45
Site # 8 Zug 2016 Single Guided Implant Placement Corrections 1 mm offset toward buccal 10° correction buccal 6° correction mesial Featuring 2-Piece Offset Guide Posts Invivo5 Software Bending Tool 4.5 MM Serial Guide Sleeves Guide Sleeve Insert Handles Versah Condensing Drills to enlarge the osteotomy Astra Implant

Single guided implant surgery #8 9.2016 z

Embed Size (px)

Citation preview

Page 1: Single guided implant surgery #8 9.2016 z

Site # 8

Zug 2016

Single Guided Implant Placement

Corrections

• 1 mm offset toward buccal

• 10° correction buccal

• 6° correction mesial

Featuring

2-Piece Offset Guide Posts Invivo5 Software

Bending Tool4.5 MM Serial Guide SleevesGuide Sleeve Insert Handles

Versah Condensing Drills to enlarge the osteotomyAstra Implant

Page 2: Single guided implant surgery #8 9.2016 z

Start With Precision. Place With Confidence.™

1.800.314.0065 • www.deplaque.com

DéPlaque

A Geometric Approach to Guided Surgery

Guide Right™ Surgical Guide Systemfabricate ▪ evaluate ▪ correct ▪ verify ▪ re-fabricate ▪ place

Page 3: Single guided implant surgery #8 9.2016 z

Guide Right™ Surgical Guide System

DéPlaque

▪ fabricate DIAGNOSTIC guide ▪ evaluate ▪ correct ▪ re-fabricate surgical guide▪ verify▪ place

Page 4: Single guided implant surgery #8 9.2016 z

Planned implant position made at the site # 8 with a 3/32” drill

Page 5: Single guided implant surgery #8 9.2016 z

3/32” hole in cast

Page 6: Single guided implant surgery #8 9.2016 z

3 mm straight guide post in hole

Page 7: Single guided implant surgery #8 9.2016 z

Articulated models with 3 mm short guide post

Page 8: Single guided implant surgery #8 9.2016 z

3 mm guide post

Page 9: Single guided implant surgery #8 9.2016 z

3 mm guide sleeve positioned on the 3 mm guide post with cleat to the palatal surface.

Page 10: Single guided implant surgery #8 9.2016 z

Triad® gel is added to the guide sleeve & adjacent teeth on either side of the site to form the DIAGNOSTIC guide

Page 11: Single guided implant surgery #8 9.2016 z

DéPlaque

▪ fabricate diagnostic guide ▪ evaluate ▪ correct ▪ re-fabricate surgical guide▪ verify▪ place

Guide Right™ Surgical Guide System

Page 12: Single guided implant surgery #8 9.2016 z

Invivo5 Anatomage

Cone Beam Analysis

Page 13: Single guided implant surgery #8 9.2016 z

HOME POSITION

Invivo5 Anatomage

Page 14: Single guided implant surgery #8 9.2016 z

Using Invivo5 Software

CHANGES in the proposed implant position can be made

Changing the LINEAR POSITION… … in either mesio-distal plane or bucco-lingual plane is made by selecting an offset guide post to alter the position of the guide sleeve

Changing the ANGLE of the TRAJECTORY by measuring the angle of the proposed implant location & bending the guide post to the corrected angle.

After making a correction (angle or linear) in one plane YOU MUST RETURN TO THE HOME POSITION To measure and make the correction needed in the other plane.

.

Page 15: Single guided implant surgery #8 9.2016 z

Use of Invivo5 Software

The HOME POSITION WARNING:

Measuring or making angle or linear corrections from any other positions (cross sectional or tangential)

will lead to errors, & the corrections recorded will not be accurate

BECAUSE the angle is not measured from the HOME POSITION

Page 16: Single guided implant surgery #8 9.2016 z

# 8 HOME POSITION

axial cross sectional

tangential volumetric

Page 17: Single guided implant surgery #8 9.2016 z

axial cross sectional

# 8

tangential volumetric

LINEAR CORRECITON: 1 mm offset buccalANGULAR CORRECTION: 10° buccal

Page 18: Single guided implant surgery #8 9.2016 z

Measure the angle formed by the 2 intersecting lines

use the angle tool in invivo5 to measure & correct the angle

of the guide post/sleeve

When … the long axis of the sleeves & the virtual implants are not parallel.

Page 19: Single guided implant surgery #8 9.2016 z

# 8 LINEAR CORRECITON: 1 mm offset buccalANGULAR CORRECTION: 10° buccal

Page 20: Single guided implant surgery #8 9.2016 z

# 8 LINEAR CORRECITON: noneANGULAR CORRECTION: 10° buccal & 6° mesial

Page 21: Single guided implant surgery #8 9.2016 z

DéPlaque

▪ fabricate [diagnostic guide] ▪ evaluate ▪ correct ▪ re-fabricate [surgical guide]▪ verify▪ place

Guide Right™ Surgical Guide System

Page 22: Single guided implant surgery #8 9.2016 z

LINEAR CORRECTIONS are made with the

1 mm offset guide post

Guide Right™ Offset Guide Posts

LINEAR CORRECTION

The bottom half of the offset guide posts are designed with 8 flat sides so they can be rotated 45° & corrected half way between the buccal & mesial surface, allowing corrections in 8 directions: mesial, mesio-buccal, buccal, mesio-lingual, distal, disto-lingual, lingual, mesio-lingual

Page 23: Single guided implant surgery #8 9.2016 z

DéPlaque

Guide Right™ Surgical Guide System

Offset Guide Posts

Page 24: Single guided implant surgery #8 9.2016 z

2-Piece Guide Post Straight Lower Piece with Upper Removable Piece

The bottom half of the straight Lower Pieces of the 2-Piece Guide Post are designed with 4 flat sides corresponding with the mesial, distal, buccal & lingual surfaces of the tooth.

Page 25: Single guided implant surgery #8 9.2016 z

Guide Right™ Bending Tool

1 mm offset Guide Post Bent 1st time to 10°

Stylus

1st BEND

10°

1 mm Offset Guide Post

Page 26: Single guided implant surgery #8 9.2016 z

Guide Right™ Bending Tool

1 mm Offset Guide Post Bending Sequence

1 mm Offset Guide Post Stylus positioned on 1 mm Offset Guide Post

Stylus positioned on 1 mm Offset Guide Post& bent 10°

1 mm Offset Guide Post Shown with 10° bend

Steps to Bending Tool use at end of slideshow

Page 27: Single guided implant surgery #8 9.2016 z

Guide Right™ Bending Tool with Support Bar

1 mm offset Guide Post with a 10° bend rotated back to 0°

• 1 mm offset guide post with a 10° bend rotated 90°

in block hole

• stylus support bar moved into place to support the 10° angle of the 1st bend

Before 2nd BEND

• stylus removed & guide post rotated 90° pointing to 0° before 2nd bend is made

Page 28: Single guided implant surgery #8 9.2016 z

Guide Right™ Bending Tool with Support Bar

1 mm offset Guide Post with 10° bend bent a 2nd time to 6°

stylus

• 1 mm offset guide post with 10° angle rotated 90° in block hole & stylus replaced

2nd BEND

• stylus removed & the guide post rotated 90° before 2nd bend is made

• stylus support bar moved into place to support the angle of the 1st bend

Page 29: Single guided implant surgery #8 9.2016 z

DéPlaque

▪ fabricate [diagnostic guide] ▪ evaluate ▪ correct ▪ re-fabricate [surgical guide]▪ verify▪ place

Guide Right™ Surgical Guide System

Page 30: Single guided implant surgery #8 9.2016 z

Angle corrected lower-half of 2-Piece 1 mm offset guide post placed in the reference hole in the cast

Page 31: Single guided implant surgery #8 9.2016 z

2-Piece 1 mm offset guide post with 3 mm upper piece

facial view

Page 32: Single guided implant surgery #8 9.2016 z

1 mm offset guide post with fabrication insert used to center the 4.5 mm guide sleeve

4.5 mm fabrication insert 4.5 mm guide sleevewith cleat to palatal

1 mm offset guide post

Page 33: Single guided implant surgery #8 9.2016 z

4.5 Fabrication Insert (FABIN)

Purposes & Use:

• To center the 4.5 mm guide sleeve on the 3 mm guide post

• To prevent acrylic gel from entering inside sleeve during guide fabrication by placing the FABIN inside the

4.5 mm diameter guide sleeve with 3 mm guide post.

• Place the FABIN over the guide post, then apply gel.

• 5.3 mm FABIN available for the 5.3 guide sleeve

Page 34: Single guided implant surgery #8 9.2016 z

Triad® gel is added to the lubricated cast to form the corrected SURGICAL GUIDE

Page 35: Single guided implant surgery #8 9.2016 z

The fabrication insert has been removed leaving the 4.5 mm guide sleeve centered on the lower-half of the 2-Piece guide post.

Page 36: Single guided implant surgery #8 9.2016 z

Facial View • Corrected SURGICAL GUIDE

Page 37: Single guided implant surgery #8 9.2016 z

SURGICAL GUIDE is trimmed to be sure no acrylic is occluding the surface of the guide sleeve.

Page 38: Single guided implant surgery #8 9.2016 z

In this case, the anterior insert handle with inserts were used to position the increasing diameter drills in the surgical guide sleeve without changing the SURGICAL GUIDE for the placement of a 4 mm X 11 mm implant.

Anterior Insert Handles with 2.2 & 3.0 mm inserts used with the 4.5 mm guide sleeves

Page 39: Single guided implant surgery #8 9.2016 z

Versah Densah Bur Kit used to prepare & widen the osteotomy

Page 40: Single guided implant surgery #8 9.2016 z

Guide Right™ Surgical Guide System

DéPlaque

▪ fabricate [diagnostic guide] ▪ evaluate ▪ correct ▪ re-fabricate [surgical guide]▪ verify▪ place

Guide Right™ Surgical Guide System

Page 41: Single guided implant surgery #8 9.2016 z

4.0 mm Astra Implant Placement

Page 42: Single guided implant surgery #8 9.2016 z

Guide Right™

GUIDE POST BENDING TOOL

Steps to SINGLE BEND

Step 1 Place bending tool plate on a secure flat surface with the degree increments at the top & the stainless steel bar at the bottom.

Step 2 Locate the two 3/32” holes in the top of the stainless steel bar. The hole closest to the plate is used for all guide posts up to 3.0 mm offsets. The front hole is used for offsets 3.5 mm or larger. Place the bottom half of the guide post to be bent into the appropriate hole. Position the offset guide post in the hole determined by the planned direction of the correction. Tighten the set screw.

Step 3 Select the appropriate stylus to fit over the top half of the guide post.

Step 4 Fit the stylus over the guide post securely with the point directed at 0 ° & the bottom of the stylus in contact with the platform of 2-piece guide post or if no platform on post, in contact with block.

Step 5 Using the stylus as a lever, bend the guide post to the degree of angle of correction. You may need to ease the point of the stylus beyond the point of the desired degree.

Step 6 Remove the stylus. Loosen screw to remove the guide post bent to the desired angle.

Page 43: Single guided implant surgery #8 9.2016 z

Steps to COMPOUND BEND

Guide Right™

GUIDE POST BENDING TOOL

Step 1 Position a straight or offset guide post in the bending plate, tightening the set screw against one of the 8 flat surfaces on the lower half of the guide post.

Step 2 The 1st bend can be made to the right or left direction.

Step 3 The set screw is loosened & the guide post is rotated 90° next flat surface.

Step 4 The 2nd bend in the 2nd plane is made after rotating the guide post up away from the surface of the bending plate to register the stylus point back at 0º.

Step 5 Slide the stylus support bar down under the stylus until it supports the stylus. Tighten the side screws before making the 2nd bend.

Step 6 The 2nd bend can be made in either direction according to the X-ray.

Step 7 Remove the stylus and place the guide post back in the cast with the appropriate side indicated by a mark with a felt tip pen facing the buccal or lingual surface. Be sure the post is in the correct position.

Linear Corrections, using an offset guide post, must be determined & made before Angle Corrections.Offset 2-Piece Guide Posts available in the 3 mm guide post: 0.5,1, 1.5, 2.0, 2.5 & 3.0 mm.

Page 44: Single guided implant surgery #8 9.2016 z

view

Guide Right™

VIDEO Case KZ for actual site preparation & implant placement

www.deplaque.com

Featuring

4.5 MM Serial Guide SleevesGuide Sleeve Insert Handles

Versah Condensing Drillsused to enlarge the osteotomy

Astra Implant

Page 45: Single guided implant surgery #8 9.2016 z

Guide Right™ Surgical Guide System

Start With Precision. Place With Confidence.™

1.800.314.0065 • www.deplaque.com

fabricate ▪ evaluate ▪ correct ▪ verify ▪ re-fabricate ▪ place

DéPlaque

A Geometric Approach to Guided Surgery