Upload
nidiam2
View
219
Download
0
Embed Size (px)
Citation preview
8/2/2019 0709_coverstory
1/7
c o v e r s t o r y
10 July 2009 MLO www.mlo-online.com
Detecting monoclonal gammopathies, or plasma cell dis-orders, usually involves serum protein electrophoresis(SPEP) and immunoelectrophoresis (IFE) to test bothserum and urine. But the growing clinical acceptance o a
serum ree light chain assay has all but eliminated urine tests
in identiying such plasma cell disorders as multiple myeloma
(MM), smoldering myeloma, monoclonal gammopathy o
undetermined signifcance (MGUS) and primary systemic
amyloidosis (AL); and because the assay has proven to be
more sensitive than IFE or detecting ree or unbound im-
munoglobulin light chains when it is used in conjunction with
SPEP, up to 99% o myelomas can be detected.
The light chain connection
Each clonal plasma cell undergoes heavy and light chain rear-
rangements to produce an immunoglobulin molecule. And
it is this rearrangement that determines not only the antigen
binding site o the immunoglobulin, but also identifes each
plasma cell clone.
Five types o immunoglobulin heavy chains have been
identifed: gamma, alpha, mu, delta and epsilon. Light chainsare identifed as either kappa or lambda. When a heavy chain
combines with a light chain, they produce molecules o IgG,
IgA, IgM, IgD, or IgE. Because plasma cells produce a larger
quantity o light chains than heavy chains, the excess light
chains enter the bloodstream as ree light chains (FLC).
In instances where plasma cell clones prolierate too rapidly,
immunoglobulin concentrations increase. These molecules
are then called monoclonal immunoglobulins and are directly
related to malignant or potentially malignant disorders such as
MM and MGUS.
Homing in on FLC
The signifcance o ree light chains led U.K.-based The BindingSite Ltd. to begin work on a new assay, explains Graham Mead,
PhD, director o research and development. Starting in the
1970s, there have been many studies published looking at di-
sum fligh hainaa:Ding plama ll didBy Richard R. Rogoski
To earn CEUs, see current test on page 22 or at
www.mlo-online.comunderthe CE Tests tab.
LEARNING OBJECTIVES
Upon completion of this article, the reader will be able to:Name fve types o immunoglobulin heavy chains and1.
two types o immunoglobulin light chains.
State specifc advantage(s)/disadvantage(s) o serum2.
ree light chain assessment.
Identiy three platorms on which the serum ree light3.
chain assay currently can be run.
State three pathologies detectable by serum ree light4.
chain assessment.Identiy a possible additional use or serum ree light5.
chain assessment.
c o N t I N U I N G e D U c A t I o N
http://www.mlo-online.com/http://www.mlo-online.com/http://www.mlo-online.com/http://www.mlo-online.com/http://www.mlo-online.com/8/2/2019 0709_coverstory
2/7
c o v e r s t o r y
www.mlo-online.com
erent methods or measuring serum ree
light chains. For most o these experimen-
tal assays, a lack o suitable specifcity
was apparent (i.e., they cross-reacted with
intact immunoglobulin), and none were
adopted or routine clinical use.Work on developing our own assays
was started in 1996. Our sta already had
a number o years experience producing
highly specifc polyclonal antisera or
measuring IgG subclasses. We hoped to
build on this experience to develop se-
rum ree light chain assays and improve
upon the studies previously published. It
took several years o trial beore we wereable to produce antibodies adequate or
developing nephelometric assays, and
our frst study o their application was
not published until 2001.
So ar, the product, Freelite, is the only
Food and Drug Administration (FDA) ap-
proved FLC assay on the market, although
a similar product manuactured by an
Italian company is being used in a ewEuropean countries to test urine samples.
New guidelines in the United States, how-
ever, specifcally recommend the use o
FLC serum tests or diagnosis.
Since the serum FLC test involves
the action o antibodies, the underlying
technology is simple and is the same as
many other nephelometric/turbidimetric
immunoassays, Mead says. The anti-
bodies orm immune complexes with
the ree light chains in a test serum and
the size/speed o complex ormation is
detected by a laser shone through the
reaction vessel. To ampliy the signal,
the antibodies are bound to microscopic
polystyrene particles (requently calledlatex). The main challenge o develop-
ing and producing the assays is the
production o suitable antibodies which
must have:
a high degree o specifcity, so they
recognize immunoglobulin light
chains when they are ree but not when
they are bound to heavy chains in in-
tact immunoglobulin molecules; and
a balanced response against the va-
riety o dierent monoclonal FLCs
produced by patients.
With regard to specifcity, simplyimmunizing with ree light chains and
absorbing with intact immunoglobulin
does not produce antisera with adequate
avidity or titer. We use proprietary tech-
niques to ocus the antibody production
on the parts o the light chain molecule
which are hidden in intact immunoglobu-
lin but exposed on ree light chains.
Producing antisera with a balanced
response against the ree light chains rom
all patients is a ormidable challenge and
one o the reasons that monoclonal anti-bodies are not suitable or these assays.
Careul control o the range o proteins
used or immunizations and antibody
Electrophoresis Simplified
Open door. Insert sample. Walk away.The MINICAP is so simple to use,its like having a new employee.
Electrophoresis of serum and urine is recommendedto screen for and monitor monoclonal gammopathies.
(800) 835-6497 www.sebia-usa.com
... serum ree light chain
elevations were associated
with an increased risk o
progression to lymphoma
among HIV-positive patients.
http://www.mlo-online.com/http://www.mlo-online.com/8/2/2019 0709_coverstory
3/7www.mlo-online.com
purifcations as well as the use o antisera
pools o >100 liters, has allowed us to
optimize the assay response.
As with any lab test, there are advan-
tages and disadvantages. Mead stresses
that the major advantage o running a FLCtest is that it measures concentrations in
serum rather than in urine. One o the
important unctions o the kidneys is to re-
absorb and catabolize small proteins, such
as ree light chains, which have been fl-
tered rom the blood in the glomeruli. It is
only when this capacity or re-absorption
is overwhelmed that signifcant quantities
o ree light chains can pass through the
kidney tubules and into the urine. There-
ore, many patients with small plasma cell
tumors are ound to have abnormal serumree light chain results while their urine
appears normal.
The sensitivity o the assay, however,is largely disease specifc (i.e., or only
those diseases with higher likelihoods
o having ree light chains will the as-
say be more sensitive than IFE alone).
For example, Katzmann, et al, evaluated
1877 patients with monoclonal gammo-
pathy using fve assays: serum and urine
protein electrophoresis (PEL), serum and
urine IFE, and the serum FLC assay. For
all comers, the sensitivity o the serum
IFE and the FLC were 87% and 74%,
respectively. I one breaks down the
sensitivity analysis by disease, however,the respective sensitivities are as ollows:
multiple myeloma 94% vs. 97%; macro-
globulinemia 100% vs. 73%; smoldering
myeloma 98% vs. 81%,; MGUS 93% vs.
42%; plasmacytoma 72% vs. 55%; AL
amyloidosis 74% vs. 88%; and light chain
deposition disease 56% vs. 78%.1
The International Myeloma Working
Groups published guidelines recom-
mend 24-hour urine samples or some
patients. First published online in Leu-
kemia in November 2008, the article byDispenzieri, et al, states that or the
purpose o screening or monoclonal
proteins or all diagnoses except AL, the
FLC can replace the 24-hour urine IFE.
Once a diagnosis o monoclonal gam-
mopathy is made, however, the 24-hour
protein IFE should be perormed. For AL
screening, however, the urine IFE should
still be done in addition to the serumtests, including the serum FLC.2
Howard Robin, MD, medical director
o laboratory services at Sharp Memo-
rial Hospital in San Diego, CA, says he
has been using the FLC assay or more
than two years or screening, diagnoses,
and prognostications. Yet, he says urine
collections are a problem. One o the
problems with urine is that we seldomget a true 24-hour urine. Mostly, it is
random or spot urine samples. And labs
do not like working with urine because
With regard to specifcity,
simply immunizing with ree
light chains and absorbing with
intact immunoglobulin does not
produce antisera with adequate
avidity or titer.
F L c A s s A y s
http://www.mlo-online.com/mailto:[email protected]://www.a2la.org/http://www.mlo-online.com/8/2/2019 0709_coverstory
4/7
c o v e r s t o r y
www.mlo-online.com
they are not getting 24-hour urine.
The rationale or emphasizing the
need or the 24-hour urine protein elec-
trophorsis in ollowing patients with light
chain myeloma is that there is a poor cor-
relation between the serum FLC and theurinary monoclonal protein as measured
by urine PEL3and or patients with amy-loidosis, serial 24-hour urine measure-
ments are critical or monitoring the status
o a patients nephrotic syndrome.
As or any disadvantages in using an
FLC assay, Mead points to two. The
material cost o running serum Freelite
assays is greater than that o a simple
urine electrophoresis gel. When costing
analysis has included storage, process-
ing, urine immunofxation, time or in-terpretation, or the Medicare reimburse-
ment costs, however, there are benefts
to using the serum assays. Freelite has
been shown to be more cost eective
than urine testing. This is evidenced by
the act that the Medicare reimburse-
ment costs or the urine panel o tests
is higher than or the alternative serum
panel which includes serum Freelite.
This cost analysis was substantiated by
separate studies published in 2006 by
Katzmann et al4, and Hill, et al.5
Another issue that is sometimes
raised is the accuracy o FLC assayscompared to other tests. Some aspects
o analytical perormance have been
criticized; and it is true that the precision
and accuracy does not equal that o a C3
assay, or example, says Mead. This is
understandable when you consider that
ree light chains are monoclonal proteins
that can vary by more than a thousand-
old in concentration and may exist in
dierent polymeric orms.
David Keren, MD, medical director
o Warde Medical Laboratory, a private
reerence lab in Ann Arbor, MI, says he
uses FLC assays every day with excellentresults. But he agrees there can sometimes
be a computation problem. In a ew
cases, we have gotten a alsely low value
because o higher levels o antigen. It is
uncommon, but it is an issue.
A simple test to run
For the laboratory proessional, running
a FLC assay is simple. As long as the
nephelometer/turbidimeter is correctly
programmed and appropriately main-
tained, running Freelite assays is aseasy as running other serum tests, says
Mead. No special training is required to
run the tests but a basic understanding o
the biology o ree light chains is helpul
when interpreting results.
Dr. Keren concurs: It is an automated
test that can be run on the same machines
used to measure IgG, IgA, and IgM.
Currently, these machines include
Dade Behring BNII and ProSpec; Beck-
man IMMAGE; Roche/Hitachi 911, 912,
917, and Modular P; Olympus AV400,
640, 2700, 5400; Radim Delta; andBayer Advia. And since it is the kappa/
lambda ratio which is read, Dr. Robin
notes that it is the systems sotware that
fgures out the ratio. The entire assay
takes between fve and 18 minutes to run,
depending on the analyzer used.
Katzmann, et al, determined in 2002
the normal range using both resh and
rozen sera rom 282 individuals aged 21
to 90. By including 100% o donors, the
normal diagnostic range or FLC kappa/
lambda was set at 0.26 mg/L to 1.65 mg/L.
Normal kappa FLC levels are 3.3 mg/Lto 19.4 mg/L, and normal lambda FLC
levels are 5.7 mg/L to 26.3 mg/L. Patients
with kappa/lambda ratios greater than
1.65 mg/L have higher levels o kappa
FLC. Those ratios less than 0.26 mg/L
have higher levels o lambda FLC.6
The normal range or kappa/lambda
ratios also is greater than those used or
most other tests in order to provide a larg-
er saety margin or normal patients.
Validating studies
When I irst saw the data, I was
skeptical, admits Dr. Keren, adding
When costing analysis has
included storage, processing,
urine immunofxation, time or
interpretation, or the Medicare
reimbursement costs, however,
there are benefts to using the
serum assays.
http://www.mlo-online.com/http://www.mlo-online.com/8/2/2019 0709_coverstory
5/7www.mlo-online.com MLO July 2009
Bio-Rad Laboratories C L I N I C A L D I A G N O S T I C S
MONOLISA
,all that and more
With outstanding performance and standardized procedures,the MONOLISA Hepatitis product line is only one of the
great buys you get shopping with Bio-Rad.
MONOLISA
HEPATITIS ASSAYS
AND MORE
that he wanted to see more empirical
data than just those in the original
Katzmann study. To date, there have
been more than 200 published studies
using Freelite as an FLC assay. As a
result, this assay is now recommendedor the initial evaluation o suspected
myeloma; or prognosis o plasma cell
dyscrasias; and or monitoring oligose-
cretory myeloma and AL amyloidosis.
Although the 2002 Katzmann study
remains the landmark, another study
he published in 2006 concluded that
urine analysis was not necessary i only
serum was analyzed using an FLC assay
combined with PE and IFE. Even as ar
back as 2001, research by Bradwell,
et al, published in Clinical Chemistry,concluded that the automated immu-
noassay then being studied could be
used to assay FLC concentrations in a
routine clinical laboratory setting.7
Other studies also have supported the
use o Freelite or screening, diagnosing,
and monitoring patients.
In a 2007 article published in Clini-
cal Lymphoma & Myeloma, Sundar
Jagannath looked at ree light chain
measurements at short sampling inter-
vals given that the hal-lie o FLC is
less than six hours. He said such testing
could allow real-time measurement o
treatment-induced tumor kill and could
possibly provide prompt indications ochemosensitivity, dose adequacy and the
need or alternative approaches.8
Hutchison, et al, in a 2008 study
published in BMC Nephrology, looked
at myeloma patients with renal ailure.
The researchers concluded: The diag-
nostic accuracy o these assays and their
rapid laboratory turnaround time should
aid nephrologists in their assessment o
acute renal ailure.9
Interestingly, Dr. Robin recommends
that patients over 50 with renal ailureshould also be screened or monoclonal
gammopathies.
... this assay is now recommended
or the initial evaluation o
suspected myeloma; or prognosis
o plasma cell dyscrasias;
and or monitoring oligosecretory
myeloma and AL amyloidosis.
F L c A s s A y s
http://www.mlo-online.com/http://www.mlo-online.com/8/2/2019 0709_coverstory
6/7July 2009 MLO www.mlo-online.com
c o v e r s t o r y
Many labs are moving to Lean processing and single piece flow. Now you dont need to wait for samples tospin in a slow conventional centrifuge. The StatSpin Express family of small high speed centrifuges is youranswer to rapid sample processing.
The high speed horizontal StatSpin Express 4 provides flat gel separation in only 3 minutes. Samples can beput directly on the analyzer. Both the StatSpin Express 3 and StatSpin Express 4 hold 8 tubes up to 10 mLThey have a rugged brushless motor for added reliability and the entire unit is backed by a 2-year warranty.Plus they are incredibly easy to use. Just push the button and walk away.
Anywhere a STAT sample isneeded, the StatSpin Express 2provides Platelet Poor Plasma injust 2 minutes. The small foot-print lets you put it anywhere inthe lab.
Tired of waiting for samplesto spin? Place any Express
centrifuge next to youranalyzer and get results fast!
To get up and runningcontact Iris Sample Processing or your Iris distributor .
Tired of waiting for samples to spin?In just 3 minutes a StatSpin
Express can get you up and running!
A Division of IRIS International, Inc.
New developments
While it is a certainty that other indepen-
dent studies will be done using Freelite,
Mead says his company will continue to
ocus on ree light chains as well as heavy
chains. The current ocus o research islooking at dierent applications or serum
ree light chain analysis. For example,
a presentation at a recent conerence
reported that serum ree light chain eleva-
tions were associated with an increased
risk o progression to lymphoma among
HIV-positive patients. Here in the U.K.,
we are currently providing support or a
trial o extended dialysis to remove ree
light chains and improve the prognosis or
myeloma patients with acute renal ailure
(caused by light chain cast nephropathy).We are also investigating whether ree
light chains contribute to renal damage in
patients without myeloma or other plasma
cell tumors, he says.
From the early studies with our prod-
uct, it became apparent that abnormalities
o the ree light chain ratio (kappa/lambda)
could provide a more sensitive indication
o monoclonal disease than simple eleva-
tions o one light chain. This is because
a tumor will usually express a surplus
o one light chain but also suppress theproduction o the alternate light chain. As
a result o this observation, we are now
developing assays that will determine these
ratios or intact immunoglobulins (e.g.,
the IgGkappa/IgGlambda ratio). Our frst
ull report o these assays has now been
accepted or publication. The preliminary
results indicate that these assays will
be useul or diagnosis, prognosis and
monitoring o plasma cell tumors. They
will be complementary to our serum ree
light chain assays because they provide asensitive marker or those tumors which
produce little ree light chain.
Richard R. Rogoski is a reelance journalist based inDurham, NC. His extensive list o published articleshave dealt with new developments in the felds ocardiology and cardiac surgery; imaging technol-ogy; inormation technology; and the business side ohealthcare. Contact him at [email protected].
Reerences:
Katzmann JA, Kyle RA, Benson J, Larson DR,1.Snyder MR, et al. Screening Panels or Detectiono Monoclonal Gammopathies Clin Chem. June
2009; doi:10.1373/clinchem.126664.
Dispenzieri A, Kyle R, Merlini G, Miguel JS, Ludwig H,2.et al. International Myeloma Working Group guide-lines or serum ree light chain analysis in multiplemyeloma and related disorders. Leukemia. 2009.
Dispenzieri A, Zhang L, Katzmann JA, et al. Ap-3.praisal o immunoglobulin ree light chain as amarker o response. Blood. 2008;111:4908-4915.
Katzmann JA, Dispenzieri A, Kyle RA, Snyder MR,4.Plevak MF, Larson DR, Abraham RS, Lust JA, MeltonIII LJ, Rajkumar SV. Elimination o the Need or UrineStudies in the Screening Algorithm or MonoclonalGammopathies by Using Serum Immunofxation andFree Light Chain Assays. Mayo Clin Proc. 2006.
Hill PG, Forsyth JM, Rai B, Mayne S. Serum Free5.Light Chains: An Alternative Test to Urine Bence
Jones Proteins When Screening or MonoclonalGammopathies. Clin Chem. 2006.
Katzmann JA. Clark RJ, Abraham RS, Bryant S,6.Lymp JF, Bradwell AR et al. Serum ReerenceIntervals and Diagnostic Ranges or Free Kappaand Free Lambda Immunoglobulin Light Chains:Relative Sensitivity or Detection o MonoclonalLight Chains. Clin Chem. 2002.
Bradwell AR, Carr-Smith HD, Mead GP, Tang LX,7.Showell PJ, Drayson MT, Drew RL. Highly sensitive,automated immunoassay or immunoglobulin reelight chains in serum and urine. Clin Chem. 2001.
Jagannath S. Value o Serum Free Light Chain Test-8.ing or the Diagnosis and Monitoring o MonoclonalGammopathies in Hematology. Clinical Lymohoma& Myeloma; 2007;7(8).
Hutchison CA, Plant T, Drayson M, Cockwell P,9. Kountouri M, Basnayke K, Harding S, Bradwell AR,Mead G. Serum ree light chain measurement aids
the diagnosis o myeloma in patients with severerenal ailure. BMC Nephrology. 2008.
http://www.mlo-online.com/mailto:[email protected]://www.mlo-online.com/mailto:[email protected]8/2/2019 0709_coverstory
7/7www.mlo-online.com MLO July 2009 43