10
5 th International Conference on NDT of HSNT- IC MINDT 2013 Athens Greece, Eugenides Foundation, May 20-22, 2013 Ultrasonics System for Arbitrary Pulse Width and Position Trains Application for Imaging L. Svilainis, V. Dumbrava, S. Kitov, A. Aleksandrovas, A. Chaziachmetovas, V.Eidukynas. D. Kybartas, K. Lukoseviciute Kaunas University of Technology, Department of Signal Processing, Studentu Str. 50-340, LT-51368, Kaunas, Lithuania, [email protected]. Abstract Performance improvement of ultrasound application in measurement or imaging demands high energy and wideband signals for excitation. Spread spectrum signals are used to increase the signal energy and maintain the resolution. We suggest using novel spread spectrum signals generation technique: trains of arbitrary pulse width and position (APWP) pulses. It is expected that APWP should have properties of chirp (wideband and compressible) and of a single pulse (low correlation sidelobes). In order to complete such study, ultrasonics signals acquisition system has been developed, capable of APWP pulse trains generation and reception. System structure and parameters investigation are presented. Acquisition system was developed to satisfy the specifics of the APWP: ability to generate constant amplitude APWP pulse trains, acceptable pulser performance, and pulser recovery time. Investigation of system performance indicates that it can be used for 0.5MHz to 30MHz frequency range: both pulser and preamplifier have balanced operation range. Excitation signal amplitude can reach 200V for unipolar signals and 400V p-p for bipolar signals. Variation of pulser output impedance measured according to EN 12668-1 standard is within 7 according to EN 12668- Keywords: ultrasonic imaging, iterative deconvolution, composite imaging, ultrasonic spectroscopy. Introduction Ultrasound application in measurement, imaging or navigation systems is mainly based on time of flight estimate. Propagation time estimate accuracy depend on signal energy, input noise and signal spectrum. Resolution of the imaging system depends on the envelope bandwidth [1]. Spread spectrum signals are used to increase the signal energy and maintain the resolution [2]. Such signals can turn useful when applied for structural noise reduction in composite materials imaging [3]. We suggest using novel spread spectrum signals generation technique [4,5]: trains of arbitrary pulse width and position (APWP) pulses. It is expected that APWP should have properties of chirp (wideband and compressible) and of a single pulse (low correlation sidelobes). In order to complete such study, ultrasonics signals acquisition system has been developed, capable of APWP pulse trains generation and reception. System structure and parameters investigation are presented. ToF Estimation

Ultrasonics system for arbitrary pulse width and position trains application for imaging

  • Upload
    ktu-lt

  • View
    0

  • Download
    0

Embed Size (px)

Citation preview

5th

International Conference on NDT of HSNT- IC MINDT 2013

Athens Greece, Eugenides Foundation, May 20-22, 2013

Ultrasonics System for Arbitrary Pulse Width and Position Trains

Application for Imaging

L. Svilainis, V. Dumbrava, S. Kitov, A. Aleksandrovas, A. Chaziachmetovas,

V.Eidukynas. D. Kybartas, K. Lukoseviciute

Kaunas University of Technology, Department of Signal Processing,

Studentu Str. 50-340, LT-51368, Kaunas, Lithuania,

[email protected].

Abstract

Performance improvement of ultrasound application in measurement or imaging demands high

energy and wideband signals for excitation. Spread spectrum signals are used to increase the

signal energy and maintain the resolution. We suggest using novel spread spectrum signals

generation technique: trains of arbitrary pulse width and position (APWP) pulses. It is expected

that APWP should have properties of chirp (wideband and compressible) and of a single pulse

(low correlation sidelobes). In order to complete such study, ultrasonics signals acquisition

system has been developed, capable of APWP pulse trains generation and reception. System

structure and parameters investigation are presented. Acquisition system was developed to

satisfy the specifics of the APWP: ability to generate constant amplitude APWP pulse trains,

acceptable pulser performance, and pulser recovery time. Investigation of system performance

indicates that it can be used for 0.5MHz to 30MHz frequency range: both pulser and

preamplifier have balanced operation range. Excitation signal amplitude can reach 200V for

unipolar signals and 400V p-p for bipolar signals. Variation of pulser output impedance

measured according to EN 12668-1 standard is within 7

according to EN 12668-

Keywords: ultrasonic imaging, iterative deconvolution, composite imaging, ultrasonic

spectroscopy.

Introduction

Ultrasound application in measurement, imaging or navigation systems is mainly based on time

of flight estimate. Propagation time estimate accuracy depend on signal energy, input noise and

signal spectrum. Resolution of the imaging system depends on the envelope bandwidth [1].

Spread spectrum signals are used to increase the signal energy and maintain the resolution [2].

Such signals can turn useful when applied for structural noise reduction in composite materials

imaging [3]. We suggest using novel spread spectrum signals generation technique [4,5]: trains

of arbitrary pulse width and position (APWP) pulses. It is expected that APWP should have

properties of chirp (wideband and compressible) and of a single pulse (low correlation

sidelobes). In order to complete such study, ultrasonics signals acquisition system has been

developed, capable of APWP pulse trains generation and reception. System structure and

parameters investigation are presented.

ToF Estimation

5th

International Conference on NDT of HSNT- IC MINDT 2013

Athens Greece, Eugenides Foundation, May 20-22, 2013

The time of flight (ToF) of the ultrasonics signal is an essential parameter in temperature

measurement, chemistry, navigation or velocity measurement systems [6-9]. It’s estimate is

influenced by the quality of the signal. If estimation is done by locating the correlation

function’s peak [1] in the presence of noise with power spectral density N0, the of ToF variance

is defined by Cramer - Rao lower error bound:

0

22

1

N

EFe

ToF

, (1)

where E is the received signal energy, delivered to impedance Z:

0

*2 21dffSfS

Zdtts

ZE T , (2)

Fe is the effective bandwidth of the signal, constituted by center frequency f0 and envelope

bandwidth :

2

2

222

0

22

0

2

E

dffSf

E

dffSff

fFe

. (3)

The amplifier input noise power density can be simplified as [1]

Z

ZieeN nnVnT

222

0

, (4)

where enT is the thermal noise of the transducer impedance Re(ZT) at temperature T, 4kRe(ZT)T,

enV is the amplifier’s input voltage noise, in is the amplifier’s current noise. Here, SNR is treated

as the signal energy and the noise power spectral density ratio:

0

2

N

ESNR . (5)

The signal energy can be improved by increasing the excitation amplitude (usually limited by

the transducer design or electronics), or by increasing the signal duration.

APWP Signals

From equations (1)-(3) it can be seen that accuracy can be improved by increasing the signal

bandwidth or increasing the energy. Energy enhancement is possible by a probing signal

amplitude or duration increase. The excitation amplitude is limited by the transducer

construction and the electronics capabilities. Another approach would be to increase the duration

of the exciting signal. But if it is a simple CW burst then the envelope of the correlation function

is not sharp and resolution will be low. Spread spectrum (SS) signals allow to achieve

compression using matched filtering [2,6,10] thanks to wide bandwidth and long duration.

Wideband signals can also turn useful if applied for structural noise dispersion increase in

composites imaging [3]. Arbitrary waveform signals are not easy to generate if large voltages

have to be delivered to capacitive load [10,11]. If linear frequency modulation signal (Fig.1) is

used, such signal is significantly compressed during the matched filtering.

5th

International Conference on NDT of HSNT- IC MINDT 2013

Athens Greece, Eugenides Foundation, May 20-22, 2013

Fig.1. APWP pulse trains comparison against conventional excitation signals

But the correlation sidelobes of the chirp are higher than for simple rectangular pulse [10].

Nonlinear frequency modulation (NFM) signals are gaining popularity thanks to the ability to

control the shape of the spectrum and so the sidelobe level [11]. In turn rectangular pulses are

easy to generate and exhibit low correlation sidelobes, but do not possess the high energy and

orthogonality set of SS. Novel spectrum spread technique did not receive the proper attention in

ultrasound: trains of the arbitrary position and width pulses (APWP) [4, 5]. Technique is using a

chaotically placed train of square pulses with arbitrary position and width (Fig.1). We suggest

using this technique novel spread spectrum signals generation technique [10]: trains of arbitrary

pulse width and position (APWP) pulses.

Initial study carried out in [10] suggests that APWP should have wide, controllable bandwidth,

similar to chirp and same time should have slightly lower correlation sidelobes compared to

rectangular pulse (Fig.2).

Fig.2. Comparison of chirp, rectangular pulse and APWP signals’ correlation function

Analysis of correlation functions of the signals presented in Fig.2 indicates that sidelobes-

optimized APWP signal maintains the mainlobe width (similar to pulse). Meanwhile, APWP

signal energy is the same as chirp signal. But chirp has much higher correlation sidelobes if

rectangular slope is used.

Since APWP signal is constructed by the set of constant amplitude but arbitrary duration pulses

with arbitrary spacing, generation of such signals should be almost as simple as rectangular

pulses. But, new, better than pulse correlation properties can be obtained. We are aiming to

investigate such signals in ultrasonic imaging and ToF measurement applications. In order to

have the ability to generate such signals and carry out the research of APWP application,

dedicated acquisition system was developed.

System Structure

Ultrasonics signals acquisition system developed is capable of APWP signals generation and 3D

spatially correlated data acquisition. System structure is presented in Fig.3.

Pulse

CW burst

Chirp

APWP

-400.0n -200.0n 0.0 200.0n 400.0n

-1.0

-0.8

-0.6

-0.4

-0.2

0.0

0.2

0.4

0.6

0.8

1.0

Chirp

PulseAPWP

Ma

tch

ed

filt

er

ou

tpu

t, A

.U.

Time, s

5th

International Conference on NDT of HSNT- IC MINDT 2013

Athens Greece, Eugenides Foundation, May 20-22, 2013

Fig.3. Acquisition system’s structure and the 3D transducer positioning equipment

System is composed of the main acquisition module, analog front-end and the transducer

positioning equipment. Main module (Fig.4) is responsible for low power excitation signals

generation and analog signals digitization and transfer to PC using USB connection.

Fig.4. Main acquisition module structure and PCB design view

It is located at host PC end. It is using two 10 bit resolution analog-to-digit converters (ADC)

with programmable conversion frequency up to 100 MHz. Resolution and sampling rate of ADC

was chosen basing the analog electronics noise level and is balanced against expected ToF

estimation systematic errors [1].

Analog front-end (Fig.5) is located at scanner end and is used for low power excitation signal

conversion into high voltage excitation signal and received signals preamplification.

Fig.5. Analog front-end structure

Excitation signals amplitude is controlled by varying the high voltage DC/DC converter output.

It can be varied up to 200 V. High voltage is fed to pulser which basically chops the high

voltage and feeds it to RF transformer output, forming switched push-pull pulser [12].

Excitation signal’s rise/fall times of 10 ns ensure up to 30 MHz system bandwidth. Pulse train is

formed in main acquisition block at 100 MHz rate. Toggling step rate of APWP pulse train is

such way is 10 ns. Reception channel’s gain is controlled at analog front-end side (Fig.5) and at

the main acquisition block (Fig.4). It can be varied in 80 dB range.

Water tank

X

Scanner

controller

Main

acquisition

moduleHost PC

Analog

front-end

Transducer

Z

3D scanner

Y

ADCUSB

communication

and

controlADC

RS485

conn

conn

To front-end

USB,

to PC

APWP RAM Drivers

Pulser

Local

uControllerTo main

acquisition

module

conn

HV DCDC

RS485

5th

International Conference on NDT of HSNT- IC MINDT 2013

Athens Greece, Eugenides Foundation, May 20-22, 2013

Scanner (Fig. 3) is controlled by dedicated controller with USB or RS485 connectivity. Every

axis (X,Y,Z) has it’s own step motor driver. Minimum scanning step along X and Y axes is

10 Z axis.

Pulser Performance Measurement

Since essential parameters of APWP generation are defined by quality of the high voltage pulser

output, investigation of pulser output impedance and attainable bandwidth was carried out.

Standard EN_12668-1 defines the ultrasonic pulser output impedance Zo measurement

procedure by using two output voltages obtained at 50 V50) and 75 V75) pulser load:

07550

5075

5075

7550

NVV

VVZo

. (6)

We have slightly modified this procedure in order to investigate the output impedance at

particular frequency. Since system is capable of automated data collection using several

excitation signals, set of CW burst signals was formed and stored for automated generation and

acquisition. Voltages V50 and V75 were recorded using 1:100 voltage divider formed by 5

50 The amplitude and phase of the harmonic signal can be extracted by using the

Sine wave correlation (SWC) technique. SWC can be treated as correlation coefficient of the

signal’s sampled version sn with the sine and cosine signals:

1

1

0

1

1

0

2sin2

2cos2

WN

wsft

fS

WN

wsft

fC

N

n

nnn

N

n

nnn

, (7)

where, N is length of the observation window wn, W1 is a L1 norm of the window. It was

decided to use the excitation generator using the direct digital synthesizer (DDS). The CW

bursts used were 10 periods long and their frequency was in fixed proportion to system clock

frequency. Since data acquisition and excitation used common reference oscillator, frequency-

related amplitude estimation errors were minimized. Results for output impedance magnitude

variation with output amplitude are presented in Fig.6.

Fig.6. Pulser output impedance vs. frequency

500k 1M 10M 30M

3

6

9

12

15

18

21

24

27

30

Zo

ut,

Frequency, Hz

@ 50V

@ 100V

@ 150V

@ 200V

5th

International Conference on NDT of HSNT- IC MINDT 2013

Athens Greece, Eugenides Foundation, May 20-22, 2013

It can be seen that output impedance magnitude is stable with output voltage and is rising with

frequency. Variation is within 7 Obtained output impedance now can be used

for pulser performance together with ultrasonic transducer estimation. Measured transducer

impedance Zin can be used to calculate the complex power delivery:

*

2

inoino

ing

TZZZZ

ZeS

, (8)

where eg is the intrinsic generator voltage. Assuming that losses in matching circuit are

negligible then real part PT of the complex power delivered ST is equal to the power transmitted

by transducer (in case of the optimal transducer design). Then the power delivery to load

efficiency criteria can be established, which indicates the ratio of the real power conveyed to

load and the power available from the generator:

%100

4%100

Re4*2

inging

ing

g

Tg

ZRZR

ZR

e

SR

, (9)

Same equation (7) was used to obtain the pulser output voltage. Pulser main harmonic output

voltage plot against load type and frequency is presented in Fig.7.

Fig.7. Pulser output amplitude vs. frequency and load type

Pulser is using active switches for both positive and negative front’s generation. Such topology

should allow for quite efficient energy use: energy is consumed only during the pulse fronts.

Energy consumption during positive front and negative front is still present, since parasitic

capacitance of output MOSFET’s has to be charged. If capacitive load is used (piezoelectric

transducer will exhibit such load) additional losses occur. Efficiency of pulser was evaluated by

comparing the energy consumption per pulse against theoretically expected energy consumption

if only capacitive load was the cause of losses. Energy stored in capacitor C0 which is charged to

voltage V is

2

0

2 CVE HV , (8)

Then, power consumed by pulser generating burst of N pulses over 1 s at pulse repetition

frequency PRF is:

NPRFCVP HVMM 0

2

21 , (9)

500k 1M 10M 40M

0

50

100

150

200

250

Vo

ut, V

Frequency, Hz

50 load:

@ 50V

@ 100V

@ 150V

@ 200V

1000pF load:

@ 50V

@ 100V

@ 150V

@ 200V

5th

International Conference on NDT of HSNT- IC MINDT 2013

Athens Greece, Eugenides Foundation, May 20-22, 2013

Power consumption on HV power supply was monitored and equation (9) was used to obtain the

power per pulse consumption. Results for 200 V output with unloaded pulser (Fig.8 left) and

50 .

0.9 1 2 3 4 5 6 7 8 9 10 2020

10

20

30

40

50

60708090

100

200

300300 Theory SPD02N60C3

IRFRC20 SPD02N60S5

IRFRC20 without cancel winding

Ene

rgy p

er

pu

lse

(

J)

Frequency (MHz)

1 2 3 4 5 6 7 8 9 10 20 3030

2020

30

40

50

60708090

100

200

300

400

500

600700700 Theory SPD02N60C3

IRFRC20 SPD02N60S5

Ene

rgy p

er

pu

lse

(

J)

Frequency (MHz) Fig.8. Energy per pulse consumption at various pulser components combinations for unloaded case

(left) and 50

Unfortunately, pulser can not be used for frequencies below 1 MHz: transformer output is

beneficial at high frequencies, but at low frequencies magnetizing inductance of the output

transformer is tampering the output, core saturates and efficiency and output amplitude is

dropping.

Preamplifier Performance Measurement

Preampifier parameters define the reception performance of the system. AC response of the

preamplifier was investigated using SWC technique: input was fed by harmonic signal from

direct digital synthesizer, driven by the same clock source as the acquisition channel. Signal at

the input was registered by channel 1 (Fig.4) and the output signal was fed to channel 2.

Obtained frequency response is presented in Fig. 9.

Fig.9. Preamplifier AC response for the gain (left) and phase (right).

Fig.9 shows that at 45 dB gain passband frequencies are from 318 kHz to 33.2 MHz. With gain

AC response available, input noise measurement can be established. Usual approach for

amplifier noise estimation could have been using spectrum analyzer. But since preamplifier

output sampling by ADC was readily available, more simple noise estimation procedure was

100k 1M 10M 40M

38

39

40

41

42

43

44

45

33.2MHz

318kHz

Ga

in, d

B

Frequency, Hz

GdB

F1

100k 1M 10M 40M

-90

0

90

ph

ase

, d

eg

Frequency, Hz

phaseGmax

phaseGavg

phaseGmin

5th

International Conference on NDT of HSNT- IC MINDT 2013

Athens Greece, Eugenides Foundation, May 20-22, 2013

used: amplifier output noise was recorder using sampling ADC. Record length of 32 k samples

was chosen (equivalent to 3 kHz resolution bandwidth). FFT has been used for noise power

spectral density estimation. In addition multiple noise power density measurements were power-

wise averaged Cmax times to obtain the statistical noise estimate:

NCf

es

PSDs

C

c

N

n

nkN

i

n

max

1

21

0

2max

2

, (10)

where N is the record length, 32 k samples; sn is the recorded noise signal. Measured gain was

used for output noise density conversion into input noise voltage density (Fig. 10).

Fig.10. Preamplifier input voltage noise density AC response

Noise whiteness now can be estimated: it can be seen that 5 MHz to 7 MHz frequency range

frequency response for input noise is sufficiently flat and can be considered as for white noise.

Moreover, measurements confirm that input voltage noise is 2 nV/Hz which produces right

balance with 10 bit ADC conversion resolution [1].

Recovery time of the ultrasonic preamplifier is important in pulse-echo mode when operation in

close vicinity is required. If not essential for conventional pulse systems, APWP signal requires

special attention on preamplifier recovery time: APWP signals suppose to be much longer than a

single pulse. Recovery time was estimated using similar to EN 12668-1 procedure: preamplifier

input was fed with sinusoidal signal via high impedance circuitry and probing rectangular pulse

was fed into same preamplifier. Recorded signal’s envelope was taken and recovery time

estimated (Fig.11).

Fig.11. Preamplifier recovery time measurement, using output envelope

100k 1M 10M 40M

0.0

5.0n

10.0n

Volta

ge

no

ise

de

nsity, V

/sq

rt(H

z)

Frequency, Hz

Max gain

Nin150

Ninopen

NinShort

Mid gain

Nin150

Ninopen

NinShort

Min gain

Nin150

Ninopen

NinShort

-5.0µ 0.0 5.0µ 10.0µ 15.0µ0.0

0.1

0.2

0.3

0.4

0.5

0.6

0.7

0.8

0.9

1.0

1.1

8.8us

Norm

alis

ed

envelo

pe

, A

.U.

Time, s

5th

International Conference on NDT of HSNT- IC MINDT 2013

Athens Greece, Eugenides Foundation, May 20-22, 2013

Output pulse was deliberately delayed so envelope magnitude before the pulse can be noted.

According to the EN 12668-1, recovery time should be estimated at 0.5 level. Taking

measurements at this level, with pulse duration subtracted, recovery time is 2 If much

stricter requirement, 0.95 level is used, recovery time is 8.8 .

Conclusions

Design of dedicated system was needed in order to start the investigation of novel APWP class

of signals. Acquisition system was developed to satisfy the specifics of the APWP: ability to

generate constant amplitude APWP pulse trains, acceptable pulser performance, and pulser

recovery time. Investigation of system performance indicates that it can be used for 0.5MHz to

30MHz frequency range: both pulser and preamplifier have balanced operation range. Excitation

signal amplitude can reach 200V for unipolar signals and 400V p-p for bipolar signals.

Variation of pulser output impedance measured according to EN 12668-1 standard is within 7

to 21 -

Acknowledgements

This research was funded by a grant (No. MIP-058/2012) from the Research Council of

Lithuania.

References

[1] L.Svilainis, V.Dumbrava, S.Kitov, A.Chaziachmetovas. The influence of digital domain on time of

flight estimation performance. In AIP Conference Proceedings, ICU 2011, vol.1433, 2012, pp.479-482.

[2] T.Virolainen, J.Eskelinen, E.Haeggstrom. Frequency domain low time-bandwidth product chirp

synthesis for pulse compression side lobe reduction. Proc. IEEE Ultrasonics Symp., 2009, pp. 1526-

1528.

[3] L.Svilainis, S.Kitov, A.Rodríguez, L.Vergara, V.Dumbrava, A.Chaziachmetovas. Comparison of

Spread Spectrum and Pulse Signal Excitation for Split Spectrum Techniques Composite Imaging. Proc.

UCIP 2012, IOP Conference Series: Materials Science and Engineering (MSE), 2012, pp.1-4.

[4] Z.-J.Yao, Q.-H.Meng, et al., Non-crosstalk real-time ultrasonic range system with optimized chaotic

pulse position-width modulation excitation. Proc. IEEE Ultrasonics Symp., 2008, pp. 729-732.

[5] L. Fortuna, M. Frasca, and A. Rizzo. Chaotic pulse position modulation to improve the efficiency of

sonar sensors. IEEE Trans. Instrum. and Meas, vol.52, 2003, pp.1809–1814.

[6] M.Zapf, B.F.Derouiche, N.V.Ruiter. Evaluation of chirp and binary code based excitation pulses for

3D USCT. Proc. IEEE Ultrasonics Symp., 2009, pp. 1996-1999.

[7] V.O.Buckin, B. Driscoll, C.Smyth. Ultrasonic spectroscopy for material analysis. Recent advances.

Spectroscopy Europe, vol.15(1), 2003, pp. 20-25.

[8] A.N.Kalashnikov, V.Ivchenko, R.E.Challis, A.K.Holmes. Compensation for temperature variation in

ultrasonic chemical process monitoring. IEEE Ultrasonics Symp., 2005, pp.1151 - 1154.

[9] F.Honarvar, H.Sheikhzadeh, M.Moles, A. N. Sinclair. Improving the time-resolution and signal-to-

noise ratio of ultrasonic NDE signals. Ultrasonics, vol. 41, 2004, pp. 755–763.

[10] L.Svilainis, K.Lukoseviciute, V.Dumbrava, A.Chaziachmetovas, A.Aleksandrovas. Application of

Arbitrary Position and Width Pulse Trains Signals in Ultrasonic Imaging: Correlation Performance

Study. Electronics and Electrical Engineering. vol.19(5), 2013, in press.

[11] M. Pollakowski, H. Ermert, Chirp signal matching and signal power optimization in pulse-echo

mode ultrasonic nondestructive testing. IEEE Trans. Ultrason. Ferroelect. Freq. Contr., vol. 41, 1994, pp.

655 – 659.

5th

International Conference on NDT of HSNT- IC MINDT 2013

Athens Greece, Eugenides Foundation, May 20-22, 2013

[12] L. Svilainis, V. Dumbrava, A. Chaziachmetovas and A.Aleksandrovas. Pulser for Arbitrary Width

and Position Square Pulse Trains Generation. In proc. IEEE IUS 2012. in press.