker

From: Chronic Performance of a Leadless Cardiac Pacemaker: 1-Year Follow-Up of the LEADLESS Trial

J Am Coll Cardiol. 2015;65(15):1497-1504. doi:10.1016/j.jacc.2015.02.022

(Left) Kaplan-Meier survival curve represents freedom from device-related complications: 2 patients had device-related complications (6%, 31 of 33), both periprocedurally. During further follow-up, no complications occurred in the remaining patients (n = 31). (Right) Device performance measurements of the leadless cardiac pacemaker. The mean value ± SD of pacing threshold (at 0.4 ms [V]) (right top); the R-wave amplitude (mV) (right middle); and the pacing impedance (Ω) (right bottom) at each follow-up assessment.

Page 21: Leadless pacemaker

Leadless II Trial• Prospective, multicenter, non-randomized, FDA IDE study• Objective: – To evaluate the clinical safety and efficacy of non-

surgical implantation of the leadless cardiac pacemaker in patients indicated for a VVI(R) pacemaker.

• Primary Cohort: 1st 300 patients followed for 6mo (June 2015) • Total Cohort: All patients enrolled by June 2015 (n=526) • Primary Endpoints (by ITT): • – Safety: Freedom from Serious Adverse Device Effects at 6 months

– Efficacy: Acceptable pacing capture threshold (d2.0 V at 0.4 msec) and a therapeutically acceptable sensing amplitude (R wave e5.0 mV, or a value equal to or greater than the value at implantation) through 6 mo.

• 56 Centers in the US, Canada and Australia• – 100 Operators (only one had prior experience with leadless

pacing)

Page 22: Leadless pacemaker

Outcomes of Leadless II Trial

• Device was successfully implanted in ~96% of patients

• Primary Safety Endpoint (Intent-to-Treat Analysis) • – 280 of the 300 patients achieved endpoint (93.3%; 95% CI = 89.9 to 95.9) • – This exceeded the performance goal of 86% (P<0.001)• Primary Efficacy Endpoint (Intent-to-Treat Analysis) – 270 of the 300 patients achieved endpoint (90.0%; 95% CI = 86.0 to 93.2) – This exceeded the performance goal of 85% (P = 0.007) • • Efficacy (Successful implants) – 289 patients with successful device implant• – 270 of the 289 patients achieved endpoint (93.4%; 95% CI = 89.9 to 96.0) • – This exceeded the performance goal of 85% (P <0.001)

Page 23: Leadless pacemaker

Device related complications

Page 24: Leadless pacemaker

Conclusion

• The Leadless Pacemaker was successfully implanted in ~96% of attempted patients.

• The trial met the pre-specified Safety and Efficacy endpoints Complication rate similar to that seen with conventional pacemakers

• Complication rate likely to improve with operator experience • The device was shown to be retrievable in a subgroup of

patients (n=7) who needed a replacement (Time from implant = 160±180 days; Range = 1 to 413 days)

• The estimated device longevity based on the 6-month follow up duration is encouraging

Page 25: Leadless pacemaker

A Leadless Intracardiac Transcatheter Pacing System

BACKGROUND A leadless intracardiac transcatheter pacing system has been designed to avoid the need for a pacemaker pocket and transvenous lead. METHODSA prospective multicenter study . The analysis of the primary end points began when 300 patients reached 6 months of follow- up. The primary safety end point was freedom from system-related or procedure related major complications. The primary efficacy end point was the percentage of patients with low and stable pacing capture thresholds at 6 months (≤2.0 V at a pulse width of 0.24 msec and an increase of ≤1.5 V from the time of implantation). The safety and efficacy end points were evaluated against performance goals (based on historical data) of 83% and 80%, respectively. We also performed a post hoc analysis in which the rates of major complications were compared with those in a control cohort of 2667 patients with transvenous pacemakers from six previously published studies.RESULTSThe device was successfully implanted in 719 of 725 patients (99.2%). The Kaplan– Meier estimate of the rate of the primary safety end point was 96.0% (95% confidence interval [CI], 93.9 to 97.3; P<0.001 for the comparison with the safety performance goal of 83%); there were 28 major complications in 25 of 725 patients, and no dislodgements. The rate of the primary efficacy end point was 98.3% (95% CI, 96.1 to 99.5; P<0.001 for the comparison with the efficacy performance goal of 80%) among 292 of 297 patients with paired 6-month data. Although there were 28 major complications in 25 patients, patients with transcatheter pacemakers had significantly fewer major complications than did the control patients (hazard ratio, 0.49; 95% CI, 0.33 to 0.75; P = 0.001).CONCLUSIONSIn this historical comparison study, the transcatheter pacemaker met the prespecified safety and efficacy goals; it had a safety profile similar to that of a transvenous system while providing low and stable pacing thresholds

Reynolds D et al. NEJM. Feb 2016

Page 26: Leadless pacemaker
Page 27: Leadless pacemaker

Electrical Performance, Characteristics of the Transcatheter Pacemaker,according toStudy Visit.

Page 28: Leadless pacemaker

WiCSTM clinical data: Wise-CRT study

• Non-randomized, safety/performance (1month) + efficacy (6 months) (N=17).

Auricchio, A et al (2014) Feasibility,Safety and short-term outcome of leadless ultrasound based endocardial LV synchronisation in heart failure patients: Result of wireless stimulation endocardially for CRT (WiSE-CRT study) Europace 16(5):681-688

Page 29: Leadless pacemaker

Device Characteristics:• Woven polyester outer jacket(enhance endothelialisation)• Aortic retrograde approach• Tissue viability should be confirmed( intracardiac ECG

and external electrophysiological stimulator)

Page 30: Leadless pacemaker

Out of 17 patients- • 7 pts had failed CS lead implantation for CRT• 2 pts were non-responders to CRT• 8 pts had previous Pacemaker or ICD in place but had CRT

indications

• Only able to implant in 13 patients.(3 pts had Pericardial effusion and in1 patient no sufficient pacing threshold achieved.)

• No thromboembolism noted.

• Study was terminated.• BiV pacing documented in 83% and clinical efficacy in 92%.

Page 31: Leadless pacemaker

Conclusion

• A promising alternative to conventional single chamber pacemaker systems.

• Safety and long-term outcome of leadless pacemaker need to be evaluated in more detail.

Page 32: Leadless pacemaker

Thank You